MProbe system integrated with specialized microscope combines the best of the both world: small spot and thickness measurement. Spot size down to 2 um.
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
MProbe 40 (MSP) brochure
1. MProbe MSP system
MProbeUVVisNIR-MSP system
Precision <0.01nm or 0.01%
Accuracy <0.2% or 1 nm
Stability <0.02nm or 0.03%
Spot Size 50 mm to 4 mm
Sample Size from 100 mm to 200mm x 200mm
Objectives 10x,20x,50x (Vis), 8x(UV-NIR).
95 parfocal, long WD objectives
Majority of translucent or lightly absorbing films can be
measured quickly and reliably:Oxides,Nitrides, Photore-sists,
Polymers, Semiconductors (Si, aSi, polySi), Compound
Semiconductors (AlGaAs, InGaAs, CdTe,CIGS),Hard coat-ings
(SiC, DLC), Polymer coatings (Paralene, PMMA, Poly-amides),
thin metal films and many more.
Thickness Range: 1 nm - 1 mm
Wavelength Range: 200nm -1700nm
Spot size: 50mm to 4 mm
Thin Solar Cells applications: aSi, TCO, CIGS, CdS, CdTe
- full solar stack measurement. LCD, FPD application: ITO,
Cell Gaps, Polyamides. Optical Coatings: dielectric filters,
hardness coating, anti-reflection coating Semiconductor and
dielectics: Oxides, Nitrides, OLED stack
Real time measurement and analysis. Multi-layer,
thin,thick, freestanding and nonuniform layers.
Extensive materials library (500+ materials) - new materials
easily added. Support of parameterized materials: Cauchy,
Tauc-Lorentz, Cody-Lorentz, EMA
and many more....
Flexible: Easy integration with external system using TCP
interface
Measurement: thickness, optical constants, surface rough-ness
User friedly and powerful: One-click measurement and
analysis. Powerful tools: simulation & sensitivity, back-ground
and scaling correction,linked layers and materials,
multisample measurements, dynamic measurement and
production batch processing.
MProbe MSP
Microscope Thin Film Measurement System
It is easy to be an expert with MProbe
Oxide (100nm) measurement, 40um spot size
Wavelength range: 200nm -1000nm
2. Basic Options/ Specification
Model
Wavelength
range
Spectrometer/Detector/Light source
Thickness range*
VIS-MSP
400-1100 nm
Spectrometer F4/Si 3600 pixels/ Tungsten - Halogen light source
15 nm to 20 mm
(option:up to 50 mm)
UVVisSR-MSP
200-1000 nm
Spectrometer F4/ Si CCD 3600 pixels/ Deuterium & Tungsten-Halogen light source
1 nm to 20 mm
(option:up to 50 mm)
HRVIS-MSP
700-1000 nm
HR Spectrometer F4/Si 3600 pixels/ Tungsten - Halogen light source
1 mm to 400 mm
NIR-MSP
900-1700nm
Transmission Spectrometer (TVG) F2/512 InGaAs/Tungsten-Halogen light source
100 nm-200 mm
VISNIR-MSP
400-1700 nm
Spectrometer F4 Si CCD 3600 pixels(Vis channel);Transmission Spectrometer (TVG)F2/512 InGaAs PDA( NIR channel)
Tungsten-Halogen light source
15 nm to 200 mm
UVVISNIR- MSP
200 -1700 nm
Spectrometer F4 Si CCD 3600 pixels(Vis channel);Transmission (TVG) F2/512 InGaA ( NIR channel)
Deuterium & Tungsten-Halogen light source
1 nm -300 mm
XT-MSP
1590nm -1650nm
Transmission Spectrometer (TVG) F2/512 InGaAs/Tungsten-Halogen light source
10 mm- 1 mm
* T, n & k measurement in 25nm - 5mm thickness range
Other configuration are available. OEM inquiries and custom development projects are welcome.
One year limited warranty on labor and materials for all system.
Semiconsoft, Inc, 2012
tel. +1.617.388.6832 email: info@semiconsoft.com fax.+1.508.858.5473
Visit us : http://www.semconsoft.com
Option
Description
Comments
-MXY[6 or 8]
Motorized XY stage 6” x6” (150mm) or 8”x8” (200mm). Controller and software support for mapping is included. 0.5 mm step size, 1 mm repeatability
8”x8” manual stage is included as standard with all models
-TOM
-TOSwitch
Transmittance measurement configuration. Includes: glass insert for stage, light source/condenser, fiber optics .
Optional -TOSWITCH to combine reflectance and trasmittance
for Vis and IR models
-RetCCD
Camera/reticle block icludes SCMOS camera and reticle. Allows imaging of the sample and marking of the measurement spot.
Can be used with any upright microscope
-APO100
APO 100X objective (visible), 95mm parforcal,
R=0.7 mm
10x,20x,50x and 8x UV-NIR are included as standard
-CCD
Eyepiece 2MP camera
Other cameras:UV, Vis, IR available