CALsys 650 Autocal is Dry Block temperature calibrator for calibrating contact type temperature sensor and provide automatic solution for temperature measurement with auto temp step function, Live temperature data and recording, Automatic Test certification generation facility ISO Traceable. Can take 4 sensor input at a time.
2. CALsys 650 AUTOCAL
• Easy–to-read color 5 Inch LCD Display
with perfect overview of actual Temp
calibrator status.
• Intuitive, Fast and User-Friendly
navigation.
• 4 Channel Calibration (4 No's Easy to
use Universal input connector suitable
for thermocouple and Rtd).
• Internal CJC Compensation.
• Ethernet (LAN) Communication with
CALsys 650 AUTOCAL Model for
PC/Laptop Interface.
• USB Connector for Data saving
(Optional)
3. CALsys 650 AUTOCAL
Sensor Connection for CALsys 650 AUTOCAL : The Calibration system provides
calibration up to four channels i.e. one master and three test sensors . We use high quality
universal LEMO connector i.e. suitable both for T/C and Rtd.
Connection for MASTER Sensor
Connection for First TEST Sensor
Connection for Second TEST Sensor
Connection for Third TEST Sensor
5. USER INTERFACE
Home Screen: In this screen user can select sensor configuration (for selection of type of
sensors), mode of operation (auto / manual) and data transfer (file transfer). This window
also shows the ongoing process.
6. USER INTERFACE
Sensor Configuration: In this screen user select the sensors (TC / RTD) for calibration with
their serial number and temperature unit (C/F/K).
7. USER INTERFACE
Working : CALsys 650 AUTOCAL have two operating modes i.e. Manual and Auto mode.
Manual Mode: (Process is taken at single calibration point)
a. Define the type of Thermocouple or RTD, calibration point and type of unit (oC/F/K).
b. System will take the calibration bath to predefined calibration temperature point.
c. System waits for the furnace/bath to be stabilized and then takes reading of master and
test sensors.
Auto Mode: (Process is taken at five calibration point)
a. Define the Serial Number, type of Thermocouple or RTD, temperature unit, calibration points.
b. System will take the calibration bath to predefined calibration temperature points.
c. System waits for the furnace/bath to be stabilized and then take reading of master sensor and
test sensors.
d. The system takes the furnace to the next temperature point.
Step c is repeated until all points are completed.
8. USER INTERFACE
Manual Mode: In this screen user set the temp. Point for calibration and on clicking start
process button the process of calibration starts.
9. USER INTERFACE
Auto Mode: In this screen user sets the temperature Points for calibration (Max 5 Points)
10. USER INTERFACE
Data Transfer: In this screen the files are transferred to pc for report generation
11. PC INTERFACE
• CALsys 650 AUTOCAL shall be connected to PC/Laptop through Ethernet (LAN) or USB
Connector (Optional).
• Tempsens will provide Customized Software tool AUTOCAL (on CD ROM ) along with CALsys
650 AUTOCAL Calibrator for access recorded data both for Manual mode and Auto Mode.
• Saved data for specific set point will be saved as .xls file in PC side both for manual mode (One
set point) and Automode ( Max 5 Set points).
Ethernet (LAN) Connection for interfacing
with PC/Laptop
12. PC INTERFACE
SOFTWARE : Tempsens make Easy to use Customized software enables end user to access
temperature data both for Manual mode and Automode.
13. DATA SAVING
Manual Mode : Temperature Data will be saved for one specific set point in .Xls format
Auto Mode : Temperature Data will be saved for specific set points (Max 5 Points) in .Xls
format.
14. AUTOMATIC CALIBRATION REPORT
Tempsens can offer customized data saving option both for manual and Automode.
After completion manual/Automode automatic calibration report can be generated at PC
side based on pre define format.