SlideShare a Scribd company logo
1 of 23
Download to read offline
Does EMG Noise Affect the Readings of Heart Rate Monitors?
Sistania Bong, Andrew Choe, Lauren Gardner, Hsiang Hsu, Nikki Jackson, Tyler Rice,
Malvika Sanghvi, Jin Eun Shin
Facilitators: Dr. Julia Babensee, Rhoades Sturkie
Group C6
Abstract
Objective: This study was carried out to determine whether EMG noise causes an
error in heart rate measurement when using a household heart rate monitor. The Sportline
S7 heart rate monitor watch was the device used to determine the effects. Our null
hypothesis is that the EMG noise will not cause a significant difference in the heart rate
measurements. Our alternative hypothesis is that the EMG noise will significantly
increase the heart rate measurements. Methods: Our procedure will include 16
participants. To have a constant EMG noise we will be using a TENS unit to cause the
muscles to contract. Results: Our results show that there is a significant increase in heart
rate caused by muscle contraction. Our p-value is 1.13745e-0.6, which is lower than our
 value is 0.05. Because of this we are able to reject our null hypothesis. Conclusion: In
conclusion, our first recommendation is that the company should improve the filter used
in the wrist watch that is suppose to filter out the EMG noise caused by muscle
contraction. Our second recommendation is that the user should not be using the device
while exercising, which causes muscle contraction.
Introduction
The Electrical Conductivity of the Heart, Mechanical Events in The Heart, and The
Electrocardiograms
The electrical and mechanical events in the heart correspond to a waveform called
electrocardiogram (ECG)(OpenStax College,2013). This is used to record the electrical
activity of the circulatory system through the use of electrodes placed on several different
locations on the body. The ECG signal is comprised of P, Q, R, S, and T waves which
visually reflects the sum of electrical activity found in the heart. This triggers the heart to
contract and relax, producing heartbeats (Figure 1).
Figure 1: Normal ECGs (Silverthorn, D. U., 2013).
The electrical conductivity in the heart is divided into two major events:
depolarization and repolarization. Depolarization starts at the sinoatrial node (SA node)
in the right atrium of the heart. The SA node is a pacemaker cell, and has the ability to
depolarize the cell by itself without being triggered by the action potential from adjacent
cells. Since the SA node is located in the right atrium, depolarization first occurs in the
right atrium and is followed by the left atrium. This causes the both atria to contract.
Atrial depolarization and contraction is indicated in the P wave (Phase 2 and Phase 3).
The depolarization wave is then transferred to the atrioventicular node (AV node)
through internodal tracts. The AV node is composed of bundle of His, also called the
atrioventricular bundle, which will branches out into two further bundles: the left and
right bundle. The right and left bundles consist of branches called Purkinje fibers. The
right and left bundles consist of branches called Purkinje fibers. The depolarization wave
is then transferred to the right and left ventricle through the Purkinje fibers that then
release electrical signals into ventricular myocardium and depolarize the ventricles. The
depolarization of the myocardium will trigger the ventricles to contract. The events in
which the depolarization wave is transferred to the ventricle and allows for the ventricles
to contract and eject blood out of the heart are represented in the QRS wave (QRS
complex) (Phase 4) and S wave (Phase 5). The atrial repolarization is not specifically
emphasized but incorporated in the starting of the QRS complex. After the blood
ejection, the ventricles start to repolarize which causes the ventricles to relax. The
repolarization of the ventricles followed by the ventricular relaxation is indicated in the T
wave. Lastly, the interval between the T and P wave indicates that there is no electrical
activity in the heart during this brief time because the SA node will begin the
depolarization process over again (Phase1) (OpenStax College,2013) (Silverthorn, D.
U.,2013).
Figure 2: Electrical and Mechanical Events of the Cardiac Cycle (OpenStax College,
2013)
The electrical signal that is detected by the heart rate monitor reflects the change
of cell membrane potential due to the charged ions activities in the cardiac muscle cell.
The changes of the cell membrane potential generate an action potential that enables the
cell to depolarize. There are two types of cardiac muscle cells, myocardial contractile cell
and myocardial autorhythmic cells, each with different action potentials. The action
potential in the myocardial contractile cells (Figure 3) is initiated by the depolarization
wave from the adjacent cells. Because of the wave of depolarization, the membrane
potential becomes more positive. The voltage-gated Na+
channels open, and Na+
ions
diffuse into the cell and depolarize the cell until the membrane potential reaches its peak
at +20mV and Na+
channels close. When the Na+
channels close, the voltage-gated K+
channels open, allowing K+
ions to leave the cell, thus repolarizing the cell. This
repolarization occurs for a brief moment because two events take place: the membrane
permeability of K+
decreases while the membrane permeability of Ca2+
increases. The
voltage-gated Ca2+
channels open and Ca2+
ions enter the cell. The increased of Ca2+
influx and the decreased of K+
efflux causes the action potential to flatten as it is shown
in the diagram (Figure 3). After the plateau, the permeability of K+
increases and
permeability of Ca2+
decreases. The voltage-gated K+
channels open, and K+
ions rapidly
leave the cell repolarizing it back to its resting membrane potential.
The action potential in the myocardial autorhythmic cells (Figure 4) are initially
generated by the SA node in the right atrium. The membrane potential of the SA node
(pacemaker potential) starts at -60mV and depolarizes the cell to threshold at -40mV. As
soon as the cell reaches -40mV, the SA node will fire the action potential and trigger the
voltage-gated Ca2+
channels to open allowing Ca2+
ions to flow in to the cell and
depolarize the cell to the peak of its membrane potential, which is almost +20mV. At the
peak of the membrane potential, the Ca2+
channels close and the voltage-gated K+
channels open. The opening of the K+
channel allows the K+ to exit the cell resulting in
repolarization of the cell back to -60mV and stimulates the SA node to start the
depolarization process all over again.
Figure 3: Action Potential of a Cardiac Contractile Cell
Figure 4: Action Potentials in Cardiac Autorhythmic Cells (Silverthorn, D.
U.,2013)
Muscle Contractions (Electromyographic Interference / EMG Noise)
The electromyographic (EMG) signal is associated with the electrical activity in
the muscle fibers that reflects muscle contraction and relaxation. EMG signals have a
high potential of corrupting ECG signals. This is called electromyographic intereference
or EMG noise. According to the U.S. Patent for S-Pulse Technology wristband heart rate
monitor, EMG noise potentially interferes with the ECG signal due to the similarity of
EMG signal characteristics with the characteristics of the ECG signal. EMG noise falls
into the same range of frequency as the ECG signal (Lo, T. Y.-C., & Tsai, Y. S., 1998).
Consequently, EMG noise overlaps the frequency spectrum of the QRS complex
(Friesen, G. M., et al, 1990).
Figure 5: ECG Corrupted with EMG (Friesen, G. M., et al, 1990)
The Device: Sportline S7 Heart Rate Watch
Sportline S7 Heart Rate Watch is an Any Touch Heart Rate Monitor,
commercially used to track the intensity of workouts. It is essentially like being
connected to a low-cost, portable ECG/EKG. It is based on Any-touch technology, a
patented technology of the Sportline company, which makes use of S-PULSE
technology. Further, it also helps in monitoring calorie burn and the fat-burning zone.
Mechanism of the Device
The digital watch employs a three contact approach, in which two of the electrical
contacts are positioned on the front of the watch, where the user places his fingers, while
the third contact is at the back of the watch. All three contacts are connected to a differential
amplifier. The heartbeat signal is picked up when the front contacts are brought in contact
with the user’s touch and then made to pass through the differential amplifier. This
amplifier amplifies signals while also suppressing common mode noise ( unwanted input
signals). Referring to Figure 6 the differential amplifier is represented by 68. Next, the
output of the differential amplifier enters the pass-band filter (#64). This filter consists of
the following two types of filters, which essentially filter out noises above and below the
frequency range in which the desired heartbeat rate will lie:
1. Low Pass Filter –which has a threshold of 25-40 Hz; all signals with frequencies that
fall in that range get filtered
2. High Pass Filter-which has a threshold of 5-15 Hz; all signals with frequencies that
fall in this range get filtered.
Next, the analog signal is amplified in amplifier 66 dB that has a gain of 50-1000 dB so
that the overall gain is about 1000-10,000 dB. Finally, after filtering the signals so that they
may be refined, the signals will be eliminated as much noise as possible. Then the analog
output of the amplifier 66 is applied to the input of the analog-to-digital converter (68),
which is integrated onto the microcontroller integrated circuit. From here, the digital
signals will be input into the microcontroller (#70) for further signal processing. The digital
samples are then filtered to further remove remnants of frequencies above and below the
range of frequencies in which heartbeat will lie. This is done to suppress power-line hum at
approximately 50-60 Hz so as to generate filtered digital samples. Finally, the signals are
subject to the enhancement signal processor. This processor enhances heartbeat peaks in
the filtered digital samples to generate enhanced digital samples. It comprises of 3 different
units:
1. Differentiator : Determines the slope of peaks in the filtered data and generates a
slope signal which defines the magnitudes and signs of the slopes of each portion
of each peak.
2. The squaring processor : Squares the results from the differentiator by looking up
results in a lookup table that shows the squares of possible values that could be
output from the differentiator
3. The moving average processor : Computes the moving average of the positive
values signal and outputting a moving average signal which defines the moving
average over time. Finally, the enhanced digital samples are again processed to
determine the individual’s heartbeat rate.
4.
Figure 6: Detailed Flowchart of the Mechanics of the Heart Rate Monitor
Figure 7: Detailed Flowchart of the Enhancement Signal Processor
Figure 8: Sportline S7 Heart Rate Monitor
Transcutaneous Electrical Nerve Stimulation (TENS)
For our experiment we wanted to create a constant EMG regulation and eliminate
any experimental variations, such as the difference of muscle contraction by using stress
balls. In order to achieve the regulation, we selected a supplementary device that was
provided by one of our group members, Omron HV-F128.
The HV-F128 uses the TENS technology to massage the muscles and relieve the
pain. The TENS technology works similar to how an electronic muscle stimulator (EMS)
does. Both stimulators generate electrical impulses that stimulate the targeting nerves
through the skin, which in turn cause the muscles controlled by those nerves to react and
contract (Jones, I., & Johnson, M. I., 2009). However, the only difference between EMS
and TENS is their targeting nerves. While EMS are designed to stimulate muscle motor
nerves, TENS devices are designed to stimulate sensory nerve endings. Even though the
TENS device targets to stimulate the sensory nerves, it also stimulates the muscle motor
nerves, which lie near the sensory nerves, and cause the muscles to contract passively.
The TENS device, HV-F128, makes muscle contract passively and generate a constant
EMG noise that we desire.
From the user manual, we obtained that the frequency HV-F128 generates is 1 ~
1200Hz and the power consumption is about 85 mA. Maximum output voltage is less or
equal to 90V and maximum output current is less or equal to 10 mA (during 1kilo-omega
load).
Method and Materials
Materials
This experiment required the use of two separate material sets: materials to
prepare the participant and devices for reading heart rate and stimulating the participant’s
forearm.
To perform the study, the participant was sprayed with one spray bottle of tap
water and was then wiped down with sterile cotton balls. This adhered to our proper
method of preparing the participant. In terms of devices used for reading heart rate and
stimulating muscle contraction, the Sportline S7 heart rate watch and the Omron HV-
F128 TENS device were used. The Sportline S7 watch uses one-touch technology and
was purchased at Walmart (Walmart.com, 2013). The Omron HV-F128 massager was
imported from Japan (Omron Healthcare Co., Ltd.), and the English manual was found at
Omron-healthcare.com.
Preparation of the Participant
The participant was prepared by being notified of the exclusion criteria of the
experiment. Participants with metal implants or pacemakers and those who were not
members of the Georgia Institute of Technology’s BMED 1300 course were asked to not
take part in the study. Eligible candidates to be participants were given a consent form
that outlined the risks and benefits of participating in this study. Once consent was given,
the participant gained entrance to the testing area where they were asked to roll up their
sleeves (if necessary) and expose their left forearm. From here, the experimenter sprayed
the wrist of the participant and wiped the area with a cotton ball to wipe off any materials
that would affect the conductivity of participant’s wrist and forearm.
Control Measurement
The participant was asked to take a control measurement. The experimenter
placed the Sportline S7 wristwatch on the left wrist of the participant, and the participant
was asked to place their index and middle finger of their right hand onto the device’s
touch sensor. The experimenter recorded this heart rate measurement (heartbeats/minute)
in a secure document for analysis at a later point. The watch was then reset using a reset
button on the side of the watch.
Experimental Measurement
The watch remained on the participant as a second modified measurement was
taken. The experimenter applied the two silicone pads of the TENS device onto the
forearm of the participant. Using the device settings (tap-mode, 35 Hz), the experimenter
turned on the TENS device massager and set the intensity to a level of four. After twenty-
five seconds, the participant again was asked to place their index and middle finger of
their right hand onto the device’s touch sensor. The experimenter then recorded the heart
rate (heartbeats/minute) and removed the watch and massager from the participant’s left
arm. This concluded the experiment, and the participant was released.
Statistical analysis
Based on the literature previously described, the sample size of this experiment
used was 16. This was found by using a power of .97, an effect size of .95, and a level of
significance or alpha of 0.05 when using a statistical program called G*Power (Faul,
Erdfelder, Lang & Buchner, 2007).
To determine the significance of the difference between the control and
experimental heart rate measurements, a one-tailed matched pairs t-test was used. Alpha
was set to 0.05 and p values smaller than alpha were considered significant.
Results
Based on the results our raw data in Fig. 1 ( see Appendix) of the 16 participants
of the study, the P-value, the mean differences, the standard deviation of differences and
standard error of differences between the experimental group and control group were
found to be 1.1375e-6, 63.6875, 34.4978, and 8.6244, respectively as shown in the table 1
(see Appendix for formulas).We decided to use a one-sided matched pairs t-test to
analyze our data, as we needed a good test for a smaller samples size (<30) with a large
difference. We used the match-pairs t-test because the same participant was used for both
the control and experimental trials. We used the one-sided t-test based on the one-sided,
positive results we found in the research. In addition, our results also further endorse our
assumption for the one-sided t-test with a positive, one-sided trend.
The box plot and graph 2 depicts that our data was slightly skewed to the left
based on the five number summary in reference to the mean. Table 2 shows the five
number summary: median, minimum, maximum, first quartile, and third quartile which
were, respectively, 72, 2, 110, 45.75, and 92.5. When the mean of 63.6875 lies to the left
of the median of 72, the bulk of the points are to the right of the center, giving a left
skew. The interquartile range of the box plot can be easily seen in the histogram where
the dense region occurs (heart rate between ~45.75 and ~92.5). The histogram uses
intervals of 2 to receive the most normal-looking distribution. The general appearance of
a normal distribution in our data is shown as a dotted line in the graph. Our Q-Q plot,
Fig.3 (see Appendix), had a r2
value of .96, giving a high correlation with little variance
from the trend line and therefore, a fairly normal distribution.
Graph 3 shows that the control mean without EMG noise, is almost half of the
experimental mean with EMG noise. The significance level of <.05 is denoted with ***.
Our error bars represent standard error.
Graph 4 shows the p-value, the area under the curve in the direction of the hypothesis, in
our case, to the right of the t-statistic. The p-value is the fraction of the total area under
the curve where the null hypothesis is correct. The blue region is the area to the right of
the t-score and the red region represents the area of our p-value (our p-value is too small
to be graphically visible). Our p-value was too small for the program we used to
calculate the graph (the lower threshold on this program is p=.0001). We are able to
reject the null hypothesis with 99.99988625% ((1-p)X100%) confidence.
One-tailed Matched
Pairs t-test
p-value 1.1275e-6
Standard Deviation of
Differences
34.4978
Standard Error of
Differences
8.6244
Variance of Differences 1190.096
Mean of Differences 63.6875
t 7.384536226
Table 1
Population Size 16
Median 72
Minimum 2
Maximum 110
First Quartile 45.75
Third Quartile 92.5
Interquartile
Range
46.75
Outerliers None
Table 2
Graph 1: Box and Whisker Plot
Graph 2: Histogram
Graph 3: Mean Change
Graph 4: T-Score (Hypothesis Test Graph Generator)
Discussion
After the data was all collected, we noticed a few trends in the data. All of the
experimental points were either reoccurring values of 150, 152, or 189, with the
exception of three lower points with the values of 68, 75, and 79; while the control values
were almost completely unique. These reoccurring values could have been explained by
error in the TENS device or the wristband. One potential error we considered in using the
TENS device was that the frequency, translating close to that of ECG signal and
hypothesized EMG noise, could have directly influenced the heart rate reading. We could
only standardize the TENS device within a frequency range, so the slight change in those
recurring frequencies could have been just that, for example the heart rate monitor could
have picked up the electrical signal directly from the TENS device, and not from the
actual EMG noise created by skeletal muscles from the TENS stimulation, which is what
we were attempting to mimic. Another concern with the use of the TENS device was
whether or not the device stimulated only skeletal muscles or also increased the heart rate
by stimulating the heart muscle as well. In the case that the TENS device stimulated the
actual heart muscle, the heart rate monitor could have given a true reading without error.
Although we did not collect data to calibrate the actual heart rate, the participants stayed
in a sitting position throughout the experiment and no one showed any sign of increased
heart rate while taking the second reading. Had a participant’s heart rate actually
increased by about 70 beats per minute in approximately 10 seconds (about the time in
between readings) , physical changes would be apparent.
Another potential explanation for the three reoccurring values could have been the
wristband and its filtering systems. It may be possible that over a certain frequency
reading, over the normal range, the watch may error. There could be a chance that the
filters could error and display randomly those three values. This seemed very unlikely,
however, as the device is meant for exercise and values around 150 beats/minute are not
completely out of the question while exercising. The smaller three experimental values
also struck curiosity when analyzing the data. When taking these readings, we noticed the
participants complained of not feeling the TENS massager’s stimulation. These
participants also had more hair on their arms, which could have led to an error in
applying the muscle stimuli. Though the smaller values did skew the data slightly left,
they were still positive differences and were not calculated as outliers. Some of the
smaller increases show a more realistic change for heart rate in about a 10 second
interval, but considering that the participants did not seem to be affected by the TENS
device, the increase was most likely due to other variables such as anxiety from the test.
Regardless, the use of a TENS device caused error in the wristband monitor, increasing
the heart rate readings above the actual level.
Conclusion
According to articles and patents we researched, the EMG noise which is
produced by muscle contractions significantly influences the heart rate that is measured
by our device due to the one of the characteristics of EMG signals, which is EMG noise
falls in the same range of frequency of the ECG signal. Using the data gathered in our
experiment, we conducted statistical analyses, and found our p value to be 1.1375e-6.
Because the p value was substantially lower than our value (0.05), we rejected the null
hypothesis in favor of alternative hypothesis, therefore supporting the idea that EMG
noise causes a significant increase in heart rate measurements. The origin of the error,
which is EMG interference that causes the device to give a less accurate heart rate
reading is the limitation of the learning process in the mechanism to suppress the noise.
Because the learning process of our device has a low threshold, it is set to allow for other
signals like the EMG noise to be considered as heart rates. Therefore, we recommend that
the producer of this device increases the ability of the learning processor to better
distinguish ECG signals from any other ECG-like noise such as EMG noise and to
improve the sensitivity of the filter to suppress any unwanted noise.
Citations
American Heart Association (8 august 2013). Target Heart Rates. Retrieved 27
september, 2013, from
http://www.heart.org/HEARTORG/GettingHealthy/PhysicalActivity/Target-
Heart-Rates_UCM_434341_Article.jsp
Burke, M. J., & Whelan, M. V. (1987). The Accuracy and Reliability of Commercial
Heart Rate Monitors. Brit.J.Sports Med., 21(1), 29-32.
Cincinnatichildrens. (2012, 04/2012). Electromyogram / EMG and Nerve Conduction
Test. Retrieved 10/31/2013, 2013, from
http://www.cincinnatichildrens.org/health/e/emg/
Drews, C. (2000). Electromyography: Recording Electrical Signals from Human Muscle.
http://ableweb.org/volumes/vol-21/12-drewes.pdf
Friesen, G. M., Jannett, T. C., Jadallah, M. A., Yates, S. L., Quint, S. R., & Nagle,
H. T. (1990). A comparison of the Noise Sensitivity of Nine QRS Detection
Algorithms. IEEE Transactions On Biomedical Engineering, 37(1).
Faul, F., Erdfelder, E., Lang, A., & Buchner, A. (2007, December 01). G*power 3.
Retrieved from http://www.psycho.uni-duesseldorf.de/abteilungen/aap/gpower3
Hypothesis Test Graph Generator. Hypothesis Test Graph Gnerator. Retrieved from
http://www.imathas.com/stattools/norm.html
Johnson, M. (2007). Transcutaneous Electrical Nerve Stimulation: Mechanisms, Clinical
Application and Evidence. British Journal of Pain, 1(1), 7-11. doi:
10.1177/204946370700100103
Lo, T. Y.-C., & Tsai, Y. S. (1998). The United States Patent No. 5,738,104.
Morris, V. Box and Whisker Plot Maker. Math Warehouse. (2013) Retrieved from
http://www.mathwarehouse.com/charts/box-and-whisker-plot-
maker.php#boxwhiskergraph
Patrick S. Hamilton, W. J. T. (1986). Quantitative Investigation of QRS Detection Rules
Using the MIT/BIH Arrhythmia Database. IEEE Transactions of Biomedical
Engineering, 33(12).
Science, A. F. S. (2013). ECG Accurate Pulse® Strapless Heart Rate Monitor
Sportswatches. Retrieved, 30 September 2013, from
http://www.aussiefitsport.com.au/wp-content/uploads/2010/12/AFSS-
PulseQTGeneralInformation.pdf
Silverthorn, D. U. (2013). Human Physiology: An Integrated Approach (6 ed.).Boston:
Pearson Education.Print
Villasenor, J. F. (2009). How to Create a Heart Rate Monitor and One-Lead EKG.
Freescale Technology Forum.
Jones, I., & Johnson, M. I. (2009). Transcutaneous electrical nerve stimulation. Education
in Anaesthesia, Critical Care, & Pain, 9(4)
Sportsline, Inc. (2006). SPORTLINE Solo 910 Heart Rate Watch. Retrieved October
2013, from: http://www.sportline.com/manuals/SP3637BK.pdf
Thomas Ying-Ching Lo, Y. S. T. (1998). United States of America Patent No. US
5738104A
Young, D. K. (n.d.). Statistics Formula Sheet. University of Surrey. Retrieved October
16, 2013, from http://personal.maths.surrey.ac.uk/st/K.Young/form_sheet.pdf
Walmart.com. Sportline S7 Any Touch Heart Rate Monitor Watch -Walmart.com.
Retrieved Oct 2013, from http://www.walmart.com/ip/Sportline-S7-Any-Touch-
Heart-Rate-Monitor-Watch-Black/21672223
Appendix
Subject Number Control Reading TENS Reading Difference
1 70 150 80
2 92 189 97
3 77 152 75
4 79 189 110
5 74 79 5
6 95 189 94
7 105 150 45
8 73 75 2
9 96 189 93
10 86 152 66
11 84 150 66
12 61 152 91
13 102 150 48
14 79 150 71
15 79 152 73
16 65 68 3
Table 3: Raw data
Formulas:
Standard deviation: s = where the mean is mean of our data
Mean:
Standard Error:
t-statistic:
t-score: This was found using the t-table in the appendix. The t-score was taken at the .05
alpha level with 15 degrees of freedom.
P value: P-value was calculated in excel as a 2 array, matched pairs, one-sided TTEST.
(Young, K.)
Graph 5: Bar graph of plotted differences
Graph 6: Q-Q plot of data

More Related Content

What's hot

ECG Machine
ECG MachineECG Machine
ECG Machinemans4ani
 
Chapter9 safety
Chapter9 safetyChapter9 safety
Chapter9 safetyVin Voro
 
Introduction to ECG
Introduction to ECGIntroduction to ECG
Introduction to ECGdelkol
 
Ecg basics electric activity within heart
Ecg basics electric activity within heartEcg basics electric activity within heart
Ecg basics electric activity within heartToufiqur Rahman
 
BASIC ECG
BASIC ECG BASIC ECG
BASIC ECG ctayue80
 
topic ECG and heart diseases+treatments
topic ECG and heart diseases+treatmentstopic ECG and heart diseases+treatments
topic ECG and heart diseases+treatmentsmariarani2
 
ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)Anwar Siddiqui
 
ECG-Electrocardiography
ECG-ElectrocardiographyECG-Electrocardiography
ECG-ElectrocardiographyAnsho Anto
 
Electrocardiography (ECG or EKG)
Electrocardiography (ECG or EKG)Electrocardiography (ECG or EKG)
Electrocardiography (ECG or EKG)Milan Taradi
 
Cardiac electrophysiology part ii lecture 4-1
Cardiac electrophysiology part ii lecture 4-1Cardiac electrophysiology part ii lecture 4-1
Cardiac electrophysiology part ii lecture 4-1Department Of Environment
 
10.31.08: Physiological Basis of ECG
10.31.08: Physiological Basis of ECG10.31.08: Physiological Basis of ECG
10.31.08: Physiological Basis of ECGOpen.Michigan
 
ECG ARTIFACTS AND PITFALLS
ECG ARTIFACTS AND PITFALLS  ECG ARTIFACTS AND PITFALLS
ECG ARTIFACTS AND PITFALLS Padmanabh Kamath
 

What's hot (19)

ECG Machine
ECG MachineECG Machine
ECG Machine
 
Chapter9 safety
Chapter9 safetyChapter9 safety
Chapter9 safety
 
Introduction to ECG
Introduction to ECGIntroduction to ECG
Introduction to ECG
 
Ecg basics electric activity within heart
Ecg basics electric activity within heartEcg basics electric activity within heart
Ecg basics electric activity within heart
 
BASIC ECG
BASIC ECG BASIC ECG
BASIC ECG
 
topic ECG and heart diseases+treatments
topic ECG and heart diseases+treatmentstopic ECG and heart diseases+treatments
topic ECG and heart diseases+treatments
 
ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)
 
Basics of ECG
Basics of ECGBasics of ECG
Basics of ECG
 
ECG
ECGECG
ECG
 
ECG-Electrocardiography
ECG-ElectrocardiographyECG-Electrocardiography
ECG-Electrocardiography
 
Cardiac electrophysiology part ii lecture 4
Cardiac electrophysiology part ii lecture 4Cardiac electrophysiology part ii lecture 4
Cardiac electrophysiology part ii lecture 4
 
Electrocardiography (ECG or EKG)
Electrocardiography (ECG or EKG)Electrocardiography (ECG or EKG)
Electrocardiography (ECG or EKG)
 
Ecg ppt
Ecg pptEcg ppt
Ecg ppt
 
Basics of-ekg-1219758859115951-9
Basics of-ekg-1219758859115951-9Basics of-ekg-1219758859115951-9
Basics of-ekg-1219758859115951-9
 
Cardiac electrophysiology part ii lecture 4-1
Cardiac electrophysiology part ii lecture 4-1Cardiac electrophysiology part ii lecture 4-1
Cardiac electrophysiology part ii lecture 4-1
 
10.31.08: Physiological Basis of ECG
10.31.08: Physiological Basis of ECG10.31.08: Physiological Basis of ECG
10.31.08: Physiological Basis of ECG
 
ecg machine
ecg machineecg machine
ecg machine
 
Lecture 3 cardiac electrophysiology part i
Lecture 3   cardiac electrophysiology part iLecture 3   cardiac electrophysiology part i
Lecture 3 cardiac electrophysiology part i
 
ECG ARTIFACTS AND PITFALLS
ECG ARTIFACTS AND PITFALLS  ECG ARTIFACTS AND PITFALLS
ECG ARTIFACTS AND PITFALLS
 

Similar to Does EMG Noise Affect the Readings of Heart Rate Monitors?

ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAP
ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAPANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAP
ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAPMAHESH SONI
 
3rd Year Project - Design and build of an Electrocardiogram
3rd Year Project - Design and build of an Electrocardiogram3rd Year Project - Design and build of an Electrocardiogram
3rd Year Project - Design and build of an ElectrocardiogramM. Yahia Al Kahf
 
Physiological Basis of Electrocardiogram
Physiological Basis of ElectrocardiogramPhysiological Basis of Electrocardiogram
Physiological Basis of ElectrocardiogramUDUAKABASI JAMES
 
Denoising of ECG -- A discrete time approach using DWT
Denoising of ECG -- A discrete time approach using DWTDenoising of ECG -- A discrete time approach using DWT
Denoising of ECG -- A discrete time approach using DWTIJERA Editor
 
Denoising of ECG -- A discrete time approach using DWT.
Denoising of ECG -- A discrete time approach using DWT.Denoising of ECG -- A discrete time approach using DWT.
Denoising of ECG -- A discrete time approach using DWT.IJERA Editor
 
ECG, step by step approach (Updated)
ECG, step by step approach (Updated)ECG, step by step approach (Updated)
ECG, step by step approach (Updated)Kerolus Shehata
 
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORMDENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORMIJEEE
 
Heart physiology, blood physiology
Heart physiology, blood physiologyHeart physiology, blood physiology
Heart physiology, blood physiologyMaruMengeshaWorku18B
 
ECG FBISE HSSC-I Chap. 12 XI - Biology
ECG FBISE HSSC-I Chap. 12 XI - BiologyECG FBISE HSSC-I Chap. 12 XI - Biology
ECG FBISE HSSC-I Chap. 12 XI - Biologyejaz khichi
 
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Manu Mitra
 
Electrocardiogram (ECG) / ECG interpretation
Electrocardiogram (ECG) / ECG interpretationElectrocardiogram (ECG) / ECG interpretation
Electrocardiogram (ECG) / ECG interpretationRahul Jaga
 

Similar to Does EMG Noise Affect the Readings of Heart Rate Monitors? (20)

ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAP
ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAPANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAP
ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAP
 
3rd Year Project - Design and build of an Electrocardiogram
3rd Year Project - Design and build of an Electrocardiogram3rd Year Project - Design and build of an Electrocardiogram
3rd Year Project - Design and build of an Electrocardiogram
 
Physiological Basis of Electrocardiogram
Physiological Basis of ElectrocardiogramPhysiological Basis of Electrocardiogram
Physiological Basis of Electrocardiogram
 
Ecg part i
Ecg part iEcg part i
Ecg part i
 
Ix3515341540
Ix3515341540Ix3515341540
Ix3515341540
 
Denoising of ECG -- A discrete time approach using DWT
Denoising of ECG -- A discrete time approach using DWTDenoising of ECG -- A discrete time approach using DWT
Denoising of ECG -- A discrete time approach using DWT
 
Denoising of ECG -- A discrete time approach using DWT.
Denoising of ECG -- A discrete time approach using DWT.Denoising of ECG -- A discrete time approach using DWT.
Denoising of ECG -- A discrete time approach using DWT.
 
Basic ECG notes
Basic ECG notesBasic ECG notes
Basic ECG notes
 
ECG, step by step approach (Updated)
ECG, step by step approach (Updated)ECG, step by step approach (Updated)
ECG, step by step approach (Updated)
 
ECG by Adil.pptx
ECG by Adil.pptxECG by Adil.pptx
ECG by Adil.pptx
 
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORMDENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
 
ecg.pptx
ecg.pptxecg.pptx
ecg.pptx
 
Lec51()
Lec51()Lec51()
Lec51()
 
Heart physiology, blood physiology
Heart physiology, blood physiologyHeart physiology, blood physiology
Heart physiology, blood physiology
 
ECG
ECGECG
ECG
 
Ecg interpretation
Ecg interpretation Ecg interpretation
Ecg interpretation
 
ECG FBISE HSSC-I Chap. 12 XI - Biology
ECG FBISE HSSC-I Chap. 12 XI - BiologyECG FBISE HSSC-I Chap. 12 XI - Biology
ECG FBISE HSSC-I Chap. 12 XI - Biology
 
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
 
Af4103186190
Af4103186190Af4103186190
Af4103186190
 
Electrocardiogram (ECG) / ECG interpretation
Electrocardiogram (ECG) / ECG interpretationElectrocardiogram (ECG) / ECG interpretation
Electrocardiogram (ECG) / ECG interpretation
 

Recently uploaded

Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxupamatechverse
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Analog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog ConverterAnalog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog ConverterAbhinavSharma374939
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130Suhani Kapoor
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingrakeshbaidya232001
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxwendy cai
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Dr.Costas Sachpazis
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxDeepakSakkari2
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024hassan khalil
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxupamatechverse
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerAnamika Sarkar
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSCAESB
 

Recently uploaded (20)

Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptx
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
Analog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog ConverterAnalog to Digital and Digital to Analog Converter
Analog to Digital and Digital to Analog Converter
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writing
 
What are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptxWhat are the advantages and disadvantages of membrane structures.pptx
What are the advantages and disadvantages of membrane structures.pptx
 
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINEDJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptx
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptx
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentation
 

Does EMG Noise Affect the Readings of Heart Rate Monitors?

  • 1. Does EMG Noise Affect the Readings of Heart Rate Monitors? Sistania Bong, Andrew Choe, Lauren Gardner, Hsiang Hsu, Nikki Jackson, Tyler Rice, Malvika Sanghvi, Jin Eun Shin Facilitators: Dr. Julia Babensee, Rhoades Sturkie Group C6 Abstract Objective: This study was carried out to determine whether EMG noise causes an error in heart rate measurement when using a household heart rate monitor. The Sportline S7 heart rate monitor watch was the device used to determine the effects. Our null hypothesis is that the EMG noise will not cause a significant difference in the heart rate measurements. Our alternative hypothesis is that the EMG noise will significantly increase the heart rate measurements. Methods: Our procedure will include 16 participants. To have a constant EMG noise we will be using a TENS unit to cause the muscles to contract. Results: Our results show that there is a significant increase in heart rate caused by muscle contraction. Our p-value is 1.13745e-0.6, which is lower than our  value is 0.05. Because of this we are able to reject our null hypothesis. Conclusion: In conclusion, our first recommendation is that the company should improve the filter used in the wrist watch that is suppose to filter out the EMG noise caused by muscle contraction. Our second recommendation is that the user should not be using the device while exercising, which causes muscle contraction. Introduction The Electrical Conductivity of the Heart, Mechanical Events in The Heart, and The Electrocardiograms The electrical and mechanical events in the heart correspond to a waveform called electrocardiogram (ECG)(OpenStax College,2013). This is used to record the electrical
  • 2. activity of the circulatory system through the use of electrodes placed on several different locations on the body. The ECG signal is comprised of P, Q, R, S, and T waves which visually reflects the sum of electrical activity found in the heart. This triggers the heart to contract and relax, producing heartbeats (Figure 1). Figure 1: Normal ECGs (Silverthorn, D. U., 2013). The electrical conductivity in the heart is divided into two major events: depolarization and repolarization. Depolarization starts at the sinoatrial node (SA node) in the right atrium of the heart. The SA node is a pacemaker cell, and has the ability to depolarize the cell by itself without being triggered by the action potential from adjacent cells. Since the SA node is located in the right atrium, depolarization first occurs in the right atrium and is followed by the left atrium. This causes the both atria to contract. Atrial depolarization and contraction is indicated in the P wave (Phase 2 and Phase 3). The depolarization wave is then transferred to the atrioventicular node (AV node) through internodal tracts. The AV node is composed of bundle of His, also called the atrioventricular bundle, which will branches out into two further bundles: the left and right bundle. The right and left bundles consist of branches called Purkinje fibers. The right and left bundles consist of branches called Purkinje fibers. The depolarization wave is then transferred to the right and left ventricle through the Purkinje fibers that then release electrical signals into ventricular myocardium and depolarize the ventricles. The depolarization of the myocardium will trigger the ventricles to contract. The events in which the depolarization wave is transferred to the ventricle and allows for the ventricles
  • 3. to contract and eject blood out of the heart are represented in the QRS wave (QRS complex) (Phase 4) and S wave (Phase 5). The atrial repolarization is not specifically emphasized but incorporated in the starting of the QRS complex. After the blood ejection, the ventricles start to repolarize which causes the ventricles to relax. The repolarization of the ventricles followed by the ventricular relaxation is indicated in the T wave. Lastly, the interval between the T and P wave indicates that there is no electrical activity in the heart during this brief time because the SA node will begin the depolarization process over again (Phase1) (OpenStax College,2013) (Silverthorn, D. U.,2013). Figure 2: Electrical and Mechanical Events of the Cardiac Cycle (OpenStax College, 2013) The electrical signal that is detected by the heart rate monitor reflects the change of cell membrane potential due to the charged ions activities in the cardiac muscle cell. The changes of the cell membrane potential generate an action potential that enables the cell to depolarize. There are two types of cardiac muscle cells, myocardial contractile cell
  • 4. and myocardial autorhythmic cells, each with different action potentials. The action potential in the myocardial contractile cells (Figure 3) is initiated by the depolarization wave from the adjacent cells. Because of the wave of depolarization, the membrane potential becomes more positive. The voltage-gated Na+ channels open, and Na+ ions diffuse into the cell and depolarize the cell until the membrane potential reaches its peak at +20mV and Na+ channels close. When the Na+ channels close, the voltage-gated K+ channels open, allowing K+ ions to leave the cell, thus repolarizing the cell. This repolarization occurs for a brief moment because two events take place: the membrane permeability of K+ decreases while the membrane permeability of Ca2+ increases. The voltage-gated Ca2+ channels open and Ca2+ ions enter the cell. The increased of Ca2+ influx and the decreased of K+ efflux causes the action potential to flatten as it is shown in the diagram (Figure 3). After the plateau, the permeability of K+ increases and permeability of Ca2+ decreases. The voltage-gated K+ channels open, and K+ ions rapidly leave the cell repolarizing it back to its resting membrane potential. The action potential in the myocardial autorhythmic cells (Figure 4) are initially generated by the SA node in the right atrium. The membrane potential of the SA node (pacemaker potential) starts at -60mV and depolarizes the cell to threshold at -40mV. As soon as the cell reaches -40mV, the SA node will fire the action potential and trigger the voltage-gated Ca2+ channels to open allowing Ca2+ ions to flow in to the cell and depolarize the cell to the peak of its membrane potential, which is almost +20mV. At the peak of the membrane potential, the Ca2+ channels close and the voltage-gated K+ channels open. The opening of the K+ channel allows the K+ to exit the cell resulting in
  • 5. repolarization of the cell back to -60mV and stimulates the SA node to start the depolarization process all over again. Figure 3: Action Potential of a Cardiac Contractile Cell Figure 4: Action Potentials in Cardiac Autorhythmic Cells (Silverthorn, D. U.,2013) Muscle Contractions (Electromyographic Interference / EMG Noise) The electromyographic (EMG) signal is associated with the electrical activity in the muscle fibers that reflects muscle contraction and relaxation. EMG signals have a high potential of corrupting ECG signals. This is called electromyographic intereference or EMG noise. According to the U.S. Patent for S-Pulse Technology wristband heart rate
  • 6. monitor, EMG noise potentially interferes with the ECG signal due to the similarity of EMG signal characteristics with the characteristics of the ECG signal. EMG noise falls into the same range of frequency as the ECG signal (Lo, T. Y.-C., & Tsai, Y. S., 1998). Consequently, EMG noise overlaps the frequency spectrum of the QRS complex (Friesen, G. M., et al, 1990). Figure 5: ECG Corrupted with EMG (Friesen, G. M., et al, 1990) The Device: Sportline S7 Heart Rate Watch Sportline S7 Heart Rate Watch is an Any Touch Heart Rate Monitor, commercially used to track the intensity of workouts. It is essentially like being
  • 7. connected to a low-cost, portable ECG/EKG. It is based on Any-touch technology, a patented technology of the Sportline company, which makes use of S-PULSE technology. Further, it also helps in monitoring calorie burn and the fat-burning zone. Mechanism of the Device The digital watch employs a three contact approach, in which two of the electrical contacts are positioned on the front of the watch, where the user places his fingers, while the third contact is at the back of the watch. All three contacts are connected to a differential amplifier. The heartbeat signal is picked up when the front contacts are brought in contact with the user’s touch and then made to pass through the differential amplifier. This amplifier amplifies signals while also suppressing common mode noise ( unwanted input signals). Referring to Figure 6 the differential amplifier is represented by 68. Next, the output of the differential amplifier enters the pass-band filter (#64). This filter consists of the following two types of filters, which essentially filter out noises above and below the frequency range in which the desired heartbeat rate will lie: 1. Low Pass Filter –which has a threshold of 25-40 Hz; all signals with frequencies that fall in that range get filtered 2. High Pass Filter-which has a threshold of 5-15 Hz; all signals with frequencies that fall in this range get filtered. Next, the analog signal is amplified in amplifier 66 dB that has a gain of 50-1000 dB so that the overall gain is about 1000-10,000 dB. Finally, after filtering the signals so that they may be refined, the signals will be eliminated as much noise as possible. Then the analog output of the amplifier 66 is applied to the input of the analog-to-digital converter (68), which is integrated onto the microcontroller integrated circuit. From here, the digital signals will be input into the microcontroller (#70) for further signal processing. The digital samples are then filtered to further remove remnants of frequencies above and below the
  • 8. range of frequencies in which heartbeat will lie. This is done to suppress power-line hum at approximately 50-60 Hz so as to generate filtered digital samples. Finally, the signals are subject to the enhancement signal processor. This processor enhances heartbeat peaks in the filtered digital samples to generate enhanced digital samples. It comprises of 3 different units: 1. Differentiator : Determines the slope of peaks in the filtered data and generates a slope signal which defines the magnitudes and signs of the slopes of each portion of each peak. 2. The squaring processor : Squares the results from the differentiator by looking up results in a lookup table that shows the squares of possible values that could be output from the differentiator 3. The moving average processor : Computes the moving average of the positive values signal and outputting a moving average signal which defines the moving average over time. Finally, the enhanced digital samples are again processed to determine the individual’s heartbeat rate.
  • 9. 4. Figure 6: Detailed Flowchart of the Mechanics of the Heart Rate Monitor Figure 7: Detailed Flowchart of the Enhancement Signal Processor
  • 10. Figure 8: Sportline S7 Heart Rate Monitor Transcutaneous Electrical Nerve Stimulation (TENS) For our experiment we wanted to create a constant EMG regulation and eliminate any experimental variations, such as the difference of muscle contraction by using stress balls. In order to achieve the regulation, we selected a supplementary device that was provided by one of our group members, Omron HV-F128. The HV-F128 uses the TENS technology to massage the muscles and relieve the pain. The TENS technology works similar to how an electronic muscle stimulator (EMS) does. Both stimulators generate electrical impulses that stimulate the targeting nerves through the skin, which in turn cause the muscles controlled by those nerves to react and contract (Jones, I., & Johnson, M. I., 2009). However, the only difference between EMS and TENS is their targeting nerves. While EMS are designed to stimulate muscle motor nerves, TENS devices are designed to stimulate sensory nerve endings. Even though the TENS device targets to stimulate the sensory nerves, it also stimulates the muscle motor nerves, which lie near the sensory nerves, and cause the muscles to contract passively.
  • 11. The TENS device, HV-F128, makes muscle contract passively and generate a constant EMG noise that we desire. From the user manual, we obtained that the frequency HV-F128 generates is 1 ~ 1200Hz and the power consumption is about 85 mA. Maximum output voltage is less or equal to 90V and maximum output current is less or equal to 10 mA (during 1kilo-omega load). Method and Materials Materials This experiment required the use of two separate material sets: materials to prepare the participant and devices for reading heart rate and stimulating the participant’s forearm. To perform the study, the participant was sprayed with one spray bottle of tap water and was then wiped down with sterile cotton balls. This adhered to our proper method of preparing the participant. In terms of devices used for reading heart rate and stimulating muscle contraction, the Sportline S7 heart rate watch and the Omron HV- F128 TENS device were used. The Sportline S7 watch uses one-touch technology and was purchased at Walmart (Walmart.com, 2013). The Omron HV-F128 massager was imported from Japan (Omron Healthcare Co., Ltd.), and the English manual was found at Omron-healthcare.com. Preparation of the Participant The participant was prepared by being notified of the exclusion criteria of the experiment. Participants with metal implants or pacemakers and those who were not members of the Georgia Institute of Technology’s BMED 1300 course were asked to not
  • 12. take part in the study. Eligible candidates to be participants were given a consent form that outlined the risks and benefits of participating in this study. Once consent was given, the participant gained entrance to the testing area where they were asked to roll up their sleeves (if necessary) and expose their left forearm. From here, the experimenter sprayed the wrist of the participant and wiped the area with a cotton ball to wipe off any materials that would affect the conductivity of participant’s wrist and forearm. Control Measurement The participant was asked to take a control measurement. The experimenter placed the Sportline S7 wristwatch on the left wrist of the participant, and the participant was asked to place their index and middle finger of their right hand onto the device’s touch sensor. The experimenter recorded this heart rate measurement (heartbeats/minute) in a secure document for analysis at a later point. The watch was then reset using a reset button on the side of the watch. Experimental Measurement The watch remained on the participant as a second modified measurement was taken. The experimenter applied the two silicone pads of the TENS device onto the forearm of the participant. Using the device settings (tap-mode, 35 Hz), the experimenter turned on the TENS device massager and set the intensity to a level of four. After twenty- five seconds, the participant again was asked to place their index and middle finger of their right hand onto the device’s touch sensor. The experimenter then recorded the heart rate (heartbeats/minute) and removed the watch and massager from the participant’s left arm. This concluded the experiment, and the participant was released.
  • 13. Statistical analysis Based on the literature previously described, the sample size of this experiment used was 16. This was found by using a power of .97, an effect size of .95, and a level of significance or alpha of 0.05 when using a statistical program called G*Power (Faul, Erdfelder, Lang & Buchner, 2007). To determine the significance of the difference between the control and experimental heart rate measurements, a one-tailed matched pairs t-test was used. Alpha was set to 0.05 and p values smaller than alpha were considered significant. Results Based on the results our raw data in Fig. 1 ( see Appendix) of the 16 participants of the study, the P-value, the mean differences, the standard deviation of differences and standard error of differences between the experimental group and control group were found to be 1.1375e-6, 63.6875, 34.4978, and 8.6244, respectively as shown in the table 1 (see Appendix for formulas).We decided to use a one-sided matched pairs t-test to analyze our data, as we needed a good test for a smaller samples size (<30) with a large difference. We used the match-pairs t-test because the same participant was used for both the control and experimental trials. We used the one-sided t-test based on the one-sided, positive results we found in the research. In addition, our results also further endorse our assumption for the one-sided t-test with a positive, one-sided trend. The box plot and graph 2 depicts that our data was slightly skewed to the left based on the five number summary in reference to the mean. Table 2 shows the five number summary: median, minimum, maximum, first quartile, and third quartile which
  • 14. were, respectively, 72, 2, 110, 45.75, and 92.5. When the mean of 63.6875 lies to the left of the median of 72, the bulk of the points are to the right of the center, giving a left skew. The interquartile range of the box plot can be easily seen in the histogram where the dense region occurs (heart rate between ~45.75 and ~92.5). The histogram uses intervals of 2 to receive the most normal-looking distribution. The general appearance of a normal distribution in our data is shown as a dotted line in the graph. Our Q-Q plot, Fig.3 (see Appendix), had a r2 value of .96, giving a high correlation with little variance from the trend line and therefore, a fairly normal distribution. Graph 3 shows that the control mean without EMG noise, is almost half of the experimental mean with EMG noise. The significance level of <.05 is denoted with ***. Our error bars represent standard error. Graph 4 shows the p-value, the area under the curve in the direction of the hypothesis, in our case, to the right of the t-statistic. The p-value is the fraction of the total area under the curve where the null hypothesis is correct. The blue region is the area to the right of the t-score and the red region represents the area of our p-value (our p-value is too small to be graphically visible). Our p-value was too small for the program we used to calculate the graph (the lower threshold on this program is p=.0001). We are able to reject the null hypothesis with 99.99988625% ((1-p)X100%) confidence.
  • 15. One-tailed Matched Pairs t-test p-value 1.1275e-6 Standard Deviation of Differences 34.4978 Standard Error of Differences 8.6244 Variance of Differences 1190.096 Mean of Differences 63.6875 t 7.384536226 Table 1 Population Size 16 Median 72 Minimum 2 Maximum 110 First Quartile 45.75 Third Quartile 92.5 Interquartile Range 46.75 Outerliers None Table 2 Graph 1: Box and Whisker Plot
  • 16. Graph 2: Histogram Graph 3: Mean Change Graph 4: T-Score (Hypothesis Test Graph Generator)
  • 17. Discussion After the data was all collected, we noticed a few trends in the data. All of the experimental points were either reoccurring values of 150, 152, or 189, with the exception of three lower points with the values of 68, 75, and 79; while the control values were almost completely unique. These reoccurring values could have been explained by error in the TENS device or the wristband. One potential error we considered in using the TENS device was that the frequency, translating close to that of ECG signal and hypothesized EMG noise, could have directly influenced the heart rate reading. We could only standardize the TENS device within a frequency range, so the slight change in those recurring frequencies could have been just that, for example the heart rate monitor could have picked up the electrical signal directly from the TENS device, and not from the actual EMG noise created by skeletal muscles from the TENS stimulation, which is what we were attempting to mimic. Another concern with the use of the TENS device was whether or not the device stimulated only skeletal muscles or also increased the heart rate by stimulating the heart muscle as well. In the case that the TENS device stimulated the actual heart muscle, the heart rate monitor could have given a true reading without error. Although we did not collect data to calibrate the actual heart rate, the participants stayed in a sitting position throughout the experiment and no one showed any sign of increased heart rate while taking the second reading. Had a participant’s heart rate actually increased by about 70 beats per minute in approximately 10 seconds (about the time in between readings) , physical changes would be apparent. Another potential explanation for the three reoccurring values could have been the wristband and its filtering systems. It may be possible that over a certain frequency
  • 18. reading, over the normal range, the watch may error. There could be a chance that the filters could error and display randomly those three values. This seemed very unlikely, however, as the device is meant for exercise and values around 150 beats/minute are not completely out of the question while exercising. The smaller three experimental values also struck curiosity when analyzing the data. When taking these readings, we noticed the participants complained of not feeling the TENS massager’s stimulation. These participants also had more hair on their arms, which could have led to an error in applying the muscle stimuli. Though the smaller values did skew the data slightly left, they were still positive differences and were not calculated as outliers. Some of the smaller increases show a more realistic change for heart rate in about a 10 second interval, but considering that the participants did not seem to be affected by the TENS device, the increase was most likely due to other variables such as anxiety from the test. Regardless, the use of a TENS device caused error in the wristband monitor, increasing the heart rate readings above the actual level. Conclusion According to articles and patents we researched, the EMG noise which is produced by muscle contractions significantly influences the heart rate that is measured by our device due to the one of the characteristics of EMG signals, which is EMG noise falls in the same range of frequency of the ECG signal. Using the data gathered in our experiment, we conducted statistical analyses, and found our p value to be 1.1375e-6. Because the p value was substantially lower than our value (0.05), we rejected the null hypothesis in favor of alternative hypothesis, therefore supporting the idea that EMG noise causes a significant increase in heart rate measurements. The origin of the error,
  • 19. which is EMG interference that causes the device to give a less accurate heart rate reading is the limitation of the learning process in the mechanism to suppress the noise. Because the learning process of our device has a low threshold, it is set to allow for other signals like the EMG noise to be considered as heart rates. Therefore, we recommend that the producer of this device increases the ability of the learning processor to better distinguish ECG signals from any other ECG-like noise such as EMG noise and to improve the sensitivity of the filter to suppress any unwanted noise.
  • 20. Citations American Heart Association (8 august 2013). Target Heart Rates. Retrieved 27 september, 2013, from http://www.heart.org/HEARTORG/GettingHealthy/PhysicalActivity/Target- Heart-Rates_UCM_434341_Article.jsp Burke, M. J., & Whelan, M. V. (1987). The Accuracy and Reliability of Commercial Heart Rate Monitors. Brit.J.Sports Med., 21(1), 29-32. Cincinnatichildrens. (2012, 04/2012). Electromyogram / EMG and Nerve Conduction Test. Retrieved 10/31/2013, 2013, from http://www.cincinnatichildrens.org/health/e/emg/ Drews, C. (2000). Electromyography: Recording Electrical Signals from Human Muscle. http://ableweb.org/volumes/vol-21/12-drewes.pdf Friesen, G. M., Jannett, T. C., Jadallah, M. A., Yates, S. L., Quint, S. R., & Nagle, H. T. (1990). A comparison of the Noise Sensitivity of Nine QRS Detection Algorithms. IEEE Transactions On Biomedical Engineering, 37(1). Faul, F., Erdfelder, E., Lang, A., & Buchner, A. (2007, December 01). G*power 3. Retrieved from http://www.psycho.uni-duesseldorf.de/abteilungen/aap/gpower3 Hypothesis Test Graph Generator. Hypothesis Test Graph Gnerator. Retrieved from http://www.imathas.com/stattools/norm.html Johnson, M. (2007). Transcutaneous Electrical Nerve Stimulation: Mechanisms, Clinical Application and Evidence. British Journal of Pain, 1(1), 7-11. doi: 10.1177/204946370700100103 Lo, T. Y.-C., & Tsai, Y. S. (1998). The United States Patent No. 5,738,104. Morris, V. Box and Whisker Plot Maker. Math Warehouse. (2013) Retrieved from http://www.mathwarehouse.com/charts/box-and-whisker-plot- maker.php#boxwhiskergraph Patrick S. Hamilton, W. J. T. (1986). Quantitative Investigation of QRS Detection Rules Using the MIT/BIH Arrhythmia Database. IEEE Transactions of Biomedical Engineering, 33(12). Science, A. F. S. (2013). ECG Accurate Pulse® Strapless Heart Rate Monitor Sportswatches. Retrieved, 30 September 2013, from http://www.aussiefitsport.com.au/wp-content/uploads/2010/12/AFSS- PulseQTGeneralInformation.pdf Silverthorn, D. U. (2013). Human Physiology: An Integrated Approach (6 ed.).Boston: Pearson Education.Print Villasenor, J. F. (2009). How to Create a Heart Rate Monitor and One-Lead EKG. Freescale Technology Forum. Jones, I., & Johnson, M. I. (2009). Transcutaneous electrical nerve stimulation. Education in Anaesthesia, Critical Care, & Pain, 9(4) Sportsline, Inc. (2006). SPORTLINE Solo 910 Heart Rate Watch. Retrieved October 2013, from: http://www.sportline.com/manuals/SP3637BK.pdf Thomas Ying-Ching Lo, Y. S. T. (1998). United States of America Patent No. US 5738104A Young, D. K. (n.d.). Statistics Formula Sheet. University of Surrey. Retrieved October 16, 2013, from http://personal.maths.surrey.ac.uk/st/K.Young/form_sheet.pdf
  • 21. Walmart.com. Sportline S7 Any Touch Heart Rate Monitor Watch -Walmart.com. Retrieved Oct 2013, from http://www.walmart.com/ip/Sportline-S7-Any-Touch- Heart-Rate-Monitor-Watch-Black/21672223 Appendix Subject Number Control Reading TENS Reading Difference 1 70 150 80 2 92 189 97 3 77 152 75 4 79 189 110 5 74 79 5 6 95 189 94 7 105 150 45 8 73 75 2 9 96 189 93 10 86 152 66 11 84 150 66 12 61 152 91 13 102 150 48 14 79 150 71 15 79 152 73 16 65 68 3 Table 3: Raw data Formulas: Standard deviation: s = where the mean is mean of our data
  • 22. Mean: Standard Error: t-statistic: t-score: This was found using the t-table in the appendix. The t-score was taken at the .05 alpha level with 15 degrees of freedom. P value: P-value was calculated in excel as a 2 array, matched pairs, one-sided TTEST. (Young, K.) Graph 5: Bar graph of plotted differences
  • 23. Graph 6: Q-Q plot of data