Basic anesthesia monitoring- Emad Zarief
1. BAS C
EMAD ZARIEF 2019 1
MONITORINGEmad Zarief MD
A. Professor of Anesthesia & ICU dpt.
Faculty of Medicine , Assiut University
School of medicine , Windhoek
2. Monitor is a Latin word “to warn”
... interpret available clinical data to help recognize present or
future mishaps or unfavorable system conditions
EMAD ZARIEF 2019 2
BAS C
3. ❖To maintain the normal patient physiology &
homeostasis To ensure the well being of the pt.
❖Surgery or trauma is a very stressful condition →
severe sympathetic stimulation, HTN, tachycardia,
arrhythmias.
EMAD ZARIEF 2019 3
WHY?
4. ❖Most drugs used for general & regional anaesthesia cause
hemodynamic instability, myocardial depression,
hypotension & arrhythmias.
❖Under GA the pt. may be hypo or hyperventilated and
may develop hypothermia.
❖Blood loss → anaemia, hypotension. So it is necessary to
recognise when the pt. is in need of blood transfusion
(transfusion point).
EMAD ZARIEF 2019 4
WHY?
5. Monitoring in the Past
Visual monitoring of
respiration and overall
clinical appearance
Finger on pulse
Blood pressure
(sometimes)
EMAD ZARIEF 2019 5
6. STANDARD I
Qualified anesthesia personnel shall be present in the room
throughout the conduct of all general anesthetics, regional anesthetics
and monitored anesthesia care.
http://www.asahq.org/publicationsAndServices/standards/02.pdf
EMAD ZARIEF 2019 6
BAS C
7. STANDARD II
Perioperatively , the patient’s :
❖Oxygenation,
❖Ventilation,
❖Circulation
❖Temperature
should be continually evaluated.
http://www.asahq.org/publicationsAndServices/standards/02.pdf
EMAD ZARIEF 2019 7
BAS C
10. Tissue perfusion
▪Capillary refill <5 sec
▪Core temperature <5 diff from surface
▪Gastric mucosa pH
▪Acid base and Lactate
▪Mixed venous SvO2
▪Urine out put
EMAD ZARIEF 2019 10
13. Auto-regulation
means that the blood
flow of an organ
is kept constant
within a specific
range
of pressure .
EMAD ZARIEF 2019 13
AUTOREGULATION
14. ❖Correct cuff size: width of the cuff should be 1.5 times limb
diameter and should occupy at least 2/3 of the arm.
❖2 cuff sizes for adult: blue: for most adult individuals (60-90 Kg),
red: for morbid obese.
❖Selection of appropriate cuff size is important because a tight cuff
leads to false high readings, while a Loose cuff gives false Low
readings.
EMAD ZARIEF 2019 14
How to apply CUFF
22. ◦On bony prominence.
◦AVOID hairy areas
◦Position them far away from each other to give e higher
voltage and better gain.
◦Ensure good contact with the skin: by using KY-Gel.
EMAD ZARIEF 2019 26
23. ◦ QRS beep ON must be heard at all times. NO silent
monitors.
◦ clinical judgement is much more superior to the
monitor. Check peripheral pulsations.
◦ Arrythmias → check radial (peripheral) pulsations.
EMAD ZARIEF 2019 27
24. LEADS
❖How many leads ?
❖Which lead(s) shall we
use for monitoring ?
EMAD ZARIEF 2019 28
43. plethysmography
detect only pulsatile blood waves ,so that not deal with venous or
capillary blood And only reads arterial blood !!
Saturation: using Infra red – and red light waves absorption reflection
EMAD ZARIEF 2019 57
SPO2 = oxy HB / Oxy HB + Deoxy HB
45. 1- No wave → don’t rely upon Spo2
2- you can use this correlation
EMAD ZARIEF 2019 59
P O2 40 mmHg 50 mmHg 60 mmHg
Saturation % 70 80 90
46. RULES:
❖Keep the sound of the pulse oximeter ON at ALL times.
❖clinical judgement is much more superior to the monitor. Check pt
colour for cyanosis: lips, nails.
❖If hypoxemia occurs immediately check the correct position of the probe
on the pt and check the pts colour: nails & lips, then manage accordingly
& CALL 4 HELP.
EMAD ZARIEF 2019 60
47. Fallacies & Inaccuracies occur when:
◦ Misplaced on the pts finger, slipped.
◦ Pt movement, shivering.
◦ Poor tissue perfusion (cold extremities, hypotension & shock).
◦ Carboxy HB , Mythelene blue … etc
◦ Sometimes by electrical interference from cautery in some monitors.
EMAD ZARIEF 2019 61
52. Continuous CO2 measurement displayed as a waveform sampled
from the patient’s airway during ventilation.
◦What is EtCO2?
A point on the capnogram. It is the final measurement at the
endpoint of the pts expiration before inspiration begins again. It
is usually the highest CO2 measurement during ventilation.
EMAD ZARIEF 2019 66
59. LISTEN
Listen to the monitor the whole time:
◦ To the pulse oximeter tone to identify: 1-Heart rate 2- Spo2
from the tone (pitch) of pulse oximeter.
◦ To the sound of the ventilator, to any abnormal sounds, any alarms.
RULE: NO silent monitors. ALWAYS keep the HR sound on. If ur
monitor is silent (sound is not working) u have to look at your
monitor the WHOLE time.
EMAD ZARIEF 2019 73