Master in Medical Physics 2015 to 2016
Practical Report of Interventionism x-ray
Francisco J. Hernandez Flores
International Centre for Theoretical Physics
franciscohernandez_f2010@hotmail.com
December 21, 2015
Abstract
This task it is about the quality control in Conventional X-Ray, The aim of
the present Quality Control (QC) testing is to detect changes that may result in
a clinically significant degradation in image quality or a significant increase in
radiation exposures to both personnel and patients. of conventional x-ray such
as reproducibility of tube voltage, dose out put, time, and x-ray tube efficiency,
Accuracy of kVp, mA, time. Examinations of these factors are studied using
adequate instrumentation for measure different parameter in the hospital Majore
of Trieste Italy. in which were discuses during this practice important concepts as:
Automatic Exposure Control (AEC), KAP or DAP, the protection of the staff and
patient, the diagnostic reference levels (DRLs), the Interventional Reference Point
and the regular quality control in interventional radiology.
I. Introduction
Quality control (QC). has been acknowledged as a powerful and effective tool
in optimizing radiation protection of the patients in diagnostic and interven-
tional radiology (IR). Equipment performance testing is necessary to meet the
aims of diagnostic radiology, i.e. to obtain images adequate for the clinical
purpose with minimum radiation dose to patient.
QC in conventional diagnostic radiology has been well established. However,
QC of modern digital radiology systems requires different approaches owing
to differences in image receptor, image display devices, image format and
equipment configuration. Available QC testing methods should be updated to
encompass the developments in the field and to include digital imaging systems.
The objective of this report was to evaluate the parameters performance of
PHILIPS fluoroscopy machine used in Angiography procedures by perform
monthly quality control tests to check the radiation output consistency.
II. Theory
The x-ray system includes a collimator with motorized blades to adjust to the
FOV, and Kerma Air Product (KAP) meter to monitor radiation output for each
1
Master in Medical Physics 2015 to 2016
patient is often available on interventional systems.
Interventional X-ray system re-
quires: [3]
• Constant potential generator.
• C-arm system (under table x-ray
tube)
• High efficiency intensifier or flat
panel imaging system
• Digital image storage and re-
trieval
Generally different dose levels in
fluoroscopy are available: Low, Nor-
mal and High
I. Quality Controls
The Quality Controls in interventional
radiology comprises (Recommenda-
tions of Joint WHO-IRH-CE work-
shoop):
• Dosimetrical Controls
• Image quality
• Detector characterisation
• Geometrical controls
• KAP calibration
• CBCT
• DICOM/PACS
II. Constancy tests (monthly)
The basic constancy test are:
• Reference dose, dose rates
• Resolution
• Field diameter
• Collimation
• Contrast resolution
• Tube and generator parameters
• Hard copy devices
III. Useful quantities for patient and staff risk evaluation:
• Dose area product (for stochastic effect). [3]
• Entrance surface dose (for deterministic effect. [3]
• Staff dose per procedure (in more than one location). [3]
The factors affecting the staff doses are: the main source of radiation for the staff
in a fluoroscopy room is the patient due to the scattered radiation produced
in the interaction of the beam with the patient. The scattered radiation is not
uniform around the patient. The dose rate around the patient is a complex
function of a number of factors. The scattered dose rate at 1 m from the patient
can be higher than 1 mGy/min for some C-arm positions.
2
Master in Medical Physics 2015 to 2016
IV. Diagnostic Reference Levels (DRLs)
"Diagnostic reference levels" means dose levels in medical radiodiagnostic or
interventional radiology practices, or, in the case of radio-pharmaceuticals,
levels of activity, for typical examinations for groups of standard-sized patients
or standard phantoms for broadly defined types of equipment. [3]
DRLs are based on dose measurements in various types of hospitals, clinics
and practices. These values represent the 75th percentile KAP values recorded
in a survey carried out in different Member of European Community States. [4]
This parameter is useful to have a guidance on the level of dose, lower o higher
than the DRL, given to the patient for a specific treatments in a particular clinic.
V. Interventional Reference Point
For isocentric fluoroscopic systems, the interventional reference point is located
along the central x-ray beam at a distance of 15 cm from the isocenter in the
direction of the focal spot (see the representation of this point in fig. 1). The
interventional reference point is close to the entrance of the patient skin surface.
Cumulative Air Kerma (CK) at the Interventional Reference Point estimates
maximum skin dose. It is difficult to measure with accuracy but it give us an
idea of the maximum dose given to the patient.
Figure 1: Interventional Reference Point [1]
III. Materials and Methods
For quality control they were used the following measuring instruments: me-
ter, ionization chamber parallel plane, kVp, Electrometer, Philips angiography
fluoroscopy unit (C-arm fluoroscopy) was tested, Cu filters were used in this
3
Master in Medical Physics 2015 to 2016
experiment to check the exposure consistency.
This practice was made following the local protocol frequently used in qual-
ity control in the Catinara Hospital, The chamber of ionization was positioned
at 100 cm of distance at focal spot, connected at the electrometer for evaluated
the Ka. then measurements began varying the kV with the same mAs at same
distance focal spot surface with different field sizes were measured (31, 25, 20
and 17 cm) in different flouroscopy modes (low, normal and high) were used
to stimulate the automatic exposure control (AEC) to check the kVp and mA
appeared in control panel of the system.
The value of kVp and mA were registered from the control panel and com-
pared with the reference values of acceptation test for the machine.
The Quality Control (QC) in radiography is a central part of QA programme,
which deals with equipment maintenance and monitoring. QA in diagnostic
radiology is a mean of maintaining standards in imaging and working towards
minimizing patient and staff doses. To accomplish these objectives, a number
of physical parameters that affect the performance of X-ray imaging system are
to be measured.
The equipment taking in to account for Radiation protection were lead
aprons and thyroid shields also were visually checked.
IV. Results and discussion
During the practice the monthly test regarding the constancy of the parameter
of exposition was measured to compare with the reference base state. The data
obtained were similar to the data showed in the Fig. 2. This test is important to
control the well functionality of the Automatic Exposure Control (AEC).
The AEC controls the Incident Air Kerma to the image detector prevent fluc-
tuation in image brightness and SNR that would make diagnosis or navigation
of instruments difficult [2].
Fluoroscopic AEC may use the signal from a sensor to determine necessary
adjustments of fluoroscopic technique factors such as tube voltage and tube
current which depends of the Field of View (FOV) and the dose level (Low,
Normal and High) like as showed in the fig. 2.
4
Master in Medical Physics 2015 to 2016
Figure 2: Monthly test to verify the parameter of exposition
5
Master in Medical Physics 2015 to 2016
V. Conclusion
• One of the most important part of quality control in interventional radiol-
ogy is to measure the output of the Automatic Exposure Control in order
to avoid high unnecessary dose to the patient.
• Other important aspect in radiology obtained the best quality image useful
for the radiologist using the ALARA concept in radiation protection for to
reduce the dose of staff and damage in the skin of the patient due to the
high dose.
• Due to the receive high dose or low dose produce effect they can to
be deterministic or probabilistic even the absorbed dose in this point
is very useful that the staff use the adequate protection like: aprons,
lead eyeglasses, thyroid shield and radiation shields in order to reduce
significant this effects
• The reference point air kerma is the most approximate quantity to cal-
culate the skin dose to the patient but it is not accurate and usually
underestimates or overestimates the real dose.
References
[1] Annalisa Trianni, Lecture 10 Skin dose calculation of Information Technology
in Medicine, ICTP Trieste Italy, 3rd trimester 2015.
[2] Dance, et.al. Diagnostic Radiology Physics. IAEA, Vienna, 2014.
[3] Paola Bregant , Lecture Physics of Diagnostic with x-ray 2, ICTP Trieste Italy,
2015
[4] R.Padovani et al Reference Levels at European Level for Cardiac Interventional
Procedures, Vol:129;105-107, 2008.
6

More Related Content

What's hot

Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]radiation oncology
 
Imrt Treatment Planning And Dosimetry
Imrt Treatment Planning And DosimetryImrt Treatment Planning And Dosimetry
Imrt Treatment Planning And Dosimetryfondas vakalis
 
Computed tomography - radiation exposure
Computed tomography - radiation exposureComputed tomography - radiation exposure
Computed tomography - radiation exposureYan Vargas
 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in RadiotherapySantam Chakraborty
 
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha Avinesh Shrestha
 
introduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapyintroduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapyRahim Gohar
 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationMiami Cancer Institute
 
Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)Dilshad Kottuparamban
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct Muntaser S.Ahmad
 
Dosimetric errors of radiotherapy techniques involving small fields
Dosimetric errors of radiotherapy techniques involving small fieldsDosimetric errors of radiotherapy techniques involving small fields
Dosimetric errors of radiotherapy techniques involving small fieldsBiplab Sarkar
 
Past, present and future of radiotherapy technique in different sites: Breast...
Past, present and future of radiotherapy technique in different sites: Breast...Past, present and future of radiotherapy technique in different sites: Breast...
Past, present and future of radiotherapy technique in different sites: Breast...Biplab Sarkar
 
Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Ravindra Shende
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose IndexAnjan Dangal
 
Lung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing TechniquesLung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing TechniquesIRJET Journal
 
Patient positional correction stategies in radiotherapy
Patient positional correction stategies   in radiotherapyPatient positional correction stategies   in radiotherapy
Patient positional correction stategies in radiotherapyBiplab Sarkar
 
3DCRT. radiation oncology conference in Tanzania
3DCRT. radiation oncology conference in Tanzania3DCRT. radiation oncology conference in Tanzania
3DCRT. radiation oncology conference in Tanzaniareuben mringo
 

What's hot (20)

Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]
 
Imrt Treatment Planning And Dosimetry
Imrt Treatment Planning And DosimetryImrt Treatment Planning And Dosimetry
Imrt Treatment Planning And Dosimetry
 
Computed tomography - radiation exposure
Computed tomography - radiation exposureComputed tomography - radiation exposure
Computed tomography - radiation exposure
 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in Radiotherapy
 
Inverse Planning
Inverse PlanningInverse Planning
Inverse Planning
 
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
 
IMRT and 3DCRT
IMRT and 3DCRT IMRT and 3DCRT
IMRT and 3DCRT
 
introduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapyintroduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapy
 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verification
 
Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)Intensity Modulated Radiation Therapy (IMRT)
Intensity Modulated Radiation Therapy (IMRT)
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 
Dosimetric errors of radiotherapy techniques involving small fields
Dosimetric errors of radiotherapy techniques involving small fieldsDosimetric errors of radiotherapy techniques involving small fields
Dosimetric errors of radiotherapy techniques involving small fields
 
Past, present and future of radiotherapy technique in different sites: Breast...
Past, present and future of radiotherapy technique in different sites: Breast...Past, present and future of radiotherapy technique in different sites: Breast...
Past, present and future of radiotherapy technique in different sites: Breast...
 
Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy
 
IMRT
IMRTIMRT
IMRT
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose Index
 
Radiotherapy Treatment Simulation
Radiotherapy Treatment SimulationRadiotherapy Treatment Simulation
Radiotherapy Treatment Simulation
 
Lung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing TechniquesLung Cancer Detection using Image Processing Techniques
Lung Cancer Detection using Image Processing Techniques
 
Patient positional correction stategies in radiotherapy
Patient positional correction stategies   in radiotherapyPatient positional correction stategies   in radiotherapy
Patient positional correction stategies in radiotherapy
 
3DCRT. radiation oncology conference in Tanzania
3DCRT. radiation oncology conference in Tanzania3DCRT. radiation oncology conference in Tanzania
3DCRT. radiation oncology conference in Tanzania
 

Similar to Quality Control of X-Ray Parameters and Automatic Exposure

Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stroundhydrologywebsite1
 
Sutton Leiden Presentation Dec 2008 Ds4
Sutton Leiden Presentation Dec 2008 Ds4Sutton Leiden Presentation Dec 2008 Ds4
Sutton Leiden Presentation Dec 2008 Ds4drn
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality ControlECRD IN
 
Quality assurance and quality control
Quality assurance and quality controlQuality assurance and quality control
Quality assurance and quality controlECRD2015
 
Chapter 12
Chapter 12Chapter 12
Chapter 12ECRD IN
 
Radiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precisionRadiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precisionAmin Amin
 
12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO procedures12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO proceduresEuro CTO Club
 
Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyRamzee Small
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2ThorikulHuda2
 
Dose Evaluation and Reporting
Dose Evaluation and Reporting Dose Evaluation and Reporting
Dose Evaluation and Reporting Carestream
 
EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649 EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649 Randox
 
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaPatient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaRamzee Small
 
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATIONTHE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATIONMoustafa Rezk
 

Similar to Quality Control of X-Ray Parameters and Automatic Exposure (20)

Session 1 conventional x ray
Session 1  conventional x raySession 1  conventional x ray
Session 1 conventional x ray
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 
Digital flouroscopy
Digital flouroscopyDigital flouroscopy
Digital flouroscopy
 
Sutton Leiden Presentation Dec 2008 Ds4
Sutton Leiden Presentation Dec 2008 Ds4Sutton Leiden Presentation Dec 2008 Ds4
Sutton Leiden Presentation Dec 2008 Ds4
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality Control
 
Quality assurance and quality control
Quality assurance and quality controlQuality assurance and quality control
Quality assurance and quality control
 
Chapter 12
Chapter 12Chapter 12
Chapter 12
 
Radiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precisionRadiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precision
 
12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO procedures12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO procedures
 
Summary of embrace protocol
Summary of embrace protocolSummary of embrace protocol
Summary of embrace protocol
 
Radiation dose reduction
Radiation dose reduction Radiation dose reduction
Radiation dose reduction
 
HUMC IMPROVE TRIFOLD
HUMC IMPROVE TRIFOLDHUMC IMPROVE TRIFOLD
HUMC IMPROVE TRIFOLD
 
ICRU 83
ICRU 83ICRU 83
ICRU 83
 
Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2
 
Image Guided Radiotherapy
Image Guided RadiotherapyImage Guided Radiotherapy
Image Guided Radiotherapy
 
Dose Evaluation and Reporting
Dose Evaluation and Reporting Dose Evaluation and Reporting
Dose Evaluation and Reporting
 
EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649 EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649
 
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaPatient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
 
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATIONTHE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
 

More from Centro Nacional de Radioterapia Nora Astorga (12)

Flattening filter Free
Flattening filter FreeFlattening filter Free
Flattening filter Free
 
Session 12 imaging instrumentation (spect) qc
Session 12 imaging instrumentation (spect) qcSession 12 imaging instrumentation (spect) qc
Session 12 imaging instrumentation (spect) qc
 
Session 11 non imaging instrumentation qc
Session 11 non imaging instrumentation qcSession 11 non imaging instrumentation qc
Session 11 non imaging instrumentation qc
 
Session 10 radiation survey of a clinical installation
Session 10 radiation survey of a clinical installationSession 10 radiation survey of a clinical installation
Session 10 radiation survey of a clinical installation
 
Session 9 radiation oncology
Session 9 radiation oncologySession 9 radiation oncology
Session 9 radiation oncology
 
Session 5 mammography
Session 5 mammographySession 5 mammography
Session 5 mammography
 
Session 4 plan verification prostate
Session 4 plan verification prostateSession 4 plan verification prostate
Session 4 plan verification prostate
 
Session 3 measure beam profile rt
Session 3 measure beam profile rtSession 3 measure beam profile rt
Session 3 measure beam profile rt
 
Session 2 radiotherapy
Session 2 radiotherapySession 2 radiotherapy
Session 2 radiotherapy
 
Monografia
MonografiaMonografia
Monografia
 
Física de la radiología.
Física de la radiología.Física de la radiología.
Física de la radiología.
 
Icru 62
Icru 62Icru 62
Icru 62
 

Recently uploaded

Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Nehru place Escorts
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls ServiceMiss joya
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowRiya Pathan
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...narwatsonia7
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...narwatsonia7
 
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Nehru place Escorts
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 

Recently uploaded (20)

Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
 
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 

Quality Control of X-Ray Parameters and Automatic Exposure

  • 1. Master in Medical Physics 2015 to 2016 Practical Report of Interventionism x-ray Francisco J. Hernandez Flores International Centre for Theoretical Physics franciscohernandez_f2010@hotmail.com December 21, 2015 Abstract This task it is about the quality control in Conventional X-Ray, The aim of the present Quality Control (QC) testing is to detect changes that may result in a clinically significant degradation in image quality or a significant increase in radiation exposures to both personnel and patients. of conventional x-ray such as reproducibility of tube voltage, dose out put, time, and x-ray tube efficiency, Accuracy of kVp, mA, time. Examinations of these factors are studied using adequate instrumentation for measure different parameter in the hospital Majore of Trieste Italy. in which were discuses during this practice important concepts as: Automatic Exposure Control (AEC), KAP or DAP, the protection of the staff and patient, the diagnostic reference levels (DRLs), the Interventional Reference Point and the regular quality control in interventional radiology. I. Introduction Quality control (QC). has been acknowledged as a powerful and effective tool in optimizing radiation protection of the patients in diagnostic and interven- tional radiology (IR). Equipment performance testing is necessary to meet the aims of diagnostic radiology, i.e. to obtain images adequate for the clinical purpose with minimum radiation dose to patient. QC in conventional diagnostic radiology has been well established. However, QC of modern digital radiology systems requires different approaches owing to differences in image receptor, image display devices, image format and equipment configuration. Available QC testing methods should be updated to encompass the developments in the field and to include digital imaging systems. The objective of this report was to evaluate the parameters performance of PHILIPS fluoroscopy machine used in Angiography procedures by perform monthly quality control tests to check the radiation output consistency. II. Theory The x-ray system includes a collimator with motorized blades to adjust to the FOV, and Kerma Air Product (KAP) meter to monitor radiation output for each 1
  • 2. Master in Medical Physics 2015 to 2016 patient is often available on interventional systems. Interventional X-ray system re- quires: [3] • Constant potential generator. • C-arm system (under table x-ray tube) • High efficiency intensifier or flat panel imaging system • Digital image storage and re- trieval Generally different dose levels in fluoroscopy are available: Low, Nor- mal and High I. Quality Controls The Quality Controls in interventional radiology comprises (Recommenda- tions of Joint WHO-IRH-CE work- shoop): • Dosimetrical Controls • Image quality • Detector characterisation • Geometrical controls • KAP calibration • CBCT • DICOM/PACS II. Constancy tests (monthly) The basic constancy test are: • Reference dose, dose rates • Resolution • Field diameter • Collimation • Contrast resolution • Tube and generator parameters • Hard copy devices III. Useful quantities for patient and staff risk evaluation: • Dose area product (for stochastic effect). [3] • Entrance surface dose (for deterministic effect. [3] • Staff dose per procedure (in more than one location). [3] The factors affecting the staff doses are: the main source of radiation for the staff in a fluoroscopy room is the patient due to the scattered radiation produced in the interaction of the beam with the patient. The scattered radiation is not uniform around the patient. The dose rate around the patient is a complex function of a number of factors. The scattered dose rate at 1 m from the patient can be higher than 1 mGy/min for some C-arm positions. 2
  • 3. Master in Medical Physics 2015 to 2016 IV. Diagnostic Reference Levels (DRLs) "Diagnostic reference levels" means dose levels in medical radiodiagnostic or interventional radiology practices, or, in the case of radio-pharmaceuticals, levels of activity, for typical examinations for groups of standard-sized patients or standard phantoms for broadly defined types of equipment. [3] DRLs are based on dose measurements in various types of hospitals, clinics and practices. These values represent the 75th percentile KAP values recorded in a survey carried out in different Member of European Community States. [4] This parameter is useful to have a guidance on the level of dose, lower o higher than the DRL, given to the patient for a specific treatments in a particular clinic. V. Interventional Reference Point For isocentric fluoroscopic systems, the interventional reference point is located along the central x-ray beam at a distance of 15 cm from the isocenter in the direction of the focal spot (see the representation of this point in fig. 1). The interventional reference point is close to the entrance of the patient skin surface. Cumulative Air Kerma (CK) at the Interventional Reference Point estimates maximum skin dose. It is difficult to measure with accuracy but it give us an idea of the maximum dose given to the patient. Figure 1: Interventional Reference Point [1] III. Materials and Methods For quality control they were used the following measuring instruments: me- ter, ionization chamber parallel plane, kVp, Electrometer, Philips angiography fluoroscopy unit (C-arm fluoroscopy) was tested, Cu filters were used in this 3
  • 4. Master in Medical Physics 2015 to 2016 experiment to check the exposure consistency. This practice was made following the local protocol frequently used in qual- ity control in the Catinara Hospital, The chamber of ionization was positioned at 100 cm of distance at focal spot, connected at the electrometer for evaluated the Ka. then measurements began varying the kV with the same mAs at same distance focal spot surface with different field sizes were measured (31, 25, 20 and 17 cm) in different flouroscopy modes (low, normal and high) were used to stimulate the automatic exposure control (AEC) to check the kVp and mA appeared in control panel of the system. The value of kVp and mA were registered from the control panel and com- pared with the reference values of acceptation test for the machine. The Quality Control (QC) in radiography is a central part of QA programme, which deals with equipment maintenance and monitoring. QA in diagnostic radiology is a mean of maintaining standards in imaging and working towards minimizing patient and staff doses. To accomplish these objectives, a number of physical parameters that affect the performance of X-ray imaging system are to be measured. The equipment taking in to account for Radiation protection were lead aprons and thyroid shields also were visually checked. IV. Results and discussion During the practice the monthly test regarding the constancy of the parameter of exposition was measured to compare with the reference base state. The data obtained were similar to the data showed in the Fig. 2. This test is important to control the well functionality of the Automatic Exposure Control (AEC). The AEC controls the Incident Air Kerma to the image detector prevent fluc- tuation in image brightness and SNR that would make diagnosis or navigation of instruments difficult [2]. Fluoroscopic AEC may use the signal from a sensor to determine necessary adjustments of fluoroscopic technique factors such as tube voltage and tube current which depends of the Field of View (FOV) and the dose level (Low, Normal and High) like as showed in the fig. 2. 4
  • 5. Master in Medical Physics 2015 to 2016 Figure 2: Monthly test to verify the parameter of exposition 5
  • 6. Master in Medical Physics 2015 to 2016 V. Conclusion • One of the most important part of quality control in interventional radiol- ogy is to measure the output of the Automatic Exposure Control in order to avoid high unnecessary dose to the patient. • Other important aspect in radiology obtained the best quality image useful for the radiologist using the ALARA concept in radiation protection for to reduce the dose of staff and damage in the skin of the patient due to the high dose. • Due to the receive high dose or low dose produce effect they can to be deterministic or probabilistic even the absorbed dose in this point is very useful that the staff use the adequate protection like: aprons, lead eyeglasses, thyroid shield and radiation shields in order to reduce significant this effects • The reference point air kerma is the most approximate quantity to cal- culate the skin dose to the patient but it is not accurate and usually underestimates or overestimates the real dose. References [1] Annalisa Trianni, Lecture 10 Skin dose calculation of Information Technology in Medicine, ICTP Trieste Italy, 3rd trimester 2015. [2] Dance, et.al. Diagnostic Radiology Physics. IAEA, Vienna, 2014. [3] Paola Bregant , Lecture Physics of Diagnostic with x-ray 2, ICTP Trieste Italy, 2015 [4] R.Padovani et al Reference Levels at European Level for Cardiac Interventional Procedures, Vol:129;105-107, 2008. 6