2. CCNPWT provides a Closed Sterile &
-
-
-
Moist Wound healing environment
Reduces bacterial burden
Reduces oedema
Enhances dermal perfusion
Promotes micro angiogenesis
Stimulates cell proliferation
Promotes granulation tissue formation.
Studies suggest mechanical stretching
may result in increased mitosis (Cell
Replication)
Controlled
Negative
Pressure Therapy decreases wound
volume approximating wound edges,
helps uniformly draw wounds closed
Promotes
continuous
wound
debridement by removing wound micro
debris & bacteria colonized wound
output / interstitial fluid allowing tissue
decompression
Protects wound from re-infection, cross
infection or spreading further infection
Promotes Skin Graft/Flap uptake (The
sterile flexible foam dressing gently
presses the skin graft or flap ensuring full
contact with the wound bed &
simultaneously removes wound output
ensuring moist healing environment)
3. SOFTWARE SETTINGS
Switch on the CCNPWT System, You will see
the home screen showing
SELF TEST MESSAGE (SEE: snap shot
image 1).
After self test system will try achieving the
target pressure set already or during previous
therapy session. Pump will stop once the
actual pressure equals target pressure, when
it will fall again pump will restart, process will
continue like this (SEE: snap shot image 2).
To Change pressure settings click the menu
button, select the setting as in (SEE: snap shot
image 3) by pressing UP / DOWN switch, the
< in the left will move up/Down
To select the desired setting press ENTER
Switch the arrow will turn >, (SEE: snap shot
image 4) this shows the desired setting has
been selected, by pressing UP / DOWN switch
you can change the pressure or other settings.
After doing the changes again press ENTER
switch to save and then the MENU switch to
exit and restart therapy.
4. DISPLAY FRONT PANEL
(VTG1901)
LCD DISPLAY
(4 Lines)
Power Indicator
LED (GREEN)
Battery Indicator
RED LED
UP Cursor Switch
DOWN Cursor Switch
ENTER Switch
MENU Switch
6. REAR VIEW
BATTERY ON / OFF
SWITCH
PRESS 0 for OFF
PRESS I for ON
CANISTER HOLDER
SEE IMAGE in NEXT PAGE
CUSTOMER CARE NO.
MACHINE SERIAL NO. LABEL
7. ACCESSORIES
INSTALLATION STEP
Open Canister Pack
& hook it in the
canister holder
(As shown in the above
picture)
Open Filter Pack
Connect the female
connector
of filter with the male
vacuum outlet in the
system
Connect the distal end
of the filter silicon tube
with the L Connector of
the canister as shown
in the picture
(push & glide the L
Connector inside the
silicon tube)
8. AFTER ALL SETUP
Fig. 1
Fig. 2
CROSS CHECK AFTER
COMPLETE INSTALLATION
Fig. 3
CONNECT THE MALE CONNECTOR FROM CANISTER & FEMALE CONNECTOR FROM
DRESSING/DRAINAGE TUBE TO COMPLETE THE VACUUM CIRCUIT (as shown in fig 2)
10. TROUBLE SHOOTING
PRESSURE NOT ACHIEVING TO TARGET PRESSURE
Unplug the filter tube and clamp & see Actual Pressure =
Target Pressure
if OK
If NOT Check the VACUUM PORT Connector or it could be a
machine problem
Proceed to Next Step
Unplug the other end of the filter and clamp & see if Actual
Pressure = Target Pressure
if OK
If NOT, possibly filter is malfunctioning
Proceed to Next Step
Clamp the Canister tube & see Actual Pressure = Target
Pressure
if OK
If NOT, possibly canister is leaking or faulty.
Proceed to Next Step
Connect the dressing tube with the canister tube & Clamp the
dressing tube & see Actual Pressure = Target Pressure
if OK
If NOT, possibly the canister & dressing tube joint is leaking
or not properly connected.
then CHECK DRESSING SEAL
11. ERROR – NO DISPLAY
Check the MAINS Cable / POWER
Cable is connected to the power
socket & is switched ON.
See the POWER SWITCH in the
CCNPWT System is RED GLOWING
See the BATTERY SWITCH in the
CCNPWT System is ON (PRESSED I)
(ON POWER)GREEN LED
(ON BATTERY) RED LED
IF ONLY RED is ON, that
shows/indicates the system is working
on battery
IF ONLY GREEN is ON, that shows/indicates the system is on MAIN supply
IF BOTH RED & GREEN is ON, that shows/indicates the system is working
on MAINS supply & battery is charging. In absence of MAIN supply power
sytem will automatically switch over to battery standby. (IDEAL CONDITION)
ENSURE the RED & GREEN LED/LIGHT in the front panel or in the left side of
the LCD Display is glowing every time
12. ERROR : BLOCKAGE / CANISTER FULL
Ensure the canister gets replaced between 450ml -500ml
accumulation of fluid
or this may choke the filters or even can damage the system
pump (if continued further or for longer period).
IF canister is not full remove the filter tube & run the system.
IF OK, then connect the filter & CHECK for any blockage in
the filter.
IF OK connect the canister without connecting the dressing
tube to CHECK any tube blockage or tube folds/kink.
IF OK, there could be drainage tube blockage or mismatch
hole with the dressing/drape sealed.
Check with a needle prick in the dressing (Close to Drainage
Tube junction point).
OR CONTACT A CCNPWT REPRESENTATIVE / SUPPORT
CENTRE
13. CCNPWT WOUND DRAINAGE KIT
SMALL (10 x 7.5 x 3 cm)
MEDIUM (18 x 12.5 x 3 cm)
LARGE (26 x 15 x 3 cm)
EXTRA LARGE (60 x 30 x 2 cm)
ABDOMINAL KIT
PVA (15 x 10 x 1 cm)
CAT NO. 9005D
CAT NO. 9004D
CAT NO. 9002D
CAT NO. 9007D
CAT NO. 9009D
CAT NO. 9008D
OTHER CONSUMABLES
DRAINAGE TUBING
CANISTER 500ml (Disposable)
with aqua gel pack
FILTER
TRANSPARENT PU DRAPE
CONTENTS OF DRESSING KIT
DRESSING FOAM
TRANSPARENT PU DRAPE
DRAINAGE TUBING
DISPOSABLE MEASURING SCALE
CAT NO. 9018D
CAT NO. 9010D
CAT NO. 111D
CAT NO. 9017D
14. ALL CCNPWT CONSUMABLES ARE SINGLE USE
TO ENSURE MAXIMUM PROTECTION & STERILITY
STERILITY
: GAMMA RADIATED
SHELF LIFE
: 3 YEARS
CERTIFICATIONS : CE / GMP / ISO 13485:2003
24 x 7 CUSTOMER CARE : +91 9650440444
MANUFACTURED BY or ON BEHALF OF
TRIAGE HEALTHCARE TECHNOLOGIES
REGD. OFF.: 17, Aashirwaad Appts.
Prop. No. 886-C/8, Mehrauli
New Delhi-110030 (INDIA)
WORKS: H-41, Shahjahanpur Indl. Area,
Distt. Alwar, Rajasthan (INDIA)
CUSTOMER CARE: +91 9650440444 / 9971722111
Email : info@npwtindia.com
www.npwtindia.com
USAGE : STRICTLY BY PHYSICIAN OR WOUND CARE NURSE‘s ADVICE ONLY