Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Pacemakers and implantable cardiac defibrillators
1. Cardiac pacemakers and
implanted defibrillators
Dr Andrew Crofton
Emergency Registrar
Disclaimer: https://criticalcarecollaborative.com/disclaimer/
2. Indications for emergency pacing
Symptomatic bradycardia
Symptomatic high-grade AV block (Mobitz II or complete)
Severe sick sinus syndrome with asystole >3s and syncope
Overdrive pacing for torsades de pointes
Overdrive pacing for recurrent monomorphic VT
Risk of inducing VF and limited by machine pacing limit of 180
Overdrive pacing of unstable SVT
Only once pharmacological and electrical cardioversion have failed
3. Emergency pacing
Transcutaneous pacing for severe hypotension with bradycardia
70% survive with good neurological outcome vs. 15% with non-pacing modalities
May be able to pace RV infarct when transvenous pacing fails (as can directly pace LV)
4. Technique for transcutaneous pacing
External pads and electrodes for monitoring required
If bradyasystolic peri-arrest, ramp up current to 100mA and titrate down once stabilised
If less severe compromise, can titrate up from 10mA to usually 50-100mA and then maintain at 1.25x the threshold
Set rate to 80bpm and increase by 10bpm until perfusing (up to 100/min)
May be fixed or demand (synchronous)
Fixed (asynchronous) pacing carries risk of R on T, however, there is little outcome data to support a preference for either
Need electrical and mechanical capture
US can be helpful for ensuring mechanical capture if poorly perfused
Reasons for failed capture may include
Inadequate current
Faulty electrical contact
Electrode placement
Patient size
Underlying pathology
PTX, severe ischaemia, pericardial effusion or metabolic derangement
5. Resuscitation in patients with permanent
pacemakers
If countershock required, place pads at least 8cm from the pulse generator
After countershock, interrogate the pacemaker to ensure still functioning as can
suffer from:
Pacemaker inhibition due to reversion to noise mode
Deletion (reprogramming)
Circuit damage
Myocardial damage near lead tip caused by current transmission via the electrode to the
myocardial interface
Defibrillation can lead to global myocardial ischaemia which in turn also increases
the pacing threshold thus causing capture failure
If this occurs, try transcutaneous pacing at a higher current
6. Nomenclature
Letter position I II III IV V
Category Chamber(s) paced Chamber(s) sensed Response to sensing Programmability, rate Antiarrhythmic
functions
O, none
A, atria
V, ventricle
D, dual (A+V)
S, single chamber
O, none
A, atria
V, ventricle
D, dual (A+V)
S, single chamber
O, none
T, triggered
I, inhibited
D, dual (D + I)
O, none
P, simple program
M, multiprogram
C, communicating
(telemetry)
R, rate modulation
O, none
P, pacing
S, shock
D, dual (P+S)
Most commonly VVI and DDD
The fifth letter is rarely used
Third position D indicates both triggered and inhibition responses can occur
i.e. If atrial activity sensed, atrial pacing is inhibited but triggers ventricular pacing
7. Programmability
Activity sensors: Vibration detectors (accelerometers or piezoelectric crystal)
May respond to certain activities e.g. using a drill
Minute ventilation sensors
Measures impedence between pacemaker unit and electrode
8. Magnet inhibition
Placing magnet over permanent pacemaker causes sensing to be inhibited and
results in asynchronous mode (AOO, VOO, DOO)
Usually at rate of 100
Risk of R on T
9. Pacing modes for bradycardia
Single-chamber pacing
AOO and VOO
Asynchronous and virtually obsolete except in a few emergency situations
AAI
Atrial demand pacing
Indicated in sinus bradycardia providing AV conduction is intact
Only paces if does not sense atrial activity
VVI
Ventricular demand pacing
Most commonly used mode in life-threatening bradycardias
Spontaneous ventricular activity is sensed and low risk of R on T
AV synchrony is lost
10. Pacing modes for bradycardia
Dual chamber pacing
DVI (AV sequential pacing)
Will pace atria, wait and then pace ventricle if no ventricular activity sensed
To maintain AV synchrony in the absence of atrial sensing, pacing rate must be higher than
intrinsic spontaneous atrial rate
Indicated if impaired AV conduction and atrial bradycardia
Not useful if atrial tachyarrhythmias exist as cannot pace faster than them
VDD (Atrial synchronous ventricular inhibited)
Paces only the ventricle but senses both
Sensed P-wave triggers ventricular pacing
11. Pacing modes for bradycardia
Dual chamber pacing
DDD (Dual pacing and sensing)
Atrial impulse triggers ventricular pacing unless senses autonomous ventricular activity
Upper rate limiters prevent ventricular pacing of atrial tachyarrhythmias
If atrial bradycardia with intact AV conduction – Atrial pacing
If sinus rhythm with AV block – synchronised ventricular pacing of intrinsic P waves
If sinus bradycardia with AV block – Sequential atrial then ventricular pacing
Normal sinus rhythm and AV conduction – Inhibition of both atrial and ventricular pacing
Can suffer re-entry loops where PVC is transmitted retrogradely to atria, where it is sensed and
results in ventricular pacing with endless loop
12. Pacing modes for bradycardia
Dual chamber pacing
DDI (AV sequential, non-P-wave synchronous)
Sensing of both but sensed atrial events do not trigger ventricular pacing
Prevents endless loop phenomenon or tracking of SVTs
Useful for SA node dysfunction with episodic atrial tachyarrhythmias
Will simply pace ventricle at backup rate in the setting of atrial tachyarrhythmia
13. Complete AV block in acute MI
Feature Inferior Anterior
Onset Slow Sudden
Type Mobitz I Mobitz 2
Ventricular rate >45 <45
Escape pacemaker Stable unstable
Response to atropine Yes No
Haemodynamic effects No Yes
Permanent pacing No Yes (if high-degree AV
block)
Prognosis Good Terrible
14. Indications for permanent pacing
Class I
Chronic symptomatic 2nd or 3rd degree AV block
SA node dysfunction with documented sinus bradycardia
Recurrent syncope associated with carotid sinus hypersensitivity
Class II-IIa
Asymptomatic complete AV block with average ventricular rate >40 in awake patient
Class IIb (weak supportive evidence)
1st degree block with depressed LV function and symptoms of LV failure
Class III (not indicated)
Asymptomatic 1st degree block or reversible AV block due to drugs
15. Indications for permanent pacing
Other
HOCM (Class IIb)
Usually for symptomatic patients with high gradient in LVOT
DDD pacing with short AV interval caused RV apical activation with altered septal activation to
reduce LVOT gradient and systolic anterior motion of mitral valve
Heart failure
Class IIa for NYHA III/IV patients with dilated or ischaemic cardiomyopathy, QRS >130ms
Cardiac resynchronisation therapy (CRT) is indicated in these patients via biventricular pacing to
improve survival and symptoms
16. Fixed-rate vs. demand
Fixed-rate pulse generators produce an electrical signal regardless of patients own
intrinsic electrical rhythm
Can result in serious arrhythmias if discharges during vulnerable period (T wave)
Demand pacing
Typically discharges if no sensed electrical activity after certain time period
May be inhibited by intrinsic sensed electrical activity or triggered to discharge during
absolute refractory period
17. Pacing in tachyarrhythmias
May be useful for:
SVT: AVNRT, AVRT (rarely required)
Atrial flutter (rapid atrial overdrive pacing)
Unifocal atrial tachycardia (rapid atrial overdrive pacing helpful if re-entry pathway driven
but less so if autonomous focus firing at rapid rate)
Ventricular tachycardia
Should not be used if very rapid ventricular rates >300 or significant haemodynamic instability
exists
No value in sinus tachycardia, AF or VF
18. Pacing in tachyarrhythmias
Torsades
Pace atria or ventricle at 110-120/min
SVT
Atrial pacing at 60-80 and slowly increase to 10-20% faster than spontaneous atrial rate
Atria then paced for around 30 seconds then switched off with ensuing sinus rhythm
If fails, try different atrial pacing site and faster rate
19. Pacing in tachyarrhythmias
VT
Ventricular burst pacing
Pace ventricle at 120% of spontaneous VT rate for 5-10 beats then stop
Can precipitate faster VT and VF
Underdrive ventricular pacing at rate <50% of VT rate is sometimes successful
Overdrive atrial pacing may be useful if 1:1 AV conduction and ventricular rate relatively
slow (120-180/min)
20. Features
Lithium batteries have lifespan of 8-12 years
Most units preset for rates near 70, with pacing interval of 0.84 seconds
Demand pacemakers have built-in refractory period of 0.2 to 0.4 seconds during
which it will not sense, to prevent it being inhibited by its own stimulus
Magnets held over most units will convert them from demand to fixed-rate mode
Can quickly ascertain paced rate, however, should only be performed for short periods
due to risk of arrhythmia in fixed-rate mode (due to stimulus in vulnerable period)
More sophisticated units can be interrogated
21. Pacemaker malfunctions
1) Problems with the pocket
2) Problems with the leads
3) Failure to pace
4) Failure to sense – Leads to fixed rate pacing and risk of arrhythmia
5) Malfunction causing overpacing or runaway pacing
22. Pacemaker evaluation
Evaluation
Examination of pocket
CXR to confirm lead placement and device itself
Electrolytes
Cardiac enzymes
ECG
Interrogation
Magnet fixed-rate testing
23. Pacemaker syndrome
Seen in 20% of patients in early phase with symptoms of syncope, near-syncope, orthostatic dizziness, exercise intolerance, dizziness,
uncomfortable pulsations over neck/abdomen and RUQ pain
AV synchrony and presence of ventriculoatrial conduction are most common in VVI but can be seen with DDI mode
In VVI
If the sinus node is intact, an atrial impulse can occur when the tricuspid and mitral valves are closed with increased jugular and pulmonary
venous pressure leading to symptoms of CCF
Atrial distension can lead to reflex vasodepressor effects mediated by the CNS
If contribution of atrial contraction to late diastolic ventricular filling is important, then orthostatic hypotension can occur
In DDI with AV block
If sinus node discharge rate exceeds the programmed rate of the pacemaker, can get atrial contraction against closed valves with subsequent
pacemaker syndrome
Typically symptoms are mild and patients adapt to them over time
Unfortunately, symptoms are severe in 1/3 and may warrant upgrade from VVI to dual-chamber pacing or lowering the rate of VVI
pacing
If symptoms occur in DDI, then optimising the timing of atrial and ventricular pacing is often required
24. Pacemaker infection
Seen in <1% soon after placement
Mostly S. aureus or S. epidermidis early on
Presents as local inflammation or abscess at site
Skin adherence to the device with discolouration is highly suggestive
If only superficial infection suspected, antibiotics and analgesics with early review is
indicated
If erosion of skin occurs, requires surgical replacement
Cardiac device-related IE can occur presenting with sepsis and positive BC without
local inflammation at pocket
25. Thrombophlebitis and venous obstruction
Extensive venous collaterals exist making this quite rare (0.3-3% of patients)
Site of insertion makes no difference
Symptoms include oedema, pain or venous engorgement ipsilaterally
Treatment is IV heparin then oral NOAC/warfarin
26. Pneumothorax
Seen in 1% of patients after insertion
More common with subclavian insertion technique
27. Undersensing
Failure to sense intrinsic cardiac activity with subsequent asynchronous, fixed-rate
pacing
Causes include increased threshold (exit block), poor lead contact, new BBB, inferior
MI or programming issues
Pacing spikes within QRS complexes is highly suggestive of this
Lead dislodgement usually occurs within 2 days of insertion
28. Oversensing
Pacemaker senses electrical activity not from atria or ventricles; it is thus inhibited, and
generation of pacemaker impulse suppressed leading to bradycardia
Unipolar electrodes are more prone to environmental sensing than bipolar leads
Suxamethonium-induced depolarisation fasciculations
Crosstalk (atrial output sensed by ventricular lead)
Why might tachyarrhythmia arise?
Intrinsic depolarisation occurring during pacemaker refractory period, this not being sensed,
and pacemaker firing soon after in the vulnerable period to initiate a re-entrant tachycardia
Maintenance of the tachyarrhythmia does then not involve the pacemaker at all
Emergency therapy can involve re-programming, or more commonly, magnet conversion to
asynchronous, fixed-rate mode
29. Output failure
Paced stimulus not generated when expected
Results in decreased or absent pacemaker activity on ECG
Causes include oversensing, wire fracture, lead displacement
Steps:
Make sure pacemaker box is on and connected
Increase pacemaker current (up to 200mA for transcutaneous or 20mA for transvenous)
Asynchronous DOO/VOO mode selected to avoid oversensing
Convert to transcutaneous pacing while new pacing system inserted
30. Failure to capture
Paced stimulation does not cause myocardial depolarisation
Electrode displacement, wire fracture, electrolyte disturbance, MI at lead tip or exit
block
Seen with amiodarone, flecainide, hyperkalaemia, acidosis, alkalosis, cardiac
perforation and improper settings also
If patients native heart rate is higher than the paced rate, no pacemaker activity is
expected and output failure/capture failure cannot be recognised on ECG
Suxamethonium causes depolarisation fasciculations that can lead to oversensing
and subsequent failure to capture
31. Failure to capture
Rx
Place in left lateral position (maximises electrode contact with endocardium)
Maximise pacemaker output
Use asynchronous mode
Transcutaneous if failing transvenous (as can directly pace LV)
Isoprenaline/adrenaline infusion to increase intrinsic HR
32. Pacemaker-associated dysrhythmias
Pacemaker-mediated tachycardia (PMT)
Re-entry tachycardia with antegrade pathway via pacemaker and retrograde via AV node
Caused by retrograde P waves being sensed as native atrial activity by dual chamber
pacemaker
Get paced tachycardia at rate determined by pacemaker
Can be terminated by AV block e.g. adenosine OR magnet application
Sensor-induced tachycardia
Modern sensors respond to exercise, tachypnoea.etc.
Sensors may misfire in the presence of vibrations, loud noise, fever, limb movement or
electrocautery with subsequent inappropriately fast rate
Ventricular rate cannot exceed programmed max of 160-180
Terminated by magnet application
33. Pacemaker-associated dysrhythmias
Runaway pacemaker
Low battery voltage in older units results in pacing spikes at 2000bpm with risk of
subsequent VF
Paradoxically, may be failure to capture due to low amplitude pacing spikes and
subsequent bradycardia
Application of magnet may be lifesaving
Lead displacement arrhythmia
If dislodged, lead may float in RV causing tickling of myocardium causing ventricular
ectopics, possible VT and failure to capture
CXR helps confirm diagnosis
If changes from typical LBBB pattern (indicating RV placement) to RBBB pattern, suggests
erosion through interventricular septum
34. Pacemaker-associated dysrhythmias
Twiddler’s syndrome
Pulse generator rotation in pocket with dislodgement of leads and subsequent
diaphragmatic or brachial plexus pacing
RV perforation
RBBB instead of LBBB
Pacing of diaphragm
Haemopericardium rarely
35. ICD
Reduce mortality from 30-45% to <2% per year in those at risk of sudden cardiac
death
Mostly follow a tiered approach to ventricular arrhythmias
Antitachycardia pacing
Low-energy cardioversion
Defibrillation
Projected lifespan of 6-9 years
Most common cause of death is CCF and this should be managed as usual
36. ED evaluation of ICD
Causes of inappropriate shock delivery
False sensing
SVT with RVR
Muscular activity (shivering, diaphragm contraction)
Extraneous source (tapping of chest wall, vibrations)
Sensing T waves as QRS (double counting)
Sensing lead fracture or migration
Unsustained tachyarrhythmia
ICD-pacemaker interactions
Component failure
37. ED evaluation
Ask about symptoms around event, number of shocks, activity at the time and any recent anti-arrhythmic
drug changes
Look for signs of trauma
12-lead
Any shock-related ST changes should resolve within 15 minutes and if not suggests new ischaemia
CXR
Electrode migration, displacement or fracture
Anti-arrhythmic drug levels and serum electrolytes
If patient is receiving repeated inappropriate shocks, temporarily deactive with a magnet over the device
(can simply remove magnet if want to shock again)
All ICD’s should then be evaluated by a cardiologist if been magnetised
If in cardiac arrest:
Follow normal protocols
Place pads at least 8cm from generator
38. Disposition
Need to discuss with treating cardiologist
Admission generally warranted if:
Cardiovascular instability
2 or more shocks in a 1-week period
Correctable causes of dysrhythmia
Any sign of infection or mechanical disruption of the system