Urilyzer® Cell
agile - affordable - accurate
Agenda
Specifications
1
Setup
2
Software
3
4 Workflow
5 Maintenance
agile - affordable - accurate
Specifications
• Up to 60 samples/hour
• 15 images/sample
Throughput
• Storage of 10.000 measurements with 15
images each
Database
• Only 175 µl per measurement
• Low space requirement
Volume
• 2x Ethernet, 2x USB 3.0, 2x USB 3.1, 1x RS232
• Specified Serial interfaces for Barcode Reader,
• Chemical Analyzer and LIS
Connectivity
• Easy connection to urine test strip analyzer
Urilyzer® 100 Pro / 500 Pro
• Whole Urinalysis result report on one device
Compatibility
• Urilyzer® Cell cuvettes (closed system,
disposable) Pipetting tips
• No reagents needed
Consumables
agile - affordable - accurate
Delivery contents
• Urilyzer® Cell
• Monitor
• Keyboard
• Mouse
agile - affordable - accurate
Setup
Load the cuvettes
Connect cabling
Start the device
agile - affordable - accurate
agile - affordable - accurate
Software
Lets switch to the device
agile - affordable - accurate
Connection to Chemical Analyzer
agile - affordable - accurate
Connection to Chemical Analyzer
The
Urilyzer®
Cell can be
easily
connected to
the Urilyzer
100/500 Pro
Backside of
the Urilyzer®
Cell (CHEM)
via serial
cable to
chemical
analyzer
Urilyzer®
100 Pro: Go
into Main 
Options 
Settings 
Output 
LIS2  set
Baudrate
Urilyzer®
500 Pro: Go
into Main 
Settings 
Output 
LIS2  set
Baudrate
Urilyzer®
Cell: Go into
Settings 
Chemical
Analyzer 
choose
„serial“ and
the
baudrate 
Activate
The
chemical
analyser has
to be
measured
before the
Urilyzer®
Cell
Use the
same ID on
both devices
for each
sample
• Best with
barcode reader
(2 required)
agile - affordable - accurate
Workflow – Manual procedure
Pressing
the Start
button to
load an
empty
cuvette to
pipetting
position
Barcode
scanning
Sample
mixing
Aspirating
urine
sample
using
manual
pipette
Injecting
urine
sample into
cuvette
Pressing
the Start
button to
start the
measurem
ent
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging process
Results are forwarded to Data
Manager Software
Cuvette is moved to next
microscope position
Focusing process
Cuvette is placed into first
microscope position
Recorded image is evaluated
HPF-like image is taken by
built-in camera
Used cuvette is placed into the
waste bin
Last image?
NO
YES
agile - affordable - accurate
agile - affordable - accurate
Pipetting practice
agile - affordable - accurate
Pipetting practice
agile - affordable - accurate
Workflow – Manual procedure
• Press the Start-Button on top of the device
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
agile - affordable - accurate
Workflow – Manual procedure
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
agile - affordable - accurate
Workflow – Manual procedure
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
agile - affordable - accurate
Workflow – Manual procedure
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
• Pick a pipette with 175 µl sample volume and aspirate the sample
• Outer needle diameter of 1,2 – 2,8 mm
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
agile - affordable - accurate
Workflow – Manual procedure
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
• Pick a pipette with 175 µl sample volume and aspirate the sample
• Outer needle diameter of 1,2 – 2,8 mm
• Fill the sample into the cuvette
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
agile - affordable - accurate
Workflow – Manual procedure
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
• Pick a pipette with 175 µl sample volume and aspirate the sample
• Outer needle diameter of 1,2 – 2,8 mm
• Fill the sample into the cuvette
• Press the Start-Button again to start the measurement process
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
agile - affordable - accurate
Workflow – Automated procedure
• Creating a monolayer of particles by
centrifuging the cuvette
10 seconds with 2000 RPM
Centrifuging
process
Results are
forwarded to Data
Manager Software
Cuvette is
moved to next
microscope
position
Focusing
process
Cuvette is placed
into first
microscope
position
Recorded
image is
evaluated
HPF-like image
is taken by
built-in camera
Used cuvette is
placed into the
waste bin
Last image?
NO
YES
agile - affordable - accurate
Workflow – Automated procedure
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
Centrifuging
process
Results are
forwarded to Data
Manager Software
Cuvette is
moved to next
microscope
position
Focusing
process
Cuvette is placed
into first
microscope
position
Recorded
image is
evaluated
HPF-like image
is taken by
built-in camera
Used cuvette is
placed into the
waste bin
Last image?
NO
YES
agile - affordable - accurate
Workflow – Automated procedure
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
• Focus series is captured in different
heights
Centrifuging
process
Results are
forwarded to Data
Manager Software
Cuvette is
moved to next
microscope
position
Focusing
process
Cuvette is placed
into first
microscope
position
Recorded
image is
evaluated
HPF-like image
is taken by
built-in camera
Used cuvette is
placed into the
waste bin
Last image?
NO
YES
agile - affordable - accurate
Workflow – Automated procedure
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
• Focus series is captured in different
heights
• Camera takes a picture of the
sharpest plane
Centrifuging
process
Results are
forwarded to Data
Manager Software
Cuvette is
moved to next
microscope
position
Focusing
process
Cuvette is placed
into first
microscope
position
Recorded
image is
evaluated
HPF-like image
is taken by
built-in camera
Used cuvette is
placed into the
waste bin
Last image?
NO
YES
agile - affordable - accurate
Workflow – Automated procedure
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
• Focus series is captured in different
heights
• Camera takes a picture of the
sharpest plane
• The system checks if this was the last
picture
• The image gets automatically
evaluated and is saved in the
database
Centrifuging
process
Results are
forwarded to Data
Manager Software
Cuvette is
moved to next
microscope
position
Focusing
process
Cuvette is placed
into first
microscope
position
Recorded
image is
evaluated
HPF-like image
is taken by
built-in camera
Used cuvette is
placed into the
waste bin
Last image?
NO
YES
agile - affordable - accurate
Workflow – Automated procedure
• The system checks if this was the last
picture
• YES: The cuvette is moved into the
waste bin
• NO: the cuvette position moves to
next microscope position
Centrifuging
process
Results are
forwarded to Data
Manager Software
Cuvette is
moved to next
microscope
position
Focusing
process
Cuvette is placed
into first
microscope
position
Recorded
image is
evaluated
HPF-like image
is taken by
built-in camera
Used cuvette is
placed into the
waste bin
Last image?
NO
YES
agile - affordable - accurate
Results
agile - affordable - accurate
Auto classification
• Red blood cells
• White blood cells
• White blood cell clumps
• Crystals
• Calcium-oxalate monohydrate
• Calcium-oxalate dihydrate
• Triple phosphate
• Uric acid
• Hyaline casts
• Pathological casts
• Non-squamous epithelial cells
• Squamous epithelial cells
• Yeast
• Bacteria
• Bacteria rods
• Bacteria cocci
• Mucus
• Spermatozoa
agile - affordable - accurate
Auto classification
Particle type:
Subclasses and other particles
can be added
Semi-quantitative category:
The ranges and category name
can be changed
Quantitative results:
p/µl or p/HPF
Negative result limits
RBC and WBC are real quantitative results
Every other result is semi-quantitative even
if the result is displayed quantitative
agile - affordable - accurate
Calculation
HPF-like
HPF
LPF-like
LPF
• 𝐻𝑃𝐹 − 𝑙𝑖𝑘𝑒 ∗ 1,5 = 𝐻𝑃𝐹
• 𝐿𝑃𝐹 − 𝑙𝑖𝑘𝑒 ∗ 0,5 = 𝐻𝑃𝐹
Semi-quantitative parameter
• 𝑝
µ𝑙 =
𝑝
15
∗ 6,6
• 𝑝
𝐻𝑃𝐹 =
𝑝
15
∗ 1,5
RBC and WBC calculation
• RBC: 𝑝
𝐻𝑃𝐹 ∗ 2
• WBC: 𝑝
𝐻𝑃𝐹 ∗ 1,5
Urilyzer® Cell p/HPF is correlated with 20x concentrated manual measurement.
agile - affordable - accurate
Focussing process
• 2000 steps ≡ 2,5 mm
• 1 step ≡ 1,25 µm
• Hard Focus Start ~ 800
• Set with the Reference Cuvette in
Service Software
2000 steps
Pic. 1
800
}320 steps
Taking 1 image every two steps 
chooses the best
Pic. 2 Pic. 3
} 16 steps up &
16 steps down
} 16 steps up &
16 steps down
Best image heigth of the
previous one defines the start
 taking 16 images in total and
choses the best again
Best image heigth of the
previous one defines the start
 taking 16 images in total and
choses the best again
And so on with the remaining 12 images
agile - affordable - accurate
Manual addable sediments
• Up to 50 parameters
are manual addable
• The red ones are only
manual addable
• Can be used for
further information
• Parameters must be
activated to add them
agile - affordable - accurate
Review criteria
0-2 review images  normal measurement:
• The results are calculated based on the non-review images  can be
auto-validated
≥ 3 review images and ≥ 5 O.K. images: review measurement:
• The results are calculated based on the non-review images  cannot
be auto-validated
< 5 O.K. images: review measurement:
• No automated result  cannot be auto-validated
• User can modify the final sample results based on the manual review
of the images or remeasure the sample with a dilution
MUC review:
• Concentration limit can be set for MUC: if the sample concentration is
above this limit, the measurement will be flagged for review
• The results are calculated based on the images  cannot be auto-
validated
agile - affordable - accurate
Maintenance
• Daily routine should maintain a
cleaning process after every workshift,
24 hours or according to the local
regulations
• What should be cleaned:
• Cuvette feeder arm
• Cuvette hold down assembly
• Drop tray
• Centrifuge cover
• Centrifuge arm
• The path of the cuvette, especially the
top of the microscope
• Waste bin
• The outer surface, especially the
pipetting window
agile - affordable - accurate
agile - affordable - accurate
We look forward to the opportunity to work with you.

More Related Content

Similar to IUCell001_Basic Handling_A.ppsx

Automation in Biochemistry (Autoanalyzers)
Automation in Biochemistry (Autoanalyzers)Automation in Biochemistry (Autoanalyzers)
Automation in Biochemistry (Autoanalyzers)Pradeep Singh Narwat
 
Item 4 aspac brisbane_workshop_qaqc_oct19
Item 4 aspac brisbane_workshop_qaqc_oct19Item 4 aspac brisbane_workshop_qaqc_oct19
Item 4 aspac brisbane_workshop_qaqc_oct19Soils FAO-GSP
 
Water Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check ProceduresWater Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check ProceduresSeaBirdScientific
 
A-PQA-Brochure-v1.2.4
A-PQA-Brochure-v1.2.4A-PQA-Brochure-v1.2.4
A-PQA-Brochure-v1.2.4Khairul Alam
 
Real time Vital Signs Monitorong Using NI LabVIEW
Real time Vital Signs Monitorong Using NI LabVIEWReal time Vital Signs Monitorong Using NI LabVIEW
Real time Vital Signs Monitorong Using NI LabVIEWTalha Patel
 
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...Chromatography & Mass Spectrometry Solutions
 
Final Presentation
Final PresentationFinal Presentation
Final PresentationBrady Bury
 
Lean Management and Kanban System
Lean Management and Kanban SystemLean Management and Kanban System
Lean Management and Kanban SystemGaurav Singh
 
4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINAL
4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINAL4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINAL
4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINALIlham Birkan
 
Poka yoke error proofing
Poka yoke error proofing Poka yoke error proofing
Poka yoke error proofing ssusercee91d
 
Lean Supply Chain Management--by JULIAN KALAC
Lean Supply Chain Management--by JULIAN KALACLean Supply Chain Management--by JULIAN KALAC
Lean Supply Chain Management--by JULIAN KALACJulian Kalac P.Eng
 
Introduction of Flow Management in Manufacturing - ADDVALUE - Nilesh Arora
Introduction of Flow Management in Manufacturing - ADDVALUE - Nilesh AroraIntroduction of Flow Management in Manufacturing - ADDVALUE - Nilesh Arora
Introduction of Flow Management in Manufacturing - ADDVALUE - Nilesh AroraADD VALUE CONSULTING Inc
 
Automation in cytology.
Automation in cytology.Automation in cytology.
Automation in cytology.Manan Shah
 
InterSystems test automation
InterSystems test automationInterSystems test automation
InterSystems test automationMassTLC
 
Puppet Camp Melbourne 2014:
Puppet Camp Melbourne 2014: Puppet Camp Melbourne 2014:
Puppet Camp Melbourne 2014: Puppet
 
CS3600 Dentist Presentation
CS3600 Dentist PresentationCS3600 Dentist Presentation
CS3600 Dentist Presentation360 Visualise
 

Similar to IUCell001_Basic Handling_A.ppsx (20)

Automation in Biochemistry (Autoanalyzers)
Automation in Biochemistry (Autoanalyzers)Automation in Biochemistry (Autoanalyzers)
Automation in Biochemistry (Autoanalyzers)
 
Item 4 aspac brisbane_workshop_qaqc_oct19
Item 4 aspac brisbane_workshop_qaqc_oct19Item 4 aspac brisbane_workshop_qaqc_oct19
Item 4 aspac brisbane_workshop_qaqc_oct19
 
Water Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check ProceduresWater Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check Procedures
 
A-PQA-Brochure-v1.2.4
A-PQA-Brochure-v1.2.4A-PQA-Brochure-v1.2.4
A-PQA-Brochure-v1.2.4
 
Real time Vital Signs Monitorong Using NI LabVIEW
Real time Vital Signs Monitorong Using NI LabVIEWReal time Vital Signs Monitorong Using NI LabVIEW
Real time Vital Signs Monitorong Using NI LabVIEW
 
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
 
Final Presentation
Final PresentationFinal Presentation
Final Presentation
 
Chromatography Data System: Control your Instruments
Chromatography Data System: Control your Instruments Chromatography Data System: Control your Instruments
Chromatography Data System: Control your Instruments
 
Beyond shingo pdca
Beyond shingo pdcaBeyond shingo pdca
Beyond shingo pdca
 
Qlcl bao cao
Qlcl bao caoQlcl bao cao
Qlcl bao cao
 
Lean Management and Kanban System
Lean Management and Kanban SystemLean Management and Kanban System
Lean Management and Kanban System
 
4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINAL
4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINAL4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINAL
4.15.14_BEC-218 Iris_Brochure_READERS SPREADS, FINAL
 
Poka yoke error proofing
Poka yoke error proofing Poka yoke error proofing
Poka yoke error proofing
 
Lean Supply Chain Management--by JULIAN KALAC
Lean Supply Chain Management--by JULIAN KALACLean Supply Chain Management--by JULIAN KALAC
Lean Supply Chain Management--by JULIAN KALAC
 
Introduction of Flow Management in Manufacturing - ADDVALUE - Nilesh Arora
Introduction of Flow Management in Manufacturing - ADDVALUE - Nilesh AroraIntroduction of Flow Management in Manufacturing - ADDVALUE - Nilesh Arora
Introduction of Flow Management in Manufacturing - ADDVALUE - Nilesh Arora
 
Automation in cytology.
Automation in cytology.Automation in cytology.
Automation in cytology.
 
INPECO TTP.pdf
INPECO TTP.pdfINPECO TTP.pdf
INPECO TTP.pdf
 
InterSystems test automation
InterSystems test automationInterSystems test automation
InterSystems test automation
 
Puppet Camp Melbourne 2014:
Puppet Camp Melbourne 2014: Puppet Camp Melbourne 2014:
Puppet Camp Melbourne 2014:
 
CS3600 Dentist Presentation
CS3600 Dentist PresentationCS3600 Dentist Presentation
CS3600 Dentist Presentation
 

Recently uploaded

Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatorenarwatsonia7
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls ServiceMiss joya
 
Call Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service ChennaiCall Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service ChennaiNehru place Escorts
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...narwatsonia7
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
 
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Nehru place Escorts
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 

Recently uploaded (20)

Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
 
Call Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service ChennaiCall Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
 
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
 

IUCell001_Basic Handling_A.ppsx

  • 1. agile - affordable - accurate agile - affordable - accurate Fast. Reliable. Close to patients. Analyticon Biotechnologies GmbH – Germany IUCell001_en_30_001_03.01_20220829
  • 2. agile - affordable - accurate agile - affordable - accurate Urilyzer® Cell
  • 3. agile - affordable - accurate Agenda Specifications 1 Setup 2 Software 3 4 Workflow 5 Maintenance
  • 4. agile - affordable - accurate Specifications • Up to 60 samples/hour • 15 images/sample Throughput • Storage of 10.000 measurements with 15 images each Database • Only 175 µl per measurement • Low space requirement Volume • 2x Ethernet, 2x USB 3.0, 2x USB 3.1, 1x RS232 • Specified Serial interfaces for Barcode Reader, • Chemical Analyzer and LIS Connectivity • Easy connection to urine test strip analyzer Urilyzer® 100 Pro / 500 Pro • Whole Urinalysis result report on one device Compatibility • Urilyzer® Cell cuvettes (closed system, disposable) Pipetting tips • No reagents needed Consumables
  • 5. agile - affordable - accurate Delivery contents • Urilyzer® Cell • Monitor • Keyboard • Mouse
  • 6. agile - affordable - accurate Setup Load the cuvettes Connect cabling Start the device
  • 7. agile - affordable - accurate agile - affordable - accurate Software Lets switch to the device
  • 8. agile - affordable - accurate Connection to Chemical Analyzer
  • 9. agile - affordable - accurate Connection to Chemical Analyzer The Urilyzer® Cell can be easily connected to the Urilyzer 100/500 Pro Backside of the Urilyzer® Cell (CHEM) via serial cable to chemical analyzer Urilyzer® 100 Pro: Go into Main  Options  Settings  Output  LIS2  set Baudrate Urilyzer® 500 Pro: Go into Main  Settings  Output  LIS2  set Baudrate Urilyzer® Cell: Go into Settings  Chemical Analyzer  choose „serial“ and the baudrate  Activate The chemical analyser has to be measured before the Urilyzer® Cell Use the same ID on both devices for each sample • Best with barcode reader (2 required)
  • 10. agile - affordable - accurate Workflow – Manual procedure Pressing the Start button to load an empty cuvette to pipetting position Barcode scanning Sample mixing Aspirating urine sample using manual pipette Injecting urine sample into cuvette Pressing the Start button to start the measurem ent
  • 11. agile - affordable - accurate Workflow – Automated procedure Centrifuging process Results are forwarded to Data Manager Software Cuvette is moved to next microscope position Focusing process Cuvette is placed into first microscope position Recorded image is evaluated HPF-like image is taken by built-in camera Used cuvette is placed into the waste bin Last image? NO YES
  • 12. agile - affordable - accurate agile - affordable - accurate Pipetting practice
  • 13. agile - affordable - accurate Pipetting practice
  • 14. agile - affordable - accurate Workflow – Manual procedure • Press the Start-Button on top of the device Pressing the Start button: to load an empty cuvette to pipetting position Barcode scanning Sample mixing Aspirating urine sample using manual pipette Injecting urine sample into cuvette Pressing the Start button: to start the measurement
  • 15. agile - affordable - accurate Workflow – Manual procedure • Press the Start-Button on top of the device • Enter ID - either manual or by a barcode scanner Pressing the Start button: to load an empty cuvette to pipetting position Barcode scanning Sample mixing Aspirating urine sample using manual pipette Injecting urine sample into cuvette Pressing the Start button: to start the measurement
  • 16. agile - affordable - accurate Workflow – Manual procedure • Press the Start-Button on top of the device • Enter ID - either manual or by a barcode scanner • Invert the sample several times – do not shake! Pressing the Start button: to load an empty cuvette to pipetting position Barcode scanning Sample mixing Aspirating urine sample using manual pipette Injecting urine sample into cuvette Pressing the Start button: to start the measurement
  • 17. agile - affordable - accurate Workflow – Manual procedure • Press the Start-Button on top of the device • Enter ID - either manual or by a barcode scanner • Invert the sample several times – do not shake! • Pick a pipette with 175 µl sample volume and aspirate the sample • Outer needle diameter of 1,2 – 2,8 mm Pressing the Start button: to load an empty cuvette to pipetting position Barcode scanning Sample mixing Aspirating urine sample using manual pipette Injecting urine sample into cuvette Pressing the Start button: to start the measurement
  • 18. agile - affordable - accurate Workflow – Manual procedure • Press the Start-Button on top of the device • Enter ID - either manual or by a barcode scanner • Invert the sample several times – do not shake! • Pick a pipette with 175 µl sample volume and aspirate the sample • Outer needle diameter of 1,2 – 2,8 mm • Fill the sample into the cuvette Pressing the Start button: to load an empty cuvette to pipetting position Barcode scanning Sample mixing Aspirating urine sample using manual pipette Injecting urine sample into cuvette Pressing the Start button: to start the measurement
  • 19. agile - affordable - accurate Workflow – Manual procedure • Press the Start-Button on top of the device • Enter ID - either manual or by a barcode scanner • Invert the sample several times – do not shake! • Pick a pipette with 175 µl sample volume and aspirate the sample • Outer needle diameter of 1,2 – 2,8 mm • Fill the sample into the cuvette • Press the Start-Button again to start the measurement process Pressing the Start button: to load an empty cuvette to pipetting position Barcode scanning Sample mixing Aspirating urine sample using manual pipette Injecting urine sample into cuvette Pressing the Start button: to start the measurement
  • 20. agile - affordable - accurate Workflow – Automated procedure • Creating a monolayer of particles by centrifuging the cuvette 10 seconds with 2000 RPM Centrifuging process Results are forwarded to Data Manager Software Cuvette is moved to next microscope position Focusing process Cuvette is placed into first microscope position Recorded image is evaluated HPF-like image is taken by built-in camera Used cuvette is placed into the waste bin Last image? NO YES
  • 21. agile - affordable - accurate Workflow – Automated procedure • Creating a monolayer of particles by centrifuging the cuvette • Cuvette is moved above microscope Centrifuging process Results are forwarded to Data Manager Software Cuvette is moved to next microscope position Focusing process Cuvette is placed into first microscope position Recorded image is evaluated HPF-like image is taken by built-in camera Used cuvette is placed into the waste bin Last image? NO YES
  • 22. agile - affordable - accurate Workflow – Automated procedure • Creating a monolayer of particles by centrifuging the cuvette • Cuvette is moved above microscope • Focus series is captured in different heights Centrifuging process Results are forwarded to Data Manager Software Cuvette is moved to next microscope position Focusing process Cuvette is placed into first microscope position Recorded image is evaluated HPF-like image is taken by built-in camera Used cuvette is placed into the waste bin Last image? NO YES
  • 23. agile - affordable - accurate Workflow – Automated procedure • Creating a monolayer of particles by centrifuging the cuvette • Cuvette is moved above microscope • Focus series is captured in different heights • Camera takes a picture of the sharpest plane Centrifuging process Results are forwarded to Data Manager Software Cuvette is moved to next microscope position Focusing process Cuvette is placed into first microscope position Recorded image is evaluated HPF-like image is taken by built-in camera Used cuvette is placed into the waste bin Last image? NO YES
  • 24. agile - affordable - accurate Workflow – Automated procedure • Creating a monolayer of particles by centrifuging the cuvette • Cuvette is moved above microscope • Focus series is captured in different heights • Camera takes a picture of the sharpest plane • The system checks if this was the last picture • The image gets automatically evaluated and is saved in the database Centrifuging process Results are forwarded to Data Manager Software Cuvette is moved to next microscope position Focusing process Cuvette is placed into first microscope position Recorded image is evaluated HPF-like image is taken by built-in camera Used cuvette is placed into the waste bin Last image? NO YES
  • 25. agile - affordable - accurate Workflow – Automated procedure • The system checks if this was the last picture • YES: The cuvette is moved into the waste bin • NO: the cuvette position moves to next microscope position Centrifuging process Results are forwarded to Data Manager Software Cuvette is moved to next microscope position Focusing process Cuvette is placed into first microscope position Recorded image is evaluated HPF-like image is taken by built-in camera Used cuvette is placed into the waste bin Last image? NO YES
  • 26. agile - affordable - accurate Results
  • 27. agile - affordable - accurate Auto classification • Red blood cells • White blood cells • White blood cell clumps • Crystals • Calcium-oxalate monohydrate • Calcium-oxalate dihydrate • Triple phosphate • Uric acid • Hyaline casts • Pathological casts • Non-squamous epithelial cells • Squamous epithelial cells • Yeast • Bacteria • Bacteria rods • Bacteria cocci • Mucus • Spermatozoa
  • 28. agile - affordable - accurate Auto classification Particle type: Subclasses and other particles can be added Semi-quantitative category: The ranges and category name can be changed Quantitative results: p/µl or p/HPF Negative result limits RBC and WBC are real quantitative results Every other result is semi-quantitative even if the result is displayed quantitative
  • 29. agile - affordable - accurate Calculation HPF-like HPF LPF-like LPF • 𝐻𝑃𝐹 − 𝑙𝑖𝑘𝑒 ∗ 1,5 = 𝐻𝑃𝐹 • 𝐿𝑃𝐹 − 𝑙𝑖𝑘𝑒 ∗ 0,5 = 𝐻𝑃𝐹 Semi-quantitative parameter • 𝑝 µ𝑙 = 𝑝 15 ∗ 6,6 • 𝑝 𝐻𝑃𝐹 = 𝑝 15 ∗ 1,5 RBC and WBC calculation • RBC: 𝑝 𝐻𝑃𝐹 ∗ 2 • WBC: 𝑝 𝐻𝑃𝐹 ∗ 1,5 Urilyzer® Cell p/HPF is correlated with 20x concentrated manual measurement.
  • 30. agile - affordable - accurate Focussing process • 2000 steps ≡ 2,5 mm • 1 step ≡ 1,25 µm • Hard Focus Start ~ 800 • Set with the Reference Cuvette in Service Software 2000 steps Pic. 1 800 }320 steps Taking 1 image every two steps  chooses the best Pic. 2 Pic. 3 } 16 steps up & 16 steps down } 16 steps up & 16 steps down Best image heigth of the previous one defines the start  taking 16 images in total and choses the best again Best image heigth of the previous one defines the start  taking 16 images in total and choses the best again And so on with the remaining 12 images
  • 31. agile - affordable - accurate Manual addable sediments • Up to 50 parameters are manual addable • The red ones are only manual addable • Can be used for further information • Parameters must be activated to add them
  • 32. agile - affordable - accurate Review criteria 0-2 review images  normal measurement: • The results are calculated based on the non-review images  can be auto-validated ≥ 3 review images and ≥ 5 O.K. images: review measurement: • The results are calculated based on the non-review images  cannot be auto-validated < 5 O.K. images: review measurement: • No automated result  cannot be auto-validated • User can modify the final sample results based on the manual review of the images or remeasure the sample with a dilution MUC review: • Concentration limit can be set for MUC: if the sample concentration is above this limit, the measurement will be flagged for review • The results are calculated based on the images  cannot be auto- validated
  • 33. agile - affordable - accurate Maintenance • Daily routine should maintain a cleaning process after every workshift, 24 hours or according to the local regulations • What should be cleaned: • Cuvette feeder arm • Cuvette hold down assembly • Drop tray • Centrifuge cover • Centrifuge arm • The path of the cuvette, especially the top of the microscope • Waste bin • The outer surface, especially the pipetting window
  • 34. agile - affordable - accurate agile - affordable - accurate We look forward to the opportunity to work with you.