OXYMAT 61
Gas Analyzers
for the Determination
of Oxygen

2
2
2
2
2
5
7

General
Overview
Benefits
Application
Design
Function
Integration

10
10
11
12
13
16

19“ unit
Technical specifications
Ordering data
Dimensional drawings
Schematics
More information
- Documentation
- Spare parts for a 2-year and
a 5-year service

17

Conditions of sale and delivery
Export regulations
Contact addresses

Siemens PA 10 News · May 2004
Gas Analysis
OXYMAT 61
General
GasGas Analytics

■ Overview

■ Design

The measuring principle of the OXYMAT 61 gas analyzers is
based on the paramagnetic alternating pressure method and is
used to measure oxygen in gases in standard applications.

Display and control panel
• Large LCD panel for simultaneous display of:
- Measured value
- Status line
- Measuring ranges
• Contrast of LCD panel adjustable via menu
• Permanent LED backlighting
• Washable membrane keyboard with five softkeys
• Menu-based operation for configuration, test functions and
calibration
• User help in plain text
• Graphic display of concentration trend; programmable time
intervals.

OXYMAT 61

■ Benefits
• Integrated pump for reference gas circulation
(ambient air)
• High linearity
• Compact design
• Possibility of physical zero suppression.

■ Application
Application areas
• Environmental protection
• Boiler control in firing systems
• Quality monitoring
• Process exhaust monitoring
• Process optimization.
Further applications
• Chemical plants
• Gas manufacturers
• Research and development.

2

• 19“ unit with 4 HU for installation
- in swing frames
- in cabinets, with or without slide rails
• Front panel for service can be hinged down
(laptop connection)
• Gas connections for sample gas input and output: pipe
diameter 6 mm or 1/4"
• Gas and electrical connections at the rear.

Siemens Katalog PA 10 News · May 2004

Inputs and outputs
• Six binary inputs freely configurable (e.g. for range switching,
processing of external signals from sample preparation)
• Six relay outputs freely configurable (failure, maintenance request, maintenance switch, limit alarm, external solenoid
valves)
• Two programmable analog inputs (e.g. cross interference correction, external pressure sensor)
• Extension with eight additional binary inputs and eight additional relay outputs for automatic calibration with up to four
calibration gases.
Communication
• RS 485 included in basic unit
(connection at the rear).
Options
• RS 485/RS 232 converter
• RS 485/Ethernet converter
• Linking to networks via PROFIBUS DP/PA interface
• SIPROM GA software as service and maintenance tool.
Gas Analysis
OXYMAT 61
General

LED backlit graphic
display and
membrane keyboard
with noticeable click

Status line to display
the analyzer status
(programmable)

Two code levels
according to NAMUR
(maintenance and
specialist level)

Easy operation
menu control
using the softkeys

Display of
concentrations as
numbers and bargraph

Display of current
measuring ranges

Display of
start-of-scale and
full-scale values

ESC key
to abort inputs
Keyboard to
enter values
INFO key
for help in plain text

CLEAR key
to delete inputs

ENTER key
to accept
input values

MEAS key
to return to
measurement mode

OXYMAT 61, membrane keyboard and graphic display

Executions of the wetted parts
Standard

Options

Gas path
with hoses

19“ unit
Nipple
Hose
Sample cell
Stub sample cell
Restrictor
O-rings
Hose coupling

SS, type No. 1.4571
Viton
SS, type No. 1.4571
SS, type No. 1.4571
Teflon
Viton
Polyamide 6

Flowmeter

Metering pipe
Float
Float limit
Elbows

Duran glass
Duran glass, black
Teflon
Viton

Sample gas
pressure switch

Membrane
Enclosure

Viton
PA 6.3T

Siemens Katalog PA 10 News · May 2004

3
Gas Analysis
OXYMAT 61
General
Gas path
Keys to the gas path figures
1 Sample gas inlet

8 Pressure switch in sample gas path (option)

2 Sample gas outlet

9 Flowmeter in sample gas path (option)

4 Reference gas inlet

10 Restrictor im in reference gas path (outlet)

5 Restrictor in reference gas path

11 Pressure switch for reference gas monitoring

6 O2 physics

12 Pump

7 Restrictor in sample gas path

13 Filter

F

9

P

7

8
1
Sample gas

6

2

10
4

Reference gas

5
12

P

13

11

Gas path OXYMAT 61 with integrated reference gas pump (connection for 100 hPa)

F

9

P

7

8
1
Sample gas

6

2

4

5
13

Gas path OXYMAT 61 with reference gas connection 3000 to 4000 hPa

4

Siemens Katalog PA 10 News · May 2004

Reference gas
Gas Analysis
OXYMAT 61
General

■ Function

1

In contrast to almost all other gases, oxygen is paramagnetic.
This property is utilized as the measuring principle by the
OXYMAT 61 gas analyzers.
Oxygen molecules in an inhomogeneous magnetic field are
drawn in the direction of increased field strength due to their
paramagnetism. When two gases with different oxygen concentrations meet in a magnetic field, a pressure difference is produced between them.

2

In the case of OXYMAT 61, one gas (1) is a reference gas (N2,
O2 or air), the other is the sample gas (5). The reference gas is
introduced into the sample chamber (6) through two channels
(3). One of these reference gas streams meets the sample gas
within the area of a magnetic field (7). Because the two channels
are connected, the pressure, which is proportional to the oxygen
concentration, causes a cross flow. This flow is converted into an
electric signal by a microflow sensor (4).

2

4
3

3

DP

5
6

The microflow sensor consists of two nickel grids heated to approx. 120 ºC which form a Wheatstone bridge together with two
supplementary resistors. The pulsating flow results in a change
in the resistance of the Ni grids. This results in a bridge offset
which depends on the oxygen concentration in the sample gas.
Because the microflow sensor is located in the reference gas
stream, the measurement is not influenced by the thermal conductivity, the specific heat or the internal friction of the sample
gas. This also provides a high degree of corrosion resistance
because the flow sensor is not exposed to the direct influence of
the sample gas.

O2

7

O2

O2
O2

8

O2

By using a magnetic field with alternating strength (8), the effect
of the background flow in the microflow sensor is not detected,
and the measurement is thus independent of the instrument orientation.
The sample cell is directly in the sample path and has a small
volume. The microflow sensor thus responds quickly, resulting in
a very short response time for the OXYMAT 61.

9
1
2
3
4
5
6
7
8
9

Reference gas inlet
Restrictors
Reference gas channels
Microflow sensor for measurement
Sample gas inlet
Sample cell
Paramagnetic effect
Electromagnet with alternating field strength
Sample gas and reference gas outlet

OXYMAT 61, mode of operation

Siemens Katalog PA 10 News · May 2004

5
Gas Analysis
OXYMAT 61
General
Special characteristics
• Four freely-parameterizable measuring ranges, also with zero
offset, all measuring ranges linear
• Electrically isolated signal output selectable as 0/2/4 to 20 mA
(also inverted)
• Autoranging or manual range switching possible; remote
switching is also possible
• Storage of measured values possible during calibration
• Time constants selectable within wide limits (static/dynamic
noise suppression); i.e. the response time of the analyzer can
be matched to the respective application
• Simple handling using menu-based operation
• Low long-term drift
• Two operation levels with separate access code to prevent
unintentional and unauthorized inputs
• Automatic range calibration can be parameterized
• Operation based on NAMUR Recommendation
• Monitoring of sample gas (option)

• Customer-specific analyzer options such as e.g.:
- Customer acceptance
- TAG labels
- Drift recording
• Simple handling using a digital membrane keyboard and
menu-based operation
• Short response time
• Reference gas supply external (N2, O2 or air, approx.
3000 hPa) or via an integrated reference gas pump (ambient
air, approx. 100 hPa)
• Monitoring of reference gas with reference gas connection
3000 to 4000 hPa
• Different smallest spans, depending on version 2.0% or
5.0% O2
• Internal pressure sensor to correct sample gas variations.

Correction of zero error / Cross interferences
Residual gas
(concentration 100 % v/v)

Zero deviation
in % v/v O2 absolute

Organic gases

Residual gas
(concentration 100 % v/v)

Zero deviation
in % v/v O2 absolute

Inert gases
CH3COOH

-0.64

Argon

Ar

-0.25

Acetylene

C2H2

-0.29

Helium

He

+0.33

1,2 butadiene

C4H6

-0.65

Krypton

Kr

-0.55

1,3 butadiene

C4H6

-0.49

Neon

Ne

+0.17

iso-butane

C4H10

-1.30

Xenon

Xe

-1.05

n-butane

C4H10

-1.26

1-butene

C4H6

-0.96

iso-butene

C4H8

-1.06

Ammonia

NH3

-0.20

C6H12

-1.84

Carbon dioxide

CO2

-0.30

Dichlorodifluoromethane (R12) CCl2F2

-1.32

Carbon monoxide

CO

+0.07

Ethane

C2H6

-0.49

Chlorine

Cl2

-0.94

Ethylene

C2H4

-0.22

Dinitrogen monoxide

n-heptane

C7H16

-2.4

Hydrogen

n-hexane

C6H14

-2.02

Methane

CH4

-0.18

Methanol

CH3OH

-0.31

Hydrogen fluoride

Acetic acid

Cyclo-hexane

Anorganic gases

N2 O

-0.23

H2

+0.26

Hydrogen bromide

HBr

-0.76

Hydrogen chloride

HCl

-0.35

HF

-0.10

HI

-1.19

n-octane

C8H18

-2.78

Hydrogen iodide

n-pentane

C5H12

-1.68

Hydrogen sulphide

iso-pentane

C5H12

-1.49

Oxygen

Propane

C3H8

-0.87

Nitrogen

Propylene

C3H6

-0.64

Nitrogen dioxide

Trichlorofluoromethane (R11)

H2S

-0.44

O2

+100

N2

0.00

NO2

+20.00

CCl3F

-1.63

Nitrogen oxide

NO

+42.94

Vinyl chloride

C2H3Cl

-0.77

Sulphur dioxide

SO2

-0.20

Vinyl fluoride

C2H3F

-0.55

Sulphur hexafluoride

SF6

-1.05

C2H2Cl2

-1.22

Water

H2O

-0.03

1,1 vinylidene chloride

Zero error due to diamagnetism or paramagnetism of residual gases with nitrogen as the reference gas at 60 °C and 1000 hPa absolute
(according to IEC 1207/3)
Conversion to other temperatures:

The zero errors mentionned in the table must be multiplied with a
correction factor (k):
• with diamagnetic gases: k = 333 K / (θ [°C] + 273 K)
• with paramagnetic gases: k = [333 K / (θ [°C] + 273 K)]2
(all diamagnetic gases have a negative zero error).

6

Siemens Katalog PA 10 News · May 2004
Gas Analysis
OXYMAT 61
General
Interface parameters

■ Integration

Level

Communication
Communications facilities

8

Stop bit

1

Start bit

1

Parity

None

No echo mode

The standard integrated series interface permits to communicate between several gas analyzers via the internal bus (ELAN).
Up to 12 gas analyzers with max. four components can be networked.
The operating principle is shown in the following figure.

2

RS485
Converter
RS232 RS485

RS485
Converter
RS232 RS485

8
5
6
7

5

Typical design of an RS 485 network
Item

Designation

1
2

Computer
RS 485/RS 232 converter
with RS 485/RS 232 cable
RS 485 bus connector with jumper
Analyzers
RS 485 cable
RS 485 bus connector
RS 485 network
9-pin SUB-D plug
Option: RS 485 repeater

A5E00054148
C79451-Z1589-U1

RS 485/Ethernet converter

C79451-A3364-D61

Further accessories (e.g. cable,
connectors, repeater, etc.)

see Catalog IK PI

RS 485 interface (SIPROM GA)

Up to 12 gas analyzers with max. four components can be networked.

3
4

RS232

Order No.

Interface description (d)

SIPROM GA is a software tool for service and maintenance. All
analyzer functions (except factory functions) can be remotecontrolled and monitored via RS 485/RS 232 converters.

1
RS232

Ordering data
RS 485/RS 232 converter

RS 485 interface (ELAN)

3
4
5
6
7
8
9

9600

Data bits

The gas analyzers of series 6 (ULTRAMAT 6,
ULTRAMAT/OXYMAT 6, OXYMAT 6, OXYMAT 61, FIDAMAT 6
and CALOMAT 6), as well as the ULTRAMAT 23 offer the following communications facilities:
• RS 485 interface
• PROFIBUS DP/PA
• AK interface (only OXYMAT 6, ULTRAMAT 6 and
ULTRAMAT/OXYMAT 6).

9

RS 485

Baud rate

Networking of several gateways is possible when using the
RS 485/Ethernet converter (gateway). The number of operatable
analyzers is increased correspondingly.
Functions
• Display and saving of all analyzer data
• Remote operation of all analyzer functions
• Parameter and configuration settings
• Comprehensive diagnostics information
• Remote calibration
• Online help
• Cyclic saving of measured values and status on hard disk
• Exporting to commercially available application programs
• Downloading of new software.
Hardware requirements:
• PC/Laptop Pentium 133 MHz, RAM 32 MB, CD-ROM drive
• Free disk capacity min. 10 MB
• Free COM-Port: RS 232 or RS 485:
- for coupling to ELAN network RS 485/RS 232
- Linking the RS 485/Ethernet converter requires a 10 MB
standard network (RJ 45 connection) with TCP/IP. The distance should not exceed 500 m with an RS 485 network, a
repeater should be inserted for a longer distance.
Software requirements:
• Windows 95
• Windows 98
• Windows NT 4.0
• Windows 2000
• Windows XP.

Siemens Katalog PA 10 News · May 2004

7
Gas Analysis
OXYMAT 61
General
Ordering data

Order No.

SIPROM GA software
German/English selectable
during installation, comprising
1 CD, with installation
instructions, software product
certificate and registration form

S79610-B4014-A1

Firmware retrofitting sets for
older analyzers:
ULTRAMAT 23
(prior to SW version 4.1)
all langages

C79451-A3494-S501

ULTRAMAT 6
(prior to SW version 4.1)
• German

C79451-A3478-S501

• English

C79451-A3478-S502

• French

C79451-A3478-S503

• Spanish

C79451-A3478-S504

• Italian

C79451-A3478-S505

OXYMAT 6
(prior to SW version 4.1)
• German

C79451-A3480-S501

• English

C79451-A3480-S502

• French

C79451-A3480-S503

• Spanish

C79451-A3480-S504

• Italian

C79451-A3480-S505

Basic structure of a PROFIBUS system

8

Siemens Katalog PA 10 News · May 2004

PROFIBUS DP/PA
PROFIBUS DP/PA is the leading field bus on the market. All
Siemens gas analyzers are suitable for PROFIBUS when
equipped with an optional plug-in card (retrofitting also possible)
and satisfy the binding "Device profile for analyzers" defined by
the PNO (PROFIBUS user organization). Central access to the
analyzers in the system is possible using the SIMATIC PDM operator input software.
The term field bus describes a digital communications system
with which distributed field devices in a plant are networked together via one single cable, and connected at the same time to
programmable controllers or to a process control system.
PROFIBUS is the leading field bus on the market. The
PROFIBUS DP version is widely used for production automation
because of its high transmission ratefor relatively small data
quantities per device, whereas PROFIBUS PA particularly takes
into account the features required for process engineering, e.g.
large data quantities and application in potentially explosive atmospheres.
User benefits can be found in the extremely high potentials for
cost savings in all areas of the plant, covering configuring and
commissioning, operation and maintenance, and up to later
plant extensions.
Operation of the gas analyzers from a control system or separate
PC is possible using the SIMATIC PDM (Process Device Manager) operator input tool which is software executing under Windows and which can also be incorporated into the
SIMATIC PCS 7 process control system. This permits clear display of both the incorporation of devices into the system and the
complex parameter structure of the analyzers, permitting operation to be carried out simply by clicking.
Gas Analysis
OXYMAT 61
General
The PROFIBUS user organization (PNO) is an independent international institution, and represents the interests of many vendors
and users. In addition to services such as consultation, training
and device certification, its prime task is the further development, standardization and promotion of the PROFIBUS technology. The definition of a binding functionality for a device class in
a profile is a prerequisite for the uniform response of devices
from different vendors, the so-called interoperability.The profile
for analyzers was defined as binding at the end of 1999, thus
guaranteeing the interaction of all PROFIBUS-based devices in
a plant.

A differentiation is made between cyclic and acyclic services.
Cyclic services are used to transmit time-critical data such as
measured values and statuses. The acyclic services permit the
scanning or modification of device parameters during operation.

This profile defines the functionality of the analyzers in a block
model: e.g. the physical block describes the measuring procedure, analyzer and vendor names, serial number and operating
state (operation, maintenance). Various functional blocks contain the execution of specific functions such as the processing
of measured values or alarms. The transducer blocks describe
the functionality of the actual measuring procedure and its control, e.g. preprocessing of a measured value, correction of crossinterferences, characteristics, measuring ranges as well as
switching and control procedures. Protocols define the data
transmission between the stations on the bus.

The user benefits of numerous functions especially in the automotive industry, e.g. to operate a re-linearization

All gas analyzers of Series 6, (ULTRAMAT 6,
ULTRAMAT/OXYMAT 6, OXYMAT 6, OXYMAT 61, FIDAMAT 6
and CALOMAT 6), as well as the ULTRAMAT 23 are suitable for
PROFIBUS when fitted with the optional plug-in card.
AK interface (only OXYMAT 6, ULTRAMAT 6 and
ULTRAMAT/OXYMAT 6)

Unlike to PROFIBUS and ELAN, communication between only
one unit and one PC is possible and operates according to the
master-slave principle. The unit only transmits data on request
with a command message, but always only command can be
processed and answered.
Funktion 88 permits to call the menu and to set the parameters.

Siemens Katalog PA 10 News · May 2004

9
Gas Analysis
OXYMAT 61
19" unit

■ Technical specifications
General
Measuring ranges

Smallest possible measuring
span (referred to 1000 hPa
absolute sample gas pressure,
0.5 l/min sample gas flow and
25 °C ambient temperature)
Largest possible measuring
span
Measuring ranges with
zero offset
Position of use
Conformity
Design, enclosure
Degree of protection
Weight
Electrical characteristics
Power supply

Power consumption
EMC interference immunity
(ElectroMagnetic
Compatibility)
Electrical safety
Fuses
Gas inlet conditions
Perm. sample gas pressure
• External reference gas supply
• With integrated pump
Sample gas flow
Sample gas temperature
Sample gas humidity
Time response
Warm-up period

Response time (T90)
Damping
(electric time constant)
Dead time (purging time of gas
path in analyzer at 1 l/min)
Time for internal signal
processing
Pressure correction range
Pressure sensor, internal

4, switchable internally and externally;
autoranging is also possible
2% v/v or 5% v/v O2

100% v/v O2
Any zero point is possible between
0 to 100% v/v as long as a suitable
reference gas is used
Front panel vertical
CE identification EN 50081-1,
EN 50082-2
IP20 according to EN 60529
Approx. 13 kg
100 to 120 V AC (rated range
90 V to 132 V), 48 to 63 Hz or
200 to 240 V AC (rated range
180 V to 264 V), 48 to 63 Hz
Approx. 37 VA
According to standard requirements
of NAMUR NE21 (08/98)
According to EN 61010-1,
overvoltage category III
100...120 V: 1.0T/250
200...240 V: 0.63T/250

Measuring response (referred to 1000 hPa absolute sample gas
pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Output signal fluctuation
< 0.75% of smallest possible measuring range specified on rating
plate with an electronic time
constant of 1 s (corresponds to
± 0.25% with 2 σ)
Zero drift
< 0.5%/month of smallest possible
meas. span specified on rating plate
Measured-value drift
< 0.5%/month of respective
measuring span
Repeatability
< 1% of respective measuring span
Linearity error
< 1% of respective measuring span
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
Ambient temperature
< 0.1% [% O2] /1 K referred to the
smallest possible measuring span
according to rating plate
Sample gas pressure with air
Without pressure compensation:
(approx. 100 hPa) as refer< 2% of measuring span / 1%
ence gas supply, a correction
change in pressure
of the atmospheric pressure
With pressure compensation:
fluctuations is only possible
< 0,2% dof measuring span /
when the sample gas is vented 1% hange in pressure
to ambient air)
Residual gases
Deviation in zero point corresponding to paramagnetic or diamagnetic
deviation of residual gas (see Table)
Sample gas flow
< 1% of smallest possible measuring
span according to rating plate with a
change in flow of 0.1 l/min within the
permissible flow range
Power supply
< 0.1% of output signal span with
rated voltage ± 10%
Electric inputs and outputs
Analog output
Relay outputs

800 to 1200 hPa
atmospheric pressure ± 50 hPa
18 to 60 l/h (0.3 to 1 l/min)
0 to 50 °C
< 90% relative humidity

Analog inputs

With ambient temperature < 30 min
(maximum accuracy achieved after
2 hours)
3.5 s
0 to 100 s, programmable
Approx. 0.5 to 2.5 s,
depending on version
<1s

Binary inputs
Serial interface
Options

Ambient conditions
Perm. ambient temperature

500 to 2000 hPa
(permitted sample gas pressure see
gas inlet conditions)

Permissible humidity

0/2/4 to 20 mA, floating;
max. load 750 Ω
6, with changeover contacts, freely
selectable, e.g. for range identification; loading capacity:
24 V AC/DC/ 1 A, floating
2, designed for 0/2/4 to 20 mA, for
external pressure sensor and correction of influence of residual gas (correction of cross interference)
6, designed for 24 V, floating, freely
selectable, e.g. for range switching
RS 485
Autocal function with 8 additional
binary inputs and 8 relay outputs;
also with PROFIBUS PA or
PROFIBUS DP
-30 to +70 °C during storage and
transport,
+5 to +45 ºC during operation
< 90% relative humidity as annual
average, during storage and transport (dew point must not be fallen
below)

Reference gases
Measuring range

N2

. . . to 100% v/v O2
(suppressed zero with full-scale value
100% v/v O2)

O2

Around 21% v/v O2
(suppressed zero with
21% v/v O2 within the span)

10

Recommended reference gas

0 to . . . % v/v O2

Air

Siemens Katalog PA 10 News · May 2004

Reference gas pressure
3000 to 4000 hPa
absolute, reference gas
from cylinder
atmospheric pressure with
internal reference gas pump

Remarks

The reference gas flow
is set automatically
to 5 to 10 ml/min
Gas Analysis
OXYMAT 61
19" unit
Ordering data

Order No..

OXYMAT 61 gas analyzer
19“ unit for installation in cabinets

7 MB 2 0 0 1 - 77 A 0 0 - 7777

Gas connections for sample gas and reference gas
Piping with outer diameter 6 mm
Piping with outer diameter ¼"

0
1

Smallest possible span O2
2% Reference gas pressure 3000 hPa
2% Reference gas supply with internal pump
5% Reference gas pressure 3000 hPa
5% Reference gas supply with internal pump

C
D
E
F

Power supply
100 V to 120 V AC, 48 to 63 Hz

0

200 V to 240 V AC, 48 to 63 Hz

1

Sample gas monitoring
Without

A

With (incl. flowmeter and pressure switch)

D

Supplementary electronics
A

Without
Autocal function
• With additional 8 binary inputs/outputs

B

• With serial interface for the automotive industry (AK)

D

• With additional 8 binary inputs/outputs and PROFIBUS PA interface

E

• With additional 8 binary inputs/outputs and PROFIBUS DP interface

F

Language
German

0

English

1

French

2

Spanish

3

Italian

4

Order code

Further versions
Please add „-Z“ to Order No. and specify Order code
RS 485/RS 232 converter

A11

Slide rails (2 rails)

A31

Set of Torx tools, socket spanner

A32

TAG labels (customer-defined inscriptions)

B03

Damping element for sample gas
Customer acceptance (in factory before delivery)

B04
1)

Clean for O2-Service (special-cleaned gas path)
Drift

recording 2)

Y01
Y02
Y03

Measuring range in plain text, if different from standard setting 3)

Y11

Retrofitting sets

Order No.

RS 485/Ethernet converter

C79451-A3364-D61

RS 485/RS 232 converter

C79451-Z1589-U1

Autocal function with 8 binary inputs/outputs

C79451-A3480-D511

Autocal function with 8 binary inputs/outputs and PROFIBUS-PA

A5E00057307

Autocal function with 8 binary inputs/outputs and PROFIBUS-DP

A5E00057312

1)

Customer acceptance: ½ day at factory in presence of customer.
The following work is carried out: comparison of analyzer with ordering data, linearization check (zero, mid-point value and full-scale value),
reproducibility check with calibration gas (recording in each case on XT recorder, logging of results).
2) Drift recording: an XT recording is supplied when the analyzer is delivered;
zero drift with 16 hours continuous operation and sensitivity drift (largest measuring range) with 6 hours continuous operation.
3) Standard setting: Measuring range 1: 0 to smallest possible span
Measuring range 2: 0 to 10%
Measuring range 3: 0 to 25%
Measuring range 4: 0 to 100%.
Siemens Katalog PA 10 News · May 2004

11
Gas Analysis
OXYMAT 61
19" unit

■ Dimensional drawings
384.5
351.5
27.5
15.6
4

82

128

149.5

172

37.7

101.6

177

173

M4

86.5

OXYMAT 61

465
483

4

OXYMAT 61, 19“ unit, dimensions in mm

12

Siemens Katalog PA 10 News · May 2004

2

1

411.5

320
345

132

104

36

0

440
Gas Analysis
OXYMAT 61
19" unit

■ Schematics
Pin assignment (electrical connections)

Connector SUB-D 9F (RS 485)
M

GND

9
8

+5 V

7
6

M

5
4
3
2
1

GND
R_Level-NNC
RD/TD-N
RD/TD-P
R_Level-PNC
NC
GND

Possibility for connection
of bus terminating resistors
to pins 7 and 9.

Connector SUB-D 15F
M

15
14
13
12

M

11

M

10
9

M

8
7
6
5
4
3
2
1

GND
NC
NC
Analog output 1-P
Analog output 1-N
NC
NC
Analog input 2-P
Analog input 2-N
Analog input 1-P
Analog input 1-N
Binary input 6-P
Binary input 5-P
Binary input 5 to 6-N
GND

Floating analog outputs (also with
respect to one another), RL £ 750 W
Pressure correction
Pressure correction
Interfering gas correction
Interfering gas correction

Non-floating
analog inputs,
0 to 20 mA/500 W
or 0 to 10 V
(low-resistance)

Connector SUB-D 25F
M

25
24
23
22
21
20
19
18
17
16
15

M

14

13
12
11
10
9
8
7

GND
Binary input 4-P
Binary input 3-P
Binary input 2-P
Binary input 1-P
Binary input 1 to 4-N
Relay 6
Relay 5
Relay 4

6
5
4
3

Floating via opto isolator
"0" = 0 V (0 to 4.5 V)
"1" = 24 V (13 to 33 V)

Relay 3

Contact loading
max. 24 V/1 A, AC/DC;
relay contacts shown:
de-energized relay coil

Relay 2

2

Relay 1

1

GND

Note:
Cable and plug must
be shielded and connected
to chassis potential.

OXYMAT 61, 19“ unit, pin assignment

Siemens Katalog PA 10 News · May 2004

13
Gas Analysis
OXYMAT 61
19" unit
Pin assignment (electrical connections)

Connector SUB-D 37F (option)

M

37
36
35
34
33
32
31
30
29
28
27
26
25
24
23
22
21
20

M

19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3

GND
NC
NC
Binary input 14-P
Binary input 13-P
Binary input 12-P
Binary input 11-P
Binary input 10-P
Binary input 9-P
Binary input 8-P
Binary input 7-P
Binary input 7 to 14-N

floating via
opto coupler
"0" = 0 V (0 to 4.5 V)
"1" = 24 V (13 to 33 V)

Relay 14
Relay 13
Relay 12
Relay 11

Contact loading
max. 24 V/1 A, AC/DC;
relay contacts shown:
de-energized relay coil

Relay 10
Relay 9
Relay 8

2

Relay 7

1

GND
Note:
Cable and plug must
be shielded and connected
to chassis potential.

Connector SUB-D 9F -X90
PROFIBUS DP

9
8
7
6

5
4
3
2
1

optional

Connector SUB-D 9M -X90
PROFIBUS PA

DGND
CNTR-N
CNTR-P/direction control
RxD/TxD-N (A)
RxD/TxD-P (B)
VP /+ 5 V

OXYMAT 61, 19“ unit, pin assignment of the AUTOCAL board and PROFIBUS connectors

14

Siemens Katalog PA 10 News · May 2004

9
8
7
6

5
4
3
2
1

Pa-N(-)
PA-P(+)
Gas Analysis
OXYMAT 61
19" unit
Gas and electrical connections
Sample gas outlet

Sample gas inlet

Reference gas inlet

15-pin connector:
Binary inputs and
analog inputs/outputs

9-pin
connector:
RS 485

9-pin interface
connector
(option): e.g.
PROFIBUS

Power supply
and fuses

Gas outlet

Gas connections: stubs 6 mm or ¼"

25-pin connector:
Binary inputs and
relay outputs

37-pin connector:
Binary inputs and
relay outputs (option)

OXYMAT 61, 19“ unit, gas and electrical connections

Siemens Katalog PA 10 News · May 2004

15
Gas Analysis
OXYMAT 61
19" unit

■ More information
Documentation
Manual

Order No.

OXYMAT 61
Gasanalysengerät für IR-absorbierende Gase und Sauerstoff
(German)

A5E00123066

OXYMAT 61
Gas Analyzers for IR-absorbing
Gases and Oxygen
(English)

A5E00123067

OXYMAT 61
Analyseurs de gaz pour la
mesure de composants infrarouges et doxygène
(French)

A5E00123068

OXYMAT 61
Analizadores para gases absorbentes de infrarrojo y oxígeno
(Spanish)

A5E00123069

OXYMAT 61
Analizzatori per i gas assorbenti
raggi infrarossi ed ossigeno
(Italian)

A5E00123070

Proposition of spare parts for a 2-year and 5-year service
Qty for
2 years

Qty for
5 years

Order No.

Reference gas supply (pump, restrictor, pressure switch, hose)

1

1

A5E00114838

O-ring

1

2

C74121-Z100-A6

Description
Analyzer section

Pressure switch (sample gas)

1

2

C79302-Z1210-A2

Flowmeter (only version with pump)

1

2

C79402-Z560-T1

Measuring cell
• SS, type No. 1.4571, without flow-type compensation branch

-

1

C79451-A3277-B535

• O-ring (measuring head)

2

4

C79121-Z100-A32

• O-ring (stub)

2

4

C71121-Z100-A159

Measuring head (non flow-type compensation branch)

1

1

C79451-A3460-B525

Restrictor for sample gas path, hose

2

2

C79451-A3480-C10

Reference gas path, 3000 hPa (set of parts)

1

1

C79451-A3480-D518

Electronics
Front panel with keyboard

1

1

A5E00259978

Baseplate (without firmware)

-

1

C79451-A3474-B601

Adapter board LCD/keyboard

1

1

C79451-A3474-B605

Magnet terminal board

-

1

C79451-A3474-B606

LC display

1

1

W75025-B5001-B1

Connector filter

-

1

W75041-E5602-K2

Fuse
• 0,63 A / 250 V (230-V version)

2

4

W79054-L1010-T630

• 1,0 A / 250 V (110-V version)

2

4

W79054-L1011-T100

16

Siemens Katalog PA 10 News · May 2004
Gas Analysis
Conditions of sale and delivery
Export regulations, contact addresses
Comprehensive Terms and Conditions of Sale and Delivery are
available free of charge from your local Siemens business office
under the following Order Nos.:
•6ZB5310-0KR30-0BA0
(for customers based in the Federal Republic of Germany)
•6ZB5310-0KS53-0BA0
(for customers based outside of theFederal Republic of
Germany)
or download them from the Internet:
www.siemens.com/automation/mall
(Germany: A&D Mall Online-Help System)

■ Terms and Conditions of Sale and Delivery
in the Federal Republic of Germany
By using this catalog you can acquire hardware and software
products described therein from the Siemens AG subject to the
following terms. Please note! The scope, the quality and the conditions for supplies and services, including software products,
by any Siemens entity having a registered office outside the
Federal Republic of Germany, shall be subject exclusively to the
General Terms and Conditions of the respective Siemens entity.
for customers with a seat or registered office in the Federal
Republic of Germany
The General Terms of Payment as well as the General Conditions
for the Supply of Products and Services of the Electrical and
Electronics Industry shall apply.
For software products, the General License Conditions for Software Products for Automation and Drives for Customers with
Seat or registered Office in Germany shall apply.

■ Export regulations

for customers with a seat or registered office outside of
Germany
The General Terms of Payment as well as the General Conditions
for Supplies of Siemens, Automation and Drives for Customers
with a Seat or registered Office outside of Germany shall apply.
For software products, the General License Conditions for Software Products for Automation and Drives for Customers with
Seat or registered Office outside of Germany shall apply.
General
The prices are in € (Euro) ex works, exclusive packaging.
The sales tax (value added tax) is not included in the prices.
It shall be debited separately at the respective rate according to
the applicable legal regulations.
In addition to the prices of products which include silver and/or
copper, surcharges may be calculated if the respective limits of
the notes are exceeded.
Prices are subject to change without prior notice. We will debit
the prices valid at the time of delivery.
The dimensions are in mm. Illustrations are not binding.
Insofar as there are no remarks on the corresponding pages,
- especially with regard to data, dimensions and weights given these are subject to change without prior notice.

The products listed in this catalog / price list may be subject to
European / German and/or US export regulations.
Therefore, any export requiring a license is subject to approval
by the competent authorities.
According to current provisions, the following export regulations
must be observed with respect to the products featured in this
catalog / price list:
AL

Number of the German Export List.
Products marked other than “N“ require an export
license.
In the case of software products, the export designations of the relevant data medium must also
be generally adhered to.
Goods labeled with an “AL not equal to N“ are
subject to a European or German export authorization when being exported out of the EU.

ECCN

Export Control Classification Number.
Products marked other than “N“ are subject to a
reexport license to specific countries.
In the case of software products, the export
designations of the relevant data medium must
also be generally adhered to.
Goods labeled with an “ECCN not equal to N“ are
subject to a US re-export authorization.

Even without a label or with an “AL: N“ or “ECCN: N“, authorization may be required due to the final destination and purpose for
which the goods are to be used.
The deciding factors are the AL or ECCN export authorization
indicated on order confirmations, delivery notes and invoices.
Subject to change without prior notice.

7
If you have any questions, please contact your local sales representative or any of the contact addresses below.
Siemens AG
A&D PI 2M Process Analytics
Oestliche Rheinbrueckenstr. 50
D-76187 Karlsruhe
Germany

Siemens Applied Automation
500 West Highway 60
Bartlesville, OK 74003
USA

Siemens Pte. Limited
A&D PI2 Regional Head Quarter
The Siemens Center
60 MacPherson Road
Singapore 348615

Tel.: +49 721 595 4234
Fax: +49 721 595 6375
E-Mail:processanalytics@siemens.com

Tel.:
+1 918 662 7000
Fax:
+1 918 662 7052
E-Mail: saaisales@sea.siemens.com

Tel.:
+65 6490 8702
Fax:
+65 6490 8703
E-Mail: splanalytics.sgp@siemens.com

www.processanalytics.com

www.sea.siemens.com/ia

www.siemens.com/processanalytics

Siemens AG
Automation and Drives
Process Instrumentation and Analytics
D-76181 Karlsruhe
Federal Republic of Germany

Order No.: E86060-K3510-B171-A3-7600
Printed in the Federal Republic of Germany
KG K 0604 3.0 ROT 16 En / 415110
Siemens Katalog PA 10 News · May 2004

17

More Related Content

What's hot

Monicon 4 channel datasheet
Monicon 4 channel datasheetMonicon 4 channel datasheet
Monicon 4 channel datasheet
a1-cbiss
 

What's hot (20)

DMA Direct Mercury Analyzer
DMA  Direct Mercury AnalyzerDMA  Direct Mercury Analyzer
DMA Direct Mercury Analyzer
 
Flow, Pressure & Vapor Instrumentation
Flow, Pressure & Vapor InstrumentationFlow, Pressure & Vapor Instrumentation
Flow, Pressure & Vapor Instrumentation
 
New Thermal Mass Flow Meter From Fox Instruments
New Thermal Mass Flow Meter From Fox InstrumentsNew Thermal Mass Flow Meter From Fox Instruments
New Thermal Mass Flow Meter From Fox Instruments
 
Yokogawa CX 2000 DAQ Station
Yokogawa CX 2000 DAQ StationYokogawa CX 2000 DAQ Station
Yokogawa CX 2000 DAQ Station
 
In Situ Oxygen Transmitter For Combustion Monitoring
In Situ Oxygen Transmitter For Combustion MonitoringIn Situ Oxygen Transmitter For Combustion Monitoring
In Situ Oxygen Transmitter For Combustion Monitoring
 
Process Gas Chromatograph - Yokogawa GC8000
Process Gas Chromatograph - Yokogawa GC8000Process Gas Chromatograph - Yokogawa GC8000
Process Gas Chromatograph - Yokogawa GC8000
 
Solutions Manual for Process Control Modeling Design And Simulation 1st Editi...
Solutions Manual for Process Control Modeling Design And Simulation 1st Editi...Solutions Manual for Process Control Modeling Design And Simulation 1st Editi...
Solutions Manual for Process Control Modeling Design And Simulation 1st Editi...
 
Codes
CodesCodes
Codes
 
OBD2 Powertrain Code List (P0100- P0199)
OBD2 Powertrain Code List (P0100- P0199)OBD2 Powertrain Code List (P0100- P0199)
OBD2 Powertrain Code List (P0100- P0199)
 
OBD2 Powertrain Code List (P0400- P0499)
OBD2 Powertrain Code List (P0400- P0499)OBD2 Powertrain Code List (P0400- P0499)
OBD2 Powertrain Code List (P0400- P0499)
 
OBD2 Powertrain Code List (P0A00- P0A99)
OBD2 Powertrain Code List (P0A00- P0A99)OBD2 Powertrain Code List (P0A00- P0A99)
OBD2 Powertrain Code List (P0A00- P0A99)
 
High integrity pressure protection system ( HIPPS)
High integrity  pressure protection  system  ( HIPPS)High integrity  pressure protection  system  ( HIPPS)
High integrity pressure protection system ( HIPPS)
 
Automatic process controls in a Thermal Power Station
Automatic process controls in a Thermal Power StationAutomatic process controls in a Thermal Power Station
Automatic process controls in a Thermal Power Station
 
Centrifugal Compressor System Design & Simulation
Centrifugal Compressor System Design & SimulationCentrifugal Compressor System Design & Simulation
Centrifugal Compressor System Design & Simulation
 
Monicon 4 channel datasheet
Monicon 4 channel datasheetMonicon 4 channel datasheet
Monicon 4 channel datasheet
 
OBD2 Powertrain Code List (P2700- P2799)
OBD2 Powertrain Code List (P2700- P2799)OBD2 Powertrain Code List (P2700- P2799)
OBD2 Powertrain Code List (P2700- P2799)
 
Yokogawa Vibration Type Density Meter
Yokogawa Vibration Type Density MeterYokogawa Vibration Type Density Meter
Yokogawa Vibration Type Density Meter
 
OBD2 Powertrain Code List (P2400- P2499)
OBD2 Powertrain Code List (P2400- P2499)OBD2 Powertrain Code List (P2400- P2499)
OBD2 Powertrain Code List (P2400- P2499)
 
875 kf gas analyzer
875 kf gas analyzer875 kf gas analyzer
875 kf gas analyzer
 
Instrumentation and process control fundamentals
Instrumentation and  process control fundamentalsInstrumentation and  process control fundamentals
Instrumentation and process control fundamentals
 

Similar to Ultramat analyzer

Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0
a1-cbiss
 
Egas datasheet(1)
Egas datasheet(1)Egas datasheet(1)
Egas datasheet(1)
a1-cbiss
 
Egas datasheet
Egas datasheetEgas datasheet
Egas datasheet
a1-cbiss
 
Gerador~Extrator de Nitrogênio CKD CORP. - Série: NS・NSU・PNA・NS-QFS
Gerador~Extrator de Nitrogênio CKD CORP. -  Série: NS・NSU・PNA・NS-QFSGerador~Extrator de Nitrogênio CKD CORP. -  Série: NS・NSU・PNA・NS-QFS
Gerador~Extrator de Nitrogênio CKD CORP. - Série: NS・NSU・PNA・NS-QFS
Júlio Gengo
 

Similar to Ultramat analyzer (20)

Easy line EL3000 series ABB CEMS Datasheet-continuous gas analyzers
Easy line EL3000 series ABB CEMS Datasheet-continuous gas analyzersEasy line EL3000 series ABB CEMS Datasheet-continuous gas analyzers
Easy line EL3000 series ABB CEMS Datasheet-continuous gas analyzers
 
Project training report
Project training report Project training report
Project training report
 
In Situ Infrared Gas Analyzer
In Situ Infrared Gas AnalyzerIn Situ Infrared Gas Analyzer
In Situ Infrared Gas Analyzer
 
Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0
 
Egas datasheet(1)
Egas datasheet(1)Egas datasheet(1)
Egas datasheet(1)
 
Egas datasheet
Egas datasheetEgas datasheet
Egas datasheet
 
Waste Water aeration system challenges solutions
Waste Water aeration system challenges solutionsWaste Water aeration system challenges solutions
Waste Water aeration system challenges solutions
 
Clean air sorbent samplers
Clean air sorbent samplersClean air sorbent samplers
Clean air sorbent samplers
 
MRU SWG 100 Fixed Biogas Analyzer
MRU SWG 100 Fixed Biogas Analyzer MRU SWG 100 Fixed Biogas Analyzer
MRU SWG 100 Fixed Biogas Analyzer
 
Ec900 Process Oxygen Analysers
Ec900 Process Oxygen AnalysersEc900 Process Oxygen Analysers
Ec900 Process Oxygen Analysers
 
Real time exhaust analyzer - icat 2014
Real time exhaust analyzer - icat 2014Real time exhaust analyzer - icat 2014
Real time exhaust analyzer - icat 2014
 
All you Need to Know About Sampling and Analysis - OGSI
All you Need to Know About Sampling and Analysis - OGSIAll you Need to Know About Sampling and Analysis - OGSI
All you Need to Know About Sampling and Analysis - OGSI
 
542 Programmable Gas Analyser
542 Programmable Gas Analyser542 Programmable Gas Analyser
542 Programmable Gas Analyser
 
Full report (Petronas Prai)
Full report (Petronas Prai)Full report (Petronas Prai)
Full report (Petronas Prai)
 
Gerador~Extrator de Nitrogênio CKD CORP. - Série: NS・NSU・PNA・NS-QFS
Gerador~Extrator de Nitrogênio CKD CORP. -  Série: NS・NSU・PNA・NS-QFSGerador~Extrator de Nitrogênio CKD CORP. -  Série: NS・NSU・PNA・NS-QFS
Gerador~Extrator de Nitrogênio CKD CORP. - Série: NS・NSU・PNA・NS-QFS
 
IMR 7500_eng(1)
IMR 7500_eng(1)IMR 7500_eng(1)
IMR 7500_eng(1)
 
GAS CHROMATOGRAPHY
GAS CHROMATOGRAPHYGAS CHROMATOGRAPHY
GAS CHROMATOGRAPHY
 
Dissolved Oxygen Sensors-Oxymax COS22/22D
Dissolved Oxygen Sensors-Oxymax COS22/22DDissolved Oxygen Sensors-Oxymax COS22/22D
Dissolved Oxygen Sensors-Oxymax COS22/22D
 
Nt216 c gb-1-minisonic-isd-psd
Nt216 c gb-1-minisonic-isd-psdNt216 c gb-1-minisonic-isd-psd
Nt216 c gb-1-minisonic-isd-psd
 
ATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdfATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdf
 

Recently uploaded

+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
panagenda
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
WSO2
 

Recently uploaded (20)

"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ..."I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
Vector Search -An Introduction in Oracle Database 23ai.pptx
Vector Search -An Introduction in Oracle Database 23ai.pptxVector Search -An Introduction in Oracle Database 23ai.pptx
Vector Search -An Introduction in Oracle Database 23ai.pptx
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
 
Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..
 
ICT role in 21st century education and its challenges
ICT role in 21st century education and its challengesICT role in 21st century education and its challenges
ICT role in 21st century education and its challenges
 
Six Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal OntologySix Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal Ontology
 
Corporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptxCorporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptx
 
Platformless Horizons for Digital Adaptability
Platformless Horizons for Digital AdaptabilityPlatformless Horizons for Digital Adaptability
Platformless Horizons for Digital Adaptability
 
Exploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with MilvusExploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with Milvus
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
WSO2's API Vision: Unifying Control, Empowering Developers
WSO2's API Vision: Unifying Control, Empowering DevelopersWSO2's API Vision: Unifying Control, Empowering Developers
WSO2's API Vision: Unifying Control, Empowering Developers
 
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
 
DEV meet-up UiPath Document Understanding May 7 2024 Amsterdam
DEV meet-up UiPath Document Understanding May 7 2024 AmsterdamDEV meet-up UiPath Document Understanding May 7 2024 Amsterdam
DEV meet-up UiPath Document Understanding May 7 2024 Amsterdam
 
FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
 
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWEREMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
 
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemkeProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
 

Ultramat analyzer

  • 1. OXYMAT 61 Gas Analyzers for the Determination of Oxygen 2 2 2 2 2 5 7 General Overview Benefits Application Design Function Integration 10 10 11 12 13 16 19“ unit Technical specifications Ordering data Dimensional drawings Schematics More information - Documentation - Spare parts for a 2-year and a 5-year service 17 Conditions of sale and delivery Export regulations Contact addresses Siemens PA 10 News · May 2004
  • 2. Gas Analysis OXYMAT 61 General GasGas Analytics ■ Overview ■ Design The measuring principle of the OXYMAT 61 gas analyzers is based on the paramagnetic alternating pressure method and is used to measure oxygen in gases in standard applications. Display and control panel • Large LCD panel for simultaneous display of: - Measured value - Status line - Measuring ranges • Contrast of LCD panel adjustable via menu • Permanent LED backlighting • Washable membrane keyboard with five softkeys • Menu-based operation for configuration, test functions and calibration • User help in plain text • Graphic display of concentration trend; programmable time intervals. OXYMAT 61 ■ Benefits • Integrated pump for reference gas circulation (ambient air) • High linearity • Compact design • Possibility of physical zero suppression. ■ Application Application areas • Environmental protection • Boiler control in firing systems • Quality monitoring • Process exhaust monitoring • Process optimization. Further applications • Chemical plants • Gas manufacturers • Research and development. 2 • 19“ unit with 4 HU for installation - in swing frames - in cabinets, with or without slide rails • Front panel for service can be hinged down (laptop connection) • Gas connections for sample gas input and output: pipe diameter 6 mm or 1/4" • Gas and electrical connections at the rear. Siemens Katalog PA 10 News · May 2004 Inputs and outputs • Six binary inputs freely configurable (e.g. for range switching, processing of external signals from sample preparation) • Six relay outputs freely configurable (failure, maintenance request, maintenance switch, limit alarm, external solenoid valves) • Two programmable analog inputs (e.g. cross interference correction, external pressure sensor) • Extension with eight additional binary inputs and eight additional relay outputs for automatic calibration with up to four calibration gases. Communication • RS 485 included in basic unit (connection at the rear). Options • RS 485/RS 232 converter • RS 485/Ethernet converter • Linking to networks via PROFIBUS DP/PA interface • SIPROM GA software as service and maintenance tool.
  • 3. Gas Analysis OXYMAT 61 General LED backlit graphic display and membrane keyboard with noticeable click Status line to display the analyzer status (programmable) Two code levels according to NAMUR (maintenance and specialist level) Easy operation menu control using the softkeys Display of concentrations as numbers and bargraph Display of current measuring ranges Display of start-of-scale and full-scale values ESC key to abort inputs Keyboard to enter values INFO key for help in plain text CLEAR key to delete inputs ENTER key to accept input values MEAS key to return to measurement mode OXYMAT 61, membrane keyboard and graphic display Executions of the wetted parts Standard Options Gas path with hoses 19“ unit Nipple Hose Sample cell Stub sample cell Restrictor O-rings Hose coupling SS, type No. 1.4571 Viton SS, type No. 1.4571 SS, type No. 1.4571 Teflon Viton Polyamide 6 Flowmeter Metering pipe Float Float limit Elbows Duran glass Duran glass, black Teflon Viton Sample gas pressure switch Membrane Enclosure Viton PA 6.3T Siemens Katalog PA 10 News · May 2004 3
  • 4. Gas Analysis OXYMAT 61 General Gas path Keys to the gas path figures 1 Sample gas inlet 8 Pressure switch in sample gas path (option) 2 Sample gas outlet 9 Flowmeter in sample gas path (option) 4 Reference gas inlet 10 Restrictor im in reference gas path (outlet) 5 Restrictor in reference gas path 11 Pressure switch for reference gas monitoring 6 O2 physics 12 Pump 7 Restrictor in sample gas path 13 Filter F 9 P 7 8 1 Sample gas 6 2 10 4 Reference gas 5 12 P 13 11 Gas path OXYMAT 61 with integrated reference gas pump (connection for 100 hPa) F 9 P 7 8 1 Sample gas 6 2 4 5 13 Gas path OXYMAT 61 with reference gas connection 3000 to 4000 hPa 4 Siemens Katalog PA 10 News · May 2004 Reference gas
  • 5. Gas Analysis OXYMAT 61 General ■ Function 1 In contrast to almost all other gases, oxygen is paramagnetic. This property is utilized as the measuring principle by the OXYMAT 61 gas analyzers. Oxygen molecules in an inhomogeneous magnetic field are drawn in the direction of increased field strength due to their paramagnetism. When two gases with different oxygen concentrations meet in a magnetic field, a pressure difference is produced between them. 2 In the case of OXYMAT 61, one gas (1) is a reference gas (N2, O2 or air), the other is the sample gas (5). The reference gas is introduced into the sample chamber (6) through two channels (3). One of these reference gas streams meets the sample gas within the area of a magnetic field (7). Because the two channels are connected, the pressure, which is proportional to the oxygen concentration, causes a cross flow. This flow is converted into an electric signal by a microflow sensor (4). 2 4 3 3 DP 5 6 The microflow sensor consists of two nickel grids heated to approx. 120 ºC which form a Wheatstone bridge together with two supplementary resistors. The pulsating flow results in a change in the resistance of the Ni grids. This results in a bridge offset which depends on the oxygen concentration in the sample gas. Because the microflow sensor is located in the reference gas stream, the measurement is not influenced by the thermal conductivity, the specific heat or the internal friction of the sample gas. This also provides a high degree of corrosion resistance because the flow sensor is not exposed to the direct influence of the sample gas. O2 7 O2 O2 O2 8 O2 By using a magnetic field with alternating strength (8), the effect of the background flow in the microflow sensor is not detected, and the measurement is thus independent of the instrument orientation. The sample cell is directly in the sample path and has a small volume. The microflow sensor thus responds quickly, resulting in a very short response time for the OXYMAT 61. 9 1 2 3 4 5 6 7 8 9 Reference gas inlet Restrictors Reference gas channels Microflow sensor for measurement Sample gas inlet Sample cell Paramagnetic effect Electromagnet with alternating field strength Sample gas and reference gas outlet OXYMAT 61, mode of operation Siemens Katalog PA 10 News · May 2004 5
  • 6. Gas Analysis OXYMAT 61 General Special characteristics • Four freely-parameterizable measuring ranges, also with zero offset, all measuring ranges linear • Electrically isolated signal output selectable as 0/2/4 to 20 mA (also inverted) • Autoranging or manual range switching possible; remote switching is also possible • Storage of measured values possible during calibration • Time constants selectable within wide limits (static/dynamic noise suppression); i.e. the response time of the analyzer can be matched to the respective application • Simple handling using menu-based operation • Low long-term drift • Two operation levels with separate access code to prevent unintentional and unauthorized inputs • Automatic range calibration can be parameterized • Operation based on NAMUR Recommendation • Monitoring of sample gas (option) • Customer-specific analyzer options such as e.g.: - Customer acceptance - TAG labels - Drift recording • Simple handling using a digital membrane keyboard and menu-based operation • Short response time • Reference gas supply external (N2, O2 or air, approx. 3000 hPa) or via an integrated reference gas pump (ambient air, approx. 100 hPa) • Monitoring of reference gas with reference gas connection 3000 to 4000 hPa • Different smallest spans, depending on version 2.0% or 5.0% O2 • Internal pressure sensor to correct sample gas variations. Correction of zero error / Cross interferences Residual gas (concentration 100 % v/v) Zero deviation in % v/v O2 absolute Organic gases Residual gas (concentration 100 % v/v) Zero deviation in % v/v O2 absolute Inert gases CH3COOH -0.64 Argon Ar -0.25 Acetylene C2H2 -0.29 Helium He +0.33 1,2 butadiene C4H6 -0.65 Krypton Kr -0.55 1,3 butadiene C4H6 -0.49 Neon Ne +0.17 iso-butane C4H10 -1.30 Xenon Xe -1.05 n-butane C4H10 -1.26 1-butene C4H6 -0.96 iso-butene C4H8 -1.06 Ammonia NH3 -0.20 C6H12 -1.84 Carbon dioxide CO2 -0.30 Dichlorodifluoromethane (R12) CCl2F2 -1.32 Carbon monoxide CO +0.07 Ethane C2H6 -0.49 Chlorine Cl2 -0.94 Ethylene C2H4 -0.22 Dinitrogen monoxide n-heptane C7H16 -2.4 Hydrogen n-hexane C6H14 -2.02 Methane CH4 -0.18 Methanol CH3OH -0.31 Hydrogen fluoride Acetic acid Cyclo-hexane Anorganic gases N2 O -0.23 H2 +0.26 Hydrogen bromide HBr -0.76 Hydrogen chloride HCl -0.35 HF -0.10 HI -1.19 n-octane C8H18 -2.78 Hydrogen iodide n-pentane C5H12 -1.68 Hydrogen sulphide iso-pentane C5H12 -1.49 Oxygen Propane C3H8 -0.87 Nitrogen Propylene C3H6 -0.64 Nitrogen dioxide Trichlorofluoromethane (R11) H2S -0.44 O2 +100 N2 0.00 NO2 +20.00 CCl3F -1.63 Nitrogen oxide NO +42.94 Vinyl chloride C2H3Cl -0.77 Sulphur dioxide SO2 -0.20 Vinyl fluoride C2H3F -0.55 Sulphur hexafluoride SF6 -1.05 C2H2Cl2 -1.22 Water H2O -0.03 1,1 vinylidene chloride Zero error due to diamagnetism or paramagnetism of residual gases with nitrogen as the reference gas at 60 °C and 1000 hPa absolute (according to IEC 1207/3) Conversion to other temperatures: The zero errors mentionned in the table must be multiplied with a correction factor (k): • with diamagnetic gases: k = 333 K / (θ [°C] + 273 K) • with paramagnetic gases: k = [333 K / (θ [°C] + 273 K)]2 (all diamagnetic gases have a negative zero error). 6 Siemens Katalog PA 10 News · May 2004
  • 7. Gas Analysis OXYMAT 61 General Interface parameters ■ Integration Level Communication Communications facilities 8 Stop bit 1 Start bit 1 Parity None No echo mode The standard integrated series interface permits to communicate between several gas analyzers via the internal bus (ELAN). Up to 12 gas analyzers with max. four components can be networked. The operating principle is shown in the following figure. 2 RS485 Converter RS232 RS485 RS485 Converter RS232 RS485 8 5 6 7 5 Typical design of an RS 485 network Item Designation 1 2 Computer RS 485/RS 232 converter with RS 485/RS 232 cable RS 485 bus connector with jumper Analyzers RS 485 cable RS 485 bus connector RS 485 network 9-pin SUB-D plug Option: RS 485 repeater A5E00054148 C79451-Z1589-U1 RS 485/Ethernet converter C79451-A3364-D61 Further accessories (e.g. cable, connectors, repeater, etc.) see Catalog IK PI RS 485 interface (SIPROM GA) Up to 12 gas analyzers with max. four components can be networked. 3 4 RS232 Order No. Interface description (d) SIPROM GA is a software tool for service and maintenance. All analyzer functions (except factory functions) can be remotecontrolled and monitored via RS 485/RS 232 converters. 1 RS232 Ordering data RS 485/RS 232 converter RS 485 interface (ELAN) 3 4 5 6 7 8 9 9600 Data bits The gas analyzers of series 6 (ULTRAMAT 6, ULTRAMAT/OXYMAT 6, OXYMAT 6, OXYMAT 61, FIDAMAT 6 and CALOMAT 6), as well as the ULTRAMAT 23 offer the following communications facilities: • RS 485 interface • PROFIBUS DP/PA • AK interface (only OXYMAT 6, ULTRAMAT 6 and ULTRAMAT/OXYMAT 6). 9 RS 485 Baud rate Networking of several gateways is possible when using the RS 485/Ethernet converter (gateway). The number of operatable analyzers is increased correspondingly. Functions • Display and saving of all analyzer data • Remote operation of all analyzer functions • Parameter and configuration settings • Comprehensive diagnostics information • Remote calibration • Online help • Cyclic saving of measured values and status on hard disk • Exporting to commercially available application programs • Downloading of new software. Hardware requirements: • PC/Laptop Pentium 133 MHz, RAM 32 MB, CD-ROM drive • Free disk capacity min. 10 MB • Free COM-Port: RS 232 or RS 485: - for coupling to ELAN network RS 485/RS 232 - Linking the RS 485/Ethernet converter requires a 10 MB standard network (RJ 45 connection) with TCP/IP. The distance should not exceed 500 m with an RS 485 network, a repeater should be inserted for a longer distance. Software requirements: • Windows 95 • Windows 98 • Windows NT 4.0 • Windows 2000 • Windows XP. Siemens Katalog PA 10 News · May 2004 7
  • 8. Gas Analysis OXYMAT 61 General Ordering data Order No. SIPROM GA software German/English selectable during installation, comprising 1 CD, with installation instructions, software product certificate and registration form S79610-B4014-A1 Firmware retrofitting sets for older analyzers: ULTRAMAT 23 (prior to SW version 4.1) all langages C79451-A3494-S501 ULTRAMAT 6 (prior to SW version 4.1) • German C79451-A3478-S501 • English C79451-A3478-S502 • French C79451-A3478-S503 • Spanish C79451-A3478-S504 • Italian C79451-A3478-S505 OXYMAT 6 (prior to SW version 4.1) • German C79451-A3480-S501 • English C79451-A3480-S502 • French C79451-A3480-S503 • Spanish C79451-A3480-S504 • Italian C79451-A3480-S505 Basic structure of a PROFIBUS system 8 Siemens Katalog PA 10 News · May 2004 PROFIBUS DP/PA PROFIBUS DP/PA is the leading field bus on the market. All Siemens gas analyzers are suitable for PROFIBUS when equipped with an optional plug-in card (retrofitting also possible) and satisfy the binding "Device profile for analyzers" defined by the PNO (PROFIBUS user organization). Central access to the analyzers in the system is possible using the SIMATIC PDM operator input software. The term field bus describes a digital communications system with which distributed field devices in a plant are networked together via one single cable, and connected at the same time to programmable controllers or to a process control system. PROFIBUS is the leading field bus on the market. The PROFIBUS DP version is widely used for production automation because of its high transmission ratefor relatively small data quantities per device, whereas PROFIBUS PA particularly takes into account the features required for process engineering, e.g. large data quantities and application in potentially explosive atmospheres. User benefits can be found in the extremely high potentials for cost savings in all areas of the plant, covering configuring and commissioning, operation and maintenance, and up to later plant extensions. Operation of the gas analyzers from a control system or separate PC is possible using the SIMATIC PDM (Process Device Manager) operator input tool which is software executing under Windows and which can also be incorporated into the SIMATIC PCS 7 process control system. This permits clear display of both the incorporation of devices into the system and the complex parameter structure of the analyzers, permitting operation to be carried out simply by clicking.
  • 9. Gas Analysis OXYMAT 61 General The PROFIBUS user organization (PNO) is an independent international institution, and represents the interests of many vendors and users. In addition to services such as consultation, training and device certification, its prime task is the further development, standardization and promotion of the PROFIBUS technology. The definition of a binding functionality for a device class in a profile is a prerequisite for the uniform response of devices from different vendors, the so-called interoperability.The profile for analyzers was defined as binding at the end of 1999, thus guaranteeing the interaction of all PROFIBUS-based devices in a plant. A differentiation is made between cyclic and acyclic services. Cyclic services are used to transmit time-critical data such as measured values and statuses. The acyclic services permit the scanning or modification of device parameters during operation. This profile defines the functionality of the analyzers in a block model: e.g. the physical block describes the measuring procedure, analyzer and vendor names, serial number and operating state (operation, maintenance). Various functional blocks contain the execution of specific functions such as the processing of measured values or alarms. The transducer blocks describe the functionality of the actual measuring procedure and its control, e.g. preprocessing of a measured value, correction of crossinterferences, characteristics, measuring ranges as well as switching and control procedures. Protocols define the data transmission between the stations on the bus. The user benefits of numerous functions especially in the automotive industry, e.g. to operate a re-linearization All gas analyzers of Series 6, (ULTRAMAT 6, ULTRAMAT/OXYMAT 6, OXYMAT 6, OXYMAT 61, FIDAMAT 6 and CALOMAT 6), as well as the ULTRAMAT 23 are suitable for PROFIBUS when fitted with the optional plug-in card. AK interface (only OXYMAT 6, ULTRAMAT 6 and ULTRAMAT/OXYMAT 6) Unlike to PROFIBUS and ELAN, communication between only one unit and one PC is possible and operates according to the master-slave principle. The unit only transmits data on request with a command message, but always only command can be processed and answered. Funktion 88 permits to call the menu and to set the parameters. Siemens Katalog PA 10 News · May 2004 9
  • 10. Gas Analysis OXYMAT 61 19" unit ■ Technical specifications General Measuring ranges Smallest possible measuring span (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) Largest possible measuring span Measuring ranges with zero offset Position of use Conformity Design, enclosure Degree of protection Weight Electrical characteristics Power supply Power consumption EMC interference immunity (ElectroMagnetic Compatibility) Electrical safety Fuses Gas inlet conditions Perm. sample gas pressure • External reference gas supply • With integrated pump Sample gas flow Sample gas temperature Sample gas humidity Time response Warm-up period Response time (T90) Damping (electric time constant) Dead time (purging time of gas path in analyzer at 1 l/min) Time for internal signal processing Pressure correction range Pressure sensor, internal 4, switchable internally and externally; autoranging is also possible 2% v/v or 5% v/v O2 100% v/v O2 Any zero point is possible between 0 to 100% v/v as long as a suitable reference gas is used Front panel vertical CE identification EN 50081-1, EN 50082-2 IP20 according to EN 60529 Approx. 13 kg 100 to 120 V AC (rated range 90 V to 132 V), 48 to 63 Hz or 200 to 240 V AC (rated range 180 V to 264 V), 48 to 63 Hz Approx. 37 VA According to standard requirements of NAMUR NE21 (08/98) According to EN 61010-1, overvoltage category III 100...120 V: 1.0T/250 200...240 V: 0.63T/250 Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) Output signal fluctuation < 0.75% of smallest possible measuring range specified on rating plate with an electronic time constant of 1 s (corresponds to ± 0.25% with 2 σ) Zero drift < 0.5%/month of smallest possible meas. span specified on rating plate Measured-value drift < 0.5%/month of respective measuring span Repeatability < 1% of respective measuring span Linearity error < 1% of respective measuring span Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) Ambient temperature < 0.1% [% O2] /1 K referred to the smallest possible measuring span according to rating plate Sample gas pressure with air Without pressure compensation: (approx. 100 hPa) as refer< 2% of measuring span / 1% ence gas supply, a correction change in pressure of the atmospheric pressure With pressure compensation: fluctuations is only possible < 0,2% dof measuring span / when the sample gas is vented 1% hange in pressure to ambient air) Residual gases Deviation in zero point corresponding to paramagnetic or diamagnetic deviation of residual gas (see Table) Sample gas flow < 1% of smallest possible measuring span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range Power supply < 0.1% of output signal span with rated voltage ± 10% Electric inputs and outputs Analog output Relay outputs 800 to 1200 hPa atmospheric pressure ± 50 hPa 18 to 60 l/h (0.3 to 1 l/min) 0 to 50 °C < 90% relative humidity Analog inputs With ambient temperature < 30 min (maximum accuracy achieved after 2 hours) 3.5 s 0 to 100 s, programmable Approx. 0.5 to 2.5 s, depending on version <1s Binary inputs Serial interface Options Ambient conditions Perm. ambient temperature 500 to 2000 hPa (permitted sample gas pressure see gas inlet conditions) Permissible humidity 0/2/4 to 20 mA, floating; max. load 750 Ω 6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC/ 1 A, floating 2, designed for 0/2/4 to 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference) 6, designed for 24 V, floating, freely selectable, e.g. for range switching RS 485 Autocal function with 8 additional binary inputs and 8 relay outputs; also with PROFIBUS PA or PROFIBUS DP -30 to +70 °C during storage and transport, +5 to +45 ºC during operation < 90% relative humidity as annual average, during storage and transport (dew point must not be fallen below) Reference gases Measuring range N2 . . . to 100% v/v O2 (suppressed zero with full-scale value 100% v/v O2) O2 Around 21% v/v O2 (suppressed zero with 21% v/v O2 within the span) 10 Recommended reference gas 0 to . . . % v/v O2 Air Siemens Katalog PA 10 News · May 2004 Reference gas pressure 3000 to 4000 hPa absolute, reference gas from cylinder atmospheric pressure with internal reference gas pump Remarks The reference gas flow is set automatically to 5 to 10 ml/min
  • 11. Gas Analysis OXYMAT 61 19" unit Ordering data Order No.. OXYMAT 61 gas analyzer 19“ unit for installation in cabinets 7 MB 2 0 0 1 - 77 A 0 0 - 7777 Gas connections for sample gas and reference gas Piping with outer diameter 6 mm Piping with outer diameter ¼" 0 1 Smallest possible span O2 2% Reference gas pressure 3000 hPa 2% Reference gas supply with internal pump 5% Reference gas pressure 3000 hPa 5% Reference gas supply with internal pump C D E F Power supply 100 V to 120 V AC, 48 to 63 Hz 0 200 V to 240 V AC, 48 to 63 Hz 1 Sample gas monitoring Without A With (incl. flowmeter and pressure switch) D Supplementary electronics A Without Autocal function • With additional 8 binary inputs/outputs B • With serial interface for the automotive industry (AK) D • With additional 8 binary inputs/outputs and PROFIBUS PA interface E • With additional 8 binary inputs/outputs and PROFIBUS DP interface F Language German 0 English 1 French 2 Spanish 3 Italian 4 Order code Further versions Please add „-Z“ to Order No. and specify Order code RS 485/RS 232 converter A11 Slide rails (2 rails) A31 Set of Torx tools, socket spanner A32 TAG labels (customer-defined inscriptions) B03 Damping element for sample gas Customer acceptance (in factory before delivery) B04 1) Clean for O2-Service (special-cleaned gas path) Drift recording 2) Y01 Y02 Y03 Measuring range in plain text, if different from standard setting 3) Y11 Retrofitting sets Order No. RS 485/Ethernet converter C79451-A3364-D61 RS 485/RS 232 converter C79451-Z1589-U1 Autocal function with 8 binary inputs/outputs C79451-A3480-D511 Autocal function with 8 binary inputs/outputs and PROFIBUS-PA A5E00057307 Autocal function with 8 binary inputs/outputs and PROFIBUS-DP A5E00057312 1) Customer acceptance: ½ day at factory in presence of customer. The following work is carried out: comparison of analyzer with ordering data, linearization check (zero, mid-point value and full-scale value), reproducibility check with calibration gas (recording in each case on XT recorder, logging of results). 2) Drift recording: an XT recording is supplied when the analyzer is delivered; zero drift with 16 hours continuous operation and sensitivity drift (largest measuring range) with 6 hours continuous operation. 3) Standard setting: Measuring range 1: 0 to smallest possible span Measuring range 2: 0 to 10% Measuring range 3: 0 to 25% Measuring range 4: 0 to 100%. Siemens Katalog PA 10 News · May 2004 11
  • 12. Gas Analysis OXYMAT 61 19" unit ■ Dimensional drawings 384.5 351.5 27.5 15.6 4 82 128 149.5 172 37.7 101.6 177 173 M4 86.5 OXYMAT 61 465 483 4 OXYMAT 61, 19“ unit, dimensions in mm 12 Siemens Katalog PA 10 News · May 2004 2 1 411.5 320 345 132 104 36 0 440
  • 13. Gas Analysis OXYMAT 61 19" unit ■ Schematics Pin assignment (electrical connections) Connector SUB-D 9F (RS 485) M GND 9 8 +5 V 7 6 M 5 4 3 2 1 GND R_Level-NNC RD/TD-N RD/TD-P R_Level-PNC NC GND Possibility for connection of bus terminating resistors to pins 7 and 9. Connector SUB-D 15F M 15 14 13 12 M 11 M 10 9 M 8 7 6 5 4 3 2 1 GND NC NC Analog output 1-P Analog output 1-N NC NC Analog input 2-P Analog input 2-N Analog input 1-P Analog input 1-N Binary input 6-P Binary input 5-P Binary input 5 to 6-N GND Floating analog outputs (also with respect to one another), RL £ 750 W Pressure correction Pressure correction Interfering gas correction Interfering gas correction Non-floating analog inputs, 0 to 20 mA/500 W or 0 to 10 V (low-resistance) Connector SUB-D 25F M 25 24 23 22 21 20 19 18 17 16 15 M 14 13 12 11 10 9 8 7 GND Binary input 4-P Binary input 3-P Binary input 2-P Binary input 1-P Binary input 1 to 4-N Relay 6 Relay 5 Relay 4 6 5 4 3 Floating via opto isolator "0" = 0 V (0 to 4.5 V) "1" = 24 V (13 to 33 V) Relay 3 Contact loading max. 24 V/1 A, AC/DC; relay contacts shown: de-energized relay coil Relay 2 2 Relay 1 1 GND Note: Cable and plug must be shielded and connected to chassis potential. OXYMAT 61, 19“ unit, pin assignment Siemens Katalog PA 10 News · May 2004 13
  • 14. Gas Analysis OXYMAT 61 19" unit Pin assignment (electrical connections) Connector SUB-D 37F (option) M 37 36 35 34 33 32 31 30 29 28 27 26 25 24 23 22 21 20 M 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 GND NC NC Binary input 14-P Binary input 13-P Binary input 12-P Binary input 11-P Binary input 10-P Binary input 9-P Binary input 8-P Binary input 7-P Binary input 7 to 14-N floating via opto coupler "0" = 0 V (0 to 4.5 V) "1" = 24 V (13 to 33 V) Relay 14 Relay 13 Relay 12 Relay 11 Contact loading max. 24 V/1 A, AC/DC; relay contacts shown: de-energized relay coil Relay 10 Relay 9 Relay 8 2 Relay 7 1 GND Note: Cable and plug must be shielded and connected to chassis potential. Connector SUB-D 9F -X90 PROFIBUS DP 9 8 7 6 5 4 3 2 1 optional Connector SUB-D 9M -X90 PROFIBUS PA DGND CNTR-N CNTR-P/direction control RxD/TxD-N (A) RxD/TxD-P (B) VP /+ 5 V OXYMAT 61, 19“ unit, pin assignment of the AUTOCAL board and PROFIBUS connectors 14 Siemens Katalog PA 10 News · May 2004 9 8 7 6 5 4 3 2 1 Pa-N(-) PA-P(+)
  • 15. Gas Analysis OXYMAT 61 19" unit Gas and electrical connections Sample gas outlet Sample gas inlet Reference gas inlet 15-pin connector: Binary inputs and analog inputs/outputs 9-pin connector: RS 485 9-pin interface connector (option): e.g. PROFIBUS Power supply and fuses Gas outlet Gas connections: stubs 6 mm or ¼" 25-pin connector: Binary inputs and relay outputs 37-pin connector: Binary inputs and relay outputs (option) OXYMAT 61, 19“ unit, gas and electrical connections Siemens Katalog PA 10 News · May 2004 15
  • 16. Gas Analysis OXYMAT 61 19" unit ■ More information Documentation Manual Order No. OXYMAT 61 Gasanalysengerät für IR-absorbierende Gase und Sauerstoff (German) A5E00123066 OXYMAT 61 Gas Analyzers for IR-absorbing Gases and Oxygen (English) A5E00123067 OXYMAT 61 Analyseurs de gaz pour la mesure de composants infrarouges et doxygène (French) A5E00123068 OXYMAT 61 Analizadores para gases absorbentes de infrarrojo y oxígeno (Spanish) A5E00123069 OXYMAT 61 Analizzatori per i gas assorbenti raggi infrarossi ed ossigeno (Italian) A5E00123070 Proposition of spare parts for a 2-year and 5-year service Qty for 2 years Qty for 5 years Order No. Reference gas supply (pump, restrictor, pressure switch, hose) 1 1 A5E00114838 O-ring 1 2 C74121-Z100-A6 Description Analyzer section Pressure switch (sample gas) 1 2 C79302-Z1210-A2 Flowmeter (only version with pump) 1 2 C79402-Z560-T1 Measuring cell • SS, type No. 1.4571, without flow-type compensation branch - 1 C79451-A3277-B535 • O-ring (measuring head) 2 4 C79121-Z100-A32 • O-ring (stub) 2 4 C71121-Z100-A159 Measuring head (non flow-type compensation branch) 1 1 C79451-A3460-B525 Restrictor for sample gas path, hose 2 2 C79451-A3480-C10 Reference gas path, 3000 hPa (set of parts) 1 1 C79451-A3480-D518 Electronics Front panel with keyboard 1 1 A5E00259978 Baseplate (without firmware) - 1 C79451-A3474-B601 Adapter board LCD/keyboard 1 1 C79451-A3474-B605 Magnet terminal board - 1 C79451-A3474-B606 LC display 1 1 W75025-B5001-B1 Connector filter - 1 W75041-E5602-K2 Fuse • 0,63 A / 250 V (230-V version) 2 4 W79054-L1010-T630 • 1,0 A / 250 V (110-V version) 2 4 W79054-L1011-T100 16 Siemens Katalog PA 10 News · May 2004
  • 17. Gas Analysis Conditions of sale and delivery Export regulations, contact addresses Comprehensive Terms and Conditions of Sale and Delivery are available free of charge from your local Siemens business office under the following Order Nos.: •6ZB5310-0KR30-0BA0 (for customers based in the Federal Republic of Germany) •6ZB5310-0KS53-0BA0 (for customers based outside of theFederal Republic of Germany) or download them from the Internet: www.siemens.com/automation/mall (Germany: A&D Mall Online-Help System) ■ Terms and Conditions of Sale and Delivery in the Federal Republic of Germany By using this catalog you can acquire hardware and software products described therein from the Siemens AG subject to the following terms. Please note! The scope, the quality and the conditions for supplies and services, including software products, by any Siemens entity having a registered office outside the Federal Republic of Germany, shall be subject exclusively to the General Terms and Conditions of the respective Siemens entity. for customers with a seat or registered office in the Federal Republic of Germany The General Terms of Payment as well as the General Conditions for the Supply of Products and Services of the Electrical and Electronics Industry shall apply. For software products, the General License Conditions for Software Products for Automation and Drives for Customers with Seat or registered Office in Germany shall apply. ■ Export regulations for customers with a seat or registered office outside of Germany The General Terms of Payment as well as the General Conditions for Supplies of Siemens, Automation and Drives for Customers with a Seat or registered Office outside of Germany shall apply. For software products, the General License Conditions for Software Products for Automation and Drives for Customers with Seat or registered Office outside of Germany shall apply. General The prices are in € (Euro) ex works, exclusive packaging. The sales tax (value added tax) is not included in the prices. It shall be debited separately at the respective rate according to the applicable legal regulations. In addition to the prices of products which include silver and/or copper, surcharges may be calculated if the respective limits of the notes are exceeded. Prices are subject to change without prior notice. We will debit the prices valid at the time of delivery. The dimensions are in mm. Illustrations are not binding. Insofar as there are no remarks on the corresponding pages, - especially with regard to data, dimensions and weights given these are subject to change without prior notice. The products listed in this catalog / price list may be subject to European / German and/or US export regulations. Therefore, any export requiring a license is subject to approval by the competent authorities. According to current provisions, the following export regulations must be observed with respect to the products featured in this catalog / price list: AL Number of the German Export List. Products marked other than “N“ require an export license. In the case of software products, the export designations of the relevant data medium must also be generally adhered to. Goods labeled with an “AL not equal to N“ are subject to a European or German export authorization when being exported out of the EU. ECCN Export Control Classification Number. Products marked other than “N“ are subject to a reexport license to specific countries. In the case of software products, the export designations of the relevant data medium must also be generally adhered to. Goods labeled with an “ECCN not equal to N“ are subject to a US re-export authorization. Even without a label or with an “AL: N“ or “ECCN: N“, authorization may be required due to the final destination and purpose for which the goods are to be used. The deciding factors are the AL or ECCN export authorization indicated on order confirmations, delivery notes and invoices. Subject to change without prior notice. 7 If you have any questions, please contact your local sales representative or any of the contact addresses below. Siemens AG A&D PI 2M Process Analytics Oestliche Rheinbrueckenstr. 50 D-76187 Karlsruhe Germany Siemens Applied Automation 500 West Highway 60 Bartlesville, OK 74003 USA Siemens Pte. Limited A&D PI2 Regional Head Quarter The Siemens Center 60 MacPherson Road Singapore 348615 Tel.: +49 721 595 4234 Fax: +49 721 595 6375 E-Mail:processanalytics@siemens.com Tel.: +1 918 662 7000 Fax: +1 918 662 7052 E-Mail: saaisales@sea.siemens.com Tel.: +65 6490 8702 Fax: +65 6490 8703 E-Mail: splanalytics.sgp@siemens.com www.processanalytics.com www.sea.siemens.com/ia www.siemens.com/processanalytics Siemens AG Automation and Drives Process Instrumentation and Analytics D-76181 Karlsruhe Federal Republic of Germany Order No.: E86060-K3510-B171-A3-7600 Printed in the Federal Republic of Germany KG K 0604 3.0 ROT 16 En / 415110 Siemens Katalog PA 10 News · May 2004 17