Ashcroft | Accuracy White Paper
1
How accurate is your
accuracy statement?
Sam Martin - Ashcroft Product Specialist, Transducers
MS Engineering & Operations Management
BS Chemical Engineering
Accuracy – White Paper
Ashcroft | Accuracy White Paper
2
	
Accuracy Statements Are
Not Created Equal.
Accuracy specifications for pressure instruments don’t
always describe performance in the same manner. From one
manufacturer to the next, they can be as different as apples and
oranges. Unlike gauges, in many regions of the world there are
no governing bodies or societies to establish a consistent set of
standards for pressure sensors and transducers. That leaves the
door open for manufacturers to publish performance data and
statistically derived accuracy specifications that appear more
precise than they actually are.
LOOK FOR THE RED FLAGS.
Root of the sum squared (RSS) and best fit straight line
(BFSL) specifications are red flags that specifying engineers
and procurement professionals should look for in accuracy
statements. RSS and BFSL specifications cannot guarantee that
every pressure point indicated by the instrument falls within the
stated error band. Since these methods do not require the zero
or span point to be fixed during testing, the zero and span will
often have to be reset after the instrument has been installed to
achieve the desired accuracy. To set the span, a portable high
accuracy secondary standard will be required. This could have
a negative impact on:
n 	Initial production and set-up costs
	
n Field interchangeability
n 	The additive overall system performance specification
What about the effect of external factors such as ambient
temperature deviation? Temperature errors are usually listed
as separate additive inaccuracy statements. These statements
typically describe the resulting consequences to accuracy
when the instrument operates in temperatures that diverge from
standard room temperature. These additional specification is
an important part of understanding accuracy under extreme
operating conditions and should be taken into consideration.
However, since temperature is a unique variable and not part of
the accuracy specification, we will omit it from this discussion
and consider that all of the examples and calculations herein are
based on an operating environment of room temperature
(70-72° F / 21.1-22.2°C).
WE’LL HELP YOU DECIPHER THE NUMBERS.
This paper will address different methods of deriving accuracy
specifications, with a concentration on pressure transducers
and transmitters, where this phenomenon is most prevalent.
However, the concepts may be universally applied to other
pressure instruments. The goal is to inform engineers and end
users how to decipher accuracy statements so they can ensure
that the instrument they are choosing honestly satisfies their
accuracy specification requirements.
Accuracy. How It’s Defined.
How It’s Shown. What To
Really Look For.
WHAT IS ACCURACY?
Accuracy is the departure of a measured value of a quantity
from the accepted standard for that quantity (Gillum, p.395). The
accuracy of a transducer is derived from its inaccuracy, which
is defined as the maximum positive and negative difference
between any measured value and the ideal value. This includes
increasing and decreasing inputs over multiple test cycles. This
value is expressed as a percentage of the ideal output span
(IEC 61298-2, p.12). The accuracy of a transducer does not
completely describe how the transducer measures pressure. For
this reason, accuracy should be incorporated into an accuracy
statement.
WHAT ARE THE COMPONENTS OF AN ACCURACY
STATEMENT?
A transducer’s accuracy statement should include all potential
sources of error relevant to the specific application. The most
common sources of error for a transmitter are non-linearity error,
hysteresis error, non-repeatability error, zero offset error, and
span setting error. Zero and span temperature coefficient errors
are common when a sensor is operated outside of its reference
temperature. To ensure accuracy, specifying engineers need to
know how accuracy is tested.
Non-linearity errors are defined as the deviation between
a transducer’s linear trend line and the ideal output of the
transducer. To calculate the non-linearity error one must first
calculate the linearity. This is accomplished by measuring the
output of the transducer over its calibrated full span to create
Ashcroft | Accuracy White Paper
3
0.30
0.25
0.20
0.15
0.10
0.05
0
-0.05
-0.10
-0.15
-0.20
-0.25
-0.30
Average
error,
percent
of
output
average up average down average BFSL linearity
Percent span
0 10 20 30 40 50 60 70 80 90 100
BEST FIT STRAIGHT LINE LINEARITY
a calibration curve. The calibration curve is then plotted and fit
with a trend line. The three most common transducer linearity
trend lines are terminal point, best fit straight line (BFSL), and
zero-based best fit straight line.
	
Terminal point linearity is determined by drawing a straight
line that coincides with the calibration curve at the full span
point and at the zero point (IEC 61298-2, p.13). One of the major
benefits of terminal point linearity is that it includes zero offset
error and span setting error. This is because terminal point
linearity is fixed to the sensor’s full span point and zero point.
A benefit of sensors tested this way is that they can be easily
interchanged with other units in the field because their accuracy
comes set from the factory and includes errors associated with
setting the zero and span points.
Best fit straight line (BFSL) linearity is determined by drawing
a trend line that minimizes the maximum deviation between
the trend line and any data point on the calibration curve (IEC
61298-2, p.13). A best fit straight line does not need to pass
through the zero or span points. The only requirements are that
it is a straight line and that it minimizes the maximum deviation
of the calibration curve. This is problematic because sensors
calibrated using the best fit straight line method will need to be
re-calibrated once installed in the field to correct for zero offsets
and span setting errors. Since these errors can have a noticeable
effect on the sensor’s accuracy, they should be included in the
calculation.
Zero offset and span setting errors are typically written as
either a percentage of span or a percentage of output. Some
sensor manufacturers list the zero offset and span setting errors
under the sensor’s electrical data or specifications. For a sensor
with a voltage output, the zero and span offset error will likely be
shown as a ± mV reading. For sensors with a current output the
zero and span offset error will likely be shown as a ± mA reading.
To convert this value into a percentage, the ± mV or ± mA error
should be divided by the sensor’s output span.
	
Correcting zero offset and span setting errors will have an effect
on the overall accuracy of the transducer. Changes made to
the zero offset will cause a linear shift of the trend line, while
changes to the span will affect the slope of the line. If zero and
span setting errors are corrected, the accuracy of the transducer
should be reevaluated.
0.30
0.25
0.20
0.15
0.10
0.05
0
-0.05
-0.10
-0.15
-0.20
-0.25
-0.30
Average
error,
percent
of
output
average up average down average Terminal Point linearity
Percent span
0 10 20 30 40 50 60 70 80 90 100
TERMINAL POINT LINEARITY
Output Span = (output at full span) - (output at zero span)
(± zero output error)
(± span output error)
(output span)
(output span)
Zero Offset Error =
Span Setting Error =
Ashcroft | Accuracy White Paper
4
Zero based best fit straight line linearity is determined
by drawing a trend line that is fixed to the zero value of the
calibration curve but also minimizes the potential deviation
between the trend line and all the other data points on the curve.
Zero based best fit straight line linearity is less common than
terminal point linearity or standard best fit straight line linearity.
It provides a more accurate measurement of the sensor’s non-
linearity error than the best fit straight line method, but it is a less
inclusive measurement of error than the terminal point method.
Similar to the standard best fit straight line linearity, the zero
based best fit straight line linearity does not account for span
offset error. Thus, the span offset error must also be calculated
and added to the total error statement when calculating a
sensor’s true accuracy.
Hysteresis is defined as “property of a device or instrument
whereby it gives different output values in relation to its input
values depending on the directional sequence in which the
input values have been applied” (IEC 61298-2, p. 8). More
simply, hysteresis describes the difference between the pressure
sensor’s output when pressure is increasing compared to when
it is decreasing. This error is commonly given as a percent of
full span or output span. Hysteresis is considered a property
of a sensor and should only account for a small portion of an
accuracy statement. Hysteresis is considered to be a cumulative
error when referring to an accuracy statement. It is independent
of linearity, and non-repeatability.
Repeatability is defined as closeness of agreement between the
results of successive measurements of the same measurement,
carried out under the same conditions, using the same
procedure, by the same observer, with the same instrument,
used under the same condition, in the same laboratory, over a
relatively short interval of time (IEC 60050, 311-06-05). The error
is commonly calculated by testing a sensor over its pressure
range three times in a row, under identical conditions, and
finding the maximum difference between measurements, which
is often defined as a percent of full span. This is commonly
referred to as the non-repeatability error. Similar to hysteresis,
repeatability is considered to be a cumulative error. When
analyzing an accuracy statement it should be considered
independent of linearity, and hysteresis.
0.30
0.25
0.20
0.15
0.10
0.05
0
-0.05
-0.10
-0.15
-0.20
-0.25
-0.30
Average
error,
percent
of
output
average up average down average Zero Based linearity
Percent of span
0 10 20 30 40 50 60 70 80 90 100
ZERO BASED BEST FIT STRAIGHT LINE LINEARITY
first
cycle up
first
cycle down
second
cycle up
second
cycle down
third
cycle up
third
cycle down
0.30
0.25
0.20
0.15
0.10
0.05
0
-0.05
-0.10
-0.15
-0.20
-0.25
-0.30
Average
error,
percent
of
output
Percent of span
REPEATABILITY
0 10 20 30 40 50 60 70 80 90 100
0.30
0.25
0.20
0.15
0.10
0.05
0
-0.05
-0.10
-0.15
-0.20
-0.25
-0.30
Average
error,
percent
of
output
average up average down
Percent of span
0 10 20 30 40 50 60 70 80 90 100
HYSTERESIS
Ashcroft | Accuracy White Paper
5
ashcroft.com info@ashcroft.com 1.800.328.8258
THE COMPONENTS OF ACCURACY ARE WELL DEFINED.
METHODS FOR COMBINING THE INDIVIDUAL ERRORS OF
A SENSOR’S ACCURACY STATEMENT ARE NOT.
The summed error of a sensor’s accuracy statement is
commonly calculated using the additive method (a.k.a. worst
case method) or the root of the sum squared (RSS) method. For
the summed error to be useful, it should include non-linearity
errors, non-repeatability errors, and hysteresis errors. If the non-
linearity error is calculated using a best fit straight line (BFSL)
method, the zero setting error and span setting errors should
also be included in the accuracy statement. The summed error
of a sensor’s accuracy statement is different from the sensor’s
real accuracy. In other words, a sensor’s accuracy will always
be defined by the maximum positive and negative deviations
from any measured value to the ideal value, for increasing and
decreasing inputs for any test cycle.
	
The worst case method is a summation of the individual
components of accuracy that make up the accuracy statement.
The worst case accuracy method is based on the fact that
a sensor’s accuracy could be equal to the sum of all of the
possible errors that make up the accuracy statement. This is
very different compared to the assumption the root of the sum
squared (RSS) method makes.
	
The root of the sum squared (RSS) method is a statistical
calculation commonly used to determine the sum of errors in
an assembly. Taking the square root of the sum of the squares
provides a few statistical benefits. First, squaring individual error
values guarantees that all of the error values being added will
be positive, meaning that negative errors will not offset positive
errors. Second, squaring also weights the error values; errors
greater than 1% are more heavily weighted, errors less than 1%
are minimized and errors equal to 1% will remain the same. Most
errors associated with the components of accuracy are typically
less than 1%. Thus, the individual errors will be minimized before
they are summed and then the square root will be taken after the
minimized errors are summed. The final result is that the total
error calculated with the root of the sum squared (RSS)
method will be less than the worst case method that focuses on
summing each of the individual errors.
THE BOTTOM LINE: CRITICAL APPLICATIONS
REQUIRE ACCURATE SENSORS.
	
It’s important to understand the accuracy statement for the
sensors used in an application. These statements vary from
manufacturer to manufacturer. The sensor’s error should not only
be determined by a percent of span but also by what that error
includes or excludes.
	
The most conservative accuracy statements will be a
summation of terminal point linearity errors, which includes
zero and span errors, non-repeatability errors, and hysteresis
errors. These errors all contribute to defining how a transducer
will perform but a transducer’s accuracy will always be defined
by its maximum positive and negative deviation from an ideal
standard.
	
The least conservative accuracy statements will utilize
an RSS accuracy statement and best fit straight line (BFSL)
linearity errors. Repeatability errors and hysteresis errors
may not be listed, and zero and span setting errors will likely
be intentionally omitted or disguised as an electrical output.
If accuracy and error data are not clearly defined on the
manufacturer’s datasheet, contact them and inquire about their
accuracy statement	and compare it to statements from other
manufacturers.
	Remember, not all accuracy statements are created equal.
WORKS CITED.
“IEC 60050 - International Electrotechnical Vocabulary -
Welcome.” IEC 60050 - International
	
Gillum, D. (1995). Industrial pressure, level & density
measurement. Research Triangle Park, N.C., U.S.A.: ISA.
	
Process Measurement and Control Devices--general Methods
and Procedures for Evaluating Performance Dispositifs De Mesure
Et De Commande De Processus--méthodes Et Procédures
Générales D’évaluation Des Performances,. Ed. 2.0. ed. Geneva:
International Electrotechnical Commission, 2008. Print.
(non-linearity error) + (non-repeatability error) + (hysteresis error)
Worst Case Error =
(Non-Linearity)2
+ (Non-Repeatability)2
+ (Hysteresis)2
RSS Error =
Ash WP-1 Accuracy
©Ashcroft Inc. 2015-07/15
All sales subject to standard terms and conditions.
All specifications are subject to change without notice.

More Related Content

What's hot

Instrument lesson
Instrument lessonInstrument lesson
Instrument lessonHung Thai
 
Comparators...4
Comparators...4Comparators...4
Comparators...4Nouman Ali
 
L7 measurement system
L7 measurement systemL7 measurement system
L7 measurement systemtaruian
 
Basics Of Instrumentation
Basics Of InstrumentationBasics Of Instrumentation
Basics Of InstrumentationVinoth Ganesh
 
Basic concepts of measurement
Basic concepts of measurementBasic concepts of measurement
Basic concepts of measurementkaroline Enoch
 
Basic instrumentation
Basic instrumentationBasic instrumentation
Basic instrumentationss61
 
Basic concepts of instrumentation
Basic concepts of instrumentationBasic concepts of instrumentation
Basic concepts of instrumentationG. Udhaya Sankar
 
Basic to Site Specific Instrument Training
Basic to Site Specific Instrument TrainingBasic to Site Specific Instrument Training
Basic to Site Specific Instrument TrainingTheunis Venter
 
THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM
THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEMTHE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM
THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEMieijjournal
 
How to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization EquipmentHow to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization EquipmentSetra Systems
 
Practical Uncertainty Estimation in Load Cell Calibration
Practical Uncertainty Estimation in Load Cell CalibrationPractical Uncertainty Estimation in Load Cell Calibration
Practical Uncertainty Estimation in Load Cell CalibrationInterface Inc.
 

What's hot (17)

Instrument lesson
Instrument lessonInstrument lesson
Instrument lesson
 
Comparators...4
Comparators...4Comparators...4
Comparators...4
 
L7 measurement system
L7 measurement systemL7 measurement system
L7 measurement system
 
Dial indicator
Dial indicatorDial indicator
Dial indicator
 
Basics instrument andcontrol SYSTEMS
Basics instrument andcontrol SYSTEMSBasics instrument andcontrol SYSTEMS
Basics instrument andcontrol SYSTEMS
 
Basics Of Instrumentation
Basics Of InstrumentationBasics Of Instrumentation
Basics Of Instrumentation
 
Basic concepts of measurement
Basic concepts of measurementBasic concepts of measurement
Basic concepts of measurement
 
Basic instrumentation
Basic instrumentationBasic instrumentation
Basic instrumentation
 
Mechanical measurements and Measurement systems
Mechanical measurements and Measurement systemsMechanical measurements and Measurement systems
Mechanical measurements and Measurement systems
 
Basic concepts of instrumentation
Basic concepts of instrumentationBasic concepts of instrumentation
Basic concepts of instrumentation
 
Basic to Site Specific Instrument Training
Basic to Site Specific Instrument TrainingBasic to Site Specific Instrument Training
Basic to Site Specific Instrument Training
 
THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM
THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEMTHE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM
THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST SYSTEM
 
How to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization EquipmentHow to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization Equipment
 
pressure transmitter
pressure transmitterpressure transmitter
pressure transmitter
 
Flowmeter course
Flowmeter courseFlowmeter course
Flowmeter course
 
Practical Uncertainty Estimation in Load Cell Calibration
Practical Uncertainty Estimation in Load Cell CalibrationPractical Uncertainty Estimation in Load Cell Calibration
Practical Uncertainty Estimation in Load Cell Calibration
 
i to p conveter
i to p  conveteri to p  conveter
i to p conveter
 

Similar to Accuracy whitepaper

Fundamentals on Measurement and uncertainty
Fundamentals on Measurement  and  uncertaintyFundamentals on Measurement  and  uncertainty
Fundamentals on Measurement and uncertaintyBerkefede
 
Mechanical measurement 1
Mechanical measurement 1Mechanical measurement 1
Mechanical measurement 1Sangram Petkar
 
Chapter 1 Measurement System
Chapter 1  Measurement  SystemChapter 1  Measurement  System
Chapter 1 Measurement SystemChe Ku Sabri
 
Instrumentation_ Measurment_course_Chapter_2_.pdf
Instrumentation_ Measurment_course_Chapter_2_.pdfInstrumentation_ Measurment_course_Chapter_2_.pdf
Instrumentation_ Measurment_course_Chapter_2_.pdfNatnaelTamrat2
 
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.ppt
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.pptSENSORS_AND_INST.UNIT_1_REG_17__Presentation.ppt
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.pptSenthilKumarP45
 
calabration-principles-chapter1.pdf
calabration-principles-chapter1.pdfcalabration-principles-chapter1.pdf
calabration-principles-chapter1.pdfsharonmarishkawilfre
 
Calabration principles-chapter1
Calabration principles-chapter1Calabration principles-chapter1
Calabration principles-chapter1Complaint2015
 
Calibrating the Measuring Chain
Calibrating the Measuring ChainCalibrating the Measuring Chain
Calibrating the Measuring ChainTacuna Systems
 
01. Sensors and Actuators.ppt
01. Sensors and Actuators.ppt01. Sensors and Actuators.ppt
01. Sensors and Actuators.pptprashant360377
 
Static terms in measurement
Static terms in measurementStatic terms in measurement
Static terms in measurementChetan Mahatme
 
Beie unit - 5
Beie unit - 5Beie unit - 5
Beie unit - 5W3Edify
 
Measurements kkkkkkkkkkkkkkkkkkkch(1)1.pptx
Measurements kkkkkkkkkkkkkkkkkkkch(1)1.pptxMeasurements kkkkkkkkkkkkkkkkkkkch(1)1.pptx
Measurements kkkkkkkkkkkkkkkkkkkch(1)1.pptxAhmedWail2
 
Simulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_Testing
Simulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_TestingSimulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_Testing
Simulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_TestingRobert Hicks
 
Characterics of measuring system.pptx
Characterics of measuring system.pptxCharacterics of measuring system.pptx
Characterics of measuring system.pptxmoviefreak5
 
2_Measurement+and+Error.pdf
2_Measurement+and+Error.pdf2_Measurement+and+Error.pdf
2_Measurement+and+Error.pdfMahamad Jawhar
 
Pressure_Measurement.ppt
Pressure_Measurement.pptPressure_Measurement.ppt
Pressure_Measurement.pptKaviyarasanS9
 
Mechanical measurement chapter 1
Mechanical measurement chapter 1Mechanical measurement chapter 1
Mechanical measurement chapter 1sudhanvavk
 
Introduction about Instrumentation.ppt
Introduction about Instrumentation.pptIntroduction about Instrumentation.ppt
Introduction about Instrumentation.pptMuhammadShoaibRabban2
 

Similar to Accuracy whitepaper (20)

Fundamentals on Measurement and uncertainty
Fundamentals on Measurement  and  uncertaintyFundamentals on Measurement  and  uncertainty
Fundamentals on Measurement and uncertainty
 
Mechanical measurement 1
Mechanical measurement 1Mechanical measurement 1
Mechanical measurement 1
 
Chapter 1 Measurement System
Chapter 1  Measurement  SystemChapter 1  Measurement  System
Chapter 1 Measurement System
 
Instrumentation_ Measurment_course_Chapter_2_.pdf
Instrumentation_ Measurment_course_Chapter_2_.pdfInstrumentation_ Measurment_course_Chapter_2_.pdf
Instrumentation_ Measurment_course_Chapter_2_.pdf
 
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.ppt
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.pptSENSORS_AND_INST.UNIT_1_REG_17__Presentation.ppt
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.ppt
 
calabration-principles-chapter1.pdf
calabration-principles-chapter1.pdfcalabration-principles-chapter1.pdf
calabration-principles-chapter1.pdf
 
Calabration principles-chapter1
Calabration principles-chapter1Calabration principles-chapter1
Calabration principles-chapter1
 
Calibrating the Measuring Chain
Calibrating the Measuring ChainCalibrating the Measuring Chain
Calibrating the Measuring Chain
 
01. Sensors and Actuators.ppt
01. Sensors and Actuators.ppt01. Sensors and Actuators.ppt
01. Sensors and Actuators.ppt
 
Static terms in measurement
Static terms in measurementStatic terms in measurement
Static terms in measurement
 
ZERO100L ultrasonic level meter manual. four lines
ZERO100L  ultrasonic level meter manual. four linesZERO100L  ultrasonic level meter manual. four lines
ZERO100L ultrasonic level meter manual. four lines
 
Beie unit - 5
Beie unit - 5Beie unit - 5
Beie unit - 5
 
Measurements kkkkkkkkkkkkkkkkkkkch(1)1.pptx
Measurements kkkkkkkkkkkkkkkkkkkch(1)1.pptxMeasurements kkkkkkkkkkkkkkkkkkkch(1)1.pptx
Measurements kkkkkkkkkkkkkkkkkkkch(1)1.pptx
 
Simulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_Testing
Simulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_TestingSimulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_Testing
Simulation_of_Fluid_Mediums_for_Suction_Tube_Degradation_Testing
 
Characterics of measuring system.pptx
Characterics of measuring system.pptxCharacterics of measuring system.pptx
Characterics of measuring system.pptx
 
2_Measurement+and+Error.pdf
2_Measurement+and+Error.pdf2_Measurement+and+Error.pdf
2_Measurement+and+Error.pdf
 
Pressure_Measurement.ppt
Pressure_Measurement.pptPressure_Measurement.ppt
Pressure_Measurement.ppt
 
Mechanical measurement chapter 1
Mechanical measurement chapter 1Mechanical measurement chapter 1
Mechanical measurement chapter 1
 
Chapter 1
Chapter 1Chapter 1
Chapter 1
 
Introduction about Instrumentation.ppt
Introduction about Instrumentation.pptIntroduction about Instrumentation.ppt
Introduction about Instrumentation.ppt
 

Recently uploaded

Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxupamatechverse
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxpurnimasatapathy1234
 
Analog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog ConverterAnalog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog ConverterAbhinavSharma374939
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...ranjana rawat
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Christo Ananth
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Dr.Costas Sachpazis
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptxDecoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptxJoão Esperancinha
 

Recently uploaded (20)

Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptx
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCRCall Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptx
 
Analog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog ConverterAnalog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog Converter
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
 
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINEDJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptxDecoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 

Accuracy whitepaper

  • 1. Ashcroft | Accuracy White Paper 1 How accurate is your accuracy statement? Sam Martin - Ashcroft Product Specialist, Transducers MS Engineering & Operations Management BS Chemical Engineering Accuracy – White Paper
  • 2. Ashcroft | Accuracy White Paper 2 Accuracy Statements Are Not Created Equal. Accuracy specifications for pressure instruments don’t always describe performance in the same manner. From one manufacturer to the next, they can be as different as apples and oranges. Unlike gauges, in many regions of the world there are no governing bodies or societies to establish a consistent set of standards for pressure sensors and transducers. That leaves the door open for manufacturers to publish performance data and statistically derived accuracy specifications that appear more precise than they actually are. LOOK FOR THE RED FLAGS. Root of the sum squared (RSS) and best fit straight line (BFSL) specifications are red flags that specifying engineers and procurement professionals should look for in accuracy statements. RSS and BFSL specifications cannot guarantee that every pressure point indicated by the instrument falls within the stated error band. Since these methods do not require the zero or span point to be fixed during testing, the zero and span will often have to be reset after the instrument has been installed to achieve the desired accuracy. To set the span, a portable high accuracy secondary standard will be required. This could have a negative impact on: n Initial production and set-up costs n Field interchangeability n The additive overall system performance specification What about the effect of external factors such as ambient temperature deviation? Temperature errors are usually listed as separate additive inaccuracy statements. These statements typically describe the resulting consequences to accuracy when the instrument operates in temperatures that diverge from standard room temperature. These additional specification is an important part of understanding accuracy under extreme operating conditions and should be taken into consideration. However, since temperature is a unique variable and not part of the accuracy specification, we will omit it from this discussion and consider that all of the examples and calculations herein are based on an operating environment of room temperature (70-72° F / 21.1-22.2°C). WE’LL HELP YOU DECIPHER THE NUMBERS. This paper will address different methods of deriving accuracy specifications, with a concentration on pressure transducers and transmitters, where this phenomenon is most prevalent. However, the concepts may be universally applied to other pressure instruments. The goal is to inform engineers and end users how to decipher accuracy statements so they can ensure that the instrument they are choosing honestly satisfies their accuracy specification requirements. Accuracy. How It’s Defined. How It’s Shown. What To Really Look For. WHAT IS ACCURACY? Accuracy is the departure of a measured value of a quantity from the accepted standard for that quantity (Gillum, p.395). The accuracy of a transducer is derived from its inaccuracy, which is defined as the maximum positive and negative difference between any measured value and the ideal value. This includes increasing and decreasing inputs over multiple test cycles. This value is expressed as a percentage of the ideal output span (IEC 61298-2, p.12). The accuracy of a transducer does not completely describe how the transducer measures pressure. For this reason, accuracy should be incorporated into an accuracy statement. WHAT ARE THE COMPONENTS OF AN ACCURACY STATEMENT? A transducer’s accuracy statement should include all potential sources of error relevant to the specific application. The most common sources of error for a transmitter are non-linearity error, hysteresis error, non-repeatability error, zero offset error, and span setting error. Zero and span temperature coefficient errors are common when a sensor is operated outside of its reference temperature. To ensure accuracy, specifying engineers need to know how accuracy is tested. Non-linearity errors are defined as the deviation between a transducer’s linear trend line and the ideal output of the transducer. To calculate the non-linearity error one must first calculate the linearity. This is accomplished by measuring the output of the transducer over its calibrated full span to create
  • 3. Ashcroft | Accuracy White Paper 3 0.30 0.25 0.20 0.15 0.10 0.05 0 -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 Average error, percent of output average up average down average BFSL linearity Percent span 0 10 20 30 40 50 60 70 80 90 100 BEST FIT STRAIGHT LINE LINEARITY a calibration curve. The calibration curve is then plotted and fit with a trend line. The three most common transducer linearity trend lines are terminal point, best fit straight line (BFSL), and zero-based best fit straight line. Terminal point linearity is determined by drawing a straight line that coincides with the calibration curve at the full span point and at the zero point (IEC 61298-2, p.13). One of the major benefits of terminal point linearity is that it includes zero offset error and span setting error. This is because terminal point linearity is fixed to the sensor’s full span point and zero point. A benefit of sensors tested this way is that they can be easily interchanged with other units in the field because their accuracy comes set from the factory and includes errors associated with setting the zero and span points. Best fit straight line (BFSL) linearity is determined by drawing a trend line that minimizes the maximum deviation between the trend line and any data point on the calibration curve (IEC 61298-2, p.13). A best fit straight line does not need to pass through the zero or span points. The only requirements are that it is a straight line and that it minimizes the maximum deviation of the calibration curve. This is problematic because sensors calibrated using the best fit straight line method will need to be re-calibrated once installed in the field to correct for zero offsets and span setting errors. Since these errors can have a noticeable effect on the sensor’s accuracy, they should be included in the calculation. Zero offset and span setting errors are typically written as either a percentage of span or a percentage of output. Some sensor manufacturers list the zero offset and span setting errors under the sensor’s electrical data or specifications. For a sensor with a voltage output, the zero and span offset error will likely be shown as a ± mV reading. For sensors with a current output the zero and span offset error will likely be shown as a ± mA reading. To convert this value into a percentage, the ± mV or ± mA error should be divided by the sensor’s output span. Correcting zero offset and span setting errors will have an effect on the overall accuracy of the transducer. Changes made to the zero offset will cause a linear shift of the trend line, while changes to the span will affect the slope of the line. If zero and span setting errors are corrected, the accuracy of the transducer should be reevaluated. 0.30 0.25 0.20 0.15 0.10 0.05 0 -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 Average error, percent of output average up average down average Terminal Point linearity Percent span 0 10 20 30 40 50 60 70 80 90 100 TERMINAL POINT LINEARITY Output Span = (output at full span) - (output at zero span) (± zero output error) (± span output error) (output span) (output span) Zero Offset Error = Span Setting Error =
  • 4. Ashcroft | Accuracy White Paper 4 Zero based best fit straight line linearity is determined by drawing a trend line that is fixed to the zero value of the calibration curve but also minimizes the potential deviation between the trend line and all the other data points on the curve. Zero based best fit straight line linearity is less common than terminal point linearity or standard best fit straight line linearity. It provides a more accurate measurement of the sensor’s non- linearity error than the best fit straight line method, but it is a less inclusive measurement of error than the terminal point method. Similar to the standard best fit straight line linearity, the zero based best fit straight line linearity does not account for span offset error. Thus, the span offset error must also be calculated and added to the total error statement when calculating a sensor’s true accuracy. Hysteresis is defined as “property of a device or instrument whereby it gives different output values in relation to its input values depending on the directional sequence in which the input values have been applied” (IEC 61298-2, p. 8). More simply, hysteresis describes the difference between the pressure sensor’s output when pressure is increasing compared to when it is decreasing. This error is commonly given as a percent of full span or output span. Hysteresis is considered a property of a sensor and should only account for a small portion of an accuracy statement. Hysteresis is considered to be a cumulative error when referring to an accuracy statement. It is independent of linearity, and non-repeatability. Repeatability is defined as closeness of agreement between the results of successive measurements of the same measurement, carried out under the same conditions, using the same procedure, by the same observer, with the same instrument, used under the same condition, in the same laboratory, over a relatively short interval of time (IEC 60050, 311-06-05). The error is commonly calculated by testing a sensor over its pressure range three times in a row, under identical conditions, and finding the maximum difference between measurements, which is often defined as a percent of full span. This is commonly referred to as the non-repeatability error. Similar to hysteresis, repeatability is considered to be a cumulative error. When analyzing an accuracy statement it should be considered independent of linearity, and hysteresis. 0.30 0.25 0.20 0.15 0.10 0.05 0 -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 Average error, percent of output average up average down average Zero Based linearity Percent of span 0 10 20 30 40 50 60 70 80 90 100 ZERO BASED BEST FIT STRAIGHT LINE LINEARITY first cycle up first cycle down second cycle up second cycle down third cycle up third cycle down 0.30 0.25 0.20 0.15 0.10 0.05 0 -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 Average error, percent of output Percent of span REPEATABILITY 0 10 20 30 40 50 60 70 80 90 100 0.30 0.25 0.20 0.15 0.10 0.05 0 -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 Average error, percent of output average up average down Percent of span 0 10 20 30 40 50 60 70 80 90 100 HYSTERESIS
  • 5. Ashcroft | Accuracy White Paper 5 ashcroft.com info@ashcroft.com 1.800.328.8258 THE COMPONENTS OF ACCURACY ARE WELL DEFINED. METHODS FOR COMBINING THE INDIVIDUAL ERRORS OF A SENSOR’S ACCURACY STATEMENT ARE NOT. The summed error of a sensor’s accuracy statement is commonly calculated using the additive method (a.k.a. worst case method) or the root of the sum squared (RSS) method. For the summed error to be useful, it should include non-linearity errors, non-repeatability errors, and hysteresis errors. If the non- linearity error is calculated using a best fit straight line (BFSL) method, the zero setting error and span setting errors should also be included in the accuracy statement. The summed error of a sensor’s accuracy statement is different from the sensor’s real accuracy. In other words, a sensor’s accuracy will always be defined by the maximum positive and negative deviations from any measured value to the ideal value, for increasing and decreasing inputs for any test cycle. The worst case method is a summation of the individual components of accuracy that make up the accuracy statement. The worst case accuracy method is based on the fact that a sensor’s accuracy could be equal to the sum of all of the possible errors that make up the accuracy statement. This is very different compared to the assumption the root of the sum squared (RSS) method makes. The root of the sum squared (RSS) method is a statistical calculation commonly used to determine the sum of errors in an assembly. Taking the square root of the sum of the squares provides a few statistical benefits. First, squaring individual error values guarantees that all of the error values being added will be positive, meaning that negative errors will not offset positive errors. Second, squaring also weights the error values; errors greater than 1% are more heavily weighted, errors less than 1% are minimized and errors equal to 1% will remain the same. Most errors associated with the components of accuracy are typically less than 1%. Thus, the individual errors will be minimized before they are summed and then the square root will be taken after the minimized errors are summed. The final result is that the total error calculated with the root of the sum squared (RSS) method will be less than the worst case method that focuses on summing each of the individual errors. THE BOTTOM LINE: CRITICAL APPLICATIONS REQUIRE ACCURATE SENSORS. It’s important to understand the accuracy statement for the sensors used in an application. These statements vary from manufacturer to manufacturer. The sensor’s error should not only be determined by a percent of span but also by what that error includes or excludes. The most conservative accuracy statements will be a summation of terminal point linearity errors, which includes zero and span errors, non-repeatability errors, and hysteresis errors. These errors all contribute to defining how a transducer will perform but a transducer’s accuracy will always be defined by its maximum positive and negative deviation from an ideal standard. The least conservative accuracy statements will utilize an RSS accuracy statement and best fit straight line (BFSL) linearity errors. Repeatability errors and hysteresis errors may not be listed, and zero and span setting errors will likely be intentionally omitted or disguised as an electrical output. If accuracy and error data are not clearly defined on the manufacturer’s datasheet, contact them and inquire about their accuracy statement and compare it to statements from other manufacturers. Remember, not all accuracy statements are created equal. WORKS CITED. “IEC 60050 - International Electrotechnical Vocabulary - Welcome.” IEC 60050 - International Gillum, D. (1995). Industrial pressure, level & density measurement. Research Triangle Park, N.C., U.S.A.: ISA. Process Measurement and Control Devices--general Methods and Procedures for Evaluating Performance Dispositifs De Mesure Et De Commande De Processus--méthodes Et Procédures Générales D’évaluation Des Performances,. Ed. 2.0. ed. Geneva: International Electrotechnical Commission, 2008. Print. (non-linearity error) + (non-repeatability error) + (hysteresis error) Worst Case Error = (Non-Linearity)2 + (Non-Repeatability)2 + (Hysteresis)2 RSS Error = Ash WP-1 Accuracy ©Ashcroft Inc. 2015-07/15 All sales subject to standard terms and conditions. All specifications are subject to change without notice.