3. Motivation and Purpose
Understand how stimulation strategies
and electrode spacing relate to the
shape of voltage distributions and
electric fields in order to inform and
improve future innovations
3
4. Voltage and Electric Field
𝐸 = −𝛻𝑉
𝐸𝑥, 𝐸𝑦, 𝐸𝑧 = −
𝑑𝑉
𝑑𝑥
,
𝑑𝑉
𝑑𝑦
,
𝑑𝑉
𝑑𝑧
The strength of the electric field must take into
account all components of the electric field
𝑬
𝐸𝑥 𝑖 = −dV/dx
𝐸𝑦 𝑗 = −dV/dy
+x
+y
4
5. Voltage/Electric Field and
Activating Function (AF)
The activating function is the second derivative
of voltage in the direction of the neuron fiber
(Rattay 1986, Litvak 2007, Falcone 2011)
V-pp
Neuron
AF =
𝑑2 𝑉𝑝𝑝
𝑑𝑥2
+x
+y
5
6. More on Activating Function
If the electrodes are simplified to point
electrodes, the activating function can be
modeled by the following equation, for a
given point electrode at point (u,v,w):
Using this equation, one can obtain the discrete AF in space by
plugging in a discrete spatial voltage distribution for each
electrode and summing the AF’s
𝐴𝐹 𝑥, 𝑦, 𝑧 = 𝑉 𝑥, 𝑦, 𝑧
3𝑧
2((𝑥 − 𝑢)2+(𝑦 − 𝑣)2+(𝑧 − 𝑤)2)2
−
1
(𝑥 − 𝑢)2+(𝑦 − 𝑣)2+(𝑧 − 𝑤)2
6
7. Derivation of AF using
equations from Falcone 2011:
Equation 1: 𝐴𝐹 𝑥, 𝑦, 𝑧, 𝐼 =
ρ𝐼
4π
3𝑧
2(𝑥2+𝑦2+𝑧2)5/2 −
1
(𝑥2+𝑦2+𝑧2)3/2
Equation 2: 𝑉 𝑥, 𝑦, 𝑧, 𝐼 =
ρ𝐼
4π 𝑥2+𝑦2+𝑧2
Equation 1 can be factored to obtain:
𝐴𝐹 𝑥, 𝑦, 𝑧, 𝐼 =
ρ𝐼
4π 𝑥2 + 𝑦2 + 𝑧2
3𝑧
2(𝑥2 + 𝑦2 + 𝑧2)2
−
1
𝑥2 + 𝑦2 + 𝑧2
The factored expression can be replaced by V(x,y,z,I) from Equation 2:
𝐴𝐹 𝑥, 𝑦, 𝑧 = 𝑉 𝑥, 𝑦, 𝑧
3𝑧
2(𝑥2 + 𝑦2 + 𝑧2)2
−
1
𝑥2 + 𝑦2 + 𝑧2
Where ρ is resisitivity, 𝐼 is current, and x, y, and z define the location of
a point relative to a point electrode at (0,0,0)
7
13. Data Acquisition Software
Previous acquisition software
WinCommander to manually control the microprobe
stepper motor
LabVIEW software to record voltage measurements
Phantom to automate mouse clicking of
WinCommander controls and record-voltage push
buttons in LabVIEW
New acquisition software uses only one
LabVIEW program that automates stepper
motor movement and voltage recording
Enables automated 1D, 2D and 3D scans
Saves time, easier to use
13
16. Data Processing for 2D XZ
Scans
% Initialize Vpp before running the script. Vpp contains the scan of
Vpp
% values. The script rearranges the Vpp data ready to be plotted in 3
% Space. It also gives the x and z components of the electric field
% Initialize the matrix that will contain the Vpp data. It will be
called
% "VppGrid"
x = 0:.02:4;
z=0:.05:.5;
[X,Z] = meshgrid(x,z);
% Insert Vpp data into VppGrid
VppGrid = X';
VppGrid(1:end) = Vpp(1:end);
VppGrid = VppGrid';
% Take the gradient to find the X and Z components of the electric
field
[Ex,Ez] = gradient(VppGrid,X',Z);
Ex = -1*Ex;
Ez = -1*Ez;
16
20. Data Processing for 3D
Scans
x = xLim(1):xStep:xLim(2);
y = yLim(1):yStep:yLim(2);
z = zLim(1):zStep:zLim(2);
[X,Y,Z] = meshgrid(x,y,z);
gridDimensions = size(X);
VppGrid =
zeros(gridDimensions(2),gridDimensions(1),gridDimensions(3));
VppGrid(1:end) = Vpp(1:end);
VppGrid = permute(VppGrid, [2,1,3]);
[Ex,Ey,Ez] = gradient(VppGrid, xStep, yStep, zStep);
Ex = -1*Ex;
Ez = -1*Ez;
Ey = -1*Ey;
20
21. Data Processing Cont’d
Peak-to-peak voltages corresponding to
biphasic pulses beginning with the
cathodic phase were assigned as “+”
Anodic phase-leading signals were
assigned as “-”
+Vpp -Vpp
21
*All current pulses in the experiments have pulse widths of .5ms at 120 Hz
26. Voltage Distribution, Electric
Fields, and AF Shapes
The voltage, electric field, and activating function follow the
same shape given a stimulation pattern
Since the E Field and AF follow the same shape as the
Vpp distribution, we can use any one of the 3 to
visualize the shape of the stimulation signal
26
27. Voltage Sum
0
0.005
0.01
0.015
0.02
0.025
-2 -1 0 1 2
Vpp(V)
X Distance (mm)
Monopolar Vpp Distribution
Vpp at Z = 100 um
E6: 50 uA, Cathodic
Pulse First
E11: 0 uA
0
0.005
0.01
0.015
0.02
0.025
-2 -1 0 1 2
Vpp(V)
X Distance (mm)
Monopolar Vpp Distribution
Vpp at Z = 100um
E6: 0 uA
E11: 50 uA, Cathodic
Pulse First
0.015
0.02
0.025
0.03
0.035
0.04
0.045
-2 -1 0 1 2
Vpp(V)
X Distance (mm)
Sum of Monopolar Vpp Distributions
Vpp at Z = 100 um
E6: 50uA, Cathodic Pulse
First
E11: 50uA, Cathodic Pulse
First
27
28. 0.015
0.02
0.025
0.03
0.035
0.04
0.045
0.05
-2 -1.5 -1 -0.5 0 0.5 1 1.5 2
Vpp(V)
X Distance (mm)
Calculated (Summed) Monopolar Vpp Distribution
Vpp at Z = 100 um
E6: 50uA, Cathodic Pulse First
E11: 50uA, Cathodic Pulse First
28
29. 0.015
0.02
0.025
0.03
0.035
0.04
0.045
0.05
-2 -1.5 -1 -0.5 0 0.5 1 1.5 2
Vpp(V)
X Distance (mm)
Measured Monopolar Vpp Distribution
Vpp at Z = 100um
E6, 50uA, Cathodic Pulse First
E11, 50 uA, Cathodic Pulse First
29
30. Summing Voltages
When multiple electrodes are on, the resulting
voltage distribution in space is equal to the sum of
the voltages coming from each electrode
0.015
0.02
0.025
0.03
0.035
0.04
0.045
0.05
-2 -1 0 1 2
Vpp(V)
X Distance (mm)
Measured Monopolar Vpp
Distribution
0.015
0.02
0.025
0.03
0.035
0.04
0.045
0.05
-2 -1 0 1 2
Vpp(V)
X Distance (mm)
Calculated Monopolar Vpp
Distribution
30Raw Data
38. Relative Vpp Peak Widths
Mean Stdev (n = 3)
Monopolar 1 0
Tripolar 0.25 0.028
Pentapolar 0.41 0.031
Relative Peak Width in X Direction
Mean Stdev (n = 3)
Monopolar 1 0
Tripolar 0.33 0.036
Pentapolar 0.37 0.053
Relative Peak Width in Y Direction
38Spreadsheet Data
39. Relative Vpp Peak Heights
Mean Relative
Height Stdev (n = 3)
Monopolar (100 uA) 1 0
Monopolar (50 uA) 0.51 0.0057
Monopolar (20 uA) 0.20 0.0093
Tripolar (50 uA ↓, 100 uA ↑, 50 uA ↓) 0.20 0.0077
Pentapolar (100 uA ↑, 25 uA ↓ 4x) 0.20 0.0093
Tripolar (100 uA ↓, 200 uA ↑, 100 uA ↓) 0.43 0.0048
Pentapolar (200 uA ↑, 50 uA ↓ 4x) 0.48 0.014
39Spreadsheet Data
40. Spread and Peak
Comments
Tripolar and Pentapolar stimulation focus current
in the X and Y directions
Tripolar focuses the current better than Pentapolar
in the X direction
Tripolar and Pentapolar have similar current
spread in the Y direction, but these results may be
due to experimental error
Tripolar and Pentapolar have peaks that are ~20%
of the monopolar
40
46. Steering Strategy #1
Comments
The E Field and AF follow the same
shape as Vpp
None of the signals peak in the space
between the electrodes given the test
conditions
The signal can peak in the middle if the
electrodes are close enough
46
47. Steering Strategy #2
120 Hz, .5 ms PW
Middle Electrode: 100uA Amplitude
Left Electrode: 75uA Amplitude
Right Electrode: 25uA Amplitude
+x
+y
47
48. Steering Strategy #2
48Spreadsheet Data
-0.01
-0.005
0
0.005
0.01
0.015
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
Vpp(V)
X Distance (mm)
Asymetric Tripolar Vpp Curve
Asymetric Tripolar Vpp, Probe Height = 50um
E6 25uA Anodic Pulse First
E11 100uA Cathodic Pulse First
E15 25uA Anodic Pulse First
0.015
0.02
0.025
0.03
0.035
0.04
0.045
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
Vpp(V)
X Distance (mm)
Monopolar Vpp Curve
Asymetric Tripolar Vpp, Probe Height = 50um
E6 0uA
E11 100uA Cathodic Pulse First
E15 0uA
Peak occurs at x = .04 mm
Peak occurs at x = -0.02 mm
50. Steering Strategy #3
0
0.005
0.01
0.015
0.02
0.025
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
Vpp(V)
X Distance (mm)
Monopolar Vpp Distribution Vpp at Z = 100um
E6: 0 uA
E11: 50 uA, Cathodic Pulse First
-0.006
-0.004
-0.002
0
0.002
0.004
0.006
0.008
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
Vpp(V)
X Distance (mm)
Observed Bipolar Vpp Distribution Vpp at Z = 100um
E6, 50uA, Anodic Pulse First
E11, 50 uA, Cathodic Pulse
First+Sheet1!$A$3
Peak occurs at
x = .08 mm
Peak occurs at
x = .01 mm
50Spreadsheet Data
51. Current Steering Comments
Current is not necessarily “steered”
additively. This is possible only when the
electrodes are close (i.e. Strategy 1)
May be possible through subtractive
strategies (i.e. Strategies 2 and 3)
Limited current steering effects
Unclear how much the peak height is affected
The extent to which the current is steered for
strategies 2 and 3 is inconclusive without
repeated experiments
51
53. Peak Fusion
At various heights above the array,
determine the spacing between two
electrodes at which the peaks fuse
50
uA
50
uA
L
120 Hz, .5 ms PW
53
54. Peak Fusion
At various heights above the array,
determine the spacing between two
electrodes at which the peaks fuse
V-pp 1 V-pp 2
L
54
55. Peak Fusion
At various heights above the array,
determine the spacing between two
electrodes at which the peaks fuse
V-pp 1 V-pp 2
L
55
56. Peak Fusion
At various heights above the array,
determine the spacing between two
electrodes at which the peaks fuse
Critical L
V-pp 1 + 2
56
57. “Critical L” vs. Height Above
Array
y = 0.0028x + 0.0587
R² = 0.9897
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
0 100 200 300 400 500 600 700
ElectrodeCenter-to-CenterSpacingatWhichPeaksFuse
(mm)
Z Distance Away from Electrode Surface (um)
50 uA
100 uA
150 uA
200 uA
Linear (100 uA)
57Spreadsheet Data
60. Distinguishable Peaks
At various heights above the array, determine
the spacing between two electrodes at which
the signals become two peaks
L
V-pp 1 + 2
60
61. Distinguishable Peaks
At various heights above the array, determine
the spacing between two electrodes at which
the signals become two peaks
V-pp 1 V-pp 2
L
61
62. Distinguishable Peaks
At various heights above the array, determine
the spacing between two electrodes at which
the signals become two peaks
V-pp 1 V-pp 2
Critical L
h1
h2
ℎ1 − ℎ2 ≥ 2𝑚𝑉
62
64. Critical L for Distinguishable
Peaks vs. Z Distance
64
y = 8.04x + 0.248
R² = 0.9977
0
0.5
1
1.5
2
2.5
0 0.05 0.1 0.15 0.2 0.25 0.3
CriticalL(mm)
Z Distance Above Electrode Surface (mm)
50 uA
Linear (50 uA)
Spreadsheet Data
65. Critical L for Distinguishable
Peaks vs. Z Distance
y = 4.98x + 0.2403
R² = 0.994
0
0.5
1
1.5
2
2.5
0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4
CriticalL(mm)
Z Distance Above Electrode Surface (mm)
100 uA
Linear (100 uA)
65Spreadsheet Data
66. Critical L for Distinguishable
Peaks vs. Z Distance
66
y = 4.8971x + 0.2898
R² = 0.9933
0
0.5
1
1.5
2
2.5
0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4
CriticalL(mm)
Z Distance Above Electrode Surface (mm)
200 uA
Linear (200 uA)
Spreadsheet Data
67. Fused Critical L’s and
Distinguishable Critical L’s
67
y = 0.0028x + 0.0587
R² = 0.9897
y = 0.008x + 0.248
R² = 0.9977
y = 0.0051x + 0.185
R² = 0.9912
0
0.5
1
1.5
2
2.5
0 100 200 300 400 500 600 700
ElectrodeCenter-toCentersSpacingat(mm)
Z Distance Away from Electrode Surface (um)
50 uA Fused
100 uA Fused
150 uA Fused
200 uA Fused
50 uA Distinguishable
100 uA Distinguishable
200 uA Distinguishable
Linear (100 uA Fused)
Linear (50 uA Distinguishable)
Linear (100 uA Distinguishable)
Fused Region
Distinguishable
Peaks Region
Spreadsheet Data
In-Between
Fused and
Distinguishable
68. Tripolar Focusing Electrode
Spacing
Determine the relationship between electrode spacing and
peak width in tripolar focusing
Determine the relationship between electrode spacing and
peak height in tripolar focusing
Peak heights and widths are normalized to the Voltage
distribution of monopolar current pulses
50
uA
100
uA
50
uA 120 Hz, .5 ms PW
68
72. Tripolar Focusing Electrode
Spacing: Peak Height vs. L
y = -0.0849x2 + 0.3423x - 0.0175
R² = 0.9995
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6
NormalizedPeakHeight
Electrode Center to Center Spacing (mm)
Series1
Poly. (Series1)
72Spreadsheet Data Raw Data
73. Tripolar Focusing Electrode
Spacing: Peak Width vs. L
y = -0.063x2 + 0.2526x + 0.0232
R² = 0.9936
0
0.05
0.1
0.15
0.2
0.25
0.3
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6
NormalizedPeakWidth
Electrode Center to Center Spacing (mm)
Series1
Poly. (Series1)
73Spreadsheet Data Raw Data
74. Summary of Results
Tripolar stimulation focuses the current better than
pentapolar and monopolar
All multipolar focusing requires more current
Current steering is not possible with the given array
design, but can be achieved if electrodes are close
enough
Current steering could be achieved subtractively, but
the extent is not clear
Electrode spacing is linearly related to the peak
separation when two electrodes are stimulated
Electrode spacing is linearly related to peak width
Electrode spacing has a polynomial relationship with
the tipolar-focused Vpp peak
74
76. Limitations and Sources of
Error
AF oversimplifies electrode geometry
Overestimates the steepness and sharpness of AF
distributions
Arbitrarily assigning +/- Vpp values creates noise
through human error
Qualitative nature of determining “peak fusion”
Surface/array tilt
Inherent noise within each scan especially when
measured Vpp < 10mV
Quality of micro-probe tip
Variability across measurements that aren’t conducted
on the same day
Varying concentration of PBS
76
77. Future Studies
Repeated asymmetric current steering
trials
Other stimulation patterns
Running all future experiments with
higher stimulation amplitudes
Computational tissue models
Approximate the direction of neuron
fibers so the second derivative of
voltage can be taken to find AF
77
Editor's Notes
Include file path for raw data and graphs
Show this slide very briefly
Maybe include disk electrodes to show scale relative to return electrode
The probe tip was changed during the last week of July 2016. Thus, all experiments either used the old probe tip (before the last week of July 2016) or the new probe tip (after the last week of July 2016)
Briefly show this
To achieve a better approximation of the activating function, one would need to integrate over the areas that each stimulating electrode occupies
This graph was obtained from measurements made by the old probe tip
This graph was obtained from measurements made by the old probe tip
This graph was obtained from measurements made by the old probe tip
These graphs were obtained from measurements made by the new probe tip
delete
delete
These graphs were obtained from measurements made by the new probe tip
In the next three slides, talk about hypotheses: Expected that the peak heights would be similar to control 1 but worse than control 3 and better than control 2, and that the peak widths would be better than control group 1 – need to compare to other control groups?
Remove the frequency and PW information – just mention it once at the beginning that all experiments use these
Peak width is a measure of how focused the signal is
Monopolar amplitude = 50uA
These results were obtained from measurements made by the new probe tip
These results were obtained from measurements made by the new probe tip
Future experiments should look at the peak spread of Tripolar and Pentapolar with 200uA center electrodes
Make the pentapolar stimulation strategy format the same as the tripolar
First bullet: Monopolar group has an amplitude of 50uA
These results were obtained from measurements made by the old probe tip
Units of E Field are in 10^-3 N/C
These results were obtained from measurements made by the old probe tip
These results were obtained from measurements made by the old probe tip
These results were obtained from measurements made by the new probe tip
These results were obtained from measurements made by the new probe tip
First bullet: How “close” exactly will be discussed in later slides for electrode spacing
Stimulation strategy for both electrodes: 50uA, Cathodic Pulse first, .5ms PW, 120 Hz
NOTE: This model cannot be applied to the AF because the AF simplifies the electrode geometry to single points
These results were obtained from measurements made by the new probe tip
These results were obtained from measurements made by the old probe tip
These results were obtained from measurements made by the old probe tip
Stimulation strategy for both electrodes: 50uA, Cathodic Pulse first, .5ms PW, 120 Hz
These results were obtained from measurements made by the new probe tip
These results were obtained from measurements made by the new probe tip
These results were obtained from measurements made by the new probe tip
These results were obtained from measurements made by the new probe tip
Data normalized to monopolar stimulation receiving 100uA amplitude pulses, .5ms PW, and 120 Hz
These results were obtained from measurements made by the new probe tip
These results were obtained from measurements made by the new probe tip