Integrated ΔB0/Rx coil array for
improved spinal cord imaging at 3T
ISMRM 26th Annual Meeting
June 20, 2018, Paris, France
Topfer R1, Foias A1, Lopez Rios N1,2, Chauffray A1,3, Germain G1, Arango N4,
Wald LL5,6, Stockmann JP5,6, Cohen-Adad J1,7
1. NeuroPoly Lab, Institute of Biomedical Engineering, Polytechnique Montreal, Montreal, QC, Canada, 2. Centro de Biofisica
Medica, Universidad de Oriente, Santiago de Cuba, Cuba, 3. École polytechnique fédérale de Lausanne, Lausanne, Switzerland, 4.
Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA, United States, 5. A. A.
Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 6. Harvard Medical
School, Boston, MA, United States, 7. Functional Neuroimaging Unit, CRIUGM, Université de Montréal, Montreal, QC, Canada
Introduction
Methods
Results
Discussion
Problems
2
[4] Vannesjo S, Neuroimage 167:191-202
T2*-w @7T: ghosting from ΔB0
dynamic ΔB0
(e.g. respiration &
swallowing [3,4])
[1] Stroman P, Neuroimage 84:1070-81
[2] Saritas E , Chapter 2.3 of Quantitative MRI of the Spinal Cord
[3] Verma T, Magn Res Med 72:1629-36
distortions
[2]
static ΔB0
(e.g vertebrae, airways)
inspired
expired
phase encoding
(AP)
[3]
10 mm
“The greatest challenge for acquiring MR images in the spinal cord is [ΔB0] the
inhomogeneous magnetic field in this region.” [1]
Introduction
Methods
Results
Discussion
Solutions
How to address these issues?
Design
principle
Rx benefit Shim benefit
Positioning
coils near
body
↑sensitivity,
↑SNR
↑efficiency (Hz/A)
Using many
coils
enhances
parallel
imaging
enables higher
order ΔB0
correction [3]
[1] Stockmann JP, Magn Res Med 75:441-51
[2] Truong TK, Neuroimage 103, 235–40
[3] Juchem C, J Mag Reson 236: 95-104 3
Parallel imaging + high-order dynamic shimming
Solving the “real-estate problem” [1,2]
Introduction
Methods
Results
Discussion
4
Conceptualization
[1] Germain G, ISMRM 2016. Abstract #3490.
Optimal coil arrangement from simulations
3. ΔB0 + Rx SNR performance assessed for various coil arrangements
1. Sagittal ΔB0 maps acquired (n=5)
+75 Hz
-75 Hz
𝛾𝛥B0Avg.ΔB0
Inspired Expired
}
}
2. Auto-segmentation of cervical spinal cord ROI using the Spinal Cord Toolbox 𝛾𝛥B0
+30 Hz
-30 Hz
single coil
optimal
Introduction
Methods
Results
Discussion
5
Construction
RF only
5.5 cm
[1] Stockmann JP, Magn Res Med 75:441-51
RF + shim
Introduction
Methods
Results
Discussion
6
Construction
DC twisted pairs routed
away from RF outputs
Preamps atop plastic supports,
away from coils
TODO: periodic cable traps +
chokes along DC cables [1]
[1] Stockmann JP, Magn Res Med 75:441-51
Introduction
Methods
Results
Discussion
+75 Hz/A-75 Hz/A
ch1
ch5
ch2 ch3
ch6 ch7
ch4
ch8
ΔB0 reference maps used in shim
optimization
7
Coil performance
Coupling:
min = 2 %
max = 48 %
avg = 18%
0 %
100%
50 %
Noise correlation (Tx = 0V)
Max shift per ch.
avg = ±25 Hz/A
over SC
Rx measurements
Introduction
Methods
Results
Discussion
Computer in control room
shim update
via USB
+fibre optic
• images transferred over local network
• runs MATLAB + Spinal Cord Toolbox
8
Experimental setup
[1] Arango N, ISMRM 2016. Abstract #1157.
RF common
mode chokes
Power supply
Equipment room
Current driver [1]
+ heatsink
Microcontroller
+DAC
pressure sensor +
microcontroller
3T Prisma
MRI room
Pressure(t)
respiratory
bellows
shim DC
[2] Topfer R, Magn Res Med 80:935-46
Introduction
Methods
Results
Discussion
9
In vivo results
Expired GRE mag
+100 Hz-100 Hz 𝛾𝛥B0
Original
shim
VOI
|Pred. − Actual|
25 Hz0 Hz 𝛾𝛥B0
Inspired
σ 𝛥B0
=54 Hz
σ 𝛥B0
=31 Hz
σ 𝛥B0
=32 Hz
σ 𝛥B0
=53 Hz
Predicted
Shimmed
σ 𝛥B0
=29 Hz
Actual
σ 𝛥B0
=29 Hz
46% improvement σ 𝛥B0
40% improvement σ 𝛥B0
Introduction
Methods
Results
Discussion
10
In vivo results
14 mm-14 mm
Original
Shift EPI
InspiredExpired
mean |shift| ± st.dev
2.7±2.1mm
mean |shift| ± st.dev
2.5±2.1mm
Shimmed
Shift EPI
10 cm
mean |shift| ± st.dev
1.4±1.3mm
mean |shift| ± st.dev
1.8±1.2mm
max |shift| = 13 mm max |shift| = 8 mm
Introduction
Methods
Results
Discussion
11
Future directions
1. Real-time shimming
2. Making the most of the low inductance coils
—> Potential for slicewise-dynamic shimming
+100 Hz-100 Hz 𝛾𝛥B0
Inspired Expired
Pressure(t)
Inputs
static
shim
respiration-
correction
shim(t)
+20 Hz
-20 Hz
𝛾𝛥B0
static
ΔB0
respiration-
induced
ΔB0(t)
+100 Hz
-100 Hz
𝛾𝛥B0
Static Dynamic
Outputs
real-time
shimmed
real-time
shimmed
+
=
+
=
3. Clinical app.: T2* for quantitative spinal-cord
assessment
healthy cord =
healthy WM/GM
contrast
[1] Martin AR, Am J Neuroradiol 38:1266–73
Introduction
Methods
Results
Discussion
Acknowledgements
Agah Karakuzu
Aldo Zaimi
Ariane Saliani
Atef Badji
Axel Guittonneau
Benjamin De Leener
Charley Gros
Christian S. Perone
Gabriel Mangeat
Jennifer Campbell
Stephanie Alley
Tommy Boshkovski
Alexandru Foias
Nibardo Lopez-Rios
Grégoire Germain
Angel Chauffray
Julien Doyon
Carollyn Hurst
André Cyr
Jason Stockmann
Larry Wald
Nick Arango
Robert Barry
Thomas Witzel
Special thanks to collaborators +
co-authors!
See you at ISMRM 2019 in Montreal
— Creatively addressing the “real-estate problem” since 1967
Introduction
Methods
Results
Discussion
References
Topfer R et al., 2018, ’Integrated ΔB0/Rx coil array for improved spinal cord
imaging at 3T’, in ISMRM 26th Annual Meeting, Paris, France
link : http://cds.ismrm.org/protected/18MPresentations/abstracts/0834.html

More Related Content

Similar to Integrated RF and Shim coils for MRI

3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...
3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...
3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...Uzay Emir
 
Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...
Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...
Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...sugiuralab
 
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITION
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITIONECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITION
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITIONSarang Joshi
 
Introduction to compressed sensing MRI
Introduction to compressed sensing MRIIntroduction to compressed sensing MRI
Introduction to compressed sensing MRISairam Geethanath
 
APPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINES
APPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINESAPPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINES
APPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINEScseij
 
Application of particle swarm optimization to microwave tapered microstrip lines
Application of particle swarm optimization to microwave tapered microstrip linesApplication of particle swarm optimization to microwave tapered microstrip lines
Application of particle swarm optimization to microwave tapered microstrip linescseij
 
Computer aided detection of pulmonary nodules using genetic programming
Computer aided detection of pulmonary nodules using genetic programmingComputer aided detection of pulmonary nodules using genetic programming
Computer aided detection of pulmonary nodules using genetic programmingWookjin Choi
 
A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...
A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...
A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...sugiuralab
 
An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...
An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...
An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...IJRESJOURNAL
 
Optimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionOptimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionWookjin Choi
 
Image-guided management of uncertainties in scanned particle therapy
Image-guided management of uncertainties in scanned particle therapyImage-guided management of uncertainties in scanned particle therapy
Image-guided management of uncertainties in scanned particle therapyGiovanni Fattori
 
Petar Petrov MSc thesis defense
Petar Petrov MSc thesis defensePetar Petrov MSc thesis defense
Petar Petrov MSc thesis defensePetar Petrov
 
A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...
A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...
A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...CSCJournals
 
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisAn Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisCSCJournals
 
Biosensors And Bioelectronics Presentation by Sijung Hu
Biosensors And Bioelectronics Presentation by Sijung HuBiosensors And Bioelectronics Presentation by Sijung Hu
Biosensors And Bioelectronics Presentation by Sijung HuConferenceMind
 
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithm
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square AlgorithmImpedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithm
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithmijcisjournal
 

Similar to Integrated RF and Shim coils for MRI (20)

3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...
3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...
3D-UTE Rosette MRI and MRSI at UHF human (7T) and animal scanners (9.4T) usin...
 
Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...
Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...
Wearable Accelerometer Optimal Positions for Human Motion Recognition(LifeTec...
 
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITION
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITIONECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITION
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITION
 
Introduction to compressed sensing MRI
Introduction to compressed sensing MRIIntroduction to compressed sensing MRI
Introduction to compressed sensing MRI
 
APPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINES
APPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINESAPPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINES
APPLICATION OF PARTICLE SWARM OPTIMIZATION TO MICROWAVE TAPERED MICROSTRIP LINES
 
Application of particle swarm optimization to microwave tapered microstrip lines
Application of particle swarm optimization to microwave tapered microstrip linesApplication of particle swarm optimization to microwave tapered microstrip lines
Application of particle swarm optimization to microwave tapered microstrip lines
 
Computer aided detection of pulmonary nodules using genetic programming
Computer aided detection of pulmonary nodules using genetic programmingComputer aided detection of pulmonary nodules using genetic programming
Computer aided detection of pulmonary nodules using genetic programming
 
A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...
A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...
A Study of Wearable Accelerometers Layout for Human Activity Recognition(Asia...
 
I010415255
I010415255I010415255
I010415255
 
An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...
An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...
An Improved Adaptive Multi-Objective Particle Swarm Optimization for Disassem...
 
Laser drivenplasma
Laser drivenplasmaLaser drivenplasma
Laser drivenplasma
 
Optimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionOptimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detection
 
Image-guided management of uncertainties in scanned particle therapy
Image-guided management of uncertainties in scanned particle therapyImage-guided management of uncertainties in scanned particle therapy
Image-guided management of uncertainties in scanned particle therapy
 
Petar Petrov MSc thesis defense
Petar Petrov MSc thesis defensePetar Petrov MSc thesis defense
Petar Petrov MSc thesis defense
 
ISMRM_2006-2015_compressed
ISMRM_2006-2015_compressedISMRM_2006-2015_compressed
ISMRM_2006-2015_compressed
 
Final Project
Final ProjectFinal Project
Final Project
 
A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...
A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...
A Study of Total-Variation Based Noise-Reduction Algorithms For Low-Dose Cone...
 
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisAn Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
 
Biosensors And Bioelectronics Presentation by Sijung Hu
Biosensors And Bioelectronics Presentation by Sijung HuBiosensors And Bioelectronics Presentation by Sijung Hu
Biosensors And Bioelectronics Presentation by Sijung Hu
 
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithm
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square AlgorithmImpedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithm
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithm
 

More from NeuroPoly

Design and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI applicationDesign and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI applicationNeuroPoly
 
20200120 SCT workshop
20200120 SCT workshop20200120 SCT workshop
20200120 SCT workshopNeuroPoly
 
20190718 7 t_spine_coil
20190718 7 t_spine_coil20190718 7 t_spine_coil
20190718 7 t_spine_coilNeuroPoly
 
20190122_cohenadad_sc-mri-workshop
20190122_cohenadad_sc-mri-workshop20190122_cohenadad_sc-mri-workshop
20190122_cohenadad_sc-mri-workshopNeuroPoly
 
Design and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI applicationDesign and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI applicationNeuroPoly
 
qMRLab Introduction
qMRLab IntroductionqMRLab Introduction
qMRLab IntroductionNeuroPoly
 

More from NeuroPoly (7)

Design and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI applicationDesign and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI application
 
20200120 SCT workshop
20200120 SCT workshop20200120 SCT workshop
20200120 SCT workshop
 
SCT course
SCT courseSCT course
SCT course
 
20190718 7 t_spine_coil
20190718 7 t_spine_coil20190718 7 t_spine_coil
20190718 7 t_spine_coil
 
20190122_cohenadad_sc-mri-workshop
20190122_cohenadad_sc-mri-workshop20190122_cohenadad_sc-mri-workshop
20190122_cohenadad_sc-mri-workshop
 
Design and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI applicationDesign and manufacturing of coils for MRI application
Design and manufacturing of coils for MRI application
 
qMRLab Introduction
qMRLab IntroductionqMRLab Introduction
qMRLab Introduction
 

Recently uploaded

Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxOH TEIK BIN
 
Blooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docxBlooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docxUnboundStockton
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxthorishapillay1
 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxpboyjonauth
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsanshu789521
 
History Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptxHistory Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptxsocialsciencegdgrohi
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxRaymartEstabillo3
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxAvyJaneVismanos
 
Science lesson Moon for 4th quarter lesson
Science lesson Moon for 4th quarter lessonScience lesson Moon for 4th quarter lesson
Science lesson Moon for 4th quarter lessonJericReyAuditor
 
Biting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfBiting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfadityarao40181
 
ENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptx
ENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptxENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptx
ENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptxAnaBeatriceAblay2
 
भारत-रोम व्यापार.pptx, Indo-Roman Trade,
भारत-रोम व्यापार.pptx, Indo-Roman Trade,भारत-रोम व्यापार.pptx, Indo-Roman Trade,
भारत-रोम व्यापार.pptx, Indo-Roman Trade,Virag Sontakke
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...Marc Dusseiller Dusjagr
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdfssuser54595a
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxNirmalaLoungPoorunde1
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 

Recently uploaded (20)

Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptx
 
Blooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docxBlooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docx
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptx
 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptx
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha elections
 
History Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptxHistory Class XII Ch. 3 Kinship, Caste and Class (1).pptx
History Class XII Ch. 3 Kinship, Caste and Class (1).pptx
 
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
 
Staff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSDStaff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSD
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptx
 
Science lesson Moon for 4th quarter lesson
Science lesson Moon for 4th quarter lessonScience lesson Moon for 4th quarter lesson
Science lesson Moon for 4th quarter lesson
 
Biting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfBiting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdf
 
ENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptx
ENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptxENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptx
ENGLISH5 QUARTER4 MODULE1 WEEK1-3 How Visual and Multimedia Elements.pptx
 
भारत-रोम व्यापार.pptx, Indo-Roman Trade,
भारत-रोम व्यापार.pptx, Indo-Roman Trade,भारत-रोम व्यापार.pptx, Indo-Roman Trade,
भारत-रोम व्यापार.pptx, Indo-Roman Trade,
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptx
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 

Integrated RF and Shim coils for MRI

  • 1. Integrated ΔB0/Rx coil array for improved spinal cord imaging at 3T ISMRM 26th Annual Meeting June 20, 2018, Paris, France Topfer R1, Foias A1, Lopez Rios N1,2, Chauffray A1,3, Germain G1, Arango N4, Wald LL5,6, Stockmann JP5,6, Cohen-Adad J1,7 1. NeuroPoly Lab, Institute of Biomedical Engineering, Polytechnique Montreal, Montreal, QC, Canada, 2. Centro de Biofisica Medica, Universidad de Oriente, Santiago de Cuba, Cuba, 3. École polytechnique fédérale de Lausanne, Lausanne, Switzerland, 4. Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA, United States, 5. A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 6. Harvard Medical School, Boston, MA, United States, 7. Functional Neuroimaging Unit, CRIUGM, Université de Montréal, Montreal, QC, Canada
  • 2. Introduction Methods Results Discussion Problems 2 [4] Vannesjo S, Neuroimage 167:191-202 T2*-w @7T: ghosting from ΔB0 dynamic ΔB0 (e.g. respiration & swallowing [3,4]) [1] Stroman P, Neuroimage 84:1070-81 [2] Saritas E , Chapter 2.3 of Quantitative MRI of the Spinal Cord [3] Verma T, Magn Res Med 72:1629-36 distortions [2] static ΔB0 (e.g vertebrae, airways) inspired expired phase encoding (AP) [3] 10 mm “The greatest challenge for acquiring MR images in the spinal cord is [ΔB0] the inhomogeneous magnetic field in this region.” [1]
  • 3. Introduction Methods Results Discussion Solutions How to address these issues? Design principle Rx benefit Shim benefit Positioning coils near body ↑sensitivity, ↑SNR ↑efficiency (Hz/A) Using many coils enhances parallel imaging enables higher order ΔB0 correction [3] [1] Stockmann JP, Magn Res Med 75:441-51 [2] Truong TK, Neuroimage 103, 235–40 [3] Juchem C, J Mag Reson 236: 95-104 3 Parallel imaging + high-order dynamic shimming Solving the “real-estate problem” [1,2]
  • 4. Introduction Methods Results Discussion 4 Conceptualization [1] Germain G, ISMRM 2016. Abstract #3490. Optimal coil arrangement from simulations 3. ΔB0 + Rx SNR performance assessed for various coil arrangements 1. Sagittal ΔB0 maps acquired (n=5) +75 Hz -75 Hz 𝛾𝛥B0Avg.ΔB0 Inspired Expired } } 2. Auto-segmentation of cervical spinal cord ROI using the Spinal Cord Toolbox 𝛾𝛥B0 +30 Hz -30 Hz single coil optimal
  • 5. Introduction Methods Results Discussion 5 Construction RF only 5.5 cm [1] Stockmann JP, Magn Res Med 75:441-51 RF + shim
  • 6. Introduction Methods Results Discussion 6 Construction DC twisted pairs routed away from RF outputs Preamps atop plastic supports, away from coils TODO: periodic cable traps + chokes along DC cables [1] [1] Stockmann JP, Magn Res Med 75:441-51
  • 7. Introduction Methods Results Discussion +75 Hz/A-75 Hz/A ch1 ch5 ch2 ch3 ch6 ch7 ch4 ch8 ΔB0 reference maps used in shim optimization 7 Coil performance Coupling: min = 2 % max = 48 % avg = 18% 0 % 100% 50 % Noise correlation (Tx = 0V) Max shift per ch. avg = ±25 Hz/A over SC Rx measurements
  • 8. Introduction Methods Results Discussion Computer in control room shim update via USB +fibre optic • images transferred over local network • runs MATLAB + Spinal Cord Toolbox 8 Experimental setup [1] Arango N, ISMRM 2016. Abstract #1157. RF common mode chokes Power supply Equipment room Current driver [1] + heatsink Microcontroller +DAC pressure sensor + microcontroller 3T Prisma MRI room Pressure(t) respiratory bellows shim DC [2] Topfer R, Magn Res Med 80:935-46
  • 9. Introduction Methods Results Discussion 9 In vivo results Expired GRE mag +100 Hz-100 Hz 𝛾𝛥B0 Original shim VOI |Pred. − Actual| 25 Hz0 Hz 𝛾𝛥B0 Inspired σ 𝛥B0 =54 Hz σ 𝛥B0 =31 Hz σ 𝛥B0 =32 Hz σ 𝛥B0 =53 Hz Predicted Shimmed σ 𝛥B0 =29 Hz Actual σ 𝛥B0 =29 Hz 46% improvement σ 𝛥B0 40% improvement σ 𝛥B0
  • 10. Introduction Methods Results Discussion 10 In vivo results 14 mm-14 mm Original Shift EPI InspiredExpired mean |shift| ± st.dev 2.7±2.1mm mean |shift| ± st.dev 2.5±2.1mm Shimmed Shift EPI 10 cm mean |shift| ± st.dev 1.4±1.3mm mean |shift| ± st.dev 1.8±1.2mm max |shift| = 13 mm max |shift| = 8 mm
  • 11. Introduction Methods Results Discussion 11 Future directions 1. Real-time shimming 2. Making the most of the low inductance coils —> Potential for slicewise-dynamic shimming +100 Hz-100 Hz 𝛾𝛥B0 Inspired Expired Pressure(t) Inputs static shim respiration- correction shim(t) +20 Hz -20 Hz 𝛾𝛥B0 static ΔB0 respiration- induced ΔB0(t) +100 Hz -100 Hz 𝛾𝛥B0 Static Dynamic Outputs real-time shimmed real-time shimmed + = + = 3. Clinical app.: T2* for quantitative spinal-cord assessment healthy cord = healthy WM/GM contrast [1] Martin AR, Am J Neuroradiol 38:1266–73
  • 12. Introduction Methods Results Discussion Acknowledgements Agah Karakuzu Aldo Zaimi Ariane Saliani Atef Badji Axel Guittonneau Benjamin De Leener Charley Gros Christian S. Perone Gabriel Mangeat Jennifer Campbell Stephanie Alley Tommy Boshkovski Alexandru Foias Nibardo Lopez-Rios Grégoire Germain Angel Chauffray Julien Doyon Carollyn Hurst André Cyr Jason Stockmann Larry Wald Nick Arango Robert Barry Thomas Witzel Special thanks to collaborators + co-authors! See you at ISMRM 2019 in Montreal — Creatively addressing the “real-estate problem” since 1967
  • 13. Introduction Methods Results Discussion References Topfer R et al., 2018, ’Integrated ΔB0/Rx coil array for improved spinal cord imaging at 3T’, in ISMRM 26th Annual Meeting, Paris, France link : http://cds.ismrm.org/protected/18MPresentations/abstracts/0834.html

Editor's Notes

  1. How to address these issues? Well, to reduce distortion in EPI from extended readouts and echo trains, we want to use parallel imaging. At the same time, to eliminate susceptibility artifacts at the source, we want a high-order shim system capable of correcting both the small-scale static and dynamic changes to B0. The efficiency of both RF receive and shim subsystems improves with the proximity of the coils to the tissue of interest. So, to solve what Stockmann et. al dubbed "the real-estate problem", a natural choice is to integrate these 2 functionalities -- receive and B0 shim -- into the same set of coils.