SlideShare a Scribd company logo
1 of 3
Download to read offline
Master in Medical Physics ICTP 2015-2016
Radiotherapy treatment verification with Octavius phantom
Francisco J.Hernández Flores∗
International Centre for Theoretical Physics
franciscohernandez_f2010@hotmail.com
August 31, 2015
Abstract
The present task is mainly focused on the verification of the calculated dose. During a
radiotherapy treatment, a dosimetric verification or a geometric localization can be done,
in order to assess the quality of the treatment. The dosimetric verification is generally
performed measuring the dose at some points inside or outside the patient, and comparing
it to the dose at the same points calculated and predicted by the treatment planning system.
This was evaluated with phantom Octavius 4D. Dose distributions were calculated for
volumetric modulated arc therapy (VMAT) treatment plans with Monaco the results of
gamma analysis was 98% passing criteria.
I. Introduction
Volumetric Modulated Arc Therapy (VMAT) is a new intensity-modulated radiation
therapy (IMRT) technique that improves critical structures and healthy tissue sparing,
in dramatically shorter treatment times all without comprising target coverage and
patient safety [2]
when advanced techniques used in radiotherapy and to finalize a plan of VMAT
in the TPS is mandatory dosimetric verification before clinical delivery using diodes
or ionization chambers, in our case we used the Octavius phantom to check the
effectiveness of the plan dose delivered to the patient and compare the passes of the
plan take into account the gamma factor and DTA.
OCTAVIUS 4D is a new 4D dosimetry system designed to verify IMRT treatment
plans based on true independent measurements. It rotates with the gantry, measuring
the dose inside the entire phantom volume, always per pendicular to the incident
beam. As a consequence, OCTAVIUS 4D requires no angular corrections or detector
calibrations to compensate for the directional response of its detectors which makes it
perfectly suited for rotational delivery techniques and easy-to-use in clinical routine. [1]
II. Theory
Distance To Agreement (DTA) is the distance between a measured data point and the
nearest point in the calculated dose distribution that exhibits the same dose [2].
The gamma function evaluates the difference between measured and calculated
dose distribution using both dose and spatial coordinates [3].
Γ(rm, rc) =
r2(rm, rc)
∆d2
m
+
δ2(rm, rc)
∆D2
m
(1)
∗Radiotherapy Practical 2
1
Master in Medical Physics ICTP 2015-2016
with δ(rm, rc) = Dm − Dc; r(rm, rc) = |rm, rc|
γ(rm) = min{Γ(rm, rc)}∀{rc} (2)
γ(rm) = min
r2(rm, rc)
∆d2
m
+
δ2(rm, rc)
∆D2
m
∀rc (3)
γ(rm) function, the minimum generalized Γ(rm, rc) function in the set of evaluated
points. ∆dm and ∆Dm the acceptance criteria for distance to dose r(rm, rc) is the distance
between the measured point rm and a point rc chosen in calculated dose distribution δ
is the corresponding dose difference. As a general rule, acceptance criteria of 3% dose
difference and 3 mm distance to agreement. The dose-difference criterion is ∆Dm , and
the DTA criterion is ∆dm.
III. Material and method
Verification of Treatment planning with dose measure using Treatment Planning sistem
Monaco, red mosaiq, phantom Octavius 4D (see fig. 1) with respective software,
motorized cylindrical phantom specially designed for rotational delivery techniques,
includes control unit and wireless inclinometer.
The plan used in this practice was calculated, teacher only explain how she made
the plan because the practice is only verification plan, she told us how is the procedure
for align the phantom and connected all the tools, the software use the equation 3 to
calculate the gamma factor for the quantitative evaluation of dose distributions.
Figure 1: Phantom Octavius 4D
IV. Analysis and discussion of results
The rsult obtained during the verification planning treatment was 98% passing criteria
based in clinical standard for photon beam. This plan was analyzed using the two
parameter DTA (limit 3%), (limit 3 mm), and index with the 3% dose tolerance and
3 mm distance to agreement in relation to the treatment planning system. The range
of acceptant pass of the gamma criterion depend anatomical site, Results confirmed a
good agreement between the two distribution with high and conformed dose to the
target and low dose to the organ at risk.
2
Master in Medical Physics ICTP 2015-2016
V. Conclusion
• Clinical quality assurance practice based on linac treatment log files for verifi-
cation of delivered 3D dose distributions in the patients geometries represents
a paradigm shift from dose measurements in a phantom. This approach offers
independent verification of VMAT dose distributions encompassing treatment
planning beam modeling differences as well as the linac uncertainties during
treatment delivery of the plan.
• Verification of a planned dose distribution is a complex and time consuming
procedure because all described parameters DTA and γ-index are very important
and using only one of them is not enough to accept a treatment plan.
References
[1] Elekta Manual of the phantom Octavius 4D,Viena 2000
[2] Daniel A. Low, William B. Harms, et al A technique for the quantitative evaluation
of dose distributions, Division of Radiation Oncology, 510 South Kingshighway Blvd.
[3] Michele Avanzo, Lecture of Oncology 3, ICTP, 2015.
3

More Related Content

What's hot

Abrigo and love_2015_
Abrigo and love_2015_Abrigo and love_2015_
Abrigo and love_2015_Murtaza Khan
 
International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)IJERD Editor
 
Medical Image Segmentation Based on Level Set Method
Medical Image Segmentation Based on Level Set MethodMedical Image Segmentation Based on Level Set Method
Medical Image Segmentation Based on Level Set MethodIOSR Journals
 
AIAA-Aviation-2015-Mehmani
AIAA-Aviation-2015-MehmaniAIAA-Aviation-2015-Mehmani
AIAA-Aviation-2015-MehmaniOptiModel
 
A Bayesian approach to estimate probabilities in classification trees
A Bayesian approach to estimate probabilities in classification treesA Bayesian approach to estimate probabilities in classification trees
A Bayesian approach to estimate probabilities in classification treesNTNU
 
Indirect inference for the application of random regret minimisation to large...
Indirect inference for the application of random regret minimisation to large...Indirect inference for the application of random regret minimisation to large...
Indirect inference for the application of random regret minimisation to large...Institute for Transport Studies (ITS)
 
Computational intelligence systems in industrial engineering
Computational intelligence systems in industrial engineeringComputational intelligence systems in industrial engineering
Computational intelligence systems in industrial engineeringSpringer
 
Analytical method development,validation by uv spectroscopy
Analytical method development,validation by uv spectroscopyAnalytical method development,validation by uv spectroscopy
Analytical method development,validation by uv spectroscopythota lakshmi bhavani
 
international article- semnan
international article- semnaninternational article- semnan
international article- semnanM Jafari
 

What's hot (9)

Abrigo and love_2015_
Abrigo and love_2015_Abrigo and love_2015_
Abrigo and love_2015_
 
International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)
 
Medical Image Segmentation Based on Level Set Method
Medical Image Segmentation Based on Level Set MethodMedical Image Segmentation Based on Level Set Method
Medical Image Segmentation Based on Level Set Method
 
AIAA-Aviation-2015-Mehmani
AIAA-Aviation-2015-MehmaniAIAA-Aviation-2015-Mehmani
AIAA-Aviation-2015-Mehmani
 
A Bayesian approach to estimate probabilities in classification trees
A Bayesian approach to estimate probabilities in classification treesA Bayesian approach to estimate probabilities in classification trees
A Bayesian approach to estimate probabilities in classification trees
 
Indirect inference for the application of random regret minimisation to large...
Indirect inference for the application of random regret minimisation to large...Indirect inference for the application of random regret minimisation to large...
Indirect inference for the application of random regret minimisation to large...
 
Computational intelligence systems in industrial engineering
Computational intelligence systems in industrial engineeringComputational intelligence systems in industrial engineering
Computational intelligence systems in industrial engineering
 
Analytical method development,validation by uv spectroscopy
Analytical method development,validation by uv spectroscopyAnalytical method development,validation by uv spectroscopy
Analytical method development,validation by uv spectroscopy
 
international article- semnan
international article- semnaninternational article- semnan
international article- semnan
 

Similar to Session 4 plan verification prostate

IMRT by Musaib Mushtaq.ppt
IMRT by Musaib Mushtaq.pptIMRT by Musaib Mushtaq.ppt
IMRT by Musaib Mushtaq.pptMusaibMushtaq
 
IMRT in Head & Neck Cancer
IMRT in Head & Neck CancerIMRT in Head & Neck Cancer
IMRT in Head & Neck CancerJyotirup Goswami
 
ICRU 83 report on dose prescription in IMRT
ICRU 83 report on dose prescription in IMRTICRU 83 report on dose prescription in IMRT
ICRU 83 report on dose prescription in IMRTAnagha pachat
 
Comparative dosimetry of forward and inverse treatment planning for Intensity...
Comparative dosimetry of forward and inverse treatment planning for Intensity...Comparative dosimetry of forward and inverse treatment planning for Intensity...
Comparative dosimetry of forward and inverse treatment planning for Intensity...iosrjce
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62Dhiman Das
 
P14_BRAKHITERAPI.pptx
P14_BRAKHITERAPI.pptxP14_BRAKHITERAPI.pptx
P14_BRAKHITERAPI.pptxdimaswardani1
 
introduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapyintroduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapyRahim Gohar
 
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. SachinIntensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. SachinSACHINS700327
 
Radiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precisionRadiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precisionAmin Amin
 
Brain Tumor Detection using Clustering Algorithms in MRI Images
Brain Tumor Detection using Clustering Algorithms in MRI ImagesBrain Tumor Detection using Clustering Algorithms in MRI Images
Brain Tumor Detection using Clustering Algorithms in MRI ImagesIRJET Journal
 
Icru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationIcru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationalthaf jouhar
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentationSathish Kumar
 
Tomosurgery Presentation
Tomosurgery PresentationTomosurgery Presentation
Tomosurgery Presentationngowdar
 

Similar to Session 4 plan verification prostate (20)

ICRU 83
ICRU 83ICRU 83
ICRU 83
 
Session 9 radiation oncology
Session 9 radiation oncologySession 9 radiation oncology
Session 9 radiation oncology
 
IMRT by Musaib Mushtaq.ppt
IMRT by Musaib Mushtaq.pptIMRT by Musaib Mushtaq.ppt
IMRT by Musaib Mushtaq.ppt
 
F04433538
F04433538F04433538
F04433538
 
IMRT in Head & Neck Cancer
IMRT in Head & Neck CancerIMRT in Head & Neck Cancer
IMRT in Head & Neck Cancer
 
C04631115
C04631115C04631115
C04631115
 
ICRU 83 report on dose prescription in IMRT
ICRU 83 report on dose prescription in IMRTICRU 83 report on dose prescription in IMRT
ICRU 83 report on dose prescription in IMRT
 
Session 7 interventional radiology
Session 7 interventional radiologySession 7 interventional radiology
Session 7 interventional radiology
 
Comparative dosimetry of forward and inverse treatment planning for Intensity...
Comparative dosimetry of forward and inverse treatment planning for Intensity...Comparative dosimetry of forward and inverse treatment planning for Intensity...
Comparative dosimetry of forward and inverse treatment planning for Intensity...
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62
 
P14_BRAKHITERAPI.pptx
P14_BRAKHITERAPI.pptxP14_BRAKHITERAPI.pptx
P14_BRAKHITERAPI.pptx
 
Session 6 image quality ct and session 8 dose ct
Session 6  image quality ct and  session 8 dose ctSession 6  image quality ct and  session 8 dose ct
Session 6 image quality ct and session 8 dose ct
 
introduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapyintroduction to Intensity modulated radiation therapy
introduction to Intensity modulated radiation therapy
 
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. SachinIntensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
 
Radiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precisionRadiobiological aspects of radiotherapy precision
Radiobiological aspects of radiotherapy precision
 
Brain Tumor Detection using Clustering Algorithms in MRI Images
Brain Tumor Detection using Clustering Algorithms in MRI ImagesBrain Tumor Detection using Clustering Algorithms in MRI Images
Brain Tumor Detection using Clustering Algorithms in MRI Images
 
Total Body Irradiation (Ttbi) Treatments Using A Tomotherapy Unit
Total Body Irradiation (Ttbi) Treatments Using A Tomotherapy UnitTotal Body Irradiation (Ttbi) Treatments Using A Tomotherapy Unit
Total Body Irradiation (Ttbi) Treatments Using A Tomotherapy Unit
 
Icru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationIcru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniation
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentation
 
Tomosurgery Presentation
Tomosurgery PresentationTomosurgery Presentation
Tomosurgery Presentation
 

More from Centro Nacional de Radioterapia Nora Astorga (10)

Flattening filter Free
Flattening filter FreeFlattening filter Free
Flattening filter Free
 
Session 12 imaging instrumentation (spect) qc
Session 12 imaging instrumentation (spect) qcSession 12 imaging instrumentation (spect) qc
Session 12 imaging instrumentation (spect) qc
 
Session 11 non imaging instrumentation qc
Session 11 non imaging instrumentation qcSession 11 non imaging instrumentation qc
Session 11 non imaging instrumentation qc
 
Session 10 radiation survey of a clinical installation
Session 10 radiation survey of a clinical installationSession 10 radiation survey of a clinical installation
Session 10 radiation survey of a clinical installation
 
Session 3 measure beam profile rt
Session 3 measure beam profile rtSession 3 measure beam profile rt
Session 3 measure beam profile rt
 
Session 2 radiotherapy
Session 2 radiotherapySession 2 radiotherapy
Session 2 radiotherapy
 
Session 1 conventional x ray
Session 1  conventional x raySession 1  conventional x ray
Session 1 conventional x ray
 
Monografia
MonografiaMonografia
Monografia
 
Física de la radiología.
Física de la radiología.Física de la radiología.
Física de la radiología.
 
Icru 62
Icru 62Icru 62
Icru 62
 

Recently uploaded

Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiNehru place Escorts
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Miss joya
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Servicenarwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls ServiceMiss joya
 

Recently uploaded (20)

Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 

Session 4 plan verification prostate

  • 1. Master in Medical Physics ICTP 2015-2016 Radiotherapy treatment verification with Octavius phantom Francisco J.Hernández Flores∗ International Centre for Theoretical Physics franciscohernandez_f2010@hotmail.com August 31, 2015 Abstract The present task is mainly focused on the verification of the calculated dose. During a radiotherapy treatment, a dosimetric verification or a geometric localization can be done, in order to assess the quality of the treatment. The dosimetric verification is generally performed measuring the dose at some points inside or outside the patient, and comparing it to the dose at the same points calculated and predicted by the treatment planning system. This was evaluated with phantom Octavius 4D. Dose distributions were calculated for volumetric modulated arc therapy (VMAT) treatment plans with Monaco the results of gamma analysis was 98% passing criteria. I. Introduction Volumetric Modulated Arc Therapy (VMAT) is a new intensity-modulated radiation therapy (IMRT) technique that improves critical structures and healthy tissue sparing, in dramatically shorter treatment times all without comprising target coverage and patient safety [2] when advanced techniques used in radiotherapy and to finalize a plan of VMAT in the TPS is mandatory dosimetric verification before clinical delivery using diodes or ionization chambers, in our case we used the Octavius phantom to check the effectiveness of the plan dose delivered to the patient and compare the passes of the plan take into account the gamma factor and DTA. OCTAVIUS 4D is a new 4D dosimetry system designed to verify IMRT treatment plans based on true independent measurements. It rotates with the gantry, measuring the dose inside the entire phantom volume, always per pendicular to the incident beam. As a consequence, OCTAVIUS 4D requires no angular corrections or detector calibrations to compensate for the directional response of its detectors which makes it perfectly suited for rotational delivery techniques and easy-to-use in clinical routine. [1] II. Theory Distance To Agreement (DTA) is the distance between a measured data point and the nearest point in the calculated dose distribution that exhibits the same dose [2]. The gamma function evaluates the difference between measured and calculated dose distribution using both dose and spatial coordinates [3]. Γ(rm, rc) = r2(rm, rc) ∆d2 m + δ2(rm, rc) ∆D2 m (1) ∗Radiotherapy Practical 2 1
  • 2. Master in Medical Physics ICTP 2015-2016 with δ(rm, rc) = Dm − Dc; r(rm, rc) = |rm, rc| γ(rm) = min{Γ(rm, rc)}∀{rc} (2) γ(rm) = min r2(rm, rc) ∆d2 m + δ2(rm, rc) ∆D2 m ∀rc (3) γ(rm) function, the minimum generalized Γ(rm, rc) function in the set of evaluated points. ∆dm and ∆Dm the acceptance criteria for distance to dose r(rm, rc) is the distance between the measured point rm and a point rc chosen in calculated dose distribution δ is the corresponding dose difference. As a general rule, acceptance criteria of 3% dose difference and 3 mm distance to agreement. The dose-difference criterion is ∆Dm , and the DTA criterion is ∆dm. III. Material and method Verification of Treatment planning with dose measure using Treatment Planning sistem Monaco, red mosaiq, phantom Octavius 4D (see fig. 1) with respective software, motorized cylindrical phantom specially designed for rotational delivery techniques, includes control unit and wireless inclinometer. The plan used in this practice was calculated, teacher only explain how she made the plan because the practice is only verification plan, she told us how is the procedure for align the phantom and connected all the tools, the software use the equation 3 to calculate the gamma factor for the quantitative evaluation of dose distributions. Figure 1: Phantom Octavius 4D IV. Analysis and discussion of results The rsult obtained during the verification planning treatment was 98% passing criteria based in clinical standard for photon beam. This plan was analyzed using the two parameter DTA (limit 3%), (limit 3 mm), and index with the 3% dose tolerance and 3 mm distance to agreement in relation to the treatment planning system. The range of acceptant pass of the gamma criterion depend anatomical site, Results confirmed a good agreement between the two distribution with high and conformed dose to the target and low dose to the organ at risk. 2
  • 3. Master in Medical Physics ICTP 2015-2016 V. Conclusion • Clinical quality assurance practice based on linac treatment log files for verifi- cation of delivered 3D dose distributions in the patients geometries represents a paradigm shift from dose measurements in a phantom. This approach offers independent verification of VMAT dose distributions encompassing treatment planning beam modeling differences as well as the linac uncertainties during treatment delivery of the plan. • Verification of a planned dose distribution is a complex and time consuming procedure because all described parameters DTA and γ-index are very important and using only one of them is not enough to accept a treatment plan. References [1] Elekta Manual of the phantom Octavius 4D,Viena 2000 [2] Daniel A. Low, William B. Harms, et al A technique for the quantitative evaluation of dose distributions, Division of Radiation Oncology, 510 South Kingshighway Blvd. [3] Michele Avanzo, Lecture of Oncology 3, ICTP, 2015. 3