Finite Element Analysis and Static Loading in Dental Implant with
Innovative Crown Design (IN VITRO STUDY)
Dr. Twana Omar Bakr 1, Dr. Fahd S. Ikram 2 and Dr. Nabeel S. Martani 3
1
B.D.S. M.Sc. Prosthodontic, College of Dentistry, Hawler Medical University.
2, 3
B.D.S. M.Sc. PhD Prosthodontic, College of Dentistry, Hawler Medical University.
Abstract
Objective: The purpose of this study is to examine how different crown materials affect stress
transformation and distribution around both dental implants. To do this, integrated and combined
crowns were built, modelled, and tested under static axial loads using various material combinations.
The biomechanical response was then examined. Methods: A validated three-dimensional finite
element (FE) models of crown supported by implant was developed by to evaluate the effect of the
different type of material (E max, zirconia, composite) on the short implant. After the FE models had
been validated, simulations utilizing various configurations of various crowns fixed to two distinct
types of implants were run and subjected to static loading to ascertain the distribution of stresses
inside the bone around the implants. Result: The comparative results showed that manufacturing the
crown using softer material (i.e., materials with lower elastic modulus) reduced the stress distribution
in crown, implant and cancellous bone. It may refer to this phenomenon that softer material can
absorb more energy from the applied compressive load, and result in transferring less energy to the
implant and jaw bone. However, this effect was not significant on cortical bone compared to the
cancellous bone. Combination of different materials for design and manufacturing the crown can
alter the biomechanical response and could be beneficial for decreasing the stress distribution in
implant and spongy region of jaw bone when stiffer material is needed to be covered in upper surface
of the crown. In addition, the results suggests that shorter implant can increase the stress distribution
in both cortical and cancellous bone. Conclusion: by using stiff material the stress will increase on
the parts of implant and the surrounding bone which may led to failure of implant or bone resorption
around of the implant, in other way by using less stiffer material the possibility of success will be
increased and also the success rate of the implant is increased, also before deciding which implant
size and length are used you select which type of the prosthetic will be used.
Keywords: Dental Biomechanics, Dental Implant, Innovative Crown, Zirconia, E-Max, Composite,
Implant Length, Finite Element Modeling
INTRODUCTION
In order to regain complete, pleasant masticatory function and face esthetic, humans have tried
using a variety of methods to restore lost teeth throughout history. Before to the osseointegration
period, there were several designs of dental implants and frameworks that were used to support
dentures and partial dentures with varying degrees of effectiveness. Among the various materials used
in implants are porcelain, cobalt chromium, and radioplatinum, but the discovery of titanium changed
the course of implant history. There are several techniques for comprehending the force distribution,
including analytical mathematical models that include strain gauges, the photo elastic model, and
studies like the finite element analysis (vandana and kartik, 2004).
FEA has been extensively used in implant dentistry to foresee the biomechanical behavior of
various dental implant designs as well as the influence of clinical factors on clinical success. There
has been a great deal of study done on the stress patterns in implant components and the surrounding
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
15 www.jnr.tw
bone. The accuracy of the used simulating structures influences the outcome of every FE
investigation. These are the material properties of the implant and bone, the surface features and
shapes of the implant and its parts, the force applied, and the biomechanical behavior of the implant-
bone relationship. (Trivedi S, 2014)
In FEA, a specific physical system's behavior is mathematically modelled. The FEA separated
the structure of the model into several distinct parts, each of which retains the structure's original
features. Each component requires a unique equation that must be solved using mathematical models
chosen in accordance with the facts being examined. (Boccaccio et al, 2011, Dejak and Mlotkowski,
2011).
By altering the parameters of those geometries, FEA makes it possible to apply a force or a
system of forces to any point and/or in any direction, providing knowledge of the movement, amount
of tension, and compression force on each area. It also converts natural or artificial tissue into complex
structures that accurately represent the original one. (Vasudeva G, 2009)
In addition to providing a forecast of a desirable outcome with the lowest chance of failure,
FEA assists in the study of stress direction and distribution between natural teeth and the kind of
material used to reconstruct the original tooth or dental structure if they are lost. Zarone et al, 2005
researchers looked at how stress surrounding teeth from everyday load on the maxillary central incisor
was affected by finishing line preparation with an alumina porcelain veneer in 2005. They advised
utilizing a chamfer with palatal overlap design when repairing with porcelain veneers because it
achieves the advised stress distribution more effectively than the window approach. (Srirekha and
Bashetty, 2010)
MATERIAL METHODS
1. Model preparation
For the model preparation in the first ready-made dental arch is used which have all the teeth
and in full occlusion, the teeth are attached to the base by the screw, first lower molar is choice and
replaced by the implant.
After the teeth is removed the socket of the teeth is grinded to become wider for the fixation,
for the parallelism of the implant dental surveyor is used (dental farm S.r.I. Via Susa, 9/A-10138
torino-italy). Table of dental surveyor is balanced until become parallel with floor by the bubble level,
then the dental arch is locked on the table by model clamp lock nut of the device laterally like each
corner of the triangular. Implant fixed to the tool adapter holder and perpendicular to the table. The
head of the surveyor pushed implant to the socket and placed in the center without touching any
border and the level of the implant neck below the level of the gum about 2-3 mm then fixed the tool
holder in the place, Cold cure acrylic (vertex type, Vertex Dental ByV, 872.3760 EBI, ISO020795)
is choice for fixation the implant which is mixed according to the manufacture instruction and used
before reach to dough stage, in flow stage. Cold cure acrylic is putted to the socket and waited until
the material become set then the tool holder is opened and the cast removed for the table and watched
for any problem present. Then the surface of the acrylic around the implant is grinded with low-speed
engine to provide smooth surface of the acrylic and free of projection and irregularities. After that the
length of the implant adjusted and shortened until provide good space for occlusal surface of the
crown.
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
16 www.jnr.tw
2. Scanning the model
After the model is prepared TRIOS, scanner is used (3Shape TRIOS® 3, Copenhagen,
Denmark) to scan the model, scanning is done for all the lower arch and opposite arch in respect to
occlusion and making STL file, then the design of the samples is made which consist of full crown
and two-part crown of Emax, zircon and composite. Figure 1
Figure 1: TRIOS3 intraoral scanner
3. Design of samples
Crown are made in two forms first form is made in full anatomical crown without any changing
and second form made in to two parts which is first part 1/3 occlusal part and second part 2/3
gingivally which are connect to each other by resin in between the two parts there is 4 projection in
the first part and 4 holes in the second part which orient the two parts to each other and prevent
movement of two parts, the projection are orient in square shape each corner has one projection,
Length of the projection is about 2mm.
4. Development of the geometrical models
The STL input data which was received based on scanned parts (Figure 2) were evaluated. To
develop fine geometrical models which can be used in finite element (FE) analyses, the STL files
were imported to Solid works (Dassault Système©
, Vélizy-Villacoublay, France) and the surfaces
were trimmed and smoothed to achieve acceptable geometries for exporting to FE package. To
simplify the simulation, a sliced part of the jaw bone was mimicked in simulations.
Figure 2: Samples of the scanned parts in STL format
Based on objective of this study, four general geometrical models were developed based on
combination of (1) integrated and (2) combined crown implanted using short implants. The diameters
for short implants were 4.5mm and their lengths were 6mm, respectively. Figure 3 show the details
of the aforementioned models.
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
17 www.jnr.tw
5. Development of the FE models
The geometrical models were imported to ABAQUS package (SIMULIA, Providence, RI,
USA) to simulate the effect of material properties of the crown and implant size on biomechanical
response. For this purpose, 18 FE models were developed as following scenario:
Model 1: E-Max
Model 2: Composite Resin
Model 3: Zirconia
Model 4: E-Max / Composite Resin
Model 5: E-Max / Zirconia
Model 6: Composite Resin / E-Max
Model 7: Composite Resin / Zirconia
Model 8: Zirconia / E-max
Model 9: Zirconia / Composite Resin
Figure 3: The arrangement of different FE models
Mechanical properties of the model were considered as isotropic elastic theory, which was
extracted from the available data in literature (Table 1)
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
18 www.jnr.tw
Table 1: Mechanical properties of different components in FE models
The analyses were classified as static and the solution examined the governing equation system
using the element diagonal mass matrix and the law of explicit integration. The applied boundary
condition between different components were considered using the discretization method of surface
to surface with tie contact property to simulate perfect connection to constrain equal degrees of
freedom (i.e., equal translational and rotational motions). The static compressive load equal to 200 N
[7] was vertically applied on the upper surface of the crown for comparative analysis using the
reference point positioning technique as shown in Figure 4.
Figure 4: The applied axial compressive concentrated force to FE model
10-node quadratic tetrahedron elements were used to generate the meshed models (Figure 5 to
6) and sensitivity analyses to assess the independency of the results from meshing were performed.
The finalized number of elements after mesh sensitivity analyses were considered as Table 2 and 3.
Table 2: Number of elements for finalized FE model after convergence analyses for integrated
crown models with short implants
Components Number of Elements
Short Implant 13810
Crown 25955
Cancellous Bone 38381
Cortical Bone 33134
Components Elastic modulus (GPa) Poisson’s ratio
Implant 110 [1] 0.3 [1]
Abutment 110 [1] 0.3 [1]
Cancellous Bone 1.37 [2, 3] 0.3 [2, 3]
Cortical Bone 12.6 [2, 3] 0.3 [2, 3]
E-Max 95 [3, 4] 0.2 [3, 4]
Zirconia 210 [5] 0.33 [5]
Composite resin 21 [6] 0.24 [6]
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
19 www.jnr.tw
Table 3: Number of elements for finalized FE model after convergence analyses for combined
crown models with short implants
Components Number of Elements
Short Implant 13810
Crown (Upper Part) 28398
Crown (Lower Part) 25063
Cancellous Bone 38381
Cortical Bone 33134
Figure 5: Meshing the FE models based on sensitivity analyses
(Integrated crown with short implant)
Figure 6: Meshing the FE models based on sensitivity analyses
(Combined crown with short implant)
The simulations were performed as described previously by comparing the results for 18
different models (i.e., Based on combination of (1) integrated and (2) combined crown implanted
using short implants.) subjected to vertical static force equal to 200N. The values of Von-Mises stress
were calculated in critical nodes in different components of the models (i.e., cortical bone, cancellous
bone, implants, and crown) for comparative study.
RESULT
The results of sensitivity analyses verified the accuracy of the FE models for investigation of
the main objective of this study. The calculated values of Von-Mises stress in different FE models
subjected to static loading (i.e., vertical compressive force equal to 200N) over the upper surface of
the crown were extracted. The maximum equivalent stress in critical regions achieved results were
calculated in different components of different models (i.e., cortical bone, cancellous bone, implants,
abutments, and crown) and the calculated results were compared.
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
20 www.jnr.tw
The comparative results are summarized in figures 9 to 16 in which the values of Von-Mises
stress were shown for different implant configurations with different materials for the crowns. The
result from the static loading appears that the stress formation inside the crown for the short implant
with zircon crown has higher value while followed by the E-max and the last one composite crown.
The details are show in figure (7). The most highly statistically significant one is composite VS
zirconia and the least one which is non statically significant is E-Max VS. Composite and other are
between them as shown in table (3):
Figure 7: Comparative results for values of maximum Von-Mises stress in crown for the
models with integrated crown using short implant.
The result from the static loading appears that the stress formation inside crown while using
zirconia as the superior part of the combined crown in both Z/E, Z/C crown while followed by the
E/Z and E/C and the last one C/E and C/Z. The details are show in figure (8). The most statistically
significant one is Composite/Zirconia vs. Zirconia/Composite and last non statically significant one
is Composite/E-max vs. Composite/Zirconia and other are between them as shown in table (4):
Figure 8: Comparative results for values of maximum Von-Mises stress in crown for the
models with combined crown using short implant.
0
10
20
30
40
50
60
E-Max Composite
Resin
Zirconia
Maximum
Equivalent
Stresses
(MPa)
E-Max Composite Resin Zirconia
0
10
20
30
40
50
60
Maximum
Equivalent
Stresses
(MPa)
E-Max/Composite E-max/Zirconia
Composite/E-max Composite/Zirconia
Zirconia/E-max Zirconia/Composite
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
21 www.jnr.tw
Table 4: statistical analysis of crown
Pair (Short Implant) Crown P- value Statistical analysis
E-Max vs. Composite Resin 0.07329
E-Max vs. Zirconia 0.0447 *
Composite Resin vs. Zirconia 0.0001785 *
E-Max/Composite vs. E-max/Zirconia 0.9998
E-Max/Composite vs. Composite/E-max 0.001219 *
E-Max/Composite vs. Composite/Zirconia 0.00111 *
E-Max/Composite vs. Zirconia/E-max 0.04841 *
E-Max/Composite vs. Zirconia/Composite 0.01694 *
E-max/Zirconia vs. Composite/E-max 0.0005098 *
E-max/Zirconia vs. Composite/Zirconia 0.0004629 *
E-max/Zirconia vs. Zirconia/E-max 0.09124
E-max/Zirconia vs. Zirconia/Composite 0.03459 *
Composite/E-max vs. Composite/Zirconia 1
Composite/E-max vs. Zirconia/E-max 2.91E-08 *
Composite/E-max vs. Zirconia/Composite 6.65E-09 *
Composite/Zirconia vs. Zirconia/E-max 2.61E-08 *
Composite/Zirconia vs. Zirconia/Composite 5.98E-09 *
Zirconia/E-max vs. Zirconia/Composite 0.9987
Pair (Short Implant) Crown Mean SD
Zirconia 40.496 9.980093743
E-Max 31.853 7.423834065
Composite Resin 23.706 5.821193463
E-Max/Composite 35.348 7.160511776
E-max/Zirconia 36.143 6.533477975
Composite/E-max 22.6344 4.3443233
Composite/Zirconia 22.5478 4.069622363
Zirconia/E-max 44.265 8.556589728
Zirconia/Composite 45.455 8.26326711
The result forms the static loading appear that the stress formation in implant itself for the short
implant more for the full anatomical zirconia crown while followed by the E-max and the least stress
formation is composite crown. The details are show in figure (9). The most statistically significant
one is Composite Resin vs. Zirconia and the least one which is non statically significant is E-Max vs.
Zirconia and other are between them as shown in table (5):
Figure 9: Comparative results for values of maximum Von-Mises stress in implants for the
models with integrated crown using short implant.
0
5
10
15
20
25
30
35
40
E-Max Composite Resin Zirconia
Maximum
Equivalent
Stresses
(MPa)
E-Max Composite Resin Zirconia
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
22 www.jnr.tw
The result from the static loading appears that the stress formation in implant itself for the short
implant the result show more stress formation while using zirconia crown as superior part in both
Z/E, Z/C while followed by the E/Z and E/C then C/E and C/Z. The details are show in figure (12).
The most highly statistically significant one is Composite/E-max vs. Zirconia/E-max and the least
one which is non statically significant is E-Max/Composite vs. E-max/Zirconia and other are between
them as shown in table (10):
Figure 10: Comparative results for values of maximum Von-Mises stress in implants for the
models with combined crown using short implant.
Table 5: statistical analysis of implant
Pair (Short Implant) implant P- value Statistical analysis
E-Max vs. Composite Resin 0.1475
E-Max vs. Zirconia 0.4093
Composite Resin vs. Zirconia 0.008718 *
E-Max/Composite vs. E-max/Zirconia 1
E-Max/Composite vs. Composite/E-max 0.4445
E-Max/Composite vs. Composite/Zirconia 0.8594
E-Max/Composite vs. Zirconia/E-max 0.03974 *
E-Max/Composite vs. Zirconia/Composite 0.2354
E-max/Zirconia vs. Composite/E-max 0.4254
E-max/Zirconia vs. Composite/Zirconia 0.8451
E-max/Zirconia vs. Zirconia/E-max 0.04306 *
E-max/Zirconia vs. Zirconia/Composite 0.2493
Composite/E-max vs. Composite/Zirconia 0.981
Composite/E-max vs. Zirconia/E-max 0.0001312 *
Composite/E-max vs. Zirconia/Composite 0.001906 *
Composite/Zirconia vs. Zirconia/E-max 0.001382 *
Composite/Zirconia vs. Zirconia/Composite 0.01571 *
Zirconia/E-max vs. Zirconia/Composite 0.966
0
10
20
30
40
Maximum
Equivalent
Stresses
(MPa)
E-Max/Composite E-max/Zirconia Composite/E-max
Composite/Zirconia Zirconia/E-max Zirconia/Composite
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
23 www.jnr.tw
Pair (Short Implant) Implant Mean SD
Zirconia 25.231 6.94377331
E-Max 21.924 5.506868842
Composite Resin 16.981 4.363598031
E-Max/Composite 20.174 4.414886937
E-max/Zirconia 20.233 4.309258895
Composite/E-max 16.709 3.525354041
Composite/Zirconia 18.024 3.074111254
Zirconia/E-max 25.883 4.574829809
Zirconia/Composite 24.38 4.956457068
DISCUSSION
The result of the study show that the stress formation on parts of crown of the two part crown
in short implant consist of zirconia base 2/3 gingival crown and 1/3 composite or e max occlusal part
are more than the other types of the crown formed by other hybrid crowns this because of the high
modulus of elasticity, stress induced toughening mechanism and high fracture resistance this lead to
more stress formation inside zirconia part of crown, this finding is similar with Güngör et al, 2018
where compared the fracture strength of all-ceramic crowns fabricated from zirconia and different
other materials and found that the highest fracture strength and stress formation inside the crown
mean value was recorded by crowns fabricated from zirconia.
When the base 2/3 of the crown made form the composite for short implant the stress formation
on the implant is less and more in zirconia base because material has law modulus of elasticity (less
stiffer) so lead to more absorption of the force coming from the mastication and less stress formation,
Inconsistent with our results, Duan and Griggs (Duan and Griggs, 2015) compared the stress
distribution in lithium disilicate ceramic and resin nanoceramic CAD-CAM crowns and reported that
resin nanoceramic crowns showed lower stress values under vertical loading on the implant. The
polymer-infiltrated hybrid ceramic crown exhibited low stress values in the cervical region. The resin
content (Gracis et al, 2015), (Duarte et al, 2016) of these materials is thought to provide a favorable
stress distribution in restorative crowns. Opposite to our result (Skalak, 1983) found that resin-based
restorative materials may reduce the stress on implants and peripheral bone.
For the all position of stress formation on (crown, implant and bone), full anatomical zirconia
crown form the greatest stress formation among all the types of the prosthetic part because material
is hard and have no property of elasticity so they did not absorb any stress, when the stress formed
either fracture or failure of the material occur or transmitted to the other part of the structure, then
followed by E max material and the last one is resin composite which absorb most of the stress
because of the spongy properties related to filler particle of material which provide more space inside
structure so only transmitted little of the stress to the other part of implant structure.
This founding is similar to study done by De kok P et al 2012 concluded that monolithic implant
supported crowns have a higher initial load to failure than conventional veneered ceramic crowns and
are expected to show clinically fewer fractures. Monolithic ceramic restorations perform better than
composite resin crowns in force standing but composite is better in stress absorption. Composite
resins with a low modulus of elasticity show promising results in withstanding the force absorption,
in other way opposite to Jassim and Majeed (2018) which concluded that the highest fracture strength
mean value was recorded by monolithic crowns fabricated from zirconia (CEREC Zirconia) followed
by crowns fabricated from reinforced composite (BRILLIANT Crios), zirconia-reinforced lithium
silicate (CELTRA DUO), and lithium disilicate (IPS e.max CAD), while the lowest fracture strength
mean value was recorded by crowns fabricated from hybrid dental ceramic (VITA ENAMIC).
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
24 www.jnr.tw
CONCLUSION
In summary, the comparative results showed that manufacturing the crown using softer material
(i.e., materials with lower elastic modulus) reduced the stress distribution in crown, implant and
cancellous bone. It may refer to this phenomenon that softer material can absorb more energy from
the applied compressive load, and result in transferring less energy to the implant and jaw bone.
However, this effect was not significant on cortical bone compared to the cancellous bone.
Combination of different materials for design and manufacturing the crown can alter the
biomechanical response and could be beneficial for decreasing the stress distribution in implant and
spongy region of jaw bone when stiffer material is needed to be covered in upper surface of the crown.
In addition, the results suggests that shorter implant can increase the stress distribution in both cortical
and cancellous bone.
References
1) Ueda N, Takayama Y, Yokoyama A. Minimization of dental implant diameter and length
according to bone quality determined by finite element analysis and optimized calculation.
Journal of prosthodontic research. 2017; 61(3):324-32.
2) Kitagawa T, Tanimoto Y, Nemoto K, Aida M. Influence of cortical bone quality on stress
distribution in bone around dental implant. Dental materials journal. 2005; 24(2):219-24.
3) Kaleli N, Sarac D, Külünk S, Öztürk Ö. Effect of different restorative crown and customized
abutment materials on stress distribution in single implants and peripheral bone: A three-
dimensional finite element analysis study. The Journal of prosthetic dentistry. 2018; 119(3):437-
45.
4) Alqahtani M. Mechanical properties enhancement of self-cured PMMA reinforced with zirconia
and boron nitride nanopowders for high-performance dental materials. Journal of the Mechanical
Behavior of Biomedical Materials. 2020; 110:103937.
5) Kaleli N, Saraç D and Kulunk S (2017). Effect of different restorative crown and customized
abutment materials on stress distribution in single implants and peripheral bone: A three-
dimensional finite element analysis. The Journal of prosthetic dentistry; 17:17-23.
6) Kátia R, Paula Benetti, Álvaro D, Pedro H, Juliana A. Medeiros D, Ediléia L and Márcia B
(2018). How does the piston material affect the in vitro mechanical behavior of dental ceramics.
The Journal of prosthetic dentistry; 20:862-72.
7) Ozkir SE (2018). Effect of restoration material on stress distribution on partial crowns: A 3D
finite element analysis. Journal of dental science; 13, 311-317
8) Himmlova L, Dostalova T, Kacovsky A and Konvickova S (2004). Influence of implant length
and diameter on stress distribution: a finite element analysis. Journal Prosthet Dent; 91: 20-25
9) Bijjargi S, Chowdhary R (2013). Stress dissipation in the bone through various crown materials
of dental implant a 2-D finite element analysis. J Investig Clin Dent; 4:172-7.
10) Güngör, Merve B and Secil KN (2018). “Fracture resistance of CAD-CAM monolithic ceramic
and veneered zirconia molar crowns after aging in a mastication simulator.” Journal of Prosthetic
Dentistry; 119(3):473-480.
11) Duan Y & Griggs JA (2015). Effect of elasticity on stress distribution in CAD/CAM dental
crowns: Glass ceramic vs. polymer–matrix composite. Journal of Dentistry; 43: 742-749.
护理杂志
Journal of Nursing
Volume 70 Number 11
ISSN: 0047-262X
DOI: 10.5281/zenodo.10278986
25 www.jnr.tw
12) Duarte S, Sartori N & Phark J-H (2016). Ceramic-reinforced polymers: CAD/CAM hybrid
restorative materials. Current Oral Health Reports; 3: 198-202.
13) Gracis S, Thompson VP, Ferencz JL, Silva NR & Bonfante EA (2015). A new classification
system for all-ceramic and ceramic-like restorative materials. International Journal of
prosthodontics; 28.
14) De Kok p, Kleverlaan CJ, de Jager N, Kuijs R and Feilzer AJ (2012). Mechanical performance
of implant-supported posterior crowns. The journal of prosthetic dentistry; V (114): 59-66
15) Jassim ZM and Majeed MA (2018). Comparative Evaluation of the Fracture Strength of
Monolithic Crowns Fabricated from Different all-ceramic CAD/CAM Materials (an in vitro
study). Biomedical & Pharmacology Journal; Vol. 11(3); 1689-1697

More Related Content

Similar to FINITE ELEMENT ANALYSIS AND STATIC LOADING IN DENTAL IMPLANT WITH INNOVATIVE CROWN DESIGN (IN VITRO STUDY)

La Jolla Dental Implants
La Jolla Dental ImplantsLa Jolla Dental Implants
La Jolla Dental ImplantsPermadontics
 
A Retrospective Study of the AL Technology Implant System used for Single-To...
 A Retrospective Study of the AL Technology Implant System used for Single-To... A Retrospective Study of the AL Technology Implant System used for Single-To...
A Retrospective Study of the AL Technology Implant System used for Single-To...Abu-Hussein Muhamad
 
The use of angulated abutments in
The use of angulated abutments inThe use of angulated abutments in
The use of angulated abutments inZardasht Bradosty
 
Implant structure, components, and surface modification
Implant structure, components, and surface modificationImplant structure, components, and surface modification
Implant structure, components, and surface modificationZainab Mahmood
 
“One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central Incisor
 “One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central Incisor “One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central Incisor
“One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central IncisorAbu-Hussein Muhamad
 
Central incisor implant
Central incisor implantCentral incisor implant
Central incisor implantNader Elbokle
 
Bio hybrid the next gen implant
Bio hybrid the next gen implantBio hybrid the next gen implant
Bio hybrid the next gen implantIJARIIT
 
Implant stability intech
Implant stability intechImplant stability intech
Implant stability intechAsmita Sodhi
 
Lecture BDS IV Implant Dentistry
Lecture BDS IV Implant DentistryLecture BDS IV Implant Dentistry
Lecture BDS IV Implant DentistryRakesh Chandran
 
Effect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegrationEffect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegrationNavneet Randhawa
 
An 2/2 Implant Overdenture
An 2/2 Implant OverdentureAn 2/2 Implant Overdenture
An 2/2 Implant Overdentureasclepiuspdfs
 
Stress Reduction in Implant using Improved Abutment
Stress Reduction in Implant using Improved AbutmentStress Reduction in Implant using Improved Abutment
Stress Reduction in Implant using Improved AbutmentIRJET Journal
 
EXTRACTION,IMMEDIATE IMPLANT-A CASE REPORT
EXTRACTION,IMMEDIATE IMPLANT-A CASE REPORTEXTRACTION,IMMEDIATE IMPLANT-A CASE REPORT
EXTRACTION,IMMEDIATE IMPLANT-A CASE REPORTAbu-Hussein Muhamad
 
Implant esthetics - Integrated approach
Implant esthetics -  Integrated approachImplant esthetics -  Integrated approach
Implant esthetics - Integrated approachLaju Mahesh
 
Impact of dental implant surface modifications on Osseo-integration
Impact of dental implant surface modifications on Osseo-integrationImpact of dental implant surface modifications on Osseo-integration
Impact of dental implant surface modifications on Osseo-integrationNaveed AnJum
 
Experimental Study on Strength Evaluation Applied for Teeth.pptx
Experimental Study on Strength Evaluation Applied for Teeth.pptxExperimental Study on Strength Evaluation Applied for Teeth.pptx
Experimental Study on Strength Evaluation Applied for Teeth.pptxNandhakumar28773
 

Similar to FINITE ELEMENT ANALYSIS AND STATIC LOADING IN DENTAL IMPLANT WITH INNOVATIVE CROWN DESIGN (IN VITRO STUDY) (20)

La Jolla Dental Implants
La Jolla Dental ImplantsLa Jolla Dental Implants
La Jolla Dental Implants
 
A Retrospective Study of the AL Technology Implant System used for Single-To...
 A Retrospective Study of the AL Technology Implant System used for Single-To... A Retrospective Study of the AL Technology Implant System used for Single-To...
A Retrospective Study of the AL Technology Implant System used for Single-To...
 
The use of angulated abutments in
The use of angulated abutments inThe use of angulated abutments in
The use of angulated abutments in
 
Implant structure, components, and surface modification
Implant structure, components, and surface modificationImplant structure, components, and surface modification
Implant structure, components, and surface modification
 
“One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central Incisor
 “One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central Incisor “One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central Incisor
“One-Piece” Immediate-Load Post-Extraction Implant In Maxillary Central Incisor
 
Central incisor implant
Central incisor implantCentral incisor implant
Central incisor implant
 
Bio hybrid the next gen implant
Bio hybrid the next gen implantBio hybrid the next gen implant
Bio hybrid the next gen implant
 
Implant stability intech
Implant stability intechImplant stability intech
Implant stability intech
 
Lecture BDS IV Implant Dentistry
Lecture BDS IV Implant DentistryLecture BDS IV Implant Dentistry
Lecture BDS IV Implant Dentistry
 
Effect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegrationEffect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegration
 
An 2/2 Implant Overdenture
An 2/2 Implant OverdentureAn 2/2 Implant Overdenture
An 2/2 Implant Overdenture
 
Ecde 13-00466
Ecde 13-00466Ecde 13-00466
Ecde 13-00466
 
Stress Reduction in Implant using Improved Abutment
Stress Reduction in Implant using Improved AbutmentStress Reduction in Implant using Improved Abutment
Stress Reduction in Implant using Improved Abutment
 
EXTRACTION,IMMEDIATE IMPLANT-A CASE REPORT
EXTRACTION,IMMEDIATE IMPLANT-A CASE REPORTEXTRACTION,IMMEDIATE IMPLANT-A CASE REPORT
EXTRACTION,IMMEDIATE IMPLANT-A CASE REPORT
 
3rd publication JCDR-8th name.pdf
3rd publication JCDR-8th name.pdf3rd publication JCDR-8th name.pdf
3rd publication JCDR-8th name.pdf
 
Implant esthetics - Integrated approach
Implant esthetics -  Integrated approachImplant esthetics -  Integrated approach
Implant esthetics - Integrated approach
 
Impact of dental implant surface modifications on Osseo-integration
Impact of dental implant surface modifications on Osseo-integrationImpact of dental implant surface modifications on Osseo-integration
Impact of dental implant surface modifications on Osseo-integration
 
4
44
4
 
21 palermo, minetti 2
21   palermo, minetti 221   palermo, minetti 2
21 palermo, minetti 2
 
Experimental Study on Strength Evaluation Applied for Teeth.pptx
Experimental Study on Strength Evaluation Applied for Teeth.pptxExperimental Study on Strength Evaluation Applied for Teeth.pptx
Experimental Study on Strength Evaluation Applied for Teeth.pptx
 

More from indexPub

CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...
CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...
CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...indexPub
 
LEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTS
LEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTSLEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTS
LEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTSindexPub
 
THE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED WITH SCHIZOPHRENIA
THE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED  WITH SCHIZOPHRENIATHE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED  WITH SCHIZOPHRENIA
THE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED WITH SCHIZOPHRENIAindexPub
 
RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)
RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)
RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)indexPub
 
INNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORY
INNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORYINNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORY
INNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORYindexPub
 
CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...
CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...
CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...indexPub
 
ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH CURVED ARTERY WITH MIL...
ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH  CURVED ARTERY WITH MIL...ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH  CURVED ARTERY WITH MIL...
ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH CURVED ARTERY WITH MIL...indexPub
 
ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...
ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...
ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...indexPub
 
IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...
IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...
IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...indexPub
 
ACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINT
ACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINTACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINT
ACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINTindexPub
 
IMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTAN
IMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTANIMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTAN
IMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTANindexPub
 
A NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASE
A NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASEA NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASE
A NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASEindexPub
 
INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...
INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...
INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...indexPub
 
DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...
DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...
DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...indexPub
 
ENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACH
ENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACHENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACH
ENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACHindexPub
 
WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...
WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...
WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...indexPub
 
ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...indexPub
 
ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...indexPub
 
STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...
STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...
STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...indexPub
 
EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...
EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...
EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...indexPub
 

More from indexPub (20)

CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...
CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...
CORRELATION BETWEEN EMPATHY AND FRIENDSHIP QUALITY AMONG HIGH SCHOOL STUDENTS...
 
LEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTS
LEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTSLEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTS
LEVELS OF DEPRESSION AND SELF-ESTEEM IN STUDENTS
 
THE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED WITH SCHIZOPHRENIA
THE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED  WITH SCHIZOPHRENIATHE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED  WITH SCHIZOPHRENIA
THE IMPACT OF SOCIAL FACTORS ON INDIVIDUALS DIAGNOSED WITH SCHIZOPHRENIA
 
RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)
RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)
RETURN ON EQUITY (ROE) AS MEDIATION OF BANK'S CAPITAL ADEQUATION RATIO (CAR)
 
INNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORY
INNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORYINNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORY
INNOVATIVE DESIGN FOR KIDS MASTERY IMPROVEMENT OF LANGUAGE FEATURES IN A STORY
 
CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...
CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...
CARDIOVASCULAR DISEASE DETECTION USING MACHINE LEARNING AND RISK CLASSIFICATI...
 
ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH CURVED ARTERY WITH MIL...
ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH  CURVED ARTERY WITH MIL...ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH  CURVED ARTERY WITH MIL...
ANALYSIS OF FLOW CHARACTERISTICS OF THE BLOOD THROUGH CURVED ARTERY WITH MIL...
 
ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...
ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...
ANALYSIS OF STUDENT ACADEMIC PERFORMANCE USING MACHINE LEARNING ALGORITHMS:– ...
 
IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...
IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...
IMPLEMENTATION OF COMPUTER TECHNOLOGY IN BLENDED LEARNING MODELS: EFFECTS ON ...
 
ACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINT
ACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINTACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINT
ACADEMIC BANK OF CREDIT: A WORLDWIDE VIEWPOINT
 
IMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTAN
IMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTANIMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTAN
IMPACT OF MICROFINANCE ON POVERTY REDUCTION IN SOUTHERN PUNJAB PAKISTAN
 
A NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASE
A NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASEA NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASE
A NOVEL DENSITY-BASED CLUSTERING ALGORITHM FOR PREDICTING CARDIOVASCULAR DISEASE
 
INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...
INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...
INFLUENCE OF ADDITIVES ON THE MECHANICAL PROPERTIES OF HIGH-MODULUS ASPHALT C...
 
DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...
DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...
DIALECTAL VARIABILITY IN SPOKEN LANGUAGE: A COMPREHENSIVE SURVEY OF MODERN TE...
 
ENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACH
ENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACHENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACH
ENHANCING ACCURACY IN HEART DISEASE PREDICTION: A HYBRID APPROACH
 
WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...
WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...
WEB-BASED APPLICATION LAYER DISTRIBUTED DENIAL-OF-SERVICE ATTACKS: A DATA-DRI...
 
ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
 
ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
ANALYSIS OF BANK MANDIRI'S HEALTH LEVEL BASED ON RISK PROFILE, GOOD CORPORATE...
 
STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...
STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...
STRATEGIC FINANCIAL PERFORMANCE ANALYSIS USING ALTMAN’S Z SCORE MODEL: A STUD...
 
EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...
EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...
EVALUATING REFLECTIVE SPECTROSCOPY FOR PREDICTING SOIL PROPERTIES IN GAJAPATI...
 

Recently uploaded

Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...
Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...
Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...home
 
Government polytechnic college-1.pptxabcd
Government polytechnic college-1.pptxabcdGovernment polytechnic college-1.pptxabcd
Government polytechnic college-1.pptxabcdshivubhavv
 
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai DouxDubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Douxkojalkojal131
 
CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service 🧵
CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service  🧵CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service  🧵
CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service 🧵anilsa9823
 
The_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdf
The_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdfThe_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdf
The_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdfAmirYakdi
 
Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)amitlee9823
 
Delhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip Call
Delhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip CallDelhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip Call
Delhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip Callshivangimorya083
 
Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...amitlee9823
 
Top Rated Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...
Top Rated  Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...Top Rated  Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...
Top Rated Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...Call Girls in Nagpur High Profile
 
Editorial design Magazine design project.pdf
Editorial design Magazine design project.pdfEditorial design Magazine design project.pdf
Editorial design Magazine design project.pdftbatkhuu1
 
SD_The MATATAG Curriculum Training Design.pptx
SD_The MATATAG Curriculum Training Design.pptxSD_The MATATAG Curriculum Training Design.pptx
SD_The MATATAG Curriculum Training Design.pptxjanettecruzeiro1
 
The history of music videos a level presentation
The history of music videos a level presentationThe history of music videos a level presentation
The history of music videos a level presentationamedia6
 
Tapestry Clothing Brands: Collapsing the Funnel
Tapestry Clothing Brands: Collapsing the FunnelTapestry Clothing Brands: Collapsing the Funnel
Tapestry Clothing Brands: Collapsing the Funneljen_giacalone
 
WAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past QuestionsWAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past QuestionsCharles Obaleagbon
 
VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...Call Girls in Nagpur High Profile
 
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130Suhani Kapoor
 
Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...
Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...
Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...babafaisel
 
Best VIP Call Girls Noida Sector 44 Call Me: 8448380779
Best VIP Call Girls Noida Sector 44 Call Me: 8448380779Best VIP Call Girls Noida Sector 44 Call Me: 8448380779
Best VIP Call Girls Noida Sector 44 Call Me: 8448380779Delhi Call girls
 
Top Rated Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated  Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...Top Rated  Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...Call Girls in Nagpur High Profile
 

Recently uploaded (20)

Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...
Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...
Recommendable # 971589162217 # philippine Young Call Girls in Dubai By Marina...
 
Government polytechnic college-1.pptxabcd
Government polytechnic college-1.pptxabcdGovernment polytechnic college-1.pptxabcd
Government polytechnic college-1.pptxabcd
 
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai DouxDubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
 
CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service 🧵
CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service  🧵CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service  🧵
CALL ON ➥8923113531 🔝Call Girls Kalyanpur Lucknow best Female service 🧵
 
The_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdf
The_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdfThe_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdf
The_Canvas_of_Creative_Mastery_Newsletter_April_2024_Version.pdf
 
Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Nagavara ☎ 7737669865☎ Book Your One night Stand (Bangalore)
 
Delhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip Call
Delhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip CallDelhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip Call
Delhi Call Girls Paharganj 9711199171 ☎✔👌✔ Whatsapp Hard And Sexy Vip Call
 
Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Brookefield Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
 
Top Rated Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...
Top Rated  Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...Top Rated  Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...
Top Rated Pune Call Girls Koregaon Park ⟟ 6297143586 ⟟ Call Me For Genuine S...
 
Editorial design Magazine design project.pdf
Editorial design Magazine design project.pdfEditorial design Magazine design project.pdf
Editorial design Magazine design project.pdf
 
SD_The MATATAG Curriculum Training Design.pptx
SD_The MATATAG Curriculum Training Design.pptxSD_The MATATAG Curriculum Training Design.pptx
SD_The MATATAG Curriculum Training Design.pptx
 
The history of music videos a level presentation
The history of music videos a level presentationThe history of music videos a level presentation
The history of music videos a level presentation
 
Tapestry Clothing Brands: Collapsing the Funnel
Tapestry Clothing Brands: Collapsing the FunnelTapestry Clothing Brands: Collapsing the Funnel
Tapestry Clothing Brands: Collapsing the Funnel
 
WAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past QuestionsWAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past Questions
 
B. Smith. (Architectural Portfolio.).pdf
B. Smith. (Architectural Portfolio.).pdfB. Smith. (Architectural Portfolio.).pdf
B. Smith. (Architectural Portfolio.).pdf
 
VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Dange Chowk (8250192130) Pune Escorts Nearby with Comple...
 
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
 
Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...
Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...
Kala jadu for love marriage | Real amil baba | Famous amil baba | kala jadu n...
 
Best VIP Call Girls Noida Sector 44 Call Me: 8448380779
Best VIP Call Girls Noida Sector 44 Call Me: 8448380779Best VIP Call Girls Noida Sector 44 Call Me: 8448380779
Best VIP Call Girls Noida Sector 44 Call Me: 8448380779
 
Top Rated Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated  Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...Top Rated  Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated Pune Call Girls Saswad ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
 

FINITE ELEMENT ANALYSIS AND STATIC LOADING IN DENTAL IMPLANT WITH INNOVATIVE CROWN DESIGN (IN VITRO STUDY)

  • 1. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 14 www.jnr.tw Finite Element Analysis and Static Loading in Dental Implant with Innovative Crown Design (IN VITRO STUDY) Dr. Twana Omar Bakr 1, Dr. Fahd S. Ikram 2 and Dr. Nabeel S. Martani 3 1 B.D.S. M.Sc. Prosthodontic, College of Dentistry, Hawler Medical University. 2, 3 B.D.S. M.Sc. PhD Prosthodontic, College of Dentistry, Hawler Medical University. Abstract Objective: The purpose of this study is to examine how different crown materials affect stress transformation and distribution around both dental implants. To do this, integrated and combined crowns were built, modelled, and tested under static axial loads using various material combinations. The biomechanical response was then examined. Methods: A validated three-dimensional finite element (FE) models of crown supported by implant was developed by to evaluate the effect of the different type of material (E max, zirconia, composite) on the short implant. After the FE models had been validated, simulations utilizing various configurations of various crowns fixed to two distinct types of implants were run and subjected to static loading to ascertain the distribution of stresses inside the bone around the implants. Result: The comparative results showed that manufacturing the crown using softer material (i.e., materials with lower elastic modulus) reduced the stress distribution in crown, implant and cancellous bone. It may refer to this phenomenon that softer material can absorb more energy from the applied compressive load, and result in transferring less energy to the implant and jaw bone. However, this effect was not significant on cortical bone compared to the cancellous bone. Combination of different materials for design and manufacturing the crown can alter the biomechanical response and could be beneficial for decreasing the stress distribution in implant and spongy region of jaw bone when stiffer material is needed to be covered in upper surface of the crown. In addition, the results suggests that shorter implant can increase the stress distribution in both cortical and cancellous bone. Conclusion: by using stiff material the stress will increase on the parts of implant and the surrounding bone which may led to failure of implant or bone resorption around of the implant, in other way by using less stiffer material the possibility of success will be increased and also the success rate of the implant is increased, also before deciding which implant size and length are used you select which type of the prosthetic will be used. Keywords: Dental Biomechanics, Dental Implant, Innovative Crown, Zirconia, E-Max, Composite, Implant Length, Finite Element Modeling INTRODUCTION In order to regain complete, pleasant masticatory function and face esthetic, humans have tried using a variety of methods to restore lost teeth throughout history. Before to the osseointegration period, there were several designs of dental implants and frameworks that were used to support dentures and partial dentures with varying degrees of effectiveness. Among the various materials used in implants are porcelain, cobalt chromium, and radioplatinum, but the discovery of titanium changed the course of implant history. There are several techniques for comprehending the force distribution, including analytical mathematical models that include strain gauges, the photo elastic model, and studies like the finite element analysis (vandana and kartik, 2004). FEA has been extensively used in implant dentistry to foresee the biomechanical behavior of various dental implant designs as well as the influence of clinical factors on clinical success. There has been a great deal of study done on the stress patterns in implant components and the surrounding
  • 2. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 15 www.jnr.tw bone. The accuracy of the used simulating structures influences the outcome of every FE investigation. These are the material properties of the implant and bone, the surface features and shapes of the implant and its parts, the force applied, and the biomechanical behavior of the implant- bone relationship. (Trivedi S, 2014) In FEA, a specific physical system's behavior is mathematically modelled. The FEA separated the structure of the model into several distinct parts, each of which retains the structure's original features. Each component requires a unique equation that must be solved using mathematical models chosen in accordance with the facts being examined. (Boccaccio et al, 2011, Dejak and Mlotkowski, 2011). By altering the parameters of those geometries, FEA makes it possible to apply a force or a system of forces to any point and/or in any direction, providing knowledge of the movement, amount of tension, and compression force on each area. It also converts natural or artificial tissue into complex structures that accurately represent the original one. (Vasudeva G, 2009) In addition to providing a forecast of a desirable outcome with the lowest chance of failure, FEA assists in the study of stress direction and distribution between natural teeth and the kind of material used to reconstruct the original tooth or dental structure if they are lost. Zarone et al, 2005 researchers looked at how stress surrounding teeth from everyday load on the maxillary central incisor was affected by finishing line preparation with an alumina porcelain veneer in 2005. They advised utilizing a chamfer with palatal overlap design when repairing with porcelain veneers because it achieves the advised stress distribution more effectively than the window approach. (Srirekha and Bashetty, 2010) MATERIAL METHODS 1. Model preparation For the model preparation in the first ready-made dental arch is used which have all the teeth and in full occlusion, the teeth are attached to the base by the screw, first lower molar is choice and replaced by the implant. After the teeth is removed the socket of the teeth is grinded to become wider for the fixation, for the parallelism of the implant dental surveyor is used (dental farm S.r.I. Via Susa, 9/A-10138 torino-italy). Table of dental surveyor is balanced until become parallel with floor by the bubble level, then the dental arch is locked on the table by model clamp lock nut of the device laterally like each corner of the triangular. Implant fixed to the tool adapter holder and perpendicular to the table. The head of the surveyor pushed implant to the socket and placed in the center without touching any border and the level of the implant neck below the level of the gum about 2-3 mm then fixed the tool holder in the place, Cold cure acrylic (vertex type, Vertex Dental ByV, 872.3760 EBI, ISO020795) is choice for fixation the implant which is mixed according to the manufacture instruction and used before reach to dough stage, in flow stage. Cold cure acrylic is putted to the socket and waited until the material become set then the tool holder is opened and the cast removed for the table and watched for any problem present. Then the surface of the acrylic around the implant is grinded with low-speed engine to provide smooth surface of the acrylic and free of projection and irregularities. After that the length of the implant adjusted and shortened until provide good space for occlusal surface of the crown.
  • 3. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 16 www.jnr.tw 2. Scanning the model After the model is prepared TRIOS, scanner is used (3Shape TRIOS® 3, Copenhagen, Denmark) to scan the model, scanning is done for all the lower arch and opposite arch in respect to occlusion and making STL file, then the design of the samples is made which consist of full crown and two-part crown of Emax, zircon and composite. Figure 1 Figure 1: TRIOS3 intraoral scanner 3. Design of samples Crown are made in two forms first form is made in full anatomical crown without any changing and second form made in to two parts which is first part 1/3 occlusal part and second part 2/3 gingivally which are connect to each other by resin in between the two parts there is 4 projection in the first part and 4 holes in the second part which orient the two parts to each other and prevent movement of two parts, the projection are orient in square shape each corner has one projection, Length of the projection is about 2mm. 4. Development of the geometrical models The STL input data which was received based on scanned parts (Figure 2) were evaluated. To develop fine geometrical models which can be used in finite element (FE) analyses, the STL files were imported to Solid works (Dassault Système© , Vélizy-Villacoublay, France) and the surfaces were trimmed and smoothed to achieve acceptable geometries for exporting to FE package. To simplify the simulation, a sliced part of the jaw bone was mimicked in simulations. Figure 2: Samples of the scanned parts in STL format Based on objective of this study, four general geometrical models were developed based on combination of (1) integrated and (2) combined crown implanted using short implants. The diameters for short implants were 4.5mm and their lengths were 6mm, respectively. Figure 3 show the details of the aforementioned models.
  • 4. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 17 www.jnr.tw 5. Development of the FE models The geometrical models were imported to ABAQUS package (SIMULIA, Providence, RI, USA) to simulate the effect of material properties of the crown and implant size on biomechanical response. For this purpose, 18 FE models were developed as following scenario: Model 1: E-Max Model 2: Composite Resin Model 3: Zirconia Model 4: E-Max / Composite Resin Model 5: E-Max / Zirconia Model 6: Composite Resin / E-Max Model 7: Composite Resin / Zirconia Model 8: Zirconia / E-max Model 9: Zirconia / Composite Resin Figure 3: The arrangement of different FE models Mechanical properties of the model were considered as isotropic elastic theory, which was extracted from the available data in literature (Table 1)
  • 5. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 18 www.jnr.tw Table 1: Mechanical properties of different components in FE models The analyses were classified as static and the solution examined the governing equation system using the element diagonal mass matrix and the law of explicit integration. The applied boundary condition between different components were considered using the discretization method of surface to surface with tie contact property to simulate perfect connection to constrain equal degrees of freedom (i.e., equal translational and rotational motions). The static compressive load equal to 200 N [7] was vertically applied on the upper surface of the crown for comparative analysis using the reference point positioning technique as shown in Figure 4. Figure 4: The applied axial compressive concentrated force to FE model 10-node quadratic tetrahedron elements were used to generate the meshed models (Figure 5 to 6) and sensitivity analyses to assess the independency of the results from meshing were performed. The finalized number of elements after mesh sensitivity analyses were considered as Table 2 and 3. Table 2: Number of elements for finalized FE model after convergence analyses for integrated crown models with short implants Components Number of Elements Short Implant 13810 Crown 25955 Cancellous Bone 38381 Cortical Bone 33134 Components Elastic modulus (GPa) Poisson’s ratio Implant 110 [1] 0.3 [1] Abutment 110 [1] 0.3 [1] Cancellous Bone 1.37 [2, 3] 0.3 [2, 3] Cortical Bone 12.6 [2, 3] 0.3 [2, 3] E-Max 95 [3, 4] 0.2 [3, 4] Zirconia 210 [5] 0.33 [5] Composite resin 21 [6] 0.24 [6]
  • 6. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 19 www.jnr.tw Table 3: Number of elements for finalized FE model after convergence analyses for combined crown models with short implants Components Number of Elements Short Implant 13810 Crown (Upper Part) 28398 Crown (Lower Part) 25063 Cancellous Bone 38381 Cortical Bone 33134 Figure 5: Meshing the FE models based on sensitivity analyses (Integrated crown with short implant) Figure 6: Meshing the FE models based on sensitivity analyses (Combined crown with short implant) The simulations were performed as described previously by comparing the results for 18 different models (i.e., Based on combination of (1) integrated and (2) combined crown implanted using short implants.) subjected to vertical static force equal to 200N. The values of Von-Mises stress were calculated in critical nodes in different components of the models (i.e., cortical bone, cancellous bone, implants, and crown) for comparative study. RESULT The results of sensitivity analyses verified the accuracy of the FE models for investigation of the main objective of this study. The calculated values of Von-Mises stress in different FE models subjected to static loading (i.e., vertical compressive force equal to 200N) over the upper surface of the crown were extracted. The maximum equivalent stress in critical regions achieved results were calculated in different components of different models (i.e., cortical bone, cancellous bone, implants, abutments, and crown) and the calculated results were compared.
  • 7. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 20 www.jnr.tw The comparative results are summarized in figures 9 to 16 in which the values of Von-Mises stress were shown for different implant configurations with different materials for the crowns. The result from the static loading appears that the stress formation inside the crown for the short implant with zircon crown has higher value while followed by the E-max and the last one composite crown. The details are show in figure (7). The most highly statistically significant one is composite VS zirconia and the least one which is non statically significant is E-Max VS. Composite and other are between them as shown in table (3): Figure 7: Comparative results for values of maximum Von-Mises stress in crown for the models with integrated crown using short implant. The result from the static loading appears that the stress formation inside crown while using zirconia as the superior part of the combined crown in both Z/E, Z/C crown while followed by the E/Z and E/C and the last one C/E and C/Z. The details are show in figure (8). The most statistically significant one is Composite/Zirconia vs. Zirconia/Composite and last non statically significant one is Composite/E-max vs. Composite/Zirconia and other are between them as shown in table (4): Figure 8: Comparative results for values of maximum Von-Mises stress in crown for the models with combined crown using short implant. 0 10 20 30 40 50 60 E-Max Composite Resin Zirconia Maximum Equivalent Stresses (MPa) E-Max Composite Resin Zirconia 0 10 20 30 40 50 60 Maximum Equivalent Stresses (MPa) E-Max/Composite E-max/Zirconia Composite/E-max Composite/Zirconia Zirconia/E-max Zirconia/Composite
  • 8. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 21 www.jnr.tw Table 4: statistical analysis of crown Pair (Short Implant) Crown P- value Statistical analysis E-Max vs. Composite Resin 0.07329 E-Max vs. Zirconia 0.0447 * Composite Resin vs. Zirconia 0.0001785 * E-Max/Composite vs. E-max/Zirconia 0.9998 E-Max/Composite vs. Composite/E-max 0.001219 * E-Max/Composite vs. Composite/Zirconia 0.00111 * E-Max/Composite vs. Zirconia/E-max 0.04841 * E-Max/Composite vs. Zirconia/Composite 0.01694 * E-max/Zirconia vs. Composite/E-max 0.0005098 * E-max/Zirconia vs. Composite/Zirconia 0.0004629 * E-max/Zirconia vs. Zirconia/E-max 0.09124 E-max/Zirconia vs. Zirconia/Composite 0.03459 * Composite/E-max vs. Composite/Zirconia 1 Composite/E-max vs. Zirconia/E-max 2.91E-08 * Composite/E-max vs. Zirconia/Composite 6.65E-09 * Composite/Zirconia vs. Zirconia/E-max 2.61E-08 * Composite/Zirconia vs. Zirconia/Composite 5.98E-09 * Zirconia/E-max vs. Zirconia/Composite 0.9987 Pair (Short Implant) Crown Mean SD Zirconia 40.496 9.980093743 E-Max 31.853 7.423834065 Composite Resin 23.706 5.821193463 E-Max/Composite 35.348 7.160511776 E-max/Zirconia 36.143 6.533477975 Composite/E-max 22.6344 4.3443233 Composite/Zirconia 22.5478 4.069622363 Zirconia/E-max 44.265 8.556589728 Zirconia/Composite 45.455 8.26326711 The result forms the static loading appear that the stress formation in implant itself for the short implant more for the full anatomical zirconia crown while followed by the E-max and the least stress formation is composite crown. The details are show in figure (9). The most statistically significant one is Composite Resin vs. Zirconia and the least one which is non statically significant is E-Max vs. Zirconia and other are between them as shown in table (5): Figure 9: Comparative results for values of maximum Von-Mises stress in implants for the models with integrated crown using short implant. 0 5 10 15 20 25 30 35 40 E-Max Composite Resin Zirconia Maximum Equivalent Stresses (MPa) E-Max Composite Resin Zirconia
  • 9. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 22 www.jnr.tw The result from the static loading appears that the stress formation in implant itself for the short implant the result show more stress formation while using zirconia crown as superior part in both Z/E, Z/C while followed by the E/Z and E/C then C/E and C/Z. The details are show in figure (12). The most highly statistically significant one is Composite/E-max vs. Zirconia/E-max and the least one which is non statically significant is E-Max/Composite vs. E-max/Zirconia and other are between them as shown in table (10): Figure 10: Comparative results for values of maximum Von-Mises stress in implants for the models with combined crown using short implant. Table 5: statistical analysis of implant Pair (Short Implant) implant P- value Statistical analysis E-Max vs. Composite Resin 0.1475 E-Max vs. Zirconia 0.4093 Composite Resin vs. Zirconia 0.008718 * E-Max/Composite vs. E-max/Zirconia 1 E-Max/Composite vs. Composite/E-max 0.4445 E-Max/Composite vs. Composite/Zirconia 0.8594 E-Max/Composite vs. Zirconia/E-max 0.03974 * E-Max/Composite vs. Zirconia/Composite 0.2354 E-max/Zirconia vs. Composite/E-max 0.4254 E-max/Zirconia vs. Composite/Zirconia 0.8451 E-max/Zirconia vs. Zirconia/E-max 0.04306 * E-max/Zirconia vs. Zirconia/Composite 0.2493 Composite/E-max vs. Composite/Zirconia 0.981 Composite/E-max vs. Zirconia/E-max 0.0001312 * Composite/E-max vs. Zirconia/Composite 0.001906 * Composite/Zirconia vs. Zirconia/E-max 0.001382 * Composite/Zirconia vs. Zirconia/Composite 0.01571 * Zirconia/E-max vs. Zirconia/Composite 0.966 0 10 20 30 40 Maximum Equivalent Stresses (MPa) E-Max/Composite E-max/Zirconia Composite/E-max Composite/Zirconia Zirconia/E-max Zirconia/Composite
  • 10. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 23 www.jnr.tw Pair (Short Implant) Implant Mean SD Zirconia 25.231 6.94377331 E-Max 21.924 5.506868842 Composite Resin 16.981 4.363598031 E-Max/Composite 20.174 4.414886937 E-max/Zirconia 20.233 4.309258895 Composite/E-max 16.709 3.525354041 Composite/Zirconia 18.024 3.074111254 Zirconia/E-max 25.883 4.574829809 Zirconia/Composite 24.38 4.956457068 DISCUSSION The result of the study show that the stress formation on parts of crown of the two part crown in short implant consist of zirconia base 2/3 gingival crown and 1/3 composite or e max occlusal part are more than the other types of the crown formed by other hybrid crowns this because of the high modulus of elasticity, stress induced toughening mechanism and high fracture resistance this lead to more stress formation inside zirconia part of crown, this finding is similar with Güngör et al, 2018 where compared the fracture strength of all-ceramic crowns fabricated from zirconia and different other materials and found that the highest fracture strength and stress formation inside the crown mean value was recorded by crowns fabricated from zirconia. When the base 2/3 of the crown made form the composite for short implant the stress formation on the implant is less and more in zirconia base because material has law modulus of elasticity (less stiffer) so lead to more absorption of the force coming from the mastication and less stress formation, Inconsistent with our results, Duan and Griggs (Duan and Griggs, 2015) compared the stress distribution in lithium disilicate ceramic and resin nanoceramic CAD-CAM crowns and reported that resin nanoceramic crowns showed lower stress values under vertical loading on the implant. The polymer-infiltrated hybrid ceramic crown exhibited low stress values in the cervical region. The resin content (Gracis et al, 2015), (Duarte et al, 2016) of these materials is thought to provide a favorable stress distribution in restorative crowns. Opposite to our result (Skalak, 1983) found that resin-based restorative materials may reduce the stress on implants and peripheral bone. For the all position of stress formation on (crown, implant and bone), full anatomical zirconia crown form the greatest stress formation among all the types of the prosthetic part because material is hard and have no property of elasticity so they did not absorb any stress, when the stress formed either fracture or failure of the material occur or transmitted to the other part of the structure, then followed by E max material and the last one is resin composite which absorb most of the stress because of the spongy properties related to filler particle of material which provide more space inside structure so only transmitted little of the stress to the other part of implant structure. This founding is similar to study done by De kok P et al 2012 concluded that monolithic implant supported crowns have a higher initial load to failure than conventional veneered ceramic crowns and are expected to show clinically fewer fractures. Monolithic ceramic restorations perform better than composite resin crowns in force standing but composite is better in stress absorption. Composite resins with a low modulus of elasticity show promising results in withstanding the force absorption, in other way opposite to Jassim and Majeed (2018) which concluded that the highest fracture strength mean value was recorded by monolithic crowns fabricated from zirconia (CEREC Zirconia) followed by crowns fabricated from reinforced composite (BRILLIANT Crios), zirconia-reinforced lithium silicate (CELTRA DUO), and lithium disilicate (IPS e.max CAD), while the lowest fracture strength mean value was recorded by crowns fabricated from hybrid dental ceramic (VITA ENAMIC).
  • 11. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 24 www.jnr.tw CONCLUSION In summary, the comparative results showed that manufacturing the crown using softer material (i.e., materials with lower elastic modulus) reduced the stress distribution in crown, implant and cancellous bone. It may refer to this phenomenon that softer material can absorb more energy from the applied compressive load, and result in transferring less energy to the implant and jaw bone. However, this effect was not significant on cortical bone compared to the cancellous bone. Combination of different materials for design and manufacturing the crown can alter the biomechanical response and could be beneficial for decreasing the stress distribution in implant and spongy region of jaw bone when stiffer material is needed to be covered in upper surface of the crown. In addition, the results suggests that shorter implant can increase the stress distribution in both cortical and cancellous bone. References 1) Ueda N, Takayama Y, Yokoyama A. Minimization of dental implant diameter and length according to bone quality determined by finite element analysis and optimized calculation. Journal of prosthodontic research. 2017; 61(3):324-32. 2) Kitagawa T, Tanimoto Y, Nemoto K, Aida M. Influence of cortical bone quality on stress distribution in bone around dental implant. Dental materials journal. 2005; 24(2):219-24. 3) Kaleli N, Sarac D, Külünk S, Öztürk Ö. Effect of different restorative crown and customized abutment materials on stress distribution in single implants and peripheral bone: A three- dimensional finite element analysis study. The Journal of prosthetic dentistry. 2018; 119(3):437- 45. 4) Alqahtani M. Mechanical properties enhancement of self-cured PMMA reinforced with zirconia and boron nitride nanopowders for high-performance dental materials. Journal of the Mechanical Behavior of Biomedical Materials. 2020; 110:103937. 5) Kaleli N, Saraç D and Kulunk S (2017). Effect of different restorative crown and customized abutment materials on stress distribution in single implants and peripheral bone: A three- dimensional finite element analysis. The Journal of prosthetic dentistry; 17:17-23. 6) Kátia R, Paula Benetti, Álvaro D, Pedro H, Juliana A. Medeiros D, Ediléia L and Márcia B (2018). How does the piston material affect the in vitro mechanical behavior of dental ceramics. The Journal of prosthetic dentistry; 20:862-72. 7) Ozkir SE (2018). Effect of restoration material on stress distribution on partial crowns: A 3D finite element analysis. Journal of dental science; 13, 311-317 8) Himmlova L, Dostalova T, Kacovsky A and Konvickova S (2004). Influence of implant length and diameter on stress distribution: a finite element analysis. Journal Prosthet Dent; 91: 20-25 9) Bijjargi S, Chowdhary R (2013). Stress dissipation in the bone through various crown materials of dental implant a 2-D finite element analysis. J Investig Clin Dent; 4:172-7. 10) Güngör, Merve B and Secil KN (2018). “Fracture resistance of CAD-CAM monolithic ceramic and veneered zirconia molar crowns after aging in a mastication simulator.” Journal of Prosthetic Dentistry; 119(3):473-480. 11) Duan Y & Griggs JA (2015). Effect of elasticity on stress distribution in CAD/CAM dental crowns: Glass ceramic vs. polymer–matrix composite. Journal of Dentistry; 43: 742-749.
  • 12. 护理杂志 Journal of Nursing Volume 70 Number 11 ISSN: 0047-262X DOI: 10.5281/zenodo.10278986 25 www.jnr.tw 12) Duarte S, Sartori N & Phark J-H (2016). Ceramic-reinforced polymers: CAD/CAM hybrid restorative materials. Current Oral Health Reports; 3: 198-202. 13) Gracis S, Thompson VP, Ferencz JL, Silva NR & Bonfante EA (2015). A new classification system for all-ceramic and ceramic-like restorative materials. International Journal of prosthodontics; 28. 14) De Kok p, Kleverlaan CJ, de Jager N, Kuijs R and Feilzer AJ (2012). Mechanical performance of implant-supported posterior crowns. The journal of prosthetic dentistry; V (114): 59-66 15) Jassim ZM and Majeed MA (2018). Comparative Evaluation of the Fracture Strength of Monolithic Crowns Fabricated from Different all-ceramic CAD/CAM Materials (an in vitro study). Biomedical & Pharmacology Journal; Vol. 11(3); 1689-1697