3. Biomed Report
• Power up and self test ok.
• Tested with QA45III analyser.
• ECG trace ok on leads and paddles monitoring.
• Defibrillator output tested, selected energy = delivered value.
• Mains and battery in good working order.
• Message displayed ‘maintenance due’ - without pacer.
• System comes with AED cable and set of 2 single use paddles.
• Printer ok - out of paper.