ELEMENTAL ANALYSES USING EXTERNAL
STANDARDS
By Cin-Ty A. Lee
Basic principle of external calibration
There are many different ways to determine the concentration of an element in a rock
sample. The approach here is to measure the signal intensity of an element in an
unknown solution (sample) and reference it to the signal intensity of the same element in
another solution, whose elemental concentrations are known. The solution whose
concentration is known is called an external standard. For example, the concentration
of an element in the sample solution is determined bySol
SAC
ST
Sol
ST
SA
Sol
SA
I
C
IC = (1)
where is the measured intensity in counts per second (cps) of the element in the
sample solution, is the intensity in the standard, and C is the known concentration
in the standard.
SAI
STI Sol
ST
Drift and the Internal Standard
In the ideal world, Eq. 1 suffices. However, in the real world, there are some
complications. First, one never measures the sample at the same time the standard is
measured. During the elapsed time between standard and sample measurements, it is
possible that the sensitivity of the instrument has drifted (sensitivity is defined as the
number of cps obtained for a given concentration unit, e.g. cps/ppm; for the Element 2,
the sensitivity of 115
In is ~1-2 x 109
cps/ppm). Drift can be caused by many factors. For
our instrument, drift is usually caused by buildup of precipitates in the nebulizer, torch,
and/or on the sampler cone. A second complication is that there may be matrix
differences between the standard and the sample. The matrix is defined loosely as a
property which describes the bulk composition of the solution, e.g., total dissolved solids,
acidity, salinity, organic molecule content, etc. Variations in these matrices can give rise
to variations in ionization efficiency; thus, that the signal of a given trace element in two
solutions having identical concentrations of that element but different matrices could be
different. In an ideal world, it would be best to match the matrices of the standard and
sample. In practice, it is not possible to exactly match standard and sample matrices
because in general we do not know the composition of the unknown sample! Matrix
effects therefore represent another form of drift. Before applying Eq. 1, one must first
correct all measured signals for drift – instrumental and matrix induced.
We correct for drift using the following principle. Let us assume that we spiked our
sample and standard solutions with exactly the same concentration of a given element
(not the particular element whose concentrations we are trying to determine!), which we
call the internal standard (IS). If there are no instrumental- and matrix-induced drifts,
the signal intensities of the IS in the sample and standard solutions should be
identical. If there is drift, it follows that we should be able to correct for the drift. Thus,
if we wish to drift-correct our sample intensities to our standard intensities, we multiply
the intensity of the element of interest in our sample, , by a drift correction factor:
SAIS STIS
SAI






SA
ST
SA
IS
IS
I (2a)
In practice, we are often faced with the situation where we did not spike each solution
with the identical amount of IS but we know exactly how much we put in to each
solution. We can account for this by modifying Eq. 2a:






IS
ST
IS
SA
SA
ST
SA
C
C
IS
IS
I (2b)
where C is the concentration of the IS in the sample and in the standard.IS
SA
IS
STC
The element mass that is chosen for an internal standard is based on the following
criteria:
1) The element mass chosen for an IS must be absent or present in negligible
quantities in the sample and standard. Thus, we could choose an enriched
isotope of an element or we can choose an element whose natural
concentrations are extremely low. The former can be very expensive. In our
lab, we typically use Indium (115
In). The natural abundances of In in most
rocks is very small, and in addition, we spike at an amount that swamps the
natural abundance signal (but remains comparable to the concentrations of the
elements of interest).
2) The IS element must have negligible isobaric interferences.
3) The IS element should remain stable in solution for long periods of time.
Blank correction
Another complication in determining elemental concentrations in an unknown is
the possibility for contamination. When using an external rock standard (see below),
one always processes a procedural blank through the same chemistry that the rock
standard and sample unknowns undergo. The contamination it picks up represents the
integrated contamination from the various dissolution steps. The assumption is that the
procedural blank represents a constant background contamination that affects standards
and samples identically. Therefore, in order to correct for the procedural blank, one must
subtract out the intensities of the procedural blank. This is, of course, done after all
intensities have been drift-corrected:






−





= IS
ST
IS
BL
BL
ST
BLIS
ST
IS
SA
SA
ST
SASA
C
C
IS
IS
I
C
C
IS
IS
II*
(3)
where is the blank-corrected and drift-corrected intensity of the element of interest
in the sample, is the intensity of the IS element in the blank, and C is the
concentration of the IS element in the blank.
*
SAI
BLIS IS
BL
When using a synthetic standard (see below), we do not apply the blank
correction at this stage. This is because the synthetic standard did not go through the
same procedure as the samples. It is also assumed that the synthetic standard has little or
no blank problems (relative to the signal intensities of the synthetic standard, samples,
etc.). The procedural blank concentrations are determined by treating the procedural
blank as a sample and then subtracting the former from the latter at the very end of the
data reduction protocol.
External standard normalization
There are two types of external standards used in this laboratory: a certified rock
standard and a synthetic standard. The certified rock standard is a rock powder whose
concentrations have been independently determined by many different laboratories. A
synthetic standard is one that is prepared in our laboratory by mixing of certified single-
element (and sometimes multi-element) solution standards purchased from various
companies that make standards (e.g., Alfa-Aesar Johnson Matthey, NIST, Spex, etc.).
In general, we prefer to use certified rock standards to determine elemental
concentrations in an unknown sample because it is much easier to match the matrix of
our samples with a certified rock standard than a synthetic standard. The external rock
standard is taken through the exact same process as all sample unknowns. An external
rock standard only works if complete dissolution is achieved. In some cases, the matrix
of the unknown sample may be so different from all of the rock standards in our
laboratory. In such cases, we may use a synthetic standard to determine concentrations.
In all cases, one or more quality standards are analyzed in order to check the
calibration. Quality standards are other certified rock standards.
Dilution effects
In general, the total amount of rock powder dissolved for the external standard
and sample unknowns are not identical. In addition the total weight of the diluted
standards and samples are also not identical. Therefore, appropriate corrections must be
made. The actual concentration of the element of interest in the sample solution, , is
given by:
Sol
SAC
ST
Sol
ST
SA
Sol
SA
I
C
IC *
= (4)
where C is the actual concentration of the element in the standard solution and is given
by
Sol
ST
Sol
ST
Rock
ST
Rock
STSol
ST
M
MC
C = (5)
where C is the certified concentration in the rock standard, is the mass of the
certified rock standard dissolved, and is the final diluted weight of the standard
solution.
Rock
ST
Rock
STM
Sol
STM
Similarly, in order to obtain the concentration in the sample powder, one needs to
correct for the different amounts of sample powder dissolved and the total diluted
solution weight. Equation 5 then becomes
Rock
SA
Sol
SA
ST
Sol
ST
SA
Rock
SA
M
M
I
C
IC *
= (6a)
Rock
SA
Sol
SA
Sol
ST
Rock
ST
ST
Rock
ST
SA
Rock
SA
M
M
M
M
I
C
IC *
= (6b)
Multiple calibration points
In the above discussion, we have calibrated the sensitivity of our instrument using
one external standard, e.g., we have a one-point calibration. To increase the accuracy of
our results, we can generate a calibration curve using more than one point. For example,
we can run two of the same rock standards, but at different dilutions. We can then do a
linear regression through the data and empirically generate the best-fit calibration line.
Treating the intensity of the standards as the y-variable and the concentration of the
standards as the x-variable, we generate the following linear regressions y = mx + b (or y
= mx if we force the regression through zero). The concentration of an element can then
be determined by inverting the linear regression, e.g. x = (y-b)/m. In Eq. 6b, the first two
terms on the right hand side would be replaced by
mbISA /)( *
− (7)
Limit of detection and filtering of bad data
There are many ways to define the limit of detection. The most widely used
definition in the ICP-MS community is the 3-sigma definition,
S
LOD Blankσ3
= (8)
where σBlank represents 1 standard deviation of the signal intensities of the procedural
blank and S is the sensitivity of the instrument. The sensitivity is the signal intensity
obtained for a given concentration of a particular element in solution. Signal intensities
above the LOD are considered to be resolvable from the blank. Those below the LOD are
rejected.

More Related Content

Similar to Icpm sdatareduction

Research Experience
Research ExperienceResearch Experience
Research Experiencealicextt
 
Mass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPUMass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPUsiraj174
 
Trace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniquesTrace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniquesAwad Albalwi
 
Standardization and calibration -dr.mallik
Standardization and calibration -dr.mallikStandardization and calibration -dr.mallik
Standardization and calibration -dr.mallikDr. Mallikarjunaswamy C
 
Mass spectroscopy for MSc I Chemistry of SPPU
Mass spectroscopy for MSc I Chemistry of SPPUMass spectroscopy for MSc I Chemistry of SPPU
Mass spectroscopy for MSc I Chemistry of SPPUsiraj174
 
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...PerkinElmer, Inc.
 
Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Salum Mkata
 
Basic chemistry concepts
Basic chemistry conceptsBasic chemistry concepts
Basic chemistry conceptsECRD2015
 
Basic Chemistry Concepts
Basic Chemistry ConceptsBasic Chemistry Concepts
Basic Chemistry ConceptsECRD IN
 
Chapter 5
Chapter 5Chapter 5
Chapter 5ECRD IN
 
Introduction Quantitative X-Ray Diffraction Methods
Introduction Quantitative X-Ray Diffraction MethodsIntroduction Quantitative X-Ray Diffraction Methods
Introduction Quantitative X-Ray Diffraction MethodsJosé da Silva Rabelo Neto
 
Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...
Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...
Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...KBI Biopharma
 
Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy
Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy
Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy Grace Jeanpierre
 
Lab 4 Photosynthesis
Lab 4 PhotosynthesisLab 4 Photosynthesis
Lab 4 Photosynthesisericavanetten
 
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...Dr. Suman Pattanayak
 

Similar to Icpm sdatareduction (20)

Laserablationicp
LaserablationicpLaserablationicp
Laserablationicp
 
Research Experience
Research ExperienceResearch Experience
Research Experience
 
Mass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPUMass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPU
 
Trace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniquesTrace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniques
 
Standardization and calibration -dr.mallik
Standardization and calibration -dr.mallikStandardization and calibration -dr.mallik
Standardization and calibration -dr.mallik
 
Mass spectroscopy for MSc I Chemistry of SPPU
Mass spectroscopy for MSc I Chemistry of SPPUMass spectroscopy for MSc I Chemistry of SPPU
Mass spectroscopy for MSc I Chemistry of SPPU
 
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
 
Kinetics
KineticsKinetics
Kinetics
 
vortrag070704.ppt
vortrag070704.pptvortrag070704.ppt
vortrag070704.ppt
 
Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Uv vis spectroscopy practical.
Uv vis spectroscopy practical.
 
Home exam answers
Home exam answersHome exam answers
Home exam answers
 
Basic chemistry concepts
Basic chemistry conceptsBasic chemistry concepts
Basic chemistry concepts
 
Basic Chemistry Concepts
Basic Chemistry ConceptsBasic Chemistry Concepts
Basic Chemistry Concepts
 
Chapter 5
Chapter 5Chapter 5
Chapter 5
 
Introduction Quantitative X-Ray Diffraction Methods
Introduction Quantitative X-Ray Diffraction MethodsIntroduction Quantitative X-Ray Diffraction Methods
Introduction Quantitative X-Ray Diffraction Methods
 
Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...
Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...
Analysis of Aggregation, Stability, and Lot Comparability by Sedimentation Ve...
 
Mass Spesc 101
Mass Spesc 101Mass Spesc 101
Mass Spesc 101
 
Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy
Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy
Exploration of Sn- doped Ga2O3 using X-Ray Absorption Spectroscopy
 
Lab 4 Photosynthesis
Lab 4 PhotosynthesisLab 4 Photosynthesis
Lab 4 Photosynthesis
 
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
 

More from Calama, Antofagasta (15)

Sillitoe2010 copper
Sillitoe2010 copperSillitoe2010 copper
Sillitoe2010 copper
 
No ions
No ionsNo ions
No ions
 
Exeter tr caspiche_270309
Exeter tr caspiche_270309Exeter tr caspiche_270309
Exeter tr caspiche_270309
 
Exeter tr caspiche_191009
Exeter tr caspiche_191009Exeter tr caspiche_191009
Exeter tr caspiche_191009
 
Exeter tr caspiche_140909
Exeter tr caspiche_140909Exeter tr caspiche_140909
Exeter tr caspiche_140909
 
Exeter tr caspiche_130910
Exeter tr caspiche_130910Exeter tr caspiche_130910
Exeter tr caspiche_130910
 
Exeter tr caspiche_120310
Exeter tr caspiche_120310Exeter tr caspiche_120310
Exeter tr caspiche_120310
 
Exeter tr caspiche_120116
Exeter tr caspiche_120116Exeter tr caspiche_120116
Exeter tr caspiche_120116
 
Exeter tr caspiche_090309
Exeter tr caspiche_090309Exeter tr caspiche_090309
Exeter tr caspiche_090309
 
Exeter tr caspiche_080426
Exeter tr caspiche_080426Exeter tr caspiche_080426
Exeter tr caspiche_080426
 
Exeter technical report_caspiche_dec_2007
Exeter technical report_caspiche_dec_2007Exeter technical report_caspiche_dec_2007
Exeter technical report_caspiche_dec_2007
 
Dilles hqgr0164
Dilles hqgr0164Dilles hqgr0164
Dilles hqgr0164
 
Caspiche discovery history_paper
Caspiche discovery history_paperCaspiche discovery history_paper
Caspiche discovery history_paper
 
805 1379-1-pb
805 1379-1-pb805 1379-1-pb
805 1379-1-pb
 
314 318-1-pb
314 318-1-pb314 318-1-pb
314 318-1-pb
 

Recently uploaded

“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...Marc Dusseiller Dusjagr
 
MENTAL STATUS EXAMINATION format.docx
MENTAL     STATUS EXAMINATION format.docxMENTAL     STATUS EXAMINATION format.docx
MENTAL STATUS EXAMINATION format.docxPoojaSen20
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactdawncurless
 
Micromeritics - Fundamental and Derived Properties of Powders
Micromeritics - Fundamental and Derived Properties of PowdersMicromeritics - Fundamental and Derived Properties of Powders
Micromeritics - Fundamental and Derived Properties of PowdersChitralekhaTherkar
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991RKavithamani
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docxPoojaSen20
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformChameera Dedduwage
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Sapana Sha
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxContemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxRoyAbrique
 
Science 7 - LAND and SEA BREEZE and its Characteristics
Science 7 - LAND and SEA BREEZE and its CharacteristicsScience 7 - LAND and SEA BREEZE and its Characteristics
Science 7 - LAND and SEA BREEZE and its CharacteristicsKarinaGenton
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13Steve Thomason
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 
PSYCHIATRIC History collection FORMAT.pptx
PSYCHIATRIC   History collection FORMAT.pptxPSYCHIATRIC   History collection FORMAT.pptx
PSYCHIATRIC History collection FORMAT.pptxPoojaSen20
 
Introduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher EducationIntroduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher Educationpboyjonauth
 

Recently uploaded (20)

“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
 
MENTAL STATUS EXAMINATION format.docx
MENTAL     STATUS EXAMINATION format.docxMENTAL     STATUS EXAMINATION format.docx
MENTAL STATUS EXAMINATION format.docx
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 
Micromeritics - Fundamental and Derived Properties of Powders
Micromeritics - Fundamental and Derived Properties of PowdersMicromeritics - Fundamental and Derived Properties of Powders
Micromeritics - Fundamental and Derived Properties of Powders
 
Staff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSDStaff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSD
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docx
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy Reform
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxContemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
 
Science 7 - LAND and SEA BREEZE and its Characteristics
Science 7 - LAND and SEA BREEZE and its CharacteristicsScience 7 - LAND and SEA BREEZE and its Characteristics
Science 7 - LAND and SEA BREEZE and its Characteristics
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
PSYCHIATRIC History collection FORMAT.pptx
PSYCHIATRIC   History collection FORMAT.pptxPSYCHIATRIC   History collection FORMAT.pptx
PSYCHIATRIC History collection FORMAT.pptx
 
Introduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher EducationIntroduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher Education
 

Icpm sdatareduction

  • 1. ELEMENTAL ANALYSES USING EXTERNAL STANDARDS By Cin-Ty A. Lee Basic principle of external calibration There are many different ways to determine the concentration of an element in a rock sample. The approach here is to measure the signal intensity of an element in an unknown solution (sample) and reference it to the signal intensity of the same element in another solution, whose elemental concentrations are known. The solution whose concentration is known is called an external standard. For example, the concentration of an element in the sample solution is determined bySol SAC ST Sol ST SA Sol SA I C IC = (1) where is the measured intensity in counts per second (cps) of the element in the sample solution, is the intensity in the standard, and C is the known concentration in the standard. SAI STI Sol ST Drift and the Internal Standard In the ideal world, Eq. 1 suffices. However, in the real world, there are some complications. First, one never measures the sample at the same time the standard is measured. During the elapsed time between standard and sample measurements, it is possible that the sensitivity of the instrument has drifted (sensitivity is defined as the number of cps obtained for a given concentration unit, e.g. cps/ppm; for the Element 2, the sensitivity of 115 In is ~1-2 x 109 cps/ppm). Drift can be caused by many factors. For our instrument, drift is usually caused by buildup of precipitates in the nebulizer, torch, and/or on the sampler cone. A second complication is that there may be matrix differences between the standard and the sample. The matrix is defined loosely as a property which describes the bulk composition of the solution, e.g., total dissolved solids, acidity, salinity, organic molecule content, etc. Variations in these matrices can give rise to variations in ionization efficiency; thus, that the signal of a given trace element in two solutions having identical concentrations of that element but different matrices could be different. In an ideal world, it would be best to match the matrices of the standard and sample. In practice, it is not possible to exactly match standard and sample matrices because in general we do not know the composition of the unknown sample! Matrix effects therefore represent another form of drift. Before applying Eq. 1, one must first correct all measured signals for drift – instrumental and matrix induced. We correct for drift using the following principle. Let us assume that we spiked our sample and standard solutions with exactly the same concentration of a given element (not the particular element whose concentrations we are trying to determine!), which we call the internal standard (IS). If there are no instrumental- and matrix-induced drifts,
  • 2. the signal intensities of the IS in the sample and standard solutions should be identical. If there is drift, it follows that we should be able to correct for the drift. Thus, if we wish to drift-correct our sample intensities to our standard intensities, we multiply the intensity of the element of interest in our sample, , by a drift correction factor: SAIS STIS SAI       SA ST SA IS IS I (2a) In practice, we are often faced with the situation where we did not spike each solution with the identical amount of IS but we know exactly how much we put in to each solution. We can account for this by modifying Eq. 2a:       IS ST IS SA SA ST SA C C IS IS I (2b) where C is the concentration of the IS in the sample and in the standard.IS SA IS STC The element mass that is chosen for an internal standard is based on the following criteria: 1) The element mass chosen for an IS must be absent or present in negligible quantities in the sample and standard. Thus, we could choose an enriched isotope of an element or we can choose an element whose natural concentrations are extremely low. The former can be very expensive. In our lab, we typically use Indium (115 In). The natural abundances of In in most rocks is very small, and in addition, we spike at an amount that swamps the natural abundance signal (but remains comparable to the concentrations of the elements of interest). 2) The IS element must have negligible isobaric interferences. 3) The IS element should remain stable in solution for long periods of time. Blank correction Another complication in determining elemental concentrations in an unknown is the possibility for contamination. When using an external rock standard (see below), one always processes a procedural blank through the same chemistry that the rock standard and sample unknowns undergo. The contamination it picks up represents the integrated contamination from the various dissolution steps. The assumption is that the procedural blank represents a constant background contamination that affects standards and samples identically. Therefore, in order to correct for the procedural blank, one must subtract out the intensities of the procedural blank. This is, of course, done after all intensities have been drift-corrected:       −      = IS ST IS BL BL ST BLIS ST IS SA SA ST SASA C C IS IS I C C IS IS II* (3) where is the blank-corrected and drift-corrected intensity of the element of interest in the sample, is the intensity of the IS element in the blank, and C is the concentration of the IS element in the blank. * SAI BLIS IS BL When using a synthetic standard (see below), we do not apply the blank correction at this stage. This is because the synthetic standard did not go through the
  • 3. same procedure as the samples. It is also assumed that the synthetic standard has little or no blank problems (relative to the signal intensities of the synthetic standard, samples, etc.). The procedural blank concentrations are determined by treating the procedural blank as a sample and then subtracting the former from the latter at the very end of the data reduction protocol. External standard normalization There are two types of external standards used in this laboratory: a certified rock standard and a synthetic standard. The certified rock standard is a rock powder whose concentrations have been independently determined by many different laboratories. A synthetic standard is one that is prepared in our laboratory by mixing of certified single- element (and sometimes multi-element) solution standards purchased from various companies that make standards (e.g., Alfa-Aesar Johnson Matthey, NIST, Spex, etc.). In general, we prefer to use certified rock standards to determine elemental concentrations in an unknown sample because it is much easier to match the matrix of our samples with a certified rock standard than a synthetic standard. The external rock standard is taken through the exact same process as all sample unknowns. An external rock standard only works if complete dissolution is achieved. In some cases, the matrix of the unknown sample may be so different from all of the rock standards in our laboratory. In such cases, we may use a synthetic standard to determine concentrations. In all cases, one or more quality standards are analyzed in order to check the calibration. Quality standards are other certified rock standards. Dilution effects In general, the total amount of rock powder dissolved for the external standard and sample unknowns are not identical. In addition the total weight of the diluted standards and samples are also not identical. Therefore, appropriate corrections must be made. The actual concentration of the element of interest in the sample solution, , is given by: Sol SAC ST Sol ST SA Sol SA I C IC * = (4) where C is the actual concentration of the element in the standard solution and is given by Sol ST Sol ST Rock ST Rock STSol ST M MC C = (5) where C is the certified concentration in the rock standard, is the mass of the certified rock standard dissolved, and is the final diluted weight of the standard solution. Rock ST Rock STM Sol STM Similarly, in order to obtain the concentration in the sample powder, one needs to correct for the different amounts of sample powder dissolved and the total diluted solution weight. Equation 5 then becomes Rock SA Sol SA ST Sol ST SA Rock SA M M I C IC * = (6a)
  • 4. Rock SA Sol SA Sol ST Rock ST ST Rock ST SA Rock SA M M M M I C IC * = (6b) Multiple calibration points In the above discussion, we have calibrated the sensitivity of our instrument using one external standard, e.g., we have a one-point calibration. To increase the accuracy of our results, we can generate a calibration curve using more than one point. For example, we can run two of the same rock standards, but at different dilutions. We can then do a linear regression through the data and empirically generate the best-fit calibration line. Treating the intensity of the standards as the y-variable and the concentration of the standards as the x-variable, we generate the following linear regressions y = mx + b (or y = mx if we force the regression through zero). The concentration of an element can then be determined by inverting the linear regression, e.g. x = (y-b)/m. In Eq. 6b, the first two terms on the right hand side would be replaced by mbISA /)( * − (7) Limit of detection and filtering of bad data There are many ways to define the limit of detection. The most widely used definition in the ICP-MS community is the 3-sigma definition, S LOD Blankσ3 = (8) where σBlank represents 1 standard deviation of the signal intensities of the procedural blank and S is the sensitivity of the instrument. The sensitivity is the signal intensity obtained for a given concentration of a particular element in solution. Signal intensities above the LOD are considered to be resolvable from the blank. Those below the LOD are rejected.