Electrodes and
Potentiometry
Electrodes and Potentiometry
Introduction
1.) Potentiometry
 Use of Electrodes to Measure Voltages that Provide Chemical Information
- Various electrodes have been designed to respond selectively to specific analytes
 Use a Galvanic Cell
- Unknown solution becomes a ½-cell
- Add Electrode that transfers/accepts
electrons from unknown analyte
- Connect unknown solution by salt
bridge to second ½-cell at fixed
composition and potential
 Indicator Electrode: electrode that
responds to analyte and donates/accepts
electrons
 Reference Electrode: second ½ cell at a
constant potential
 Cell voltage is difference between the
indicator and reference electrode
Binding generates
potential difference.
Electrodes and Potentiometry
Introduction
2.) Example
 A Heparin Sensor
- Voltage response is proportional to heparin concentration in blood
- Sensor is selective for heparin
Negatively charged heparin
binds selectively to positively
charged membrane.
heparin
Potential is
proportional to
[heparin]
Electrodes and Potentiometry
Reference Electrodes
1.) Overview
 Potential change only dependent on one ½ cell concentrations
 Reference electrode is fixed or saturated  doesn’t change!
Reference electrode,
[Cl-] is constant
Potential of the cell
only depends on
[Fe2+] & [Fe3+]
Pt wire is indicator
electrode whose
potential responds
to [Fe2+]/[Fe3+]
 
]
Cl
log[
.
.
]
Fe
[
]
Fe
[
log
.
.
Ecell
























 05916
0
222
0
1
05916
0
771
0
3
2
Unknown solution of
[Fe2+] & [Fe3+]
Electrodes and Potentiometry
Reference Electrodes
2.) Silver-Silver Chloride Reference Electrode
 Convenient
- Common problem is porous plug becomes clogged
Eo = +0.222 V
Activity of Cl- not 1E(sat,KCl) = +0.197 V
Electrodes and Potentiometry
Reference Electrodes
3.) Saturated Calomel Reference Electrode (S.C.E)
 Saturated KCl maintains constant [Cl-] even with
some evaporation
 Standard hydrogen electrodes are cumbersome
- Requires H2 gas and freshly prepared Pt surface
Eo = +0.268 V
Activity of Cl- not 1E(sat,KCl) = +0.241 V
Electrodes and Potentiometry
Reference Electrodes
4.) Observed Voltage is Reference Electrode Dependant
 The observed potential depends on the choice of reference electrode
- Silver-silver chloride and calomel have different potentials
 Use Reference Scale to convert between Reference Electrodes
Observed potential relative to SCE
Observed potential relative to Ag|AgCl
Observed potential relative to SHE
Electrodes and Potentiometry
Junction Potential
1.) Occurs Whenever Dissimilar Electrolyte Solutions are in Contact
 Develops at solution interface (salt bridge)
 Small potential (few millivolts)
 Junction potential puts a fundamental limitation on the accuracy of direct
potentiometric measurements
- Don’t know contribution to the measured voltage
Again, an electric potential is generated by a separation of charge
Different ion mobility results
in separation in charge
Electrodes and Potentiometry
Indicator Electrodes
1.) Two Broad Classes of Indicator Electrodes
 Metal Electrodes
- Develop an electric potential in response to a redox reaction at the metal
surface
 Ion-selective Electrodes
- Selectively bind one type of ion to a membrane to generate an electric
potential
Remember an electric potential is generated by a separation of charge
Electrodes and Potentiometry
Indicator Electrodes
2.) Metal Electrodes
 Platinum
- Most common metal indicator electrode
- Inert: does not participate in many chemical reactions
- Simply used to transmit electrons
 Other electrodes include Gold and Carbon
 Metals (Ag, Cu, Zn, Cd, Hg) can be used to monitor their aqueous ions
- Most metals are not useable
- Equilibrium not readily established at the metal surface
E+
o = +799 V
E(sat,KCl) = +0.241 V
½ Reaction at Ag indicator electrode:
½ Reaction at Calomel reference electrode:
 
241
0
1
1
05916
0
799
0 .
]
Ag
[
log
.
.
E
E
Ecell 

























Potential of Ag
indicator electrode
Cell voltage changes as a function of [Ag+]
Example:
Cell Potential from Nernst Equation:
Electrodes and Potentiometry
Indicator Electrodes
3.) Example
 A 10.0 mL solution of 0.0500 M AgNO3 was titrated with 0.0250M NaBr in
the cell:
S.C.E. || titration solution | Ag(s)
Find the cell voltage for 10.0 mL of titrant
C+ diffuses across the membrane due to
concentration gradient resulting in charge
difference across membrane
Electrodes and Potentiometry
Indicator Electrodes
4.) Ion-Selective Electrodes
 Responds Selectively to one ion
- Contains a thin membrane
capable of only binding the
desired ion
 Does not involve a redox process
Membrane contains a ligand (L) that
specifically and tightly binds analyte
of interest (C+)
A difference in the concentration of C+
exists across the outer membrane.
The counter-ions (R-,A-) can’t cross the
membrane and/or have low solubility in
membrane or analyte solution
Remember an electric potential is generated by a separation of charge
Potential across outer membrane depends on
[C+] in analyte solution
Electrodes and Potentiometry
Indicator Electrodes
4.) Ion-Selective Electrodes
 Responds Selectively to one ion
- Contains a thin membrane
capable of only binding the
desired ion
 Does not involve a redox process
Remember an electric potential is generated by a separation of charge
Potential across inner membrane depends on [C+]
in filling solution, which is a known constant
C+ diffuses across the membrane due to
concentration gradient resulting in charge
difference across membrane
A difference in the concentration of C+
exists across the inner membrane.
inner
outer E
E
E 

Electrode potential is determined by the potential
difference between the inner and outer membranes:
where Einner is a constant and Eouter depends on the
concentration of C+ in analyte solution
]
[
constant 

 C
log
n
.
E
05916
0
Electrodes and Potentiometry
Indicator Electrodes
4.) Ion-Selective Electrodes
 Responds Selectively to one ion
- Contains a thin membrane
capable of only binding the
desired ion
 Does not involve a redox process
Electrode Potential is defined as:
where [C+] is actually the activity of the analyte and
n is the charge of the analyte
Electrodes and Potentiometry
pH Electrodes
1.) pH Measurement with a Glass Electrode
 Glass electrode is most common ion-selective electrode
 Combination electrode incorporates both glass and
reference electrode in one body
Ag(s)|AgCl(s)|Cl-(aq)||H+(aq,outside) H+(aq,inside),Cl-(aq)|AgCl(s)|Ag(s)
Outer reference
electrode
[H+] outside
(analyte solution)
[H+] inside Inner reference
electrode
Glass membrane
Selectively binds H+
Electric potential is generated by [H+] difference across glass membrane
Electrodes and Potentiometry
pH Electrodes
2.) Glass Membrane
 Irregular structure of silicate lattice
Cations (Na+) bind
oxygen in SiO4 structure
Electrodes and Potentiometry
pH Electrodes
2.) Glass Membrane
 Two surfaces of glass “swell” as they absorb water
- Surfaces are in contact with [H+]
Electrodes and Potentiometry
pH Electrodes
2.) Glass Membrane
 H+ diffuse into glass membrane and replace Na+ in hydrated gel region
- Ion-exchange equilibrium
- Selective for H+ because H+ is only ion that binds significantly to the
hydrated gel layer
H
(0.05916)
constant p
E 


Charge is slowly carried
by migration of Na+
across glass membrane
Potential is determined
by external [H+]
Constant and b are measured when electrode is calibrated with solution of known pH
Electrodes and Potentiometry
pH Electrodes
3.) Calibration
 A pH electrode should be calibrated with two or more standard buffers
before use.
 pH of the unknown should lie within the range of the standard buffers
Measured voltage is correlated
with a pH, which is then used to
measure an unknown.
Electrodes and Potentiometry
pH Electrodes
4.) Errors in pH Measurements
 Standards
- pH measurements cannot be more accurate than standards (±0.01)
 Junction potential
- If ionic strengths differ between analyte and standard buffer, junction potential
will differ resulting in an error of ±0.01
 Junction Potential Drift
- Caused by slow changes in [KCl] and [AgCl] re-calibrate!
 Sodium Error
- At very low [H+], electrode responds to Na+ and the apparent pH is lower than
the true pH
 Acid Error
- At high [H+], the measured pH is higher than actual pH, glass is saturated
 Equilibration Time
- Takes ~30s to minutes for electrode to equilibrate with solution
 Hydration of glass
- A dry electrode will not respond to H+ correctly
 Temperature
- Calibration needs to be done at same temperature of measurement
 Cleaning
- Contaminates on probe will cause reading to drift until properly cleaned or
equilibrated with analyte solution
Electrodes and Potentiometry
pH Electrodes
4.) Errors in pH Measurements
 pH measurements are accurate to ± 0.02 pH units
Larger errors occur at high
and low pH readings
Electrodes and Potentiometry
Other Ion-Selective Electrodes
1.) Solid-State Electrode
 Based on an inorganic crystal
- Fluoride electrode: LaF3 crystal doped with Eu2+
F- migrates across crystal by “jumping”
into crystal vacancies caused by Eu2+
Potential caused by charge imbalance from migrating ion across membrane


 F
(0.05916)
constant p
E 
Electrodes and Potentiometry
Other Ion-Selective Electrodes
2.) Liquid-Based Ion-Selective Electrodes
 Similar to solid-state electrode
- Hydrophobic membrane impregnated with hydrophobic ion exchanger
- Hydrophobic ion exchanger selective for analyte ion


 2
pCa
E )
2
0.05916
(
constant 
Binds Ca+2
Hydrophobic solvent
Hydrophobic counter-ion
Electrodes and Potentiometry
Other Ion-Selective Electrodes
2.) Liquid-Based Ion-Selective
Electrodes
Remember: ion-selective electrodes
create a potential from a charge
imbalance caused by analyte ion
migration across membrane
Electrodes and Potentiometry
Other Ion-Selective Electrodes
3.) Compound Electrodes
 Conventional electrode surrounded by a membrane that isolates or generates
the analyte to which the electrode responds
pH electrode surrounded by membrane
permeable to CO2.
As CO2 passes through membrane and
dissolves in solution, pH changes.
pH change is an indirect measure of CO2
concentration
Electrodes and Potentiometry
Other Ion-Selective Electrodes
4.) Standard Addition
 Corrects for analyte dissolved in complex or unknown matrix
- Blood, urine, biomass, etc
 Procedure:
1. Measure potential for unknown analyte solution
2. Add small (known) volume of a standard solution
3. Measure new potential
4. Repeat and graph data
s
s
S
/
k
x
o
S
/
k
S
/
E
s
o V
c
c
V
)
V
V
( 10
10
10 


y b m x
where:
Vo is the initial volume
Vs is the added volume
E is the measured potential
cx is the unknown concentration
cs is the standard concentration
s is a constant (RT/nF)ln10
Other Ion-Selective Electrodes
4.) Standard Addition
 Corrects for analyte dissolved in complex or unknown matrix
- Blood, urine, biomass, etc
 Procedure:
5. x-intercept yields the unknown (cx) concentration
Electrodes and Potentiometry
s
x
o
c
c
V
m
b
x 


 intercept
Only unknown

More Related Content

What's hot

Redox reaction (unit 4)
Redox reaction (unit 4)Redox reaction (unit 4)
Redox reaction (unit 4)ShwetaYadav150
 
Pharmaceutical Analysis &Errors
Pharmaceutical Analysis &ErrorsPharmaceutical Analysis &Errors
Pharmaceutical Analysis &ErrorsAMOL DIGHE
 
Conductometry (Pharmaceutical analysis)
Conductometry (Pharmaceutical analysis)Conductometry (Pharmaceutical analysis)
Conductometry (Pharmaceutical analysis)Yunesalsayadi
 
Ion selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterIon selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterMuhammad Asif Shaheeen
 
Minimization of errors, accuracy, precission, significant figures.
Minimization of errors, accuracy, precission, significant figures.Minimization of errors, accuracy, precission, significant figures.
Minimization of errors, accuracy, precission, significant figures.Nidhi Sharma
 
Potentiometry and electroanalysis basics
Potentiometry and electroanalysis basicsPotentiometry and electroanalysis basics
Potentiometry and electroanalysis basicsVishaalini Kamali
 
Electroanalytical Methods of analysis
Electroanalytical Methods of analysisElectroanalytical Methods of analysis
Electroanalytical Methods of analysisShanta Majumder
 
Complexometric titration.pptx
Complexometric titration.pptxComplexometric titration.pptx
Complexometric titration.pptxSKarthigaSVCP
 
Ion selective electrodes
Ion selective electrodesIon selective electrodes
Ion selective electrodesShweta Singh
 
Electroanalytical chemistry
Electroanalytical chemistryElectroanalytical chemistry
Electroanalytical chemistrywadhava gurumeet
 
Potentiometrynew 170227172008
Potentiometrynew 170227172008Potentiometrynew 170227172008
Potentiometrynew 170227172008Dorothy Nyamai
 

What's hot (20)

Lect. 2 DMG structure-uses-Advantages-limitation
Lect. 2 DMG  structure-uses-Advantages-limitationLect. 2 DMG  structure-uses-Advantages-limitation
Lect. 2 DMG structure-uses-Advantages-limitation
 
Conductometry
ConductometryConductometry
Conductometry
 
Complexometric Titration
Complexometric TitrationComplexometric Titration
Complexometric Titration
 
voltammetry basics.pptx
voltammetry basics.pptxvoltammetry basics.pptx
voltammetry basics.pptx
 
Redox reaction (unit 4)
Redox reaction (unit 4)Redox reaction (unit 4)
Redox reaction (unit 4)
 
Pharmaceutical Analysis &Errors
Pharmaceutical Analysis &ErrorsPharmaceutical Analysis &Errors
Pharmaceutical Analysis &Errors
 
Conductometry (Pharmaceutical analysis)
Conductometry (Pharmaceutical analysis)Conductometry (Pharmaceutical analysis)
Conductometry (Pharmaceutical analysis)
 
Polarography
PolarographyPolarography
Polarography
 
Ion selective electrode (hw)85
Ion selective electrode (hw)85Ion selective electrode (hw)85
Ion selective electrode (hw)85
 
Ion selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterIon selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meter
 
Minimization of errors, accuracy, precission, significant figures.
Minimization of errors, accuracy, precission, significant figures.Minimization of errors, accuracy, precission, significant figures.
Minimization of errors, accuracy, precission, significant figures.
 
Conductometry
ConductometryConductometry
Conductometry
 
Potentiometry and electroanalysis basics
Potentiometry and electroanalysis basicsPotentiometry and electroanalysis basics
Potentiometry and electroanalysis basics
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Electroanalytical Methods of analysis
Electroanalytical Methods of analysisElectroanalytical Methods of analysis
Electroanalytical Methods of analysis
 
Complexometric titration.pptx
Complexometric titration.pptxComplexometric titration.pptx
Complexometric titration.pptx
 
Ion selective electrodes
Ion selective electrodesIon selective electrodes
Ion selective electrodes
 
Electroanalytical chemistry
Electroanalytical chemistryElectroanalytical chemistry
Electroanalytical chemistry
 
INDICATOR ELECTRODE
INDICATOR ELECTRODEINDICATOR ELECTRODE
INDICATOR ELECTRODE
 
Potentiometrynew 170227172008
Potentiometrynew 170227172008Potentiometrynew 170227172008
Potentiometrynew 170227172008
 

Similar to Electrodes and Potentiometry.ppt

Potentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptxPotentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptxshwetasingh31652
 
Pharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BPPharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BPshwetasingh31652
 
Basics of Electrochemistry and Electrochemical Measurements
Basics of Electrochemistry and Electrochemical MeasurementsBasics of Electrochemistry and Electrochemical Measurements
Basics of Electrochemistry and Electrochemical MeasurementsHalavath Ramesh
 
POTENTIOMETRY.pdf
POTENTIOMETRY.pdfPOTENTIOMETRY.pdf
POTENTIOMETRY.pdfSaqibShaik2
 
Potentiometry, voltamemtry and conductometry
Potentiometry, voltamemtry and conductometryPotentiometry, voltamemtry and conductometry
Potentiometry, voltamemtry and conductometryapeksha40
 
potentiometry-180713090143.pdf
potentiometry-180713090143.pdfpotentiometry-180713090143.pdf
potentiometry-180713090143.pdfPriyanka Yadav
 
2_Biopotential Elektrode.pptx
2_Biopotential Elektrode.pptx2_Biopotential Elektrode.pptx
2_Biopotential Elektrode.pptxMimoCh
 
ppt uit 2 link -final - (1.1.23) - Copy.pptx
ppt uit 2 link -final - (1.1.23) - Copy.pptxppt uit 2 link -final - (1.1.23) - Copy.pptx
ppt uit 2 link -final - (1.1.23) - Copy.pptxKundanBhatkar
 
3rd.prof mam abida.pptx
3rd.prof mam abida.pptx3rd.prof mam abida.pptx
3rd.prof mam abida.pptxZORAIZ HAIDER
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finMUBOSScz
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finMUBOSScz
 
Second lect Potentiometry.ppt
Second lect Potentiometry.pptSecond lect Potentiometry.ppt
Second lect Potentiometry.pptMohamedWefqi
 
Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..vishnupriya456
 

Similar to Electrodes and Potentiometry.ppt (20)

Potentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptxPotentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptx
 
Pharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BPPharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BP
 
Potentiometry.pptx
Potentiometry.pptxPotentiometry.pptx
Potentiometry.pptx
 
Basics of Electrochemistry and Electrochemical Measurements
Basics of Electrochemistry and Electrochemical MeasurementsBasics of Electrochemistry and Electrochemical Measurements
Basics of Electrochemistry and Electrochemical Measurements
 
POTENTIOMETRY.pdf
POTENTIOMETRY.pdfPOTENTIOMETRY.pdf
POTENTIOMETRY.pdf
 
Potentiometry, voltamemtry and conductometry
Potentiometry, voltamemtry and conductometryPotentiometry, voltamemtry and conductometry
Potentiometry, voltamemtry and conductometry
 
potentiometry-180713090143.pdf
potentiometry-180713090143.pdfpotentiometry-180713090143.pdf
potentiometry-180713090143.pdf
 
2_Biopotential Elektrode.pptx
2_Biopotential Elektrode.pptx2_Biopotential Elektrode.pptx
2_Biopotential Elektrode.pptx
 
ppt uit 2 link -final - (1.1.23) - Copy.pptx
ppt uit 2 link -final - (1.1.23) - Copy.pptxppt uit 2 link -final - (1.1.23) - Copy.pptx
ppt uit 2 link -final - (1.1.23) - Copy.pptx
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
2 potentiometry
2 potentiometry2 potentiometry
2 potentiometry
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
3rd.prof mam abida.pptx
3rd.prof mam abida.pptx3rd.prof mam abida.pptx
3rd.prof mam abida.pptx
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-fin
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-fin
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Second lect Potentiometry.ppt
Second lect Potentiometry.pptSecond lect Potentiometry.ppt
Second lect Potentiometry.ppt
 
Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..
 

More from Meenakshi Dhanawat (20)

UV-Woodward Fischer Rules.pptx
UV-Woodward Fischer Rules.pptxUV-Woodward Fischer Rules.pptx
UV-Woodward Fischer Rules.pptx
 
UV
UVUV
UV
 
Infrared spectroscopy- Spectra intrepretation.pptx
Infrared spectroscopy- Spectra intrepretation.pptxInfrared spectroscopy- Spectra intrepretation.pptx
Infrared spectroscopy- Spectra intrepretation.pptx
 
Infrared Spectrometry-Introduction.pptx
Infrared Spectrometry-Introduction.pptxInfrared Spectrometry-Introduction.pptx
Infrared Spectrometry-Introduction.pptx
 
BASICS OF uv.pptx
BASICS OF uv.pptxBASICS OF uv.pptx
BASICS OF uv.pptx
 
POTENTIOMTRY.ppt
POTENTIOMTRY.pptPOTENTIOMTRY.ppt
POTENTIOMTRY.ppt
 
regulation of medical devices.pptx
regulation of medical devices.pptxregulation of medical devices.pptx
regulation of medical devices.pptx
 
PRIMARY AND SECONDARY.pptx
PRIMARY AND SECONDARY.pptxPRIMARY AND SECONDARY.pptx
PRIMARY AND SECONDARY.pptx
 
Introduction to Analytical Chemistry.pptx
Introduction to Analytical Chemistry.pptxIntroduction to Analytical Chemistry.pptx
Introduction to Analytical Chemistry.pptx
 
Electrophoresis.pptx
Electrophoresis.pptxElectrophoresis.pptx
Electrophoresis.pptx
 
Affinity chromatography.pptx
Affinity chromatography.pptxAffinity chromatography.pptx
Affinity chromatography.pptx
 
PROBLEMS WITH TLC.pptx
PROBLEMS WITH TLC.pptxPROBLEMS WITH TLC.pptx
PROBLEMS WITH TLC.pptx
 
Good Documentation Practice.pptx
Good Documentation Practice.pptxGood Documentation Practice.pptx
Good Documentation Practice.pptx
 
IR.ppt
IR.pptIR.ppt
IR.ppt
 
Medicinal Gases.pptx
Medicinal Gases.pptxMedicinal Gases.pptx
Medicinal Gases.pptx
 
METHODS OF IONIZATION-MASS.pptx
METHODS OF IONIZATION-MASS.pptxMETHODS OF IONIZATION-MASS.pptx
METHODS OF IONIZATION-MASS.pptx
 
Essential Trace elements.pptx
Essential Trace elements.pptxEssential Trace elements.pptx
Essential Trace elements.pptx
 
antacid.ppt
antacid.pptantacid.ppt
antacid.ppt
 
Theory of Indicators.pptx
Theory of Indicators.pptxTheory of Indicators.pptx
Theory of Indicators.pptx
 
Thermogravimetry
ThermogravimetryThermogravimetry
Thermogravimetry
 

Recently uploaded

Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...
Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...
Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...call girls in ahmedabad high profile
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatorenarwatsonia7
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...Miss joya
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 

Recently uploaded (20)

Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...
Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...
Call Girls Near Hotel Marine Plaza ✔ 9820252231 ✔For 18+ VIP Call Girl At The...
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 

Electrodes and Potentiometry.ppt

  • 2. Electrodes and Potentiometry Introduction 1.) Potentiometry  Use of Electrodes to Measure Voltages that Provide Chemical Information - Various electrodes have been designed to respond selectively to specific analytes  Use a Galvanic Cell - Unknown solution becomes a ½-cell - Add Electrode that transfers/accepts electrons from unknown analyte - Connect unknown solution by salt bridge to second ½-cell at fixed composition and potential  Indicator Electrode: electrode that responds to analyte and donates/accepts electrons  Reference Electrode: second ½ cell at a constant potential  Cell voltage is difference between the indicator and reference electrode
  • 3. Binding generates potential difference. Electrodes and Potentiometry Introduction 2.) Example  A Heparin Sensor - Voltage response is proportional to heparin concentration in blood - Sensor is selective for heparin Negatively charged heparin binds selectively to positively charged membrane. heparin Potential is proportional to [heparin]
  • 4. Electrodes and Potentiometry Reference Electrodes 1.) Overview  Potential change only dependent on one ½ cell concentrations  Reference electrode is fixed or saturated  doesn’t change! Reference electrode, [Cl-] is constant Potential of the cell only depends on [Fe2+] & [Fe3+] Pt wire is indicator electrode whose potential responds to [Fe2+]/[Fe3+]   ] Cl log[ . . ] Fe [ ] Fe [ log . . Ecell                          05916 0 222 0 1 05916 0 771 0 3 2 Unknown solution of [Fe2+] & [Fe3+]
  • 5. Electrodes and Potentiometry Reference Electrodes 2.) Silver-Silver Chloride Reference Electrode  Convenient - Common problem is porous plug becomes clogged Eo = +0.222 V Activity of Cl- not 1E(sat,KCl) = +0.197 V
  • 6. Electrodes and Potentiometry Reference Electrodes 3.) Saturated Calomel Reference Electrode (S.C.E)  Saturated KCl maintains constant [Cl-] even with some evaporation  Standard hydrogen electrodes are cumbersome - Requires H2 gas and freshly prepared Pt surface Eo = +0.268 V Activity of Cl- not 1E(sat,KCl) = +0.241 V
  • 7. Electrodes and Potentiometry Reference Electrodes 4.) Observed Voltage is Reference Electrode Dependant  The observed potential depends on the choice of reference electrode - Silver-silver chloride and calomel have different potentials  Use Reference Scale to convert between Reference Electrodes Observed potential relative to SCE Observed potential relative to Ag|AgCl Observed potential relative to SHE
  • 8. Electrodes and Potentiometry Junction Potential 1.) Occurs Whenever Dissimilar Electrolyte Solutions are in Contact  Develops at solution interface (salt bridge)  Small potential (few millivolts)  Junction potential puts a fundamental limitation on the accuracy of direct potentiometric measurements - Don’t know contribution to the measured voltage Again, an electric potential is generated by a separation of charge Different ion mobility results in separation in charge
  • 9. Electrodes and Potentiometry Indicator Electrodes 1.) Two Broad Classes of Indicator Electrodes  Metal Electrodes - Develop an electric potential in response to a redox reaction at the metal surface  Ion-selective Electrodes - Selectively bind one type of ion to a membrane to generate an electric potential Remember an electric potential is generated by a separation of charge
  • 10. Electrodes and Potentiometry Indicator Electrodes 2.) Metal Electrodes  Platinum - Most common metal indicator electrode - Inert: does not participate in many chemical reactions - Simply used to transmit electrons  Other electrodes include Gold and Carbon  Metals (Ag, Cu, Zn, Cd, Hg) can be used to monitor their aqueous ions - Most metals are not useable - Equilibrium not readily established at the metal surface E+ o = +799 V E(sat,KCl) = +0.241 V ½ Reaction at Ag indicator electrode: ½ Reaction at Calomel reference electrode:   241 0 1 1 05916 0 799 0 . ] Ag [ log . . E E Ecell                           Potential of Ag indicator electrode Cell voltage changes as a function of [Ag+] Example: Cell Potential from Nernst Equation:
  • 11. Electrodes and Potentiometry Indicator Electrodes 3.) Example  A 10.0 mL solution of 0.0500 M AgNO3 was titrated with 0.0250M NaBr in the cell: S.C.E. || titration solution | Ag(s) Find the cell voltage for 10.0 mL of titrant
  • 12. C+ diffuses across the membrane due to concentration gradient resulting in charge difference across membrane Electrodes and Potentiometry Indicator Electrodes 4.) Ion-Selective Electrodes  Responds Selectively to one ion - Contains a thin membrane capable of only binding the desired ion  Does not involve a redox process Membrane contains a ligand (L) that specifically and tightly binds analyte of interest (C+) A difference in the concentration of C+ exists across the outer membrane. The counter-ions (R-,A-) can’t cross the membrane and/or have low solubility in membrane or analyte solution Remember an electric potential is generated by a separation of charge Potential across outer membrane depends on [C+] in analyte solution
  • 13. Electrodes and Potentiometry Indicator Electrodes 4.) Ion-Selective Electrodes  Responds Selectively to one ion - Contains a thin membrane capable of only binding the desired ion  Does not involve a redox process Remember an electric potential is generated by a separation of charge Potential across inner membrane depends on [C+] in filling solution, which is a known constant C+ diffuses across the membrane due to concentration gradient resulting in charge difference across membrane A difference in the concentration of C+ exists across the inner membrane. inner outer E E E   Electrode potential is determined by the potential difference between the inner and outer membranes: where Einner is a constant and Eouter depends on the concentration of C+ in analyte solution
  • 14. ] [ constant    C log n . E 05916 0 Electrodes and Potentiometry Indicator Electrodes 4.) Ion-Selective Electrodes  Responds Selectively to one ion - Contains a thin membrane capable of only binding the desired ion  Does not involve a redox process Electrode Potential is defined as: where [C+] is actually the activity of the analyte and n is the charge of the analyte
  • 15. Electrodes and Potentiometry pH Electrodes 1.) pH Measurement with a Glass Electrode  Glass electrode is most common ion-selective electrode  Combination electrode incorporates both glass and reference electrode in one body Ag(s)|AgCl(s)|Cl-(aq)||H+(aq,outside) H+(aq,inside),Cl-(aq)|AgCl(s)|Ag(s) Outer reference electrode [H+] outside (analyte solution) [H+] inside Inner reference electrode Glass membrane Selectively binds H+ Electric potential is generated by [H+] difference across glass membrane
  • 16. Electrodes and Potentiometry pH Electrodes 2.) Glass Membrane  Irregular structure of silicate lattice Cations (Na+) bind oxygen in SiO4 structure
  • 17. Electrodes and Potentiometry pH Electrodes 2.) Glass Membrane  Two surfaces of glass “swell” as they absorb water - Surfaces are in contact with [H+]
  • 18. Electrodes and Potentiometry pH Electrodes 2.) Glass Membrane  H+ diffuse into glass membrane and replace Na+ in hydrated gel region - Ion-exchange equilibrium - Selective for H+ because H+ is only ion that binds significantly to the hydrated gel layer H (0.05916) constant p E    Charge is slowly carried by migration of Na+ across glass membrane Potential is determined by external [H+] Constant and b are measured when electrode is calibrated with solution of known pH
  • 19. Electrodes and Potentiometry pH Electrodes 3.) Calibration  A pH electrode should be calibrated with two or more standard buffers before use.  pH of the unknown should lie within the range of the standard buffers Measured voltage is correlated with a pH, which is then used to measure an unknown.
  • 20. Electrodes and Potentiometry pH Electrodes 4.) Errors in pH Measurements  Standards - pH measurements cannot be more accurate than standards (±0.01)  Junction potential - If ionic strengths differ between analyte and standard buffer, junction potential will differ resulting in an error of ±0.01  Junction Potential Drift - Caused by slow changes in [KCl] and [AgCl] re-calibrate!  Sodium Error - At very low [H+], electrode responds to Na+ and the apparent pH is lower than the true pH  Acid Error - At high [H+], the measured pH is higher than actual pH, glass is saturated  Equilibration Time - Takes ~30s to minutes for electrode to equilibrate with solution  Hydration of glass - A dry electrode will not respond to H+ correctly  Temperature - Calibration needs to be done at same temperature of measurement  Cleaning - Contaminates on probe will cause reading to drift until properly cleaned or equilibrated with analyte solution
  • 21. Electrodes and Potentiometry pH Electrodes 4.) Errors in pH Measurements  pH measurements are accurate to ± 0.02 pH units Larger errors occur at high and low pH readings
  • 22. Electrodes and Potentiometry Other Ion-Selective Electrodes 1.) Solid-State Electrode  Based on an inorganic crystal - Fluoride electrode: LaF3 crystal doped with Eu2+ F- migrates across crystal by “jumping” into crystal vacancies caused by Eu2+ Potential caused by charge imbalance from migrating ion across membrane    F (0.05916) constant p E 
  • 23. Electrodes and Potentiometry Other Ion-Selective Electrodes 2.) Liquid-Based Ion-Selective Electrodes  Similar to solid-state electrode - Hydrophobic membrane impregnated with hydrophobic ion exchanger - Hydrophobic ion exchanger selective for analyte ion    2 pCa E ) 2 0.05916 ( constant  Binds Ca+2 Hydrophobic solvent Hydrophobic counter-ion
  • 24. Electrodes and Potentiometry Other Ion-Selective Electrodes 2.) Liquid-Based Ion-Selective Electrodes Remember: ion-selective electrodes create a potential from a charge imbalance caused by analyte ion migration across membrane
  • 25. Electrodes and Potentiometry Other Ion-Selective Electrodes 3.) Compound Electrodes  Conventional electrode surrounded by a membrane that isolates or generates the analyte to which the electrode responds pH electrode surrounded by membrane permeable to CO2. As CO2 passes through membrane and dissolves in solution, pH changes. pH change is an indirect measure of CO2 concentration
  • 26. Electrodes and Potentiometry Other Ion-Selective Electrodes 4.) Standard Addition  Corrects for analyte dissolved in complex or unknown matrix - Blood, urine, biomass, etc  Procedure: 1. Measure potential for unknown analyte solution 2. Add small (known) volume of a standard solution 3. Measure new potential 4. Repeat and graph data s s S / k x o S / k S / E s o V c c V ) V V ( 10 10 10    y b m x where: Vo is the initial volume Vs is the added volume E is the measured potential cx is the unknown concentration cs is the standard concentration s is a constant (RT/nF)ln10
  • 27. Other Ion-Selective Electrodes 4.) Standard Addition  Corrects for analyte dissolved in complex or unknown matrix - Blood, urine, biomass, etc  Procedure: 5. x-intercept yields the unknown (cx) concentration Electrodes and Potentiometry s x o c c V m b x     intercept Only unknown