Lab Test 3

Loading...

Flash Player 9 (or above) is needed to view presentations.
We have detected that you do not have it on your computer. To install it, go here.

0 comments

Post a comment

    Post a comment
    Embed Video
    Edit your comment Cancel

    1 Favorite

    Lab Test 3 - Presentation Transcript

    1. Density, Contrast, Distortion & Detail Lab Test #3
    2. Contrast and the 15% Rule Objective: To  demonstrate the influence of the 15% rule on contrast 1st exposure at  recommended technique Used as control 
    3. Contrast and the 15% rule Apply the 15% rule  twice and ↑ kVp ↓ mAs accordingly All other factors stay the  same The math doesn’t quite  add up on this one, so I think the images are from 2 different groups
    4. Contrast and the 15% rule Again, this image doesn’t  match my lab sheet, I believe we lowered the kVp 15% and shot another on just for S&G At this point, we should  compare the 2 images to determine which has the longest scale of contrast See next slide 
    5. Contrast and the 15% rule Low kVp=High Contrast= Short scale High kVp=Low Contrast= Long scale More B&W More Shades of Gray
    6. Density- Control Image Objective: To  demonstrate the effect of select factors on density Control image of knee  phantom taken @ 5mAs, 60kVp, & 40” SID.
    7. Density- Filter With no change in tech  factors, a copper sheet is taped to the face of the collimator Filters have an inverse  relationship on density thus: ↑ Filtration ↓ Density 
    8. Density- Grid Copper filter removed  No change in technical  factors Grid is introduced to IR  Result: Adding a grid  decreases density this is an inverse relationship (+) ↑ Grid ↓Density 
    9. Density- SID Grid removed  No change in technical  factors SID increased from 40”  to 60”
    10. Density-Tissue Thickness Grid is removed and hand  phantom is substituted for knee No change in technical  factors Overall, the hand the tissue  thickness of the hand is much less than that of the knee ↑ tissue thickness ↓density 
    11. Density- Collimation-Control Torso phantom is  substituted for hand 14x17 collimation  Use technique  recommended by console + 20 mAs This image will be  compared with the highly collimated image.
    12. Density- Collimation Collimation is  increased to a 4x4 square. Repeat exposure of  torso with no change in technical factors ↑collimation ↓density 
    13. Density- Anode Heel Effect Use a foot phantom  and a 14x17 cassette 2 mAs, 60 kVp, 40” SID  Leave collimation open  lengthwise to 17” Orient toes over the  anode side
    14. Density- Anode Heel Effect All technical factors remain the  same Foot is moved to opposite end of  cassette to place toes on cathode side In practice, the thickest tissue  should be placed at the cathode end of tube In theory, placed thicker tissue at  anode end would effect density A visible change in density would  only be appreciated with film/screen
    15. Size Distortion: Control Objective: To  demonstrate the various types of distortion Control image of hand  shot at 40” SID Technique: Pre-  programmed
    16. Size Distortion: 6” OID 2nd image  6” OID is created with  sponges All other factors remain  the same Compare with control  image to note distortion (magnification)
    17. Size Distortion: 12” OID Image #3  OID is ↑ to 12”  Collimation is exactly the  same No other factors have  been changed Note magnified  appearance of hand Such distortion can mask  pathology
    18. Distortion: OID + SID OID is ↓ to 6”  SID is ↓ to 20”  All remaining tech factors  are unchanged Result: Size distortion  (magnification in this case) can be caused by OID, SID or both
    19. Shape Distortion: Control Control exposure of  knee phantom w/recommended tech factors This image will be used  for comparison with others
    20. Shape Distortion: Angled Tube 2nd image of knee  phantom Tube is angled  All tech factors remain  the same
    21. Shape Distortion: Angled IR 3rd exposure of knee  phantom IR is angled  All tech factors are  unchanged from control Note closed joint  spaces
    22. Shape Distortion: Angled Anatomy 4th exposure of knee  phantom All tech factors remain  constant except… Phantom is angled  Result: Shape distortion  can be caused by the angulation of tube, IR or anatomy being imaged.
    23. Density & mAs Objective is to  demonstrate the effects of overexposure & underexposure on CR images First image is made @  60kVp, 10mAs w/40” SID
    24. Density & mAs 2 mAs is ↑ 50  All other technical factors  remain the same LGM#’s of all images will be  compared with control LGM represents the # of  photons reaching IR to form the latent image LGM is proportional to mAs 
    25. Density & mAs 3 ↑ to 100 mAs  No other technical  factors are changed At this stage, LGM#’s  appear to be ↑ as mAs is ↑ (a direct relationship).
    26. Density & mAs 4 ↑ to 200 mAs  Remaining tech factors  unchanged LGM#’s increase with  each increase in mAs mAs is THE controlling  factor of density
    27. Detail and Distortion: Motion Objective: To demonstrate  the effect of motion on detail Technical factors: The  programmed technique for a foot but ↓ mA Decrease in mA is necessary  to ↑ exposure time (applicable in next image) 1st image of top taken on  10x12 w/no motion
    28. Detail and Distortion: Motion 2nd image- top is  spinning while exposure is taken Technical factors are  unchanged The lower mA ↑  exposure time, allowing the motion to be caught on film
    29. Detail and Distortion: Motion Additional image of this  experiment Compared to the 1st image  (stationary), the lead letter attached to top is blurred. Result: Increased motion  decreases detail Motion↑↓ Detail  Although motion is generally a  detriment to good films, it can be used to the RT’s advantage Ex: Using breathing technique to  blur ribs
    30. Quantum Mottle Definition: a lack of  sufficient incoming data to process an image; AKA quantum noise No idea what this  image has to do with anything, but that’s how it was labeled That would be a cool  name for a band though
    31. Contrast- Control Image Objective: To  demonstrate how selected factors effect contrast 66kVp, 10 mAs, 40”  SID on table top
    32. Contrast 66kv Skull phantom is  replaced w/step wedge Exposure taken with no  changes in technical factors Compare this image  w/second exposure using a higher technique
    33. Contrast: ↑kVp ↓mAs ↑ to 86kVp  Compensate by ↓mAs to  ¼ of its original value All other factors remain  the same
    34. Contrast- 86kv Again, step wedge  takes the place of skull phantom No change in technical  factors Compare shades of  gray in adjacent densities
    35. Comparison of Contrast Low kVp High kVp
    36. Contrast: +Grid 3rd image of skull  Grid is introduced  mAs ↑ 4x to  compensate All remaining factors  unchanged
    37. Contrast: Scatter 4th image of skull  Collimation open wide  to expose IR to scatter All remaining factors  unchanged Compare this series of  images to note differences in adjacent densities
    38. Contrast Collimation Skull is now positioned  laterally This exposure will use  bucky instead of table top 100 kVp @ 20mAs 
    39. Contrast: Collimation Repeat exposure of  lateral skull Collimated to 3”x3”  area All remaining factors  are unchanged
    40. Short vs Long Scale Contrast: kVp Objective: To  demonstrate short & long scale contrast Elbow phantom @  46kVp & 5mAs
    41. Stepwedge 2 Contrast ???? I think this is supposed  to be part of a contrast lab but I couldn’t match it up with anything Also, it was posted  twice in the images file…so there’s nothing to compare it to as far as I can tell
    42. Short vs Long Scale Contrast; kVp 2nd image of elbow  phantom ↑ to 70kVp  ↓to 3mAs 
    43. Density Chart Variables Effects Density Relationship ↑↓ ↓↑ Filtration Inverse ↑↓ ↓↑ Grid Inverse ↑↓ ↓↑ Tissue Thickness Inverse ↑↓ ↓↑ SID Inverse ↑↓ ↓↑ Collimation Inverse Cathode: ↑ Anode Heel Effect *In theory the N/A unless film Anode: ↓ anode heel effect screen technology does cause a is involved change in density but this is not evident in digital imaging
    44. Contrast Chart Variable Effect Contrast Relationship ↑↓ ↓↑ kVp inverse Ø Ø mAs none Ø Ø SID none ↑↓ ↑↓ OID direct ↑↓ ↓↑ Filtration inverse ↑↓ ↑↓ Collimation direct ↑↓ ↓↑ Tissue Thickness inverse ↑ Contrast Media + direct ↑↓ ↑↓ Grid ratio direct Ø Ø Focal spot size none ↓ Always ↓ contrast Film Processing *+/- developing time and temp beyond optimal
    45. Detail Chart Variable Effect Detail Relationship ↑↓ ↑↓ SID direct ↑↓ ↓↑ OID inverse ↑↓ ↓↑ Tissue Thickness inverse ↑↓ ↓↑ Focal Spot Size inverse ↑↓ ↓↑ Motion inverse
    SlideShare Zeitgeist 2009

    + rainesalatterainesalatte Nominate

    custom

    572 views, 1 favs, 2 embeds more stats

    Images and concepts from the labs on density, contr more

    More info about this document

    © All Rights Reserved

    Go to text version

    • Total Views 572
      • 550 on SlideShare
      • 22 from embeds
    • Comments 0
    • Favorites 1
    • Downloads 15
    Most viewed embeds
    • 17 views on http://gammakitty.blogspot.com
    • 5 views on http://www.gammakitty.blogspot.com

    more

    All embeds
    • 17 views on http://gammakitty.blogspot.com
    • 5 views on http://www.gammakitty.blogspot.com

    less

    Flagged as inappropriate Flag as inappropriate
    Flag as inappropriate

    Select your reason for flagging this presentation as inappropriate. If needed, use the feedback form to let us know more details.

    Cancel
    File a copyright complaint
    Having problems? Go to our helpdesk?

    Categories