Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016
DOI : 10.5121/ieij.2016.4201 1
THE PRESSURE SIGNAL CALIBRATION
TECHNOLOGY OF THE COMPREHENSIVE TEST
SYSTEM
Yixiong Xu
Mechatronic Engineering, Shanghai University of Engineering Science, Shanghai, China
ABSTRACT
The pressure signal calibration technology of the comprehensive test system which involved pressure
sensors was studied in this paper. The melioration of pressure signal calibration methods was elaborated.
Compared with the calibration methods in the lab and after analyzing the relevant problems,the
calibration technology online was achieved. The test datum and reasons of measuring error analyzed,the
uncertainty evaluation was given and then this calibration method was proved to be feasible and accurate.
KEYWORD
Calibration online; Pressure transducer; Pressure signal calibration
1. INTRODUCTION
Integrated test system for pressure transducer can be achieved on liquid, gas, electric and other
different media measurement and control of physical quantities, and it is mostly a complex
testing system of electric hydraulic control. It are widely used in industrial control, military,
aerospace and other fields. So the calibration of test system for pressure transmitter is
particularly important. The improvement of calibration technology ensure system stability,
accuracy and fast control of the parameters in a large extent, and provides quality assurance for
research and production.
2. THE CALIBRATION OF PRESSURE SIGNALS OF THE INTEGRATED TEST
SYSTEM
2.1 Pressure signal improvement of the calibration method
Calibration of the pressure signal is divided into two steps. The first step is a pressure transducer
calibration; Calibration of pressure signal is the second one[1]
.Conventional calibration method is
Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016
2
to test the transducer that is removed from the system in the laboratory; the transducer is
reinstalled in the test system after test qualified, and then through the pressure transducer, we
calibrate the secondary display device of the test system. This calibration method is cumbersome,
time consuming and have an accumulation of errors. All these lead to a large measurement error.
Therefore we make the following improvement of the calibration method: Firstly, the pressure
transducer is achieved on-line calibration instead of laboratory calibration; Secondly, the
secondary display instrument of test systems is calibrated, by accessing directly to the terminal
via a standard signal acquisition or signal source waveform generator.
2.2 On-line calibration of pressure transmitter
2.2.1 Realization of environmental conditions
Although pressure transducer calibration needs relatively harsh conditions, now many workshop
works has requirements of environment temperature, humidity and outside magnetic field. Thus
the calibration environment that temperature: 20℃±5℃, every 10min change is not more than
1℃, relative humidity: 45% to 75%, and there is no external magnetic field affecting the output
stable mechanical vibration and no other effects in addition to the geomagnetic off-normal
operation pressure transducer, can be achieved.
2.2.2 Build on-line calibration device
Calibration device consists of digital pressure calibrators, pressure source and composition of the
adapter, and the auxiliary equipment is a walkie-talkie. CST 2005 digital pressure calibrator,
intelligent modules and CST 1004 high pressure pump constitute the standard pressure source,
which weighs 15kg totally and is moved easily; Adapters is used to connected different thread
diameters of standard pressure sensors to the source, and common diameter is 12mm adapter
turn 20mm; Sometimes the control system is relatively far away from the pressure transducer on
site, and control system location apart more far, so two people are needed to complete the
calibration. Calibration device from the system as shown in Figure 1.
Fig.1 Composition of a test system and its calibration theory
Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016
3
2.2.3 The calibration method
According to state test procedures, we finish the appearance and basic error checks. After
calibration equipment and field equipment system get thermal equilibrium (2 hours), we installed
and preheat the equipments, as shown in Figure 1.Because the pressure standard pressure source
is difficult to maintain stable calibration points corresponding to the value in the measurement
process, we use the method of multiple readings averaged to improve the accuracy of reading.
Measurement error is calculated according to formula (1).If the test data meet the technical
requirements, it is judged as qualified pressure transducer; Otherwise, it is considered
unqualified.
△d=Yd1-Yd2 (1)
In the formula:△d-measurement error of the pressure transmitter calibrated point;Yd1—the value
of the standard digital pressure calibrator, Pa;Yd2—pressure indication of the software interface
or standard gauge display, Pa.
2.2.4 Advantages of pressure transducer calibration on-line
It is very important to achieve pressure transmitter calibration on-line, which improves the
efficiency of non-standard test equipment measurement ensures efficient conduct of scientific
research and production. Compared to calibration in the laboratory, on-line calibration has the
following advantages: firstly, through on-site demolition, on-site calibration and on-site
installation, this method has shorten the calibration period and improved the calibration
efficiency of the whole testing system; secondly, it is convenient to build the calibration device,
not complicated; Thirdly, flexible, it is not influenced by other conditions apart from the
calibration conditions.
2.3 Current signal calibration
The high-precision digital ammeter is connected in series circuit which is formed by standard
signal source and pressure signal acquisition end of. Digital ammeter works as a data monitoring
device and gets real-time data. Test software gives the current signal via the control panel;
Record the measured value and displayed value and calculate the current linearity error. Since the
electrical signal is small, in order to improve measurement accuracy and reduce human error
machine error, in the calibration process we should pay attention to the following matters: Firstly,
in the selection of monitoring digital ammeter, its resolution ratio should be more than four and a
half, the resolution should be more than four and a half, its accuracy should be more than five ten
thousandths, such as FLUKE8840;Secondly, when measuring we should try to avoid some of the
factors that caused the difference ,such as triboelectric effect, leakage current and piezoelectric
effect.
Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016
4
3. DATAANALYSIS
3.1 pressure transducer test data
Taking 10Mpa as an example, test the effect of calibration techniques improved by actual test.
Pressure transducer, measuring range: 0~10Mpa, precision: 0.5%.Test data in table 1.According
to the technical requirements, The pressure transducer is determined as: Qualified.
Tab.1 test datum of 10Mpa pressure sensor
3.2 Current test data
The test was made on the test data of the heat compressed air signal in the test station, whose
acquisition range from 4 to 20mA.The data recorded in table 2.
Tab.2 The test datum of current signal
Calibration
points((((mA))))
Current signal Found
((((mA))))
Current signal display
((((mA)))) Error((((mA))))
20 20.000 20.03 0.02
16 16.000 16.01 00
12 12.000 11.99 -0.01
8 8.000 7.98 -0.02
4 4.000 3.96 -0.04
3.3 Evaluation of uncertainty in measurement
3.3.1 Sources of uncertainty
a) The basic standard uncertainty: uA;
b) The uncertainty of the introduction of the standard pressure source: ub1;
c) The measurement uncertainty introduced by the digital current meter: ub2;
Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016
5
3.3.2 Evaluation of uncertainty
a) Basic standard type A evaluation of uncertainty
Since there are more test data, repeat the test for the pressure transducer 5Mpa profile and carry
on the uncertainty analysis; The test data record as is shown in Table 3.
Tab.4 repetitive test datum of 5Mpa of pressure sensor
Its average value:
10
1
5.0176i
i
x x n
=
= =∑ Mpa
Standard type A evaluation of uncertainty calculated according to the formula(2):
2
1
( ) ( ) ( 1)
n
i i
i
s x x x n
=
= − −∑
10
2
10
1
( ) ( ) (10 1) 0.00004548 9 0.0022i
i
s x x x
=
= − − = =∑ Mpa (2)
A class of uncertainty that measurement repeatability leads to is as follows:
uA= ( )10s x / n =0.0007 Mpa (3)
b)The uncertainty of the introduction of the standard pressure source:ub1
The main source is the intelligent pressure digital module of the standard pressure calibration
instrument and the error of the digital pressure calibration instrument, while the influence of
environmental temperature and humidity can be ignored. Type B evaluation of uncertainty.
Standard pressure calibration instrument 15Mpa pressure module allows error limit a1=0.003Mpa.
Its distribution is subject to uniform distribution, and take the factor k= 3 , calculated according
to the formula(4):
uc1= a1/k=0.001732 Mpa (4)
c) The measurement uncertainty introduced by the digital current meter:
Its main source is the uncertainty of the current introduction of the digital current meter in the
measurement of the pressure signal acquisition port and the standard signal source, and its type B
evaluation of uncertainty. The error limit of the digital meter 100mA measurement range is
0.01%, And a2=0.01mA, take the coverage factor k= 3 , calculated according to the formula
(5) :
Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016
6
A2/k =0.00577mA (5)
Converted into pressure by linear conversion:
uc2=0.00540Mpa (6)
d) Combined standard uncertainty Uc
Above all are independent and unrelated, and the combined standard uncertainty Uc is calculated
according to the formula (7) :
Uc = 2 2 2
1 2A b bu u u+ + =0.00571Mpa (7)
e) Evaluation of expanded uncertainty
All the sources of uncertainty are uniform distribution, and take k=2, according to the formula
(8):
U =2* Uc=0.0114Mpa (8)
f) The measuring result of the 5Mpa file of the pressure transducer is:
(5.0176±0.0114)Mpa.
3.3.3 Methods for reducing uncertainty of measurement
Through the analysis of the measurement uncertainty, we can see that the measurement
uncertainty u introduced by the digital current meter is relatively larger, the uncertainty of u
introduced by the standard pressure source is relatively moderate. , and the basic standard
uncertainty is relatively smaller. In order to improve the accuracy of measurement and reduce the
uncertainty of measurement, we have the following improvements: firstly, choosing high
precision digital ammeter can effectively reduce the measurement uncertainty introduced by the
digital current meter and the uncertainty of the introduction of the standard pressure source,
reducing system error; Thirdly, the A/D acquisition card could be modified to ensure consistency
between data acquisition and actual values.
4. SUMMARY
The containing pressure transducer test system calibration techniques were studied, focusing on
analyses and improvements of the system testing pressure signal calibration method, and through
the test data processing and analysis, it proves the feasibility and superiority of this calibration
method. It improves the calibration efficiency of pressure transmitter test system and the stability
and reliability of the test system in the test products, and has accumulated some calibration
experience for this type of test system.
Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016
7
REFERENCES
[1] Shanghai Institute of Measurement and Testing technology. JJG 882-2004 pressure transducer
calibration specification [s]. China measurement Publishing House, 2004
[2] China Metrology Association. Data processing and measurement professional practice [m].
Beijing: China's quality watchdog, 2013
[3] Sheng Hong Jiang, Feng Yi.Online Calibration of the Sensor [j]. Electronic Design Engineering,
2012,43 (1): 34-37.
[4] Tian Hui, Subaoming. The Measurement and Calibration of the Pressure Sensor System [j].
Electronic Component & Device Applications, 2008 (09): 44-46
Author
Yixiong Xu(1985-),postgraduate in Shanghai University of Engineering Science(SUES),
the main research interest now is Instrumentation measurement technology, address:
Room 1105,Build ing NO.243,Lane 777,Dushi road, Minhang district, Shanghai(201109),
fixed 021-34628108, mobile 13681897190,email:heroexu@126.com

More Related Content

What's hot

Ee2201 measurement-and-instrumentation-lecture-notes
Ee2201 measurement-and-instrumentation-lecture-notesEe2201 measurement-and-instrumentation-lecture-notes
Ee2201 measurement-and-instrumentation-lecture-notes
Jayakumar T
 

What's hot (20)

Measurement System
Measurement SystemMeasurement System
Measurement System
 
Ee2201 measurement-and-instrumentation-lecture-notes
Ee2201 measurement-and-instrumentation-lecture-notesEe2201 measurement-and-instrumentation-lecture-notes
Ee2201 measurement-and-instrumentation-lecture-notes
 
Measurement and instrumentation
Measurement and instrumentationMeasurement and instrumentation
Measurement and instrumentation
 
Chapter-1_Mechanical Measurement and Metrology
Chapter-1_Mechanical Measurement and MetrologyChapter-1_Mechanical Measurement and Metrology
Chapter-1_Mechanical Measurement and Metrology
 
QAL and standards in CEMS systems
QAL and standards in CEMS systemsQAL and standards in CEMS systems
QAL and standards in CEMS systems
 
Electronic Measurement and Instrumentation
Electronic Measurement and InstrumentationElectronic Measurement and Instrumentation
Electronic Measurement and Instrumentation
 
Unit 1 CONCEPT OF MEASUREMENT
Unit 1 CONCEPT OF MEASUREMENTUnit 1 CONCEPT OF MEASUREMENT
Unit 1 CONCEPT OF MEASUREMENT
 
Basic concepts of measurement
Basic concepts of measurementBasic concepts of measurement
Basic concepts of measurement
 
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...Testing and Troubleshooting 4-20 mA Control Loops  Presented by Fluke and Tra...
Testing and Troubleshooting 4-20 mA Control Loops Presented by Fluke and Tra...
 
1
11
1
 
Measure 2nd lec
Measure 2nd lecMeasure 2nd lec
Measure 2nd lec
 
Lecture 1
Lecture  1Lecture  1
Lecture 1
 
Unit i
Unit iUnit i
Unit i
 
UNIT 1 BASICS OF METRO LOGY
UNIT 1 BASICS OF METRO LOGYUNIT 1 BASICS OF METRO LOGY
UNIT 1 BASICS OF METRO LOGY
 
Metrology Assignment
Metrology AssignmentMetrology Assignment
Metrology Assignment
 
Tqa
TqaTqa
Tqa
 
ME 313 Mechanical Measurements and Instrumentation Lecture 01
ME 313 Mechanical Measurements and Instrumentation Lecture 01ME 313 Mechanical Measurements and Instrumentation Lecture 01
ME 313 Mechanical Measurements and Instrumentation Lecture 01
 
Yokogawa Vibration Type Density Meter
Yokogawa Vibration Type Density MeterYokogawa Vibration Type Density Meter
Yokogawa Vibration Type Density Meter
 
Equipment calibration PPT by Shravan Kumar
Equipment calibration PPT by Shravan KumarEquipment calibration PPT by Shravan Kumar
Equipment calibration PPT by Shravan Kumar
 
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
 

Similar to THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM

Instrument lesson
Instrument lessonInstrument lesson
Instrument lesson
Hung Thai
 
Basic method of measurement.pptx
Basic method of measurement.pptxBasic method of measurement.pptx
Basic method of measurement.pptx
SivaGovind2
 

Similar to THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM (20)

Instrument lesson
Instrument lessonInstrument lesson
Instrument lesson
 
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Errors During ...
Lecture Notes:  EEEC6430312 Measurements And Instrumentation - Errors During ...Lecture Notes:  EEEC6430312 Measurements And Instrumentation - Errors During ...
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Errors During ...
 
Ch 01 (Measurement Systems).pptx
Ch 01 (Measurement Systems).pptxCh 01 (Measurement Systems).pptx
Ch 01 (Measurement Systems).pptx
 
A05910106
A05910106A05910106
A05910106
 
Experiment #1 in Instrumentation And Dynamic Systems Lab : Calibration
Experiment #1 in Instrumentation And Dynamic Systems Lab : Calibration Experiment #1 in Instrumentation And Dynamic Systems Lab : Calibration
Experiment #1 in Instrumentation And Dynamic Systems Lab : Calibration
 
Smart Sensors.
Smart Sensors.Smart Sensors.
Smart Sensors.
 
Ch 01 (Measurement Systems).pptx
Ch 01 (Measurement Systems).pptxCh 01 (Measurement Systems).pptx
Ch 01 (Measurement Systems).pptx
 
Converter station water-cooled pump vibration monitoring and condition assess...
Converter station water-cooled pump vibration monitoring and condition assess...Converter station water-cooled pump vibration monitoring and condition assess...
Converter station water-cooled pump vibration monitoring and condition assess...
 
Instrumentation_ Measurment_course_Chapter_2_.pdf
Instrumentation_ Measurment_course_Chapter_2_.pdfInstrumentation_ Measurment_course_Chapter_2_.pdf
Instrumentation_ Measurment_course_Chapter_2_.pdf
 
Measurement and error
Measurement and errorMeasurement and error
Measurement and error
 
Introduction to Measurement Transducers.ppt
Introduction to Measurement Transducers.pptIntroduction to Measurement Transducers.ppt
Introduction to Measurement Transducers.ppt
 
rf based complex impedance measurement
rf based complex impedance measurementrf based complex impedance measurement
rf based complex impedance measurement
 
Prototype of multifunctional transmitter with Rejection of disturbances
Prototype of multifunctional transmitter with Rejection of disturbancesPrototype of multifunctional transmitter with Rejection of disturbances
Prototype of multifunctional transmitter with Rejection of disturbances
 
Mmm lab manual_10_mel47b
Mmm lab manual_10_mel47bMmm lab manual_10_mel47b
Mmm lab manual_10_mel47b
 
IRJET- Patient Monitoring System
IRJET- Patient Monitoring SystemIRJET- Patient Monitoring System
IRJET- Patient Monitoring System
 
EMI unit -1 introduction to measurements
EMI unit -1 introduction to measurementsEMI unit -1 introduction to measurements
EMI unit -1 introduction to measurements
 
Basic method of measurement.pptx
Basic method of measurement.pptxBasic method of measurement.pptx
Basic method of measurement.pptx
 
Chapter 1 Measurement System
Chapter 1  Measurement  SystemChapter 1  Measurement  System
Chapter 1 Measurement System
 
Flowmeter course
Flowmeter courseFlowmeter course
Flowmeter course
 
Basic concepts of instrumentation
Basic concepts of instrumentationBasic concepts of instrumentation
Basic concepts of instrumentation
 

More from ieijjournal1

Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...
Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...
Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...
ieijjournal1
 
CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)
CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)
CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)
ieijjournal1
 

More from ieijjournal1 (15)

Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...
Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...
Issues and Challenges in Broadcast Storm Suppression Algorithms of Vehicular ...
 
Low Power SI Class E Power Amplifier and Rf Switch for Health Care
Low Power SI Class E Power Amplifier and Rf Switch for Health CareLow Power SI Class E Power Amplifier and Rf Switch for Health Care
Low Power SI Class E Power Amplifier and Rf Switch for Health Care
 
Informatics Engineering, an International Journal (IEIJ)
 Informatics Engineering, an International Journal (IEIJ) Informatics Engineering, an International Journal (IEIJ)
Informatics Engineering, an International Journal (IEIJ)
 
A NEW HYBRID FOR SOFTWARE COST ESTIMATION USING PARTICLE SWARM OPTIMIZATION A...
A NEW HYBRID FOR SOFTWARE COST ESTIMATION USING PARTICLE SWARM OPTIMIZATION A...A NEW HYBRID FOR SOFTWARE COST ESTIMATION USING PARTICLE SWARM OPTIMIZATION A...
A NEW HYBRID FOR SOFTWARE COST ESTIMATION USING PARTICLE SWARM OPTIMIZATION A...
 
THE LEFT AND RIGHT BLOCK POLE PLACEMENT COMPARISON STUDY: APPLICATION TO FLIG...
THE LEFT AND RIGHT BLOCK POLE PLACEMENT COMPARISON STUDY: APPLICATION TO FLIG...THE LEFT AND RIGHT BLOCK POLE PLACEMENT COMPARISON STUDY: APPLICATION TO FLIG...
THE LEFT AND RIGHT BLOCK POLE PLACEMENT COMPARISON STUDY: APPLICATION TO FLIG...
 
A HYPER-HEURISTIC METHOD FOR SCHEDULING THEJOBS IN CLOUD ENVIRONMENT
A HYPER-HEURISTIC METHOD FOR SCHEDULING THEJOBS IN CLOUD ENVIRONMENTA HYPER-HEURISTIC METHOD FOR SCHEDULING THEJOBS IN CLOUD ENVIRONMENT
A HYPER-HEURISTIC METHOD FOR SCHEDULING THEJOBS IN CLOUD ENVIRONMENT
 
FROM FPGA TO ASIC IMPLEMENTATION OF AN OPENRISC BASED SOC FOR VOIP APPLICATION
FROM FPGA TO ASIC IMPLEMENTATION OF AN OPENRISC BASED SOC FOR VOIP APPLICATIONFROM FPGA TO ASIC IMPLEMENTATION OF AN OPENRISC BASED SOC FOR VOIP APPLICATION
FROM FPGA TO ASIC IMPLEMENTATION OF AN OPENRISC BASED SOC FOR VOIP APPLICATION
 
CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)
CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)
CLASSIFIER SELECTION MODELS FOR INTRUSION DETECTION SYSTEM (IDS)
 
ADAPTIVE MODELING OF URBAN DYNAMICS DURING EPHEMERAL EVENT VIA MOBILE PHONE T...
ADAPTIVE MODELING OF URBAN DYNAMICS DURING EPHEMERAL EVENT VIA MOBILE PHONE T...ADAPTIVE MODELING OF URBAN DYNAMICS DURING EPHEMERAL EVENT VIA MOBILE PHONE T...
ADAPTIVE MODELING OF URBAN DYNAMICS DURING EPHEMERAL EVENT VIA MOBILE PHONE T...
 
AN IMPROVED METHOD TO DETECT INTRUSION USING MACHINE LEARNING ALGORITHMS
AN IMPROVED METHOD TO DETECT INTRUSION USING MACHINE LEARNING ALGORITHMSAN IMPROVED METHOD TO DETECT INTRUSION USING MACHINE LEARNING ALGORITHMS
AN IMPROVED METHOD TO DETECT INTRUSION USING MACHINE LEARNING ALGORITHMS
 
CLASS D POWER AMPLIFIER FOR MEDICAL APPLICATION
CLASS D POWER AMPLIFIER FOR MEDICAL APPLICATIONCLASS D POWER AMPLIFIER FOR MEDICAL APPLICATION
CLASS D POWER AMPLIFIER FOR MEDICAL APPLICATION
 
RESEARCH IN BIG DATA – AN OVERVIEW
RESEARCH IN BIG DATA – AN OVERVIEWRESEARCH IN BIG DATA – AN OVERVIEW
RESEARCH IN BIG DATA – AN OVERVIEW
 
LOW POWER SI CLASS E POWER AMPLIFIER AND RF SWITCH FOR HEALTH CARE
LOW POWER SI CLASS E POWER AMPLIFIER AND RF SWITCH FOR HEALTH CARELOW POWER SI CLASS E POWER AMPLIFIER AND RF SWITCH FOR HEALTH CARE
LOW POWER SI CLASS E POWER AMPLIFIER AND RF SWITCH FOR HEALTH CARE
 
PRACTICE OF CALIBRATION TECHNOLOGY FOR THE SPECIAL TEST EQUIPMENT
PRACTICE OF CALIBRATION TECHNOLOGY FOR THE SPECIAL TEST EQUIPMENT PRACTICE OF CALIBRATION TECHNOLOGY FOR THE SPECIAL TEST EQUIPMENT
PRACTICE OF CALIBRATION TECHNOLOGY FOR THE SPECIAL TEST EQUIPMENT
 
Informatics Engineering, an International Journal (IEIJ)
Informatics Engineering, an International Journal (IEIJ) Informatics Engineering, an International Journal (IEIJ)
Informatics Engineering, an International Journal (IEIJ)
 

Recently uploaded

1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
QucHHunhnh
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
AnaAcapella
 
Seal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptxSeal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptx
negromaestrong
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
ciinovamais
 

Recently uploaded (20)

Asian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptxAsian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptx
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Understanding Accommodations and Modifications
Understanding  Accommodations and ModificationsUnderstanding  Accommodations and Modifications
Understanding Accommodations and Modifications
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
 
Magic bus Group work1and 2 (Team 3).pptx
Magic bus Group work1and 2 (Team 3).pptxMagic bus Group work1and 2 (Team 3).pptx
Magic bus Group work1and 2 (Team 3).pptx
 
Spatium Project Simulation student brief
Spatium Project Simulation student briefSpatium Project Simulation student brief
Spatium Project Simulation student brief
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
On National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan FellowsOn National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan Fellows
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
Seal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptxSeal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptx
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM

  • 1. Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016 DOI : 10.5121/ieij.2016.4201 1 THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM Yixiong Xu Mechatronic Engineering, Shanghai University of Engineering Science, Shanghai, China ABSTRACT The pressure signal calibration technology of the comprehensive test system which involved pressure sensors was studied in this paper. The melioration of pressure signal calibration methods was elaborated. Compared with the calibration methods in the lab and after analyzing the relevant problems,the calibration technology online was achieved. The test datum and reasons of measuring error analyzed,the uncertainty evaluation was given and then this calibration method was proved to be feasible and accurate. KEYWORD Calibration online; Pressure transducer; Pressure signal calibration 1. INTRODUCTION Integrated test system for pressure transducer can be achieved on liquid, gas, electric and other different media measurement and control of physical quantities, and it is mostly a complex testing system of electric hydraulic control. It are widely used in industrial control, military, aerospace and other fields. So the calibration of test system for pressure transmitter is particularly important. The improvement of calibration technology ensure system stability, accuracy and fast control of the parameters in a large extent, and provides quality assurance for research and production. 2. THE CALIBRATION OF PRESSURE SIGNALS OF THE INTEGRATED TEST SYSTEM 2.1 Pressure signal improvement of the calibration method Calibration of the pressure signal is divided into two steps. The first step is a pressure transducer calibration; Calibration of pressure signal is the second one[1] .Conventional calibration method is
  • 2. Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016 2 to test the transducer that is removed from the system in the laboratory; the transducer is reinstalled in the test system after test qualified, and then through the pressure transducer, we calibrate the secondary display device of the test system. This calibration method is cumbersome, time consuming and have an accumulation of errors. All these lead to a large measurement error. Therefore we make the following improvement of the calibration method: Firstly, the pressure transducer is achieved on-line calibration instead of laboratory calibration; Secondly, the secondary display instrument of test systems is calibrated, by accessing directly to the terminal via a standard signal acquisition or signal source waveform generator. 2.2 On-line calibration of pressure transmitter 2.2.1 Realization of environmental conditions Although pressure transducer calibration needs relatively harsh conditions, now many workshop works has requirements of environment temperature, humidity and outside magnetic field. Thus the calibration environment that temperature: 20℃±5℃, every 10min change is not more than 1℃, relative humidity: 45% to 75%, and there is no external magnetic field affecting the output stable mechanical vibration and no other effects in addition to the geomagnetic off-normal operation pressure transducer, can be achieved. 2.2.2 Build on-line calibration device Calibration device consists of digital pressure calibrators, pressure source and composition of the adapter, and the auxiliary equipment is a walkie-talkie. CST 2005 digital pressure calibrator, intelligent modules and CST 1004 high pressure pump constitute the standard pressure source, which weighs 15kg totally and is moved easily; Adapters is used to connected different thread diameters of standard pressure sensors to the source, and common diameter is 12mm adapter turn 20mm; Sometimes the control system is relatively far away from the pressure transducer on site, and control system location apart more far, so two people are needed to complete the calibration. Calibration device from the system as shown in Figure 1. Fig.1 Composition of a test system and its calibration theory
  • 3. Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016 3 2.2.3 The calibration method According to state test procedures, we finish the appearance and basic error checks. After calibration equipment and field equipment system get thermal equilibrium (2 hours), we installed and preheat the equipments, as shown in Figure 1.Because the pressure standard pressure source is difficult to maintain stable calibration points corresponding to the value in the measurement process, we use the method of multiple readings averaged to improve the accuracy of reading. Measurement error is calculated according to formula (1).If the test data meet the technical requirements, it is judged as qualified pressure transducer; Otherwise, it is considered unqualified. △d=Yd1-Yd2 (1) In the formula:△d-measurement error of the pressure transmitter calibrated point;Yd1—the value of the standard digital pressure calibrator, Pa;Yd2—pressure indication of the software interface or standard gauge display, Pa. 2.2.4 Advantages of pressure transducer calibration on-line It is very important to achieve pressure transmitter calibration on-line, which improves the efficiency of non-standard test equipment measurement ensures efficient conduct of scientific research and production. Compared to calibration in the laboratory, on-line calibration has the following advantages: firstly, through on-site demolition, on-site calibration and on-site installation, this method has shorten the calibration period and improved the calibration efficiency of the whole testing system; secondly, it is convenient to build the calibration device, not complicated; Thirdly, flexible, it is not influenced by other conditions apart from the calibration conditions. 2.3 Current signal calibration The high-precision digital ammeter is connected in series circuit which is formed by standard signal source and pressure signal acquisition end of. Digital ammeter works as a data monitoring device and gets real-time data. Test software gives the current signal via the control panel; Record the measured value and displayed value and calculate the current linearity error. Since the electrical signal is small, in order to improve measurement accuracy and reduce human error machine error, in the calibration process we should pay attention to the following matters: Firstly, in the selection of monitoring digital ammeter, its resolution ratio should be more than four and a half, the resolution should be more than four and a half, its accuracy should be more than five ten thousandths, such as FLUKE8840;Secondly, when measuring we should try to avoid some of the factors that caused the difference ,such as triboelectric effect, leakage current and piezoelectric effect.
  • 4. Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016 4 3. DATAANALYSIS 3.1 pressure transducer test data Taking 10Mpa as an example, test the effect of calibration techniques improved by actual test. Pressure transducer, measuring range: 0~10Mpa, precision: 0.5%.Test data in table 1.According to the technical requirements, The pressure transducer is determined as: Qualified. Tab.1 test datum of 10Mpa pressure sensor 3.2 Current test data The test was made on the test data of the heat compressed air signal in the test station, whose acquisition range from 4 to 20mA.The data recorded in table 2. Tab.2 The test datum of current signal Calibration points((((mA)))) Current signal Found ((((mA)))) Current signal display ((((mA)))) Error((((mA)))) 20 20.000 20.03 0.02 16 16.000 16.01 00 12 12.000 11.99 -0.01 8 8.000 7.98 -0.02 4 4.000 3.96 -0.04 3.3 Evaluation of uncertainty in measurement 3.3.1 Sources of uncertainty a) The basic standard uncertainty: uA; b) The uncertainty of the introduction of the standard pressure source: ub1; c) The measurement uncertainty introduced by the digital current meter: ub2;
  • 5. Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016 5 3.3.2 Evaluation of uncertainty a) Basic standard type A evaluation of uncertainty Since there are more test data, repeat the test for the pressure transducer 5Mpa profile and carry on the uncertainty analysis; The test data record as is shown in Table 3. Tab.4 repetitive test datum of 5Mpa of pressure sensor Its average value: 10 1 5.0176i i x x n = = =∑ Mpa Standard type A evaluation of uncertainty calculated according to the formula(2): 2 1 ( ) ( ) ( 1) n i i i s x x x n = = − −∑ 10 2 10 1 ( ) ( ) (10 1) 0.00004548 9 0.0022i i s x x x = = − − = =∑ Mpa (2) A class of uncertainty that measurement repeatability leads to is as follows: uA= ( )10s x / n =0.0007 Mpa (3) b)The uncertainty of the introduction of the standard pressure source:ub1 The main source is the intelligent pressure digital module of the standard pressure calibration instrument and the error of the digital pressure calibration instrument, while the influence of environmental temperature and humidity can be ignored. Type B evaluation of uncertainty. Standard pressure calibration instrument 15Mpa pressure module allows error limit a1=0.003Mpa. Its distribution is subject to uniform distribution, and take the factor k= 3 , calculated according to the formula(4): uc1= a1/k=0.001732 Mpa (4) c) The measurement uncertainty introduced by the digital current meter: Its main source is the uncertainty of the current introduction of the digital current meter in the measurement of the pressure signal acquisition port and the standard signal source, and its type B evaluation of uncertainty. The error limit of the digital meter 100mA measurement range is 0.01%, And a2=0.01mA, take the coverage factor k= 3 , calculated according to the formula (5) :
  • 6. Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016 6 A2/k =0.00577mA (5) Converted into pressure by linear conversion: uc2=0.00540Mpa (6) d) Combined standard uncertainty Uc Above all are independent and unrelated, and the combined standard uncertainty Uc is calculated according to the formula (7) : Uc = 2 2 2 1 2A b bu u u+ + =0.00571Mpa (7) e) Evaluation of expanded uncertainty All the sources of uncertainty are uniform distribution, and take k=2, according to the formula (8): U =2* Uc=0.0114Mpa (8) f) The measuring result of the 5Mpa file of the pressure transducer is: (5.0176±0.0114)Mpa. 3.3.3 Methods for reducing uncertainty of measurement Through the analysis of the measurement uncertainty, we can see that the measurement uncertainty u introduced by the digital current meter is relatively larger, the uncertainty of u introduced by the standard pressure source is relatively moderate. , and the basic standard uncertainty is relatively smaller. In order to improve the accuracy of measurement and reduce the uncertainty of measurement, we have the following improvements: firstly, choosing high precision digital ammeter can effectively reduce the measurement uncertainty introduced by the digital current meter and the uncertainty of the introduction of the standard pressure source, reducing system error; Thirdly, the A/D acquisition card could be modified to ensure consistency between data acquisition and actual values. 4. SUMMARY The containing pressure transducer test system calibration techniques were studied, focusing on analyses and improvements of the system testing pressure signal calibration method, and through the test data processing and analysis, it proves the feasibility and superiority of this calibration method. It improves the calibration efficiency of pressure transmitter test system and the stability and reliability of the test system in the test products, and has accumulated some calibration experience for this type of test system.
  • 7. Informatics Engineering, an International Journal (IEIJ), Vol.4, No.2, June 2016 7 REFERENCES [1] Shanghai Institute of Measurement and Testing technology. JJG 882-2004 pressure transducer calibration specification [s]. China measurement Publishing House, 2004 [2] China Metrology Association. Data processing and measurement professional practice [m]. Beijing: China's quality watchdog, 2013 [3] Sheng Hong Jiang, Feng Yi.Online Calibration of the Sensor [j]. Electronic Design Engineering, 2012,43 (1): 34-37. [4] Tian Hui, Subaoming. The Measurement and Calibration of the Pressure Sensor System [j]. Electronic Component & Device Applications, 2008 (09): 44-46 Author Yixiong Xu(1985-),postgraduate in Shanghai University of Engineering Science(SUES), the main research interest now is Instrumentation measurement technology, address: Room 1105,Build ing NO.243,Lane 777,Dushi road, Minhang district, Shanghai(201109), fixed 021-34628108, mobile 13681897190,email:heroexu@126.com