CONTENT
• Qualification
• Membrane filtration
• Types of filter
• Design qualification
• Operational qualification
• Installation qualification
• Performance qualification
• Validation study element
• References
2
QUALIFICATION
• Qualification is defined as an action of providing
that equipment or systems are properly installed,
work correctly and actually leads to expected
results.
• Qualification is part of validation.
Types of qualification:
I. Design qualification
II. Installation qualification
III. Operational qualification
IV. Performance qualification
3
MEMBRANE FILTRATION
• Unit operation of filtration is the separation of
solids from a liquid by passage through a filter
medium.
• Membrane filtration is used for sterilization of
drug product and used in sterilization process.
4
TYPES OF FILTER
There are two types of filter used in filtration
process:
I. Depth filters: Consist of fibrous or granular
materials.
II. Membrane filters: These are porous
membrane.
5
DESIGN QUALIFICATION (DQ)
• DQ defines the functional and operational
specifications of an instrument.
• DQ details the conscious decisions made in
the selection of the supplier.
• DQ should ensure that instruments have all
the necessary functions and performance
criteria that will enable them to meet user
requirements.
• The list shows the recommended steps should
considered during DQ
6
7
Description of the analyse problem
Description of the intended use for the equipment
Description of the intended environment
Preliminary selection of the functional and performance specifications
Preliminary selection of the supplier
Development and documentation of final functional and operational
specifications
Final selection of the supplier and equipment
OPERATIONAL QUALIFICATION
(OQ)
• OQ is the process of demonstrating that
instrument will function according to its
operational specification in the selected
environment.
• Before OQ testing is done, one should always
consider what the instrument will be used for.
• Testing may be quite extensive if the
instrument is to be used for all types of
applications.
8
INSTALLATION QUALIFICATION(IQ)
• IQ establishes that the instrument is received
as designed and specified, that it is properly
installed in the selected environment.
• IQ tell about environment that environment is
suitable for the operation and use of the
instrument.
9
Performance Qualification (PQ)
• PQ is the process of demonstrating that an
instrument consistently performs according
to a specification appropriate to its routine
use.
• PQ should be performed on a daily (or at least
a weekly) basis whenever instrument is used.
• The test frequency depends not only on the
stability of the equipment but also on
everything in the system that may contribute
to the analysis results.
10
VALIDATION STUDY ELEMENT
11
Physical
 Reproducibility of filter
Sterilization
Integrity test
Operating condition
Shedding
Microbial challenge test
 Biological
 Endotoxin
 Toxicity
Chemical
Inertness
Activity/stability
Test for antimicrobial
activities
Consistency and
reliability
1. REPRODUCIBILITY OF FILTER
Membrane filters should be routinely
discarded.
Factor that can affect in filtration process
include viscosity, surface tension of material to
be filtered, ph, flow rates, time, temperature,
osmolarity etc.
12
2. STERILIZATION
Validation of sterilization method of filter is
necessary because filter it self cause
contamination of the product.
To validate use of sterilizing grade filter it is
not only prove that the filter is adequately
sterilized but also method does not damage the
filter.
 Most preferred method is moist heat
sterilizing.
13
3. INTEGRITY TEST
It should be non destructive and provide an
indication of “ fitness for use ”.
 This include bubble point test, diffusion test,
pressure hold test, water intrusion test.
 This test of filter should be performed prior to
processing and should be performed routinely
and conducted after filtration to detect any
filter leaks or perforation that might have
occurred during the filtration.
14
4. OPERATING CONDITION
a. Time: Long processing time could allow
bacteria which have been trapped by the filter.
Filter manufacturer can provide the data on the
retention tests that have been conducted for
specific membrane and generally suggest that
filter should retain bacteria excess of 48hr.
b. Temperature: Manufacture of filter
recommended the limit and these not
exceeded.
c. Pressure: Inlet pressure to the filter must be
monitored to ensure that there is no potential for
structure damage.
15
5. Filter inertness
 There may be extraction and adsorption
phenomena occurs.
 Stability of the product should not be affected
by the filter
16
6. TEST FOR ANTIMICROBIAL
ACTIVITY
The test is performed to ensure that, any residual
of Antimicrobial Activity is satisfactory
eliminated by using the steps mentioned in this
protocol.
An inoculums of viable cells of the specific
bacteria and fungi has been passed through the
filter, inoculate filter paper in FTM & incubate at
30 to 350C or in SCDM and incubate at 20 to
250C.
If conspicuous growth does not occur within 3
days for bacteria and 5 days for fungi, the test
procedure is not valid and must be modified 17
7. ENDOTOXIN
 Validation must address filter does not add
endotoxin to drug product.
 It depend on quality control process of the
filter manufacturer, water used in
manufacturing, choice of filter vendor,
verification are not done properly.
18
8. TOXICITY
 A validation study should determine that
passage of the drug product through a filter
does not cause any toxicological effects
 Construction material off filtration system
should be non toxic
19
REFRENCES
• R. A. Nash and A. H. Watche, Pharmaceutical
process validation, Third edition
• https://www.researchgate.net/publication/27
5519051_Review_on_Validation_of_Autoclave
_Membrane_Filtration
• https://www.slideshare.net/sagarsavale1/vali
dation-of-membrane-filter
20
21

More Related Content

What's hot

Advanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyAdvanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyPriyankaSananse
 
Process validation of Tablets and Lquids
Process validation of Tablets and LquidsProcess validation of Tablets and Lquids
Process validation of Tablets and LquidsRaheem Kurikkal
 
Sterile process validation
Sterile process validationSterile process validation
Sterile process validationSagar Savale
 
Role of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalRole of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalganpat420
 
Advanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyAdvanced sterile product manufacturing technology
Advanced sterile product manufacturing technologySantosh kumar
 
Qualification of laboratory equipments
Qualification of laboratory equipmentsQualification of laboratory equipments
Qualification of laboratory equipmentsPranali Polshettiwar
 
Advance non – sterile solid product manufacturing technology
Advance non – sterile solid product manufacturing technologyAdvance non – sterile solid product manufacturing technology
Advance non – sterile solid product manufacturing technologyAbhishekJadhav189260
 
Quality by Design and Process Analytical Technology
Quality by Design and Process Analytical TechnologyQuality by Design and Process Analytical Technology
Quality by Design and Process Analytical TechnologyMANIKANDAN V
 
Auditing of vendors and production department
Auditing of vendors and production departmentAuditing of vendors and production department
Auditing of vendors and production departmentArpitSuralkar
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validationSagar Savale
 
Introduction to validation
Introduction to validation Introduction to validation
Introduction to validation Ankush Sule
 
Equipment validation of fbd
Equipment validation of fbdEquipment validation of fbd
Equipment validation of fbdpriyanka odela
 
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamJp Prakash
 
Vendor qualification
Vendor qualificationVendor qualification
Vendor qualificationShihabPatel
 
Process Validation of Liquid Orals
Process Validation of Liquid OralsProcess Validation of Liquid Orals
Process Validation of Liquid OralsAasawaree Yadav
 

What's hot (20)

Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Advanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyAdvanced sterile product manufacturing technology
Advanced sterile product manufacturing technology
 
Process validation of Tablets and Lquids
Process validation of Tablets and LquidsProcess validation of Tablets and Lquids
Process validation of Tablets and Lquids
 
Sterile process validation
Sterile process validationSterile process validation
Sterile process validation
 
Role of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalRole of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceutical
 
Advanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyAdvanced sterile product manufacturing technology
Advanced sterile product manufacturing technology
 
Qualification of laboratory equipments
Qualification of laboratory equipmentsQualification of laboratory equipments
Qualification of laboratory equipments
 
Advance non – sterile solid product manufacturing technology
Advance non – sterile solid product manufacturing technologyAdvance non – sterile solid product manufacturing technology
Advance non – sterile solid product manufacturing technology
 
Quality by Design and Process Analytical Technology
Quality by Design and Process Analytical TechnologyQuality by Design and Process Analytical Technology
Quality by Design and Process Analytical Technology
 
Auditing of vendors and production department
Auditing of vendors and production departmentAuditing of vendors and production department
Auditing of vendors and production department
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Equipment qualification
Equipment qualificationEquipment qualification
Equipment qualification
 
Pharmaceutical validation of water system
Pharmaceutical validation of  water system Pharmaceutical validation of  water system
Pharmaceutical validation of water system
 
Introduction to validation
Introduction to validation Introduction to validation
Introduction to validation
 
Equipment validation of fbd
Equipment validation of fbdEquipment validation of fbd
Equipment validation of fbd
 
Ampule filling and_sealing_machine.ppt1
Ampule filling and_sealing_machine.ppt1Ampule filling and_sealing_machine.ppt1
Ampule filling and_sealing_machine.ppt1
 
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steam
 
Vendor qualification
Vendor qualificationVendor qualification
Vendor qualification
 
Process Validation of Liquid Orals
Process Validation of Liquid OralsProcess Validation of Liquid Orals
Process Validation of Liquid Orals
 
Validation of hvac system
Validation of hvac systemValidation of hvac system
Validation of hvac system
 

Similar to Qualification of membrane filtration apparatus

Filter validation
Filter validationFilter validation
Filter validationVINOTH R
 
Audit ppt.pptx
Audit ppt.pptxAudit ppt.pptx
Audit ppt.pptxNilam71
 
solid tablet validation
solid tablet validationsolid tablet validation
solid tablet validationaqsa fatima
 
Validation utility system
Validation utility systemValidation utility system
Validation utility systemPRANJAY PATIL
 
types of validation
types of validation types of validation
types of validation AbdulNaim14
 
Validation utility system
Validation utility systemValidation utility system
Validation utility systemArpitSuralkar
 
Validation of equipments
Validation of equipmentsValidation of equipments
Validation of equipmentssuhasini
 
Qualification of manufacturing equipment.
Qualification of manufacturing equipment.Qualification of manufacturing equipment.
Qualification of manufacturing equipment.KhushbooKunkulol
 
validation of tablet and capsule formulation8
validation of tablet and capsule formulation8validation of tablet and capsule formulation8
validation of tablet and capsule formulation8tanvikumbhar
 
b.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptxb.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptxAarthi358332
 
design & operation of QC lab by nikita kakad
design & operation of QC lab by nikita kakaddesign & operation of QC lab by nikita kakad
design & operation of QC lab by nikita kakadnikita kakad
 
validation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfvalidation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfMd. Omar Faruq Khan
 
Equipment validation
Equipment validationEquipment validation
Equipment validationGaurav Kr
 
Cleaning validation
Cleaning validation Cleaning validation
Cleaning validation Alekya Chinki
 
Pharmaceutical Validation
 Pharmaceutical Validation  Pharmaceutical Validation
Pharmaceutical Validation AmnThakur
 

Similar to Qualification of membrane filtration apparatus (20)

Filter validation
Filter validationFilter validation
Filter validation
 
Audit ppt.pptx
Audit ppt.pptxAudit ppt.pptx
Audit ppt.pptx
 
solid tablet validation
solid tablet validationsolid tablet validation
solid tablet validation
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Autoclave validation maliba
Autoclave validation malibaAutoclave validation maliba
Autoclave validation maliba
 
types of validation
types of validation types of validation
types of validation
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Validation of equipments
Validation of equipmentsValidation of equipments
Validation of equipments
 
Phụ lục 2. Tiêu chuẩn GMP WHO cho phòng thí nghiệm vi sinh dược phẩm
Phụ lục 2. Tiêu chuẩn GMP WHO cho phòng thí nghiệm vi sinh dược phẩmPhụ lục 2. Tiêu chuẩn GMP WHO cho phòng thí nghiệm vi sinh dược phẩm
Phụ lục 2. Tiêu chuẩn GMP WHO cho phòng thí nghiệm vi sinh dược phẩm
 
Qualification of manufacturing equipment.
Qualification of manufacturing equipment.Qualification of manufacturing equipment.
Qualification of manufacturing equipment.
 
validation of tablet and capsule formulation8
validation of tablet and capsule formulation8validation of tablet and capsule formulation8
validation of tablet and capsule formulation8
 
b.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptxb.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptx
 
Manufacturing facility of parentarals as per schedule m
Manufacturing facility of parentarals as per schedule mManufacturing facility of parentarals as per schedule m
Manufacturing facility of parentarals as per schedule m
 
design & operation of QC lab by nikita kakad
design & operation of QC lab by nikita kakaddesign & operation of QC lab by nikita kakad
design & operation of QC lab by nikita kakad
 
validation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfvalidation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdf
 
EU GMP Revise.pdf
EU GMP Revise.pdfEU GMP Revise.pdf
EU GMP Revise.pdf
 
Equipment validation
Equipment validationEquipment validation
Equipment validation
 
Cssd ppt
Cssd pptCssd ppt
Cssd ppt
 
Cleaning validation
Cleaning validation Cleaning validation
Cleaning validation
 
Pharmaceutical Validation
 Pharmaceutical Validation  Pharmaceutical Validation
Pharmaceutical Validation
 

Recently uploaded

Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableJanvi Singh
 
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Janvi Singh
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Janvi Singh
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...dishamehta3332
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...rajnisinghkjn
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan 087776558899
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableSteve Davis
 
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...dilbirsingh0889
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryJyoti singh
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxSwetaba Besh
 
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service AvailableLucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Availablesoniyagrag336
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Janvi Singh
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsMedicoseAcademics
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...gragneelam30
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesMedicoseAcademics
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...soniyagrag336
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowtanudubay92
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...Rashmi Entertainment
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxSwetaba Besh
 

Recently uploaded (20)

Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service AvailableLucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 

Qualification of membrane filtration apparatus

  • 1. 1
  • 2. CONTENT • Qualification • Membrane filtration • Types of filter • Design qualification • Operational qualification • Installation qualification • Performance qualification • Validation study element • References 2
  • 3. QUALIFICATION • Qualification is defined as an action of providing that equipment or systems are properly installed, work correctly and actually leads to expected results. • Qualification is part of validation. Types of qualification: I. Design qualification II. Installation qualification III. Operational qualification IV. Performance qualification 3
  • 4. MEMBRANE FILTRATION • Unit operation of filtration is the separation of solids from a liquid by passage through a filter medium. • Membrane filtration is used for sterilization of drug product and used in sterilization process. 4
  • 5. TYPES OF FILTER There are two types of filter used in filtration process: I. Depth filters: Consist of fibrous or granular materials. II. Membrane filters: These are porous membrane. 5
  • 6. DESIGN QUALIFICATION (DQ) • DQ defines the functional and operational specifications of an instrument. • DQ details the conscious decisions made in the selection of the supplier. • DQ should ensure that instruments have all the necessary functions and performance criteria that will enable them to meet user requirements. • The list shows the recommended steps should considered during DQ 6
  • 7. 7 Description of the analyse problem Description of the intended use for the equipment Description of the intended environment Preliminary selection of the functional and performance specifications Preliminary selection of the supplier Development and documentation of final functional and operational specifications Final selection of the supplier and equipment
  • 8. OPERATIONAL QUALIFICATION (OQ) • OQ is the process of demonstrating that instrument will function according to its operational specification in the selected environment. • Before OQ testing is done, one should always consider what the instrument will be used for. • Testing may be quite extensive if the instrument is to be used for all types of applications. 8
  • 9. INSTALLATION QUALIFICATION(IQ) • IQ establishes that the instrument is received as designed and specified, that it is properly installed in the selected environment. • IQ tell about environment that environment is suitable for the operation and use of the instrument. 9
  • 10. Performance Qualification (PQ) • PQ is the process of demonstrating that an instrument consistently performs according to a specification appropriate to its routine use. • PQ should be performed on a daily (or at least a weekly) basis whenever instrument is used. • The test frequency depends not only on the stability of the equipment but also on everything in the system that may contribute to the analysis results. 10
  • 11. VALIDATION STUDY ELEMENT 11 Physical  Reproducibility of filter Sterilization Integrity test Operating condition Shedding Microbial challenge test  Biological  Endotoxin  Toxicity Chemical Inertness Activity/stability Test for antimicrobial activities Consistency and reliability
  • 12. 1. REPRODUCIBILITY OF FILTER Membrane filters should be routinely discarded. Factor that can affect in filtration process include viscosity, surface tension of material to be filtered, ph, flow rates, time, temperature, osmolarity etc. 12
  • 13. 2. STERILIZATION Validation of sterilization method of filter is necessary because filter it self cause contamination of the product. To validate use of sterilizing grade filter it is not only prove that the filter is adequately sterilized but also method does not damage the filter.  Most preferred method is moist heat sterilizing. 13
  • 14. 3. INTEGRITY TEST It should be non destructive and provide an indication of “ fitness for use ”.  This include bubble point test, diffusion test, pressure hold test, water intrusion test.  This test of filter should be performed prior to processing and should be performed routinely and conducted after filtration to detect any filter leaks or perforation that might have occurred during the filtration. 14
  • 15. 4. OPERATING CONDITION a. Time: Long processing time could allow bacteria which have been trapped by the filter. Filter manufacturer can provide the data on the retention tests that have been conducted for specific membrane and generally suggest that filter should retain bacteria excess of 48hr. b. Temperature: Manufacture of filter recommended the limit and these not exceeded. c. Pressure: Inlet pressure to the filter must be monitored to ensure that there is no potential for structure damage. 15
  • 16. 5. Filter inertness  There may be extraction and adsorption phenomena occurs.  Stability of the product should not be affected by the filter 16
  • 17. 6. TEST FOR ANTIMICROBIAL ACTIVITY The test is performed to ensure that, any residual of Antimicrobial Activity is satisfactory eliminated by using the steps mentioned in this protocol. An inoculums of viable cells of the specific bacteria and fungi has been passed through the filter, inoculate filter paper in FTM & incubate at 30 to 350C or in SCDM and incubate at 20 to 250C. If conspicuous growth does not occur within 3 days for bacteria and 5 days for fungi, the test procedure is not valid and must be modified 17
  • 18. 7. ENDOTOXIN  Validation must address filter does not add endotoxin to drug product.  It depend on quality control process of the filter manufacturer, water used in manufacturing, choice of filter vendor, verification are not done properly. 18
  • 19. 8. TOXICITY  A validation study should determine that passage of the drug product through a filter does not cause any toxicological effects  Construction material off filtration system should be non toxic 19
  • 20. REFRENCES • R. A. Nash and A. H. Watche, Pharmaceutical process validation, Third edition • https://www.researchgate.net/publication/27 5519051_Review_on_Validation_of_Autoclave _Membrane_Filtration • https://www.slideshare.net/sagarsavale1/vali dation-of-membrane-filter 20
  • 21. 21