Sonic IR Imaging on Bell Imact 
Panels (Hat and I-Beam) 
By: Justin Ar-Rasheed
Overview 
• Behavior of Panels Observed 
• Optimal Experimental Settings Tested 
– Frequency 
– Coupling Material 
– Tip Size/Amplitude 
• Additional Areas of Study 
– Mode Pattern Vibration 
– Automation
Legend: 
< 1” Diameter, 9 Each 
1.0” – 1.75” Diameter, 6 Each 
1.75” - 2.65” Diameter, 4 Each 
> 2.65” Diameter, 3 Each 
0.170” Skin 
0.096” Skin 
Impact Panel for Wayne State University 
60” 
9/30/2011 Final Impact Plan - RJB
Hat Panel
Test Experiments 
• Shoot gun on various locations of panel 
• Analyze results and come up with hypothesis
Ultrasound 
20KHz Handheld 
20k LPt 20:0.5:902 
40% Amplitude 
800ms Pulse 
Camera 
30 Frames/s 
IR Camera 8s 
File 
20K_handheld/40%_30fps_8s_01.fts 
Panel mounted vertically
Ultrasound 
40KHz Handheld 
40k Lpe 40:0.5:4T 
100% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
40k_handheld/100%_40fps_6s_02.fts
Ultrasound 
40KHz Table Gun 
40k Lpe 40:0.5:4T 
100% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
40k_tableGun/40fra_6s_800ms_30lbs_coin.fts
Ultrasound 
40KHz Handheld 
40k Lpe 40:0.5:4T 
100% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
40K_handheld/100%_40fps_6s_01_top.fts
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
Files 
20PartitionShot/40%_Top33%_1.2s_40fps_6s 
20PartitionShot/50%_Top_left_33%_800ms_40fps_6s_01 
20PartitionShot/50%_Top_right_33%_800ms_40fps_6s01 
Left 
Middle 
Right 
Left Middle Right
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
Files 
20PartitionShot/40%_Mid33%_1.2s_40fps_6s 
20PartitionShot/50%_Mid_left_33%_800ms_40fps_6s_02 
20PartitionShot/50%_Mid_right_33%_800ms_40fps_6s_02 
Left 
Middle 
Right 
Left Middle Right
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
Files 
20PartitionShot/50%_bottom_left_33%_800ms_40fps_6 
s_01 
20PartitionShot/50%_bottom_right_33%_800ms_40fps 
_6s_01 
20PartitionShot/50%_Mid_left_33%_800ms_40fps_6s_0 
2 
Left 
Middle 
Right Left Middle Right
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
20PartitionShot
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
Bellimpactpanel03/Bottomhalfve 
iw40fps_50%_800ms_6s_02 
Panel mounted horizontally
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
Bellimpactpanel03/Fullveiw40fps_50%_8 
00ms_6s_03
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
Bellimpactpanel03/Tophalfveiw40fps_50 
%_800ms_6s_01
Experiment Setup: 
20kHz Handheld (500W 
from backside) 
40 frames per second 
Total 6 seconds 
50% amplitude
Experimental Setup: 
20kHz Handheld (500W) 
40 frames per second 
Total 6 seconds 
50% amplitude
Experimental Setup: 
20kHz Handheld(500W) 
40 frames per second 
Total 6 seconds 
50% amplitude 
File Name : 20K_handheld_Amp50_middle_GunHandI_LogoLeftBottom.bmp
Experimental Setup: 
20kHz Handheld (500W) 
40 frames per second 
Total 6 seconds 
60% amplitude 
File Name : 20K_handheld_Amp60_middle_GunGandN_LogoLeftBottom.bmp
Experimental Setup: 
20kHz Handheld (500W) 
40 frames per second 
Total 6 seconds 
60% amplitude 
File Name : 20K_handheld_Top_Amp60_GunOnTop_4inFromEdge.bmp
Experimental Setup: 
20kHz Handheld(500w) 
40 frames per second 
Total 6 seconds 
60% amplitude 
File Name : 20K_handheld_Bottom_Amp60_GunOnBottom_2inFromEdge.bmp
Back View
Structure I - Beam Thin area Doublers Area 
Size L M S L M S L M S 
Front View(20k Table and 
Handheld) 
C,D,P,Q A G,F,N,M J,K,L I 
Back View(20kTable and 
Handheld) 
R,O S 
Never Seen 
(Both 20k &40k) 
B, E H
Frequency Experiments
20k_Amp40%_without coin 
20k_Amp40%_with coin
40k_Amp40%_without coin 
40k_Amp40%_with coin
40k_Amp100%_without coin 
40k_Amp100%_with coin
Preferred Frequency 
20K
Small Tip Experiment 
Coupling Materials and Tips
Small Tip Experiment 
• Coin + Business Card (20 lbs) 
• Coin + Cardboard (20 lbs) 
• Coin + Duct Tape (20 lbs) 
• Coin + Teflon (20 lbs) 
• Business Card 
– (20 lbs) 
– (40 lbs) 
• Cardboard 
– (20 lbs) 
– (40 lbs) 
• Duct Tape (20 lbs) 
• Teflon (20 lbs)
Coin + Business Card (20 lbs) 
60% (4.56 RMS) 80% (6.27 RMS)
Coin + Cardboard (20 lbs) 
60% (0.85 RMS) 80% (1.80 RMS)
Coin + Duct Tape (20 lbs) 
60% (4.22 RMS) 80% (3.46 RMS)
Coin + Teflon (20 lbs) 
60% (4.75 RMS)
Business Card (20 lbs) 
60% (1.35 RMS) 80% (1.27 RMS)
Business Card (40 lbs) 
60% (1.53 RMS) 80% (2.25 RMS)
Cardboard (20 lbs) 
60% (0.64 RMS) 80% (0.76 RMS)
Cardboard (40 lbs) 
60% (1.33 RMS) 80% (1.64 RMS)
Duct Tape (20 lbs) 
60% (5.64 RMS) 80% (N/A RMS) (15 lbs)
Teflon (20 lbs) 
40% (0.57 RMS)
Medium Tip Experiment 
Coupling Materials and Tips
Medium Tip Experiment 
• Coin + Duct Tape (40 lbs) 
• Coin + Duct Tape (60 lbs) 
• Coin + Teflon (40 lbs) 
• Coin + Teflon (20 lbs) 
• Business Card (40 lbs) 
• Cardboard (40 lbs) 
• Duct Tape (40 lbs)
Coin + Duct Tape (40 lbs) 
60% (0.67 RMS) 100% (0.96 RMS)
Coin + Duct Tape (60 lbs) 
60% (0.85 RMS) 100% (0.84 RMS)
Coin + Teflon (40 lbs) 
60% (0.75 RMS) 100% (1.15 RMS)
Coin + Teflon (20 lbs) 
60% (0.19 RMS) 100% (5.95 RMS)
Business Card (40 lbs) 
60% (0.84 RMS) 100% (1.53 RMS)
Cardboard (40 lbs) 
60% (0.76 RMS) 100% (0.93 RMS)
Large Tip Experiment 
Coupling Materials and Tips
Large Tip Experiment 
• Coin + Duct Tape (40 lbs) 
• Coin + Teflon (20 lbs) 
• Business Card 
– (40 lbs) 
– (60 lbs) 
• Cardboard 
– (40 lbs) 
– (60 lbs) 
• Duct Tape 
– (15 lbs) 
– (30 lbs)
Coin + Teflon (40 lbs) 
60% (0.71 RMS) 100% (2.18 RMS)
Coin + Teflon (20 lbs) 
60% (0.90 RMS) 100% (1.61 RMS)
Business Card (40 lbs) 
60% (0.69 RMS) 100% (1.07 RMS)
Business Card (60 lbs) 
60% (0.84 RMS) 100% (1.19 RMS)
Cardboard (40 lbs) 
60% (0.29 RMS) 100% (0.44 RMS)
Cardboard (60 lbs) 
60% (0.39 RMS) 100% (0.57 RMS)
Duct Tape (15 lbs) 
60% (0.81 RMS) 100% (0.93 RMS)
Duct Tape (30 lbs) 
60% (1.18 RMS) 100% (1.42 RMS)
Best Coupling Materials Setup 
1. Business Card with Large tip 
2. Cardboard with Large tip
Vibration Analysis 
(Width-Wise)
I beam panel ; 20 kHz; 20lbs; Without coin; 
Conclusion: I beam structure is very hard to vibrate effectively. 
R1 Gun 
V 
R3 
R2 
Name of shot RMS of vilocity (mm/s) 
Ref1 Vib Ref3 Ref2 
02 52.69 47.61 19.07 45.99 
03 63.02 54.45 16.67 35.12 
04 77.46 36.31 23.62 42.52 
05 70.66 31.48 20.70 36.18 
07 75.92 37.96 24.51 44.71
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
bellimpactpanel03/ref2 
Max Magnitude = 6.83 m/s RMS = 30.07mm/s
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
bellimpactpanel03/vib 
Max Magnitude = 9.95 m/s RMS = 31.05mm/s
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
bellimpactpanel03/ref1 
Max Magnitude = 31.63 m/s RMS = 84.55mm/s
Ultrasound 
20KHz 
20k LPt 20:0.5:902 
50% Amplitude 
800ms Pulse 
Camera 
40 Frames/s 
IR Camera 6s 
File 
Bellimpactpanel03/ref3 
Max Magnitude = 36.13 m/s RMS = 88.58m
15 inches 
J 
G 
N 
R 
B 
Defect J shows up 
Power Amplitude: 10% 
Load: 25lbs 
Input Frequency: 20kHz 
RMS(Root Mean Square): 41.55mm/s
15 inches 
Defect N shows up 
Power Amplitude: 20% 
Load: 25lbs 
Input Frequency: 20kHz 
RMS(Root Mean Square): 117.7mm/s 
J 
G 
N 
R 
B
25 inches 
Defect G shows up 
Power Amplitude: 45% 
Load: 25lbs 
Input Frequency: 20kHz 
RMS(Root Mean Square): 90.66mm/s 
J 
G 
N 
R 
B
25 inches 
Defect R shows up 
Power Amplitude: 45% 
Load: 25lbs 
Input Frequency: 20kHz 
RMS(Root Mean Square): 44.08mm/s 
J 
G 
N 
R 
B
30 inches 
Defect B shows up 
Power Amplitude: 35% 
Load: 25lbs 
Input Frequency: 20kHz 
RMS(Root Mean Square): 53.73mm/s 
J 
G 
N 
R 
B
Vibration Analysis 
(Length-wise)
10 inches 
180 
160 
140 
120 
100 
80 
60 
40 
20 
0 
Vibration Data in Thin Channel 
0 10 20 30 40 50 60 
Vibration Amplitude(mm/s) 
Distance from Transducer (inches) 
10% Amplitude 
20% Amplitude 
30% Amplitude 
40% Amplitude 
50% Amplitude 
Transducer
10 inches 
Transducer 
140 
120 
100 
80 
60 
40 
20 
0 
Vibration Data in Doubler Channel 
0 10 20 30 40 50 60 
Vibration Amplitude (mm/s) 
Distance from Transducer (inches) 
10% Amplitude 
20% Amplitude 
30% Amplitude 
40% Amplitude 
50% Amplitude
4 4 8 10 
Gun 
I beam panel ; Center doubler channel; 20 kHz; 20lbs; Without coin; 
Conclusion: Experiment is repeatable and vibration results show some periodicity property.
Gun 
4 4 8 10 
I beam panel ; 20 kHz; 20lbs; Without coin; 
Conclusion: Experiment is repeatable and vibration results show some periodicity property.
Hat Pattern Consistency Test 
(Vibration Study)
Mode Pattern 
Left Gun Shot
Time Lapse 
Left Gun Shot
Mode Pattern 
Middle Gun Shot
Time Lapse 
Middle Gun Shot
Mode Pattern 
Right Gun Shot
Time Lapse 
Right Gun Shot
Video
Consistency Test 
Right Shot 1
Consistency Test 
Right Shot 2
Consistency Test 
Right Shot 3
Consistency Test 
Right Shot 4
Consistency Test 
Right Shot 6
Consistency Test 
Right Shot 7
Consistency Test 
Right Shot 8
Consistency Test 
Right Shot 9
Mode Pattern Consistency 
(Discard 3) 
1 2 3 4 5 6 7 
1 O O O O O 
2 O O O O O O 
3 
4 O O O O O O O 
5 O O O O O O 
6 O O O O O O O 
7 O O O O O O 
8 O O O O O 
9 O O O O O O O 
10 O O O O O O 
11 O O O O O O O 
12 O O O O O O 
13 O O O O O O O 
1 2 3 4 5 6 7 8 9 10 11 12 13 
5 6 0 7 6 7 6 5 7 6 7 6 7
Vibrometer Test 
OFV-551 Vib 
Distance (mm) RMS (mm/s) Signal Level Stability 
135 2.60 1.00 High 
339 2.60 1.00 High 
543 2.60 1.00 High 
747 2.60 1.00 High 
951 2.59 1.00 High 
237 2.60 1.00 High 
441 2.60 1.00 High 
645 2.60 1.00 Low 
849 2.60 1.00 High 
1053 2.60 0.90 Low 
Average 2.60 0.99 High 
OFV-511 Ref3 
Distance (mm) RMS (mm/s) Signal Level Stability 
232 2.58 0.85 No 
437 2.56 0.70 No 
642 2.56 0.30 No 
847 2.55 0.35 No 
1052 2.56 0.00 No 
335 2.56 0.60 No 
540 Error 0.35 No 
745 2.56 0.40 No 
950 2.55 0.45 No 
1155 2.53 0.40 No 
Average 2.56 0.44 No
Defects 1 & 2 
1.2 
1 
0.8 
0.6 
0.4 
0.2 
0 
Source, 1 and 2 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 
RMS 
Trial # 
1 
2 
Source
Defects 4 & 11 
2.5 
2 
1.5 
1 
0.5 
0 
Source, 4 and 11 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 
Axis Title 
Axis Title 
4 
11 
Source
Defects 5 & 10 
2.5 
2 
1.5 
1 
0.5 
0 
Source, 5 and 10 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 
RMS (m/s) 
Trial # 
Source 
10 
5
Defects 6 & 13 
1.6 
1.4 
1.2 
1 
0.8 
0.6 
0.4 
0.2 
0 
Source, 6 and 13 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 
RMS (m/s) 
Trial # 
6 
13 
Source
Defects 7 & 12 
2 
1.8 
1.6 
1.4 
1.2 
1 
0.8 
0.6 
0.4 
0.2 
0 
Source, 7 and 12 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 
RMS (m/s) 
Trial # 
7 
12 
Source
Defects 8 & 9 
2 
1.8 
1.6 
1.4 
1.2 
1 
0.8 
0.6 
0.4 
0.2 
0 
Source, 8 and 9 
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 
RMS 
Trial # 
8 
9 
Source
New Approach 
• Fixed Camera 
• Enables Image Processing
High Vibration
Steps 
• Fixed Camera Images 
• Convert to Binary 
• Image Registration (Optional) 
• Binary Operation 
• Point selection
Registration
Binary Operation
Selected Points
Results 
1 2 3 4 5 6 
Point 0.45 0.14 0.18 0.10 0.10 0.14 
Source 2.29 2.25 2.32 2.49 2.43 2.38 
Ratio 0.20 0.06 0.08 0.04 0.04 0.06
Low Vibration
Low Vibration
Results 
1 2 3 4 5 6 
Point 0.09 0.08 0.04 0.06 0.09 0.06 
Source 2.08 2.28 2.00 1.86 2.50 2.22 
Ratio 0.04 0.04 0.02 0.03 0.04 0.03
Results 
High 
1 2 3 4 5 6 
Point 0.45 0.14 0.18 0.10 0.10 0.14 
Source 2.29 2.25 2.32 2.49 2.43 2.38 
Ratio 0.20 0.06 0.08 0.04 0.04 0.06 
Low 
1 2 3 4 5 6 
Point 0.09 0.08 0.04 0.06 0.09 0.06 
Source 2.08 2.28 2.00 1.86 2.50 2.22 
Ratio 0.04 0.04 0.02 0.03 0.04 0.03
Conclusion 
• High density granule areas indicate low 
vibration 
• Low density granule areas indicate high 
vibration
TT-Plot Automation 
.
Hat Defect Locations
Legend: 
< 1” Diameter, 9 Each 
1.0” – 1.75” Diameter, 6 Each 
1.75” - 2.65” Diameter, 4 Each 
> 2.65” Diameter, 3 Each 
0.170” Skin 
0.096” Skin 
Impact Panel for Wayne State University 
60” 
9/30/2011 Final Impact Plan - RJB
TT-Plot on Hat
Sample of TT-Plot from 
Defect J 
Video = V 
Frames = F1…Fn 
F1…Fn ∈ V 
Pixels = P1,1…Pn,n 
TT-Plot = P for all F 
Defect J = J (White) 
J = {P} 
J ⊆ F 
TT-Plot 
Threshold Sampling
TT-Plot for Frame Subset (Pixels) for all 
Video 
0 20 40 60 80 100 120 140 160 180 200 
1 
0.95 
0.9 
0.85 
0.8 
0.75 
0.7 
0.65 
0.6 
0.55 
All Defects in Neighborhood + Error 
Pixel Value 
Frame
Min and Max Graphs for TT-Plot 
0 20 40 60 80 100 120 140 160 180 200 
1.2 
1.1 
1 
0.9 
0.8 
0.7 
0.6 
0.5 
Max and Min Defect in Neighborhood 
Pixel Value 
Frame
Scanned Matching TT-Plots that fall 
between Min and Max TT-Plots
Repeated for Defect F
TT-Plots and Min/Max 
0 20 40 60 80 100 120 140 160 180 200 
1 
0.9 
0.8 
0.7 
0.6 
0.5 
0.4 
0.3 
0.2 
All Defects in Neighborhood + Error 
Pixel Value 
Frame 
0 20 40 60 80 100 120 140 160 180 200 
1.1 
1 
0.9 
0.8 
0.7 
0.6 
0.5 
0.4 
0.3 
0.2 
0.1 
Max and Min Defect in Neighborhood 
Pixel Value 
Frame
Matching TT-Plots
Combined Matching TT-Plots
Sample of TT-Plots on defect J 
with subtraction
TT-Plots and Min/Max 
0 20 40 60 80 100 120 140 160 180 200 
1 
0.9 
0.8 
0.7 
0.6 
0.5 
0.4 
0.3 
0.2 
0.1 
0 
All Defects in Neighborhood 
Pixel Value 
Frame 
0 20 40 60 80 100 120 140 160 180 200 
1.2 
1 
0.8 
0.6 
0.4 
0.2 
0 
-0.2 
Max and Min Defect in Neighborhood + 9% Error 
Pixel Value 
Frame
Scanned Matching TT-Plots that fall 
between Min and Max TT-Plots
4 Point Smooth, 4% Error, 31% 
Threshold 
6 Point Smooth, 6% Error, 31% 
Threshold
4 Point Smooth, 7% Error, 31% 
Threshold 
6 Point Smooth, 7% Error, 31% 
Threshold
4 Point Smooth, 4% Error, 27% 
Threshold 
6 Point Smooth, 4% Error, 27% 
Threshold
With and Without TT-Plot Automation 
Before After
Conclusion 
• Shows possibility of automatic defect 
detection 
• Does not support hypothesis that depth of 
defect is directly proportional to how slow it 
shows up on camera
Final Conclusion 
• Behavior of Panels Observed 
• Optimal Experimental Settings Tested 
– Frequency 
– Coupling Material 
– Tip Size/Amplitude 
• Additional Areas of Study 
– Mode Pattern Vibration 
– Automation
Future Work 
• Further study on mode pattern 
• Further study on TT-Plot shape of 
defects/depth of defects

More Related Content

Similar to Sonic IR Imaging on Bell Composite Panels

MagneTag Presentation Winter Quarter V3.0
MagneTag Presentation Winter Quarter V3.0MagneTag Presentation Winter Quarter V3.0
MagneTag Presentation Winter Quarter V3.0John-Paul Petersen
 
Hioki 3291 50-clamp-on_hi_tester_datasheet
Hioki 3291 50-clamp-on_hi_tester_datasheetHioki 3291 50-clamp-on_hi_tester_datasheet
Hioki 3291 50-clamp-on_hi_tester_datasheetAngus Sankaran
 
Compact goniophotometer
Compact goniophotometerCompact goniophotometer
Compact goniophotometer世满 江
 
Compact goniophotometer
Compact goniophotometerCompact goniophotometer
Compact goniophotometer世满 江
 
Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2
Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2
Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2Celeste Yang
 
Catalog 2023 | Integrix Optics | Optics Trade
Catalog 2023 | Integrix Optics | Optics TradeCatalog 2023 | Integrix Optics | Optics Trade
Catalog 2023 | Integrix Optics | Optics TradeOptics-Trade
 
High-Speed Balanced Photoreceiver
High-Speed Balanced Photoreceiver High-Speed Balanced Photoreceiver
High-Speed Balanced Photoreceiver Anne Stiegler
 
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
Textile testing-equipments-1224828041803735-9
Textile testing-equipments-1224828041803735-9Textile testing-equipments-1224828041803735-9
Textile testing-equipments-1224828041803735-9Mazedul Islam Mazed
 
Metodiautonomico2015
Metodiautonomico2015Metodiautonomico2015
Metodiautonomico2015PahPavia
 
Avigilon h5a-camera-line-datasheet-en-rev7 2
Avigilon h5a-camera-line-datasheet-en-rev7 2Avigilon h5a-camera-line-datasheet-en-rev7 2
Avigilon h5a-camera-line-datasheet-en-rev7 2GietisakRuengsawat
 
Calibration Laboratory In India - FARE Labs Pvt. Ltd.
Calibration Laboratory In India - FARE Labs Pvt. Ltd.Calibration Laboratory In India - FARE Labs Pvt. Ltd.
Calibration Laboratory In India - FARE Labs Pvt. Ltd.Fare Labs
 
Hioki IR4056-20 Digital Insulation Resistance Tester
Hioki IR4056-20 Digital Insulation Resistance Tester Hioki IR4056-20 Digital Insulation Resistance Tester
Hioki IR4056-20 Digital Insulation Resistance Tester ROnex International
 
Airscan TX Brochure
Airscan TX BrochureAirscan TX Brochure
Airscan TX BrochureImaginos NDE
 
Compact Goniophotometer
Compact GoniophotometerCompact Goniophotometer
Compact GoniophotometerLisun Group
 
LAMBDA Specifications Sheet 35, 265, 365, 465
LAMBDA Specifications Sheet 35, 265, 365, 465LAMBDA Specifications Sheet 35, 265, 365, 465
LAMBDA Specifications Sheet 35, 265, 365, 465PerkinElmer, Inc.
 
Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024Haag-Streit UK (HS-UK)
 
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62PT. Siwali Swantika
 

Similar to Sonic IR Imaging on Bell Composite Panels (20)

MagneTag Presentation Winter Quarter V3.0
MagneTag Presentation Winter Quarter V3.0MagneTag Presentation Winter Quarter V3.0
MagneTag Presentation Winter Quarter V3.0
 
Hioki 3291 50-clamp-on_hi_tester_datasheet
Hioki 3291 50-clamp-on_hi_tester_datasheetHioki 3291 50-clamp-on_hi_tester_datasheet
Hioki 3291 50-clamp-on_hi_tester_datasheet
 
Hioki series battery_tester_denkei
Hioki series battery_tester_denkeiHioki series battery_tester_denkei
Hioki series battery_tester_denkei
 
Compact goniophotometer
Compact goniophotometerCompact goniophotometer
Compact goniophotometer
 
Compact goniophotometer
Compact goniophotometerCompact goniophotometer
Compact goniophotometer
 
Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2
Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2
Fluke Multimode OTDR for Troubleshooting and Extended Certification – 2
 
Catalog 2023 | Integrix Optics | Optics Trade
Catalog 2023 | Integrix Optics | Optics TradeCatalog 2023 | Integrix Optics | Optics Trade
Catalog 2023 | Integrix Optics | Optics Trade
 
High-Speed Balanced Photoreceiver
High-Speed Balanced Photoreceiver High-Speed Balanced Photoreceiver
High-Speed Balanced Photoreceiver
 
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
 
Textile testing-equipments-1224828041803735-9
Textile testing-equipments-1224828041803735-9Textile testing-equipments-1224828041803735-9
Textile testing-equipments-1224828041803735-9
 
Metodiautonomico2015
Metodiautonomico2015Metodiautonomico2015
Metodiautonomico2015
 
Katalog produk-agilent-technologies-tridinamika
Katalog produk-agilent-technologies-tridinamikaKatalog produk-agilent-technologies-tridinamika
Katalog produk-agilent-technologies-tridinamika
 
Avigilon h5a-camera-line-datasheet-en-rev7 2
Avigilon h5a-camera-line-datasheet-en-rev7 2Avigilon h5a-camera-line-datasheet-en-rev7 2
Avigilon h5a-camera-line-datasheet-en-rev7 2
 
Calibration Laboratory In India - FARE Labs Pvt. Ltd.
Calibration Laboratory In India - FARE Labs Pvt. Ltd.Calibration Laboratory In India - FARE Labs Pvt. Ltd.
Calibration Laboratory In India - FARE Labs Pvt. Ltd.
 
Hioki IR4056-20 Digital Insulation Resistance Tester
Hioki IR4056-20 Digital Insulation Resistance Tester Hioki IR4056-20 Digital Insulation Resistance Tester
Hioki IR4056-20 Digital Insulation Resistance Tester
 
Airscan TX Brochure
Airscan TX BrochureAirscan TX Brochure
Airscan TX Brochure
 
Compact Goniophotometer
Compact GoniophotometerCompact Goniophotometer
Compact Goniophotometer
 
LAMBDA Specifications Sheet 35, 265, 365, 465
LAMBDA Specifications Sheet 35, 265, 365, 465LAMBDA Specifications Sheet 35, 265, 365, 465
LAMBDA Specifications Sheet 35, 265, 365, 465
 
Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024
 
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
 

Recently uploaded

Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...ronahami
 
Online food ordering system project report.pdf
Online food ordering system project report.pdfOnline food ordering system project report.pdf
Online food ordering system project report.pdfKamal Acharya
 
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...vershagrag
 
Computer Networks Basics of Network Devices
Computer Networks  Basics of Network DevicesComputer Networks  Basics of Network Devices
Computer Networks Basics of Network DevicesChandrakantDivate1
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaOmar Fathy
 
Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...
Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...
Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...gragchanchal546
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdfKamal Acharya
 
DeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakesDeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakesMayuraD1
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdfKamal Acharya
 
Online electricity billing project report..pdf
Online electricity billing project report..pdfOnline electricity billing project report..pdf
Online electricity billing project report..pdfKamal Acharya
 
A Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityA Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityMorshed Ahmed Rahath
 
PE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiesPE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiessarkmank1
 
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptxOrlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptxMuhammadAsimMuhammad6
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . pptDineshKumar4165
 
Ground Improvement Technique: Earth Reinforcement
Ground Improvement Technique: Earth ReinforcementGround Improvement Technique: Earth Reinforcement
Ground Improvement Technique: Earth ReinforcementDr. Deepak Mudgal
 
Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayEpec Engineered Technologies
 
UNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptxUNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptxkalpana413121
 
DC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationDC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationBhangaleSonal
 
Theory of Time 2024 (Universal Theory for Everything)
Theory of Time 2024 (Universal Theory for Everything)Theory of Time 2024 (Universal Theory for Everything)
Theory of Time 2024 (Universal Theory for Everything)Ramkumar k
 

Recently uploaded (20)

Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
 
Online food ordering system project report.pdf
Online food ordering system project report.pdfOnline food ordering system project report.pdf
Online food ordering system project report.pdf
 
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
 
Computer Networks Basics of Network Devices
Computer Networks  Basics of Network DevicesComputer Networks  Basics of Network Devices
Computer Networks Basics of Network Devices
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS Lambda
 
Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...
Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...
Ghuma $ Russian Call Girls Ahmedabad ₹7.5k Pick Up & Drop With Cash Payment 8...
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdf
 
DeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakesDeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakes
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdf
 
Online electricity billing project report..pdf
Online electricity billing project report..pdfOnline electricity billing project report..pdf
Online electricity billing project report..pdf
 
A Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityA Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna Municipality
 
PE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiesPE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and properties
 
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptxOrlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
Ground Improvement Technique: Earth Reinforcement
Ground Improvement Technique: Earth ReinforcementGround Improvement Technique: Earth Reinforcement
Ground Improvement Technique: Earth Reinforcement
 
Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power Play
 
UNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptxUNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptx
 
Integrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - NeometrixIntegrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - Neometrix
 
DC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationDC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equation
 
Theory of Time 2024 (Universal Theory for Everything)
Theory of Time 2024 (Universal Theory for Everything)Theory of Time 2024 (Universal Theory for Everything)
Theory of Time 2024 (Universal Theory for Everything)
 

Sonic IR Imaging on Bell Composite Panels

  • 1. Sonic IR Imaging on Bell Imact Panels (Hat and I-Beam) By: Justin Ar-Rasheed
  • 2. Overview • Behavior of Panels Observed • Optimal Experimental Settings Tested – Frequency – Coupling Material – Tip Size/Amplitude • Additional Areas of Study – Mode Pattern Vibration – Automation
  • 3.
  • 4. Legend: < 1” Diameter, 9 Each 1.0” – 1.75” Diameter, 6 Each 1.75” - 2.65” Diameter, 4 Each > 2.65” Diameter, 3 Each 0.170” Skin 0.096” Skin Impact Panel for Wayne State University 60” 9/30/2011 Final Impact Plan - RJB
  • 6. Test Experiments • Shoot gun on various locations of panel • Analyze results and come up with hypothesis
  • 7. Ultrasound 20KHz Handheld 20k LPt 20:0.5:902 40% Amplitude 800ms Pulse Camera 30 Frames/s IR Camera 8s File 20K_handheld/40%_30fps_8s_01.fts Panel mounted vertically
  • 8. Ultrasound 40KHz Handheld 40k Lpe 40:0.5:4T 100% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File 40k_handheld/100%_40fps_6s_02.fts
  • 9. Ultrasound 40KHz Table Gun 40k Lpe 40:0.5:4T 100% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File 40k_tableGun/40fra_6s_800ms_30lbs_coin.fts
  • 10. Ultrasound 40KHz Handheld 40k Lpe 40:0.5:4T 100% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File 40K_handheld/100%_40fps_6s_01_top.fts
  • 11. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s Files 20PartitionShot/40%_Top33%_1.2s_40fps_6s 20PartitionShot/50%_Top_left_33%_800ms_40fps_6s_01 20PartitionShot/50%_Top_right_33%_800ms_40fps_6s01 Left Middle Right Left Middle Right
  • 12. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s Files 20PartitionShot/40%_Mid33%_1.2s_40fps_6s 20PartitionShot/50%_Mid_left_33%_800ms_40fps_6s_02 20PartitionShot/50%_Mid_right_33%_800ms_40fps_6s_02 Left Middle Right Left Middle Right
  • 13. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s Files 20PartitionShot/50%_bottom_left_33%_800ms_40fps_6 s_01 20PartitionShot/50%_bottom_right_33%_800ms_40fps _6s_01 20PartitionShot/50%_Mid_left_33%_800ms_40fps_6s_0 2 Left Middle Right Left Middle Right
  • 14. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File 20PartitionShot
  • 15. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File Bellimpactpanel03/Bottomhalfve iw40fps_50%_800ms_6s_02 Panel mounted horizontally
  • 16. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File Bellimpactpanel03/Fullveiw40fps_50%_8 00ms_6s_03
  • 17. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File Bellimpactpanel03/Tophalfveiw40fps_50 %_800ms_6s_01
  • 18. Experiment Setup: 20kHz Handheld (500W from backside) 40 frames per second Total 6 seconds 50% amplitude
  • 19. Experimental Setup: 20kHz Handheld (500W) 40 frames per second Total 6 seconds 50% amplitude
  • 20. Experimental Setup: 20kHz Handheld(500W) 40 frames per second Total 6 seconds 50% amplitude File Name : 20K_handheld_Amp50_middle_GunHandI_LogoLeftBottom.bmp
  • 21. Experimental Setup: 20kHz Handheld (500W) 40 frames per second Total 6 seconds 60% amplitude File Name : 20K_handheld_Amp60_middle_GunGandN_LogoLeftBottom.bmp
  • 22. Experimental Setup: 20kHz Handheld (500W) 40 frames per second Total 6 seconds 60% amplitude File Name : 20K_handheld_Top_Amp60_GunOnTop_4inFromEdge.bmp
  • 23. Experimental Setup: 20kHz Handheld(500w) 40 frames per second Total 6 seconds 60% amplitude File Name : 20K_handheld_Bottom_Amp60_GunOnBottom_2inFromEdge.bmp
  • 24.
  • 26. Structure I - Beam Thin area Doublers Area Size L M S L M S L M S Front View(20k Table and Handheld) C,D,P,Q A G,F,N,M J,K,L I Back View(20kTable and Handheld) R,O S Never Seen (Both 20k &40k) B, E H
  • 32. Small Tip Experiment Coupling Materials and Tips
  • 33. Small Tip Experiment • Coin + Business Card (20 lbs) • Coin + Cardboard (20 lbs) • Coin + Duct Tape (20 lbs) • Coin + Teflon (20 lbs) • Business Card – (20 lbs) – (40 lbs) • Cardboard – (20 lbs) – (40 lbs) • Duct Tape (20 lbs) • Teflon (20 lbs)
  • 34. Coin + Business Card (20 lbs) 60% (4.56 RMS) 80% (6.27 RMS)
  • 35. Coin + Cardboard (20 lbs) 60% (0.85 RMS) 80% (1.80 RMS)
  • 36. Coin + Duct Tape (20 lbs) 60% (4.22 RMS) 80% (3.46 RMS)
  • 37. Coin + Teflon (20 lbs) 60% (4.75 RMS)
  • 38. Business Card (20 lbs) 60% (1.35 RMS) 80% (1.27 RMS)
  • 39. Business Card (40 lbs) 60% (1.53 RMS) 80% (2.25 RMS)
  • 40. Cardboard (20 lbs) 60% (0.64 RMS) 80% (0.76 RMS)
  • 41. Cardboard (40 lbs) 60% (1.33 RMS) 80% (1.64 RMS)
  • 42. Duct Tape (20 lbs) 60% (5.64 RMS) 80% (N/A RMS) (15 lbs)
  • 43. Teflon (20 lbs) 40% (0.57 RMS)
  • 44. Medium Tip Experiment Coupling Materials and Tips
  • 45. Medium Tip Experiment • Coin + Duct Tape (40 lbs) • Coin + Duct Tape (60 lbs) • Coin + Teflon (40 lbs) • Coin + Teflon (20 lbs) • Business Card (40 lbs) • Cardboard (40 lbs) • Duct Tape (40 lbs)
  • 46. Coin + Duct Tape (40 lbs) 60% (0.67 RMS) 100% (0.96 RMS)
  • 47. Coin + Duct Tape (60 lbs) 60% (0.85 RMS) 100% (0.84 RMS)
  • 48. Coin + Teflon (40 lbs) 60% (0.75 RMS) 100% (1.15 RMS)
  • 49. Coin + Teflon (20 lbs) 60% (0.19 RMS) 100% (5.95 RMS)
  • 50. Business Card (40 lbs) 60% (0.84 RMS) 100% (1.53 RMS)
  • 51. Cardboard (40 lbs) 60% (0.76 RMS) 100% (0.93 RMS)
  • 52. Large Tip Experiment Coupling Materials and Tips
  • 53. Large Tip Experiment • Coin + Duct Tape (40 lbs) • Coin + Teflon (20 lbs) • Business Card – (40 lbs) – (60 lbs) • Cardboard – (40 lbs) – (60 lbs) • Duct Tape – (15 lbs) – (30 lbs)
  • 54. Coin + Teflon (40 lbs) 60% (0.71 RMS) 100% (2.18 RMS)
  • 55. Coin + Teflon (20 lbs) 60% (0.90 RMS) 100% (1.61 RMS)
  • 56. Business Card (40 lbs) 60% (0.69 RMS) 100% (1.07 RMS)
  • 57. Business Card (60 lbs) 60% (0.84 RMS) 100% (1.19 RMS)
  • 58. Cardboard (40 lbs) 60% (0.29 RMS) 100% (0.44 RMS)
  • 59. Cardboard (60 lbs) 60% (0.39 RMS) 100% (0.57 RMS)
  • 60. Duct Tape (15 lbs) 60% (0.81 RMS) 100% (0.93 RMS)
  • 61. Duct Tape (30 lbs) 60% (1.18 RMS) 100% (1.42 RMS)
  • 62. Best Coupling Materials Setup 1. Business Card with Large tip 2. Cardboard with Large tip
  • 64. I beam panel ; 20 kHz; 20lbs; Without coin; Conclusion: I beam structure is very hard to vibrate effectively. R1 Gun V R3 R2 Name of shot RMS of vilocity (mm/s) Ref1 Vib Ref3 Ref2 02 52.69 47.61 19.07 45.99 03 63.02 54.45 16.67 35.12 04 77.46 36.31 23.62 42.52 05 70.66 31.48 20.70 36.18 07 75.92 37.96 24.51 44.71
  • 65. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File bellimpactpanel03/ref2 Max Magnitude = 6.83 m/s RMS = 30.07mm/s
  • 66. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File bellimpactpanel03/vib Max Magnitude = 9.95 m/s RMS = 31.05mm/s
  • 67. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File bellimpactpanel03/ref1 Max Magnitude = 31.63 m/s RMS = 84.55mm/s
  • 68. Ultrasound 20KHz 20k LPt 20:0.5:902 50% Amplitude 800ms Pulse Camera 40 Frames/s IR Camera 6s File Bellimpactpanel03/ref3 Max Magnitude = 36.13 m/s RMS = 88.58m
  • 69. 15 inches J G N R B Defect J shows up Power Amplitude: 10% Load: 25lbs Input Frequency: 20kHz RMS(Root Mean Square): 41.55mm/s
  • 70. 15 inches Defect N shows up Power Amplitude: 20% Load: 25lbs Input Frequency: 20kHz RMS(Root Mean Square): 117.7mm/s J G N R B
  • 71. 25 inches Defect G shows up Power Amplitude: 45% Load: 25lbs Input Frequency: 20kHz RMS(Root Mean Square): 90.66mm/s J G N R B
  • 72. 25 inches Defect R shows up Power Amplitude: 45% Load: 25lbs Input Frequency: 20kHz RMS(Root Mean Square): 44.08mm/s J G N R B
  • 73. 30 inches Defect B shows up Power Amplitude: 35% Load: 25lbs Input Frequency: 20kHz RMS(Root Mean Square): 53.73mm/s J G N R B
  • 75. 10 inches 180 160 140 120 100 80 60 40 20 0 Vibration Data in Thin Channel 0 10 20 30 40 50 60 Vibration Amplitude(mm/s) Distance from Transducer (inches) 10% Amplitude 20% Amplitude 30% Amplitude 40% Amplitude 50% Amplitude Transducer
  • 76. 10 inches Transducer 140 120 100 80 60 40 20 0 Vibration Data in Doubler Channel 0 10 20 30 40 50 60 Vibration Amplitude (mm/s) Distance from Transducer (inches) 10% Amplitude 20% Amplitude 30% Amplitude 40% Amplitude 50% Amplitude
  • 77. 4 4 8 10 Gun I beam panel ; Center doubler channel; 20 kHz; 20lbs; Without coin; Conclusion: Experiment is repeatable and vibration results show some periodicity property.
  • 78. Gun 4 4 8 10 I beam panel ; 20 kHz; 20lbs; Without coin; Conclusion: Experiment is repeatable and vibration results show some periodicity property.
  • 79. Hat Pattern Consistency Test (Vibration Study)
  • 80. Mode Pattern Left Gun Shot
  • 81. Time Lapse Left Gun Shot
  • 83. Time Lapse Middle Gun Shot
  • 84. Mode Pattern Right Gun Shot
  • 85. Time Lapse Right Gun Shot
  • 86. Video
  • 95. Mode Pattern Consistency (Discard 3) 1 2 3 4 5 6 7 1 O O O O O 2 O O O O O O 3 4 O O O O O O O 5 O O O O O O 6 O O O O O O O 7 O O O O O O 8 O O O O O 9 O O O O O O O 10 O O O O O O 11 O O O O O O O 12 O O O O O O 13 O O O O O O O 1 2 3 4 5 6 7 8 9 10 11 12 13 5 6 0 7 6 7 6 5 7 6 7 6 7
  • 96.
  • 97. Vibrometer Test OFV-551 Vib Distance (mm) RMS (mm/s) Signal Level Stability 135 2.60 1.00 High 339 2.60 1.00 High 543 2.60 1.00 High 747 2.60 1.00 High 951 2.59 1.00 High 237 2.60 1.00 High 441 2.60 1.00 High 645 2.60 1.00 Low 849 2.60 1.00 High 1053 2.60 0.90 Low Average 2.60 0.99 High OFV-511 Ref3 Distance (mm) RMS (mm/s) Signal Level Stability 232 2.58 0.85 No 437 2.56 0.70 No 642 2.56 0.30 No 847 2.55 0.35 No 1052 2.56 0.00 No 335 2.56 0.60 No 540 Error 0.35 No 745 2.56 0.40 No 950 2.55 0.45 No 1155 2.53 0.40 No Average 2.56 0.44 No
  • 98. Defects 1 & 2 1.2 1 0.8 0.6 0.4 0.2 0 Source, 1 and 2 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 RMS Trial # 1 2 Source
  • 99. Defects 4 & 11 2.5 2 1.5 1 0.5 0 Source, 4 and 11 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 Axis Title Axis Title 4 11 Source
  • 100. Defects 5 & 10 2.5 2 1.5 1 0.5 0 Source, 5 and 10 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 RMS (m/s) Trial # Source 10 5
  • 101. Defects 6 & 13 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 Source, 6 and 13 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 RMS (m/s) Trial # 6 13 Source
  • 102. Defects 7 & 12 2 1.8 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 Source, 7 and 12 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 RMS (m/s) Trial # 7 12 Source
  • 103. Defects 8 & 9 2 1.8 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 Source, 8 and 9 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 RMS Trial # 8 9 Source
  • 104. New Approach • Fixed Camera • Enables Image Processing
  • 106. Steps • Fixed Camera Images • Convert to Binary • Image Registration (Optional) • Binary Operation • Point selection
  • 107.
  • 111. Results 1 2 3 4 5 6 Point 0.45 0.14 0.18 0.10 0.10 0.14 Source 2.29 2.25 2.32 2.49 2.43 2.38 Ratio 0.20 0.06 0.08 0.04 0.04 0.06
  • 114. Results 1 2 3 4 5 6 Point 0.09 0.08 0.04 0.06 0.09 0.06 Source 2.08 2.28 2.00 1.86 2.50 2.22 Ratio 0.04 0.04 0.02 0.03 0.04 0.03
  • 115. Results High 1 2 3 4 5 6 Point 0.45 0.14 0.18 0.10 0.10 0.14 Source 2.29 2.25 2.32 2.49 2.43 2.38 Ratio 0.20 0.06 0.08 0.04 0.04 0.06 Low 1 2 3 4 5 6 Point 0.09 0.08 0.04 0.06 0.09 0.06 Source 2.08 2.28 2.00 1.86 2.50 2.22 Ratio 0.04 0.04 0.02 0.03 0.04 0.03
  • 116. Conclusion • High density granule areas indicate low vibration • Low density granule areas indicate high vibration
  • 119. Legend: < 1” Diameter, 9 Each 1.0” – 1.75” Diameter, 6 Each 1.75” - 2.65” Diameter, 4 Each > 2.65” Diameter, 3 Each 0.170” Skin 0.096” Skin Impact Panel for Wayne State University 60” 9/30/2011 Final Impact Plan - RJB
  • 121. Sample of TT-Plot from Defect J Video = V Frames = F1…Fn F1…Fn ∈ V Pixels = P1,1…Pn,n TT-Plot = P for all F Defect J = J (White) J = {P} J ⊆ F TT-Plot Threshold Sampling
  • 122. TT-Plot for Frame Subset (Pixels) for all Video 0 20 40 60 80 100 120 140 160 180 200 1 0.95 0.9 0.85 0.8 0.75 0.7 0.65 0.6 0.55 All Defects in Neighborhood + Error Pixel Value Frame
  • 123. Min and Max Graphs for TT-Plot 0 20 40 60 80 100 120 140 160 180 200 1.2 1.1 1 0.9 0.8 0.7 0.6 0.5 Max and Min Defect in Neighborhood Pixel Value Frame
  • 124. Scanned Matching TT-Plots that fall between Min and Max TT-Plots
  • 126. TT-Plots and Min/Max 0 20 40 60 80 100 120 140 160 180 200 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 All Defects in Neighborhood + Error Pixel Value Frame 0 20 40 60 80 100 120 140 160 180 200 1.1 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 Max and Min Defect in Neighborhood Pixel Value Frame
  • 129. Sample of TT-Plots on defect J with subtraction
  • 130. TT-Plots and Min/Max 0 20 40 60 80 100 120 140 160 180 200 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 All Defects in Neighborhood Pixel Value Frame 0 20 40 60 80 100 120 140 160 180 200 1.2 1 0.8 0.6 0.4 0.2 0 -0.2 Max and Min Defect in Neighborhood + 9% Error Pixel Value Frame
  • 131. Scanned Matching TT-Plots that fall between Min and Max TT-Plots
  • 132. 4 Point Smooth, 4% Error, 31% Threshold 6 Point Smooth, 6% Error, 31% Threshold
  • 133. 4 Point Smooth, 7% Error, 31% Threshold 6 Point Smooth, 7% Error, 31% Threshold
  • 134. 4 Point Smooth, 4% Error, 27% Threshold 6 Point Smooth, 4% Error, 27% Threshold
  • 135. With and Without TT-Plot Automation Before After
  • 136. Conclusion • Shows possibility of automatic defect detection • Does not support hypothesis that depth of defect is directly proportional to how slow it shows up on camera
  • 137. Final Conclusion • Behavior of Panels Observed • Optimal Experimental Settings Tested – Frequency – Coupling Material – Tip Size/Amplitude • Additional Areas of Study – Mode Pattern Vibration – Automation
  • 138. Future Work • Further study on mode pattern • Further study on TT-Plot shape of defects/depth of defects