Audiological fitting bulletin no. 59

                         How to measure the In-situ RECD with mind440 (m4-9) and
                         Compass V4.5




                         News from Audiological Research and Communication
9 502 1456 001 – 09-08
2
This bulletin describes how to measure the Widex in-situ RECD for mind440
using Compass V4.5. The description refers to the Acoustic conditions panel
in the Preconditions window, the RECD calibration window and the RECD
measurement windows in Compass. Please refer to Compass in order to see
details of the screen. Remember that Compass offers several help functions
to explain the different options in the program: Tooltips, Solution guide, Using
the window panels and the help manual.

mind440 m4-9 includes the possibility of measuring the RECD with the hearing aid itself. The pro-
cedure is called Widex in-situ RECD and is part of the m4-9 fitting procedure. The measurement is
made using a special RECD probe that is mounted on the hearing aid.

While the mind fitting procedure takes into account all the acoustic aspects of the hearing aid and its
coupling to the ear, the precision of the fitting may be further improved by performing an individual
RECD measurement on the client. This ensures that the hearing aid delivers the exact amount of
sound pressure at the eardrum prescribed for the given hearing loss – taking into account the shape
of the individual client’s ear canal.

To perform the RECD measurement, select Widex in-situ RECD in the RECD drop-down menu in
the Acoustic conditions panel found in the Preconditions window.




Selecting Widex in-situ RECD in the Preconditions window
3
After having selected the desired settings in the Preconditions window, press Next. Compass now
starts the Fitting wizard, with the in-situ RECD measurement as the first step. The measurement is
divided into two parts: (1) RECD calibration and (2) RECD measurement.

RECD calibration

 1. The first step is to calibrate the hearing aid; First, place the RECD probe over the rear micro-
    phone of the hearing aid. Secondly, insert the probe tube 5 mm into the hook of the hearing
    aid. Please observe that the hearing aid should never be placed on the ear of the client while the
    calibration is performed.




        Placement of the RECD probe                 Insertion depth of the probe tube for calibration
        tube for calibration

 2. Mount the orange seal sliding on the probe tube on the earhook to provide a tight seal between
    the hook and the tube.

 3. Make sure that the noise level is low enough (in the green area), before you start the calibration.

 4. Start the calibration by clicking the button Start calibration. When the measurement is done, the
    calibration result is displayed as either OK or “---“ .
4




RECD calibration in Compass showing that the calibration is OK

If the calibration failed (---), a dialog box is shown asking you to check the set-up once more and
repeat the measurement. If the calibration is OK, select Next to move on to the RECD measurement
window.
5
RECD measurement

 1. In the RECD measurement window, select whether you are measuring the RECD with an open
    or closed vent in the “Mould used for RECD measurement” drop-down menu. If there is a vent in
    the earmould, we recommend to leave it open during the measurement.

 2. Secondly, we recommend that you check your client’s ear canal status with an otoscope to make
    sure that conditions are normal and that the canal is not blocked by earwax.

 3. Place the hearing aid on the ear, and place the probe tube and the earmould in the ear of the cli-
    ent. Depending on the client, the probe should be placed at a recommended insertion depth from
    the intratragal notch:
       31 mm for an adult male
       28 mm for an adult female
       15-25 mm for infants and children

   Use the orange seal as a marker to mark the distance from the tip of the tube. You may also check
   the position of the probe tube at the eardrum with the otoscope after having inserted it into the
   ear canal.




                        Placement of the hearing aid on the ear with the probe tube
                        and earmould in place
6
  4. Make sure that the noise level is low enough (in the green area) before you start the measurement.
     Then start the RECD measurement; the progress bar indicates how far the measurement has pro-
     ceeded.

    If the measurement failed, a dialog box is shown asking you to check the set-up once more and
    repeat the measurement.

  5. The result of the in-situ RECD measurement is shown both as a graph and as values on the
     screen.

    The graph shows the in-situ RECD (in dB) as a function of frequency. The red (right ear) or blue
    (left ear) area indicates the range within which the measured in situ RECD is expected to fall ac-
    cording to the age of your client and the vent size in the earmould.




The graphics indicate the range within which the in-situ RECD measurement is expected to fall. The tables show
the binaurally measured values in dB at the main audiometric frequencies.

  6. When doing a binaural fitting, perform the measurement on the opposite ear. Alternatively you
     may choose to copy the RECD measured on one ear to the opposite ear. The best result will be
     obtained by performing the measurement on both ears, but in cases where this is not possible, the
     other approach may be used.

The measurement of the in-situ RECD is now finished and the in-situ RECD will be included in the
fitting. Remove the RECD probe and select Next to continue to the Sensogram measurement.

More Related Content

Viewers also liked

Compression View, M Manager And Pecalculate Oind 440
Compression View,  M Manager And  Pecalculate  Oind 440Compression View,  M Manager And  Pecalculate  Oind 440
Compression View, M Manager And Pecalculate Oind 440Geoffrey Cooling
 
Hear It Report October 2006
Hear It Report October 2006Hear It Report October 2006
Hear It Report October 2006Geoffrey Cooling
 
Online marketing strategy for audiologists
Online marketing strategy for audiologistsOnline marketing strategy for audiologists
Online marketing strategy for audiologistsGeoffrey Cooling
 
Citations & google my business for audiology marketing
Citations & google my business for audiology marketingCitations & google my business for audiology marketing
Citations & google my business for audiology marketingGeoffrey Cooling
 
Widex guidelines on tinnitus
Widex guidelines on tinnitusWidex guidelines on tinnitus
Widex guidelines on tinnitusGeoffrey Cooling
 
Commercial Strategy In Practice Management
Commercial Strategy In Practice ManagementCommercial Strategy In Practice Management
Commercial Strategy In Practice ManagementGeoffrey Cooling
 
Speech Enhancer Study For Facebook
Speech  Enhancer Study For FacebookSpeech  Enhancer Study For Facebook
Speech Enhancer Study For FacebookGeoffrey Cooling
 
Audibility Extender Feature from Widex
Audibility Extender Feature from WidexAudibility Extender Feature from Widex
Audibility Extender Feature from WidexGeoffrey Cooling
 
Passion Earset And Programming Notes
Passion Earset And Programming NotesPassion Earset And Programming Notes
Passion Earset And Programming NotesGeoffrey Cooling
 
Pay Per Click & social media in Audiology Marketing
Pay Per Click & social media in Audiology MarketingPay Per Click & social media in Audiology Marketing
Pay Per Click & social media in Audiology MarketingGeoffrey Cooling
 
Embedding young learners into the information society
Embedding young learners into the information societyEmbedding young learners into the information society
Embedding young learners into the information societyChristophe Guéret
 
Decentralised Open Data for World Citizens
Decentralised Open Data  for World CitizensDecentralised Open Data  for World Citizens
Decentralised Open Data for World CitizensChristophe Guéret
 
Let's downscale the semantic web !
Let's downscale the semantic web !Let's downscale the semantic web !
Let's downscale the semantic web !Christophe Guéret
 
Assessing Linked Data Mappings using Network Measures
Assessing Linked Data Mappings using Network MeasuresAssessing Linked Data Mappings using Network Measures
Assessing Linked Data Mappings using Network MeasuresChristophe Guéret
 
The road towards a Web-based data ecosystem
The road towards a Web-based data ecosystemThe road towards a Web-based data ecosystem
The road towards a Web-based data ecosystemChristophe Guéret
 
Exposing the data from NARCIS with VIVO
Exposing the data from NARCIS with VIVOExposing the data from NARCIS with VIVO
Exposing the data from NARCIS with VIVOChristophe Guéret
 
An Evolutionary Perspective on Approximate RDF Query Answering
An Evolutionary Perspective on Approximate RDF Query AnsweringAn Evolutionary Perspective on Approximate RDF Query Answering
An Evolutionary Perspective on Approximate RDF Query AnsweringChristophe Guéret
 
Evolutionary and Swarm Computing for scaling up the Semantic Web
Evolutionary and Swarm Computing for scaling up the Semantic WebEvolutionary and Swarm Computing for scaling up the Semantic Web
Evolutionary and Swarm Computing for scaling up the Semantic WebChristophe Guéret
 
Is linked data something for me?
Is linked data something for me?Is linked data something for me?
Is linked data something for me?Christophe Guéret
 

Viewers also liked (20)

Compression View, M Manager And Pecalculate Oind 440
Compression View,  M Manager And  Pecalculate  Oind 440Compression View,  M Manager And  Pecalculate  Oind 440
Compression View, M Manager And Pecalculate Oind 440
 
Hear It Report October 2006
Hear It Report October 2006Hear It Report October 2006
Hear It Report October 2006
 
Online marketing strategy for audiologists
Online marketing strategy for audiologistsOnline marketing strategy for audiologists
Online marketing strategy for audiologists
 
Citations & google my business for audiology marketing
Citations & google my business for audiology marketingCitations & google my business for audiology marketing
Citations & google my business for audiology marketing
 
Widex guidelines on tinnitus
Widex guidelines on tinnitusWidex guidelines on tinnitus
Widex guidelines on tinnitus
 
Commercial Strategy In Practice Management
Commercial Strategy In Practice ManagementCommercial Strategy In Practice Management
Commercial Strategy In Practice Management
 
Speech Enhancer Study For Facebook
Speech  Enhancer Study For FacebookSpeech  Enhancer Study For Facebook
Speech Enhancer Study For Facebook
 
Audibility Extender Feature from Widex
Audibility Extender Feature from WidexAudibility Extender Feature from Widex
Audibility Extender Feature from Widex
 
Frequency Transposition
Frequency TranspositionFrequency Transposition
Frequency Transposition
 
Passion Earset And Programming Notes
Passion Earset And Programming NotesPassion Earset And Programming Notes
Passion Earset And Programming Notes
 
Pay Per Click & social media in Audiology Marketing
Pay Per Click & social media in Audiology MarketingPay Per Click & social media in Audiology Marketing
Pay Per Click & social media in Audiology Marketing
 
Embedding young learners into the information society
Embedding young learners into the information societyEmbedding young learners into the information society
Embedding young learners into the information society
 
Decentralised Open Data for World Citizens
Decentralised Open Data  for World CitizensDecentralised Open Data  for World Citizens
Decentralised Open Data for World Citizens
 
Let's downscale the semantic web !
Let's downscale the semantic web !Let's downscale the semantic web !
Let's downscale the semantic web !
 
Assessing Linked Data Mappings using Network Measures
Assessing Linked Data Mappings using Network MeasuresAssessing Linked Data Mappings using Network Measures
Assessing Linked Data Mappings using Network Measures
 
The road towards a Web-based data ecosystem
The road towards a Web-based data ecosystemThe road towards a Web-based data ecosystem
The road towards a Web-based data ecosystem
 
Exposing the data from NARCIS with VIVO
Exposing the data from NARCIS with VIVOExposing the data from NARCIS with VIVO
Exposing the data from NARCIS with VIVO
 
An Evolutionary Perspective on Approximate RDF Query Answering
An Evolutionary Perspective on Approximate RDF Query AnsweringAn Evolutionary Perspective on Approximate RDF Query Answering
An Evolutionary Perspective on Approximate RDF Query Answering
 
Evolutionary and Swarm Computing for scaling up the Semantic Web
Evolutionary and Swarm Computing for scaling up the Semantic WebEvolutionary and Swarm Computing for scaling up the Semantic Web
Evolutionary and Swarm Computing for scaling up the Semantic Web
 
Is linked data something for me?
Is linked data something for me?Is linked data something for me?
Is linked data something for me?
 

Similar to How To Measure The In Situ R With Mind440 (M4 9

Videolaryngoscopes for tracheal intubation
Videolaryngoscopes for tracheal intubationVideolaryngoscopes for tracheal intubation
Videolaryngoscopes for tracheal intubationZIKRULLAH MALLICK
 
Chapter 17: Paralleling
Chapter 17: ParallelingChapter 17: Paralleling
Chapter 17: ParallelingKatieHenkel1
 
A Risk-Based Approach to Cleaning.pdf
A Risk-Based Approach to Cleaning.pdfA Risk-Based Approach to Cleaning.pdf
A Risk-Based Approach to Cleaning.pdfKellySoranyBerro
 
Otowave-202 Tympanometer | Henan Medical
Otowave-202 Tympanometer | Henan MedicalOtowave-202 Tympanometer | Henan Medical
Otowave-202 Tympanometer | Henan MedicalHenan Medical
 
Working length determination
Working length determinationWorking length determination
Working length determinationSaeed Bajafar
 
Instruction Manual | Maven RF1 Rangefinders | Optics Trade
 Instruction Manual | Maven RF1 Rangefinders  | Optics Trade Instruction Manual | Maven RF1 Rangefinders  | Optics Trade
Instruction Manual | Maven RF1 Rangefinders | Optics TradeOptics-Trade
 
Acoustic rhinometry
Acoustic rhinometryAcoustic rhinometry
Acoustic rhinometrySaef Moniem
 
Audio Media Monitors Headphones Guide 2014
Audio Media Monitors Headphones Guide 2014Audio Media Monitors Headphones Guide 2014
Audio Media Monitors Headphones Guide 2014Radikal Ltd.
 
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...SELINA SRAVANTHI
 
PM 10 Measurement
PM 10 MeasurementPM 10 Measurement
PM 10 MeasurementECRD IN
 
Pm10 measmnt
Pm10 measmntPm10 measmnt
Pm10 measmntECRD2015
 
Tonometery gonio scopy
Tonometery gonio scopyTonometery gonio scopy
Tonometery gonio scopyUsamaIqbal18
 
It's that small your take home sleep monitor
It's that small your take home sleep monitorIt's that small your take home sleep monitor
It's that small your take home sleep monitorDavid Schwaber
 
Introduction to OPTIK MIZEN
Introduction to OPTIK MIZENIntroduction to OPTIK MIZEN
Introduction to OPTIK MIZENRob Harnett
 
Qaulity Control of CT
Qaulity Control of CTQaulity Control of CT
Qaulity Control of CTAsim Gilani
 
Why do we all need to use CheckFox and what’s so Sigfox special about it?
Why do we all need to use CheckFox and what’s so Sigfox special about it?Why do we all need to use CheckFox and what’s so Sigfox special about it?
Why do we all need to use CheckFox and what’s so Sigfox special about it?Simple Hardware
 

Similar to How To Measure The In Situ R With Mind440 (M4 9 (20)

Ultrasonics aws
Ultrasonics awsUltrasonics aws
Ultrasonics aws
 
Videolaryngoscopes for tracheal intubation
Videolaryngoscopes for tracheal intubationVideolaryngoscopes for tracheal intubation
Videolaryngoscopes for tracheal intubation
 
Chapter 17: Paralleling
Chapter 17: ParallelingChapter 17: Paralleling
Chapter 17: Paralleling
 
A Risk-Based Approach to Cleaning.pdf
A Risk-Based Approach to Cleaning.pdfA Risk-Based Approach to Cleaning.pdf
A Risk-Based Approach to Cleaning.pdf
 
Otowave-202 Tympanometer | Henan Medical
Otowave-202 Tympanometer | Henan MedicalOtowave-202 Tympanometer | Henan Medical
Otowave-202 Tympanometer | Henan Medical
 
Working length determination
Working length determinationWorking length determination
Working length determination
 
Instruction Manual | Maven RF1 Rangefinders | Optics Trade
 Instruction Manual | Maven RF1 Rangefinders  | Optics Trade Instruction Manual | Maven RF1 Rangefinders  | Optics Trade
Instruction Manual | Maven RF1 Rangefinders | Optics Trade
 
Acoustic rhinometry
Acoustic rhinometryAcoustic rhinometry
Acoustic rhinometry
 
Audio Media Monitors Headphones Guide 2014
Audio Media Monitors Headphones Guide 2014Audio Media Monitors Headphones Guide 2014
Audio Media Monitors Headphones Guide 2014
 
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
 
PM 10 Measurement
PM 10 MeasurementPM 10 Measurement
PM 10 Measurement
 
Pm10 measmnt
Pm10 measmntPm10 measmnt
Pm10 measmnt
 
Tonometery gonio scopy
Tonometery gonio scopyTonometery gonio scopy
Tonometery gonio scopy
 
Critical Care Airway Management5 (3)
Critical Care Airway Management5 (3)Critical Care Airway Management5 (3)
Critical Care Airway Management5 (3)
 
It's that small your take home sleep monitor
It's that small your take home sleep monitorIt's that small your take home sleep monitor
It's that small your take home sleep monitor
 
Introduction to OPTIK MIZEN
Introduction to OPTIK MIZENIntroduction to OPTIK MIZEN
Introduction to OPTIK MIZEN
 
Qaulity Control of CT
Qaulity Control of CTQaulity Control of CT
Qaulity Control of CT
 
Hrt & g dx
Hrt & g dxHrt & g dx
Hrt & g dx
 
Visual Field | Humphrey Perimetry
Visual Field | Humphrey PerimetryVisual Field | Humphrey Perimetry
Visual Field | Humphrey Perimetry
 
Why do we all need to use CheckFox and what’s so Sigfox special about it?
Why do we all need to use CheckFox and what’s so Sigfox special about it?Why do we all need to use CheckFox and what’s so Sigfox special about it?
Why do we all need to use CheckFox and what’s so Sigfox special about it?
 

More from Geoffrey Cooling

Branding voor de audicienspraktijk: De klantervaring maakt uw merk
Branding voor de audicienspraktijk: De klantervaring maakt uw merkBranding voor de audicienspraktijk: De klantervaring maakt uw merk
Branding voor de audicienspraktijk: De klantervaring maakt uw merkGeoffrey Cooling
 
Twenty tips to move your healthcare practice forward
Twenty tips to move your healthcare practice forwardTwenty tips to move your healthcare practice forward
Twenty tips to move your healthcare practice forwardGeoffrey Cooling
 
Social media & marketing for audiological Practices
Social media & marketing for audiological PracticesSocial media & marketing for audiological Practices
Social media & marketing for audiological PracticesGeoffrey Cooling
 
Modern Practice Management
Modern Practice ManagementModern Practice Management
Modern Practice ManagementGeoffrey Cooling
 
Widex Clear Stand Ppt Language Corr2 V0.1
Widex Clear Stand Ppt Language Corr2 V0.1Widex Clear Stand Ppt Language Corr2 V0.1
Widex Clear Stand Ppt Language Corr2 V0.1Geoffrey Cooling
 
Acoustic identity using Tulip
Acoustic identity using TulipAcoustic identity using Tulip
Acoustic identity using TulipGeoffrey Cooling
 
DAI Programming in Compass 4.6 and above
DAI Programming in Compass 4.6 and aboveDAI Programming in Compass 4.6 and above
DAI Programming in Compass 4.6 and aboveGeoffrey Cooling
 
Using The Solution Guide In Mind440 Fittings
Using The  Solution Guide In Mind440 FittingsUsing The  Solution Guide In Mind440 Fittings
Using The Solution Guide In Mind440 FittingsGeoffrey Cooling
 
Using The Program Manager In Mind440 Fittings
Using The  Program Manager In Mind440 FittingsUsing The  Program Manager In Mind440 Fittings
Using The Program Manager In Mind440 FittingsGeoffrey Cooling
 
Widex Ireland ISHAA presentation
Widex Ireland ISHAA presentationWidex Ireland ISHAA presentation
Widex Ireland ISHAA presentationGeoffrey Cooling
 

More from Geoffrey Cooling (17)

Bshaa 2014
Bshaa 2014Bshaa 2014
Bshaa 2014
 
Branding voor de audicienspraktijk: De klantervaring maakt uw merk
Branding voor de audicienspraktijk: De klantervaring maakt uw merkBranding voor de audicienspraktijk: De klantervaring maakt uw merk
Branding voor de audicienspraktijk: De klantervaring maakt uw merk
 
Twenty tips to move your healthcare practice forward
Twenty tips to move your healthcare practice forwardTwenty tips to move your healthcare practice forward
Twenty tips to move your healthcare practice forward
 
Social media & marketing for audiological Practices
Social media & marketing for audiological PracticesSocial media & marketing for audiological Practices
Social media & marketing for audiological Practices
 
Modern Practice Management
Modern Practice ManagementModern Practice Management
Modern Practice Management
 
Zen and tinnitus
Zen and tinnitus Zen and tinnitus
Zen and tinnitus
 
Lets Talk Tinnitus
Lets Talk  TinnitusLets Talk  Tinnitus
Lets Talk Tinnitus
 
Clear440
Clear440Clear440
Clear440
 
Dex Product Presentation
Dex Product PresentationDex Product Presentation
Dex Product Presentation
 
Widex Clear Stand Ppt Language Corr2 V0.1
Widex Clear Stand Ppt Language Corr2 V0.1Widex Clear Stand Ppt Language Corr2 V0.1
Widex Clear Stand Ppt Language Corr2 V0.1
 
Acoustic identity using Tulip
Acoustic identity using TulipAcoustic identity using Tulip
Acoustic identity using Tulip
 
Mind 220
Mind 220Mind 220
Mind 220
 
Mind Comparison
Mind ComparisonMind Comparison
Mind Comparison
 
DAI Programming in Compass 4.6 and above
DAI Programming in Compass 4.6 and aboveDAI Programming in Compass 4.6 and above
DAI Programming in Compass 4.6 and above
 
Using The Solution Guide In Mind440 Fittings
Using The  Solution Guide In Mind440 FittingsUsing The  Solution Guide In Mind440 Fittings
Using The Solution Guide In Mind440 Fittings
 
Using The Program Manager In Mind440 Fittings
Using The  Program Manager In Mind440 FittingsUsing The  Program Manager In Mind440 Fittings
Using The Program Manager In Mind440 Fittings
 
Widex Ireland ISHAA presentation
Widex Ireland ISHAA presentationWidex Ireland ISHAA presentation
Widex Ireland ISHAA presentation
 

How To Measure The In Situ R With Mind440 (M4 9

  • 1. Audiological fitting bulletin no. 59 How to measure the In-situ RECD with mind440 (m4-9) and Compass V4.5 News from Audiological Research and Communication 9 502 1456 001 – 09-08
  • 2. 2 This bulletin describes how to measure the Widex in-situ RECD for mind440 using Compass V4.5. The description refers to the Acoustic conditions panel in the Preconditions window, the RECD calibration window and the RECD measurement windows in Compass. Please refer to Compass in order to see details of the screen. Remember that Compass offers several help functions to explain the different options in the program: Tooltips, Solution guide, Using the window panels and the help manual. mind440 m4-9 includes the possibility of measuring the RECD with the hearing aid itself. The pro- cedure is called Widex in-situ RECD and is part of the m4-9 fitting procedure. The measurement is made using a special RECD probe that is mounted on the hearing aid. While the mind fitting procedure takes into account all the acoustic aspects of the hearing aid and its coupling to the ear, the precision of the fitting may be further improved by performing an individual RECD measurement on the client. This ensures that the hearing aid delivers the exact amount of sound pressure at the eardrum prescribed for the given hearing loss – taking into account the shape of the individual client’s ear canal. To perform the RECD measurement, select Widex in-situ RECD in the RECD drop-down menu in the Acoustic conditions panel found in the Preconditions window. Selecting Widex in-situ RECD in the Preconditions window
  • 3. 3 After having selected the desired settings in the Preconditions window, press Next. Compass now starts the Fitting wizard, with the in-situ RECD measurement as the first step. The measurement is divided into two parts: (1) RECD calibration and (2) RECD measurement. RECD calibration 1. The first step is to calibrate the hearing aid; First, place the RECD probe over the rear micro- phone of the hearing aid. Secondly, insert the probe tube 5 mm into the hook of the hearing aid. Please observe that the hearing aid should never be placed on the ear of the client while the calibration is performed. Placement of the RECD probe Insertion depth of the probe tube for calibration tube for calibration 2. Mount the orange seal sliding on the probe tube on the earhook to provide a tight seal between the hook and the tube. 3. Make sure that the noise level is low enough (in the green area), before you start the calibration. 4. Start the calibration by clicking the button Start calibration. When the measurement is done, the calibration result is displayed as either OK or “---“ .
  • 4. 4 RECD calibration in Compass showing that the calibration is OK If the calibration failed (---), a dialog box is shown asking you to check the set-up once more and repeat the measurement. If the calibration is OK, select Next to move on to the RECD measurement window.
  • 5. 5 RECD measurement 1. In the RECD measurement window, select whether you are measuring the RECD with an open or closed vent in the “Mould used for RECD measurement” drop-down menu. If there is a vent in the earmould, we recommend to leave it open during the measurement. 2. Secondly, we recommend that you check your client’s ear canal status with an otoscope to make sure that conditions are normal and that the canal is not blocked by earwax. 3. Place the hearing aid on the ear, and place the probe tube and the earmould in the ear of the cli- ent. Depending on the client, the probe should be placed at a recommended insertion depth from the intratragal notch: 31 mm for an adult male 28 mm for an adult female 15-25 mm for infants and children Use the orange seal as a marker to mark the distance from the tip of the tube. You may also check the position of the probe tube at the eardrum with the otoscope after having inserted it into the ear canal. Placement of the hearing aid on the ear with the probe tube and earmould in place
  • 6. 6 4. Make sure that the noise level is low enough (in the green area) before you start the measurement. Then start the RECD measurement; the progress bar indicates how far the measurement has pro- ceeded. If the measurement failed, a dialog box is shown asking you to check the set-up once more and repeat the measurement. 5. The result of the in-situ RECD measurement is shown both as a graph and as values on the screen. The graph shows the in-situ RECD (in dB) as a function of frequency. The red (right ear) or blue (left ear) area indicates the range within which the measured in situ RECD is expected to fall ac- cording to the age of your client and the vent size in the earmould. The graphics indicate the range within which the in-situ RECD measurement is expected to fall. The tables show the binaurally measured values in dB at the main audiometric frequencies. 6. When doing a binaural fitting, perform the measurement on the opposite ear. Alternatively you may choose to copy the RECD measured on one ear to the opposite ear. The best result will be obtained by performing the measurement on both ears, but in cases where this is not possible, the other approach may be used. The measurement of the in-situ RECD is now finished and the in-situ RECD will be included in the fitting. Remove the RECD probe and select Next to continue to the Sensogram measurement.