Lecture 10: Nuclear Imaging-II
Shahid Younas
NUCLEAR IMAGING
The Scintillation Camera- Design Performance
Gamma Camera โ€“Collimators
Lecture 10: Nuclear Imaging-II
Which of the following statements best describes the primary purpose of a
collimator on a gamma camera?
a. It prevents scattered photons from reaching the detector.
b. It prevents cosmic radiation from reaching the detector.
c. It stops pre-detector scattered photons.
Gamma Camera โ€“Collimators
Lecture 10: Nuclear Imaging-II
๏ถ To allow photons from a given region of interest to strike the
detector and try to minimize the contribution of photons
originating from outside this region.
Gamma Camera โ€“Collimators
Lecture 10: Nuclear Imaging-II
๏ถ Most scintillation cameras are provided with a selection of
parallel-hole collimators:
๏ƒผ โ€œlow-energy, high-sensitivityโ€
๏ƒผ โ€œlow-energy, all-purposeโ€ (LEAP)
๏ƒผ โ€œlow-energy, high-resolutionโ€ (LEHR)
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
If a low-energy collimator is (incorrectly) used with a high-energy
radionuclide the results would be:
a. A reduced camera sensitivity (counting efficiency).
b. A blurred image.
c. A reduced field of view.
d. Reduced image detail.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
๏ถ Size of the image produced by parallel-hole collimator not
affected by distance of object from collimator.
๏ถ Spatial resolution degrades rapidly with increasing collimator-to-
object distance.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
The principal disadvantage in using a high resolution collimator on a
gamma camera is that it has,
a. Limited field of view.
b. More distortion.
c. Less scatter rejection.
d. Lower sensitivity.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
๏ถ R is an indicator of spatial
resolution.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
๏ถ Collimator spatial resolution of multi-hole collimators is
determined by geometry of the holes .
๏ถ Spatial resolution improves as the,
๏ƒผ Diameters of the holes is reduced
๏ƒผ Lengths of the holes (thickness of the collimator) are increased
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
๏ถ Changing hole geometry to improve spatial resolution generally
reduces the collimatorโ€™s efficiency.
๏ƒ˜ Resultant compromise is single most significant limitation on
scintillation camera performance.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
๏ถ Spatial resolution of parallel-hole collimator decreases linearly as
collimator-to-object distance increases.
๏ƒ˜ Also one of the most important factors limiting scintillation
camera performance.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
Do you know why the efficiency of a parallel-hole collimator is
nearly constant over the collimator-to-object distances used for
clinical imaging.
Number of photons passing through given hole decreases with square
of distance, number of holes through which photons can pass
increases with square of distance.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
๏ถ FWHM of LSF increased
linearly with distance.
๏ถ Total area under LSF
(photon fluence through
collimator) decreases
very little with s-c
distance.
Parallel-hole LEAP collimator
Lecture 10: Nuclear Imaging-II
๏ถ LEAP collimators have holes with a large diameter.
๏ถ The sensitivity is relatively high as where the resolution is moderate.
(larger diameter holes allow more scattered photons).
๏ถ The average sensitivity of a LEAP is approx. 500 kcpm for a 1-uCi
source, and the resolution is 1.0cm at 10cm from the patent side of the
collimator.
Parallel-hole LEHR collimator
Lecture 10: Nuclear Imaging-II
๏ถ LEHR collimators have higher resolution images than the LEAP.
๏ถ They have more holes that are both smaller and deeper.
๏ถ The sensitivity is approx. 185 kcpm for 1-uCi source, and the resolution
is higher with 0.65cm at 10cm from the patient side of the collimator.
Parallel-hole collimator
Lecture 10: Nuclear Imaging-II
Which is the ideal radionuclide to use with LEAP and LEHR?
99mTc
Parallel-hole Medium Energy Collimator
Lecture 10: Nuclear Imaging-II
๏ถ Medium Energy Collimators are used for medium energy photons of
nuclides such as Krypton81, Gallium67, Indium111. These collimators
have thinner septa than LEAP and LEHR collimators.
Parallel-hole Medium Energy Collimator
Lecture 10: Nuclear Imaging-II
๏ถ High Energy Collimators are used for Iodine131 and F-18FDG. These
collimators have thicker septa than LEAP and LEHR collimators in
order to reduce septal penetration by the higher energy photons.
Parallel-Slant -hole Collimator
Lecture 10: Nuclear Imaging-II
๏ถ A variation of the Parallel hole which has all tunnels slanted at a specific
angle.
๏ถ It generates an oblique view for better visualization of an organ, which
view is (partly) blocked by other parts of the body.
๏ถ As an advantage, this collimator can be positioned close to the body for
the maximum gain in resolution.
Pinhole collimator
Lecture 10: Nuclear Imaging-II
๏ถ Used to produce magnified views of small
objects, such as thyroid or hip joint.
๏ถ Consists of small (typically 3- to 5-mm
diameter) hole in a piece of lead or
tungsten.
๏ถ Mounted at apex of a leaded cone.
Pinhole collimator
Lecture 10: Nuclear Imaging-II
๏ถ Produces a magnified image whose
orientation is reversed.
๏ถ Magnification decreases as object is
moved away from pinhole.
Pinhole collimator
Lecture 10: Nuclear Imaging-II
๏ถ If the object is as far from the
pinhole as the pinhole is from the
crystal of the camera, the object is
not magnified.
Pinhole collimator
Lecture 10: Nuclear Imaging-II
๏ถ This creates pitfall in clinical usage.
๏ถ A thyroid nodule deep in the mediastinum can appear to be
in the thyroid itself.
๏ถ Extensive use is in pediatric nuclear medicine.
Converging collimator
Lecture 10: Nuclear Imaging-II
๏ถ In a Converging collimator the holes are
not parallel but focused toward the organ.
๏ถ The focal point is normally located in the
center of the field of view (FOV).
Converging collimator
Lecture 10: Nuclear Imaging-II
๏ถ Magnifies the image.
๏ถ Magnification increases as the object is
moved away from the collimator.
Converging collimator
Lecture 10: Nuclear Imaging-II
๏ถ Converging collimator is seldom used.
๏ถ Imaging characteristics are superior, in theory, to the parallel-
hole collimators but in practice owing to,
๏ƒ˜ Decreasing FOV and varying magnification with distance
Diverging collimator
Lecture 10: Nuclear Imaging-II
๏ถ Many holes, all aimed a focal point behind
the camera.
๏ถ Produces a minified image.
๏ถ Amount of mini-fication increases as
object is moved away from the collimator
Diverging collimator
Lecture 10: Nuclear Imaging-II
๏ถ May be used to image a large portion of a patient on a small (25-cm
diameter) or standard (30-cm diameter) FOV camera.
๏ถ You can perform lung study using a mobile gamma camera in
intensive care unit.
Diverging collimator
Lecture 10: Nuclear Imaging-II
๏ถ Compared to a diverging collimator, a converging collimator will
produce:
a. A increase in sensitivity when distance is increased.
b. Better image detail.
c. A reduced FOV as distance is increased.
Diverging collimator
Lecture 10: Nuclear Imaging-II
What collimator will you get if you reverse a diverging collimator on
a camera?
Converging
Diverging collimator
Lecture 10: Nuclear Imaging-II
๏ถ Its seldom used because of,
๏ƒ˜ Inferior imaging characteristics to the parallel hole
๏ƒ˜ Large crystal sizes of modern cameras.
Fan Beam collimator
Lecture 10: Nuclear Imaging-II
๏ถ They are designed for a rectangular
camera head or SPECT to image smaller
organs like the brain and heart.
๏ถ When viewed from one direction, the
holes are parallel. When viewed from the
other direction, the holes converge.
Itโ€™s a hybrid of parallel-hole
and converging collimator.
collimator
Lecture 10: Nuclear Imaging-II
Image magnification is changed if you change,
A. Radionuclide
B. Imaging time
C. Type of collimator
D. Patient-collimator distance
E. PHA window level
Image formation
Lecture 10: Nuclear Imaging-II
๏ถ Photons from each point in the patient are emitted isotropically.
๏ถ Some photons escape the patient without interaction, some scatter
within the patient before escaping, and some are absorbed in the
patient
Image formation
Lecture 10: Nuclear Imaging-II
๏ถ Many of the escaping photons are not detected because they are
emitted in directions away from the detector.
๏ถ Only a tiny fraction of emitted photons has trajectories permitting
passage through the collimator holes.
Image formation
Lecture 10: Nuclear Imaging-II
๏ถ Of those reaching the crystal, some are absorbed in the crystal,
some scatter from the crystal, and some pass through the crystal
without interaction
Image formation
Lecture 10: Nuclear Imaging-II
๏ƒ˜ Relative probabilities
of these events
depends on the
energies of the
photons and the
thickness of the crystal
Image formation
Lecture 10: Nuclear Imaging-II
๏ถ Of those photons absorbed in the crystal, some are absorbed by a
single photoelectric absorption, others undergo one or more
Compton scatters before a photoelectric absorption.
Image formation
Lecture 10: Nuclear Imaging-II
Is it possible for two photons to simultaneously interact with the
crystal?
Image formation
Lecture 10: Nuclear Imaging-II
๏ƒ˜ If the energy signal (Z) from the coincident interactions is within
the energy window of the energy discrimination circuit, the result
will be a single count mis-positioned in the image.
๏ถ Fraction of simultaneous interactions increases with the interaction
rate of the camera.
Image formation
Lecture 10: Nuclear Imaging-II
๏ถ Spatial resolution and image contrast reduced by:
๏ƒ˜ Interactions in crystal of photons that have been scattered in the
patient.
๏ƒ˜ Photons that have penetrated the collimator septa.
๏ƒ˜ Photons that undergo one or more scatters in the crystal.
๏ƒ˜ Coincident interactions.
Image formation
Lecture 10: Nuclear Imaging-II
Energy discrimination circuits reduce this by rejecting photons that
scatter in the patient or result in coincident interactions.
Image formation
Lecture 10: Nuclear Imaging-II
If the PHA window on a gamma camera is (incorrectly) set below the
photo-peak energy you would expect to get:
๏ƒผ Decreased sensitivity.
๏ƒผ An image of primarily scattered radiation.
๏ƒผ Decreased lesion contrast.
Image formation
Lecture 10: Nuclear Imaging-II
๏ถ Pulse Height Analyzer (PHA) reduce this loss of resolution and
contrast by rejecting scattered photons.
๏ถ However, low energy photons can scatter through large angles with
only a small energy loss.
Image formation
Lecture 10: Nuclear Imaging-II
๏ถ 140 keV photon scattering 45 degree will only lose 7.4% of its
energy.
๏ถ This causes a wide photo-peak because PHA accepts most of the
significant fraction of the scattered photons and coincident
interactions.
Image formation
Lecture 10: Nuclear Imaging-II
The full width at half maximum (FWHM) of a photo-peak is a
measure of,
a. PHA window setting.
b. Camera sensitivity.
c. Field of view.
d. Detector energy resolution.
Image formation
Lecture 10: Nuclear Imaging-II
Performance

More Related Content

What's hot

Ct Generations
Ct  GenerationsCt  Generations
Ct GenerationsDr. Mohit Goel
ย 
Nuclear Medicine
Nuclear MedicineNuclear Medicine
Nuclear MedicineSajad Buch
ย 
Radionuclide Production - Nuclear Medicine
Radionuclide Production - Nuclear MedicineRadionuclide Production - Nuclear Medicine
Radionuclide Production - Nuclear Medicine@Saudi_nmc
ย 
What is Nuclear Medicine and PET?
What is Nuclear Medicine and PET?What is Nuclear Medicine and PET?
What is Nuclear Medicine and PET?todd_charge
ย 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionparthajyotidas11
ย 
Digital Radiography
Digital RadiographyDigital Radiography
Digital Radiographysantosh neupane
ย 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkDr pradeep Kumar
ย 
Nuclear medicine 2
Nuclear medicine 2Nuclear medicine 2
Nuclear medicine 2Rad Tech
ย 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxThejaTej6
ย 
CT Artifacts
CT ArtifactsCT Artifacts
CT ArtifactsKyle Rousseau
ย 
Computed radiography
Computed radiographyComputed radiography
Computed radiographyAnjan Dangal
ย 
Nuclear medicine
Nuclear medicineNuclear medicine
Nuclear medicineShrikant Nagare
ย 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentationMumba Chilimboyi
ย 
Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiographyDurga Singh
ย 
Linear Accelerators
Linear AcceleratorsLinear Accelerators
Linear AcceleratorsVelliangiri P
ย 

What's hot (20)

Ct Generations
Ct  GenerationsCt  Generations
Ct Generations
ย 
Nuclear Medicine
Nuclear MedicineNuclear Medicine
Nuclear Medicine
ย 
Radionuclide Production - Nuclear Medicine
Radionuclide Production - Nuclear MedicineRadionuclide Production - Nuclear Medicine
Radionuclide Production - Nuclear Medicine
ย 
What is Nuclear Medicine and PET?
What is Nuclear Medicine and PET?What is Nuclear Medicine and PET?
What is Nuclear Medicine and PET?
ย 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
ย 
Digital Radiography
Digital RadiographyDigital Radiography
Digital Radiography
ย 
CT Generations and Artefacts
CT Generations and ArtefactsCT Generations and Artefacts
CT Generations and Artefacts
ย 
CT artifact
CT artifact CT artifact
CT artifact
ย 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
ย 
Nuclear medicine 2
Nuclear medicine 2Nuclear medicine 2
Nuclear medicine 2
ย 
Linear Accelerator
Linear AcceleratorLinear Accelerator
Linear Accelerator
ย 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
ย 
Ct instrument
Ct instrumentCt instrument
Ct instrument
ย 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
ย 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
ย 
Nuclear medicine
Nuclear medicineNuclear medicine
Nuclear medicine
ย 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
ย 
Fluoroscopy
 Fluoroscopy  Fluoroscopy
Fluoroscopy
ย 
Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiography
ย 
Linear Accelerators
Linear AcceleratorsLinear Accelerators
Linear Accelerators
ย 

Similar to Nuclear imaging PET CT Imaging Medical Physics Nuclear Medicine

flouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdfflouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdfVishnuDutt40
ย 
basic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdfbasic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdfnaima SENHOU
ย 
Flouroscopic imaging
Flouroscopic imagingFlouroscopic imaging
Flouroscopic imagingRuta Angel
ย 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imgingSahith Reddy
ย 
Nuclear imaging
Nuclear imagingNuclear imaging
Nuclear imagingShahid Younas
ย 
flouroscopy.pptx
flouroscopy.pptxflouroscopy.pptx
flouroscopy.pptxMdSohailAhmed7
ย 
Fluoroscopy presentation1
Fluoroscopy presentation1Fluoroscopy presentation1
Fluoroscopy presentation1Huzaifa Oxford
ย 
Traineeship Melbourne University - Michael Beljaars
Traineeship Melbourne University - Michael BeljaarsTraineeship Melbourne University - Michael Beljaars
Traineeship Melbourne University - Michael BeljaarsMichael Beljaars
ย 
Fluroscopy.pptx
Fluroscopy.pptxFluroscopy.pptx
Fluroscopy.pptxMahrukhMunawar1
ย 
Medical x ray image sensors
Medical x ray image sensorsMedical x ray image sensors
Medical x ray image sensorsIslam Kotb Ismail
ย 
Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxRohit Bansal
ย 
Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)Ravi Shekhar
ย 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentationHuzaifa Oxford
ย 
D&euv lithography final
D&euv lithography finalD&euv lithography final
D&euv lithography finalZaahir Salam
ย 
Atomic Force Microscope
Atomic Force MicroscopeAtomic Force Microscope
Atomic Force MicroscopeGhalia Nawal
ย 
Optical measurements
Optical measurementsOptical measurements
Optical measurementsAndrei Tsiboulia
ย 
Transmission Electron Microscope
Transmission Electron MicroscopeTransmission Electron Microscope
Transmission Electron Microscope20was
ย 

Similar to Nuclear imaging PET CT Imaging Medical Physics Nuclear Medicine (20)

flouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdfflouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdf
ย 
Lecture 12
Lecture 12Lecture 12
Lecture 12
ย 
basic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdfbasic of angiography physics and equipement.pdf
basic of angiography physics and equipement.pdf
ย 
Flouroscopic imaging
Flouroscopic imagingFlouroscopic imaging
Flouroscopic imaging
ย 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
ย 
Nuclear imaging
Nuclear imagingNuclear imaging
Nuclear imaging
ย 
final ppt
final pptfinal ppt
final ppt
ย 
Night vision technology by shudhanshu agarwal
Night vision technology by shudhanshu agarwalNight vision technology by shudhanshu agarwal
Night vision technology by shudhanshu agarwal
ย 
flouroscopy.pptx
flouroscopy.pptxflouroscopy.pptx
flouroscopy.pptx
ย 
Fluoroscopy presentation1
Fluoroscopy presentation1Fluoroscopy presentation1
Fluoroscopy presentation1
ย 
Traineeship Melbourne University - Michael Beljaars
Traineeship Melbourne University - Michael BeljaarsTraineeship Melbourne University - Michael Beljaars
Traineeship Melbourne University - Michael Beljaars
ย 
Fluroscopy.pptx
Fluroscopy.pptxFluroscopy.pptx
Fluroscopy.pptx
ย 
Medical x ray image sensors
Medical x ray image sensorsMedical x ray image sensors
Medical x ray image sensors
ย 
Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptx
ย 
Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)
ย 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentation
ย 
D&euv lithography final
D&euv lithography finalD&euv lithography final
D&euv lithography final
ย 
Atomic Force Microscope
Atomic Force MicroscopeAtomic Force Microscope
Atomic Force Microscope
ย 
Optical measurements
Optical measurementsOptical measurements
Optical measurements
ย 
Transmission Electron Microscope
Transmission Electron MicroscopeTransmission Electron Microscope
Transmission Electron Microscope
ย 

More from Shahid Younas

Skm symps poster 2015 spect ct
Skm symps poster 2015 spect ctSkm symps poster 2015 spect ct
Skm symps poster 2015 spect ctShahid Younas
ย 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteShahid Younas
ย 
Nuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICSNuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICSShahid Younas
ย 
2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformation2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformationShahid Younas
ย 
Radioactivity and nuclear transformation
Radioactivity and nuclear transformationRadioactivity and nuclear transformation
Radioactivity and nuclear transformationShahid Younas
ย 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matterShahid Younas
ย 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matterShahid Younas
ย 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement Shahid Younas
ย 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement Shahid Younas
ย 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement Shahid Younas
ย 
1 radiation detection and measurement
1 radiation detection and measurement 1 radiation detection and measurement
1 radiation detection and measurement Shahid Younas
ย 
L 7 ct physics
L 7 ct physics L 7 ct physics
L 7 ct physics Shahid Younas
ย 
L 6 ct physics
L 6 ct physicsL 6 ct physics
L 6 ct physicsShahid Younas
ย 
L 5 ct physics
L 5 ct physicsL 5 ct physics
L 5 ct physicsShahid Younas
ย 
L 4 ct physics
L 4  ct physicsL 4  ct physics
L 4 ct physicsShahid Younas
ย 
L 3 ct physics, Shahid Younas, Sr. Medical Physicist
L 3  ct physics, Shahid Younas, Sr. Medical PhysicistL 3  ct physics, Shahid Younas, Sr. Medical Physicist
L 3 ct physics, Shahid Younas, Sr. Medical PhysicistShahid Younas
ย 
L 2 ct physics
L 2  ct physics L 2  ct physics
L 2 ct physics Shahid Younas
ย 
L 1 ct physics
L 1  ct physicsL 1  ct physics
L 1 ct physicsShahid Younas
ย 

More from Shahid Younas (18)

Skm symps poster 2015 spect ct
Skm symps poster 2015 spect ctSkm symps poster 2015 spect ct
Skm symps poster 2015 spect ct
ย 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT Constitute
ย 
Nuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICSNuclear imaging, PET CT MEDICAL PHYSICS
Nuclear imaging, PET CT MEDICAL PHYSICS
ย 
2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformation2 radioactivity and nuclear transformation
2 radioactivity and nuclear transformation
ย 
Radioactivity and nuclear transformation
Radioactivity and nuclear transformationRadioactivity and nuclear transformation
Radioactivity and nuclear transformation
ย 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matter
ย 
1 interaction of radiation with matter
1 interaction of radiation with matter1 interaction of radiation with matter
1 interaction of radiation with matter
ย 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement
ย 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement
ย 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement
ย 
1 radiation detection and measurement
1 radiation detection and measurement 1 radiation detection and measurement
1 radiation detection and measurement
ย 
L 7 ct physics
L 7 ct physics L 7 ct physics
L 7 ct physics
ย 
L 6 ct physics
L 6 ct physicsL 6 ct physics
L 6 ct physics
ย 
L 5 ct physics
L 5 ct physicsL 5 ct physics
L 5 ct physics
ย 
L 4 ct physics
L 4  ct physicsL 4  ct physics
L 4 ct physics
ย 
L 3 ct physics, Shahid Younas, Sr. Medical Physicist
L 3  ct physics, Shahid Younas, Sr. Medical PhysicistL 3  ct physics, Shahid Younas, Sr. Medical Physicist
L 3 ct physics, Shahid Younas, Sr. Medical Physicist
ย 
L 2 ct physics
L 2  ct physics L 2  ct physics
L 2 ct physics
ย 
L 1 ct physics
L 1  ct physicsL 1  ct physics
L 1 ct physics
ย 

Recently uploaded

Most Beautiful Call Girl in Chennai 7427069034 Contact on WhatsApp
Most Beautiful Call Girl in Chennai 7427069034 Contact on WhatsAppMost Beautiful Call Girl in Chennai 7427069034 Contact on WhatsApp
Most Beautiful Call Girl in Chennai 7427069034 Contact on WhatsAppjimmihoslasi
ย 
Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...
Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...
Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...Angel
ย 
tongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacytongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacyDrMohamed Assadawy
ย 
Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...
Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...
Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...Sheetaleventcompany
ย 
๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...
๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...
๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...soniya pandit
ย 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxSwetaba Besh
ย 
๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...
๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...
๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...Sheetaleventcompany
ย 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotecjualobat34
ย 
๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...
๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...
๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...gragneelam30
ย 
Kolkata Call Girls Naktala ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ Top Class Call Girl Se...
Kolkata Call Girls Naktala  ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ  Top Class Call Girl Se...Kolkata Call Girls Naktala  ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ  Top Class Call Girl Se...
Kolkata Call Girls Naktala ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ Top Class Call Girl Se...Namrata Singh
ย 
Electrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfElectrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfMedicoseAcademics
ย 
Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...Sheetaleventcompany
ย 
Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...dishamehta3332
ย 
๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...
๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...
๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...Sheetaleventcompany
ย 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan 087776558899
ย 
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...GENUINE ESCORT AGENCY
ย 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxSwetaba Besh
ย 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...dishamehta3332
ย 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsMedicoseAcademics
ย 
Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...Sheetaleventcompany
ย 

Recently uploaded (20)

Most Beautiful Call Girl in Chennai 7427069034 Contact on WhatsApp
Most Beautiful Call Girl in Chennai 7427069034 Contact on WhatsAppMost Beautiful Call Girl in Chennai 7427069034 Contact on WhatsApp
Most Beautiful Call Girl in Chennai 7427069034 Contact on WhatsApp
ย 
Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...
Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...
Bandra East [ best call girls in Mumbai Get 50% Off On VIP Escorts Service 90...
ย 
tongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacytongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacy
ย 
Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...
Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...
Chandigarh Call Girls Service โค๏ธ๐Ÿ‘ 9809698092 ๐Ÿ‘„๐ŸซฆIndependent Escort Service Cha...
ย 
๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...
๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...
๐ŸšบLEELA JOSHI WhatsApp Number +91-9930245274 โœ” Unsatisfied Bhabhi Call Girls T...
ย 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ย 
๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...
๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...
๐Ÿ’šCall Girls In Amritsar ๐Ÿ’ฏAnvi ๐Ÿ“ฒ๐Ÿ”8725944379๐Ÿ”Amritsar Call Girl No๐Ÿ’ฐAdvance Cash...
ย 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
ย 
๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...
๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...
๐Ÿ’ฐCall Girl In Bangaloreโ˜Ž๏ธ63788-78445๐Ÿ’ฐ Call Girl service in Bangaloreโ˜Ž๏ธBangalo...
ย 
Kolkata Call Girls Naktala ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ Top Class Call Girl Se...
Kolkata Call Girls Naktala  ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ  Top Class Call Girl Se...Kolkata Call Girls Naktala  ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ  Top Class Call Girl Se...
Kolkata Call Girls Naktala ๐Ÿ’ฏCall Us ๐Ÿ” 8005736733 ๐Ÿ” ๐Ÿ’ƒ Top Class Call Girl Se...
ย 
Electrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfElectrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdf
ย 
Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} โค๏ธVVIP ANJU Call Girls in Dehradun U...
ย 
Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...
Whitefield { Call Girl in Bangalore โ‚น7.5k Pick Up & Drop With Cash Payment 63...
ย 
๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...
๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...
๐Ÿ‘‰ Amritsar Call Girls ๐Ÿ‘‰๐Ÿ“ž 8725944379 ๐Ÿ‘‰๐Ÿ“ž Just๐Ÿ“ฒ Call Ruhi Call Girl Near Me Amri...
ย 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
ย 
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
Genuine Call Girls Hyderabad 9630942363 Book High Profile Call Girl in Hydera...
ย 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ย 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
ย 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
ย 
Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} โค๏ธVVIP POOJA Call Girls in Nagpur Maha...
ย 

Nuclear imaging PET CT Imaging Medical Physics Nuclear Medicine

  • 1. Lecture 10: Nuclear Imaging-II Shahid Younas NUCLEAR IMAGING The Scintillation Camera- Design Performance
  • 2. Gamma Camera โ€“Collimators Lecture 10: Nuclear Imaging-II Which of the following statements best describes the primary purpose of a collimator on a gamma camera? a. It prevents scattered photons from reaching the detector. b. It prevents cosmic radiation from reaching the detector. c. It stops pre-detector scattered photons.
  • 3. Gamma Camera โ€“Collimators Lecture 10: Nuclear Imaging-II ๏ถ To allow photons from a given region of interest to strike the detector and try to minimize the contribution of photons originating from outside this region.
  • 4. Gamma Camera โ€“Collimators Lecture 10: Nuclear Imaging-II ๏ถ Most scintillation cameras are provided with a selection of parallel-hole collimators: ๏ƒผ โ€œlow-energy, high-sensitivityโ€ ๏ƒผ โ€œlow-energy, all-purposeโ€ (LEAP) ๏ƒผ โ€œlow-energy, high-resolutionโ€ (LEHR)
  • 5. Parallel-hole collimator Lecture 10: Nuclear Imaging-II If a low-energy collimator is (incorrectly) used with a high-energy radionuclide the results would be: a. A reduced camera sensitivity (counting efficiency). b. A blurred image. c. A reduced field of view. d. Reduced image detail.
  • 6. Parallel-hole collimator Lecture 10: Nuclear Imaging-II ๏ถ Size of the image produced by parallel-hole collimator not affected by distance of object from collimator. ๏ถ Spatial resolution degrades rapidly with increasing collimator-to- object distance.
  • 7. Parallel-hole collimator Lecture 10: Nuclear Imaging-II The principal disadvantage in using a high resolution collimator on a gamma camera is that it has, a. Limited field of view. b. More distortion. c. Less scatter rejection. d. Lower sensitivity.
  • 8. Parallel-hole collimator Lecture 10: Nuclear Imaging-II ๏ถ R is an indicator of spatial resolution.
  • 9. Parallel-hole collimator Lecture 10: Nuclear Imaging-II ๏ถ Collimator spatial resolution of multi-hole collimators is determined by geometry of the holes . ๏ถ Spatial resolution improves as the, ๏ƒผ Diameters of the holes is reduced ๏ƒผ Lengths of the holes (thickness of the collimator) are increased
  • 10. Parallel-hole collimator Lecture 10: Nuclear Imaging-II ๏ถ Changing hole geometry to improve spatial resolution generally reduces the collimatorโ€™s efficiency. ๏ƒ˜ Resultant compromise is single most significant limitation on scintillation camera performance.
  • 11. Parallel-hole collimator Lecture 10: Nuclear Imaging-II ๏ถ Spatial resolution of parallel-hole collimator decreases linearly as collimator-to-object distance increases. ๏ƒ˜ Also one of the most important factors limiting scintillation camera performance.
  • 12. Parallel-hole collimator Lecture 10: Nuclear Imaging-II Do you know why the efficiency of a parallel-hole collimator is nearly constant over the collimator-to-object distances used for clinical imaging. Number of photons passing through given hole decreases with square of distance, number of holes through which photons can pass increases with square of distance.
  • 13. Parallel-hole collimator Lecture 10: Nuclear Imaging-II ๏ถ FWHM of LSF increased linearly with distance. ๏ถ Total area under LSF (photon fluence through collimator) decreases very little with s-c distance.
  • 14. Parallel-hole LEAP collimator Lecture 10: Nuclear Imaging-II ๏ถ LEAP collimators have holes with a large diameter. ๏ถ The sensitivity is relatively high as where the resolution is moderate. (larger diameter holes allow more scattered photons). ๏ถ The average sensitivity of a LEAP is approx. 500 kcpm for a 1-uCi source, and the resolution is 1.0cm at 10cm from the patent side of the collimator.
  • 15. Parallel-hole LEHR collimator Lecture 10: Nuclear Imaging-II ๏ถ LEHR collimators have higher resolution images than the LEAP. ๏ถ They have more holes that are both smaller and deeper. ๏ถ The sensitivity is approx. 185 kcpm for 1-uCi source, and the resolution is higher with 0.65cm at 10cm from the patient side of the collimator.
  • 16. Parallel-hole collimator Lecture 10: Nuclear Imaging-II Which is the ideal radionuclide to use with LEAP and LEHR? 99mTc
  • 17. Parallel-hole Medium Energy Collimator Lecture 10: Nuclear Imaging-II ๏ถ Medium Energy Collimators are used for medium energy photons of nuclides such as Krypton81, Gallium67, Indium111. These collimators have thinner septa than LEAP and LEHR collimators.
  • 18. Parallel-hole Medium Energy Collimator Lecture 10: Nuclear Imaging-II ๏ถ High Energy Collimators are used for Iodine131 and F-18FDG. These collimators have thicker septa than LEAP and LEHR collimators in order to reduce septal penetration by the higher energy photons.
  • 19. Parallel-Slant -hole Collimator Lecture 10: Nuclear Imaging-II ๏ถ A variation of the Parallel hole which has all tunnels slanted at a specific angle. ๏ถ It generates an oblique view for better visualization of an organ, which view is (partly) blocked by other parts of the body. ๏ถ As an advantage, this collimator can be positioned close to the body for the maximum gain in resolution.
  • 20. Pinhole collimator Lecture 10: Nuclear Imaging-II ๏ถ Used to produce magnified views of small objects, such as thyroid or hip joint. ๏ถ Consists of small (typically 3- to 5-mm diameter) hole in a piece of lead or tungsten. ๏ถ Mounted at apex of a leaded cone.
  • 21. Pinhole collimator Lecture 10: Nuclear Imaging-II ๏ถ Produces a magnified image whose orientation is reversed. ๏ถ Magnification decreases as object is moved away from pinhole.
  • 22. Pinhole collimator Lecture 10: Nuclear Imaging-II ๏ถ If the object is as far from the pinhole as the pinhole is from the crystal of the camera, the object is not magnified.
  • 23. Pinhole collimator Lecture 10: Nuclear Imaging-II ๏ถ This creates pitfall in clinical usage. ๏ถ A thyroid nodule deep in the mediastinum can appear to be in the thyroid itself. ๏ถ Extensive use is in pediatric nuclear medicine.
  • 24. Converging collimator Lecture 10: Nuclear Imaging-II ๏ถ In a Converging collimator the holes are not parallel but focused toward the organ. ๏ถ The focal point is normally located in the center of the field of view (FOV).
  • 25. Converging collimator Lecture 10: Nuclear Imaging-II ๏ถ Magnifies the image. ๏ถ Magnification increases as the object is moved away from the collimator.
  • 26. Converging collimator Lecture 10: Nuclear Imaging-II ๏ถ Converging collimator is seldom used. ๏ถ Imaging characteristics are superior, in theory, to the parallel- hole collimators but in practice owing to, ๏ƒ˜ Decreasing FOV and varying magnification with distance
  • 27. Diverging collimator Lecture 10: Nuclear Imaging-II ๏ถ Many holes, all aimed a focal point behind the camera. ๏ถ Produces a minified image. ๏ถ Amount of mini-fication increases as object is moved away from the collimator
  • 28. Diverging collimator Lecture 10: Nuclear Imaging-II ๏ถ May be used to image a large portion of a patient on a small (25-cm diameter) or standard (30-cm diameter) FOV camera. ๏ถ You can perform lung study using a mobile gamma camera in intensive care unit.
  • 29. Diverging collimator Lecture 10: Nuclear Imaging-II ๏ถ Compared to a diverging collimator, a converging collimator will produce: a. A increase in sensitivity when distance is increased. b. Better image detail. c. A reduced FOV as distance is increased.
  • 30. Diverging collimator Lecture 10: Nuclear Imaging-II What collimator will you get if you reverse a diverging collimator on a camera? Converging
  • 31. Diverging collimator Lecture 10: Nuclear Imaging-II ๏ถ Its seldom used because of, ๏ƒ˜ Inferior imaging characteristics to the parallel hole ๏ƒ˜ Large crystal sizes of modern cameras.
  • 32. Fan Beam collimator Lecture 10: Nuclear Imaging-II ๏ถ They are designed for a rectangular camera head or SPECT to image smaller organs like the brain and heart. ๏ถ When viewed from one direction, the holes are parallel. When viewed from the other direction, the holes converge. Itโ€™s a hybrid of parallel-hole and converging collimator.
  • 33. collimator Lecture 10: Nuclear Imaging-II Image magnification is changed if you change, A. Radionuclide B. Imaging time C. Type of collimator D. Patient-collimator distance E. PHA window level
  • 34. Image formation Lecture 10: Nuclear Imaging-II ๏ถ Photons from each point in the patient are emitted isotropically. ๏ถ Some photons escape the patient without interaction, some scatter within the patient before escaping, and some are absorbed in the patient
  • 35. Image formation Lecture 10: Nuclear Imaging-II ๏ถ Many of the escaping photons are not detected because they are emitted in directions away from the detector. ๏ถ Only a tiny fraction of emitted photons has trajectories permitting passage through the collimator holes.
  • 36. Image formation Lecture 10: Nuclear Imaging-II ๏ถ Of those reaching the crystal, some are absorbed in the crystal, some scatter from the crystal, and some pass through the crystal without interaction
  • 37. Image formation Lecture 10: Nuclear Imaging-II ๏ƒ˜ Relative probabilities of these events depends on the energies of the photons and the thickness of the crystal
  • 38. Image formation Lecture 10: Nuclear Imaging-II ๏ถ Of those photons absorbed in the crystal, some are absorbed by a single photoelectric absorption, others undergo one or more Compton scatters before a photoelectric absorption.
  • 39. Image formation Lecture 10: Nuclear Imaging-II Is it possible for two photons to simultaneously interact with the crystal?
  • 40. Image formation Lecture 10: Nuclear Imaging-II ๏ƒ˜ If the energy signal (Z) from the coincident interactions is within the energy window of the energy discrimination circuit, the result will be a single count mis-positioned in the image. ๏ถ Fraction of simultaneous interactions increases with the interaction rate of the camera.
  • 41. Image formation Lecture 10: Nuclear Imaging-II ๏ถ Spatial resolution and image contrast reduced by: ๏ƒ˜ Interactions in crystal of photons that have been scattered in the patient. ๏ƒ˜ Photons that have penetrated the collimator septa. ๏ƒ˜ Photons that undergo one or more scatters in the crystal. ๏ƒ˜ Coincident interactions.
  • 42. Image formation Lecture 10: Nuclear Imaging-II Energy discrimination circuits reduce this by rejecting photons that scatter in the patient or result in coincident interactions.
  • 43. Image formation Lecture 10: Nuclear Imaging-II If the PHA window on a gamma camera is (incorrectly) set below the photo-peak energy you would expect to get: ๏ƒผ Decreased sensitivity. ๏ƒผ An image of primarily scattered radiation. ๏ƒผ Decreased lesion contrast.
  • 44. Image formation Lecture 10: Nuclear Imaging-II ๏ถ Pulse Height Analyzer (PHA) reduce this loss of resolution and contrast by rejecting scattered photons. ๏ถ However, low energy photons can scatter through large angles with only a small energy loss.
  • 45. Image formation Lecture 10: Nuclear Imaging-II ๏ถ 140 keV photon scattering 45 degree will only lose 7.4% of its energy. ๏ถ This causes a wide photo-peak because PHA accepts most of the significant fraction of the scattered photons and coincident interactions.
  • 46. Image formation Lecture 10: Nuclear Imaging-II The full width at half maximum (FWHM) of a photo-peak is a measure of, a. PHA window setting. b. Camera sensitivity. c. Field of view. d. Detector energy resolution.
  • 47. Image formation Lecture 10: Nuclear Imaging-II Performance