7. High-quality CPR
Rate at least 100/min
Depth at least 5cm
Release Complete chest recoil
Uninterrupted Minimize interruption
Ventilation 과도한 환기 하지 않기
Berg RA, Hemphill R, Abella BS, et al. Part 5: Adult Basic Life Support: 2010 American Heart Association
Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;12
2(18_suppl_3):S685–S705. doi:10.1161/CIRCULATIONAHA.110.970939.
9. CPR quality & feedback
• CPR quality 에 대한 현재 알고 있는 지식과 이를 실
제 적용하는 현장과의 간극이 크다.
1. CPR performance 에 대한 측정 기준
2. Monitoring & Feedback
3. Team-level logistics
4. continuous QI
• End tidal carbon dioxide
• Feedback devices
13. Oxygenation and Ventilation
Oxygenation (Pulse Ox)
– O2 for metabolism
– SpO2 measures
% of O2 in RBCs
– Reflects changes in
oxygenation within
5 minutes
Ventilation (Capnography)
– CO2 from metabolism
– EtCO2 measures exhaled
CO2 at point of exit
– Reflects changes in
ventilation within
10 seconds
18. CPR feedback devices
• 사용목적
– 교육을 위한 장비
• The Skillmeter Anne (Laerdal, Orpington, UK)
• The voice advisory manikin (VAM) (Laerdal, Orpington, UK)
– 임상에 적용하기 위한 장비
• 작동방식에 따라
– 메트로놈: guide compression rate
– 압력센서: Barometer based CPR feedback device
– 가속도센서: Accelerometer based CPR feedback device
21. CPR rate at a constant 100bpm
; >80dB Audible tone, yellow LED
LED pressure sensor
; 환자의 체중에 따라 적정 압력
CPRezy
22. Pocket CPR
• 실시간 음성 피드백
• Audible metronome
• 불빛 신호를 이용한 시각피드백
• Audible and visual feedback to perform CPR
consistent with the 2010 AHA guideline
– good compression " “good compressions” + four LED
– bad compression " “push harder” + one LED
• Audible metronome : proper rate of compression
39. The Mattress Issue:
• Mattress compression = 35 – 40% of total
compression depth
• Accelerometer feedback devices fail to
account for mattress compression
• Use of a backboard fails to compensate for
mattress compression
Perkins et al. Resuscitation 2009;80:
79-82
40. Take home message
• 흉부압박은 CPR 결과의 주요결정요인이다.
– Most common errors: depth and speed
• CPR 품질의 적절성에 대한 모니터링이 필요
– 생존율과 밀접한 관계
• 현재 사용되는 도구: EtCO2 & AV feedback장비
– CO2 : CPR 품질에 의한 결과를 감지
– 압력센서/가속도센서
• Mattress issue: 바닥재질에 따라 압박깊이 과평가 문제