Using EIT to determine the role of gestational age on
lung volume response during aeration at birth – A
feasibility study
McCall KE, Waldmann AD, Bohm SH, Rajapaksa A, Tan A,
Perkins EJ, Sourial M, Adler A, Grychtol B, Frerichs I, Tingay DG;
16th International Conference on Biomedical Applications of
Electrical Impedance Tomography, Neuchâtel Switzerland, June
2-5, 2015
82 SESSION 9. LUNGS II
Using EIT to determine the role of gestational age on lung volume
response during aeration at birth – A feasibility study
Karen E. McCall
1
, Andreas D. Waldmann
2
, Stephan H. Bohm
2
, Anushi Rajapaksa
1
, Andre Tan
1
,
Elizabeth J Perkins
1
, Magdy Sourial
1
, Andy Adler
3
, Bartlomiej Grychtol
4
, Inéz Frerichs
5
, David G. Tingay
1,6,7
1
Murdoch Children’s Research Institute, Melbourne, Australia, 2
Swisstom AG, Landquart, Switzerland, 3
Carleton University,
Ottawa, Canada, 4
Project Group for Automation in Medical Engineering, Frauenhofer IPA, Mannheim, Germany, 5
Department of
Anaesthesiology and Intensive Care Medicine, University Medical Centre Schleswig-Holstein, Campus Kiel, Kiel, Germany, 6
Royal
Children’s Hospital, Melbourne, Australia, 7
University of Melbourne, Australia
Abstract: Electrical impedance tomography (EIT)
measurements were used to study the first inflation of life
in five lambs of different gestational ages. An exponential
curve has previously been shown to describe the inflation
process. Shorter inflation times were found in term lambs,
whilst longer inflation times were calculated for preterm
and extremely preterm lambs.
1 Introduction
Lung aeration is the first process the newly born lung
needs to achieve at birth if the transition from in-utero life
is to be successful. As the fetal lung is fluid-filled, the first
inflations often need to be longer and of higher pressure to
drive fluid reabsorption. A sustained inflation (SI) has
been advocated as a method of augmenting lung aeration
at birth. Lung behaviour during a SI is dependent on the SI
pressure, duration and intrinsic lung mechanics.
Mathematically, the resistance and compliance of the
respiratory system (lung mechanics) will influence the
time needed to achieve aeration as these parameters
determine the time constant of the lung (τ), a measure of
the time required to reach steady state lung volume (5τ).
Gestational age is an indicator of the developmental status
of the lung at birth, an important factor in the mechanical
state of the lung. Presently, clinicians have no method of
knowing lung volume, resistance or compliance in real-
time at the bedside, hampering meaningful clinical
techniques. Recently, we showed that EIT could be used
to display the volumetric behaviour of the newly born lung
at birth, particularly during the early aeration process. The
aim of this study was to investigate the relationship
between gestational age (GA) and time to reach lung
volume stability (Tstable) during a SI.
2 Methods
The study protocol was approved by the Ethics Committee
of the Murdoch Children’s Research Institute, Melbourne
Australia. A 40 cmH2O SI was applied to lambs in five
GA groups, representing equivalent GA for term, 32-34
week preterm, preterm (27-29 week), extremely preterm
(25 week or less) and pre-viability (20-22 week) GA in
humans. EIT measurements were performed with a
modified EIT Pioneer Set (Swisstom AG, Switzerland). A
special lamb interface was made of an elastic band
carrying 32 contact pads made of gold plated pin arrays,
(Figure 1). A standard electrically non-conductive
ultrasound gel was used to reduce the impedance between
these pins and the skin. Each of the arrays was connected
via a cable to the active SensorBelt. The SI was applied
until 10 s after visual volume stability (Vstable), or for 180 s
maximum. To describe the mechanical status of the
respiratory system we fitted an exponential curve to the
impedance signal using Matlab (MathWorks, USA). The
time constant (τ) was calculated to describe the inflation
time and was estimated whenever the goodness of fit R2
exceeded 0.9. Complete data for 8-10 lambs/gestational
age group will be presented at meeting.
Figure 1: Lamb interface, consisting of an elastic band, 32 gold
plated arrays and SensorBelt wrapped in a tarpaulin
3 Results
Figure 2 depicts five inflation curves, calculated as the
relative impedance change over the EIT images. The
smallest τ value of 1.5 s was estimated for the 139 day old
lamb, and the highest τ of 45.8 s for the 118 day old lamb.
Figure 2: Five representative inflation curves for different
gestational ages. The table shows the estimated τ values.
4 Conclusions and Outlook
The time to achieve aeration was estimated using the time
constant τ. τ and therefore time to reach lung volume
stability (Tstable) is shorter for term then for preterm and
pre-viability lambs. In this feasibility study we estimated
the τ only in 5 different preterm lambs with various
gestational ages. Furthermore, we looked only at the
global inflation curve and not at local ventilation
distribution. In the future we should study more lambs and
analyse the EIT images also on a local basis.
GAin
humans
20-22
week
23-25
week
27-29
week
32-34
week
term
118d 123d 129d 132d 139d
τ [s] 45.8 9.7 21.6 4.1 1.5
R2
0.96 0.91 0.95 0.95 0.97
SWI SSTO M SCIE NT IFIC LIBRARY
Content:Dr.StephanBöhm;Concept&Design:ZweizeitBrandDevelopment
Made in Switzerland
electrical
impedance
tomography
Real-time tomographic images for organ
function monitoring and diagnosis
Contact us!
call: + 41 (0) 81 330 09 72
mail: info@swisstom.com
visit: www.swisstom.com
Swisstom AG
Schulstrasse 1, CH-7302
Landquart, Switzerland
Swisstom AG
Swisstom AG, located in Landquart, Switzerland,
develops and manufactures innovative medical devices.
Our new lung function monitor enables life-saving
treatments for patients in intensive care and during
general anesthesia.
Unlike traditional tomography, Swisstom´s bedside
imaging is based on non-radiating principles: Electrical
Impedance Tomography (EIT). To date, no comparable
devices can show such regional organ function
continuously and in real-time at the patient’s bedside.
Swisstom creates its competitive edge by passionate
leadership in non-invasive tomography with the goal
to improve individual lives and therapies.
© Swisstom AG

More Related Content

Similar to McCall K.E. et al.: Using EIT to determine the role of gestational age on lung volume response during aeration at birth – A feasibility study.

Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...
Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...
Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...AI Publications
 
Linee guida nava nel neonato
Linee guida nava nel neonatoLinee guida nava nel neonato
Linee guida nava nel neonatopeppepag
 
Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...
Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...
Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...InsideScientific
 
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...Hauke Sann
 
Stochastic Model of Vasopressin by the Effect of Age
Stochastic Model of Vasopressin by the Effect of AgeStochastic Model of Vasopressin by the Effect of Age
Stochastic Model of Vasopressin by the Effect of AgeIJERA Editor
 
Identifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - PilbeamIdentifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - PilbeamRiver City Symposium
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteShahid Younas
 
Integrated Textronic System to Control Health of Small Children_Crimson Publ...
 Integrated Textronic System to Control Health of Small Children_Crimson Publ... Integrated Textronic System to Control Health of Small Children_Crimson Publ...
Integrated Textronic System to Control Health of Small Children_Crimson Publ...CrimsonpublishersTTEFT
 
8737 Coclia 84 Glottic Ans Subglottic Stenosis
8737 Coclia 84 Glottic Ans Subglottic Stenosis8737 Coclia 84 Glottic Ans Subglottic Stenosis
8737 Coclia 84 Glottic Ans Subglottic StenosisMedicineAndHealthResearch
 
OM-85 Applicability in routine clinical practice - Professor Susanna Esposito
OM-85 Applicability in routine clinical practice - Professor Susanna EspositoOM-85 Applicability in routine clinical practice - Professor Susanna Esposito
OM-85 Applicability in routine clinical practice - Professor Susanna EspositoWAidid
 
Swine Spine Presentation Patricia 10_28_16
Swine Spine Presentation Patricia 10_28_16Swine Spine Presentation Patricia 10_28_16
Swine Spine Presentation Patricia 10_28_16Patricia Stan
 
Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...
Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...
Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...IOSR Journals
 
SALT Study Published in PEC 2011
SALT Study Published in PEC 2011SALT Study Published in PEC 2011
SALT Study Published in PEC 2011larry_johnson
 
weaning 986.full.pdf
weaning 986.full.pdfweaning 986.full.pdf
weaning 986.full.pdfvignesh79445
 
Comparative study of corneal thickness measures
Comparative study of corneal thickness measuresComparative study of corneal thickness measures
Comparative study of corneal thickness measuresCLINICA REMENTERIA
 
Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...
Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...
Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...semualkaira
 

Similar to McCall K.E. et al.: Using EIT to determine the role of gestational age on lung volume response during aeration at birth – A feasibility study. (20)

Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...
Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...
Ultrasonographic Cervical Length Measurement at 10-14- and 20-24-weeks’ Gesta...
 
Linee guida nava nel neonato
Linee guida nava nel neonatoLinee guida nava nel neonato
Linee guida nava nel neonato
 
Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...
Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...
Case Studies in Home Cage Monitoring: Rodent Behavior, Circadian Biology and ...
 
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
 
Stochastic Model of Vasopressin by the Effect of Age
Stochastic Model of Vasopressin by the Effect of AgeStochastic Model of Vasopressin by the Effect of Age
Stochastic Model of Vasopressin by the Effect of Age
 
Identifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - PilbeamIdentifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - Pilbeam
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT Constitute
 
Integrated Textronic System to Control Health of Small Children_Crimson Publ...
 Integrated Textronic System to Control Health of Small Children_Crimson Publ... Integrated Textronic System to Control Health of Small Children_Crimson Publ...
Integrated Textronic System to Control Health of Small Children_Crimson Publ...
 
8737 Coclia 84 Glottic Ans Subglottic Stenosis
8737 Coclia 84 Glottic Ans Subglottic Stenosis8737 Coclia 84 Glottic Ans Subglottic Stenosis
8737 Coclia 84 Glottic Ans Subglottic Stenosis
 
OM-85 Applicability in routine clinical practice - Professor Susanna Esposito
OM-85 Applicability in routine clinical practice - Professor Susanna EspositoOM-85 Applicability in routine clinical practice - Professor Susanna Esposito
OM-85 Applicability in routine clinical practice - Professor Susanna Esposito
 
-28th ISPSR POSTER
-28th ISPSR POSTER-28th ISPSR POSTER
-28th ISPSR POSTER
 
Swine Spine Presentation Patricia 10_28_16
Swine Spine Presentation Patricia 10_28_16Swine Spine Presentation Patricia 10_28_16
Swine Spine Presentation Patricia 10_28_16
 
Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...
Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...
Assessment of Entrance Skin Dose in routine x-ray examinations of chest, skul...
 
Comparative Evaluation of ENTEROSCREEN-WBTM and Widal test in Suspected Cases...
Comparative Evaluation of ENTEROSCREEN-WBTM and Widal test in Suspected Cases...Comparative Evaluation of ENTEROSCREEN-WBTM and Widal test in Suspected Cases...
Comparative Evaluation of ENTEROSCREEN-WBTM and Widal test in Suspected Cases...
 
SALT Study Published in PEC 2011
SALT Study Published in PEC 2011SALT Study Published in PEC 2011
SALT Study Published in PEC 2011
 
weaning 986.full.pdf
weaning 986.full.pdfweaning 986.full.pdf
weaning 986.full.pdf
 
tomotherapy
tomotherapytomotherapy
tomotherapy
 
tomotherapy
tomotherapytomotherapy
tomotherapy
 
Comparative study of corneal thickness measures
Comparative study of corneal thickness measuresComparative study of corneal thickness measures
Comparative study of corneal thickness measures
 
Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...
Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...
Use of the EIS (Electrical Impedance Segmentography System) in Less-Invasive ...
 

More from Hauke Sann

Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...Hauke Sann
 
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...Hauke Sann
 
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...Hauke Sann
 
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...Hauke Sann
 
Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.Hauke Sann
 
Krammer P. et al.: Electrical impedance tomography Simulator.
Krammer P. et al.: Electrical impedance tomography Simulator.Krammer P. et al.: Electrical impedance tomography Simulator.
Krammer P. et al.: Electrical impedance tomography Simulator.Hauke Sann
 
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Hauke Sann
 
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...Hauke Sann
 
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB) White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB) Hauke Sann
 
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)Hauke Sann
 
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)Hauke Sann
 
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)Hauke Sann
 
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)Hauke Sann
 

More from Hauke Sann (20)

Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
 
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
 
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
 
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
 
Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.
 
Krammer P. et al.: Electrical impedance tomography Simulator.
Krammer P. et al.: Electrical impedance tomography Simulator.Krammer P. et al.: Electrical impedance tomography Simulator.
Krammer P. et al.: Electrical impedance tomography Simulator.
 
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
 
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
 
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB) White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
 
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
 
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
 
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
 
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
 

Recently uploaded

Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012Call Girls Service Gurgaon
 
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...gurkirankumar98700
 
Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...Russian Call Girls Amritsar
 
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...Sheetaleventcompany
 
Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...
Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...
Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...Niamh verma
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknowgragteena
 
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅gragmanisha42
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...Gfnyt
 
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsiindian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana TulsiHigh Profile Call Girls Chandigarh Aarushi
 
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...gragteena
 
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130  Available With RoomVIP Kolkata Call Girl New Town 👉 8250192130  Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Roomdivyansh0kumar0
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Call Girls Noida
 
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaHot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaRussian Call Girls in Ludhiana
 
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★indiancallgirl4rent
 

Recently uploaded (20)

Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
 
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
Russian Call Girls Lucknow ₹7.5k Pick Up & Drop With Cash Payment 8923113531 ...
 
Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎️ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
 
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
 
Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...
Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...
Call Girls Amritsar 💯Call Us 🔝 8725944379 🔝 💃 Independent Escort Service Amri...
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
 
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
 
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
 
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsiindian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
 
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
 
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130  Available With RoomVIP Kolkata Call Girl New Town 👉 8250192130  Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Room
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
 
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaHot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
 
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
 

McCall K.E. et al.: Using EIT to determine the role of gestational age on lung volume response during aeration at birth – A feasibility study.

  • 1. Using EIT to determine the role of gestational age on lung volume response during aeration at birth – A feasibility study McCall KE, Waldmann AD, Bohm SH, Rajapaksa A, Tan A, Perkins EJ, Sourial M, Adler A, Grychtol B, Frerichs I, Tingay DG; 16th International Conference on Biomedical Applications of Electrical Impedance Tomography, Neuchâtel Switzerland, June 2-5, 2015
  • 2. 82 SESSION 9. LUNGS II Using EIT to determine the role of gestational age on lung volume response during aeration at birth – A feasibility study Karen E. McCall 1 , Andreas D. Waldmann 2 , Stephan H. Bohm 2 , Anushi Rajapaksa 1 , Andre Tan 1 , Elizabeth J Perkins 1 , Magdy Sourial 1 , Andy Adler 3 , Bartlomiej Grychtol 4 , Inéz Frerichs 5 , David G. Tingay 1,6,7 1 Murdoch Children’s Research Institute, Melbourne, Australia, 2 Swisstom AG, Landquart, Switzerland, 3 Carleton University, Ottawa, Canada, 4 Project Group for Automation in Medical Engineering, Frauenhofer IPA, Mannheim, Germany, 5 Department of Anaesthesiology and Intensive Care Medicine, University Medical Centre Schleswig-Holstein, Campus Kiel, Kiel, Germany, 6 Royal Children’s Hospital, Melbourne, Australia, 7 University of Melbourne, Australia Abstract: Electrical impedance tomography (EIT) measurements were used to study the first inflation of life in five lambs of different gestational ages. An exponential curve has previously been shown to describe the inflation process. Shorter inflation times were found in term lambs, whilst longer inflation times were calculated for preterm and extremely preterm lambs. 1 Introduction Lung aeration is the first process the newly born lung needs to achieve at birth if the transition from in-utero life is to be successful. As the fetal lung is fluid-filled, the first inflations often need to be longer and of higher pressure to drive fluid reabsorption. A sustained inflation (SI) has been advocated as a method of augmenting lung aeration at birth. Lung behaviour during a SI is dependent on the SI pressure, duration and intrinsic lung mechanics. Mathematically, the resistance and compliance of the respiratory system (lung mechanics) will influence the time needed to achieve aeration as these parameters determine the time constant of the lung (τ), a measure of the time required to reach steady state lung volume (5τ). Gestational age is an indicator of the developmental status of the lung at birth, an important factor in the mechanical state of the lung. Presently, clinicians have no method of knowing lung volume, resistance or compliance in real- time at the bedside, hampering meaningful clinical techniques. Recently, we showed that EIT could be used to display the volumetric behaviour of the newly born lung at birth, particularly during the early aeration process. The aim of this study was to investigate the relationship between gestational age (GA) and time to reach lung volume stability (Tstable) during a SI. 2 Methods The study protocol was approved by the Ethics Committee of the Murdoch Children’s Research Institute, Melbourne Australia. A 40 cmH2O SI was applied to lambs in five GA groups, representing equivalent GA for term, 32-34 week preterm, preterm (27-29 week), extremely preterm (25 week or less) and pre-viability (20-22 week) GA in humans. EIT measurements were performed with a modified EIT Pioneer Set (Swisstom AG, Switzerland). A special lamb interface was made of an elastic band carrying 32 contact pads made of gold plated pin arrays, (Figure 1). A standard electrically non-conductive ultrasound gel was used to reduce the impedance between these pins and the skin. Each of the arrays was connected via a cable to the active SensorBelt. The SI was applied until 10 s after visual volume stability (Vstable), or for 180 s maximum. To describe the mechanical status of the respiratory system we fitted an exponential curve to the impedance signal using Matlab (MathWorks, USA). The time constant (τ) was calculated to describe the inflation time and was estimated whenever the goodness of fit R2 exceeded 0.9. Complete data for 8-10 lambs/gestational age group will be presented at meeting. Figure 1: Lamb interface, consisting of an elastic band, 32 gold plated arrays and SensorBelt wrapped in a tarpaulin 3 Results Figure 2 depicts five inflation curves, calculated as the relative impedance change over the EIT images. The smallest τ value of 1.5 s was estimated for the 139 day old lamb, and the highest τ of 45.8 s for the 118 day old lamb. Figure 2: Five representative inflation curves for different gestational ages. The table shows the estimated τ values. 4 Conclusions and Outlook The time to achieve aeration was estimated using the time constant τ. τ and therefore time to reach lung volume stability (Tstable) is shorter for term then for preterm and pre-viability lambs. In this feasibility study we estimated the τ only in 5 different preterm lambs with various gestational ages. Furthermore, we looked only at the global inflation curve and not at local ventilation distribution. In the future we should study more lambs and analyse the EIT images also on a local basis. GAin humans 20-22 week 23-25 week 27-29 week 32-34 week term 118d 123d 129d 132d 139d τ [s] 45.8 9.7 21.6 4.1 1.5 R2 0.96 0.91 0.95 0.95 0.97
  • 3. SWI SSTO M SCIE NT IFIC LIBRARY Content:Dr.StephanBöhm;Concept&Design:ZweizeitBrandDevelopment Made in Switzerland electrical impedance tomography Real-time tomographic images for organ function monitoring and diagnosis Contact us! call: + 41 (0) 81 330 09 72 mail: info@swisstom.com visit: www.swisstom.com Swisstom AG Schulstrasse 1, CH-7302 Landquart, Switzerland Swisstom AG Swisstom AG, located in Landquart, Switzerland, develops and manufactures innovative medical devices. Our new lung function monitor enables life-saving treatments for patients in intensive care and during general anesthesia. Unlike traditional tomography, Swisstom´s bedside imaging is based on non-radiating principles: Electrical Impedance Tomography (EIT). To date, no comparable devices can show such regional organ function continuously and in real-time at the patient’s bedside. Swisstom creates its competitive edge by passionate leadership in non-invasive tomography with the goal to improve individual lives and therapies. © Swisstom AG