Your SlideShare is downloading. ×
0
Dr. Ananya
Contents   Classification   History   Introduction   Indications   Key terms- compliance , ventilatory work   Compon...
Classification According to Robert chatburn Broadly classified into Negative pressure ventilators            And        ...
Negative pressure ventilators Exert a negative pressure on the external chest Decreasing the intrathoracic pressure duri...
 The iron lung, often referred  to in the early days as the  "Drinker respirator", was  invented by Phillip  Drinker(1894...
Biphasic cuirass ventilation Biphasic cuirass ventilation (BCV) is a method  of ventilation which requires the patient to...
Disadvantages Complex and Cumbersome Difficult for transporting Difficult to access the patient in emergency claustrop...
Positive pressure ventilators Inflate the lungs by exerting positive pressure on  the airway, similar to a bellows mechan...
History Andreas Vesalius (1555) Vesalius is credited with the first description of positive-  pressure ventilation, but ...
INTRODUCTION TO MECHANICALVENTILATION: CONVENTIONAL MECHANICAL VENTILATION Mechanical ventilation is a useful modality f...
INDICATIONS FOR MECHANICALVENTILATION Respiratory Failure Cardiac Insufficiency Neurologic dysfunction Rule 1. The ind...
Key terms Ventilatory work- During inspiration , the size of the thoracic cage  increases overcoming the elastic forces ...
 This ventilatory work is proportional to the pressure required for inspiration times the tidal volume. LOAD- The press...
Equation of motion for respiratorysystem Muscle pressure + ventilator pressure =                 (volume / compliance)+ (...
Lung compliance Lung compliance: Is the change in volume per unit change in pressure                    COMPLIANCE =     ...
Types Static compliance- is measured when there is no air flow. Reflects the elastic properties of the lung and the ches...
What is a mechanical ventilator? A machine or a device that fully or partially substitute for the ventilatory work accomp...
INPUT POWER It can be Pneumatically powered(uses compressed gases) Electrically powered(uses 120 Volts AC/12Volts DC)He...
Source of Gas Supply Air - Central compressed air, compressor, turbine flow  generator, etc Oxygen – Central oxygen sour...
DRIVE MECHANISM It’s the system used by the ventilator to transmit or  convert the input power to useful ventilatory work...
Piston mechanismBellows mechanism                    Pneumatic mechanism
 Pneumatic circuits- uses pressurized gas as power  source. these are microprocessor controlled with solenoid  valves. ...
Control circuit Its the system that governs the ventilator drive  mechanism or output control valve. Classified as- Ope...
Control parameters Pressure Volume Flow Time
 Ventilators deliver gas to the lungs using positive  pressure at a certain rate. The amount of gas  delivered can be lim...
 Mechanical- employs levers or pulleys to control drive  mechanism. Pneumatic Fluidic- applies gas flows and pressure t...
Pressure controller Ventilator controls the trans-respiratory system  pressure . This trans-respiratory system gradient ...
Volume controller Volume cycled ventilation delivers a: set volume; with a variable Pressure - determined by resistance...
Flow controller Allows pressure to vary with changes in patient s  compliance and resistance while controlling flow. Thi...
Normal inspiratory time of a spontaneously breathing healthy adult is approximately 0.8- 1.2 seconds, with an inspiratory ...
Phase variablesA. Trigger …….   What causes the breath to begin?                                                     B   ...
 Phases of ventilator supported breath                    inspiration                    change from inspiration to expir...
Trigger variable It’s the variable that determines start of inspiration Triggering refers to the mechanism through which...
Time triggering
Pressure Triggering Breath is delivered when ventilator senses patients spontaneous    inspiratory effort.   sensitivity...
Flow Triggering The flow triggered system has two preset variables for    triggering, the base flow and flow sensitivity....
•Flow triggerAdvantages --The time taken for the onset of inspiratory effort to the onset ofinspiratory flow is considerab...
Limit variable
Cycle variable Defined as the length of one complete breathing cycle. Inspiration ends when a specific cycle variable is...
Baseline variable The variable controlled during expiration phase. Mostly its pressure
Basic definitions Airway Pressures   Peak Inspiratory Pressure (PIP)   Plateau pressures   Positive End Expiratory Pre...
Pressures Mechanical ventilation delivers flow and volume to the    patient’s as a result of the development of a positiv...
 Peak Inspiratory Pressure (PIP)- The peak pressure is the maximum pressure obtainable  during active gas delivery. This ...
 As the plateau pressure is the pressure when there is    no flow within the circuit and patient airways it most    close...
 It is generally believed that end inspiratory occlusion  pressure (ie plateau pressure) is the best clinically  applicab...
 Mean Airway Pressure- The mean airway pressure is an average of the system  pressure over the entire ventilatory period....
PEEP Positive end expiratory pressure (PEEP) refers to the  application of a fixed amount of positive pressure  applied d...
Advantages ability to increase functional residual capacity (FRC)  and keep FRC above Closing Capacity. The increase in ...
 Airway Pressures (Paw) For gas to flow to occur there must be a positive  pressure gradient. In spontaneous respiration...
Physiology of PEEP Reinflates collapsed alveoli and maintains alveolar  inflation during exhalation                      ...
Physiological Responses to CPAP /PEEP
Dangers of PEEP High intrathoracic pressures can cause decreased    venous return and decreased cardiac output   May pro...
AutoPEEP During expiration alveolar pressure is greater than circuit    pressure until expiratory flow ceases. If expirat...
 effects of autoPEEP can predispose the patient to: an increased risk of barotrauma; fall in cardiac output; hypotensi...
I:E ratioThis defines the inspiration to expiration ratio.I:E ratios are normally set as 1:2 as expiration requires a lo...
Tidal Volume Tidal volume refers to the size of the breath that is  delivered to the patient. Normal physiologic tidal v...
 The inspired and expired tidal volumes should generally  correlate. Expired tidal volumes may be less than inspired tid...
Flow (V) Flow rate refers to the speed at which a volume of gas  is delivered, or exhaled, per unit of time. Flow is  des...
Flow
primary factors to consider whenapplying mechanical ventilation the components of each individual breath, specifically   ...
Time (Ti) Time in mechanical ventilation is divided between  inspiratory and expiratory time. Inspiratory Time In most ...
 The following example illustrate how these parameters effect    inspiratory time.   Ventilator settings   · Tidal volu...
Output waveforms Graphical representation of the control or phase  variables in relation to time. presented as          ...
Advantages • Allows user to interpret, evaluate, and troubleshoot     the ventilator and the patient’s response to     ven...
Flow Waveforms inspiratory flow is controlled by setting the peak flow  and flow waveform. The peak flow rate is the max...
 Square waveform- The square flow waveform delivers a set flow rate  throughout ventilator inspiration. If for example t...
 Accelerating waveform- The accelerating flow waveform initially delivers a  fraction of the peak inspiratory flow and s...
Setting the Peak Flow and FlowWaveform The flow rate should be set to match the patient’s  inspiratory demand. Where the ...
 The decelerating flow waveform is the most frequently  selected flow waveform as it produces the lowest peak  inspirator...
Pressure waveforms    Rectangular    Exponential rise    Sine• Can be used to monitor-• Air trapping (auto-PEEP)• Airwa...
References Guide to mechanical ventilation- chang s Breathing and mechanical support- wolfgang oczenski Internet refere...
Thank you
Advantages of Volume Cycled Ventilation Ease of Use Set Volumes: One of the major advantages of volume cycled  ventilati...
Disadvantages The major disadvantages of volume cycled ventilation are the variable pressure and set flow rate. It is the...
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Principles of icu ventilators
Upcoming SlideShare
Loading in...5
×

Principles of icu ventilators

14,966

Published on

Published in: Career, Business, Technology
1 Comment
17 Likes
Statistics
Notes
  • nicely made and good knowledge
       Reply 
    Are you sure you want to  Yes  No
    Your message goes here
No Downloads
Views
Total Views
14,966
On Slideshare
0
From Embeds
0
Number of Embeds
2
Actions
Shares
0
Downloads
853
Comments
1
Likes
17
Embeds 0
No embeds

No notes for slide

Transcript of "Principles of icu ventilators"

  1. 1. Dr. Ananya
  2. 2. Contents Classification History Introduction Indications Key terms- compliance , ventilatory work Components Control mechanism Variables Triggering Factors to consider in mechanical ventilation Wave-forms
  3. 3. Classification According to Robert chatburn Broadly classified into Negative pressure ventilators And according to the manner in which Positive pressure ventilators they support ventilation
  4. 4. Negative pressure ventilators Exert a negative pressure on the external chest Decreasing the intrathoracic pressure during inspiration allows air to flow into the lung, filling its volume Physiologically, this type of assissted ventilation is similar to spontaneous ventilation It is used mainly in chronic respiratory failure associated with neuromascular conditions such as poliomyleitis, muscular dystrophy, a myotrophic lateral sclerosis, and mysthenia gravis.
  5. 5.  The iron lung, often referred to in the early days as the "Drinker respirator", was invented by Phillip Drinker(1894 – 1972) and Louis Agassiz Shaw Junior, professors of industrial hygiene at the Harvard School of Public Health . The machine was powered by an electric motor with air pumps from two vacuum cleaners. The air pumps changed the pressure inside a rectangular, airtight metal box, pulling air in and out of the lungs
  6. 6. Biphasic cuirass ventilation Biphasic cuirass ventilation (BCV) is a method of ventilation which requires the patient to wear an upper body shell or cuirass, so named after the body armour worn by medieval soldiers. The ventilation is biphasic because the cuirass is attached to a pump which actively controls both the inspiratory and expiratory phases of the respiratory cycle .
  7. 7. Disadvantages Complex and Cumbersome Difficult for transporting Difficult to access the patient in emergency claustrophobic
  8. 8. Positive pressure ventilators Inflate the lungs by exerting positive pressure on the airway, similar to a bellows mechanism, forcing the alveoli to expand during inspiration Expiration occurs passively. modern ventilators are mainly PPV s and are classified based on related features, principles and engineering.
  9. 9. History Andreas Vesalius (1555) Vesalius is credited with the first description of positive- pressure ventilation, but it took 400 years to apply his concept to patient care. The occasion was the polio epidemic of 1955, when the demand for assisted ventilation outgrew the supply of negative-pressure tank ventilators (known as iron lungs). In Sweden, all medical schools shut down and medical students worked in 8-hour shifts as human ventilators, manually inflating the lungs of afflicted patients. Invasive ventilation first used at Massachusetts General Hospital in 1955. Thus began the era of positive-pressure mechanical ventilation (and the era of intensive care medicine).
  10. 10. INTRODUCTION TO MECHANICALVENTILATION: CONVENTIONAL MECHANICAL VENTILATION Mechanical ventilation is a useful modality for patients who are unable to sustain the level of ventilation necessary to maintain the gas exchange functions- oxygenation and carbon dioxide elimination The first positive-pressure ventilators were designed to inflate the lungs until a preset pressure was reached. In contrast, volume-cycled ventilation, which inflates the lungs to a predetermined volume, delivers a constant alveolar volume despite changes in the mechanical properties of the lungs.
  11. 11. INDICATIONS FOR MECHANICALVENTILATION Respiratory Failure Cardiac Insufficiency Neurologic dysfunction Rule 1. The indication for intubation and mechanical ventilation is thinking of it. Rule 2. Endotracheal tubes are not a disease, and ventilators are not an addiction
  12. 12. Key terms Ventilatory work- During inspiration , the size of the thoracic cage increases overcoming the elastic forces of the lungs and the thorax and resistance of the airways. As the volume of the thoracic cage increases, intrapleural pressure becomes more negative, resulting in lung expansion. Gas flows from the atmosphere into the lungs as a result of transairway pressure gradient. During expiration, the elastic forces of the lung and thorax cause the chest to decrease in volume and exhalation occurs as a result of greater pressure at the alveolus compared to atm. Press.
  13. 13.  This ventilatory work is proportional to the pressure required for inspiration times the tidal volume. LOAD- The pressure required to deliver the tidal volume is referred to as the load that the muscles or ventilator must work against. load elastic ( α volume & inv. Prop t0 compliance) resistance (α Raw & inspiratory flow)
  14. 14. Equation of motion for respiratorysystem Muscle pressure + ventilator pressure = (volume / compliance)+ (resistance x flow) Flow- it’s the unit of volume by unit of time. Resistance- it is the force that must be overcome to move the gas through the conducting airways. It is described by the poiseulle’s law.
  15. 15. Lung compliance Lung compliance: Is the change in volume per unit change in pressure COMPLIANCE =  Volume /  Pressure
  16. 16. Types Static compliance- is measured when there is no air flow. Reflects the elastic properties of the lung and the chest wall Dynamic compliance is measured when air flow is present Reflects the airway resistance (non elastic resistance) and elastic properties of lung and chest wall Static compliance=Corrected tidal volume Plateau pressure-PEEP Dynamic compliance corrected tidal volume Peak inspiratory pressure-PEEP
  17. 17. What is a mechanical ventilator? A machine or a device that fully or partially substitute for the ventilatory work accomplished by the patients muscles. Components – INPUT POWER DRIVE MECHANISM CONTROL CIRCUIT OUTPUT WAVEFORMS ALARMS
  18. 18. INPUT POWER It can be Pneumatically powered(uses compressed gases) Electrically powered(uses 120 Volts AC/12Volts DC)Here the electric motor drives pistons and compressors to generate gas flows . Microprocessor controlled- combined.Also called as 3rd generation ventilators.
  19. 19. Source of Gas Supply Air - Central compressed air, compressor, turbine flow generator, etc Oxygen – Central oxygen source, O2 concentrator, O2 cylinder Gas mixing unit – O2 blender
  20. 20. DRIVE MECHANISM It’s the system used by the ventilator to transmit or convert the input power to useful ventilatory work. This determines the characteristic flow and pressure patterns produced by the ventilator. It includes pistons bellows reducing valves pneumatic circuits
  21. 21. Piston mechanismBellows mechanism Pneumatic mechanism
  22. 22.  Pneumatic circuits- uses pressurized gas as power source. these are microprocessor controlled with solenoid valves. use programmed algorithms in microprocessor to open and close solenoid valves to mimic any flow or pressure wave pattern.
  23. 23. Control circuit Its the system that governs the ventilator drive mechanism or output control valve. Classified as- Open circuits- desired output is selected and venti. achieves it without any further input from clinician. Closed circuits- desired output is selected and venti. Measures a specific parameter (flow/vol/press) continuously and input is constantly adjusted to match desired output.a.k.a SERVO controlled.
  24. 24. Control parameters Pressure Volume Flow Time
  25. 25.  Ventilators deliver gas to the lungs using positive pressure at a certain rate. The amount of gas delivered can be limited by time, pressure or volume. The duration can be cycled by time, pressure or flow.If volume is set, pressure varies…..if pressure is set, volume varies….. ….according to the compliance…...
  26. 26.  Mechanical- employs levers or pulleys to control drive mechanism. Pneumatic Fluidic- applies gas flows and pressure to control direction of other gas flows and to perform logic functions based on the COANDA effect. Electronic- uses resistors and diodes and integrated circuits to provide control over the drive mechanism.
  27. 27. Pressure controller Ventilator controls the trans-respiratory system pressure . This trans-respiratory system gradient determines the depth or volume of respiration. Based on this a ventilator can be positive or negative pressure ventilator.
  28. 28. Volume controller Volume cycled ventilation delivers a: set volume; with a variable Pressure - determined by resistance, compliance and inspiratory effort
  29. 29. Flow controller Allows pressure to vary with changes in patient s compliance and resistance while controlling flow. This flow is measured by vortex sensors or venturi pnemotachometers.Time controllermeasures and controls inspiratory and expiratory time.These ventilators are used in newborns and infants Inspiratory time is a combination of the inspiratory flow period and time taken for inspiratory pause. The following diagram depicts how the addition of an inspiratory pause extends total inspiratory time.
  30. 30. Normal inspiratory time of a spontaneously breathing healthy adult is approximately 0.8- 1.2 seconds, with an inspiratory expiratory (I: E) ratio of 1:1.5 to 1:2 2.Its advantageous to extend the inspiratory time in order to:• improve oxygenation - through the addition of an inspiratory pause; or to•increase tidal volume - in pressure controlled ventilationAdverse effects of excessively long inspiratory times are haemodynamic compromise,patient ventilator dysynchrony, and the development of autoPEEP.
  31. 31. Phase variablesA. Trigger …….  What causes the breath to begin? B CB. Limit ……  What regulates gas flow during the breath? AC. Cycle …….  What causes the breath to end?
  32. 32.  Phases of ventilator supported breath inspiration change from inspiration to expiration expiration change from expiration to inspirationTypes of ventilator breaths- Mandatory breath Assisted breath Spontaneous breath
  33. 33. Trigger variable It’s the variable that determines start of inspiration Triggering refers to the mechanism through which the ventilator senses inspiratory effort and delivers gas flow or a machine breath in concert with the patient’s inspiratory effort. Can use pressure or volume or time or flow as a trigger. In modern ventilators the demand valve is triggered by either a fall in pressure (pressure triggered) or a change in flow (flow triggered). With pressure triggered a preset pressure sensitivity has to be achieved before the ventilator delivers fresh gas into the inspiratory circuit. With flow triggered a preset flow sensitivity is employed as the trigger mechanism.
  34. 34. Time triggering
  35. 35. Pressure Triggering Breath is delivered when ventilator senses patients spontaneous inspiratory effort. sensitivity refers to the amount of negative pressure the patient must generate to receive a breath/gas flow. If the sensitivity is set at 1 cm then the patient must generate 1 cm H2O of negative pressure for the machine to sense the patients effort and deliver a breath. Acceptable range - -1 to -5 cm H2O below patient s baseline pressure If the sensitivity is too high the patients work of breathing will be unnecessarily increased. It is not a reasonable course of action to increase the sensitivity to reduce the patients respiratory rate as it only increases their work of breathing.
  36. 36. Flow Triggering The flow triggered system has two preset variables for triggering, the base flow and flow sensitivity. The base flow consists of fresh gas that flows continuously through the circuit. The patient’s earliest demand for flow is satisfied by the base flow. The flow sensitivity is computed as the difference between the base flow and the exhaled flow Here delivered flow= base flow- returned flow Hence the flow sensitivity is the magnitude of the flow diverted from the exhalation circuit into the patient’s lungs. As the subject inhales and the set flow sensitivity is reached the flow pressure control algorithm is activated, the proportional valve opens, and fresh gas is delivered.
  37. 37. •Flow triggerAdvantages --The time taken for the onset of inspiratory effort to the onset ofinspiratory flow is considerably less. -decreases the work involved in initiating a breath.
  38. 38. Limit variable
  39. 39. Cycle variable Defined as the length of one complete breathing cycle. Inspiration ends when a specific cycle variable is reached. This variable is used as a feedback signal to end inspiratory flow delivery which then allows exhalation to start. Most new ventilators measure flow and use it as a feedback signal. So volume becomes a function of flow and time Volume= flow x inspiratory time
  40. 40. Baseline variable The variable controlled during expiration phase. Mostly its pressure
  41. 41. Basic definitions Airway Pressures  Peak Inspiratory Pressure (PIP)  Plateau pressures  Positive End Expiratory Pressure (PEEP)  Continuous Positive Airway Pressure (CPAP) Inspiratory Time or I:E ratio Tidal Volume: amount of gas delivered with each breath
  42. 42. Pressures Mechanical ventilation delivers flow and volume to the patient’s as a result of the development of a positive pressure gradient between the ventilator circuit and the patient’s gas exchange units as illustrated in the diagram above. There are four pressures to be aware of in regards to mechanical ventilation. These are the: Peak Plateau Mean; and End expiratory pressures.
  43. 43.  Peak Inspiratory Pressure (PIP)- The peak pressure is the maximum pressure obtainable during active gas delivery. This pressure a function of the compliance of the lung and thorax and the airway resistance including the contribution made by the tracheal tube and the ventilator circuit. Maintained at <45cm H2O to minimize barotrauma Plateau Pressure-The plateau pressure is defined as the end inspiratory pressure during a period of no gas flow. The plateau pressure reflects lung and chest wall compliance.
  44. 44.  As the plateau pressure is the pressure when there is no flow within the circuit and patient airways it most closely represents the alveolar pressure and thus is of considerable significance as it desirable to limit the pressure that the alveoli are subjected to. Excessive pressure may result in extrapulmonary air (eg pneumothorax) and acute lung injury. An increase in airways resistance (including ETT resistance) will result in an increase in PIP. An increase in resistance will result in a widening of the difference between PIP and plateau pressure. A fall in compliance will elevate both PIP and plateau pressure.
  45. 45.  It is generally believed that end inspiratory occlusion pressure (ie plateau pressure) is the best clinically applicable estimate of average peak alveolar pressure. Although controversial it has been generally recommended that the plateau pressure should be limited to 35 cms H2O.
  46. 46.  Mean Airway Pressure- The mean airway pressure is an average of the system pressure over the entire ventilatory period. End Expiratory Pressure- End expiratory pressure is the airway pressure at the termination of the expiratory phase and is normally equal to atmospheric or the applied PEEP level.
  47. 47. PEEP Positive end expiratory pressure (PEEP) refers to the application of a fixed amount of positive pressure applied during mechanical ventilation cycle Continuous positive airway pressure (CPAP) refers to the addition of a fixed amount of positive airway pressure to spontaneous respirations, in the presence or absence of an endotracheal tube. PEEP and CPAP are not separate modes of ventilation as they do not provide ventilation. Rather they are used together with other modes of ventilation or during spontaneous breathing to improve oxygenation, recruit alveoli, and / or decrease the work of breathing
  48. 48. Advantages ability to increase functional residual capacity (FRC) and keep FRC above Closing Capacity. The increase in FRC is accomplished by increasing alveolar volume and through the recruitment of alveoli that would not otherwise contribute to gas exchange. Thus increasing oxygenation and lung compliance The potential ability of PEEP and CPAP to open closed lung units increases lung compliance and tends to make regional impedances to ventilation more homogenous.
  49. 49.  Airway Pressures (Paw) For gas to flow to occur there must be a positive pressure gradient. In spontaneous respiration gas flow occurs due to the generation of a negative pressure in the alveoli relative to atmospheric or circuit pressure (as in CPAP) (refer to following diagram).
  50. 50. Physiology of PEEP Reinflates collapsed alveoli and maintains alveolar inflation during exhalation PEEP Decreases alveolar distending pressure Increases FRC by alveolar recruitment Improves ventilation Increases V/Q, improves oxygenation, decreases work of breathing
  51. 51. Physiological Responses to CPAP /PEEP
  52. 52. Dangers of PEEP High intrathoracic pressures can cause decreased venous return and decreased cardiac output May produce pulmonary barotrauma May worsen air-trapping in obstructive pulmonary disease Increases intracranial pressure Alterations of renal functions and water metabolism
  53. 53. AutoPEEP During expiration alveolar pressure is greater than circuit pressure until expiratory flow ceases. If expiratory flow does not cease prior to the initiation of the next breath gas trapping may occur. Gas trapping increases the pressure in the alveoli at the end of expiration and has been termed: dynamic hyperinflation; autoPEEP; inadvertent PEEP; intrinsic PEEP; and occult PEEP
  54. 54.  effects of autoPEEP can predispose the patient to: an increased risk of barotrauma; fall in cardiac output; hypotension; fluid retention; and an increased work of breathing
  55. 55. I:E ratioThis defines the inspiration to expiration ratio.I:E ratios are normally set as 1:2 as expiration requires a longer time .In severe obstructive disease such as status asthamaticus it can be set as 1:4Factors affecting I:E Ratio-1. Tidal volume2. Respiratory rate3. Flow rate • Increasing inspiration time will increase TV, but may lead to auto-PEEP
  56. 56. Tidal Volume Tidal volume refers to the size of the breath that is delivered to the patient. Normal physiologic tidal volumes are approximately 5-7 ml / kg whereas the traditional aim for tidal volumes has been approximately 10 - 15 ml / kg. The rationale for increasing the size of the tidal volume in ventilated patients has been to prevent atelectasis and overcome the deadspace of the ventilator circuitry and endotracheal tube. Inspired and expired tidal volumes are plotted on the y axis against time as depicted in the following diagram.
  57. 57.  The inspired and expired tidal volumes should generally correlate. Expired tidal volumes may be less than inspired tidal volumes if: there is a leak in the ventilator circuit - causing some of the gas delivered to the patient to leak into the atmosphere there is a leak around the endotracheal / tracheostomy tube - due to tube position, inadequate seal or cuff leak there is a leak from the patient, such as a bronchopleural fistula Expired tidal volumes may be larger than inspired tidal volumes due to: the addition of water vapour in the ventilator circuitry from a hot water bath humidifier.
  58. 58. Flow (V) Flow rate refers to the speed at which a volume of gas is delivered, or exhaled, per unit of time. Flow is described in litres per minute . The peak (inspiratory) flow rate is therefore the maximum flow delivered to a patient per ventilator breath. Flow is plotted on the y axis of the ventilator graphics against time on the x axis . In the following diagram that inspiratory flow is plotted above the zero flow line, whereas expiratory flow is plotted as a negative deflection. When the graph depicting flow is at zero there is no gas flow going into or out of the patient.
  59. 59. Flow
  60. 60. primary factors to consider whenapplying mechanical ventilation the components of each individual breath, specifically whether pressure, flow, volume and time are set by the operator, variable or dependent on other parameters the method of triggering the mechanical ventilator breath/gas flow, how the ventilator breath is terminated: potential complications of mechanical ventilation. methods to improve patient ventilator synchrony; and the nursing observations required to provide a safe and effective level of care for the patient receiving mechanical ventilation
  61. 61. Time (Ti) Time in mechanical ventilation is divided between inspiratory and expiratory time. Inspiratory Time In most volume cycled ventilators used in the intensive care environment it is not possible to set the inspiratory time. The inspiratory time is determined by the peak inspiratory flow rate, flow waveform and inspiratory pause. Where inspiratory time is able to be set, flow becomes dependent on inspiratory time and tidal volume.
  62. 62.  The following example illustrate how these parameters effect inspiratory time. Ventilator settings · Tidal volume 1000mls · Peak Flow 60 lpm · Flow Waveform square / constant · Insp. Pause 0 secs The inspiratory time for this patient would be 1 second because gas is constantly being delivered at a flow rate of 60 lpm, which equals 1 litre per second. If an inspiratory pause of 0.5 seconds were applied then the inspiratory time would be increased to 1.5 seconds. Changing the patients flow waveform from a square to a decelerating flow waveform, without changing the flow rate, will result in an increase in inspiratory time, because the flow of gas is only initially set at 60 lpm and decreases throughout inspiration
  63. 63. Output waveforms Graphical representation of the control or phase variables in relation to time. presented as pressure flow waveforms volume The ventilator determines the shape of control variable whereas the other two depend on the patient compliance and resistance. Conventionally flow above X-axis is inspiration.
  64. 64. Advantages • Allows user to interpret, evaluate, and troubleshoot the ventilator and the patient’s response to ventilator. • Monitors the patient’s disease status (C and Raw). • Assesses patient’s response to therapy. • Monitors ventilator function • Allows fine tuning of ventilator to decrease WOB, optimize ventilation, and maximize patient comfort.
  65. 65. Flow Waveforms inspiratory flow is controlled by setting the peak flow and flow waveform. The peak flow rate is the maximum amount of flow delivered to the patient during inspiration, whereas the flow waveform determines the how quickly gas will be delivered to the patient throughout various stages of the inspiratory cycle. There are four different types of flow waveforms available. These include the square, decelerating (ramp), accelerating sine/sinusoidal waveform
  66. 66.  Square waveform- The square flow waveform delivers a set flow rate throughout ventilator inspiration. If for example the peak flow rate is set at 60 lpm then the patient will receive 60 lpm throughout ventilator inspiration. Decelerating waveform The decelerating flow waveform delivers the peak flow at the start of ventilator inspiration and slowly decreases until a percentage of the peak inspiratory flow rate is attained.
  67. 67.  Accelerating waveform- The accelerating flow waveform initially delivers a fraction of the peak inspiratory flow and steadily increasing the rate of flow until the peak flow has been reached. Sine / sinusoidal waveform- The sine waveform was designed to match the normal flow waveform of a spontaneously breathing patient.
  68. 68. Setting the Peak Flow and FlowWaveform The flow rate should be set to match the patient’s inspiratory demand. Where the patient’s inspiratory flow requirements exceed the preset flow rate there will be an imposed work of breathing which may cause the patient to fight the ventilator and become fatigued. Where flow rate is unable to match the patient’s inspiratory flow requirements the pressure waveform on the ventilator graphics screen may show a depressed or “scooped out” pressure waveform. This is often referred to as flow starvation.
  69. 69.  The decelerating flow waveform is the most frequently selected flow waveform as it produces the lowest peak inspiratory pressures of all the flow waveforms. This is because of the characteristics of alveolar expansion. Initially a high flow rate is required to open the alveoli. Once alveolar opening has occurred a lower flow rate is sufficient to procure alveolar expansion. Flow waveforms which produce a high flow rate at the end of inspiration (ie. square and accelerating flow waveforms) exceed the flow requirements for alveolar expansion, resulting in elevated peak inspiratory pressures
  70. 70. Pressure waveforms  Rectangular  Exponential rise  Sine• Can be used to monitor-• Air trapping (auto-PEEP)• Airway Obstruction• Bronchodilator Response• Respiratory Mechanics (C/Raw)• Active Exhalation• Breath Type (Pressure vs. Volume)• PIP, Pplat• CPAP, PEEP• Asynchrony• Triggering Effort
  71. 71. References Guide to mechanical ventilation- chang s Breathing and mechanical support- wolfgang oczenski Internet references
  72. 72. Thank you
  73. 73. Advantages of Volume Cycled Ventilation Ease of Use Set Volumes: One of the major advantages of volume cycled ventilation is the ability to set a tidal volume. This is of critical importance to patient’s requiring tight regulation of carbon dioxide elimination. Neurosurgical patients - post surgery / head injury and patients suffering a neurological insult (eg post cardiac arrest) often require CO2 regulation. This is because carbon dioxide is a potent vasodilator. Increased levels of carbon dioxide, in these groups of patients, may therefore increase cerebral blood volume with a concomitant elevation of intracranial pressure. A raised intracranial pressure may decrease the delivery of oxygenated blood to the brain - resulting in cerebral ischaemia. Conversely a low CO2 may cause constriction of the cerebral vasculature also resulting in decreased oxygen delivery and cerebral ischaemia. For these reasons volume cycled ventilation is often the mode of choice for patients requiring CO2 regulation.
  74. 74. Disadvantages The major disadvantages of volume cycled ventilation are the variable pressure and set flow rate. It is therefore a necessary part of nursing practice to closely monitor the patients inspiratory pressure and observe the patient for signs of “flow starvation”.
  1. A particular slide catching your eye?

    Clipping is a handy way to collect important slides you want to go back to later.

×