SlideShare a Scribd company logo
1 of 1
Download to read offline
RESEARCH POSTER PRESENTATION DESIGN © 2012
www.PosterPresentations.com
• In resource limited settings, the use of
DBS for HIV-1 and Hepatitis diagnosis
has become essential for optimal
disease management.
• Several studies have evaluated the
performance of Whatman 903 device
with the RT assay.
• The objective of this study was to
compare the performance of the
HemaSpot and Whatman 903 devices
using the RT assay.
• In addition, this study evaluated the
performance of the devices with the
m2000sp automation.
Conclusion
• Third HCV World Health Organization
(WHO) International Standard (IS)
was used to generate whole blood
(ProMedDx (Norton, Ma) panels
targeting the following
concentrations: 5000IU/mL,
2500IU/mL, 1250IU/mL and
625IU/mL.
• Seventy five microliters of whole
blood was spotted on Whatman 903
and HemaSpot devices and air dried
overnight at room temperature.
• Panels were placed at 8C overnight
for long term storage.
• One full spot from each device was
incubated in 1.3mL of m2000
proprietary elution buffer for 45min.
• Samples were gently vortexed and
transferred to m2000sp for
extraction. RNA extraction was
performed using 1ml extraction
protocol and the performance of the
RT assay was assessed.
• Intra and Inter-run reproducibility
was evaluated across 3 different
runs.
• The enhanced mPLUS capability provides the clinical laboratories
with increased efficiencies to meet the increasingly stringent
turnaround time requirements without increased costs associated
with discarding partially used amplification reagents.
• This solution offers a combination of automation, flexibility,
efficiency and test menu unmatched in the field today.
Evaluation of novel sample collection device using the Abbott RealTime HCV assay (RT)
Danijela Lucic Ph.D.1, Christine Herman1, Jeanette Hill2, Shelley Hossenlopp2, Gavin Cloherty1
1Abbott Molecular , Des Plaines, IL; 2Spot On Sciences, Austin, TX
Introduction
• Whatman device demonstrated
100% detectability at 5000IU/mL
(n=12), 90% detectability at
2500IU/mL (n=30) and 1250IU/mL
(n=20) and only 70% detectability at
625IU/mL (n=29).
• HemaSpot device demonstrated
100% detectability at 5000IU/mL
(n=15), 1250IU/mL (n=30) and
625IU/mL while 97% detectability
was observed at 2500IU/mL (n=30).
• Intra-run reproducibility did not
exceed 0.30 log copies/mL at any
level (Table 1).
• Inter-run reproducibility did not
exceed 0.33 log copies/mL at any
level (Table 2).
Results Results Continued
Figure 1: HemaSpot device
Figure 2: HCV detectability at 5000, 2500, 1250 and 625 IU/mL.
Table 1a: Whatman Intra-run Reproducibility
Table 1b: HemaSpot Intra-run Reproducibility
Table 2: Inter-run Reproducibility
Methods

More Related Content

Similar to Abbott HCV-HemaSpot final 0424

ACMG_2016_MicroFISH_Poster
ACMG_2016_MicroFISH_PosterACMG_2016_MicroFISH_Poster
ACMG_2016_MicroFISH_PosterDavid Wright
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labAnkit Raiyani
 
chlamydia nucleic acid amplification tests
chlamydia nucleic acid amplification testschlamydia nucleic acid amplification tests
chlamydia nucleic acid amplification testsmorteza karami
 
V1c 97 03 Hstm 2011 Hamburg 20110526 Dj
V1c 97 03 Hstm 2011 Hamburg 20110526 DjV1c 97 03 Hstm 2011 Hamburg 20110526 Dj
V1c 97 03 Hstm 2011 Hamburg 20110526 Djdennisjapink
 
Molecular Diagnostics.pptx
Molecular Diagnostics.pptxMolecular Diagnostics.pptx
Molecular Diagnostics.pptxAlisha Shaikh
 
Nuclear Medicine Procedures-3
Nuclear Medicine Procedures-3Nuclear Medicine Procedures-3
Nuclear Medicine Procedures-3Ashraf Zytoon
 
Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...
Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...
Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...Wisit Cheungpasitporn
 
Newer methods in diagnosis of tuberculosis in children
Newer methods in diagnosis of tuberculosis in childrenNewer methods in diagnosis of tuberculosis in children
Newer methods in diagnosis of tuberculosis in childrenDr Naveen kumar
 
MSD Poster Tau CTAD_MARS_REVIEW
MSD Poster Tau CTAD_MARS_REVIEWMSD Poster Tau CTAD_MARS_REVIEW
MSD Poster Tau CTAD_MARS_REVIEWLawrence Hwang
 
AUTOMATION IN SEROLOGY.pptx educational
AUTOMATION IN SEROLOGY.pptx  educationalAUTOMATION IN SEROLOGY.pptx  educational
AUTOMATION IN SEROLOGY.pptx educationalaryajayakottarathil
 
KIPER_SCTRP_Presentation
KIPER_SCTRP_PresentationKIPER_SCTRP_Presentation
KIPER_SCTRP_PresentationKeturah Kiper
 
Quanterix Long Sales Presentation_SK
Quanterix Long Sales Presentation_SKQuanterix Long Sales Presentation_SK
Quanterix Long Sales Presentation_SKSteve Klose
 
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03krishgen
 
Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?
Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?
Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?asclepiuspdfs
 
A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...
A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...
A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...InsideScientific
 

Similar to Abbott HCV-HemaSpot final 0424 (20)

ACMG_2016_MicroFISH_Poster
ACMG_2016_MicroFISH_PosterACMG_2016_MicroFISH_Poster
ACMG_2016_MicroFISH_Poster
 
JOURNAL CLUB PRESENTATION (20L81S0710-PA)
JOURNAL CLUB PRESENTATION (20L81S0710-PA)JOURNAL CLUB PRESENTATION (20L81S0710-PA)
JOURNAL CLUB PRESENTATION (20L81S0710-PA)
 
Nat Testing
Nat TestingNat Testing
Nat Testing
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
 
chlamydia nucleic acid amplification tests
chlamydia nucleic acid amplification testschlamydia nucleic acid amplification tests
chlamydia nucleic acid amplification tests
 
V1c 97 03 Hstm 2011 Hamburg 20110526 Dj
V1c 97 03 Hstm 2011 Hamburg 20110526 DjV1c 97 03 Hstm 2011 Hamburg 20110526 Dj
V1c 97 03 Hstm 2011 Hamburg 20110526 Dj
 
Molecular Diagnostics.pptx
Molecular Diagnostics.pptxMolecular Diagnostics.pptx
Molecular Diagnostics.pptx
 
Nuclear Medicine Procedures-3
Nuclear Medicine Procedures-3Nuclear Medicine Procedures-3
Nuclear Medicine Procedures-3
 
Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...
Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...
Normal Saline is not Normal? ; Chloride liberal vs. Chloride restrictive IV F...
 
Newer methods in diagnosis of tuberculosis in children
Newer methods in diagnosis of tuberculosis in childrenNewer methods in diagnosis of tuberculosis in children
Newer methods in diagnosis of tuberculosis in children
 
SLAS Screen Design and Assay Technology Special Interest Group SLAS2017 Prese...
SLAS Screen Design and Assay Technology Special Interest Group SLAS2017 Prese...SLAS Screen Design and Assay Technology Special Interest Group SLAS2017 Prese...
SLAS Screen Design and Assay Technology Special Interest Group SLAS2017 Prese...
 
MSD Poster Tau CTAD_MARS_REVIEW
MSD Poster Tau CTAD_MARS_REVIEWMSD Poster Tau CTAD_MARS_REVIEW
MSD Poster Tau CTAD_MARS_REVIEW
 
AUTOMATION IN SEROLOGY.pptx educational
AUTOMATION IN SEROLOGY.pptx  educationalAUTOMATION IN SEROLOGY.pptx  educational
AUTOMATION IN SEROLOGY.pptx educational
 
KIPER_SCTRP_Presentation
KIPER_SCTRP_PresentationKIPER_SCTRP_Presentation
KIPER_SCTRP_Presentation
 
Quanterix Long Sales Presentation_SK
Quanterix Long Sales Presentation_SKQuanterix Long Sales Presentation_SK
Quanterix Long Sales Presentation_SK
 
Journal club pptx
Journal club pptxJournal club pptx
Journal club pptx
 
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
 
Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?
Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?
Apoptosis Post Microwave Ablation of the Liver: Does it Change with Power?
 
A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...
A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...
A Comprehensive How-To Demonstration of Higher Throughput Excitation-Contract...
 
Axt microarrays
Axt microarraysAxt microarrays
Axt microarrays
 

Abbott HCV-HemaSpot final 0424

  • 1. RESEARCH POSTER PRESENTATION DESIGN © 2012 www.PosterPresentations.com • In resource limited settings, the use of DBS for HIV-1 and Hepatitis diagnosis has become essential for optimal disease management. • Several studies have evaluated the performance of Whatman 903 device with the RT assay. • The objective of this study was to compare the performance of the HemaSpot and Whatman 903 devices using the RT assay. • In addition, this study evaluated the performance of the devices with the m2000sp automation. Conclusion • Third HCV World Health Organization (WHO) International Standard (IS) was used to generate whole blood (ProMedDx (Norton, Ma) panels targeting the following concentrations: 5000IU/mL, 2500IU/mL, 1250IU/mL and 625IU/mL. • Seventy five microliters of whole blood was spotted on Whatman 903 and HemaSpot devices and air dried overnight at room temperature. • Panels were placed at 8C overnight for long term storage. • One full spot from each device was incubated in 1.3mL of m2000 proprietary elution buffer for 45min. • Samples were gently vortexed and transferred to m2000sp for extraction. RNA extraction was performed using 1ml extraction protocol and the performance of the RT assay was assessed. • Intra and Inter-run reproducibility was evaluated across 3 different runs. • The enhanced mPLUS capability provides the clinical laboratories with increased efficiencies to meet the increasingly stringent turnaround time requirements without increased costs associated with discarding partially used amplification reagents. • This solution offers a combination of automation, flexibility, efficiency and test menu unmatched in the field today. Evaluation of novel sample collection device using the Abbott RealTime HCV assay (RT) Danijela Lucic Ph.D.1, Christine Herman1, Jeanette Hill2, Shelley Hossenlopp2, Gavin Cloherty1 1Abbott Molecular , Des Plaines, IL; 2Spot On Sciences, Austin, TX Introduction • Whatman device demonstrated 100% detectability at 5000IU/mL (n=12), 90% detectability at 2500IU/mL (n=30) and 1250IU/mL (n=20) and only 70% detectability at 625IU/mL (n=29). • HemaSpot device demonstrated 100% detectability at 5000IU/mL (n=15), 1250IU/mL (n=30) and 625IU/mL while 97% detectability was observed at 2500IU/mL (n=30). • Intra-run reproducibility did not exceed 0.30 log copies/mL at any level (Table 1). • Inter-run reproducibility did not exceed 0.33 log copies/mL at any level (Table 2). Results Results Continued Figure 1: HemaSpot device Figure 2: HCV detectability at 5000, 2500, 1250 and 625 IU/mL. Table 1a: Whatman Intra-run Reproducibility Table 1b: HemaSpot Intra-run Reproducibility Table 2: Inter-run Reproducibility Methods