Design of Body Fat Measurement System based on the Android
Platform
Woong-Sik Kim
Medical IT Engineering Department, University of Konyang, #121, Daehak-ro, Nonsan, , Chungnam, South
Korea
Abstract As there has been recent fast development of a BIO signal measurement technology,
our computing environment is rapidly shifting from existing desktop PC to Embedded System.
Therefore, this paper has introduced an implementation of the body fat measurement system
based on the Android Platform which has the same data precision as the one used in a
hospital. The most important fact in most of the hospital equipments is that they need
separate communication means which can help connection with PC, but in this paper, as we
use the Android Platform in the Porting Embedded System, it is possible to measure body fat
signal directly without separate communication means. We also have developed an
application(App) program for the body fat measurement which can measure the body fat
more conveniently and simple.
I. Introduction
Body fat is the fat a human body contains. It is the energy storage which stores surplus calories after
the body uses up calorie intake. When the body needs it, it decomposes into energy.
The body fat percentage is the percentage of the body fat to the body weight. For a man, it is 10-20%,
and for a woman 18-28%. The percentages of visceral fat and subcutaneous fat are different individually by the
degree of obesity and the amount of exercise.
When people become adult with obesity, normally the abdominal fat percentage becomes higher due to an
excessive visceral fat. This abdominal obesity is the cause of adult diseases such as high blood pressure,
cardiovascular disorder and diabetes.
It is because the subcutaneous fat is accumulated far from the center of the human body anatomically. Yet, even
though the oxygen transfers through microvessles, the visceral fat relatively connects with large blood vessels so
that it is highly active.
For the body fat measurements, there are UWW(Under Water Weight), BIA(Bioelectrical Impedance
Analysis), and TANITA BIA(Tanita Bioelectrical Impedance Analysis.) Among these, UWW is the most
accurate method to measure the body fat.
From the differential between a weight measured in the air and a weight in the water, this method uses
Archimedes principle to calculate the body density and then calculate the body fat percentage by composition
and specific gravity of the body tissue. However, this method needs an equipment to immerge the body in water
up to head for measurement so that it is not possible for a general use.
Currently, the simplest method to measure the body fat is BIA which uses a bioelectrical impedance.
All matters have impedance electrically. BIA uses both the electrical characteristic of the human body which
consists of highly conductive tissue(Conductor) and less conductive tissue(Insulator) and the fact that the ratio
of the above two tissues is reflected in measured impedance. So it is called bioelectrical impedance analysis.
II. Main discussion
To measure the body fat, let micro alternating current with 50 kHz and 100 μA flow into the body through
two input electrode plates.
And the resistance composition of the body is calculated by measuring the voltage between two electrodes.
Besides this, the information of height, gender, weight and age of a person is needed. The amount of a body fat
is calculated by subtracting FFM(Fat Free Mass) from the weight of the user. The three existing calculating
methods to measure a fat free mass are shown in equations 4. Its unit is Kg.
FFMPAT =4.033+GPAT×(0.734×(H2
/R)+0.096×X)+0.116×A+0.878×G . (1a)
Design of Body Fat Measurement System based on the Android Platform
www.ijres.org 45 | Page
FFMHYU = 0.0005+H2
-GHYU×Z+0.392×W-0.0684×A-5.1841×G+24.678 . (1b)
FFMHOOT = 0.61×GHOOT×(H2
/R)+0.25×W+1.31 (1c)
.
Here, H(cm) denotes the height, R(Ω) the Resistance, X(Ω) the Reactance, A the Age, Z(Ω) the Impedance, W
the Weight. G denotes Gender and its value is set as 1 for male and 0 for female.
GPAT, GHYU, GHOOT are constants which values are related to the values of Impedance measured from the
body of the user. All these values are set as 1 in existing equations. In existing methods, the body fat was
measured by attaching two hands and two feet closely on electrodes.
As a voltage differential is measured by attaching two left and right fingers in the implementing system, it is
hard to apply the existing equation as it is. Therefore, in order to revise this, the impedance and its related
constants are used as tuning elements.
Fat Mass(FM) is described as in the equation 2.
FM = W - FFM(Kg) (2)
Body Fat(%BF) can be calculated as in the equation 3.
%BF = (FM/W) × 100(%) (3)
And Body Mass Index(BMI) is calculated as in the equation 4.
BMI = (W/H2
) (Kg/m2
) (4)
The amount of the body water which measures the water in a body can be calculated as in the equation 5.
TBW = 0.59×GTBW×(H2
/R)+0.065×W+1.04(Kg) (5)
Based on measured body fat, daily calory consumption can be calculated by substituting user's walking steps
into the equation. The equation 6 shows daily calory consumption(C).
C = 1.11 × (Ka × W + Kb × Wn
) (6)
Ka denotes a living activity index, W a weight, Kb a basal energy metabolic rate, and n=0.424* constant.
The measuring equipment for the body fat currently made for a commercial use measures the body fat in various
parts of the body by attaching electrodes on legs and arms or by grabbing electrodes attached on handles by two
hands.
In the implemented system, the body fat can be measured by attaching electrodes on four fingers. Therefore,
based on the results from various clinical demonstrations, tuning coefficients of the equation is necessary.
From the result of clinical demonstrations, the heart rate result is accurate, but there is much error in the body fat
measurement depending on how a user grabs the electrodes with hands.
In case when he grabs electrodes properly, there is about 10% error compared to the product for common use.
Figure 1 is a block diagram for the body fat equipment developed in this paper. This body fat equipment
consists of a signal detection component which measures impedance from the body, an amplifying component
which amplifies detected signal, a signal processing component for filtering and formatting the amplified signal,
and a transferring component which transfers the processed signal.
[Figure 1] Components Diagram for Body Fat Equipment
III. Test and Result
3.1 How to measure the body fat
Design of Body Fat Measurement System based on the Android Platform
www.ijres.org 46 | Page
[Figure 2] Circuit Configuration to measure the body fat.
The above figure shows a circuit diagram for the measurement of the body fat. In the first component of the
circuit diagram, there is a part which causes a frequency of 100 uA, 50 KHz entering into the body.
Due to a delicate difference in PCB pattern, in order to overcome the some voltage differential in each board,
namely for calibration, 5 KΩ adjustable resistance is provided.
A signal occurred in the main circuit enters into the body. And after the signal passes through the body, it comes
out by both inverting amplifier and non inverting amplifier.
The output signal is analyzed by ADC of MCU and MCU to measure the degree of the final body fat. For
Calibration, R58, R59, R60 parts should be connected in the above circuit diagram, and it should be controlled
to have +(-) 1Vp-p in TP8
3.2 Result from Implementation
This display shows the test result of the body fat measurement by DDR function of Eclipse smart display
capturing which is a tool for the development of Android. Figure 8 is the result of the elctrocardiogram signal
tested in the embeded board based on Android.
[Figure 3] raw Pass filter application display
The left signal in the figure 8 is the wave pattern before applying the filter, and the display on the right in figure
8 shows an example of an output when raw Pass filter is applied to the wave pattern of an electrocardiogram.
[Figure 4] high Pass filter application
Design of Body Fat Measurement System based on the Android Platform
www.ijres.org 47 | Page
Figure 4 shows an example of a high Pass filter application. It can be seen that the clear electrocardiogram
signal has come out by applying filter at channels 1, 2, and 3.
[Figure 5] Body fat measurement display of an embedded board based on the Android Platform.
Figure 5 shows the display of electrocardiogram. More meaningful electrocardiogram data has come out by
applying the filter previously explained.
IV. Conclusions and Next Assignment
In this paper, for the accurate electrocardiogram measurement, the filter circuit, which can remove Low
Pass Filter, High Pass Filter and Power Noise, has been implemented to reduce error caused by the noise of the
measuring sensor in the system which measures the body fat signal.
In this paper, the App(application) of the body fat measurement has been implemented and displayed, which
measures the body fat in the embedded system in which the Android has been porting
This means that it can simply diagnose not only an Android link electrocardiogram measurement system but
also a real ubiquitous through various communication equipments and their methods.
Also, it has a merit to offer effective diagnosis service with low cost through linkage with the USN base system.
As this system can transfer data by using communication means based on Android, the next study about
information security method should be done at the same time to protect information of a user.
References
[1.] Albu, J. B., Kovera, A. J., Johnson, J. A. (2000) Fat distribution and health in obesity. Ann N Y Acad Sci 904: 491–501.
[2.] Doll, S., Paccaud, F., Bovet, P., Burnier, M., Wietlisbach, V. (2002) Body mass index, abdominal adiposity and blood pressure:
consistency of their association across developing and developed countries. Int J Obes Relat Metab Disord 26: 48–57.
[3.] Visscher, T. L., Seidell, J. C., Molarius, A., van der Kuip, D., Hofman, A., Witteman, J. C. (2001) A comparison of body mass
index, waist-hip ratio and waist circumference as predictors of all-cause mortality among the elderly: the Rotterdam study. Int J
Obes Relat Metab Disord 25: 1730–1735.
[4.] Rexrode, K. M., Carey, V. J., Hennekens, C. H., et al. (1998) Abdominal adiposity and coronary heart disease in women. JAMA
280: 1843–1848.
[5.] Rexrode, K. M., Buring, J. E., Manson, J. E. (2001) Abdominal and total adiposity and risk of coronary heart disease in men. Int
J Obes Relat Metab Disord 25: 1047–1056.
[6.] Folsom, A. R., Kushi, L. H., Anderson, K. E., et al. (2000) Associations of general and abdominal obesity with multiple health
outcomes in older women: the Iowa Women's Health Study. Arch Intern Med 160: 2117–2128.
[7.] Peiris, A. N., Sothmann, M. S., Hoffmann, R. G., et al. (1989) Adiposity, fat distribution, and cardiovascular risk. Ann Intern
Med 110: 867–872.
[8.] Ansari, A., Rholl, A. O. (1986) Pseudopericardial effusion: echocardiographic and computed tomographic correlations. Clin
Cardiol 9: 551–555.
[9.] Savage, D. D., Garrison, R. J., Brand, F., et al. (1983) Prevalence and correlates of posterior extra echocardiographic spaces in a
free-living population based sample (the Framingham study). Am J Cardiol 51: 1207–1212.
[10.] Walsh, K. M., Leen, E., MacSween, R. N., Morris, A. J. (1998) Hepatic blood flow changes in chronic hepatitis C measured by
duplex Doppler color sonography: relationship to histological features. Dig Dis Sci 43: 2584–2590.
[11.] Janssen, I., Heymsfield, S. B., Allison, D. B., Kotler, D. P., Ross, R. (2002) Body mass index and waist circumference
independently contribute to the prediction of nonabdominal, abdominal subcutaneous, and visceral fat. Am J Clin Nutr 75: 683–
688.
[12.] Marchington, J. M., Mattacks, C. A., Pond, C. M. (1989) Adipose tissue in the mammalian heart and pericardium: structure, fetal
development and biochemical properties. Comp Biochem Physiol B 94: 225–232.
[13.] Sahn, D., DeMaria, A., Kisslo, J., Weyman, A. (1978) The Committee on M-mode standardization of the American Society of
Echocardiography: recommendations regarding quantitation in M-mode echocardiography: results of a survey of
echocardiography measurements. Circulation 58: 1072–1083.

More Related Content

Similar to Design of Body Fat Measurement System based on the Android Platform

Bioelectric fat analyser
Bioelectric fat analyserBioelectric fat analyser
Bioelectric fat analyserSandeep Reddy
 
Bioimpedance harwareembc20
Bioimpedance harwareembc20Bioimpedance harwareembc20
Bioimpedance harwareembc20Hoopeer Hoopeer
 
Waist-to-height ratio assessment device
Waist-to-height ratio assessment device Waist-to-height ratio assessment device
Waist-to-height ratio assessment device IJECEIAES
 
Final project ppt.pptx
Final project ppt.pptxFinal project ppt.pptx
Final project ppt.pptxMalathyN5
 
Liver segmentationwith2du net
Liver segmentationwith2du netLiver segmentationwith2du net
Liver segmentationwith2du netAmrCady
 
IRJET- Wireless Healthcare Monitoring using Android Phones
IRJET- Wireless Healthcare Monitoring using Android PhonesIRJET- Wireless Healthcare Monitoring using Android Phones
IRJET- Wireless Healthcare Monitoring using Android PhonesIRJET Journal
 
rf based complex impedance measurement
rf based complex impedance measurementrf based complex impedance measurement
rf based complex impedance measurementINFOGAIN PUBLICATION
 
Eating Habit and Health Monitoring System using Android Based Machine Learning
Eating Habit and Health Monitoring System using Android Based Machine LearningEating Habit and Health Monitoring System using Android Based Machine Learning
Eating Habit and Health Monitoring System using Android Based Machine LearningIRJET Journal
 
Non-Contact Health Monitoring System Using Image and Signal Processing
Non-Contact Health Monitoring System Using Image and Signal ProcessingNon-Contact Health Monitoring System Using Image and Signal Processing
Non-Contact Health Monitoring System Using Image and Signal ProcessingAtul Kumar Sharma
 
IRJET-Automatic RBC And WBC Counting using Watershed Segmentation Algorithm
IRJET-Automatic RBC And WBC Counting using Watershed Segmentation AlgorithmIRJET-Automatic RBC And WBC Counting using Watershed Segmentation Algorithm
IRJET-Automatic RBC And WBC Counting using Watershed Segmentation AlgorithmIRJET Journal
 
IRJET- Diabetes Prediction using Machine Learning
IRJET- Diabetes Prediction using Machine LearningIRJET- Diabetes Prediction using Machine Learning
IRJET- Diabetes Prediction using Machine LearningIRJET Journal
 
WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.
WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.
WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.ijiert bestjournal
 
IRJET- Remote Health Monitoring of Ankle Edema by using E- Textile Techno...
IRJET-  	  Remote Health Monitoring of Ankle Edema by using E- Textile Techno...IRJET-  	  Remote Health Monitoring of Ankle Edema by using E- Textile Techno...
IRJET- Remote Health Monitoring of Ankle Edema by using E- Textile Techno...IRJET Journal
 
Bioimpedance tech in breast canser
Bioimpedance tech in breast canser Bioimpedance tech in breast canser
Bioimpedance tech in breast canser athraaflower
 
Mesurment of body composition
Mesurment of body compositionMesurment of body composition
Mesurment of body compositionHarshita Bhargava
 
Review on Image-based Body Mass Index Prediction Methods
Review on Image-based Body Mass Index Prediction MethodsReview on Image-based Body Mass Index Prediction Methods
Review on Image-based Body Mass Index Prediction MethodsIRJET Journal
 
Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...
Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...
Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...toukaigi
 
Strategy to determine the foot plantar center of pressure of a person through...
Strategy to determine the foot plantar center of pressure of a person through...Strategy to determine the foot plantar center of pressure of a person through...
Strategy to determine the foot plantar center of pressure of a person through...TELKOMNIKA JOURNAL
 

Similar to Design of Body Fat Measurement System based on the Android Platform (20)

Bioelectric fat analyser
Bioelectric fat analyserBioelectric fat analyser
Bioelectric fat analyser
 
Bioimpedance harwareembc20
Bioimpedance harwareembc20Bioimpedance harwareembc20
Bioimpedance harwareembc20
 
Waist-to-height ratio assessment device
Waist-to-height ratio assessment device Waist-to-height ratio assessment device
Waist-to-height ratio assessment device
 
Final project ppt.pptx
Final project ppt.pptxFinal project ppt.pptx
Final project ppt.pptx
 
Liver segmentationwith2du net
Liver segmentationwith2du netLiver segmentationwith2du net
Liver segmentationwith2du net
 
Health Monitoring
Health MonitoringHealth Monitoring
Health Monitoring
 
IRJET- Wireless Healthcare Monitoring using Android Phones
IRJET- Wireless Healthcare Monitoring using Android PhonesIRJET- Wireless Healthcare Monitoring using Android Phones
IRJET- Wireless Healthcare Monitoring using Android Phones
 
rf based complex impedance measurement
rf based complex impedance measurementrf based complex impedance measurement
rf based complex impedance measurement
 
Eating Habit and Health Monitoring System using Android Based Machine Learning
Eating Habit and Health Monitoring System using Android Based Machine LearningEating Habit and Health Monitoring System using Android Based Machine Learning
Eating Habit and Health Monitoring System using Android Based Machine Learning
 
Non-Contact Health Monitoring System Using Image and Signal Processing
Non-Contact Health Monitoring System Using Image and Signal ProcessingNon-Contact Health Monitoring System Using Image and Signal Processing
Non-Contact Health Monitoring System Using Image and Signal Processing
 
IRJET-Automatic RBC And WBC Counting using Watershed Segmentation Algorithm
IRJET-Automatic RBC And WBC Counting using Watershed Segmentation AlgorithmIRJET-Automatic RBC And WBC Counting using Watershed Segmentation Algorithm
IRJET-Automatic RBC And WBC Counting using Watershed Segmentation Algorithm
 
IRJET- Diabetes Prediction using Machine Learning
IRJET- Diabetes Prediction using Machine LearningIRJET- Diabetes Prediction using Machine Learning
IRJET- Diabetes Prediction using Machine Learning
 
WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.
WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.
WIRELESS BODY AREA SENSOR NETWORK FOR SOLDIER SYSTEM.
 
Sub1550
Sub1550Sub1550
Sub1550
 
IRJET- Remote Health Monitoring of Ankle Edema by using E- Textile Techno...
IRJET-  	  Remote Health Monitoring of Ankle Edema by using E- Textile Techno...IRJET-  	  Remote Health Monitoring of Ankle Edema by using E- Textile Techno...
IRJET- Remote Health Monitoring of Ankle Edema by using E- Textile Techno...
 
Bioimpedance tech in breast canser
Bioimpedance tech in breast canser Bioimpedance tech in breast canser
Bioimpedance tech in breast canser
 
Mesurment of body composition
Mesurment of body compositionMesurment of body composition
Mesurment of body composition
 
Review on Image-based Body Mass Index Prediction Methods
Review on Image-based Body Mass Index Prediction MethodsReview on Image-based Body Mass Index Prediction Methods
Review on Image-based Body Mass Index Prediction Methods
 
Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...
Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...
Development of a 3D Interactive Virtual Market System with Adaptive Treadmill...
 
Strategy to determine the foot plantar center of pressure of a person through...
Strategy to determine the foot plantar center of pressure of a person through...Strategy to determine the foot plantar center of pressure of a person through...
Strategy to determine the foot plantar center of pressure of a person through...
 

More from IJRES Journal

Exploratory study on the use of crushed cockle shell as partial sand replacem...
Exploratory study on the use of crushed cockle shell as partial sand replacem...Exploratory study on the use of crushed cockle shell as partial sand replacem...
Exploratory study on the use of crushed cockle shell as partial sand replacem...IJRES Journal
 
Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...
Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...
Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...IJRES Journal
 
Review: Nonlinear Techniques for Analysis of Heart Rate Variability
Review: Nonlinear Techniques for Analysis of Heart Rate VariabilityReview: Nonlinear Techniques for Analysis of Heart Rate Variability
Review: Nonlinear Techniques for Analysis of Heart Rate VariabilityIJRES Journal
 
Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...
Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...
Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...IJRES Journal
 
Study and evaluation for different types of Sudanese crude oil properties
Study and evaluation for different types of Sudanese crude oil propertiesStudy and evaluation for different types of Sudanese crude oil properties
Study and evaluation for different types of Sudanese crude oil propertiesIJRES Journal
 
A Short Report on Different Wavelets and Their Structures
A Short Report on Different Wavelets and Their StructuresA Short Report on Different Wavelets and Their Structures
A Short Report on Different Wavelets and Their StructuresIJRES Journal
 
A Case Study on Academic Services Application Using Agile Methodology for Mob...
A Case Study on Academic Services Application Using Agile Methodology for Mob...A Case Study on Academic Services Application Using Agile Methodology for Mob...
A Case Study on Academic Services Application Using Agile Methodology for Mob...IJRES Journal
 
Wear Analysis on Cylindrical Cam with Flexible Rod
Wear Analysis on Cylindrical Cam with Flexible RodWear Analysis on Cylindrical Cam with Flexible Rod
Wear Analysis on Cylindrical Cam with Flexible RodIJRES Journal
 
DDOS Attacks-A Stealthy Way of Implementation and Detection
DDOS Attacks-A Stealthy Way of Implementation and DetectionDDOS Attacks-A Stealthy Way of Implementation and Detection
DDOS Attacks-A Stealthy Way of Implementation and DetectionIJRES Journal
 
An improved fading Kalman filter in the application of BDS dynamic positioning
An improved fading Kalman filter in the application of BDS dynamic positioningAn improved fading Kalman filter in the application of BDS dynamic positioning
An improved fading Kalman filter in the application of BDS dynamic positioningIJRES Journal
 
Positioning Error Analysis and Compensation of Differential Precision Workbench
Positioning Error Analysis and Compensation of Differential Precision WorkbenchPositioning Error Analysis and Compensation of Differential Precision Workbench
Positioning Error Analysis and Compensation of Differential Precision WorkbenchIJRES Journal
 
Status of Heavy metal pollution in Mithi river: Then and Now
Status of Heavy metal pollution in Mithi river: Then and NowStatus of Heavy metal pollution in Mithi river: Then and Now
Status of Heavy metal pollution in Mithi river: Then and NowIJRES Journal
 
The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...
The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...
The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...IJRES Journal
 
Experimental study on critical closing pressure of mudstone fractured reservoirs
Experimental study on critical closing pressure of mudstone fractured reservoirsExperimental study on critical closing pressure of mudstone fractured reservoirs
Experimental study on critical closing pressure of mudstone fractured reservoirsIJRES Journal
 
Correlation Analysis of Tool Wear and Cutting Sound Signal
Correlation Analysis of Tool Wear and Cutting Sound SignalCorrelation Analysis of Tool Wear and Cutting Sound Signal
Correlation Analysis of Tool Wear and Cutting Sound SignalIJRES Journal
 
Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...
Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...
Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...IJRES Journal
 
Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...
Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...
Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...IJRES Journal
 
A novel high-precision curvature-compensated CMOS bandgap reference without u...
A novel high-precision curvature-compensated CMOS bandgap reference without u...A novel high-precision curvature-compensated CMOS bandgap reference without u...
A novel high-precision curvature-compensated CMOS bandgap reference without u...IJRES Journal
 
Structural aspect on carbon dioxide capture in nanotubes
Structural aspect on carbon dioxide capture in nanotubesStructural aspect on carbon dioxide capture in nanotubes
Structural aspect on carbon dioxide capture in nanotubesIJRES Journal
 
Thesummaryabout fuzzy control parameters selected based on brake driver inten...
Thesummaryabout fuzzy control parameters selected based on brake driver inten...Thesummaryabout fuzzy control parameters selected based on brake driver inten...
Thesummaryabout fuzzy control parameters selected based on brake driver inten...IJRES Journal
 

More from IJRES Journal (20)

Exploratory study on the use of crushed cockle shell as partial sand replacem...
Exploratory study on the use of crushed cockle shell as partial sand replacem...Exploratory study on the use of crushed cockle shell as partial sand replacem...
Exploratory study on the use of crushed cockle shell as partial sand replacem...
 
Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...
Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...
Congenital Malaria: Correlation of Umbilical Cord Plasmodium falciparum Paras...
 
Review: Nonlinear Techniques for Analysis of Heart Rate Variability
Review: Nonlinear Techniques for Analysis of Heart Rate VariabilityReview: Nonlinear Techniques for Analysis of Heart Rate Variability
Review: Nonlinear Techniques for Analysis of Heart Rate Variability
 
Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...
Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...
Dynamic Modeling for Gas Phase Propylene Copolymerization in a Fluidized Bed ...
 
Study and evaluation for different types of Sudanese crude oil properties
Study and evaluation for different types of Sudanese crude oil propertiesStudy and evaluation for different types of Sudanese crude oil properties
Study and evaluation for different types of Sudanese crude oil properties
 
A Short Report on Different Wavelets and Their Structures
A Short Report on Different Wavelets and Their StructuresA Short Report on Different Wavelets and Their Structures
A Short Report on Different Wavelets and Their Structures
 
A Case Study on Academic Services Application Using Agile Methodology for Mob...
A Case Study on Academic Services Application Using Agile Methodology for Mob...A Case Study on Academic Services Application Using Agile Methodology for Mob...
A Case Study on Academic Services Application Using Agile Methodology for Mob...
 
Wear Analysis on Cylindrical Cam with Flexible Rod
Wear Analysis on Cylindrical Cam with Flexible RodWear Analysis on Cylindrical Cam with Flexible Rod
Wear Analysis on Cylindrical Cam with Flexible Rod
 
DDOS Attacks-A Stealthy Way of Implementation and Detection
DDOS Attacks-A Stealthy Way of Implementation and DetectionDDOS Attacks-A Stealthy Way of Implementation and Detection
DDOS Attacks-A Stealthy Way of Implementation and Detection
 
An improved fading Kalman filter in the application of BDS dynamic positioning
An improved fading Kalman filter in the application of BDS dynamic positioningAn improved fading Kalman filter in the application of BDS dynamic positioning
An improved fading Kalman filter in the application of BDS dynamic positioning
 
Positioning Error Analysis and Compensation of Differential Precision Workbench
Positioning Error Analysis and Compensation of Differential Precision WorkbenchPositioning Error Analysis and Compensation of Differential Precision Workbench
Positioning Error Analysis and Compensation of Differential Precision Workbench
 
Status of Heavy metal pollution in Mithi river: Then and Now
Status of Heavy metal pollution in Mithi river: Then and NowStatus of Heavy metal pollution in Mithi river: Then and Now
Status of Heavy metal pollution in Mithi river: Then and Now
 
The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...
The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...
The Low-Temperature Radiant Floor Heating System Design and Experimental Stud...
 
Experimental study on critical closing pressure of mudstone fractured reservoirs
Experimental study on critical closing pressure of mudstone fractured reservoirsExperimental study on critical closing pressure of mudstone fractured reservoirs
Experimental study on critical closing pressure of mudstone fractured reservoirs
 
Correlation Analysis of Tool Wear and Cutting Sound Signal
Correlation Analysis of Tool Wear and Cutting Sound SignalCorrelation Analysis of Tool Wear and Cutting Sound Signal
Correlation Analysis of Tool Wear and Cutting Sound Signal
 
Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...
Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...
Reduce Resources for Privacy in Mobile Cloud Computing Using Blowfish and DSA...
 
Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...
Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...
Resistance of Dryland Rice to Stem Borer (Scirpophaga incertulas Wlk.) Using ...
 
A novel high-precision curvature-compensated CMOS bandgap reference without u...
A novel high-precision curvature-compensated CMOS bandgap reference without u...A novel high-precision curvature-compensated CMOS bandgap reference without u...
A novel high-precision curvature-compensated CMOS bandgap reference without u...
 
Structural aspect on carbon dioxide capture in nanotubes
Structural aspect on carbon dioxide capture in nanotubesStructural aspect on carbon dioxide capture in nanotubes
Structural aspect on carbon dioxide capture in nanotubes
 
Thesummaryabout fuzzy control parameters selected based on brake driver inten...
Thesummaryabout fuzzy control parameters selected based on brake driver inten...Thesummaryabout fuzzy control parameters selected based on brake driver inten...
Thesummaryabout fuzzy control parameters selected based on brake driver inten...
 

Recently uploaded

Thermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - VThermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - VDineshKumar4165
 
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Bookingroncy bisnoi
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
Vivazz, Mieres Social Housing Design Spain
Vivazz, Mieres Social Housing Design SpainVivazz, Mieres Social Housing Design Spain
Vivazz, Mieres Social Housing Design Spaintimesproduction05
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...ranjana rawat
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . pptDineshKumar4165
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756dollysharma2066
 
Online banking management system project.pdf
Online banking management system project.pdfOnline banking management system project.pdf
Online banking management system project.pdfKamal Acharya
 
UNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICS
UNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICSUNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICS
UNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICSrknatarajan
 
Unit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdfUnit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdfRagavanV2
 
Glass Ceramics: Processing and Properties
Glass Ceramics: Processing and PropertiesGlass Ceramics: Processing and Properties
Glass Ceramics: Processing and PropertiesPrabhanshu Chaturvedi
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTbhaskargani46
 
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...Call Girls in Nagpur High Profile
 
chapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringchapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringmulugeta48
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Christo Ananth
 
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxBSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxfenichawla
 
Intze Overhead Water Tank Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank  Design by Working Stress - IS Method.pdfIntze Overhead Water Tank  Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank Design by Working Stress - IS Method.pdfSuman Jyoti
 

Recently uploaded (20)

Thermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - VThermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - V
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Walvekar Nagar Call Me 7737669865 Budget Friendly No Advance Booking
 
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
Vivazz, Mieres Social Housing Design Spain
Vivazz, Mieres Social Housing Design SpainVivazz, Mieres Social Housing Design Spain
Vivazz, Mieres Social Housing Design Spain
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
 
Online banking management system project.pdf
Online banking management system project.pdfOnline banking management system project.pdf
Online banking management system project.pdf
 
UNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICS
UNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICSUNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICS
UNIT-IFLUID PROPERTIES & FLOW CHARACTERISTICS
 
Unit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdfUnit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdf
 
Glass Ceramics: Processing and Properties
Glass Ceramics: Processing and PropertiesGlass Ceramics: Processing and Properties
Glass Ceramics: Processing and Properties
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPT
 
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...
 
chapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringchapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineering
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
 
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxBSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
 
Intze Overhead Water Tank Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank  Design by Working Stress - IS Method.pdfIntze Overhead Water Tank  Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank Design by Working Stress - IS Method.pdf
 

Design of Body Fat Measurement System based on the Android Platform

  • 1. International Journal of Research in Engineering and Science (IJRES) ISSN (Online): 2320-9364, ISSN (Print): 2320-9356 www.ijres.org Volume 2 Issue 12 ǁ December. 2014 ǁ PP.44-47 www.ijres.org 44 | Page Design of Body Fat Measurement System based on the Android Platform Woong-Sik Kim Medical IT Engineering Department, University of Konyang, #121, Daehak-ro, Nonsan, , Chungnam, South Korea Abstract As there has been recent fast development of a BIO signal measurement technology, our computing environment is rapidly shifting from existing desktop PC to Embedded System. Therefore, this paper has introduced an implementation of the body fat measurement system based on the Android Platform which has the same data precision as the one used in a hospital. The most important fact in most of the hospital equipments is that they need separate communication means which can help connection with PC, but in this paper, as we use the Android Platform in the Porting Embedded System, it is possible to measure body fat signal directly without separate communication means. We also have developed an application(App) program for the body fat measurement which can measure the body fat more conveniently and simple. I. Introduction Body fat is the fat a human body contains. It is the energy storage which stores surplus calories after the body uses up calorie intake. When the body needs it, it decomposes into energy. The body fat percentage is the percentage of the body fat to the body weight. For a man, it is 10-20%, and for a woman 18-28%. The percentages of visceral fat and subcutaneous fat are different individually by the degree of obesity and the amount of exercise. When people become adult with obesity, normally the abdominal fat percentage becomes higher due to an excessive visceral fat. This abdominal obesity is the cause of adult diseases such as high blood pressure, cardiovascular disorder and diabetes. It is because the subcutaneous fat is accumulated far from the center of the human body anatomically. Yet, even though the oxygen transfers through microvessles, the visceral fat relatively connects with large blood vessels so that it is highly active. For the body fat measurements, there are UWW(Under Water Weight), BIA(Bioelectrical Impedance Analysis), and TANITA BIA(Tanita Bioelectrical Impedance Analysis.) Among these, UWW is the most accurate method to measure the body fat. From the differential between a weight measured in the air and a weight in the water, this method uses Archimedes principle to calculate the body density and then calculate the body fat percentage by composition and specific gravity of the body tissue. However, this method needs an equipment to immerge the body in water up to head for measurement so that it is not possible for a general use. Currently, the simplest method to measure the body fat is BIA which uses a bioelectrical impedance. All matters have impedance electrically. BIA uses both the electrical characteristic of the human body which consists of highly conductive tissue(Conductor) and less conductive tissue(Insulator) and the fact that the ratio of the above two tissues is reflected in measured impedance. So it is called bioelectrical impedance analysis. II. Main discussion To measure the body fat, let micro alternating current with 50 kHz and 100 μA flow into the body through two input electrode plates. And the resistance composition of the body is calculated by measuring the voltage between two electrodes. Besides this, the information of height, gender, weight and age of a person is needed. The amount of a body fat is calculated by subtracting FFM(Fat Free Mass) from the weight of the user. The three existing calculating methods to measure a fat free mass are shown in equations 4. Its unit is Kg. FFMPAT =4.033+GPAT×(0.734×(H2 /R)+0.096×X)+0.116×A+0.878×G . (1a)
  • 2. Design of Body Fat Measurement System based on the Android Platform www.ijres.org 45 | Page FFMHYU = 0.0005+H2 -GHYU×Z+0.392×W-0.0684×A-5.1841×G+24.678 . (1b) FFMHOOT = 0.61×GHOOT×(H2 /R)+0.25×W+1.31 (1c) . Here, H(cm) denotes the height, R(Ω) the Resistance, X(Ω) the Reactance, A the Age, Z(Ω) the Impedance, W the Weight. G denotes Gender and its value is set as 1 for male and 0 for female. GPAT, GHYU, GHOOT are constants which values are related to the values of Impedance measured from the body of the user. All these values are set as 1 in existing equations. In existing methods, the body fat was measured by attaching two hands and two feet closely on electrodes. As a voltage differential is measured by attaching two left and right fingers in the implementing system, it is hard to apply the existing equation as it is. Therefore, in order to revise this, the impedance and its related constants are used as tuning elements. Fat Mass(FM) is described as in the equation 2. FM = W - FFM(Kg) (2) Body Fat(%BF) can be calculated as in the equation 3. %BF = (FM/W) × 100(%) (3) And Body Mass Index(BMI) is calculated as in the equation 4. BMI = (W/H2 ) (Kg/m2 ) (4) The amount of the body water which measures the water in a body can be calculated as in the equation 5. TBW = 0.59×GTBW×(H2 /R)+0.065×W+1.04(Kg) (5) Based on measured body fat, daily calory consumption can be calculated by substituting user's walking steps into the equation. The equation 6 shows daily calory consumption(C). C = 1.11 × (Ka × W + Kb × Wn ) (6) Ka denotes a living activity index, W a weight, Kb a basal energy metabolic rate, and n=0.424* constant. The measuring equipment for the body fat currently made for a commercial use measures the body fat in various parts of the body by attaching electrodes on legs and arms or by grabbing electrodes attached on handles by two hands. In the implemented system, the body fat can be measured by attaching electrodes on four fingers. Therefore, based on the results from various clinical demonstrations, tuning coefficients of the equation is necessary. From the result of clinical demonstrations, the heart rate result is accurate, but there is much error in the body fat measurement depending on how a user grabs the electrodes with hands. In case when he grabs electrodes properly, there is about 10% error compared to the product for common use. Figure 1 is a block diagram for the body fat equipment developed in this paper. This body fat equipment consists of a signal detection component which measures impedance from the body, an amplifying component which amplifies detected signal, a signal processing component for filtering and formatting the amplified signal, and a transferring component which transfers the processed signal. [Figure 1] Components Diagram for Body Fat Equipment III. Test and Result 3.1 How to measure the body fat
  • 3. Design of Body Fat Measurement System based on the Android Platform www.ijres.org 46 | Page [Figure 2] Circuit Configuration to measure the body fat. The above figure shows a circuit diagram for the measurement of the body fat. In the first component of the circuit diagram, there is a part which causes a frequency of 100 uA, 50 KHz entering into the body. Due to a delicate difference in PCB pattern, in order to overcome the some voltage differential in each board, namely for calibration, 5 KΩ adjustable resistance is provided. A signal occurred in the main circuit enters into the body. And after the signal passes through the body, it comes out by both inverting amplifier and non inverting amplifier. The output signal is analyzed by ADC of MCU and MCU to measure the degree of the final body fat. For Calibration, R58, R59, R60 parts should be connected in the above circuit diagram, and it should be controlled to have +(-) 1Vp-p in TP8 3.2 Result from Implementation This display shows the test result of the body fat measurement by DDR function of Eclipse smart display capturing which is a tool for the development of Android. Figure 8 is the result of the elctrocardiogram signal tested in the embeded board based on Android. [Figure 3] raw Pass filter application display The left signal in the figure 8 is the wave pattern before applying the filter, and the display on the right in figure 8 shows an example of an output when raw Pass filter is applied to the wave pattern of an electrocardiogram. [Figure 4] high Pass filter application
  • 4. Design of Body Fat Measurement System based on the Android Platform www.ijres.org 47 | Page Figure 4 shows an example of a high Pass filter application. It can be seen that the clear electrocardiogram signal has come out by applying filter at channels 1, 2, and 3. [Figure 5] Body fat measurement display of an embedded board based on the Android Platform. Figure 5 shows the display of electrocardiogram. More meaningful electrocardiogram data has come out by applying the filter previously explained. IV. Conclusions and Next Assignment In this paper, for the accurate electrocardiogram measurement, the filter circuit, which can remove Low Pass Filter, High Pass Filter and Power Noise, has been implemented to reduce error caused by the noise of the measuring sensor in the system which measures the body fat signal. In this paper, the App(application) of the body fat measurement has been implemented and displayed, which measures the body fat in the embedded system in which the Android has been porting This means that it can simply diagnose not only an Android link electrocardiogram measurement system but also a real ubiquitous through various communication equipments and their methods. Also, it has a merit to offer effective diagnosis service with low cost through linkage with the USN base system. As this system can transfer data by using communication means based on Android, the next study about information security method should be done at the same time to protect information of a user. References [1.] Albu, J. B., Kovera, A. J., Johnson, J. A. (2000) Fat distribution and health in obesity. Ann N Y Acad Sci 904: 491–501. [2.] Doll, S., Paccaud, F., Bovet, P., Burnier, M., Wietlisbach, V. (2002) Body mass index, abdominal adiposity and blood pressure: consistency of their association across developing and developed countries. Int J Obes Relat Metab Disord 26: 48–57. [3.] Visscher, T. L., Seidell, J. C., Molarius, A., van der Kuip, D., Hofman, A., Witteman, J. C. (2001) A comparison of body mass index, waist-hip ratio and waist circumference as predictors of all-cause mortality among the elderly: the Rotterdam study. Int J Obes Relat Metab Disord 25: 1730–1735. [4.] Rexrode, K. M., Carey, V. J., Hennekens, C. H., et al. (1998) Abdominal adiposity and coronary heart disease in women. JAMA 280: 1843–1848. [5.] Rexrode, K. M., Buring, J. E., Manson, J. E. (2001) Abdominal and total adiposity and risk of coronary heart disease in men. Int J Obes Relat Metab Disord 25: 1047–1056. [6.] Folsom, A. R., Kushi, L. H., Anderson, K. E., et al. (2000) Associations of general and abdominal obesity with multiple health outcomes in older women: the Iowa Women's Health Study. Arch Intern Med 160: 2117–2128. [7.] Peiris, A. N., Sothmann, M. S., Hoffmann, R. G., et al. (1989) Adiposity, fat distribution, and cardiovascular risk. Ann Intern Med 110: 867–872. [8.] Ansari, A., Rholl, A. O. (1986) Pseudopericardial effusion: echocardiographic and computed tomographic correlations. Clin Cardiol 9: 551–555. [9.] Savage, D. D., Garrison, R. J., Brand, F., et al. (1983) Prevalence and correlates of posterior extra echocardiographic spaces in a free-living population based sample (the Framingham study). Am J Cardiol 51: 1207–1212. [10.] Walsh, K. M., Leen, E., MacSween, R. N., Morris, A. J. (1998) Hepatic blood flow changes in chronic hepatitis C measured by duplex Doppler color sonography: relationship to histological features. Dig Dis Sci 43: 2584–2590. [11.] Janssen, I., Heymsfield, S. B., Allison, D. B., Kotler, D. P., Ross, R. (2002) Body mass index and waist circumference independently contribute to the prediction of nonabdominal, abdominal subcutaneous, and visceral fat. Am J Clin Nutr 75: 683– 688. [12.] Marchington, J. M., Mattacks, C. A., Pond, C. M. (1989) Adipose tissue in the mammalian heart and pericardium: structure, fetal development and biochemical properties. Comp Biochem Physiol B 94: 225–232. [13.] Sahn, D., DeMaria, A., Kisslo, J., Weyman, A. (1978) The Committee on M-mode standardization of the American Society of Echocardiography: recommendations regarding quantitation in M-mode echocardiography: results of a survey of echocardiography measurements. Circulation 58: 1072–1083.