Measuring leakage currents
in electromedical devices in
accordance with IEC EN 62353
White Paper
Introduction
The measurement of leakage currents in electro-
medical devices and systems (hereinafter EM),
in accordance with the IEC EN 62353:2010-10
Standard “Electromedical devices - Periodic
checks and tests after repairs on electromedical
devices,” (hereinafter the Standard), provides for
the use of three methods: direct, alternative and
differential. The measurements to be performed
are then classified into leakage current in the
device and leakage currents in the parts applied,
as defined in the Standard.
Where possible, the direct method is preferred,
since it is more similar to the measurement ap-
proach in the product standard IEC 60601-1-XX.
We will indicate below the criteria for selecting
the measurement method according to the mains
supply characteristics and the installation condi-
tions of the EM device or system.
One of the most severe errors committed in the
execution of electrical safety checks is to neglect
the following fundamental aspects:
•	 Failure to remove ground connections results in
measurements generally equal to zero for the
measured currents.
•	 A power supply from the isolation transformer
reduces the currents to negligible values be-
cause the voltages on the two supply polarities
are equal, and with opposite signs, compared to
the housing connected to ground.
A particular case to take into consideration is the
existence, in some hospitals and in some parts of
the LV network, of a distribution system with only
three active distributed conductors and phase-to-
phase voltage of 230 V. (The neutral is not dis-
tributed and the power supply voltage of the EM
devices or systems is 230 V phase/phase.)
It should be noted, finally, that the same
standard states if the power supply voltage of the
device is U0 ≠ 230 V, the values measured with
the different methods must be multiplied by 230/
U0 to be more correctly compared with the limits
indicated in the Standard.
Therefore, the variables in the leakage cur-
rent measurement system are: type of system, EM
installation and measurement method.
Type of mains supply
The mains supply types, in order of importance
for the purposes of periodic checking, are:
•	 TN-S with four active distributed conductors
and voltage 400/230 V
•	 TT with four active distributed conductors and
voltage 400/230 V
•	 IT-M 230 V
•	 TN-S with three active distributed conductors
and phase-to-phase voltage of 230 V
•	 TT with three active distributed conductors and
phase-to-phase voltage of 220 V
The TN-S system with four active distributed con-
ductors 400/230 V (Fig. 1) is generally present in
all large public and private health care facilities.
The TT system with four active distributed con-
ductors 400/230 V (Fig. 2) is generally present in
small public and private health care facilities that
require less than approx. 100 kW.
Vincenzo Ventimiglia, engineer, SLT srl
Gianpetro Pellizzer, self-employed engineer
2	 Fluke Biomedical	 Measuring leakage currents in electromedical devices in accordance with IEC EN 62353
The system of type IT-M (Fig. 3), i.e. with an isola-
tion transformer, voltage 230 V and isolation detec-
tor is required in all Group II medical rooms.
In some hospitals, the distribution is of the type
-TN-S with three active distributed conductors and
phase-to-phase voltage of 230 V (Fig. 4).
Some parts of the LV system are still managed
with the TT system with three active distributed
conductors and phase-to-phase voltage of 230 V
(Fig. 5).
L1
L2
L3
N
PE
EM
400 V
400 V
400 V
200 V
L1
L2
L3
N
PE PE
EM
400 V
400 V
400 V
200 V
L1
L2
L3
N
PE
TR 1:1 EM
400 V
400 V
400 V
200 V
200V
L1
L2
L3
PE PE
EM
230 V
133 V
230 V
230 V
133 V
L1
L2
L3
PE
EM
230 V
133 V
230 V
230 V
133 V
Figure 1: Electrical diagram for -TN-S type system with four
active distributed conductors and voltages of 400/230 V.
Figure 2: Electrical diagram for -TT type system with four
active distributed conductors and voltages of 400/230 V.
Figure 3: Electrical diagram for IT-M system.
Figure 5: Electrical diagram for -TT type system with three active distributed
conductors and phase-to-phase voltages of 220 V.
Figure 4: Electrical diagram for -TN-S type system with three active distributed
conductors and phase-to-phase voltages of 230 V.
3	 Fluke Biomedical	 Measuring leakage currents in electromedical devices in accordance with IEC EN 62353
Type of installation and power supply
The connection of the EM devices and systems to
their power supply can be:
•	 Movable single-phase/three-phase equipment
(with the possibility of being completely iso-
lated from the ground during the measurement)
with connection to the power supply through a
plug/socket connection
•	 Movable single-phase/three-phase equipment
with permanent connection (without plug/
socket connection to the power supply)
•	 Fixed single-phase/three-phase equipment
(which cannot be completely isolated from the
ground during the measurement) with connec-
tion to the power supply through a plug/socket
connection
•	 Fixed single-phase/three-phase equipment
with permanent connection (without a plug/
socket connection to the power supply)
•	 Internal electrical power source.
Measurement methods
All three leakage current measurement methods
required by the Standard are generally possible
with the measuring instruments available in the
market, but only for EM devices and EM systems
with a single-phase power supply. With existing
instruments, it is possible to measure both the
equipment leakage current of the devices and the
applied part leakage currents.
Direct method
Using the direct method, the device under test is
powered and the leakage currents are measured
from accessible parts and applied parts with
interruption of the protective ground. The power
supply is then inverted between the two poles
and the highest value is recorded.
Alternative method
With the alternative method, the two ELM power
supply poles are short-circuited to create the
same voltage between them and either the parts
are connected or disconnected from the protection
circuit. In this situation, isolations are strained
at a double voltage (or better with the same volt-
age in both polarities) compared to what they are
subjected to in the case of a device that is nor-
mally powered.
Differential method
This leakage current measurement method for the
device involves the use of a toroid traversed by
all of the active conductors of the power supply
circuit of the EM device or EM system. The toroid
gives a signal proportional to the differential cur-
rent flowing through it.
EM devices with internal power source
The leakage current in the device is not appli-
cable to devices with an internal electrical power
source. However, it is necessary to correctly
measure the applied part leakage current if the
applied part is floating. This will be discussed
later.
The Standard, and all of the electrical safety
analyzers’ user manuals, demonstrate the elec-
trical diagrams of the various measurement
methods.
According to the IEC EN 62353 standard, if the
system is built and maintained in compliance
with the requirements of IEC 60346-7-710 (Italian
CEI 64-8), there is no need to check the equip-
ment’s leakage current if the device is perma-
nently installed.
However, the measurement of the leakage cur-
rent of the EM device or EM system, especially if
powered with a three-phase supply, may be nec-
essary to check its compatibility with the differ-
ential protection upstream. This can be done with
a differential clamp-on ammeter, with frequency
characteristics similar to the MD (measuring
device), on all active conductors that supply the
device (see Figure 6).
Figure 6: Electrical diagram for measuring the leakage current in
a three-phase EM device or system.
L1
AP
L2
L3
N
PE
M
4	 Fluke Biomedical	 Measuring leakage currents in electromedical devices in accordance with IEC EN 62353
Selection criteria for the
measurement method
The first factor to consider is the type of system.
Many electrical safety testers, including the Fluke
Biomedical ESA612, ESA620 and ESA615, perform
a measurement of the voltages phase 1-phase
2, phase 1-ground and phase 2-ground, while
also warning the technician to pay attention to
unconventional situations. Conventional is con-
sidered to be phase-to-neutral voltage close to
230 V, phase-to-phase close to 230 V, and ground
potential close to zero.
Each technician must check these voltages, and
if an IT-M type system is identified, it is neces-
sary to give it close attention. The presence of the
isolation transformer has always been a problem
for the correct execution of the electrical safety
measurements, also performed with procedures
based on the IEC EN 60601-1 standard. The rec-
ommended solutions are to use a device power-
ing the system through the “radiology socket”,
(dedicated to power on equipment with high con-
sumption), which is not generally supplied by the
IT system. Or, supply the device via an extension
from a socket outside of the room, again not pow-
ered through the insulation transformer.
With the IEC EN 62353 standard, this is no
longer necessary if using the alternative method,
which is also independent of any grounding of
the device through additional connections other
than PE. The only limitation on the application of
the alternative method is the presence of an elec-
tronic switch. In this case, the execution of the
measurements is not possible by supplying the
electrical safety tester in these conditions.
If the power supply is of the types TN-S or TT
400/230 V with four active distributed conduc-
tors, it is advised to proceed with the direct
method if the device is isolated from the ground.
In all other situations, it is possible to use the
alternative method (only if the electronic switch is
absent), or the differential method.
If the system is of the types TN-S 230 V phase/
phase or TT 220 V phase/phase with three active
distributed conductors, you can continue to per-
form the measurements described in the direct
method. However, you must take into account that
the values to be compared with the limits set in
the Standard are the measured values multiplied
by √3 because the power supply voltages of the
device referenced to ground are reduced by a fac-
tor of √3.
Again, the direct method is only possible if the
device under test is isolated from the ground. If
the device is a fixed installation, it is possible
to use the two other methods: the alternative
method (only if the electronic switch is absent) or
differential method. In these two cases, the value
of the current will be identical to that measured
with the TN-S or TT system, without any effect of
the √3 factor.
The following diagrams may be used to make the
correct selection of the measurement method
Electronic
switch
YES
NO
No measurement is possible
with this power supply
Alternative method, even if the
device is in a fixed installation
IT-M system (simplified: IT system)
System of types TN-S or TT (phase-to-neutral system)
TN-S type system with phase-to-phase voltages
of 230 V or TT type with phase-to-phase voltage of
220 V (phase-to-phase system)
Electronic
switch
YES
NOYES
NO
Differential
Method
Direct
method
Alternative method
Device with
a fixed installation
Electronic
switch
YES
NOYES
NO
Differential
Method
Direct method
(measurements
reduced by a
factor of √3)
Alternative method
Device with
a fixed installation
5	 Fluke Biomedical	 Measuring leakage currents in electromedical devices in accordance with IEC EN 62353
In the case of measurements on three-phase
equipment with a plug/socket combination, all of
the previously mentioned considerations apply. It
is possible to perform the measurement with an
appropriate adapter connected to the electrical
safety tester, which inserts the MD in series with
the protective ground conductor using the direct
method. Additionally the measurement can be
performed using a differential clamp-on ammeter
that is connected to the active conductors using
the differential method.
The alternative method is not applicable to
devices with a three-phase power supply.
Measurement of the leakage currents in the
applied parts
The IEC EN 62353 standard also states require-
ments for the measurement of applied parts
leakage currents performed by the application of
mains voltage to the patient connections short-
circuited “by function”.
For the measurement of current in applied
parts, both the power supply circuit and any
ground connections not removed have no effect
on the result. This is because the electrical safety
tester always applies a voltage of 220 V between
the applied parts and the housing, independently
of the choice of the alternative or direct meth-
ods, not depending on whether the device has a
single-phase or a three-phase power supply, or
if it is isolated from ground or in a fixed instal-
lation. Note that even if a device is permanently
installed and has floating applied parts, such as
the detection electrodes of the ECG waveform in
an angiography system, this test must always be
performed without powering up the device under
test via the electrical safety tester.
This test is the only measurement to be per-
formed on devices with an internal power supply.
The following table shows the measurements
of equipment leakage currents performed on
an operating table using the Fluke Biomedical
ESA615 tester powered by the three types of
mains supply discussed above. The results con-
firm the effect of the power supply type on the
values of the leakage current.
Measurement from system Ground-Neutral
Test results ESA615 direct method Test results ESA615 alternative
method
Test results ESA615 differential method
Test Value High value Result Value High value Result Value High value Low value Result
Protective earth resistance 0.079 ohm 0.3 P 0.080 ohm 0.3 P 0.079 ohm 0.3 - P
Mains voltage P P P
Live to neutral 236.3 V - P 236.0 V - P 235.8 V - - P
Ground to neutral 237.2 V - P 237.2 V - P 237.1 V - - P
Ground to live 1.8 V - P 1.7 V - P 1.9 V - - P
Current consumption 0.3 A - P 0.3 A - P 0.3 A - - P
Direct device dispersion P P P
Open ground 23.1 uA 500 P 48.5 uAAC 1000 P 22 uA 500 - P
Open ground-reverse
polarity
22.0 uA 500 P 21 uA 500 - P
Measurement from IT System
Test results ESA615 direct method Test results ESA615 alternative
method
Test results ESA615 differential method
Test Value High value Result Value High value Result Value High Value Low value Result
Protective earth resistance 0.079 ohm 0.3 P 0.079 Ohm 0.3 P 0.079 Ohm 0.3 - P
Mains voltage P P P
Live to Neutral 239.5 V - P 239.5 V - P 239.4 V - - P
Ground to Neutral 122.7 V - P 122.7 V - P 123.0 V - - P
Ground to Live 116.9 V - P 116.9 V - P 117.0 V - - P
Current consumption 0.3 A - P 0.3 A - P 0.3 A - - P
Direct device dispersion P P P
Open ground 2.1 uA 500 P 48.7 uAAC 1000 P 0 uA 500 - P
Open ground-reverse
polarity
1.0 uA 500 P 0 uA 500 - P
6	 Fluke Biomedical	 Measuring leakage currents in electromedical devices in accordance with IEC EN 62353
Fluke Biomedical.
Better products. More choices. One company.
Fluke Biomedical
6045 Cochran Road
Cleveland, OH 44139-3303 U.S.A.
For more information, contact us at:
(800) 850-4608 or Fax (440) 349-2307
Email: sales@flukebiomedical.com
Web access: www.flukebiomedical.com
©2013 Fluke Biomedical. Specifications subject
to change without notice. Printed in U.S.A.
10/2013 6001325A_EN
Modification of this document is not permitted
without written permission from Fluke Corporation.
About Fluke Biomedical ESA615
Electrical Safety Analyzer
The ESA615 performs all primary electrical safety
tests, ECG simulation, point-to-point voltage, leak-
age and resistance tests to the latest electrical
safety standards. The ESA615 has automation
inside that lets you test in minutes to the electrical
safety standard of your choice. Reduce test time,
eliminate errors and standardize testing with the
latest innovation in electrical safety analyzers.
To learn more about the ESA615 or any other
electrical safety analyzer mentioned in this white
paper, click here or visit flukebiomedical.com.
Measurement from system Ground-Ground
Test results direct method Test results ESA615 alternative
method
Test results ESA615 differential method
Test Value High value Result Value High value Result Value High value Low value Result
Protective earth resistance 0.079 ohm 0.3 P 0.079 ohm 0.3 P 0.079 ohm 0.3 - P
Mains Voltage P P P
Live to Neutral 238.5 V - P 238.2 V - P 239.4 V - - P
Ground to Neutral 137.7 V - P 137.6 V - P 123.0 V - - P
Ground to Live 137.4 V - P 137.3 V - P 117.0 V - - P
Current consumption 0.3 A - P 0.3 A - P 0.3 A - - P
Direct device dispersion P P P
Open ground 13.3 uA 500 P 48.7 uAAC 1000 P 21 uA 500 - P
Open ground-reverse
polarity
12.9 uA 500 P 22 uA 500 - P
About the author, Vincenzo Ventimiglia
Vincenzo Ventimiglia is an electrical engineer with
a strong background in biomedical specialization. A
renowned professional in the biomedical business since
1991, Ventimiglia studied at the Polytechnic University
of Milan and has since developed a business of bio-
medical technician quality assurance training. He has
been a member of the Italian Electrical Committee for
over twenty years and is currently a member of the IEC
SC62A Working Group 14.

More Related Content

What's hot

Harmonics mitigation solutions
Harmonics mitigation solutionsHarmonics mitigation solutions
Harmonics mitigation solutionsFahad Shaikh
 
Electrical measurements - Sultan LeMarc
Electrical measurements - Sultan LeMarcElectrical measurements - Sultan LeMarc
Electrical measurements - Sultan LeMarcslemarc
 
Identification and minimization of Harmonics
Identification and minimization of HarmonicsIdentification and minimization of Harmonics
Identification and minimization of HarmonicsMayur Hiwale
 
Electrical measurement & measuring instruments [emmi (nee-302) -unit-2]
Electrical measurement & measuring instruments [emmi  (nee-302) -unit-2]Electrical measurement & measuring instruments [emmi  (nee-302) -unit-2]
Electrical measurement & measuring instruments [emmi (nee-302) -unit-2]Md Irshad Ahmad
 
Product: Power Factor & Harmonics: StacoSine: MGE filter presentation
Product: Power Factor & Harmonics: StacoSine: MGE filter presentationProduct: Power Factor & Harmonics: StacoSine: MGE filter presentation
Product: Power Factor & Harmonics: StacoSine: MGE filter presentationStaco Energy
 
Electrical measuring instruments
Electrical measuring instrumentsElectrical measuring instruments
Electrical measuring instrumentsMinka Grdesic
 
A Synchronizing Devicefor Power Electronic Converters
A Synchronizing Devicefor Power Electronic ConvertersA Synchronizing Devicefor Power Electronic Converters
A Synchronizing Devicefor Power Electronic ConvertersIJERA Editor
 
Microcontroller based multifunction_relay
Microcontroller based multifunction_relayMicrocontroller based multifunction_relay
Microcontroller based multifunction_relayRajeev Kumar
 
Electrical Troubleshooting
Electrical TroubleshootingElectrical Troubleshooting
Electrical TroubleshootingNabeel Hassan
 
Power Quality Measurement Devices & Monotoring
Power Quality Measurement Devices & MonotoringPower Quality Measurement Devices & Monotoring
Power Quality Measurement Devices & MonotoringParth Patel
 
Factors to be considered while selecting CT
Factors to be considered while selecting CTFactors to be considered while selecting CT
Factors to be considered while selecting CTParth Patel
 
harmonic distortion ppt
harmonic distortion pptharmonic distortion ppt
harmonic distortion pptAditi Tiwari
 
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]Md Irshad Ahmad
 
Introduction to measuring instruments (ALIV - Bangladesh)
Introduction to measuring instruments (ALIV - Bangladesh)Introduction to measuring instruments (ALIV - Bangladesh)
Introduction to measuring instruments (ALIV - Bangladesh)Md Abu Jauad Khan Aliv
 
Introduction to electrical measurements
Introduction to electrical measurementsIntroduction to electrical measurements
Introduction to electrical measurementsparneetchowdhary1
 

What's hot (18)

MAJOE PROJECT PPT
MAJOE PROJECT PPTMAJOE PROJECT PPT
MAJOE PROJECT PPT
 
Harmonics mitigation solutions
Harmonics mitigation solutionsHarmonics mitigation solutions
Harmonics mitigation solutions
 
Electrical measurements - Sultan LeMarc
Electrical measurements - Sultan LeMarcElectrical measurements - Sultan LeMarc
Electrical measurements - Sultan LeMarc
 
Identification and minimization of Harmonics
Identification and minimization of HarmonicsIdentification and minimization of Harmonics
Identification and minimization of Harmonics
 
Electrical measurement & measuring instruments [emmi (nee-302) -unit-2]
Electrical measurement & measuring instruments [emmi  (nee-302) -unit-2]Electrical measurement & measuring instruments [emmi  (nee-302) -unit-2]
Electrical measurement & measuring instruments [emmi (nee-302) -unit-2]
 
Product: Power Factor & Harmonics: StacoSine: MGE filter presentation
Product: Power Factor & Harmonics: StacoSine: MGE filter presentationProduct: Power Factor & Harmonics: StacoSine: MGE filter presentation
Product: Power Factor & Harmonics: StacoSine: MGE filter presentation
 
Electrical measuring instruments
Electrical measuring instrumentsElectrical measuring instruments
Electrical measuring instruments
 
A Synchronizing Devicefor Power Electronic Converters
A Synchronizing Devicefor Power Electronic ConvertersA Synchronizing Devicefor Power Electronic Converters
A Synchronizing Devicefor Power Electronic Converters
 
Microcontroller based multifunction_relay
Microcontroller based multifunction_relayMicrocontroller based multifunction_relay
Microcontroller based multifunction_relay
 
Electrical Troubleshooting
Electrical TroubleshootingElectrical Troubleshooting
Electrical Troubleshooting
 
Power Quality Measurement Devices & Monotoring
Power Quality Measurement Devices & MonotoringPower Quality Measurement Devices & Monotoring
Power Quality Measurement Devices & Monotoring
 
Measurements and Instrumentation - Electrical and electronics instruments
Measurements and Instrumentation - Electrical and electronics instrumentsMeasurements and Instrumentation - Electrical and electronics instruments
Measurements and Instrumentation - Electrical and electronics instruments
 
Factors to be considered while selecting CT
Factors to be considered while selecting CTFactors to be considered while selecting CT
Factors to be considered while selecting CT
 
harmonic distortion ppt
harmonic distortion pptharmonic distortion ppt
harmonic distortion ppt
 
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
 
Ect2
Ect2Ect2
Ect2
 
Introduction to measuring instruments (ALIV - Bangladesh)
Introduction to measuring instruments (ALIV - Bangladesh)Introduction to measuring instruments (ALIV - Bangladesh)
Introduction to measuring instruments (ALIV - Bangladesh)
 
Introduction to electrical measurements
Introduction to electrical measurementsIntroduction to electrical measurements
Introduction to electrical measurements
 

Viewers also liked

Future of Social Interaction
Future of Social InteractionFuture of Social Interaction
Future of Social InteractionJa-Nae Duane
 
acı ölöcer
acı ölöceracı ölöcer
acı ölöcerbkrdgny
 
Once Upon a Time um inglês. - Graziella, Laísa, Katharine.
Once Upon a Time um inglês. -  Graziella, Laísa, Katharine.Once Upon a Time um inglês. -  Graziella, Laísa, Katharine.
Once Upon a Time um inglês. - Graziella, Laísa, Katharine.Graziella Marino
 
Building institutional research capacity for integrated approaches: An exampl...
Building institutional research capacity for integrated approaches: An exampl...Building institutional research capacity for integrated approaches: An exampl...
Building institutional research capacity for integrated approaches: An exampl...ILRI
 
Perwali 21 2006 tentang jam kerja
Perwali 21   2006 tentang jam kerjaPerwali 21   2006 tentang jam kerja
Perwali 21 2006 tentang jam kerjaSDN_Perak_Barat
 
Aula06 – sistemas de proteção de dispositivos
Aula06 – sistemas de proteção de dispositivosAula06 – sistemas de proteção de dispositivos
Aula06 – sistemas de proteção de dispositivosCarlos Veiga
 
La nube en informática
La nube en informáticaLa nube en informática
La nube en informáticaAngelicaFacio
 
Presentacion medellin city
Presentacion medellin cityPresentacion medellin city
Presentacion medellin citymile7304
 
La contaminación ambiental en juliaca y política ambiental
La contaminación ambiental en juliaca y política ambientalLa contaminación ambiental en juliaca y política ambiental
La contaminación ambiental en juliaca y política ambientalCristóbal Jove
 
Makalah kelompok JJ Rousseau
Makalah kelompok JJ RousseauMakalah kelompok JJ Rousseau
Makalah kelompok JJ RousseauBilqis Putri
 
Moneycation may 2015 newsletter; volume #3, issue #11
Moneycation may 2015 newsletter; volume #3, issue #11Moneycation may 2015 newsletter; volume #3, issue #11
Moneycation may 2015 newsletter; volume #3, issue #11A.W. Berry
 
Dinâmicas e Entrevistas em processos seletivos
Dinâmicas e Entrevistas em processos seletivosDinâmicas e Entrevistas em processos seletivos
Dinâmicas e Entrevistas em processos seletivosRômulo Andrade
 

Viewers also liked (20)

Future of Social Interaction
Future of Social InteractionFuture of Social Interaction
Future of Social Interaction
 
acı ölöcer
acı ölöceracı ölöcer
acı ölöcer
 
Evaluation Question 4
Evaluation Question 4Evaluation Question 4
Evaluation Question 4
 
Once Upon a Time um inglês. - Graziella, Laísa, Katharine.
Once Upon a Time um inglês. -  Graziella, Laísa, Katharine.Once Upon a Time um inglês. -  Graziella, Laísa, Katharine.
Once Upon a Time um inglês. - Graziella, Laísa, Katharine.
 
Building institutional research capacity for integrated approaches: An exampl...
Building institutional research capacity for integrated approaches: An exampl...Building institutional research capacity for integrated approaches: An exampl...
Building institutional research capacity for integrated approaches: An exampl...
 
Perwali 21 2006 tentang jam kerja
Perwali 21   2006 tentang jam kerjaPerwali 21   2006 tentang jam kerja
Perwali 21 2006 tentang jam kerja
 
Entrega prueba
Entrega pruebaEntrega prueba
Entrega prueba
 
Aula06 – sistemas de proteção de dispositivos
Aula06 – sistemas de proteção de dispositivosAula06 – sistemas de proteção de dispositivos
Aula06 – sistemas de proteção de dispositivos
 
La nube en informática
La nube en informáticaLa nube en informática
La nube en informática
 
Evaluation Question 3
Evaluation Question 3Evaluation Question 3
Evaluation Question 3
 
Evaluation Question 2
Evaluation Question 2Evaluation Question 2
Evaluation Question 2
 
Presentacion medellin city
Presentacion medellin cityPresentacion medellin city
Presentacion medellin city
 
La contaminación ambiental en juliaca y política ambiental
La contaminación ambiental en juliaca y política ambientalLa contaminación ambiental en juliaca y política ambiental
La contaminación ambiental en juliaca y política ambiental
 
Rip doc
Rip docRip doc
Rip doc
 
Plataformas online
Plataformas onlinePlataformas online
Plataformas online
 
inter certificate
inter certificateinter certificate
inter certificate
 
Elaboración de purines
Elaboración de purinesElaboración de purines
Elaboración de purines
 
Makalah kelompok JJ Rousseau
Makalah kelompok JJ RousseauMakalah kelompok JJ Rousseau
Makalah kelompok JJ Rousseau
 
Moneycation may 2015 newsletter; volume #3, issue #11
Moneycation may 2015 newsletter; volume #3, issue #11Moneycation may 2015 newsletter; volume #3, issue #11
Moneycation may 2015 newsletter; volume #3, issue #11
 
Dinâmicas e Entrevistas em processos seletivos
Dinâmicas e Entrevistas em processos seletivosDinâmicas e Entrevistas em processos seletivos
Dinâmicas e Entrevistas em processos seletivos
 

Similar to 6001325 0000 eng_a_w

283481961-POWER-SYSTEM-FAULTS-ppt.ppt
283481961-POWER-SYSTEM-FAULTS-ppt.ppt283481961-POWER-SYSTEM-FAULTS-ppt.ppt
283481961-POWER-SYSTEM-FAULTS-ppt.pptAshaLatha456489
 
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer NIHON DENKEI SINGAPORE
 
Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...Alexander Decker
 
overvoltage in distribution network
overvoltage in distribution networkovervoltage in distribution network
overvoltage in distribution networkmegala venkatesan
 
Electrical instuments.pptx
Electrical instuments.pptxElectrical instuments.pptx
Electrical instuments.pptxRohith sai
 
Mitsubishi energy measuring unit eco monitorplus_dienhathe.vn
Mitsubishi energy measuring unit eco monitorplus_dienhathe.vnMitsubishi energy measuring unit eco monitorplus_dienhathe.vn
Mitsubishi energy measuring unit eco monitorplus_dienhathe.vnDien Ha The
 
Mitsubishi energy measuring unit eco monitorplus
Mitsubishi energy measuring unit eco monitorplusMitsubishi energy measuring unit eco monitorplus
Mitsubishi energy measuring unit eco monitorplusDien Ha The
 
POWER SYSTEM ANALYSIS_254ppt.pdf
POWER SYSTEM ANALYSIS_254ppt.pdfPOWER SYSTEM ANALYSIS_254ppt.pdf
POWER SYSTEM ANALYSIS_254ppt.pdfMillionAlemayehu2
 
Analysis and mitigation of unbalance due to load in distribution system by us...
Analysis and mitigation of unbalance due to load in distribution system by us...Analysis and mitigation of unbalance due to load in distribution system by us...
Analysis and mitigation of unbalance due to load in distribution system by us...eSAT Journals
 
Analysis and mitigation of unbalance due to load in
Analysis and mitigation of unbalance due to load inAnalysis and mitigation of unbalance due to load in
Analysis and mitigation of unbalance due to load ineSAT Publishing House
 
AC Distribution System - Generation Distribution and Transmission
AC Distribution System - Generation Distribution and TransmissionAC Distribution System - Generation Distribution and Transmission
AC Distribution System - Generation Distribution and TransmissionGowtham Cr
 
Health monitoring of transformer using plc
Health monitoring of transformer using plcHealth monitoring of transformer using plc
Health monitoring of transformer using plcRohith sai
 
IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...
IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...
IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...IRJET Journal
 
UNSYMMETRICAL FAULTS IN POWER SYSTEM
UNSYMMETRICAL FAULTS IN POWER SYSTEMUNSYMMETRICAL FAULTS IN POWER SYSTEM
UNSYMMETRICAL FAULTS IN POWER SYSTEMerramansaini1947
 
Powergrid Corporation India Limited industrial training
Powergrid Corporation India Limited industrial trainingPowergrid Corporation India Limited industrial training
Powergrid Corporation India Limited industrial trainingAkriti Gupta
 
powergrid presentation
powergrid presentationpowergrid presentation
powergrid presentationAkriti Gupta
 

Similar to 6001325 0000 eng_a_w (20)

283481961-POWER-SYSTEM-FAULTS-ppt.ppt
283481961-POWER-SYSTEM-FAULTS-ppt.ppt283481961-POWER-SYSTEM-FAULTS-ppt.ppt
283481961-POWER-SYSTEM-FAULTS-ppt.ppt
 
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
 
Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...
 
overvoltage in distribution network
overvoltage in distribution networkovervoltage in distribution network
overvoltage in distribution network
 
Electrical instuments.pptx
Electrical instuments.pptxElectrical instuments.pptx
Electrical instuments.pptx
 
Mitsubishi energy measuring unit eco monitorplus_dienhathe.vn
Mitsubishi energy measuring unit eco monitorplus_dienhathe.vnMitsubishi energy measuring unit eco monitorplus_dienhathe.vn
Mitsubishi energy measuring unit eco monitorplus_dienhathe.vn
 
Mitsubishi energy measuring unit eco monitorplus
Mitsubishi energy measuring unit eco monitorplusMitsubishi energy measuring unit eco monitorplus
Mitsubishi energy measuring unit eco monitorplus
 
Lecture (11.11.22).pptx
Lecture (11.11.22).pptxLecture (11.11.22).pptx
Lecture (11.11.22).pptx
 
POWER SYSTEM ANALYSIS_254ppt.pdf
POWER SYSTEM ANALYSIS_254ppt.pdfPOWER SYSTEM ANALYSIS_254ppt.pdf
POWER SYSTEM ANALYSIS_254ppt.pdf
 
Analysis and mitigation of unbalance due to load in distribution system by us...
Analysis and mitigation of unbalance due to load in distribution system by us...Analysis and mitigation of unbalance due to load in distribution system by us...
Analysis and mitigation of unbalance due to load in distribution system by us...
 
Analysis and mitigation of unbalance due to load in
Analysis and mitigation of unbalance due to load inAnalysis and mitigation of unbalance due to load in
Analysis and mitigation of unbalance due to load in
 
AC Distribution System - Generation Distribution and Transmission
AC Distribution System - Generation Distribution and TransmissionAC Distribution System - Generation Distribution and Transmission
AC Distribution System - Generation Distribution and Transmission
 
Ac distribution system
Ac distribution systemAc distribution system
Ac distribution system
 
Health monitoring of transformer using plc
Health monitoring of transformer using plcHealth monitoring of transformer using plc
Health monitoring of transformer using plc
 
IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...
IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...
IRJET- Fault Location using Discrete Wavelet Transform for Smart Grid Distrib...
 
EDS Unit 6.pptx
EDS Unit 6.pptxEDS Unit 6.pptx
EDS Unit 6.pptx
 
UNSYMMETRICAL FAULTS IN POWER SYSTEM
UNSYMMETRICAL FAULTS IN POWER SYSTEMUNSYMMETRICAL FAULTS IN POWER SYSTEM
UNSYMMETRICAL FAULTS IN POWER SYSTEM
 
Ct4201633637
Ct4201633637Ct4201633637
Ct4201633637
 
Powergrid Corporation India Limited industrial training
Powergrid Corporation India Limited industrial trainingPowergrid Corporation India Limited industrial training
Powergrid Corporation India Limited industrial training
 
powergrid presentation
powergrid presentationpowergrid presentation
powergrid presentation
 

More from Cao Xuân Trình

More from Cao Xuân Trình (20)

Signals And Systems.pdf
Signals And Systems.pdfSignals And Systems.pdf
Signals And Systems.pdf
 
Robocon 2007_Theme and Rules
Robocon 2007_Theme and RulesRobocon 2007_Theme and Rules
Robocon 2007_Theme and Rules
 
Aami 2015expo expanded
Aami 2015expo expandedAami 2015expo expanded
Aami 2015expo expanded
 
20130225092745
2013022509274520130225092745
20130225092745
 
6004312 0000 eng_a_w
6004312 0000 eng_a_w6004312 0000 eng_a_w
6004312 0000 eng_a_w
 
6001325 0000 eng_a_w
6001325 0000 eng_a_w6001325 0000 eng_a_w
6001325 0000 eng_a_w
 
2700
27002700
2700
 
2015 calendar
2015 calendar2015 calendar
2015 calendar
 
55 w 60126-0-tech-brief_keys-to-coherent-acq-success
55 w 60126-0-tech-brief_keys-to-coherent-acq-success55 w 60126-0-tech-brief_keys-to-coherent-acq-success
55 w 60126-0-tech-brief_keys-to-coherent-acq-success
 
High speed-interface-standards-e guide
High speed-interface-standards-e guideHigh speed-interface-standards-e guide
High speed-interface-standards-e guide
 
Rigel vital-signs-booklet-uk
Rigel vital-signs-booklet-ukRigel vital-signs-booklet-uk
Rigel vital-signs-booklet-uk
 
St3300655 lw
St3300655 lwSt3300655 lw
St3300655 lw
 
Seagate cheetah-15k.5
Seagate cheetah-15k.5Seagate cheetah-15k.5
Seagate cheetah-15k.5
 
Po cheetah 15k_5
Po cheetah 15k_5Po cheetah 15k_5
Po cheetah 15k_5
 
100384784c
100384784c100384784c
100384784c
 
Chuan thiet yeu quoc te ve chat luong ...
Chuan thiet yeu quoc te ve chat luong ...Chuan thiet yeu quoc te ve chat luong ...
Chuan thiet yeu quoc te ve chat luong ...
 
55 w 60126-0-tech-brief_keys-to-coherent-acq-success
55 w 60126-0-tech-brief_keys-to-coherent-acq-success55 w 60126-0-tech-brief_keys-to-coherent-acq-success
55 w 60126-0-tech-brief_keys-to-coherent-acq-success
 
Aami 2015 registration
Aami 2015 registrationAami 2015 registration
Aami 2015 registration
 
Da hoa b08
Da hoa b08Da hoa b08
Da hoa b08
 
Da hoa a-cd08
Da hoa a-cd08Da hoa a-cd08
Da hoa a-cd08
 

Recently uploaded

Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...
Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...
Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...Pooja Nehwal
 
Vip Noida Escorts 9873940964 Greater Noida Escorts Service
Vip Noida Escorts 9873940964 Greater Noida Escorts ServiceVip Noida Escorts 9873940964 Greater Noida Escorts Service
Vip Noida Escorts 9873940964 Greater Noida Escorts Serviceankitnayak356677
 
定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一
定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一
定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一ss ss
 
如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一
如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一
如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一ga6c6bdl
 
Real Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCR
Real Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCRReal Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCR
Real Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCRdollysharma2066
 
VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...
VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...
VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...Suhani Kapoor
 
萨斯喀彻温大学毕业证学位证成绩单-购买流程
萨斯喀彻温大学毕业证学位证成绩单-购买流程萨斯喀彻温大学毕业证学位证成绩单-购买流程
萨斯喀彻温大学毕业证学位证成绩单-购买流程1k98h0e1
 
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Call Girls Service Kolkata Aishwarya 🤌 8250192130 🚀 Vip Call Girls Kolkata
Call Girls Service Kolkata Aishwarya 🤌  8250192130 🚀 Vip Call Girls KolkataCall Girls Service Kolkata Aishwarya 🤌  8250192130 🚀 Vip Call Girls Kolkata
Call Girls Service Kolkata Aishwarya 🤌 8250192130 🚀 Vip Call Girls Kolkataanamikaraghav4
 
如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一
如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一
如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一ga6c6bdl
 
NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...
NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...
NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...Amil baba
 
Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...
Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...
Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...srsj9000
 
Alambagh Call Girl 9548273370 , Call Girls Service Lucknow
Alambagh Call Girl 9548273370 , Call Girls Service LucknowAlambagh Call Girl 9548273370 , Call Girls Service Lucknow
Alambagh Call Girl 9548273370 , Call Girls Service Lucknowmakika9823
 
Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...
Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...
Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...nagunakhan
 
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,Pooja Nehwal
 
Gaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service GayaGaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service Gayasrsj9000
 
如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查
如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查
如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查awo24iot
 
(办理学位证)多伦多大学毕业证成绩单原版一比一
(办理学位证)多伦多大学毕业证成绩单原版一比一(办理学位证)多伦多大学毕业证成绩单原版一比一
(办理学位证)多伦多大学毕业证成绩单原版一比一C SSS
 

Recently uploaded (20)

Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...
Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...
Kalyan callg Girls, { 07738631006 } || Call Girl In Kalyan Women Seeking Men ...
 
Vip Noida Escorts 9873940964 Greater Noida Escorts Service
Vip Noida Escorts 9873940964 Greater Noida Escorts ServiceVip Noida Escorts 9873940964 Greater Noida Escorts Service
Vip Noida Escorts 9873940964 Greater Noida Escorts Service
 
定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一
定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一
定制(RHUL学位证)伦敦大学皇家霍洛威学院毕业证成绩单原版一比一
 
如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一
如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一
如何办理萨省大学毕业证(UofS毕业证)成绩单留信学历认证原版一比一
 
Real Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCR
Real Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCRReal Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCR
Real Sure (Call Girl) in I.G.I. Airport 8377087607 Hot Call Girls In Delhi NCR
 
CIVIL ENGINEERING
CIVIL ENGINEERINGCIVIL ENGINEERING
CIVIL ENGINEERING
 
VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...
VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...
VIP Call Girls Kavuri Hills ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With ...
 
萨斯喀彻温大学毕业证学位证成绩单-购买流程
萨斯喀彻温大学毕业证学位证成绩单-购买流程萨斯喀彻温大学毕业证学位证成绩单-购买流程
萨斯喀彻温大学毕业证学位证成绩单-购买流程
 
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
 
Call Girls Service Kolkata Aishwarya 🤌 8250192130 🚀 Vip Call Girls Kolkata
Call Girls Service Kolkata Aishwarya 🤌  8250192130 🚀 Vip Call Girls KolkataCall Girls Service Kolkata Aishwarya 🤌  8250192130 🚀 Vip Call Girls Kolkata
Call Girls Service Kolkata Aishwarya 🤌 8250192130 🚀 Vip Call Girls Kolkata
 
如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一
如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一
如何办理伦敦大学伯贝克学院毕业证(BBK毕业证) 成绩单留信学历认证原版一比一
 
NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...
NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...
NO1 Qualified Best Black Magic Specialist Near Me Spiritual Healer Powerful L...
 
Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...
Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...
Hifi Defence Colony Call Girls Service WhatsApp -> 9999965857 Available 24x7 ...
 
Alambagh Call Girl 9548273370 , Call Girls Service Lucknow
Alambagh Call Girl 9548273370 , Call Girls Service LucknowAlambagh Call Girl 9548273370 , Call Girls Service Lucknow
Alambagh Call Girl 9548273370 , Call Girls Service Lucknow
 
Low rate Call girls in Delhi Justdial | 9953330565
Low rate Call girls in Delhi Justdial | 9953330565Low rate Call girls in Delhi Justdial | 9953330565
Low rate Call girls in Delhi Justdial | 9953330565
 
Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...
Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...
Russian Call Girls In South Delhi Delhi 9711199012 💋✔💕😘 Independent Escorts D...
 
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
 
Gaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service GayaGaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service Gaya
 
如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查
如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查
如何办理(Adelaide毕业证)阿德莱德大学毕业证成绩单Adelaide学历认证真实可查
 
(办理学位证)多伦多大学毕业证成绩单原版一比一
(办理学位证)多伦多大学毕业证成绩单原版一比一(办理学位证)多伦多大学毕业证成绩单原版一比一
(办理学位证)多伦多大学毕业证成绩单原版一比一
 

6001325 0000 eng_a_w

  • 1. Measuring leakage currents in electromedical devices in accordance with IEC EN 62353 White Paper Introduction The measurement of leakage currents in electro- medical devices and systems (hereinafter EM), in accordance with the IEC EN 62353:2010-10 Standard “Electromedical devices - Periodic checks and tests after repairs on electromedical devices,” (hereinafter the Standard), provides for the use of three methods: direct, alternative and differential. The measurements to be performed are then classified into leakage current in the device and leakage currents in the parts applied, as defined in the Standard. Where possible, the direct method is preferred, since it is more similar to the measurement ap- proach in the product standard IEC 60601-1-XX. We will indicate below the criteria for selecting the measurement method according to the mains supply characteristics and the installation condi- tions of the EM device or system. One of the most severe errors committed in the execution of electrical safety checks is to neglect the following fundamental aspects: • Failure to remove ground connections results in measurements generally equal to zero for the measured currents. • A power supply from the isolation transformer reduces the currents to negligible values be- cause the voltages on the two supply polarities are equal, and with opposite signs, compared to the housing connected to ground. A particular case to take into consideration is the existence, in some hospitals and in some parts of the LV network, of a distribution system with only three active distributed conductors and phase-to- phase voltage of 230 V. (The neutral is not dis- tributed and the power supply voltage of the EM devices or systems is 230 V phase/phase.) It should be noted, finally, that the same standard states if the power supply voltage of the device is U0 ≠ 230 V, the values measured with the different methods must be multiplied by 230/ U0 to be more correctly compared with the limits indicated in the Standard. Therefore, the variables in the leakage cur- rent measurement system are: type of system, EM installation and measurement method. Type of mains supply The mains supply types, in order of importance for the purposes of periodic checking, are: • TN-S with four active distributed conductors and voltage 400/230 V • TT with four active distributed conductors and voltage 400/230 V • IT-M 230 V • TN-S with three active distributed conductors and phase-to-phase voltage of 230 V • TT with three active distributed conductors and phase-to-phase voltage of 220 V The TN-S system with four active distributed con- ductors 400/230 V (Fig. 1) is generally present in all large public and private health care facilities. The TT system with four active distributed con- ductors 400/230 V (Fig. 2) is generally present in small public and private health care facilities that require less than approx. 100 kW. Vincenzo Ventimiglia, engineer, SLT srl Gianpetro Pellizzer, self-employed engineer
  • 2. 2 Fluke Biomedical Measuring leakage currents in electromedical devices in accordance with IEC EN 62353 The system of type IT-M (Fig. 3), i.e. with an isola- tion transformer, voltage 230 V and isolation detec- tor is required in all Group II medical rooms. In some hospitals, the distribution is of the type -TN-S with three active distributed conductors and phase-to-phase voltage of 230 V (Fig. 4). Some parts of the LV system are still managed with the TT system with three active distributed conductors and phase-to-phase voltage of 230 V (Fig. 5). L1 L2 L3 N PE EM 400 V 400 V 400 V 200 V L1 L2 L3 N PE PE EM 400 V 400 V 400 V 200 V L1 L2 L3 N PE TR 1:1 EM 400 V 400 V 400 V 200 V 200V L1 L2 L3 PE PE EM 230 V 133 V 230 V 230 V 133 V L1 L2 L3 PE EM 230 V 133 V 230 V 230 V 133 V Figure 1: Electrical diagram for -TN-S type system with four active distributed conductors and voltages of 400/230 V. Figure 2: Electrical diagram for -TT type system with four active distributed conductors and voltages of 400/230 V. Figure 3: Electrical diagram for IT-M system. Figure 5: Electrical diagram for -TT type system with three active distributed conductors and phase-to-phase voltages of 220 V. Figure 4: Electrical diagram for -TN-S type system with three active distributed conductors and phase-to-phase voltages of 230 V.
  • 3. 3 Fluke Biomedical Measuring leakage currents in electromedical devices in accordance with IEC EN 62353 Type of installation and power supply The connection of the EM devices and systems to their power supply can be: • Movable single-phase/three-phase equipment (with the possibility of being completely iso- lated from the ground during the measurement) with connection to the power supply through a plug/socket connection • Movable single-phase/three-phase equipment with permanent connection (without plug/ socket connection to the power supply) • Fixed single-phase/three-phase equipment (which cannot be completely isolated from the ground during the measurement) with connec- tion to the power supply through a plug/socket connection • Fixed single-phase/three-phase equipment with permanent connection (without a plug/ socket connection to the power supply) • Internal electrical power source. Measurement methods All three leakage current measurement methods required by the Standard are generally possible with the measuring instruments available in the market, but only for EM devices and EM systems with a single-phase power supply. With existing instruments, it is possible to measure both the equipment leakage current of the devices and the applied part leakage currents. Direct method Using the direct method, the device under test is powered and the leakage currents are measured from accessible parts and applied parts with interruption of the protective ground. The power supply is then inverted between the two poles and the highest value is recorded. Alternative method With the alternative method, the two ELM power supply poles are short-circuited to create the same voltage between them and either the parts are connected or disconnected from the protection circuit. In this situation, isolations are strained at a double voltage (or better with the same volt- age in both polarities) compared to what they are subjected to in the case of a device that is nor- mally powered. Differential method This leakage current measurement method for the device involves the use of a toroid traversed by all of the active conductors of the power supply circuit of the EM device or EM system. The toroid gives a signal proportional to the differential cur- rent flowing through it. EM devices with internal power source The leakage current in the device is not appli- cable to devices with an internal electrical power source. However, it is necessary to correctly measure the applied part leakage current if the applied part is floating. This will be discussed later. The Standard, and all of the electrical safety analyzers’ user manuals, demonstrate the elec- trical diagrams of the various measurement methods. According to the IEC EN 62353 standard, if the system is built and maintained in compliance with the requirements of IEC 60346-7-710 (Italian CEI 64-8), there is no need to check the equip- ment’s leakage current if the device is perma- nently installed. However, the measurement of the leakage cur- rent of the EM device or EM system, especially if powered with a three-phase supply, may be nec- essary to check its compatibility with the differ- ential protection upstream. This can be done with a differential clamp-on ammeter, with frequency characteristics similar to the MD (measuring device), on all active conductors that supply the device (see Figure 6). Figure 6: Electrical diagram for measuring the leakage current in a three-phase EM device or system. L1 AP L2 L3 N PE M
  • 4. 4 Fluke Biomedical Measuring leakage currents in electromedical devices in accordance with IEC EN 62353 Selection criteria for the measurement method The first factor to consider is the type of system. Many electrical safety testers, including the Fluke Biomedical ESA612, ESA620 and ESA615, perform a measurement of the voltages phase 1-phase 2, phase 1-ground and phase 2-ground, while also warning the technician to pay attention to unconventional situations. Conventional is con- sidered to be phase-to-neutral voltage close to 230 V, phase-to-phase close to 230 V, and ground potential close to zero. Each technician must check these voltages, and if an IT-M type system is identified, it is neces- sary to give it close attention. The presence of the isolation transformer has always been a problem for the correct execution of the electrical safety measurements, also performed with procedures based on the IEC EN 60601-1 standard. The rec- ommended solutions are to use a device power- ing the system through the “radiology socket”, (dedicated to power on equipment with high con- sumption), which is not generally supplied by the IT system. Or, supply the device via an extension from a socket outside of the room, again not pow- ered through the insulation transformer. With the IEC EN 62353 standard, this is no longer necessary if using the alternative method, which is also independent of any grounding of the device through additional connections other than PE. The only limitation on the application of the alternative method is the presence of an elec- tronic switch. In this case, the execution of the measurements is not possible by supplying the electrical safety tester in these conditions. If the power supply is of the types TN-S or TT 400/230 V with four active distributed conduc- tors, it is advised to proceed with the direct method if the device is isolated from the ground. In all other situations, it is possible to use the alternative method (only if the electronic switch is absent), or the differential method. If the system is of the types TN-S 230 V phase/ phase or TT 220 V phase/phase with three active distributed conductors, you can continue to per- form the measurements described in the direct method. However, you must take into account that the values to be compared with the limits set in the Standard are the measured values multiplied by √3 because the power supply voltages of the device referenced to ground are reduced by a fac- tor of √3. Again, the direct method is only possible if the device under test is isolated from the ground. If the device is a fixed installation, it is possible to use the two other methods: the alternative method (only if the electronic switch is absent) or differential method. In these two cases, the value of the current will be identical to that measured with the TN-S or TT system, without any effect of the √3 factor. The following diagrams may be used to make the correct selection of the measurement method Electronic switch YES NO No measurement is possible with this power supply Alternative method, even if the device is in a fixed installation IT-M system (simplified: IT system) System of types TN-S or TT (phase-to-neutral system) TN-S type system with phase-to-phase voltages of 230 V or TT type with phase-to-phase voltage of 220 V (phase-to-phase system) Electronic switch YES NOYES NO Differential Method Direct method Alternative method Device with a fixed installation Electronic switch YES NOYES NO Differential Method Direct method (measurements reduced by a factor of √3) Alternative method Device with a fixed installation
  • 5. 5 Fluke Biomedical Measuring leakage currents in electromedical devices in accordance with IEC EN 62353 In the case of measurements on three-phase equipment with a plug/socket combination, all of the previously mentioned considerations apply. It is possible to perform the measurement with an appropriate adapter connected to the electrical safety tester, which inserts the MD in series with the protective ground conductor using the direct method. Additionally the measurement can be performed using a differential clamp-on ammeter that is connected to the active conductors using the differential method. The alternative method is not applicable to devices with a three-phase power supply. Measurement of the leakage currents in the applied parts The IEC EN 62353 standard also states require- ments for the measurement of applied parts leakage currents performed by the application of mains voltage to the patient connections short- circuited “by function”. For the measurement of current in applied parts, both the power supply circuit and any ground connections not removed have no effect on the result. This is because the electrical safety tester always applies a voltage of 220 V between the applied parts and the housing, independently of the choice of the alternative or direct meth- ods, not depending on whether the device has a single-phase or a three-phase power supply, or if it is isolated from ground or in a fixed instal- lation. Note that even if a device is permanently installed and has floating applied parts, such as the detection electrodes of the ECG waveform in an angiography system, this test must always be performed without powering up the device under test via the electrical safety tester. This test is the only measurement to be per- formed on devices with an internal power supply. The following table shows the measurements of equipment leakage currents performed on an operating table using the Fluke Biomedical ESA615 tester powered by the three types of mains supply discussed above. The results con- firm the effect of the power supply type on the values of the leakage current. Measurement from system Ground-Neutral Test results ESA615 direct method Test results ESA615 alternative method Test results ESA615 differential method Test Value High value Result Value High value Result Value High value Low value Result Protective earth resistance 0.079 ohm 0.3 P 0.080 ohm 0.3 P 0.079 ohm 0.3 - P Mains voltage P P P Live to neutral 236.3 V - P 236.0 V - P 235.8 V - - P Ground to neutral 237.2 V - P 237.2 V - P 237.1 V - - P Ground to live 1.8 V - P 1.7 V - P 1.9 V - - P Current consumption 0.3 A - P 0.3 A - P 0.3 A - - P Direct device dispersion P P P Open ground 23.1 uA 500 P 48.5 uAAC 1000 P 22 uA 500 - P Open ground-reverse polarity 22.0 uA 500 P 21 uA 500 - P Measurement from IT System Test results ESA615 direct method Test results ESA615 alternative method Test results ESA615 differential method Test Value High value Result Value High value Result Value High Value Low value Result Protective earth resistance 0.079 ohm 0.3 P 0.079 Ohm 0.3 P 0.079 Ohm 0.3 - P Mains voltage P P P Live to Neutral 239.5 V - P 239.5 V - P 239.4 V - - P Ground to Neutral 122.7 V - P 122.7 V - P 123.0 V - - P Ground to Live 116.9 V - P 116.9 V - P 117.0 V - - P Current consumption 0.3 A - P 0.3 A - P 0.3 A - - P Direct device dispersion P P P Open ground 2.1 uA 500 P 48.7 uAAC 1000 P 0 uA 500 - P Open ground-reverse polarity 1.0 uA 500 P 0 uA 500 - P
  • 6. 6 Fluke Biomedical Measuring leakage currents in electromedical devices in accordance with IEC EN 62353 Fluke Biomedical. Better products. More choices. One company. Fluke Biomedical 6045 Cochran Road Cleveland, OH 44139-3303 U.S.A. For more information, contact us at: (800) 850-4608 or Fax (440) 349-2307 Email: sales@flukebiomedical.com Web access: www.flukebiomedical.com ©2013 Fluke Biomedical. Specifications subject to change without notice. Printed in U.S.A. 10/2013 6001325A_EN Modification of this document is not permitted without written permission from Fluke Corporation. About Fluke Biomedical ESA615 Electrical Safety Analyzer The ESA615 performs all primary electrical safety tests, ECG simulation, point-to-point voltage, leak- age and resistance tests to the latest electrical safety standards. The ESA615 has automation inside that lets you test in minutes to the electrical safety standard of your choice. Reduce test time, eliminate errors and standardize testing with the latest innovation in electrical safety analyzers. To learn more about the ESA615 or any other electrical safety analyzer mentioned in this white paper, click here or visit flukebiomedical.com. Measurement from system Ground-Ground Test results direct method Test results ESA615 alternative method Test results ESA615 differential method Test Value High value Result Value High value Result Value High value Low value Result Protective earth resistance 0.079 ohm 0.3 P 0.079 ohm 0.3 P 0.079 ohm 0.3 - P Mains Voltage P P P Live to Neutral 238.5 V - P 238.2 V - P 239.4 V - - P Ground to Neutral 137.7 V - P 137.6 V - P 123.0 V - - P Ground to Live 137.4 V - P 137.3 V - P 117.0 V - - P Current consumption 0.3 A - P 0.3 A - P 0.3 A - - P Direct device dispersion P P P Open ground 13.3 uA 500 P 48.7 uAAC 1000 P 21 uA 500 - P Open ground-reverse polarity 12.9 uA 500 P 22 uA 500 - P About the author, Vincenzo Ventimiglia Vincenzo Ventimiglia is an electrical engineer with a strong background in biomedical specialization. A renowned professional in the biomedical business since 1991, Ventimiglia studied at the Polytechnic University of Milan and has since developed a business of bio- medical technician quality assurance training. He has been a member of the Italian Electrical Committee for over twenty years and is currently a member of the IEC SC62A Working Group 14.