SlideShare a Scribd company logo
1 of 19
Download to read offline
PARAMEE PENGPRECHA
THAILAND INSTITUTE OF SCIENTIFIC AND TECHNOLOGICAL RESEARCH
Preparation of matrix sample for Food Testing Laboratory
INTRODUCTION
No
Traceability
NIMT (IDMS
method)
Traceability
to SI unit via
NIMT
TISTR has been designated by NIMT
Be able to
assign true
value using
IDMS
Preservative is widely used in food and beverage
Its use has to be controlled and labeled
“Preservatives added” .
Its amounts has to be declared when submitted to
Authority by TEST REPORT.
In fact, most of laboratories can provide additives
testing in food.
BUT in the condition of ISO/IEC 17025 accredited.
Paramee CP 2009
สถาบันวิจัยวิทยาศาสตรและเทคโนโลยีแหงประเทศไทย(วว.)
ti t th
สถาบันวิจัยวิทยาศาสตรและเทคโนโลยีแหงประเทศไทยสถาบันวิจัยวิทยาศาสตรและเทคโนโลยีแหงประเทศไทย((วววว.).)
i hti t th
Spiked benzoic acid in (500mg/kg)
and sorbic acid (250mg/kg) in
70%EtOH
CRM Production : Preservatives in Food
2kg
(40 bags)
Preparation of Matrix food sample containing preservatives content for
food testing Laboratory (PT sample and CRM Production)
Characteristic study
Test Items Sample blank Spiked sample
Appearance Red curry smooth paste
pH 5.07-5.10 5.06-5.07
Water activity (%) 0.891-0.917 0.881-0.876
Total plate count
(CFU/g)
3.6x106 12.0x106
Homogeneity : Feasibility Study
No of
Prep
SA %RSDSA BA %RSDBA
Lot 1
(2kg)
273 5.3 549 8.1
Lot 2
(10kg)
257 2.2 500 2.5
Lot 3
(10kg)
237 0.7 482 1.2
Lot 1 : Blending No 1
Lot 2 : Blending
No 2Random Sampling : N=10-12 (n=2) by HPLC-UV
Method : Modified Determination of benzoic acid,
sorbic acid and methyl, ethyl, propyl and butyl
parabens by HPLC, SOP No. CLG-BSO.01,
US Department of Agriculture, Food Safety and
Inspection Service, Office of Public Health Science
Lot 3 :Blending
and kneading
ISO 13528 : Statistical methods for use in proficiency testing by inter laboratory comparisons)
Sampling N = 12
(Between Sample)
n1 n2
Test portion n=2
(Within sample)
Lot No.3
N2
n1 n2
N3
n1 n2
N4
n1 n2
N5
n1 n2
N6
n1 n2
N1
Homogeneity Study
HPLC UV-VIS
Modified Determination of benzoic acid, sorbic acid and methyl, ethyl, propyl and
butyl parabens by HPLC, SOP No. CLG-BSO.01, US Department of Agriculture, Food
Safety and Inspection Service
Measurement results for homogeneity
check, Sorbic acid in curry paste
(mg/kg)
Test portion
Avera
ge (A-B)
(A-
B)^2
%RP
D
Sample No A B
mg/k
g mg/kg
mg/k
g
mg/k
g
1026 237.5 241.5 239.5 4.0 16.0 1.67
1029 238.2 238.5 238.4 0.3 0.1 0.13
1036 236.2 234.8 235.5 1.4 2.0 0.59
1065 237.4 235.7 236.6 1.7 2.9 0.72
1067 239.7 238.1 238.9 1.6 2.6 0.67
1071 235.5 241.5 238.5 6.0 36.0 2.52
1072 238.8 240.1 239.5 1.3 1.7 0.54
1079 233.4 237.9 235.7 4.5 20.3 1.91
1088 239.3 237.4 238.4 1.9 3.6 0.80
1099 237.4 237.1 237.3 0.3 0.1 0.13
1121 232.0 235.9 234.0 3.9 15.2 1.67
1122 237.8 238.6 238.2 0.8 0.6 0.34
Mean (N=12)
237.
5
SD
1.75
0
% CV
0.73
7
Within-sample standard deviation
2.05
1
Between-sample standard deviation
0.97
9SD robust statistic from Proficiency
assessment Robust statistic (3 outliers
excluded)
19.0
6
0.3xSD (robust statistic)
5.74
8
One-way ANOVA between-sample (F) 1.46One-way ANOVA between-sample (F
critical) 2.71
Measurement results for homogeneity
check, Sodium benzoate in curry paste
(mg/kg)
Sample No
Test portion
Averag
e (A-B)
(A-
B)^2 %RPD
A B mg/kg mg/kg mg/kg mg/kg
1026 480.0 490.3 485.2 10.3106.09 2.12
1029 480.4 482.2 481.3 1.8 3.24 0.37
1036 478.7 475.1 476.9 3.6 12.96 0.75
1065 472.3 482.5 477.4 10.2104.04 2.14
1067 491.8 485.6 488.7 6.2 38.44 1.27
1071 482.1 489.1 485.6 7 49 1.44
1072 486.7 492.8 489.8 6.1 37.21 1.25
1079 484.9 481.8 483.4 3.1 9.61 0.64
1088 464.4 482.3 473.4 17.9320.41 3.78
1099 483.4 483.6 483.5 0.2 0.04 0.04
1121 473.9 480.7 477.3 6.8 46.24 1.42
1122 490.7 491.7 491.2 1 1 0.20
Mean (N=12) 482.8
SD 5.670
% CV 1.174
Within-sample standard deviation 5.509
Between-sample standard deviation 4.120
SD robust statistic from Proficiency assessment
Robust statistic (4 outliers excluded) 41.430
0.3xSD (robust statistic) 12.429
One-way ANOVA between-sample (F) 2.11
One-way ANOVA between-sample (F critical) 2.71
ISO Guide 35 -Reference materials-General and statistical principles for certification
HOMOGENEITY (☺)
Short term Stability Study
Storage condition : in Nylon
transparent plastic bag : Dark
color were observed after 3
months at 25OC.
ySod.BA@4C = -0.0625x + 503.78 R² = 0.0672 y Sod.BA@25C= -0.2285x + 494.93 R² = 0.20
ySA@4C = -0.3473x + 254.96 R² = 0.7725
ySA@25C = -0.9157x + 244.88 R² = 0.8918
Instability of sorbic acid
PT Food Sample : Stability Study
Stability study and shelf life for short term study (3months)
Data
(Na.benzoate)
25C 4C Remark
Slope (b1) -0.2285 -0.0625
Intercept(b0) 494.931 503.779
S2 (sd of the point
among the line)
662.56 822.16
S(b1) (uncertainty
associated with the
slope)
0.175 0.195
Significant of slope
(t0.95, n-2=26 ,=2.056)
0.228<0.360 0.062<0.401
2
)(
1
2
110
2
−
−−
=
∑=
n
xXbbY
s
n
i
i
∑=
−
=
n
i
i XX
s
s
1
2
)(
)( 12,95.01 bxstb n−<
No instability was observed at 25C and 4C of storage condition for 90days
ISO Guide 35 – Reference materials-General and statistical principles for certification (Annex B5)
PT(Local Labs) : Sodium benzoate in curry paste
MTC-PT-1000-2550
-assigned value from 6 Labs (ISO/IEC
17025 accredited),
- Evaluation of the PT is according to
ISO 13528 : Statistical methods for use
in proficiency testing by inter laboratory
comparisons)
-Reference Standards were from
NIST, RMIPO, ACROSS, BDH, Carlo
Erbra, Fluka, MERCK, Riedel-
dehaenAG, SIGMA Aldich, Supelco,
UNILAB,and USP
Oct 2007-Feb 2008
77.11175.0553.9509.5120.4 2222
=+++=PTu
2222
stabPTwbbPT uuuuu +++=
Sodium benzoate PT Sample =474+24 mg/kg
U= 4.9%
where k = 2 is the coverage factor for a 95 %
confidence level and uCRM is the combined standard
uncertainty calculated according to ISO Guide 35
22
hom* assignoPT uuu +=
APMP –P14 : BENZOIC ACID IN CURRY PASTE
CRM Production : Benzoic acid in curry paste
Material No of bags
(50g)
PT Sample 40
CRM needed 50
Feasibility study 20
Characteristic study 20
Homogeneity study 10
Stability study 60
10kg
Gamma radiation
Eliminate bacteria*
*Curry paste samples were sterilized by gamma ray irradiation at Thailand Institute of Nuclear and
Technology (TINT)
•Decreasing of sorbic acid was observed and inhomogeneity was found.
•Study on stability was found that 39.5% of sorbic acid was decreased @
1000 mg/l) after 1h at 105C.
•Therefore sorbic acid was not considered for the APMP.QM-P14
*the homogeneity were carried on before and after radiation
Analysis Blank sample PT sample
Benzoic acid Not detected
(less than 3 mg/kg)
See homogeneity study
pH (10%w/v) 5.09 5.02
Water activity 0.89% 0.88%
Total plate count 1.0x106 CFU/g 4.2x106 CFU/g (before irradiation)
8.3x103 CFU/g (after irradiation)
Solubility 80% w/w
Non soluble matter 4% w/w on 1 micron membrane filter
16% w/w on 22 micron membrane filter
Characterization Study : Benzoic acid in curry paste
Result is similar to the previous study
CRM Homogeneity Study : Benzoic acid in Curry Paste
Homogeneity: Benzoic acid(mg/kg)
n 1 2 3 Average SD
1 381.6 387.7 388.6 386.0 3.8
2 390.7 387.1 390.9 389.6 2.1
3 377.3 388.9 390.0 385.4 7.0
4 386.6 388.9 389.7 388.4 1.6
5 388.2 389.5 391.1 389.6 1.4
6 389.9 388.3 391.0 389.7 1.3
7 388.6 390.3 392.9 390.6 2.2
8 389.8 392.7 392.8 391.8 1.7
9 391.2 393.9 389.3 391.5 2.3
10 388.0 382.3 384.9 385.1 2.8
Figure 1: Participants’ results for benzoic acid content in curry paste sample. The
horizontal blue lines represent the proposed reference value (median) and its associated
uncertainty. The horizontal brown line is the arithmetic mean value and its associated
uncertainty.
assigned Value for CRM =378 mg/kg
U=12 mg/kg
CRM Stability Study : Benzoic Acid in Curry Paste
BA -22C
Slope (b1) 0.262
Intercept(b0) 379.334
S
2
(sd of the point among the line) 9.162
S(b1) (unc associated with the
slope)
0.143
Significant of slope
(t0.95, n-2=10 ,=2.776)
0.261<0.396
)( 12,95.01 bxstb n−<
No instability was observed at -22C for 2 years
CRM : Benzoic acid in curry paste
22222
ltsstsassignedwbbCRM uuuuuu ++++=
Sources of unc ui
ui
2
usts 0.175 0.031
ults 0.143 0.020
u(assigned)
6.500 42.250
Sbb
2.928 8.572
S within gr
3.097 9.591
SQRT(ui
2
combined) 7.776
U (k=2) 15.6
%U 3.2
Benzoic acid = 378 +15.6 mg/kg
•The certified value is based on isotope dilution mass
spectrometry method which is analysed by National
Metrology Institute of Thailand
•Benzoic acid SRM is used for matrix matching
standard curve
CRMCRM kxuU
•uncertainty is the combined standard uncertainty
multiply by k calculated according to ISO Guide 35,
•where k = 2 is the coverage factor for a 95 %
confidence level
=
ACKNOWLWDGEMENT
Dr. Chainarong Cherdchu NIMT
Dr. Charun Yafa NIMT
Ms.Pranee Phukphatthanachai NIMT
Dr. Luxsamee Plungsangmas TISTR
Mr. Cheumsak Sinchaisri TISTR
Ms.Vilaval Pongpitak et.al. TISTR (BCL)
Ms.Thippaya Junvee Fortune et.al. TISTR (ACL)
ACL team fromTISTR

More Related Content

What's hot

CPSC Phthalate SOP
CPSC Phthalate SOPCPSC Phthalate SOP
CPSC Phthalate SOPjaykg64
 
Mdo2014 takeda
Mdo2014 takedaMdo2014 takeda
Mdo2014 takedaEd Tessier
 
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Shimadzu Scientific Instruments
 
Biosimilar Development Regulatory, Analytical, and Clinical Considerations
Biosimilar Development Regulatory, Analytical, and Clinical Considerations Biosimilar Development Regulatory, Analytical, and Clinical Considerations
Biosimilar Development Regulatory, Analytical, and Clinical Considerations SGS
 
NYSAS Seminar LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...
NYSAS Seminar  LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...NYSAS Seminar  LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...
NYSAS Seminar LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...mzhou45
 
Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...Ratnakaram Venkata Nadh
 
Valganciclovir Tablet
Valganciclovir Tablet Valganciclovir Tablet
Valganciclovir Tablet google
 
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111Shimadzu Scientific Instruments
 
Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4Nádia Paracampo
 
GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)
GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)
GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)Shimadzu Scientific Instruments
 
VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...
VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...
VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...Ruqsar Fatima
 
Stability indicating method development and validation for the simultaneous e...
Stability indicating method development and validation for the simultaneous e...Stability indicating method development and validation for the simultaneous e...
Stability indicating method development and validation for the simultaneous e...pharmaindexing
 
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...Shimadzu Scientific Instruments
 
Diclofenac rabeprazole hplc
Diclofenac rabeprazole hplcDiclofenac rabeprazole hplc
Diclofenac rabeprazole hplcDeepak Gadade
 
SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014Kenneth Wong
 
An Experimental Design Approach for Method Development and Impurity Profiling...
An Experimental Design Approach for Method Development and Impurity Profiling...An Experimental Design Approach for Method Development and Impurity Profiling...
An Experimental Design Approach for Method Development and Impurity Profiling...pharmaindexing
 
Portugalska 2005 A
Portugalska 2005 APortugalska 2005 A
Portugalska 2005 AFNian
 

What's hot (20)

CPSC Phthalate SOP
CPSC Phthalate SOPCPSC Phthalate SOP
CPSC Phthalate SOP
 
Mdo2014 takeda
Mdo2014 takedaMdo2014 takeda
Mdo2014 takeda
 
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
 
Biosimilar Development Regulatory, Analytical, and Clinical Considerations
Biosimilar Development Regulatory, Analytical, and Clinical Considerations Biosimilar Development Regulatory, Analytical, and Clinical Considerations
Biosimilar Development Regulatory, Analytical, and Clinical Considerations
 
Pesticides
PesticidesPesticides
Pesticides
 
Presentation 3
Presentation 3Presentation 3
Presentation 3
 
NYSAS Seminar LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...
NYSAS Seminar  LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...NYSAS Seminar  LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...
NYSAS Seminar LC-IR To Characterize Polymeric Excipients In Pharmaceutical F...
 
Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...
 
Valganciclovir Tablet
Valganciclovir Tablet Valganciclovir Tablet
Valganciclovir Tablet
 
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
 
Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4
 
Final PPT
Final PPTFinal PPT
Final PPT
 
GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)
GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)
GC Analysis of Residual Solvents in Pharmaceuticals (USP&lt;467>)
 
VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...
VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...
VALIDATION AND DETERMINATION OF CAFFEINE CONTENT IN ENERGY DRINKS BY USING HP...
 
Stability indicating method development and validation for the simultaneous e...
Stability indicating method development and validation for the simultaneous e...Stability indicating method development and validation for the simultaneous e...
Stability indicating method development and validation for the simultaneous e...
 
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
 
Diclofenac rabeprazole hplc
Diclofenac rabeprazole hplcDiclofenac rabeprazole hplc
Diclofenac rabeprazole hplc
 
SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014
 
An Experimental Design Approach for Method Development and Impurity Profiling...
An Experimental Design Approach for Method Development and Impurity Profiling...An Experimental Design Approach for Method Development and Impurity Profiling...
An Experimental Design Approach for Method Development and Impurity Profiling...
 
Portugalska 2005 A
Portugalska 2005 APortugalska 2005 A
Portugalska 2005 A
 

Similar to Preparation and Characterization of Benzoic Acid in Curry Paste CRM

Analysis of organic acids in e-vapor products by ion chromatography
Analysis of organic acids in e-vapor products by ion chromatographyAnalysis of organic acids in e-vapor products by ion chromatography
Analysis of organic acids in e-vapor products by ion chromatographyRana Tayyarah
 
Nutritional Metrology: The Role of Reference Materials in Improving Quality o...
Nutritional Metrology: The Role of Reference Materials in Improving Quality o...Nutritional Metrology: The Role of Reference Materials in Improving Quality o...
Nutritional Metrology: The Role of Reference Materials in Improving Quality o...Malaysian Palm Oil Council (MPOC Egypt)
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESPerkinElmer, Inc.
 
sawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptxsawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptxEhabAdel29
 
Improvement in Protein A Resin Lifetime
Improvement in Protein A Resin LifetimeImprovement in Protein A Resin Lifetime
Improvement in Protein A Resin LifetimeMilliporeSigma
 
Improvement in Protein A Resin Lifetime
Improvement in Protein A Resin LifetimeImprovement in Protein A Resin Lifetime
Improvement in Protein A Resin LifetimeMerck Life Sciences
 
FT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturingFT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturingGalaxy Scientific
 
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...PerkinElmer, Inc.
 
FOCUS 2016 Poster: Specific Proteins
FOCUS 2016 Poster: Specific ProteinsFOCUS 2016 Poster: Specific Proteins
FOCUS 2016 Poster: Specific ProteinsRandox
 
PVR P1-60 mcg.doc
PVR P1-60 mcg.docPVR P1-60 mcg.doc
PVR P1-60 mcg.doc9160676107
 
Nitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNguyenMinhChau22
 
Topical Niosomal Preparation Presentation 1
Topical Niosomal Preparation Presentation  1Topical Niosomal Preparation Presentation  1
Topical Niosomal Preparation Presentation 1guest52ea7
 
Green Site Project results
Green Site Project resultsGreen Site Project results
Green Site Project resultsPetra Scanferla
 
Solid Lipid Nano particles of Alendronate Sodium
 Solid Lipid Nano particles of Alendronate Sodium Solid Lipid Nano particles of Alendronate Sodium
Solid Lipid Nano particles of Alendronate SodiumRohit Suman
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsSCIEX
 

Similar to Preparation and Characterization of Benzoic Acid in Curry Paste CRM (20)

Chromatography: Trends and Developments in MAb Screening and Characterization
Chromatography: Trends and Developments in MAb Screening and CharacterizationChromatography: Trends and Developments in MAb Screening and Characterization
Chromatography: Trends and Developments in MAb Screening and Characterization
 
Analysis of organic acids in e-vapor products by ion chromatography
Analysis of organic acids in e-vapor products by ion chromatographyAnalysis of organic acids in e-vapor products by ion chromatography
Analysis of organic acids in e-vapor products by ion chromatography
 
Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...
Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...
Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...
 
Nutritional Metrology: The Role of Reference Materials in Improving Quality o...
Nutritional Metrology: The Role of Reference Materials in Improving Quality o...Nutritional Metrology: The Role of Reference Materials in Improving Quality o...
Nutritional Metrology: The Role of Reference Materials in Improving Quality o...
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 
sawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptxsawsan-mohamed-amer-cairo-university-egypt.pptx
sawsan-mohamed-amer-cairo-university-egypt.pptx
 
Improvement in Protein A Resin Lifetime
Improvement in Protein A Resin LifetimeImprovement in Protein A Resin Lifetime
Improvement in Protein A Resin Lifetime
 
Improvement in Protein A Resin Lifetime
Improvement in Protein A Resin LifetimeImprovement in Protein A Resin Lifetime
Improvement in Protein A Resin Lifetime
 
Analysis of Milk and Egg Allergens in Wine using UPLC-MS - Waters Corporation...
Analysis of Milk and Egg Allergens in Wine using UPLC-MS - Waters Corporation...Analysis of Milk and Egg Allergens in Wine using UPLC-MS - Waters Corporation...
Analysis of Milk and Egg Allergens in Wine using UPLC-MS - Waters Corporation...
 
FT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturingFT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturing
 
The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...
The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...
The Advantages of Mass Detection for the Food Testing Laboratory - Waters Cor...
 
Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water
 
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
 
FOCUS 2016 Poster: Specific Proteins
FOCUS 2016 Poster: Specific ProteinsFOCUS 2016 Poster: Specific Proteins
FOCUS 2016 Poster: Specific Proteins
 
PVR P1-60 mcg.doc
PVR P1-60 mcg.docPVR P1-60 mcg.doc
PVR P1-60 mcg.doc
 
Nitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdf
 
Topical Niosomal Preparation Presentation 1
Topical Niosomal Preparation Presentation  1Topical Niosomal Preparation Presentation  1
Topical Niosomal Preparation Presentation 1
 
Green Site Project results
Green Site Project resultsGreen Site Project results
Green Site Project results
 
Solid Lipid Nano particles of Alendronate Sodium
 Solid Lipid Nano particles of Alendronate Sodium Solid Lipid Nano particles of Alendronate Sodium
Solid Lipid Nano particles of Alendronate Sodium
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminants
 

Preparation and Characterization of Benzoic Acid in Curry Paste CRM

  • 1. PARAMEE PENGPRECHA THAILAND INSTITUTE OF SCIENTIFIC AND TECHNOLOGICAL RESEARCH
  • 2. Preparation of matrix sample for Food Testing Laboratory INTRODUCTION No Traceability NIMT (IDMS method) Traceability to SI unit via NIMT TISTR has been designated by NIMT Be able to assign true value using IDMS Preservative is widely used in food and beverage Its use has to be controlled and labeled “Preservatives added” . Its amounts has to be declared when submitted to Authority by TEST REPORT. In fact, most of laboratories can provide additives testing in food. BUT in the condition of ISO/IEC 17025 accredited.
  • 3. Paramee CP 2009 สถาบันวิจัยวิทยาศาสตรและเทคโนโลยีแหงประเทศไทย(วว.) ti t th สถาบันวิจัยวิทยาศาสตรและเทคโนโลยีแหงประเทศไทยสถาบันวิจัยวิทยาศาสตรและเทคโนโลยีแหงประเทศไทย((วววว.).) i hti t th Spiked benzoic acid in (500mg/kg) and sorbic acid (250mg/kg) in 70%EtOH CRM Production : Preservatives in Food 2kg (40 bags) Preparation of Matrix food sample containing preservatives content for food testing Laboratory (PT sample and CRM Production)
  • 4. Characteristic study Test Items Sample blank Spiked sample Appearance Red curry smooth paste pH 5.07-5.10 5.06-5.07 Water activity (%) 0.891-0.917 0.881-0.876 Total plate count (CFU/g) 3.6x106 12.0x106
  • 5. Homogeneity : Feasibility Study No of Prep SA %RSDSA BA %RSDBA Lot 1 (2kg) 273 5.3 549 8.1 Lot 2 (10kg) 257 2.2 500 2.5 Lot 3 (10kg) 237 0.7 482 1.2 Lot 1 : Blending No 1 Lot 2 : Blending No 2Random Sampling : N=10-12 (n=2) by HPLC-UV Method : Modified Determination of benzoic acid, sorbic acid and methyl, ethyl, propyl and butyl parabens by HPLC, SOP No. CLG-BSO.01, US Department of Agriculture, Food Safety and Inspection Service, Office of Public Health Science Lot 3 :Blending and kneading ISO 13528 : Statistical methods for use in proficiency testing by inter laboratory comparisons)
  • 6. Sampling N = 12 (Between Sample) n1 n2 Test portion n=2 (Within sample) Lot No.3 N2 n1 n2 N3 n1 n2 N4 n1 n2 N5 n1 n2 N6 n1 n2 N1 Homogeneity Study HPLC UV-VIS Modified Determination of benzoic acid, sorbic acid and methyl, ethyl, propyl and butyl parabens by HPLC, SOP No. CLG-BSO.01, US Department of Agriculture, Food Safety and Inspection Service
  • 7. Measurement results for homogeneity check, Sorbic acid in curry paste (mg/kg) Test portion Avera ge (A-B) (A- B)^2 %RP D Sample No A B mg/k g mg/kg mg/k g mg/k g 1026 237.5 241.5 239.5 4.0 16.0 1.67 1029 238.2 238.5 238.4 0.3 0.1 0.13 1036 236.2 234.8 235.5 1.4 2.0 0.59 1065 237.4 235.7 236.6 1.7 2.9 0.72 1067 239.7 238.1 238.9 1.6 2.6 0.67 1071 235.5 241.5 238.5 6.0 36.0 2.52 1072 238.8 240.1 239.5 1.3 1.7 0.54 1079 233.4 237.9 235.7 4.5 20.3 1.91 1088 239.3 237.4 238.4 1.9 3.6 0.80 1099 237.4 237.1 237.3 0.3 0.1 0.13 1121 232.0 235.9 234.0 3.9 15.2 1.67 1122 237.8 238.6 238.2 0.8 0.6 0.34 Mean (N=12) 237. 5 SD 1.75 0 % CV 0.73 7 Within-sample standard deviation 2.05 1 Between-sample standard deviation 0.97 9SD robust statistic from Proficiency assessment Robust statistic (3 outliers excluded) 19.0 6 0.3xSD (robust statistic) 5.74 8 One-way ANOVA between-sample (F) 1.46One-way ANOVA between-sample (F critical) 2.71 Measurement results for homogeneity check, Sodium benzoate in curry paste (mg/kg) Sample No Test portion Averag e (A-B) (A- B)^2 %RPD A B mg/kg mg/kg mg/kg mg/kg 1026 480.0 490.3 485.2 10.3106.09 2.12 1029 480.4 482.2 481.3 1.8 3.24 0.37 1036 478.7 475.1 476.9 3.6 12.96 0.75 1065 472.3 482.5 477.4 10.2104.04 2.14 1067 491.8 485.6 488.7 6.2 38.44 1.27 1071 482.1 489.1 485.6 7 49 1.44 1072 486.7 492.8 489.8 6.1 37.21 1.25 1079 484.9 481.8 483.4 3.1 9.61 0.64 1088 464.4 482.3 473.4 17.9320.41 3.78 1099 483.4 483.6 483.5 0.2 0.04 0.04 1121 473.9 480.7 477.3 6.8 46.24 1.42 1122 490.7 491.7 491.2 1 1 0.20 Mean (N=12) 482.8 SD 5.670 % CV 1.174 Within-sample standard deviation 5.509 Between-sample standard deviation 4.120 SD robust statistic from Proficiency assessment Robust statistic (4 outliers excluded) 41.430 0.3xSD (robust statistic) 12.429 One-way ANOVA between-sample (F) 2.11 One-way ANOVA between-sample (F critical) 2.71 ISO Guide 35 -Reference materials-General and statistical principles for certification HOMOGENEITY (☺)
  • 8. Short term Stability Study Storage condition : in Nylon transparent plastic bag : Dark color were observed after 3 months at 25OC. ySod.BA@4C = -0.0625x + 503.78 R² = 0.0672 y Sod.BA@25C= -0.2285x + 494.93 R² = 0.20 ySA@4C = -0.3473x + 254.96 R² = 0.7725 ySA@25C = -0.9157x + 244.88 R² = 0.8918 Instability of sorbic acid
  • 9. PT Food Sample : Stability Study Stability study and shelf life for short term study (3months) Data (Na.benzoate) 25C 4C Remark Slope (b1) -0.2285 -0.0625 Intercept(b0) 494.931 503.779 S2 (sd of the point among the line) 662.56 822.16 S(b1) (uncertainty associated with the slope) 0.175 0.195 Significant of slope (t0.95, n-2=26 ,=2.056) 0.228<0.360 0.062<0.401 2 )( 1 2 110 2 − −− = ∑= n xXbbY s n i i ∑= − = n i i XX s s 1 2 )( )( 12,95.01 bxstb n−< No instability was observed at 25C and 4C of storage condition for 90days ISO Guide 35 – Reference materials-General and statistical principles for certification (Annex B5)
  • 10. PT(Local Labs) : Sodium benzoate in curry paste MTC-PT-1000-2550 -assigned value from 6 Labs (ISO/IEC 17025 accredited), - Evaluation of the PT is according to ISO 13528 : Statistical methods for use in proficiency testing by inter laboratory comparisons) -Reference Standards were from NIST, RMIPO, ACROSS, BDH, Carlo Erbra, Fluka, MERCK, Riedel- dehaenAG, SIGMA Aldich, Supelco, UNILAB,and USP Oct 2007-Feb 2008 77.11175.0553.9509.5120.4 2222 =+++=PTu 2222 stabPTwbbPT uuuuu +++= Sodium benzoate PT Sample =474+24 mg/kg U= 4.9% where k = 2 is the coverage factor for a 95 % confidence level and uCRM is the combined standard uncertainty calculated according to ISO Guide 35 22 hom* assignoPT uuu +=
  • 11. APMP –P14 : BENZOIC ACID IN CURRY PASTE
  • 12. CRM Production : Benzoic acid in curry paste Material No of bags (50g) PT Sample 40 CRM needed 50 Feasibility study 20 Characteristic study 20 Homogeneity study 10 Stability study 60 10kg Gamma radiation Eliminate bacteria* *Curry paste samples were sterilized by gamma ray irradiation at Thailand Institute of Nuclear and Technology (TINT) •Decreasing of sorbic acid was observed and inhomogeneity was found. •Study on stability was found that 39.5% of sorbic acid was decreased @ 1000 mg/l) after 1h at 105C. •Therefore sorbic acid was not considered for the APMP.QM-P14 *the homogeneity were carried on before and after radiation
  • 13. Analysis Blank sample PT sample Benzoic acid Not detected (less than 3 mg/kg) See homogeneity study pH (10%w/v) 5.09 5.02 Water activity 0.89% 0.88% Total plate count 1.0x106 CFU/g 4.2x106 CFU/g (before irradiation) 8.3x103 CFU/g (after irradiation) Solubility 80% w/w Non soluble matter 4% w/w on 1 micron membrane filter 16% w/w on 22 micron membrane filter Characterization Study : Benzoic acid in curry paste Result is similar to the previous study
  • 14. CRM Homogeneity Study : Benzoic acid in Curry Paste Homogeneity: Benzoic acid(mg/kg) n 1 2 3 Average SD 1 381.6 387.7 388.6 386.0 3.8 2 390.7 387.1 390.9 389.6 2.1 3 377.3 388.9 390.0 385.4 7.0 4 386.6 388.9 389.7 388.4 1.6 5 388.2 389.5 391.1 389.6 1.4 6 389.9 388.3 391.0 389.7 1.3 7 388.6 390.3 392.9 390.6 2.2 8 389.8 392.7 392.8 391.8 1.7 9 391.2 393.9 389.3 391.5 2.3 10 388.0 382.3 384.9 385.1 2.8
  • 15. Figure 1: Participants’ results for benzoic acid content in curry paste sample. The horizontal blue lines represent the proposed reference value (median) and its associated uncertainty. The horizontal brown line is the arithmetic mean value and its associated uncertainty. assigned Value for CRM =378 mg/kg U=12 mg/kg
  • 16. CRM Stability Study : Benzoic Acid in Curry Paste BA -22C Slope (b1) 0.262 Intercept(b0) 379.334 S 2 (sd of the point among the line) 9.162 S(b1) (unc associated with the slope) 0.143 Significant of slope (t0.95, n-2=10 ,=2.776) 0.261<0.396 )( 12,95.01 bxstb n−< No instability was observed at -22C for 2 years
  • 17. CRM : Benzoic acid in curry paste 22222 ltsstsassignedwbbCRM uuuuuu ++++= Sources of unc ui ui 2 usts 0.175 0.031 ults 0.143 0.020 u(assigned) 6.500 42.250 Sbb 2.928 8.572 S within gr 3.097 9.591 SQRT(ui 2 combined) 7.776 U (k=2) 15.6 %U 3.2 Benzoic acid = 378 +15.6 mg/kg •The certified value is based on isotope dilution mass spectrometry method which is analysed by National Metrology Institute of Thailand •Benzoic acid SRM is used for matrix matching standard curve CRMCRM kxuU •uncertainty is the combined standard uncertainty multiply by k calculated according to ISO Guide 35, •where k = 2 is the coverage factor for a 95 % confidence level =
  • 18. ACKNOWLWDGEMENT Dr. Chainarong Cherdchu NIMT Dr. Charun Yafa NIMT Ms.Pranee Phukphatthanachai NIMT Dr. Luxsamee Plungsangmas TISTR Mr. Cheumsak Sinchaisri TISTR Ms.Vilaval Pongpitak et.al. TISTR (BCL) Ms.Thippaya Junvee Fortune et.al. TISTR (ACL)