3. After setting the high voltage for both PMTs, we can begin making transparency
measurements. Transparency is the amount of light passing through the sample (RS) divided by
the amount of light passing through an empty holder (RE).
. (1)
We measure the light passing through the empty holder as the ratio of PMTA/PMTB pulse
heights.
. (2)
We then measure the light passing through a sample as the ratio of PMTA/PMTB pulse
heights.
. (3)
Taking the ratio of the sample divided by the ratio of the empty holder results in the
transparency of the optical cement. Care must be taken not to alter the alignment of the setup
during a measurement to assure consistent results.
We evaluate optical cement from two manufactures. The BC-600 from St.-Gobain
Crystals [4] and the EJ-500 from Eljen Technologies [5]. Both are two component adhesives.
When preparing the adhesives, one must keep in mind not to stir too quickly to prevent air
bubbles. Once the optical cement is mixed well, we wait approximately 20 minuets to let air
bubbles rise to the surface. After the bubbles have risen, the cement was poured into a ½ inch
long, ¾ inch section of clear PVC pipe and set aside to cure. After a few days, we removed the
samples from the molds and had their optical surface diamond polished. A few imperfections
such as air bubbles and swirls caused by the difference in index of refraction from incomplete
mixing of cements remained.
Optical Transparency Measurements
Figure 3 shows the graphs of the light transmission of the samples as a function of the
orientation of the sample. The left graph is for the BC-600. The right graph is for the EJ-500. We
evaluated the variation in transmission due to these defects by rotating the sample in the
holder. An initial angle was set by a mark on the sample. The figure shows that most of the
optical cements have an approximately 80% transmission. About 10% of the light was lost due
to imperfections in the optical cement. The remaining 10% was calculated. We calculated the
amount of light that should be reflected due to the index of reflection. BC-600 should reflect
around 10.37% of the light passing through and the EJ-500 should reflect about 9.83%.
Neglecting the reflection at the optical surfaces, ~90% of the light is transmitted though each
sample.