Basics of calibrations &
performance verification
of non analytical instruments
Dr K Krishnan
5.3.1.4 Equipment calibration and
metrological traceability
The laboratory shall have a documented procedure for the calibration of
equipment that directly or indirectly affects examination results. This
procedure includes:
a) taking into account conditions of use and the manufacturer’s
instructions;
b) recording the metrological traceability of the calibration standard and
the traceable calibration of the item of equipment;
c) verifying the required measurement accuracy and the functioning of
the measuring system at defined intervals;
d) recording the calibration status and date of recalibration;
e) ensuring that, where calibration gives rise to a set of correction
factors, the previous calibration factors are correctly updated;
f) safeguards to prevent adjustments or tampering that might
invalidate examination results.
22/02/2014 FQI ISO 15189 IA & QMS certificate
course
• Metrological traceability shall be to a reference material or
reference procedure of the higher metrological order available.
• NOTE Documentation of calibration traceability to a higher order
reference material or reference procedure may be provided by
an examination system manufacturer. Such documentation is
acceptable as long as the manufacturer’s examination system
and calibration procedures are used without modification.
• Where this is not possible or relevant, other means for providing
confidence in the results shall be applied, including but not
limited to the following:
— use of certified reference materials;
— examination or calibration by another procedure;
— mutual consent standards or methods which are clearly
established, specified, characterized and mutually agreed
upon by all parties concerned.
22/02/2014 FQI ISO 15189 IA & QMS certificate
course
Equipments that indirectly affect
results
• Parts of Analyzer which may / may not be integrated
but contribute indirectly to results:
• Pre Analytical Phase
– Centrifuge
– Water bath
– Autoclave
– Pipette
– Storage chambers, Environment
– HIT indices measuring component
• Automation of non – analyzer component
– Engen™
Calibration requirements - general
• Appropriate to the
equipment
• SOP of calibration defined
• Certification a must
• Signed off by the qualified
engineer / agency
• Traceability certificate copy
required for the instrument
used for any calibration
Calibration certificate contents
The calibration certificates issued by external agency must be checked for
following details:
• Name of Certificate issuing agency
• Date of issuance
• Description of the equipment under calibration
• Standards / Calibrators used for the calibration of the reference
• Date of Calibration
• Calibration results
• Acceptance criteria
• Next calibration due date
• Traceability to national/international standards
• Copies of calibration certificate of the standard / calibrator with traceability
to national / international standards (if applicable).
Centrifuge
• Indicators on the instruments
– Speed
– Temperature ?
• Calibration of speed
– Tachometer ( Traceability to reference standard and calibrated)
• Indicator display should match the actual reading with
tolerance limits prescribed by the manufacturer
• Frequency : every PM event / as required
• Centrifuge should be fully loaded
• Speed requirement specific for the tests done
• Speed required for tests defined:
– as per tests
– As per type of tubes used
– Temperature:
• Calibrated thermometer equipment used for calibrating temp
display
• Tolerance limit defined
Source: tech talk BD Diagnostics Vol 6, 2008
Water baths calibration
• Correction Factor = the difference in the NIST traceable
thermometer and the Observed Temperature measured
by the NIST traceable thermometer. This is calculated
and reported during the annual certification of the
thermometers
• The True Temperature measured by the laboratory
thermometer = the Observed Temperature plus the
correction factor for that thermometer
• Equipment used: NIST traceable thermometer
• Frequency: Annual
Source: EPA/OPP MICROBIOLOGY LABORATORY
ESC, Ft. Meade, MD
Standard Operating Procedure for
Calibration of Thermometers
Autoclave calibration
Frequency of controls:
• With chemical Indicator : Every load
• With Biological indicator: Every month
Calibration Procedure:
• Temperature indicator, Pressure gauge and probe
should be calibrated quarterly / on extensive
maintenance
– Temperature gauge: Benzoic acid Grade AR: melts at 121°C:
– Pressure: A standard calibrated gauge attached to a T
connection and pressure compared – Limits: +5 % psi,
– Temperature indicator calibrated using silicon oil bath and std
thermometer at 100° C and 125° C (Limit: + 1°C)
Source: S S Jadiya, School of Pharma Sciences,
RGTU UIT, Bhopal
Calibrating a pipette for actual
volume
VOLUME
Pipette :
Water Temperature:
Density From Table: 0.9982
Technicia
n:
Date :
Balance :
Measure
Actual
Volume Weight Volume calc. Limits
No (µl) (g) (µl) (%)
1 1000 0.988 989.8 -
2 0.0 -
3 0.0 -
4 0.0 -
5 0.0 -
°C Density
18 0.9986
19 0.9984
20 0.9982
21 0.998
22 0.9978
23 0.9975
24 0.9973
25 0.997
26 0.9968
27 0.9965
28 0.9962
Storage chambers:
calibration and validation
Specify range of temperature for storage
The objectives:
• To calibrate temperature display with calibrated
equipment (as for water baths with NIST
thermometers)
• To identify the hot and cold zones in storage chambers,
• To challenge the storage temperature by keeping the
chambers open for fixed time
• To find the maximum time allowable for power shut
downs
Calibrating storage chambers
Three types of validation to be
conducted with the sensors
located at different places:
• The Static Mapping
• The Dynamic mapping
• The Power cut-off and
recovery mapping
Equipment ID Accuracy Calibration
validity
Data Logger
(36 CH)
Make: Chino
+ 0.5 °C 31st Oct
2014
RTD Pt-100
Sensor 4
wire
+ 0.5 °C 24th Nov
2014
Characterization of equipment
used to calibrate storage
Measuring HIT indices
• Manual inspection
– Color index grade I to IV
– Visual grading as per a given chart
– Range of hemolysis, turbidity and icterus to be determined for
each grading using dedicated reagents
• Automated Systems
– Integrated with Analyzer
• Usually done with dedicated reagents for measuring
hemolysis, turbidity and Bilirubin
• Throughput may be affected
– VITROS Microsensor technology is unique method
• No reagents used
• Hence Throughput not affected
VITROS MicroSensor™
MicroSensor™ - How it Works
Reflectometer
Spectrophotometer
Grating &
Photodiode Array
Electronics
A/D, DSP,
S/W
Collect Light from
Reflectometer
Lamp
Shine Light
Through Patient
Sample
Measure Spectrum
of Transmitted
Light
Master
Computer
Software
Communicate
Absorbance
Spectrum
Convert Spectra a
to Semi-
Quantitative
Indice Values
Absorbance Spectrum
0
0.5
1
1.5
2
400 450 500 550 600 650 700 750 800
Wavelength
Absorbance
Bilirubin
Turbidity
Hemoglobin
Absorbance Spectrum is acquired for each sample.
Spectrum is used to make predictions for the three indices.
Controls to be run weekly to cross check
No calibration required as no reagent used
Cost of not calibrating non- analytical
equipments
Automation of pre analyticals
enGen™
•Sample processing – decapping, centrifugation
•Sample transport with intelligent routing
•Rules-based reflex & repeat testing
•Result verification
The only calibration required is for the
centrifuge
Krishnan basics of calibrations and pv non analytical instruments
Krishnan basics of calibrations and pv non analytical instruments

More Related Content

What's hot

Calibration of Instruments
Calibration of InstrumentsCalibration of Instruments
Calibration of Instrumentskaroline Enoch
 
Introduction to validation
Introduction to validation Introduction to validation
Introduction to validation Ankush Sule
 
Instrument qualification
Instrument qualificationInstrument qualification
Instrument qualificationArti Thakkar
 
Calibration of Analytical instrument
Calibration of Analytical instrument Calibration of Analytical instrument
Calibration of Analytical instrument Imdad H. Mukeri
 
calibration of analytical instruments
calibration of analytical instruments calibration of analytical instruments
calibration of analytical instruments Lavanya Pinneboina
 
Calibration and validation ,difference , ph meter and weigh balance calibrati...
Calibration and validation ,difference , ph meter and weigh balance calibrati...Calibration and validation ,difference , ph meter and weigh balance calibrati...
Calibration and validation ,difference , ph meter and weigh balance calibrati...Divy Patel
 
Calibration as a risk management strategy
Calibration as a risk management strategyCalibration as a risk management strategy
Calibration as a risk management strategyAlliance Calibration
 
Control of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgradControl of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgradMedhatEldeeb2
 
PET - Total Quality Management
PET - Total Quality ManagementPET - Total Quality Management
PET - Total Quality Management@Saudi_nmc
 
PET - Calibration of QC Instruments for PET Radiopharmaceuticals Testing
PET - Calibration of QC Instruments for PET Radiopharmaceuticals TestingPET - Calibration of QC Instruments for PET Radiopharmaceuticals Testing
PET - Calibration of QC Instruments for PET Radiopharmaceuticals Testing@Saudi_nmc
 
PET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET RadiopharmaceutialsPET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET Radiopharmaceutials@Saudi_nmc
 
Revalidation presentation
Revalidation presentationRevalidation presentation
Revalidation presentationDenny Walthers
 
Quality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryQuality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryDr. Bikash Kumar Chaudhury
 

What's hot (20)

Overview of Instrument Calibration
Overview of Instrument CalibrationOverview of Instrument Calibration
Overview of Instrument Calibration
 
Calibration of Instruments
Calibration of InstrumentsCalibration of Instruments
Calibration of Instruments
 
Calibration systems
Calibration systemsCalibration systems
Calibration systems
 
Introduction to validation
Introduction to validation Introduction to validation
Introduction to validation
 
Instrument qualification
Instrument qualificationInstrument qualification
Instrument qualification
 
Calibration of Analytical instrument
Calibration of Analytical instrument Calibration of Analytical instrument
Calibration of Analytical instrument
 
CALIBRATION INSTRUMENTS
CALIBRATION INSTRUMENTSCALIBRATION INSTRUMENTS
CALIBRATION INSTRUMENTS
 
calibration of analytical instruments
calibration of analytical instruments calibration of analytical instruments
calibration of analytical instruments
 
Calibration and validation ,difference , ph meter and weigh balance calibrati...
Calibration and validation ,difference , ph meter and weigh balance calibrati...Calibration and validation ,difference , ph meter and weigh balance calibrati...
Calibration and validation ,difference , ph meter and weigh balance calibrati...
 
Calibration as a risk management strategy
Calibration as a risk management strategyCalibration as a risk management strategy
Calibration as a risk management strategy
 
Calibration
CalibrationCalibration
Calibration
 
Presentation for APPON
Presentation for APPONPresentation for APPON
Presentation for APPON
 
QUALIFICATION OF HPTLC,HPTLC
QUALIFICATION  OF HPTLC,HPTLCQUALIFICATION  OF HPTLC,HPTLC
QUALIFICATION OF HPTLC,HPTLC
 
Control of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgradControl of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgrad
 
PET - Total Quality Management
PET - Total Quality ManagementPET - Total Quality Management
PET - Total Quality Management
 
PET - Calibration of QC Instruments for PET Radiopharmaceuticals Testing
PET - Calibration of QC Instruments for PET Radiopharmaceuticals TestingPET - Calibration of QC Instruments for PET Radiopharmaceuticals Testing
PET - Calibration of QC Instruments for PET Radiopharmaceuticals Testing
 
PET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET RadiopharmaceutialsPET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET Radiopharmaceutials
 
Doc1 copy
Doc1   copyDoc1   copy
Doc1 copy
 
Revalidation presentation
Revalidation presentationRevalidation presentation
Revalidation presentation
 
Quality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryQuality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing Laboratory
 

Similar to Krishnan basics of calibrations and pv non analytical instruments

3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptx
3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptx3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptx
3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptxGayuMahe
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemAkshay Trivedi , Maliba Pharmacy College
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instrumentsSanathoiba Singha
 
Module_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.pptModule_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.pptssuserbf5fe71
 
Validation utility system
Validation utility systemValidation utility system
Validation utility systemArpitSuralkar
 
Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)Mohammed Alzhrani
 
Pharmaceutical Validation
Pharmaceutical ValidationPharmaceutical Validation
Pharmaceutical Validationsilambarasan I
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of resultsGH Yeoh
 
validation-1.pptx
validation-1.pptxvalidation-1.pptx
validation-1.pptxSridharA50
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollutionECRD2015
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollutionECRD IN
 
Metrology manager resume
Metrology manager resumeMetrology manager resume
Metrology manager resumeKarthick G
 
Calibration and Qualification
Calibration and QualificationCalibration and Qualification
Calibration and QualificationPriyanka Kandhare
 
Validation utility system
Validation utility systemValidation utility system
Validation utility systemPRANJAY PATIL
 

Similar to Krishnan basics of calibrations and pv non analytical instruments (20)

calibration.pptx
calibration.pptxcalibration.pptx
calibration.pptx
 
3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptx
3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptx3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptx
3.ICH &WHO Guidelines for Calibration And Validation of Equipments.pptx
 
Calibration
Calibration Calibration
Calibration
 
Calibration
CalibrationCalibration
Calibration
 
Validation of gc instrument
Validation of gc instrumentValidation of gc instrument
Validation of gc instrument
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility System
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Module_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.pptModule_4_GCLP_LABORATORY_EQUIPMENT.ppt
Module_4_GCLP_LABORATORY_EQUIPMENT.ppt
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)Laboratory Equipment (from CLSI)
Laboratory Equipment (from CLSI)
 
Pharmaceutical Validation
Pharmaceutical ValidationPharmaceutical Validation
Pharmaceutical Validation
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of results
 
validation-1.pptx
validation-1.pptxvalidation-1.pptx
validation-1.pptx
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollution
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollution
 
Metrology manager resume
Metrology manager resumeMetrology manager resume
Metrology manager resume
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
Calibration and Qualification
Calibration and QualificationCalibration and Qualification
Calibration and Qualification
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Dipak Prajapati CV_16
Dipak Prajapati CV_16Dipak Prajapati CV_16
Dipak Prajapati CV_16
 

Recently uploaded

Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhVip call girls In Chandigarh
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...Gfnyt.com
 
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...Call Girls Noida
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliHigh Profile Call Girls Chandigarh Aarushi
 
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...Vip call girls In Chandigarh
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Memriyagarg453
 
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...High Profile Call Girls Chandigarh Aarushi
 
Dehradun Call Girls Service 7017441440 Real Russian Girls Looking Models
Dehradun Call Girls Service 7017441440 Real Russian Girls Looking ModelsDehradun Call Girls Service 7017441440 Real Russian Girls Looking Models
Dehradun Call Girls Service 7017441440 Real Russian Girls Looking Modelsindiancallgirl4rent
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girls Service Chandigarh Ayushi
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591adityaroy0215
 
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...High Profile Call Girls Chandigarh Aarushi
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxAyush Gupta
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Niamh verma
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...Gfnyt.com
 

Recently uploaded (20)

Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
 
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
 
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
 
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
 
Dehradun Call Girls Service 7017441440 Real Russian Girls Looking Models
Dehradun Call Girls Service 7017441440 Real Russian Girls Looking ModelsDehradun Call Girls Service 7017441440 Real Russian Girls Looking Models
Dehradun Call Girls Service 7017441440 Real Russian Girls Looking Models
 
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service DehradunCall Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
 
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
 
#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi
#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi
#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi ❤️🍑 9907093804 👄🫦 Independent Escort ...
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptx
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
 

Krishnan basics of calibrations and pv non analytical instruments

  • 1. Basics of calibrations & performance verification of non analytical instruments Dr K Krishnan
  • 2. 5.3.1.4 Equipment calibration and metrological traceability The laboratory shall have a documented procedure for the calibration of equipment that directly or indirectly affects examination results. This procedure includes: a) taking into account conditions of use and the manufacturer’s instructions; b) recording the metrological traceability of the calibration standard and the traceable calibration of the item of equipment; c) verifying the required measurement accuracy and the functioning of the measuring system at defined intervals; d) recording the calibration status and date of recalibration; e) ensuring that, where calibration gives rise to a set of correction factors, the previous calibration factors are correctly updated; f) safeguards to prevent adjustments or tampering that might invalidate examination results. 22/02/2014 FQI ISO 15189 IA & QMS certificate course
  • 3. • Metrological traceability shall be to a reference material or reference procedure of the higher metrological order available. • NOTE Documentation of calibration traceability to a higher order reference material or reference procedure may be provided by an examination system manufacturer. Such documentation is acceptable as long as the manufacturer’s examination system and calibration procedures are used without modification. • Where this is not possible or relevant, other means for providing confidence in the results shall be applied, including but not limited to the following: — use of certified reference materials; — examination or calibration by another procedure; — mutual consent standards or methods which are clearly established, specified, characterized and mutually agreed upon by all parties concerned. 22/02/2014 FQI ISO 15189 IA & QMS certificate course
  • 4. Equipments that indirectly affect results • Parts of Analyzer which may / may not be integrated but contribute indirectly to results: • Pre Analytical Phase – Centrifuge – Water bath – Autoclave – Pipette – Storage chambers, Environment – HIT indices measuring component • Automation of non – analyzer component – Engen™
  • 5. Calibration requirements - general • Appropriate to the equipment • SOP of calibration defined • Certification a must • Signed off by the qualified engineer / agency • Traceability certificate copy required for the instrument used for any calibration
  • 6. Calibration certificate contents The calibration certificates issued by external agency must be checked for following details: • Name of Certificate issuing agency • Date of issuance • Description of the equipment under calibration • Standards / Calibrators used for the calibration of the reference • Date of Calibration • Calibration results • Acceptance criteria • Next calibration due date • Traceability to national/international standards • Copies of calibration certificate of the standard / calibrator with traceability to national / international standards (if applicable).
  • 7. Centrifuge • Indicators on the instruments – Speed – Temperature ? • Calibration of speed – Tachometer ( Traceability to reference standard and calibrated) • Indicator display should match the actual reading with tolerance limits prescribed by the manufacturer • Frequency : every PM event / as required • Centrifuge should be fully loaded • Speed requirement specific for the tests done • Speed required for tests defined: – as per tests – As per type of tubes used – Temperature: • Calibrated thermometer equipment used for calibrating temp display • Tolerance limit defined
  • 8. Source: tech talk BD Diagnostics Vol 6, 2008
  • 9. Water baths calibration • Correction Factor = the difference in the NIST traceable thermometer and the Observed Temperature measured by the NIST traceable thermometer. This is calculated and reported during the annual certification of the thermometers • The True Temperature measured by the laboratory thermometer = the Observed Temperature plus the correction factor for that thermometer • Equipment used: NIST traceable thermometer • Frequency: Annual Source: EPA/OPP MICROBIOLOGY LABORATORY ESC, Ft. Meade, MD Standard Operating Procedure for Calibration of Thermometers
  • 10. Autoclave calibration Frequency of controls: • With chemical Indicator : Every load • With Biological indicator: Every month Calibration Procedure: • Temperature indicator, Pressure gauge and probe should be calibrated quarterly / on extensive maintenance – Temperature gauge: Benzoic acid Grade AR: melts at 121°C: – Pressure: A standard calibrated gauge attached to a T connection and pressure compared – Limits: +5 % psi, – Temperature indicator calibrated using silicon oil bath and std thermometer at 100° C and 125° C (Limit: + 1°C) Source: S S Jadiya, School of Pharma Sciences, RGTU UIT, Bhopal
  • 11. Calibrating a pipette for actual volume VOLUME Pipette : Water Temperature: Density From Table: 0.9982 Technicia n: Date : Balance : Measure Actual Volume Weight Volume calc. Limits No (µl) (g) (µl) (%) 1 1000 0.988 989.8 - 2 0.0 - 3 0.0 - 4 0.0 - 5 0.0 - °C Density 18 0.9986 19 0.9984 20 0.9982 21 0.998 22 0.9978 23 0.9975 24 0.9973 25 0.997 26 0.9968 27 0.9965 28 0.9962
  • 12. Storage chambers: calibration and validation Specify range of temperature for storage The objectives: • To calibrate temperature display with calibrated equipment (as for water baths with NIST thermometers) • To identify the hot and cold zones in storage chambers, • To challenge the storage temperature by keeping the chambers open for fixed time • To find the maximum time allowable for power shut downs
  • 13. Calibrating storage chambers Three types of validation to be conducted with the sensors located at different places: • The Static Mapping • The Dynamic mapping • The Power cut-off and recovery mapping Equipment ID Accuracy Calibration validity Data Logger (36 CH) Make: Chino + 0.5 °C 31st Oct 2014 RTD Pt-100 Sensor 4 wire + 0.5 °C 24th Nov 2014 Characterization of equipment used to calibrate storage
  • 14. Measuring HIT indices • Manual inspection – Color index grade I to IV – Visual grading as per a given chart – Range of hemolysis, turbidity and icterus to be determined for each grading using dedicated reagents • Automated Systems – Integrated with Analyzer • Usually done with dedicated reagents for measuring hemolysis, turbidity and Bilirubin • Throughput may be affected – VITROS Microsensor technology is unique method • No reagents used • Hence Throughput not affected
  • 16. MicroSensor™ - How it Works Reflectometer Spectrophotometer Grating & Photodiode Array Electronics A/D, DSP, S/W Collect Light from Reflectometer Lamp Shine Light Through Patient Sample Measure Spectrum of Transmitted Light Master Computer Software Communicate Absorbance Spectrum Convert Spectra a to Semi- Quantitative Indice Values
  • 17. Absorbance Spectrum 0 0.5 1 1.5 2 400 450 500 550 600 650 700 750 800 Wavelength Absorbance Bilirubin Turbidity Hemoglobin Absorbance Spectrum is acquired for each sample. Spectrum is used to make predictions for the three indices. Controls to be run weekly to cross check No calibration required as no reagent used
  • 18.
  • 19. Cost of not calibrating non- analytical equipments
  • 20. Automation of pre analyticals enGen™ •Sample processing – decapping, centrifugation •Sample transport with intelligent routing •Rules-based reflex & repeat testing •Result verification The only calibration required is for the centrifuge