SYNOPTOPHORE
Information guide
(MAJOR AMBLYOSCOPE)
02 | 03 SYNOPTOPHORE - A USERS GUIDE
Synoptophore
Versatile, accurate & easy-to-
use
02 | 03 |
For generations, the Synoptophore has been THE standard
instrument of choice for the busy Orthoptic clinic. It is ideal
for the assessment and treatment of ocular motility disorders
by reliably performing the most comprehensive binocular
vision assessment available today.
The versatile and accurate Synoptophore provides precise
measurements in a single, effortless examination.
Easy-to-use
This information has been designed for eye care professionals and is intended as a quick
and easy companion, not as a formal teaching aid.
As each assessment is highly individual the proficient use of the Synoptophore comes
from experience and skill in interpreting the patient responses correctly.
Introduction
The Synoptophore design has undergone many changes over the years. In 1838 Sir Charles
Wheatstone constructed the first stereoscope. In the early 20th century Claud Worth
produced the Amblyoscope, to evaluate and stimulate binocular vision. To determine the
extent of simultaneous perception and measure the area of suppression M.C. Maddox
first developed slides to be used in these early devices.
Initially based on these instruments the Synoptophore (Major Amblyoscope) has evolved
from measuring angles of deviation, treating binocular anomalies and has utilised
pleoptics since Clement Clarke’s first patent in 1929.
History
The 2001 is the most comprehensive model in the range. All standard measurements
and treatments are possible, and the model boasts an automatic flashing unit allowing
each eye to be stimulated alternately or simultaneously. A removable pair of Haidinger’s
Brushes motor units are provided as standard, allowing users to employ pleoptic
techniques to assess and treat eccentric fixation and abnormal retinal correspondance.
3 models available
2001 model
The 2002 model combines all the features of the 2001, with the exception of the
Haidinger’s Brushes. It is ideal for the assessment and treatment of patients needing
orthoptic care or management.
2002 model
The 2003 model provides all the standard measurements and treatments including the
manual flashing switches and independent dimming rheostats.
2003 model
Synoptophore (sin-op-to-phore) Greek: syn = with, ops =
eye, phoros = bearing
Standard measurements and treatments include the
assessment of; the objective and subjective angle of deviation,
abnormal retinal correspondence, cyclophoria, hyperphoria
and horizontal and vertical vergences.
04 |
Each slide size has been calculated to subtend a
different angle at the nodal point of the eye.
05 |
The eye-pieces consist of +6.50 DS colli-
mating lenses, which require the patient to
relax their accommodation, as if looking into the
infinite distance.A plane, silvered, mirror reflects
Basic principles
DEGREE & DIOPTRE SCALE
All grades of Binocular Single Vision (BSV) can
be assessed on the Synoptophore.
As shown in the picture above - 1Δ = 100 X tan (°) - Therefore 1° = 1.75 Δ
Basic operation
The first grade of BSV, Simultaneous percep-
tion (red binding), is tested using two dissimilar
pictures, such as a lion and a cage.
Simultaneous Perception (SP)
90° along the two, cylindrical-conical, optical
tubes. These tubes pivot horizontally, vertically
and torsionally on their supports so that the two
presented images may be moved in relation to
each other.
At the end of the tubes there is a light source,
which evenly illuminates the transparencies, as a
plastic screen diffuses light from the lamp bulbs.
As the Synoptophore is adjusted, to pro-
duce the required measurements, a series of
alternately sized slides as well as various scales
are required.
The scales on the Synoptophore, to measure the
displacement, are degrees and Prism Dioptres
(Δ).
The tubes are objectively (by the examiner) and
subjectively (by the patient) adjusted so that
either the lion is perceived to be inside the cage
or one image is suppressed.
Foveal
Macular
Paramacular
Slide size
Peripheral
1°
1° -3°
3° -5°
> 5°
Angle subtended
06 |
07 |
The patient should be seated at the
Synoptophore
The inter-pupillary distance (IPD) should be
adjusted so the lines on the eyepiece line-up
with the corneal reflections
The smaller picture, i.e. the lion, should be
placed in the slide holder in front of the fixing
eye, whilst the larger picture should be placed
in front of the fellow eye.
IPD SCALE
Subjective - The patient pulls/pushes the han-
dle controlling the non-fixing eye’s tube until the
two images are superimposed. If this is difficult
or not possible suppression may be present and
a larger target should be introduced, however, if
superimposition is not achieved with peripher-
al slides then the patient has no potential BSV.
Near - To calculate the amount of deviation at
near the process can be replicated with -3.00
DS lenses placed in the eye-piece’s lens hold-
er This induces 3.00 DS of accommodation,
simulating near viewing.
FUSION SCALE
Objective - By alternatively switching off the lights
illuminating the slides an alternate cover test is
performed. The patient’s eyes are dissociated
and as the single illuminated picture is projected
To assess the second grade of BSV, sensory
fusion (green slides), two similar pictures each
with an incomplete “control” (i.e. rabbit’s tail
and a bunch of flowers) are presented to the
patient. One tube is locked, at zero, and the
patient is instructed to move the image, as
with SP, and create a composite image of the
rabbit holding a bunch of flowers. It is important
to question the patient about the “controls” to
prove sensory fusion or assess for the pres-
ence of suppression. Yet again, if this is difficult
for the patient, larger slides may be beneficial.
Sensory (SF) & Motor Fusion (MF)
NB: When testing motor fusion the images
may become de-focused or alter in size
before fusion fails.
TIP: It is worthwhile asking questions
about the pictures to stimulate the
patient’s attention i.e. “does the lion look
hungry?”
onto each fovea alternately a re-fixation move-
ment occurs. The direction of the eye movement
is examined and the tube before the non fixing
eye is adjusted until no eye movement is seen, or
reversal of movement is noticed. The measure-
ment is then read off the scale in degrees. This
can be repeated for vertical deviations however
measurements should be read off the scale in
prism dioptres (Δ).
NB: If the fixation is poor the corneal reflections
alone may be used to assess the angle of
deviation.
The position sensory fusion was achieved
is read off the scale and the range of
motor fusion is then tested by locking the
columns at this corrected angle and converging/
diverging the tubes until either control disap-
pears or the image splits into two. The vergenc-
es may then be read off the scale in degrees.
TIP: Observing the corneal reflections is
beneficial when assessing fusion, as it
will indicate when suppression occurs.
A gross qualitative stereopsis (yellow slides)
assessment can be obtained using two
images of the same object, hypothetically
taken from slightly different angles, to
indicate depth perception. The slides are
inserted into the slide holders with the controls
of each slide positioned towards or away from
the subject. The patient is then required to
describe the apparent effect i.e. a swing moving
towards or away from them. The test should be
repeated, with the slide controls changing
position, to ensure the stereoscopic effect
is authentic and the patient is not guessing.
08 | 09 |
Stereopsis
Whilst assessing simultaneous perception
retinal correspondence may also be exam-
ined. When the objective and subjective
angles are equal normal retinal correspond-
ence (NRC) is present, however, should the
objective and objective differ an “angle of anom-
aly” is present. If this equals the objective angle
abnormal retinal correspondence (ARC) is
“harmonious”, if the objective angle is larger than
the angle of anomaly ARC is “unharmonious”.
Retinal Correspondence
NB: Retinal correspondence can also be
determined on the Synoptophore using special
Bielchowsky after image test slides and high
intensity lamps.
SLIDE PROJECTORS
The gradient method is most often used,
when accommodation, minus without
accommodation is divided by the change in
accommodation. This can be assessed on
the Synoptophore by adjusting the IPD to the
patient, placing foveal SP slides into the slide
holders and measuring the objective angle (Δ),
then repeating with -3.00DS (or a lower power).
Accommodative convergence to
The variety of movements the Synopto-
phore allows permits several other tests to be
undertaken. Some of the most prevalent are
AC/A ratio, plotting the area and density of
suppression scotomas, assessing 9 positions
of gaze and the subjective measure of torsion.
accommodation ratio (AC/A ratio)
i.e. +18Δ - +9Δ / 3DS = 3:1
The Graphic method however plots com-
parisons of normal convergence against the
subjective angle, as minus lenses of increasing
power (usually -1.00DS to -4.00DS) are used.
As the test is standardised (fixation distance,
illumination etc) Accommodative Convergence
is the only variable affecting the deviation.
Objective angle
Subjective angle
Fusion at subj.
Example report
Stereopsis at subj.
+6° (macular slides)
+2°
5 conv - 3 div (macular slides)
Gross only
In complex ocular motility cases, all 9 car-
dinal positions of gaze can be subjectively
measured along with unilateral ductions with
repeatable, standardised conditions.
Maddox slides (white binding) can aid the
assessment of 9 positions of gaze. Horizon-
tal and vertical deviations are assessed in the
normal way. However, the examiner may rotate
the torsional control until the patient is satisfied
that it superimposes in the centre of the green
surround and all lines should run parallel.
Want to find out more? For further
information on the Synoptophore,
please telephone (01279) 456255 or email
synoptophore@haag-streit-uk.com.
10 | 11 |
As the rheostat controls the illumination pre-
sented to the fixing eye lowering the illumination
until simultaneous perception is achievable
gives an estimate of the density of suppres-
sion. This is not however recommended for
9 Positions of gaze
HEIGHT CONTROL
Torsion
The area of suppression may be mapped out by
initially recording the angle at which the image is
suppressed, then as the tube is rotated horizon-
tally or vertically record when both pictures are
again apparent and subtract one from the other.
Suppression
TIP: In dense suppression it may
be required to press the slide ejec-
tion lever or tap the light hous-
ing to draw attention to the image.
TIP: It may be useful to hold the pa-
tient’s head whilst testing inconcomitant
deviations.
Further information
The subjective measurement can be performed
fixing either eye in the primary position, when
the central lock is released on lateral versions
and using the elevation and depression controls
up to +/- 30° vertically.
Using an appropriate size slide, the arm of
the Synoptophore may be progressively moved
from zero degrees into the defective position
of gaze, whilst the patient maintains
fixation. When either the patient responds or
it becomes apparent the eyes have stopped
following the fixation target a duction measure-
ment is recorded, in degrees.
dense suppression.
E&OE. 1902072/Issue 2/November 2014
HAAG-STREIT UK
Unit C
Woodside Industrial Estate
Dunmow Road
Bishop’s Stortford
Hertfordshire
CM23 5RG
(01279) 456255
sales@haag-streit-uk.com
Common terms
Angle alpha – The angle between optical axis
and the visual axis. A positive angle, (corneal
reflection placed nasally) causes pseudo
divergent squint; negative angle (corneal
reflections placed temporally) causes
pseudo-convergent squint.
Angle gamma – The angle between the optical
axis and the fixation axis.
Angle kappa – The angle between the
mid-pupillarly line and visual axis.
Angle of anomaly – The difference
between objective and subjective angles of
deviation.
Motor fusion – The ability to
maintain sensory fusion through a range of
vergence, which may be horizontal, vertical or
cyclovergence.
Sensory fusion – The ability to
perceive two similar images, one formed on
each retina, and interpret them as one.
Simultaneous perception – The ability to
perceive simultaneously two images, one formed
on each retina.
Stereopsis – The perception of relative depth of
objects on the basis of binocular disparity.
Synoptophore (Major Amblyoscope) –
Equipment which can be used to assess the
angle of deviation and binocular potential at a
theoretical distance fixation.
Superimposition – The simultaneous
perception of the two images formed on
corresponding areas, with the projection of
these images to the same poison in space. This
may occur whether retinal correspondence is
normal or abnormal. If fusion is absent two
similar images are seen as separate but
superimposed and no fusion range is
demonstrable.

More Related Content

Similar to Synoptophore Information Guide

Methods of subjective refraction
Methods of subjective refractionMethods of subjective refraction
Methods of subjective refractionMaryam Fashola
 
Optom Anamul Haq( synoptophore)
Optom Anamul Haq( synoptophore) Optom Anamul Haq( synoptophore)
Optom Anamul Haq( synoptophore) OptomAnamulHaq
 
Keratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKeratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKamal Luitel
 
Ophthalmoscopy
OphthalmoscopyOphthalmoscopy
OphthalmoscopySalalKhan5
 
Contact Lenses and LVAs.pptx
Contact Lenses and LVAs.pptxContact Lenses and LVAs.pptx
Contact Lenses and LVAs.pptxZaid Azhar
 
Motor evaluation of strabismus
Motor evaluation of strabismusMotor evaluation of strabismus
Motor evaluation of strabismusGauriSShrestha
 
Objective refraction
Objective refractionObjective refraction
Objective refractionMOHAMMEDJN
 
Visual function assessment in low vision
Visual function assessment in low visionVisual function assessment in low vision
Visual function assessment in low visionPuneet
 
Lenses of slit lamp biomicroscope & indirect ophthalmoscope.
Lenses of slit lamp biomicroscope & indirect ophthalmoscope.Lenses of slit lamp biomicroscope & indirect ophthalmoscope.
Lenses of slit lamp biomicroscope & indirect ophthalmoscope.Ayat AbuJazar
 
Correcting presbyopia - Modern Options
Correcting presbyopia - Modern OptionsCorrecting presbyopia - Modern Options
Correcting presbyopia - Modern OptionsJason Higginbotham
 
Objective retinoscopy
Objective retinoscopyObjective retinoscopy
Objective retinoscopyBipin Koirala
 
Indirect Ophthalmoscopy and slit lamp biomicroscopy
Indirect Ophthalmoscopy and slit lamp biomicroscopyIndirect Ophthalmoscopy and slit lamp biomicroscopy
Indirect Ophthalmoscopy and slit lamp biomicroscopyLakshmi Murthy
 

Similar to Synoptophore Information Guide (20)

Methods of subjective refraction
Methods of subjective refractionMethods of subjective refraction
Methods of subjective refraction
 
Optom Anamul Haq( synoptophore)
Optom Anamul Haq( synoptophore) Optom Anamul Haq( synoptophore)
Optom Anamul Haq( synoptophore)
 
Synaptophore in ophthalmology
Synaptophore in ophthalmologySynaptophore in ophthalmology
Synaptophore in ophthalmology
 
Keratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKeratometry and Dynamic Retinoscopy
Keratometry and Dynamic Retinoscopy
 
OPTHALMOSCOPY.pdf
OPTHALMOSCOPY.pdfOPTHALMOSCOPY.pdf
OPTHALMOSCOPY.pdf
 
refractive technique
refractive techniquerefractive technique
refractive technique
 
Optometry instruments
Optometry instruments Optometry instruments
Optometry instruments
 
Ophthalmoscopy
OphthalmoscopyOphthalmoscopy
Ophthalmoscopy
 
Ophthalmoscopy
OphthalmoscopyOphthalmoscopy
Ophthalmoscopy
 
Ophthalmoscopy
OphthalmoscopyOphthalmoscopy
Ophthalmoscopy
 
Contact Lenses and LVAs.pptx
Contact Lenses and LVAs.pptxContact Lenses and LVAs.pptx
Contact Lenses and LVAs.pptx
 
4 RETINOSCOPY.pptx
4 RETINOSCOPY.pptx4 RETINOSCOPY.pptx
4 RETINOSCOPY.pptx
 
Motor evaluation of strabismus
Motor evaluation of strabismusMotor evaluation of strabismus
Motor evaluation of strabismus
 
Objective refraction
Objective refractionObjective refraction
Objective refraction
 
Visual function assessment in low vision
Visual function assessment in low visionVisual function assessment in low vision
Visual function assessment in low vision
 
Lenses of slit lamp biomicroscope & indirect ophthalmoscope.
Lenses of slit lamp biomicroscope & indirect ophthalmoscope.Lenses of slit lamp biomicroscope & indirect ophthalmoscope.
Lenses of slit lamp biomicroscope & indirect ophthalmoscope.
 
Correcting presbyopia - Modern Options
Correcting presbyopia - Modern OptionsCorrecting presbyopia - Modern Options
Correcting presbyopia - Modern Options
 
Iol
IolIol
Iol
 
Objective retinoscopy
Objective retinoscopyObjective retinoscopy
Objective retinoscopy
 
Indirect Ophthalmoscopy and slit lamp biomicroscopy
Indirect Ophthalmoscopy and slit lamp biomicroscopyIndirect Ophthalmoscopy and slit lamp biomicroscopy
Indirect Ophthalmoscopy and slit lamp biomicroscopy
 

More from Haag-Streit UK (HS-UK)

Haag-Streit UK InFocus Newsletter Issue 12
Haag-Streit UK InFocus Newsletter Issue 12Haag-Streit UK InFocus Newsletter Issue 12
Haag-Streit UK InFocus Newsletter Issue 12Haag-Streit UK (HS-UK)
 
Haag-Streit Simulation Eyesi Indirect ROP Brochure
Haag-Streit Simulation Eyesi Indirect ROP BrochureHaag-Streit Simulation Eyesi Indirect ROP Brochure
Haag-Streit Simulation Eyesi Indirect ROP BrochureHaag-Streit UK (HS-UK)
 
John Weiss - How to Care for your surgical instruments
John Weiss - How to Care for your surgical instrumentsJohn Weiss - How to Care for your surgical instruments
John Weiss - How to Care for your surgical instrumentsHaag-Streit UK (HS-UK)
 
Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024Haag-Streit UK (HS-UK)
 
MedOne MicroDose Silicone Free Injection Kit
MedOne MicroDose Silicone Free Injection KitMedOne MicroDose Silicone Free Injection Kit
MedOne MicroDose Silicone Free Injection KitHaag-Streit UK (HS-UK)
 
John Weiss Microsurgical Instruments Catalogue
John Weiss Microsurgical Instruments CatalogueJohn Weiss Microsurgical Instruments Catalogue
John Weiss Microsurgical Instruments CatalogueHaag-Streit UK (HS-UK)
 
Convert your slit lamp today - swap tungsten for LED
Convert your slit lamp today - swap tungsten for LEDConvert your slit lamp today - swap tungsten for LED
Convert your slit lamp today - swap tungsten for LEDHaag-Streit UK (HS-UK)
 
John Weiss & HS-UK Newsletter - Issue 5
John Weiss & HS-UK Newsletter - Issue 5John Weiss & HS-UK Newsletter - Issue 5
John Weiss & HS-UK Newsletter - Issue 5Haag-Streit UK (HS-UK)
 
John Weiss & HS-UK Newsletter - Issue 6
John Weiss & HS-UK Newsletter - Issue 6John Weiss & HS-UK Newsletter - Issue 6
John Weiss & HS-UK Newsletter - Issue 6Haag-Streit UK (HS-UK)
 
HS-UK Guide to Checking Tonometer Calibration
HS-UK Guide to Checking Tonometer CalibrationHS-UK Guide to Checking Tonometer Calibration
HS-UK Guide to Checking Tonometer CalibrationHaag-Streit UK (HS-UK)
 

More from Haag-Streit UK (HS-UK) (20)

Haag-Streit UK InFocus Newsletter Issue 12
Haag-Streit UK InFocus Newsletter Issue 12Haag-Streit UK InFocus Newsletter Issue 12
Haag-Streit UK InFocus Newsletter Issue 12
 
Haag-Streit Simulation Eyesi Indirect ROP Brochure
Haag-Streit Simulation Eyesi Indirect ROP BrochureHaag-Streit Simulation Eyesi Indirect ROP Brochure
Haag-Streit Simulation Eyesi Indirect ROP Brochure
 
John Weiss - How to Care for your surgical instruments
John Weiss - How to Care for your surgical instrumentsJohn Weiss - How to Care for your surgical instruments
John Weiss - How to Care for your surgical instruments
 
Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024Haag-Streit UK Diagnostics Price List 2024
Haag-Streit UK Diagnostics Price List 2024
 
MedOne MicroDose Silicone Free Injection Kit
MedOne MicroDose Silicone Free Injection KitMedOne MicroDose Silicone Free Injection Kit
MedOne MicroDose Silicone Free Injection Kit
 
John Weiss Microsurgical Instruments Catalogue
John Weiss Microsurgical Instruments CatalogueJohn Weiss Microsurgical Instruments Catalogue
John Weiss Microsurgical Instruments Catalogue
 
Convert your slit lamp today - swap tungsten for LED
Convert your slit lamp today - swap tungsten for LEDConvert your slit lamp today - swap tungsten for LED
Convert your slit lamp today - swap tungsten for LED
 
HS-UK UK Diagnostics Catalogue 2024
HS-UK UK Diagnostics Catalogue 2024HS-UK UK Diagnostics Catalogue 2024
HS-UK UK Diagnostics Catalogue 2024
 
HS-UK Newsletter Issue 11
HS-UK Newsletter Issue 11HS-UK Newsletter Issue 11
HS-UK Newsletter Issue 11
 
MedOne brochure
MedOne brochureMedOne brochure
MedOne brochure
 
KERARING brochure
KERARING brochureKERARING brochure
KERARING brochure
 
Tonosafe brochure
Tonosafe brochureTonosafe brochure
Tonosafe brochure
 
MR Q SLT brochure
MR Q SLT brochureMR Q SLT brochure
MR Q SLT brochure
 
HS-UK & John Weiss Newsletter Issue 7
HS-UK & John Weiss Newsletter Issue 7HS-UK & John Weiss Newsletter Issue 7
HS-UK & John Weiss Newsletter Issue 7
 
John Weiss & HS-UK Newsletter - Issue 5
John Weiss & HS-UK Newsletter - Issue 5John Weiss & HS-UK Newsletter - Issue 5
John Weiss & HS-UK Newsletter - Issue 5
 
John Weiss & HS-UK Newsletter - Issue 6
John Weiss & HS-UK Newsletter - Issue 6John Weiss & HS-UK Newsletter - Issue 6
John Weiss & HS-UK Newsletter - Issue 6
 
HS-UK Newsletter Issue 10
HS-UK Newsletter Issue 10HS-UK Newsletter Issue 10
HS-UK Newsletter Issue 10
 
Hill RBF Calculator
Hill RBF CalculatorHill RBF Calculator
Hill RBF Calculator
 
Synoptophore Slide Catalogue
Synoptophore Slide CatalogueSynoptophore Slide Catalogue
Synoptophore Slide Catalogue
 
HS-UK Guide to Checking Tonometer Calibration
HS-UK Guide to Checking Tonometer CalibrationHS-UK Guide to Checking Tonometer Calibration
HS-UK Guide to Checking Tonometer Calibration
 

Recently uploaded

Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...narwatsonia7
 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...Miss joya
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...Miss joya
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Servicenarwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night EnjoyCall Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoynarwatsonia7
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls ServiceMiss joya
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoybabeytanya
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 

Recently uploaded (20)

Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night EnjoyCall Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 

Synoptophore Information Guide

  • 2. 02 | 03 SYNOPTOPHORE - A USERS GUIDE Synoptophore Versatile, accurate & easy-to- use 02 | 03 | For generations, the Synoptophore has been THE standard instrument of choice for the busy Orthoptic clinic. It is ideal for the assessment and treatment of ocular motility disorders by reliably performing the most comprehensive binocular vision assessment available today. The versatile and accurate Synoptophore provides precise measurements in a single, effortless examination. Easy-to-use This information has been designed for eye care professionals and is intended as a quick and easy companion, not as a formal teaching aid. As each assessment is highly individual the proficient use of the Synoptophore comes from experience and skill in interpreting the patient responses correctly. Introduction The Synoptophore design has undergone many changes over the years. In 1838 Sir Charles Wheatstone constructed the first stereoscope. In the early 20th century Claud Worth produced the Amblyoscope, to evaluate and stimulate binocular vision. To determine the extent of simultaneous perception and measure the area of suppression M.C. Maddox first developed slides to be used in these early devices. Initially based on these instruments the Synoptophore (Major Amblyoscope) has evolved from measuring angles of deviation, treating binocular anomalies and has utilised pleoptics since Clement Clarke’s first patent in 1929. History The 2001 is the most comprehensive model in the range. All standard measurements and treatments are possible, and the model boasts an automatic flashing unit allowing each eye to be stimulated alternately or simultaneously. A removable pair of Haidinger’s Brushes motor units are provided as standard, allowing users to employ pleoptic techniques to assess and treat eccentric fixation and abnormal retinal correspondance. 3 models available 2001 model The 2002 model combines all the features of the 2001, with the exception of the Haidinger’s Brushes. It is ideal for the assessment and treatment of patients needing orthoptic care or management. 2002 model The 2003 model provides all the standard measurements and treatments including the manual flashing switches and independent dimming rheostats. 2003 model Synoptophore (sin-op-to-phore) Greek: syn = with, ops = eye, phoros = bearing Standard measurements and treatments include the assessment of; the objective and subjective angle of deviation, abnormal retinal correspondence, cyclophoria, hyperphoria and horizontal and vertical vergences.
  • 3. 04 | Each slide size has been calculated to subtend a different angle at the nodal point of the eye. 05 | The eye-pieces consist of +6.50 DS colli- mating lenses, which require the patient to relax their accommodation, as if looking into the infinite distance.A plane, silvered, mirror reflects Basic principles DEGREE & DIOPTRE SCALE All grades of Binocular Single Vision (BSV) can be assessed on the Synoptophore. As shown in the picture above - 1Δ = 100 X tan (°) - Therefore 1° = 1.75 Δ Basic operation The first grade of BSV, Simultaneous percep- tion (red binding), is tested using two dissimilar pictures, such as a lion and a cage. Simultaneous Perception (SP) 90° along the two, cylindrical-conical, optical tubes. These tubes pivot horizontally, vertically and torsionally on their supports so that the two presented images may be moved in relation to each other. At the end of the tubes there is a light source, which evenly illuminates the transparencies, as a plastic screen diffuses light from the lamp bulbs. As the Synoptophore is adjusted, to pro- duce the required measurements, a series of alternately sized slides as well as various scales are required. The scales on the Synoptophore, to measure the displacement, are degrees and Prism Dioptres (Δ). The tubes are objectively (by the examiner) and subjectively (by the patient) adjusted so that either the lion is perceived to be inside the cage or one image is suppressed. Foveal Macular Paramacular Slide size Peripheral 1° 1° -3° 3° -5° > 5° Angle subtended
  • 4. 06 | 07 | The patient should be seated at the Synoptophore The inter-pupillary distance (IPD) should be adjusted so the lines on the eyepiece line-up with the corneal reflections The smaller picture, i.e. the lion, should be placed in the slide holder in front of the fixing eye, whilst the larger picture should be placed in front of the fellow eye. IPD SCALE Subjective - The patient pulls/pushes the han- dle controlling the non-fixing eye’s tube until the two images are superimposed. If this is difficult or not possible suppression may be present and a larger target should be introduced, however, if superimposition is not achieved with peripher- al slides then the patient has no potential BSV. Near - To calculate the amount of deviation at near the process can be replicated with -3.00 DS lenses placed in the eye-piece’s lens hold- er This induces 3.00 DS of accommodation, simulating near viewing. FUSION SCALE Objective - By alternatively switching off the lights illuminating the slides an alternate cover test is performed. The patient’s eyes are dissociated and as the single illuminated picture is projected To assess the second grade of BSV, sensory fusion (green slides), two similar pictures each with an incomplete “control” (i.e. rabbit’s tail and a bunch of flowers) are presented to the patient. One tube is locked, at zero, and the patient is instructed to move the image, as with SP, and create a composite image of the rabbit holding a bunch of flowers. It is important to question the patient about the “controls” to prove sensory fusion or assess for the pres- ence of suppression. Yet again, if this is difficult for the patient, larger slides may be beneficial. Sensory (SF) & Motor Fusion (MF) NB: When testing motor fusion the images may become de-focused or alter in size before fusion fails. TIP: It is worthwhile asking questions about the pictures to stimulate the patient’s attention i.e. “does the lion look hungry?” onto each fovea alternately a re-fixation move- ment occurs. The direction of the eye movement is examined and the tube before the non fixing eye is adjusted until no eye movement is seen, or reversal of movement is noticed. The measure- ment is then read off the scale in degrees. This can be repeated for vertical deviations however measurements should be read off the scale in prism dioptres (Δ). NB: If the fixation is poor the corneal reflections alone may be used to assess the angle of deviation. The position sensory fusion was achieved is read off the scale and the range of motor fusion is then tested by locking the columns at this corrected angle and converging/ diverging the tubes until either control disap- pears or the image splits into two. The vergenc- es may then be read off the scale in degrees. TIP: Observing the corneal reflections is beneficial when assessing fusion, as it will indicate when suppression occurs.
  • 5. A gross qualitative stereopsis (yellow slides) assessment can be obtained using two images of the same object, hypothetically taken from slightly different angles, to indicate depth perception. The slides are inserted into the slide holders with the controls of each slide positioned towards or away from the subject. The patient is then required to describe the apparent effect i.e. a swing moving towards or away from them. The test should be repeated, with the slide controls changing position, to ensure the stereoscopic effect is authentic and the patient is not guessing. 08 | 09 | Stereopsis Whilst assessing simultaneous perception retinal correspondence may also be exam- ined. When the objective and subjective angles are equal normal retinal correspond- ence (NRC) is present, however, should the objective and objective differ an “angle of anom- aly” is present. If this equals the objective angle abnormal retinal correspondence (ARC) is “harmonious”, if the objective angle is larger than the angle of anomaly ARC is “unharmonious”. Retinal Correspondence NB: Retinal correspondence can also be determined on the Synoptophore using special Bielchowsky after image test slides and high intensity lamps. SLIDE PROJECTORS The gradient method is most often used, when accommodation, minus without accommodation is divided by the change in accommodation. This can be assessed on the Synoptophore by adjusting the IPD to the patient, placing foveal SP slides into the slide holders and measuring the objective angle (Δ), then repeating with -3.00DS (or a lower power). Accommodative convergence to The variety of movements the Synopto- phore allows permits several other tests to be undertaken. Some of the most prevalent are AC/A ratio, plotting the area and density of suppression scotomas, assessing 9 positions of gaze and the subjective measure of torsion. accommodation ratio (AC/A ratio) i.e. +18Δ - +9Δ / 3DS = 3:1 The Graphic method however plots com- parisons of normal convergence against the subjective angle, as minus lenses of increasing power (usually -1.00DS to -4.00DS) are used. As the test is standardised (fixation distance, illumination etc) Accommodative Convergence is the only variable affecting the deviation. Objective angle Subjective angle Fusion at subj. Example report Stereopsis at subj. +6° (macular slides) +2° 5 conv - 3 div (macular slides) Gross only
  • 6. In complex ocular motility cases, all 9 car- dinal positions of gaze can be subjectively measured along with unilateral ductions with repeatable, standardised conditions. Maddox slides (white binding) can aid the assessment of 9 positions of gaze. Horizon- tal and vertical deviations are assessed in the normal way. However, the examiner may rotate the torsional control until the patient is satisfied that it superimposes in the centre of the green surround and all lines should run parallel. Want to find out more? For further information on the Synoptophore, please telephone (01279) 456255 or email synoptophore@haag-streit-uk.com. 10 | 11 | As the rheostat controls the illumination pre- sented to the fixing eye lowering the illumination until simultaneous perception is achievable gives an estimate of the density of suppres- sion. This is not however recommended for 9 Positions of gaze HEIGHT CONTROL Torsion The area of suppression may be mapped out by initially recording the angle at which the image is suppressed, then as the tube is rotated horizon- tally or vertically record when both pictures are again apparent and subtract one from the other. Suppression TIP: In dense suppression it may be required to press the slide ejec- tion lever or tap the light hous- ing to draw attention to the image. TIP: It may be useful to hold the pa- tient’s head whilst testing inconcomitant deviations. Further information The subjective measurement can be performed fixing either eye in the primary position, when the central lock is released on lateral versions and using the elevation and depression controls up to +/- 30° vertically. Using an appropriate size slide, the arm of the Synoptophore may be progressively moved from zero degrees into the defective position of gaze, whilst the patient maintains fixation. When either the patient responds or it becomes apparent the eyes have stopped following the fixation target a duction measure- ment is recorded, in degrees. dense suppression.
  • 7. E&OE. 1902072/Issue 2/November 2014 HAAG-STREIT UK Unit C Woodside Industrial Estate Dunmow Road Bishop’s Stortford Hertfordshire CM23 5RG (01279) 456255 sales@haag-streit-uk.com Common terms Angle alpha – The angle between optical axis and the visual axis. A positive angle, (corneal reflection placed nasally) causes pseudo divergent squint; negative angle (corneal reflections placed temporally) causes pseudo-convergent squint. Angle gamma – The angle between the optical axis and the fixation axis. Angle kappa – The angle between the mid-pupillarly line and visual axis. Angle of anomaly – The difference between objective and subjective angles of deviation. Motor fusion – The ability to maintain sensory fusion through a range of vergence, which may be horizontal, vertical or cyclovergence. Sensory fusion – The ability to perceive two similar images, one formed on each retina, and interpret them as one. Simultaneous perception – The ability to perceive simultaneously two images, one formed on each retina. Stereopsis – The perception of relative depth of objects on the basis of binocular disparity. Synoptophore (Major Amblyoscope) – Equipment which can be used to assess the angle of deviation and binocular potential at a theoretical distance fixation. Superimposition – The simultaneous perception of the two images formed on corresponding areas, with the projection of these images to the same poison in space. This may occur whether retinal correspondence is normal or abnormal. If fusion is absent two similar images are seen as separate but superimposed and no fusion range is demonstrable.