DIFFERENT
CORONARY STENT
DESIGN
DR AWADHESH KUMAR SHARMA
DEPARTMENT OF CARDIOLOGY
PGIMER & DR RML HOSPITAL,NEW DELHI
BACKGROUND
 The introduction of angioplasty led to the
development of a completely new approach to treat
CAD.
 Until 1994, the percutaneous transluminal coronary
angioplasty (PTCA) was the alone treatment for
coronary artery disease.
 However, the incidence of restenosis of coronary
arteries was an important problem, necessitating
repeated interventional procedures in 30% of
patients treated with PTCA alone.
NEJM 1994;331:489-95
 Coronary artery stents were developed to provide a
metal scaffolding for the angio-plastied vessel, in an
attempt to limit negative re-modelling.
 Sigwart et al first reported the efficacy of stents in
reducing restenosis rates in 1987.
 By 1994, the Food and Drug Administration (FDA)
had approved two stents (Gianturco-Roubin stent
and the Palmaz- SchatzTM stent).
Journal of Invasiv Cardiology.2001;13:634-639
GIANTURCO-ROUBIN II
 Flat wire coil attached
to a single longitudinal
strut
 316 L stainless steel
 The first coronary stent
approved by the FDA in
June 1993.
J. clinical pathology.2005,aug;58(8):795-804
PALMAZ-SCHATZ
 Balloon expandable;
slotted tube
 316 L stainless steel
J. clinical pathology.2005,aug;58(8):795-804
 The wide acceptance of coronary stenting was based on
the results of the BElgian NEtherlands STENT
(BENESTENT) and the STent REStenosis Study
(STRESS) trials, which showed the superiority of
stenting over balloon angioplasty.
 After the wide acceptance of coronary stents the primary
concern of stent development is the need to reduce
device profiles and to increase flexibility to facilitate safe
delivery.
N Engl J Med,1994;331:489-95.
STENT SELECTION
STENT SELECTION-PERFORMANCE &
EASE OF USE
STENT SELECTION-GOOD ACUTE
ANGIOGRAPHIC RESULTS
JACC 2003;41:1283-1288
TYPES OF STENTS
 Mechanism of expansion (self-expanding or balloon-
expandable)
 Materials (stainless steel, cobalt-based alloy, tantalum,
nitinol, Pt,Ir,Cr, inert coating, biodegradable)
 Forms (sheet, wire or tube)
 Manufacturing methods (laser cut, water-jet cutting, photo-
etching)
 Geometrical configurations/design (mesh structure, coil,
slotted tube, ring, multi-design)
 Addition to stent (grafts, radio-opaque markers, coatings)
Min Invas Ther & Allied Technol
2002;11:137-47.
STENT GEOMETRIC DESIGN
 MECHANISM OF EXPANSION
 Balloon-expandable stents
 The stent is pre-mounted on a balloon and the inflation
of the balloon plastically expands the stent with respect
to the balloon diameter.
 Self-expanding stents- The smart material auto expands
to a calculated size.
Journal of invasive cardiology 1995;7:127-
134
MATERIALS
 Corrosion resistance
 Biocompatibility
 Adequately radio-opaque
 Create minimal artifacts during MRI
STENT PLATFORMS
STENT MATERIALS- NON DEGRADABLE MATERIAL
 316L stainless steel-
 Excellent mechanical properties and corrosion
resistance
 Ferromagnetic nature and low density make it a
non-MRI compatible
 Poorly visible fluoroscopic material
 First generation DESs, Cypher (sirolimus-eluting
stent, Cordis, Warren, NJ) and Taxus (paclitaxel-
eluting stent, Boston Scientific, Natick, MA)
JACC 1996;27:53
CO-CR
 Superior radial strength and improved radiopacity
 Thinner stent struts
 The second generation DES, Xience V (everolimus-
eluting stent, Abott Vascular, CA) and Endeavor
(zotarolimus-eluting stent,Medtronic Vascular,
Santa Rosa, CA).
JACC 1996;27:53
TA- TANTALUM
 Excellent corrosion resistant material
 Coated on 316L SS to improve corrosion properties
and biocompatibility
 High density and non-ferromagnetic properties
 Fluoroscopically visible and MRI compatible
 Higher rates of recoil- poor mechanical properties
JACC 1996;27:53
TI
 Excellent biocompatibility and corrosion resistance
 Low tensile strength and ductility
 Ti alloys in combination with Ni-Ti
 Ti-nitride oxide coating on 316L SS
JACC 1996;27:53
NI-TI
 Good biocompatibility, radial force and shape
memory
 Coated by some materials such as polyurethane, Ti
nitride and polycrystalline oxides to improve the
corrosion resistance
 Inadequate visibility under fluoroscopy
American J of cardiology.2008;86:1073-
1079
PT-IR
 Pt-Ir alloy of 90% platinum and 10% iridium
 Excellent radiopacity and a reduction in both
thrombosis and neointimal proliferation with less
inflammatory reactions
 Recoiling percentage was much higher (16%) than
the 316L SS stents
Journal of invasive cardiology 1995;7:127-
134
BIODEGRADABLE METALLIC MATERIALS
 Pure Fe
 Oxidation of Fe into ferrous and ferric irons
 Mg alloys
 There are two Mg alloys, AE2153 and WE4357,
used for making stents
 Radiolucent
Biomaterials.2006;27:1728-1734
RATIONAL FOR BIODEGRADABLE STENTS
Metal stent drawbacks
 Cause permanent physical
irritation
 Risk of long term endothelial
dysfunction and chronic
inflammation
 Metal have thrombogenic
properties
 Inability for the vessel to
restore its a normal
physiology
Biodegradable stent advantages
 May eliminate early and late
complications of bare-metal
stents
 Restore the vasoreactivity
 Allow a gradual transfer of the
mechanical load to the vessel
 Higher capacity for drug
incorporation and complex
release kinetics
The need for a permanent prosthesis decreases
dramatically 6 months post-implantation
STENT DESIGN
 On the basis of design, stents can be divided into
three groups: coil, tubular mesh, and slotted tube.
 Coil stents are characterised by metallic wires or
strips formed into a circular coil shape
 Tubular mesh stents consist of wires wound
together in a meshwork, forming a tube.
 Slotted tube stents are made from tubes of metal
from which a stent design is laser cut.
Eur Heart J 1997;18:1536–47
COIL VS. TUBE
 Coil design had greater strut width with gaps and
fewer or no connections between struts
 The strut width is greater; there are gap between
struts, and no connections between struts which
give it more flexibility.
 However, the design lack radial strength, and the
wide gap allow tissues to dangle.
Singapore Medical Journal, 2004.
COIL VS TUBE
 As a result, coil design has become obsolete and
replace by the more superior in radial strength, the
tube design.
 In tubular, there are two type of specification, a
slotted tube and modular tube.
Singapore Medical Journal, 2004.
SLOTTED TUBE VS. MODULAR (TUBULAR)
MODULAR DESIGN
SLOTTED TUBE VS. MODULAR (TUBULAR)
 Slotted tube stents resisted restenosis more than
the modular stents (22.1% vs 25.2%)
 Slotted tube- Closed cell design, and open cell
design
CLOSED CELL
 Sequential ring construction
 All Internal inflection points of the structural
members are connected by bridging elements.
 Regular peak-to-peak connections.
 Optimal scaffolding and a uniform surface,
regardless of the degree of bending.
 Less flexible than a similar open-cell design.
Ann Ist Super Sanita 2007;43,no1:89-100
CYPHER STENT BY CORDIS
OPEN CELL
 Some or all the internal inflection points of the
structural members are not connected by bridging
elements.
 Periodic peak-to-peak connections, peak-to-valley
connections, and mid-strut to mid strut connections
 The unconnected structural elements contribute to
longitudinal flexibility.
Ann Ist Super Sanita 2007;43,no1:89-100
OPEN CELL DESIGN
STENT DESIGN IMPACTS DRUG
DELIEVERY
LENGTH & DIAMETER OF STENT
 Long vs. Short
 Stent length is associated with restenosis rate and
clinical events (mainly target lesion revascularization)
 Short stent has lower cases of restenosis than long
stent.
 Wide vs. Narrow
 The wide diameter stent is more favorable than the
narrow one
European Heart Journal 2001;22:1585-
1593
NUMBER OF STRUTS
 More struts vs. less
 Less struts induce less chance of restenosis
compare to more struts.
THIN STRUT VS THICK STRUT
STRUT THICKNESS
 Although the immediate stent performance may be
improved by increasing strut thickness (which increases
radiovisibility, radial strength and arterial wall support)
excessive strut thickness, on the other hand, may impart
more vascular injury, trigger more intimal hyperplasia,
and engender a higher risk for restenosis than thinner
struts.
 Strut thickness was observed to be an independent
predictor of in-stent restenosis
ISAR STEREO study(Circulation 2001;103:2816-21)
ISAR-STEREO-2 trial(J Am Coll Cardiol 2003;41:1283-8.)
 In an effort to further reduce strut thickness while
maintaining adequate radiovisibility and radial
strength, novel metallic materials such as cobalt-
chromium alloy are being used for the production of
stent.
THICK VS. THIN STRUTS
 The stents with thinner struts is preferred for the
design of new stents as they can reduce
angiographic and clinical restenosis more than
those with thicker struts
ISAR-STEREO and ISAR-STEREO 2
trials
SQUARE VS. ROUND STRUT CROSS-SECTION
 The round strut cross-section without corners or
sharp edges is popular at present
 Round strut cross-section area is ideal for
smoothness design.
 Square strut cross-section area in not recommend
because it interferes with blood flow due to their
sharp edge which can slice blood cells.
Kluwer Academic Publishers 2012
SQUARE VS. ROUND STRUT CROSS-SECTION
ROUGH VS. SMOOTH SURFACE
 Smoothness of a stent can affect the performance and
biocompatibily of the stent.
 Smooth surface can reduce thrombus adhesion and
neointimal growth.
 To obtain smoothness, the stent need to be treated with
acid-pickling and then electrochemical polishing.
 The process removes slag which includes depositions
and burrs, formed on the surface of stents due to the
laser cutting production process. Seminars in interventional
cardiology1998;3:139-144
ELEMENT OF STENT DESIGN- BALLOON
OVERHANG
DRUG DELIVERY VEHICLES – COATING
POLYMER- DRUG CARRIERS IN DESS
 Nonbiodegradable and biodegradable polymers
 Non biodegradable polymers
 First and the second generation of DESs
 The first generation of DES
 Cypher - polyethylene-co-vinyl acetate (PEVA)/poly-n-butyl methacrylate
(PBMA)
 Taxus - polystyrene-b-isobutylene-b-styrene (SIBS)
 The second generation of DES
 Xience V – fluoropolymer
 Endeavor - phosphorylcholine (PC)
Eurointervention,2005;1:266-272
 Biodegradable polymers
 Polylactic acid (PLA)
 Polyglycolic acid (PGA)
 Polylactic-co-glycolic acid (PLGA)
 NON POLYMER
 Titanium–nitric oxide alloy
 Microporous stainless steel stent (Yukon, Translumina, Germany)
 A nanoporous hydroxyapetite (a biocompatible crystalline derivative of
calcium phosphate) coating
 Magnetic nanoparticles (MNPs)
Eurointervention,2005;1:266-272
YUKON Choice DES system: Translumina modified stent surface containing micropores to
enable the adsorption of different organic substances.
Abizaid A , and Costa J R Circ Cardiovasc Interv
2010;3:384-393
Copyright © American Heart Association
THERAPEUTIC AGENTS
 Sirolimus (Rapamycin)
 A macrocyclic lactone
 Inhibits the migration and proliferation of SMCs
 Zotarolimus
 The sirolimus analogues
 Developed by Abbott laboratories
 Extremely lipophilic property and low water solubility
 Everolimus
 Sirolimus analogue
 Immunosuppressive agent
 Absorbs to local tissue more rapidly and has a longer celluar residence time and activity
 Biolimus
PACLITAXEL AND ITS ANALOGUES
 Paclitaxel
 Promoting tubulin polymerization and cell cycle arrest
 Inhibiting the migration and proliferation of SMCs
 Coroxane
 Nanoparticle albumin bound paclitaxel (nab-paclitaxel)
 To improve the solubility
 Docetaxel
 Semi-synthetic analogue
 Better anti-proliferative properties
OTHERS
 Tacrolimus
 Pimecrolimus
 Curcumin
 Resveratrol
 CD 34 antibody
 Anti-VEGF
RADIO-OPACITY ENHANCEMENTS
 Stainless steel or nitinol - hard to see
fluoroscopically
 To improve X-ray visibility, markers are often
attached to the stents.
 These additions are typically made from gold,
platinum or tantalum
 Electroplating (with gold) is also being used to
enhance X-ray visibility
COATINGS
 To increase biocompatibility
 Heparin was one of the first. Its mode of action is to
reduce the coagulation cascade (and thus possibly the
thrombogenic risk) after the deployment of a stent.
 Phosphorylcoline and silicon-carbide have been used in
order to reduce platelet activation and interaction, thus
possibly controlling their adhesion to the stent struts
during the acute phase of stent re-endothelization.
 Passive coverage has been also shown to be
useful.
 Indeed, covered stents have been created, in which
a PTFE layer was put between two stents (Jostent
graft, Jomed) or one stent was covered by a inner
and an outer layer of PTFE (Symbiot, Boston
Scientific)
COMMONLY USED CORONARY
STENTS IN CLINICAL PRACTICE
XIENCE FAMILY OF STENTS
Stent Manufactu
rer
Drug Base Form/Desi
gn
Polymer Diameter Length
XIENCE
Xpedition
Abott
vascular
FDA
Approved
Everolimus
100μg/cm2
L-605 CoCr Hybrid cell
Multilink
0.0032" strut
thickness,
laser cut
PBMA
Non erodible
SV-2.25
MV-
2.5,2.75,3.0,3.
25,3.5,4.0
LL
2.5,2.75,3.0,
3.25,3.5,4.0
8,12,15,18,23
,28
33,38
XIENCE V Abott
vascular
FDA
Approved
Everolimus
100μg/cm2
Multi-layer
Coating
MULTI-LINK
VISION CoCr
stent
Hybrid cell
Multilink
0.0032" strut
thickness,
laser cut,
PBMA
Non erodible
2.25,2.5,2.75,
3.0,3.5,4.0
8,12,15,18,23
,28
XINCE
PRIME
Abott
vascular
FDA
Approved
Everolimus
100μg/cm2
Cobalt
Chromium
Hybrid cell
Multilink
0.0032" strut
thickness,las
er cut,
biocompatibl
e fluorinated
copolymer
SV-2.25
MV
2.5,2.75,3.0,
3.5,4.0
LL-
2.5,2.75,3.0,
3.5,4.0
8,12,15,18,23
,28
Same
33,38
THE XIENCE XPEDITION EVEROLIMUS ELUTING
CORONARY STENT SYSTEM
(ABOTT VASCULAR) FDA, CE MARK
 The drug-coated stent and the balloon expandable
delivery system
 22% less force used to deliever than prime.
 Ultra low distal seal technology for outstanding
crossability.
 Unique 3.25mm diameter for more accurate vessel
sizing.
 More flexible multilayered balloon with flatter
compliance.
Stent Manufactur
er
Drug Base Form/Desi
gn
Polymer Diameter Length
Promus element
Plus
Boston scientific Everolimus Platinum
Chromium
Tubular open
cell,thin
strut,high radial
strength,good
delieverality &
trackability
Thin, fluorinated
copolymer
matrix for
controlled drug
release (100%
drug elution in
120 days)
2.25,2.5,2.75,3.0
,3.5,4.0
8,12,16,20,24,28
,32,38
Endeavor Sprint Medtronic Zotarolimus-
Eluting
10μg/mm
cobalt-based
alloy (cobalt,
nickel,
chromium, and
molybdenum)
Modular
design,Sinusoid
al form
wire,helical
wrap,laser fused
Phosphorylcholi
ne polymer
2.25,2.5,2.75,3.0
,3.5,4.0
8,12,14,18,22,26
,30,34,38
Resolut Integrity Medtronic Zotarolimus
eluting
cobalt-based
alloy (cobalt,
nickel,
chromium, and
molybdenum)
Modular
design,Sinusoidal
form wire,helical
wrap,laser fused
BioLinx
biocompatible
polymer
2.25,2.5,2.75,3.0
,3.5,4.0
8,12,14,18,22,26
,30,34,38
Stent Manufactur
er
Drug Base Form/Desi
gn
Polymer Diameter Length
Taxus Liberte Boston Scientific Paclitaxel
1 μg/mm2
paclitaxel in a
slow release
(SR)*
316L surgical
grade stainless
steel
Sinusoidal ring
modules linked
via curved link
elements
SIBS
[poly(styrene-b-
isobutylene-b-
styrene)], a tri-
block copolymer
(trade name:
Translute)
2.50, 2.75, 3.00,
3.50, 4.00
8, 12, 16, 20, 24,
28, 32
TAXUS Express Boston Scientific Paclitaxel
1μg/mm2
paclitaxel in a
slow release
(SR)
316L surgical
grade stainless
steel
modular ring
strut pattern
consists of two
separate module
designs: short,
narrow
sinusoidal Micro
elements linked
via straight
articulations to
long, wide
sinusoidal Macro
elements
SIBS
[poly(styrene-b-
isobutylene-b-
styrene)], a tri-
block copolymer
(trade name:
Translute)
2.50, 2.75, 3.00,
3.50
8, 12, 16, 20, 24,
28, 32
Taxus Element Boston Scientific Paclitaxel
1.0 μg/mm2
Platinum
Chromium
Sinusoidal ring
modules
consisting of
alternating long
and short
SIBS
[poly(styrene-b-
isobutylene-b-
styrene)], a tri-
block copolymer
2.25,2.50,2.75,3.
0,3.5,4.0,4.5
8,12,16,20,24,28
,32,38
Stent Manufactur
er
Drug Base Form/Design Polymer Diameter Length
Coracto Alvimedica Rapamycin Stainless
steel
Tubular,open cell
design
Ultrathin
polymer layer
absobes 100%
in 10-12 week
2.5,2.75,2.90,3
.00,3.5,4.0
9,13,17,21,26,
28,32
Coroflex
please
B.Braun Paclitaxel
1μg/cumm
Stainless
steel
Multicellular ring
design,Hybrid
Superb
radioopacity
P matrix-
polysulfone
coating
2.5,2.75,3.0,3.
5,4.0
8,13,16,19,25,
28,32
Cypher cordis Sirolimus
100% drug
release with in 1
month
Stainless
steel
Tubular,laser
cut,sinusoidal
pattern,closed cell
two non-erodible
polymers:
polyethylene-co-
vinyl acetate
(PEVA) and poly
n-butyl
methacrylate
(PBMA)
2.50, 2.75, 3.00,
3.50
8, 13, 18, 23, 28,
33
Stent Manufactu
rer
Drug Base Form/Desi
gn
Polymer Diameter Length
YUKON
Choice 4DES
Translumina,
German
CE mark
Sirolimus Medical
Stainless
Steel, 316
LVM, Surface
containing
micro-pores
1million
pores/sqcm
Balloon marker
material
Platinum /
Iridium
microporous
PEARL
Surface
Strut thickness
0,0034” / 87
μm
Hybrid design
Non
polymeric
Shellac resin
bio
compatible
resin
6 to 8 weeks
release
2.0,2.25,2.50,2
.75,3.0,3.5,4.0
8,12,16,18,21,
24,28,32,40
GEN X Sync MIV
therapeutics
India pvt ltd
Sirolimus Co Cr Open cell,
alternate S
link,uniform
sinusoidal strut
design
Bio resorb
PLLA-poly L
lactic acid
polymer
Ultrathin
coating(3μm)
Drug sudden
release f/b
release upto 40-
50 days.
2.0,2.25,2.50,2
.75,3.00,3.50,4
.0,4.5
8,13,16,19,24,
29,32,37
Supralimus Sahajanand
Medical
Technologies
Pvt Ltd, India
Sirolimus Sainless steel Hybrid biodegradable
drug-
carrier ,50%
drug release in
7 days next
50% in 41days
2.5,2.75,3.0,3.
5
8,12,16,20,24,
2832,36,40
Supralimus-
Core
Sahajanand
Medical
Technologies
Pvt Ltd, India
Sirolimus cobalt-
chromium
Hybrid biodegradable
drug-
carrier ,50%
drug release in
7 days next
50% in 41days
same same
Stent Manufactur
er
Drug Base Form/Desi
gn
Polymer Diameter Length
YUKON Choice
PC
Translumina,
German
CE mark
Rapamycin
(Sirolimus)
Release of
sirolimus up to 4
weeks
Medical
Stainless Steel,
316 LVM,
Surface
containing
micro-pores
1million
pores/sqcm
Favours better
endothelialisatio
n
Balloon marker
material
Platinum /
Iridium
microporous
PEARL Surface
Strut thickness
0,0034” / 87 μm
Hybrid design
The
biodegradable
components
polylactide and
shellac
2.0,2.50,2.75,3.0
,3.5,4.0
8,12,16,18,21,24
,28,32,40
Stent Manufactu
rer
Drug Base Form/Desi
gn
Polymer Diameter Length
BioMatrix Biosensors
Inc, Newport
Beach, Calif
CE mark
biolimus A9
highly
lipophilic,
semi
synthetic
sirolimus
analogue
(≈15.6 μg/mm
of stent
length)
S-Stent (316
L) stainless
steel stent
with a strut
thickness of
0.0054 inches
(137 μm)
laser-cut,
tubular stent
S-Stent
platform
Open cell,
quadrature
link
Biodegradabl
e,
Polylactic
acid (PLA)
applied to the
abluminal
surface
2.25,2.50,2.7
5,3.0,3.5,4.0
8,11,14,18,24
,28,33,36
Pronova Vascular
concepts,UK
Sirolimus Co Cr Hybrid
S shaped
articulations
Biocompatibl
e,biostable
polymer,drug
release upto
30 days
2.25,2.50,2.7
5,3.0,3.25,3.5
0,4.0
13,18,23,28,3
3,38
Biomime Meril Life
Sciences,
India
Sirolimus
1.25μgm/sqm
m of stent
surface,30 day
elution kinetics
Co Cr Hybrid cell
design
65μm strut
thickness
Biodegradabl
e polymer
2.5,2.75,3.0,3
.5,4.0,4.5
8,13,16,19,24
,29,32,37,40
Stent Manufactur
er
Drug Base Form/Desi
gn
Polymer Diameter Length
ACTIVE&
ACTVE small
IHT Paclitaxel Stainless steel Open
cell,tubular
P5 -
Biocompatible
polymer
2.0,2.25,2.5,2.
75,3.0,3.5,4.0,
4.5
9,14,18,19,23,
28,36
EVERLITE Unimark
remedies
Everolimus
Low drug dose
1.2μg/sqmm
Co Cr Open
cell,Sinosoidal
strut
design,alternativ
e S link,ultrathin
strut 65μm
Biodegradable 2.25,2.5,2.75,3.0
,3.5,4.0,4.5
8,13,16,19,24,29
,32,37,40
Flexy Rap Lancer medical
technology
Rapamycin
1μg/sqmm
Co Cr Open
cell, Radial star
segments
combined with
flexible
links,Strut 65μm,
Biodegradable
polymer
2.25,2.5,2.75,3.0
,3.5,4.0
7,10,13,15,17,20
,24,28,33,38,42
INDOLIMUS
Ce mark
Sahajanand
medical
sirolimus Co Cr Open cell,laser
cut,seamless
tube,60 micm
strut thickness
Biodegradable
polymer matrix
2.5,2.75,3.0,3.5 8,12,16,20,24,28
,32,36,40
THANKS

More Related Content

Similar to coronarystentdesignppt-131213121014-phpapp02.pdf

Novel self expanding nitinol carotid stent - dr Krzysztof Milewski
Novel self expanding nitinol carotid stent - dr Krzysztof MilewskiNovel self expanding nitinol carotid stent - dr Krzysztof Milewski
Novel self expanding nitinol carotid stent - dr Krzysztof Milewski
piodof
 
Orientation Effects of Stress Concentrators on the Material Deformation Beha...
Orientation Effects of Stress Concentrators on the Material  Deformation Beha...Orientation Effects of Stress Concentrators on the Material  Deformation Beha...
Orientation Effects of Stress Concentrators on the Material Deformation Beha...
IJMER
 
Strategies of handling side branch during pci
Strategies of handling side branch during pciStrategies of handling side branch during pci
Strategies of handling side branch during pci
Manjunath D
 
Recent advances in arch wires
Recent advances in arch wiresRecent advances in arch wires
Recent advances in arch wires
Ajinkya Patel
 
Orientation Effects of Stress Concentrators on the Material Deformation Behav...
Orientation Effects of Stress Concentrators on the Material Deformation Behav...Orientation Effects of Stress Concentrators on the Material Deformation Behav...
Orientation Effects of Stress Concentrators on the Material Deformation Behav...
IJMER
 

Similar to coronarystentdesignppt-131213121014-phpapp02.pdf (20)

biomechanics of stents
biomechanics of stentsbiomechanics of stents
biomechanics of stents
 
Biovascular scaffolds
Biovascular scaffolds Biovascular scaffolds
Biovascular scaffolds
 
Appliance selection in fixed orthodontic appliances (dr nayaungbds phd)
Appliance selection in fixed orthodontic appliances (dr nayaungbds phd)Appliance selection in fixed orthodontic appliances (dr nayaungbds phd)
Appliance selection in fixed orthodontic appliances (dr nayaungbds phd)
 
coronary artery calcification.pptx
coronary artery calcification.pptxcoronary artery calcification.pptx
coronary artery calcification.pptx
 
Coronary Calcium Modification
Coronary Calcium ModificationCoronary Calcium Modification
Coronary Calcium Modification
 
Comparative study of effect of basalt, glass and steel fiber on compressive a...
Comparative study of effect of basalt, glass and steel fiber on compressive a...Comparative study of effect of basalt, glass and steel fiber on compressive a...
Comparative study of effect of basalt, glass and steel fiber on compressive a...
 
Novel self expanding nitinol carotid stent - dr Krzysztof Milewski
Novel self expanding nitinol carotid stent - dr Krzysztof MilewskiNovel self expanding nitinol carotid stent - dr Krzysztof Milewski
Novel self expanding nitinol carotid stent - dr Krzysztof Milewski
 
Tissue engineering of stents
Tissue engineering of stentsTissue engineering of stents
Tissue engineering of stents
 
Ultrasonics endodontic tips
Ultrasonics endodontic tips Ultrasonics endodontic tips
Ultrasonics endodontic tips
 
Biosolve II
Biosolve IIBiosolve II
Biosolve II
 
Orientation Effects of Stress Concentrators on the Material Deformation Beha...
Orientation Effects of Stress Concentrators on the Material  Deformation Beha...Orientation Effects of Stress Concentrators on the Material  Deformation Beha...
Orientation Effects of Stress Concentrators on the Material Deformation Beha...
 
20320140503001
2032014050300120320140503001
20320140503001
 
Strategies of handling side branch during pci
Strategies of handling side branch during pciStrategies of handling side branch during pci
Strategies of handling side branch during pci
 
Bio vascular scaffold i tammi raju
Bio vascular scaffold i tammi rajuBio vascular scaffold i tammi raju
Bio vascular scaffold i tammi raju
 
Bio vascular scaffold i tammi raju
Bio vascular scaffold i tammi rajuBio vascular scaffold i tammi raju
Bio vascular scaffold i tammi raju
 
Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...
Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...
Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...
 
Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...
Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...
Cyclic Elastoplastic Large Displacement Analysis and Stability Evaluation of ...
 
Recent advances in arch wires
Recent advances in arch wiresRecent advances in arch wires
Recent advances in arch wires
 
Orientation Effects of Stress Concentrators on the Material Deformation Behav...
Orientation Effects of Stress Concentrators on the Material Deformation Behav...Orientation Effects of Stress Concentrators on the Material Deformation Behav...
Orientation Effects of Stress Concentrators on the Material Deformation Behav...
 
Thornton clifford sample_consulting
Thornton clifford sample_consultingThornton clifford sample_consulting
Thornton clifford sample_consulting
 

Recently uploaded

+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
levieagacer
 
development of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virusdevelopment of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virus
NazaninKarimi6
 
biology HL practice questions IB BIOLOGY
biology HL practice questions IB BIOLOGYbiology HL practice questions IB BIOLOGY
biology HL practice questions IB BIOLOGY
1301aanya
 

Recently uploaded (20)

An introduction on sequence tagged site mapping
An introduction on sequence tagged site mappingAn introduction on sequence tagged site mapping
An introduction on sequence tagged site mapping
 
GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)
 
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
 
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
 
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort ServiceCall Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
 
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit flypumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Stages in the normal growth curve
Stages in the normal growth curveStages in the normal growth curve
Stages in the normal growth curve
 
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate ProfessorThyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
 
GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)
 
Clean In Place(CIP).pptx .
Clean In Place(CIP).pptx                 .Clean In Place(CIP).pptx                 .
Clean In Place(CIP).pptx .
 
GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)
 
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICEPATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
 
Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS ESCORT SERVICE In Bhiwan...
Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS  ESCORT SERVICE In Bhiwan...Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS  ESCORT SERVICE In Bhiwan...
Bhiwandi Bhiwandi ❤CALL GIRL 7870993772 ❤CALL GIRLS ESCORT SERVICE In Bhiwan...
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
 
CURRENT SCENARIO OF POULTRY PRODUCTION IN INDIA
CURRENT SCENARIO OF POULTRY PRODUCTION IN INDIACURRENT SCENARIO OF POULTRY PRODUCTION IN INDIA
CURRENT SCENARIO OF POULTRY PRODUCTION IN INDIA
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
development of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virusdevelopment of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virus
 
biology HL practice questions IB BIOLOGY
biology HL practice questions IB BIOLOGYbiology HL practice questions IB BIOLOGY
biology HL practice questions IB BIOLOGY
 
Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.
 

coronarystentdesignppt-131213121014-phpapp02.pdf

  • 1. DIFFERENT CORONARY STENT DESIGN DR AWADHESH KUMAR SHARMA DEPARTMENT OF CARDIOLOGY PGIMER & DR RML HOSPITAL,NEW DELHI
  • 2. BACKGROUND  The introduction of angioplasty led to the development of a completely new approach to treat CAD.  Until 1994, the percutaneous transluminal coronary angioplasty (PTCA) was the alone treatment for coronary artery disease.  However, the incidence of restenosis of coronary arteries was an important problem, necessitating repeated interventional procedures in 30% of patients treated with PTCA alone. NEJM 1994;331:489-95
  • 3.
  • 4.  Coronary artery stents were developed to provide a metal scaffolding for the angio-plastied vessel, in an attempt to limit negative re-modelling.  Sigwart et al first reported the efficacy of stents in reducing restenosis rates in 1987.  By 1994, the Food and Drug Administration (FDA) had approved two stents (Gianturco-Roubin stent and the Palmaz- SchatzTM stent). Journal of Invasiv Cardiology.2001;13:634-639
  • 5. GIANTURCO-ROUBIN II  Flat wire coil attached to a single longitudinal strut  316 L stainless steel  The first coronary stent approved by the FDA in June 1993. J. clinical pathology.2005,aug;58(8):795-804
  • 6. PALMAZ-SCHATZ  Balloon expandable; slotted tube  316 L stainless steel J. clinical pathology.2005,aug;58(8):795-804
  • 7.  The wide acceptance of coronary stenting was based on the results of the BElgian NEtherlands STENT (BENESTENT) and the STent REStenosis Study (STRESS) trials, which showed the superiority of stenting over balloon angioplasty.  After the wide acceptance of coronary stents the primary concern of stent development is the need to reduce device profiles and to increase flexibility to facilitate safe delivery. N Engl J Med,1994;331:489-95.
  • 12. TYPES OF STENTS  Mechanism of expansion (self-expanding or balloon- expandable)  Materials (stainless steel, cobalt-based alloy, tantalum, nitinol, Pt,Ir,Cr, inert coating, biodegradable)  Forms (sheet, wire or tube)  Manufacturing methods (laser cut, water-jet cutting, photo- etching)  Geometrical configurations/design (mesh structure, coil, slotted tube, ring, multi-design)  Addition to stent (grafts, radio-opaque markers, coatings) Min Invas Ther & Allied Technol 2002;11:137-47.
  • 13. STENT GEOMETRIC DESIGN  MECHANISM OF EXPANSION  Balloon-expandable stents  The stent is pre-mounted on a balloon and the inflation of the balloon plastically expands the stent with respect to the balloon diameter.  Self-expanding stents- The smart material auto expands to a calculated size. Journal of invasive cardiology 1995;7:127- 134
  • 14. MATERIALS  Corrosion resistance  Biocompatibility  Adequately radio-opaque  Create minimal artifacts during MRI
  • 15. STENT PLATFORMS STENT MATERIALS- NON DEGRADABLE MATERIAL  316L stainless steel-  Excellent mechanical properties and corrosion resistance  Ferromagnetic nature and low density make it a non-MRI compatible  Poorly visible fluoroscopic material  First generation DESs, Cypher (sirolimus-eluting stent, Cordis, Warren, NJ) and Taxus (paclitaxel- eluting stent, Boston Scientific, Natick, MA) JACC 1996;27:53
  • 16. CO-CR  Superior radial strength and improved radiopacity  Thinner stent struts  The second generation DES, Xience V (everolimus- eluting stent, Abott Vascular, CA) and Endeavor (zotarolimus-eluting stent,Medtronic Vascular, Santa Rosa, CA). JACC 1996;27:53
  • 17. TA- TANTALUM  Excellent corrosion resistant material  Coated on 316L SS to improve corrosion properties and biocompatibility  High density and non-ferromagnetic properties  Fluoroscopically visible and MRI compatible  Higher rates of recoil- poor mechanical properties JACC 1996;27:53
  • 18. TI  Excellent biocompatibility and corrosion resistance  Low tensile strength and ductility  Ti alloys in combination with Ni-Ti  Ti-nitride oxide coating on 316L SS JACC 1996;27:53
  • 19. NI-TI  Good biocompatibility, radial force and shape memory  Coated by some materials such as polyurethane, Ti nitride and polycrystalline oxides to improve the corrosion resistance  Inadequate visibility under fluoroscopy American J of cardiology.2008;86:1073- 1079
  • 20. PT-IR  Pt-Ir alloy of 90% platinum and 10% iridium  Excellent radiopacity and a reduction in both thrombosis and neointimal proliferation with less inflammatory reactions  Recoiling percentage was much higher (16%) than the 316L SS stents Journal of invasive cardiology 1995;7:127- 134
  • 21.
  • 22. BIODEGRADABLE METALLIC MATERIALS  Pure Fe  Oxidation of Fe into ferrous and ferric irons  Mg alloys  There are two Mg alloys, AE2153 and WE4357, used for making stents  Radiolucent Biomaterials.2006;27:1728-1734
  • 23. RATIONAL FOR BIODEGRADABLE STENTS Metal stent drawbacks  Cause permanent physical irritation  Risk of long term endothelial dysfunction and chronic inflammation  Metal have thrombogenic properties  Inability for the vessel to restore its a normal physiology Biodegradable stent advantages  May eliminate early and late complications of bare-metal stents  Restore the vasoreactivity  Allow a gradual transfer of the mechanical load to the vessel  Higher capacity for drug incorporation and complex release kinetics The need for a permanent prosthesis decreases dramatically 6 months post-implantation
  • 24. STENT DESIGN  On the basis of design, stents can be divided into three groups: coil, tubular mesh, and slotted tube.  Coil stents are characterised by metallic wires or strips formed into a circular coil shape  Tubular mesh stents consist of wires wound together in a meshwork, forming a tube.  Slotted tube stents are made from tubes of metal from which a stent design is laser cut. Eur Heart J 1997;18:1536–47
  • 25. COIL VS. TUBE  Coil design had greater strut width with gaps and fewer or no connections between struts  The strut width is greater; there are gap between struts, and no connections between struts which give it more flexibility.  However, the design lack radial strength, and the wide gap allow tissues to dangle. Singapore Medical Journal, 2004.
  • 27.  As a result, coil design has become obsolete and replace by the more superior in radial strength, the tube design.  In tubular, there are two type of specification, a slotted tube and modular tube. Singapore Medical Journal, 2004.
  • 28. SLOTTED TUBE VS. MODULAR (TUBULAR)
  • 30.
  • 31. SLOTTED TUBE VS. MODULAR (TUBULAR)  Slotted tube stents resisted restenosis more than the modular stents (22.1% vs 25.2%)  Slotted tube- Closed cell design, and open cell design
  • 32. CLOSED CELL  Sequential ring construction  All Internal inflection points of the structural members are connected by bridging elements.  Regular peak-to-peak connections.  Optimal scaffolding and a uniform surface, regardless of the degree of bending.  Less flexible than a similar open-cell design. Ann Ist Super Sanita 2007;43,no1:89-100
  • 33. CYPHER STENT BY CORDIS
  • 34. OPEN CELL  Some or all the internal inflection points of the structural members are not connected by bridging elements.  Periodic peak-to-peak connections, peak-to-valley connections, and mid-strut to mid strut connections  The unconnected structural elements contribute to longitudinal flexibility. Ann Ist Super Sanita 2007;43,no1:89-100
  • 36.
  • 37.
  • 38.
  • 39.
  • 40. STENT DESIGN IMPACTS DRUG DELIEVERY
  • 41.
  • 42. LENGTH & DIAMETER OF STENT  Long vs. Short  Stent length is associated with restenosis rate and clinical events (mainly target lesion revascularization)  Short stent has lower cases of restenosis than long stent.  Wide vs. Narrow  The wide diameter stent is more favorable than the narrow one European Heart Journal 2001;22:1585- 1593
  • 43. NUMBER OF STRUTS  More struts vs. less  Less struts induce less chance of restenosis compare to more struts.
  • 44.
  • 45. THIN STRUT VS THICK STRUT
  • 46.
  • 47. STRUT THICKNESS  Although the immediate stent performance may be improved by increasing strut thickness (which increases radiovisibility, radial strength and arterial wall support) excessive strut thickness, on the other hand, may impart more vascular injury, trigger more intimal hyperplasia, and engender a higher risk for restenosis than thinner struts.  Strut thickness was observed to be an independent predictor of in-stent restenosis ISAR STEREO study(Circulation 2001;103:2816-21) ISAR-STEREO-2 trial(J Am Coll Cardiol 2003;41:1283-8.)
  • 48.  In an effort to further reduce strut thickness while maintaining adequate radiovisibility and radial strength, novel metallic materials such as cobalt- chromium alloy are being used for the production of stent.
  • 49. THICK VS. THIN STRUTS  The stents with thinner struts is preferred for the design of new stents as they can reduce angiographic and clinical restenosis more than those with thicker struts ISAR-STEREO and ISAR-STEREO 2 trials
  • 50.
  • 51.
  • 52. SQUARE VS. ROUND STRUT CROSS-SECTION  The round strut cross-section without corners or sharp edges is popular at present  Round strut cross-section area is ideal for smoothness design.  Square strut cross-section area in not recommend because it interferes with blood flow due to their sharp edge which can slice blood cells. Kluwer Academic Publishers 2012
  • 53. SQUARE VS. ROUND STRUT CROSS-SECTION
  • 54. ROUGH VS. SMOOTH SURFACE  Smoothness of a stent can affect the performance and biocompatibily of the stent.  Smooth surface can reduce thrombus adhesion and neointimal growth.  To obtain smoothness, the stent need to be treated with acid-pickling and then electrochemical polishing.  The process removes slag which includes depositions and burrs, formed on the surface of stents due to the laser cutting production process. Seminars in interventional cardiology1998;3:139-144
  • 55. ELEMENT OF STENT DESIGN- BALLOON OVERHANG
  • 56. DRUG DELIVERY VEHICLES – COATING POLYMER- DRUG CARRIERS IN DESS  Nonbiodegradable and biodegradable polymers  Non biodegradable polymers  First and the second generation of DESs  The first generation of DES  Cypher - polyethylene-co-vinyl acetate (PEVA)/poly-n-butyl methacrylate (PBMA)  Taxus - polystyrene-b-isobutylene-b-styrene (SIBS)  The second generation of DES  Xience V – fluoropolymer  Endeavor - phosphorylcholine (PC) Eurointervention,2005;1:266-272
  • 57.  Biodegradable polymers  Polylactic acid (PLA)  Polyglycolic acid (PGA)  Polylactic-co-glycolic acid (PLGA)  NON POLYMER  Titanium–nitric oxide alloy  Microporous stainless steel stent (Yukon, Translumina, Germany)  A nanoporous hydroxyapetite (a biocompatible crystalline derivative of calcium phosphate) coating  Magnetic nanoparticles (MNPs) Eurointervention,2005;1:266-272
  • 58. YUKON Choice DES system: Translumina modified stent surface containing micropores to enable the adsorption of different organic substances. Abizaid A , and Costa J R Circ Cardiovasc Interv 2010;3:384-393 Copyright © American Heart Association
  • 59.
  • 60. THERAPEUTIC AGENTS  Sirolimus (Rapamycin)  A macrocyclic lactone  Inhibits the migration and proliferation of SMCs  Zotarolimus  The sirolimus analogues  Developed by Abbott laboratories  Extremely lipophilic property and low water solubility  Everolimus  Sirolimus analogue  Immunosuppressive agent  Absorbs to local tissue more rapidly and has a longer celluar residence time and activity  Biolimus
  • 61. PACLITAXEL AND ITS ANALOGUES  Paclitaxel  Promoting tubulin polymerization and cell cycle arrest  Inhibiting the migration and proliferation of SMCs  Coroxane  Nanoparticle albumin bound paclitaxel (nab-paclitaxel)  To improve the solubility  Docetaxel  Semi-synthetic analogue  Better anti-proliferative properties
  • 62. OTHERS  Tacrolimus  Pimecrolimus  Curcumin  Resveratrol  CD 34 antibody  Anti-VEGF
  • 63. RADIO-OPACITY ENHANCEMENTS  Stainless steel or nitinol - hard to see fluoroscopically  To improve X-ray visibility, markers are often attached to the stents.  These additions are typically made from gold, platinum or tantalum  Electroplating (with gold) is also being used to enhance X-ray visibility
  • 64. COATINGS  To increase biocompatibility  Heparin was one of the first. Its mode of action is to reduce the coagulation cascade (and thus possibly the thrombogenic risk) after the deployment of a stent.  Phosphorylcoline and silicon-carbide have been used in order to reduce platelet activation and interaction, thus possibly controlling their adhesion to the stent struts during the acute phase of stent re-endothelization.
  • 65.  Passive coverage has been also shown to be useful.  Indeed, covered stents have been created, in which a PTFE layer was put between two stents (Jostent graft, Jomed) or one stent was covered by a inner and an outer layer of PTFE (Symbiot, Boston Scientific)
  • 66. COMMONLY USED CORONARY STENTS IN CLINICAL PRACTICE
  • 67. XIENCE FAMILY OF STENTS Stent Manufactu rer Drug Base Form/Desi gn Polymer Diameter Length XIENCE Xpedition Abott vascular FDA Approved Everolimus 100μg/cm2 L-605 CoCr Hybrid cell Multilink 0.0032" strut thickness, laser cut PBMA Non erodible SV-2.25 MV- 2.5,2.75,3.0,3. 25,3.5,4.0 LL 2.5,2.75,3.0, 3.25,3.5,4.0 8,12,15,18,23 ,28 33,38 XIENCE V Abott vascular FDA Approved Everolimus 100μg/cm2 Multi-layer Coating MULTI-LINK VISION CoCr stent Hybrid cell Multilink 0.0032" strut thickness, laser cut, PBMA Non erodible 2.25,2.5,2.75, 3.0,3.5,4.0 8,12,15,18,23 ,28 XINCE PRIME Abott vascular FDA Approved Everolimus 100μg/cm2 Cobalt Chromium Hybrid cell Multilink 0.0032" strut thickness,las er cut, biocompatibl e fluorinated copolymer SV-2.25 MV 2.5,2.75,3.0, 3.5,4.0 LL- 2.5,2.75,3.0, 3.5,4.0 8,12,15,18,23 ,28 Same 33,38
  • 68.
  • 69. THE XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM (ABOTT VASCULAR) FDA, CE MARK  The drug-coated stent and the balloon expandable delivery system  22% less force used to deliever than prime.  Ultra low distal seal technology for outstanding crossability.  Unique 3.25mm diameter for more accurate vessel sizing.  More flexible multilayered balloon with flatter compliance.
  • 70. Stent Manufactur er Drug Base Form/Desi gn Polymer Diameter Length Promus element Plus Boston scientific Everolimus Platinum Chromium Tubular open cell,thin strut,high radial strength,good delieverality & trackability Thin, fluorinated copolymer matrix for controlled drug release (100% drug elution in 120 days) 2.25,2.5,2.75,3.0 ,3.5,4.0 8,12,16,20,24,28 ,32,38 Endeavor Sprint Medtronic Zotarolimus- Eluting 10μg/mm cobalt-based alloy (cobalt, nickel, chromium, and molybdenum) Modular design,Sinusoid al form wire,helical wrap,laser fused Phosphorylcholi ne polymer 2.25,2.5,2.75,3.0 ,3.5,4.0 8,12,14,18,22,26 ,30,34,38 Resolut Integrity Medtronic Zotarolimus eluting cobalt-based alloy (cobalt, nickel, chromium, and molybdenum) Modular design,Sinusoidal form wire,helical wrap,laser fused BioLinx biocompatible polymer 2.25,2.5,2.75,3.0 ,3.5,4.0 8,12,14,18,22,26 ,30,34,38
  • 71. Stent Manufactur er Drug Base Form/Desi gn Polymer Diameter Length Taxus Liberte Boston Scientific Paclitaxel 1 μg/mm2 paclitaxel in a slow release (SR)* 316L surgical grade stainless steel Sinusoidal ring modules linked via curved link elements SIBS [poly(styrene-b- isobutylene-b- styrene)], a tri- block copolymer (trade name: Translute) 2.50, 2.75, 3.00, 3.50, 4.00 8, 12, 16, 20, 24, 28, 32 TAXUS Express Boston Scientific Paclitaxel 1μg/mm2 paclitaxel in a slow release (SR) 316L surgical grade stainless steel modular ring strut pattern consists of two separate module designs: short, narrow sinusoidal Micro elements linked via straight articulations to long, wide sinusoidal Macro elements SIBS [poly(styrene-b- isobutylene-b- styrene)], a tri- block copolymer (trade name: Translute) 2.50, 2.75, 3.00, 3.50 8, 12, 16, 20, 24, 28, 32 Taxus Element Boston Scientific Paclitaxel 1.0 μg/mm2 Platinum Chromium Sinusoidal ring modules consisting of alternating long and short SIBS [poly(styrene-b- isobutylene-b- styrene)], a tri- block copolymer 2.25,2.50,2.75,3. 0,3.5,4.0,4.5 8,12,16,20,24,28 ,32,38
  • 72. Stent Manufactur er Drug Base Form/Design Polymer Diameter Length Coracto Alvimedica Rapamycin Stainless steel Tubular,open cell design Ultrathin polymer layer absobes 100% in 10-12 week 2.5,2.75,2.90,3 .00,3.5,4.0 9,13,17,21,26, 28,32 Coroflex please B.Braun Paclitaxel 1μg/cumm Stainless steel Multicellular ring design,Hybrid Superb radioopacity P matrix- polysulfone coating 2.5,2.75,3.0,3. 5,4.0 8,13,16,19,25, 28,32 Cypher cordis Sirolimus 100% drug release with in 1 month Stainless steel Tubular,laser cut,sinusoidal pattern,closed cell two non-erodible polymers: polyethylene-co- vinyl acetate (PEVA) and poly n-butyl methacrylate (PBMA) 2.50, 2.75, 3.00, 3.50 8, 13, 18, 23, 28, 33
  • 73. Stent Manufactu rer Drug Base Form/Desi gn Polymer Diameter Length YUKON Choice 4DES Translumina, German CE mark Sirolimus Medical Stainless Steel, 316 LVM, Surface containing micro-pores 1million pores/sqcm Balloon marker material Platinum / Iridium microporous PEARL Surface Strut thickness 0,0034” / 87 μm Hybrid design Non polymeric Shellac resin bio compatible resin 6 to 8 weeks release 2.0,2.25,2.50,2 .75,3.0,3.5,4.0 8,12,16,18,21, 24,28,32,40 GEN X Sync MIV therapeutics India pvt ltd Sirolimus Co Cr Open cell, alternate S link,uniform sinusoidal strut design Bio resorb PLLA-poly L lactic acid polymer Ultrathin coating(3μm) Drug sudden release f/b release upto 40- 50 days. 2.0,2.25,2.50,2 .75,3.00,3.50,4 .0,4.5 8,13,16,19,24, 29,32,37 Supralimus Sahajanand Medical Technologies Pvt Ltd, India Sirolimus Sainless steel Hybrid biodegradable drug- carrier ,50% drug release in 7 days next 50% in 41days 2.5,2.75,3.0,3. 5 8,12,16,20,24, 2832,36,40 Supralimus- Core Sahajanand Medical Technologies Pvt Ltd, India Sirolimus cobalt- chromium Hybrid biodegradable drug- carrier ,50% drug release in 7 days next 50% in 41days same same
  • 74. Stent Manufactur er Drug Base Form/Desi gn Polymer Diameter Length YUKON Choice PC Translumina, German CE mark Rapamycin (Sirolimus) Release of sirolimus up to 4 weeks Medical Stainless Steel, 316 LVM, Surface containing micro-pores 1million pores/sqcm Favours better endothelialisatio n Balloon marker material Platinum / Iridium microporous PEARL Surface Strut thickness 0,0034” / 87 μm Hybrid design The biodegradable components polylactide and shellac 2.0,2.50,2.75,3.0 ,3.5,4.0 8,12,16,18,21,24 ,28,32,40
  • 75. Stent Manufactu rer Drug Base Form/Desi gn Polymer Diameter Length BioMatrix Biosensors Inc, Newport Beach, Calif CE mark biolimus A9 highly lipophilic, semi synthetic sirolimus analogue (≈15.6 μg/mm of stent length) S-Stent (316 L) stainless steel stent with a strut thickness of 0.0054 inches (137 μm) laser-cut, tubular stent S-Stent platform Open cell, quadrature link Biodegradabl e, Polylactic acid (PLA) applied to the abluminal surface 2.25,2.50,2.7 5,3.0,3.5,4.0 8,11,14,18,24 ,28,33,36 Pronova Vascular concepts,UK Sirolimus Co Cr Hybrid S shaped articulations Biocompatibl e,biostable polymer,drug release upto 30 days 2.25,2.50,2.7 5,3.0,3.25,3.5 0,4.0 13,18,23,28,3 3,38 Biomime Meril Life Sciences, India Sirolimus 1.25μgm/sqm m of stent surface,30 day elution kinetics Co Cr Hybrid cell design 65μm strut thickness Biodegradabl e polymer 2.5,2.75,3.0,3 .5,4.0,4.5 8,13,16,19,24 ,29,32,37,40
  • 76. Stent Manufactur er Drug Base Form/Desi gn Polymer Diameter Length ACTIVE& ACTVE small IHT Paclitaxel Stainless steel Open cell,tubular P5 - Biocompatible polymer 2.0,2.25,2.5,2. 75,3.0,3.5,4.0, 4.5 9,14,18,19,23, 28,36 EVERLITE Unimark remedies Everolimus Low drug dose 1.2μg/sqmm Co Cr Open cell,Sinosoidal strut design,alternativ e S link,ultrathin strut 65μm Biodegradable 2.25,2.5,2.75,3.0 ,3.5,4.0,4.5 8,13,16,19,24,29 ,32,37,40 Flexy Rap Lancer medical technology Rapamycin 1μg/sqmm Co Cr Open cell, Radial star segments combined with flexible links,Strut 65μm, Biodegradable polymer 2.25,2.5,2.75,3.0 ,3.5,4.0 7,10,13,15,17,20 ,24,28,33,38,42 INDOLIMUS Ce mark Sahajanand medical sirolimus Co Cr Open cell,laser cut,seamless tube,60 micm strut thickness Biodegradable polymer matrix 2.5,2.75,3.0,3.5 8,12,16,20,24,28 ,32,36,40