MENA3100
Scanning Electron Microscopy
(SEM)
Anders Werner Bredvei Skilbred
Harald Fjeld
MENA3100
Outline
• What can we use a SEM for?
• How do we get an image?
• Electron beam-sample interactions
• Signals that can be used to characterize the
microstructure
– Secondary electrons
– Backscattered electrons
– X-rays
• Components of the SEM
• Some comments on resolution
• Summary
MENA3100
The most versatile instrument for a
materials scientist?
What can we study in a SEM?
• Topography and morphology
• Chemistry
• Crystallography
• Orientation of grains
• In-situ experiments:
– Reactions with atmosphere
– Effects of temperature
“Easy” sample
preparation!!
“Big” samples!
MENA3100
Topography and morphology
• High depth of
focus
Image: Camilla Kongshaug, UiO
Image: Christian Kjølseth, UiO
MENA3100
Depth of focus
Optical microscopy vs SEM
• A SEM typically has orders of magnitude better
depth of focus than a optical microscope making
SEM suitable for studying rough surfaces
• The higher magnification, the lower depth of focus
Screw length: ~ 0.6 cm
Images: the A to Z of Materials
MENA3100
Ce
Fe Sr
Chemistry Images: Harald Fjeld, UiO
MENA3100
In-situ imaging
• A modern SEM can be equipped with various
accessories, e.g. a hot stage
MENA3100
In-situ imaging: oxidation of steel at
high temperatures
• 800 °C, pH2O = 667 Pa
• Formation of Cr2O3
Images: Anders W. B. Skilbred, UiO
2 min 10 min 90 min
MENA3100
How do we get an image?
• In brief: we shoot high-energy electrons and
analyze the outcoming electrons/x-rays
Electrons in
Electrons out
or: x-rays out
MENA3100
The instrument in brief
MENA3100
How do we get an image?
156 electrons!
Image
Detector
Electron gun
288 electrons!
MENA3100
Electron beam-sample interactions
• The incident electron beam is scattered in the sample,
both elastically and inelastically
• This gives rise to various signals that we can detect
(more on that on next slide)
• Interaction volume increases with increasing acceleration
voltage and decreases with increasing atomic number
Images: Smith College Northampton, Massachusetts
MENA3100
Signals from the sample
Incoming electrons
Secondary electrons
Backscattered
electrons
Auger electrons
X-rays
Cathodo-
luminescence (light)
Sample
MENA3100
Image: see the text book, page 281.
MENA3100
Image: Department of Geology and
Geophysics, Louisiana State University
Where does the signals come from?
• Diameter of the interaction
volume is larger than the
electron spot
 resolution is poorer than the
size of the electron spot
MENA3100
Secondary electrons (SE)
• Generated from the collision
between the incoming electrons
and the loosely bonded outer
electrons
• Low energy electrons (~10-50 eV)
• Only SE generated close to
surface escape (topographic
information is obtained)
• Number of SE is greater than the
number of incoming electrons
• We differentiate between SE1 and
SE2
MENA3100
SE1
• The secondary electrons that are
generated by the incoming electron beam
as they enter the surface
• High resolution signal with a resolution
which is only limited by the electron beam
diameter
MENA3100
SE2
• The secondary electrons that are generated by the
backscattered electrons that have returned to the surface
after several inelastic scattering events
• SE2 come from a surface area that is bigger than the spot
from the incoming electrons  resolution is poorer than for
SE1 exclusively
Sample surface
Incoming electrons
SE2
MENA3100
Factors that affect SE emission
1. Work function of the surface
2. Beam energy and beam current
• Electron yield goes through a maximum at low acc.
voltage, then decreases with increasing acc. voltage
(page 283)
Incident electron energy / kV
Secondary
electron
yield
MENA3100
Factors that affect SE emission
3. Atomic number (Z)
• More SE2 are created
with increasing Z
• The Z-dependence is
more pronounced at
lower beam energies
4. The local curvature of the
surface (the most
important factor)
Image: Smith College Northampton, Massachusetts
MENA3100
High resolution image setup
• By placing the
secondary electron
detector inside the
lens, mainly SE1
are detected
– Resolution of 1 – 2
nm is possible
Image: see the text book (page 286)
MENA3100
Backscattered electrons (BSE)
• A fraction of the incident
electrons is retarded by
the electro-magnetic field
of the nucleus and if the
scattering angle is greater
than 180 ° the electron can
escape from the surface
MENA3100
Backscattered electrons (BSE)
• High energy electrons (elastic scattering)
• Fewer BSE than SE
• We differentiate between BSE1 and BSE2
Image: See the text book, page 281.
MENA3100
BSE2
Sample surface
Incoming electrons
BSE2
• Most BSE are of BSE2 type
MENA3100
BSE as a function of atomic number
Image: University of Cape Town
• For phases containing more than one element, it is the average
atomic number that determines the backscatter coefficient h
MENA3100
Factors that affect BSE emission
• Direction of the irritated surface
– more electrons will hit the BSE detector when
the surface is aligned towards the BSE
detector
• Average atomic number
• When you want to study differences in
atomic numbers the sample should be as
levelled as possible (sample preparation is
an issue!)
MENA3100
BSE vs SE
Images: Greg Meeker, USGS
MENA3100
X-rays
• Photons not electrons
• Each element has a
fingerprint X-ray signal
• Poorer spatial resolution
than BSE and SE
• Relatively few X-ray signals
are emitted and the
detector is inefficient
 relatively long signal
collecting times are needed
MENA3100
X-rays
• Most common spectrometer: EDS
(energy-dispersive spectrometer)
• Signal overlap can be a problem
• We can analyze our sample in different
modes
– spot analysis
– line scan
– chemical concentration map (elemental
mapping)
MENA3100
Considerations when using EDS
• Dead time
– some twenty-thirty percent is ok
• Statistics
– Signal-to-noise ratio
• Drift in electron beam with time
• Build-up of a carbonaceous contamination
film after extended periods of electron
probe irradiation
MENA3100
Components of the instrument
• electron gun (filament)
• electromagnetic optics
• scan coils
• sample stage
• detectors
• vacuum system
• computer hardware and
software (not trivial!!)
MENA3100
Electron guns
• We want many electrons per
time unit per area (high current
density) and as small electron
spot as possible
• Traditional guns: thermionic
electron gun (electrons are
emitted when a solid is heated)
– W-wire, LaB6-crystal
• Modern: field emission guns
(FEG) (cold guns, a strong
electric field is used to extract
electrons)
– Single crystal of W, etched to a
thin tip
MENA3100
Electron guns
• With field emission guns we get a smaller spot
and higher current densities compared to
thermionic guns
• Vacuum requirements are tougher for a field
emission guns
Single crystal of LaB6
Tungsten wire Field emission tip
MENA3100
Detectors
Image: Anders W. B. Skilbred, UiO
Secondary electron detector:
(Everhart-Thornley)
Backscattered electron
detector:
(Solid-State Detector)
MENA3100
Our traditional detectors
• Secondary electrons: Everhart-Thornley
Detector
• Backscattered electrons: Solid State
Detector
• X-rays: Energy dispersive spectrometer
(EDS)
MENA3100
Why do we need vacuum?
• Chemical (corrosion!!) and thermal stability
is necessary for a well-functioning
filament (gun pressure)
– A field emission gun requires ~ 10-10 Torr
– LaB6: ~ 10-6 Torr
• The signal electrons must travel from the
sample to the detector (chamber pressure)
– Vacuum requirements is dependant of the
type of detector
MENA3100
Environmental SEM: ESEM
• Traditional SEM chamber pressure:
~ 10-6 Torr
• ESEM: 0.08 – 30 Torr
• Various gases can be used
• Requires different SE detector
MENA3100
Why ESEM?
• To image challenging samples such as:
– insulating samples
– vacuum-sensitive samples (e.g. biological samples)
– irradiation-sensitive samples (e.g. thin organic films)
– “wet” samples (oily, dirty, greasy)
• To study and image chemical and physical
processes in-situ such as:
– mechanical stress-testing
– oxidation of metals
– hydration/dehydration (e.g. watching paint dry)
MENA3100
Our instrument: Quanta 200, FEI
• Field emission gun, but no in-lens detector
• ESEM
• Can be equipped with various accessories
to perform in-situ experiments (more on
this on next slide)
MENA3100
Accessories on our Quanta 200:
◦ GAD – Gaseous Analytical Detector
→ for X-ray analysis in gaseous environments
◦ GSED – Gaseous Secondary Electron Detector
→ 500 μm aperture, allowing 20 Torr
chamber pressure
◦ Hot stage GSED
→ Must be used at temperatures above 500°C
◦ EBSD – Electron Backscatter Diffraction
→ Grain orientation, grain and subgrain
structures, phase identification, micro textures
◦ Hot stages – 1000°C and 1500°C
▪ ETD – Everhart-Thornley Detector
→ Secondary electron detector
▪ LFD – Large Field Detector
→ used in low vacuum and ESEM mode (SE)
▪ SSD-BSD – Solid State Backscattered Detector
→ Backscatter electrons
▪ EDS – Energy dispersive spectroscopy
→ X-ray analysis
Images: FEI
MENA3100
Some comments on resolution
• Best resolution that can be obtained: size of the
electron spot on the sample surface
– The introduction of FEG has dramatically improved
the resolution of SEM’s
• The volume from which the signal electrons are
formed defines the resolution
– SE image has higher resolution than a BSE image
• Scanning speed:
– a weak signal requires slow speed to improve signal-
to-noise ratio
– when doing a slow scan drift in the electron beam can
affect the accuracy of the analysis
MENA3100
What have we not covered in this
presentation
• Sample preparation
• Details on the electromagnetic optics
• Alternative imaging modes:
– Cathodoluminescence
– Electron beam induces current
– Orientation imaging microscopy (available at
UiO)
• Focused ion beam microscopy
MENA3100
Summary
• The scanning electron microscope is a versatile
instrument that can be used for many purposes
and can be equipped with various accessories
• An electron probe is scanned across the surface
of the sample and detectors interpret the signal
as a function of time
• A resolution of 1 – 2 nm can be obtained when
operated in a high resolution setup
• The introduction of ESEM and the field emission
gun have simplified the imaging of challenging
samples
MENA3100
Summary
• Signals:
– Secondary electrons (SE): mainly
topography
• Low energy electrons, high resolution
• Surface signal dependent on curvature
– Backscattered electrons (BSE): mainly
chemistry
• High energy electrons
• “Bulk” signal dependent on atomic number
– X-rays: chemistry
• Longer recording times are needed

More Related Content

Similar to Principle Scanning Electron Microscopy (SEM)

Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscope
Gajendra Kumar
 

Similar to Principle Scanning Electron Microscopy (SEM) (20)

Sem mahfooz
Sem   mahfoozSem   mahfooz
Sem mahfooz
 
Introduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdfIntroduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdf
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopy
 
Microscoy description by praveen
Microscoy description by praveenMicroscoy description by praveen
Microscoy description by praveen
 
2018 HM-scanning electron microscope
2018 HM-scanning electron microscope2018 HM-scanning electron microscope
2018 HM-scanning electron microscope
 
Scanning Electron Microscopy - SEM
Scanning Electron Microscopy - SEM Scanning Electron Microscopy - SEM
Scanning Electron Microscopy - SEM
 
Characterization of nanoparticles
Characterization of nanoparticlesCharacterization of nanoparticles
Characterization of nanoparticles
 
Scanning and Transmission Electron Microscope
Scanning and Transmission Electron MicroscopeScanning and Transmission Electron Microscope
Scanning and Transmission Electron Microscope
 
Chapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptxChapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptx
 
Presentation forensic micoscopy SEM microscope.pptx
Presentation forensic micoscopy SEM microscope.pptxPresentation forensic micoscopy SEM microscope.pptx
Presentation forensic micoscopy SEM microscope.pptx
 
Scanning electron microscope (sem)
Scanning electron microscope (sem)Scanning electron microscope (sem)
Scanning electron microscope (sem)
 
Modern characterization
Modern characterizationModern characterization
Modern characterization
 
scanning electron microscope for analysis
scanning electron microscope for analysisscanning electron microscope for analysis
scanning electron microscope for analysis
 
Scanning Electron Microscope
Scanning Electron MicroscopeScanning Electron Microscope
Scanning Electron Microscope
 
Sem presentation
Sem presentationSem presentation
Sem presentation
 
Sem
SemSem
Sem
 
Scanning Electron Microscopy (SEM)
Scanning Electron Microscopy (SEM)Scanning Electron Microscopy (SEM)
Scanning Electron Microscopy (SEM)
 
Sem Poster
Sem PosterSem Poster
Sem Poster
 
scanning electron microscope (SEM)
scanning electron microscope (SEM)scanning electron microscope (SEM)
scanning electron microscope (SEM)
 
Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscope
 

Recently uploaded

Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...
Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...
Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...
Klinik kandungan
 
In Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi Arabia
In Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi ArabiaIn Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi Arabia
In Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi Arabia
ahmedjiabur940
 
Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...
nirzagarg
 
Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...
Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...
Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...
HyderabadDolls
 
Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...
gajnagarg
 
Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...
Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...
Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...
HyderabadDolls
 
Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...
Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...
Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...
HyderabadDolls
 
Abortion pills in Doha {{ QATAR }} +966572737505) Get Cytotec
Abortion pills in Doha {{ QATAR }} +966572737505) Get CytotecAbortion pills in Doha {{ QATAR }} +966572737505) Get Cytotec
Abortion pills in Doha {{ QATAR }} +966572737505) Get Cytotec
Abortion pills in Riyadh +966572737505 get cytotec
 

Recently uploaded (20)

Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...
Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...
Jual obat aborsi Bandung ( 085657271886 ) Cytote pil telat bulan penggugur ka...
 
In Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi Arabia
In Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi ArabiaIn Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi Arabia
In Riyadh ((+919101817206)) Cytotec kit @ Abortion Pills Saudi Arabia
 
Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Satna [ 7014168258 ] Call Me For Genuine Models We ...
 
Fun all Day Call Girls in Jaipur 9332606886 High Profile Call Girls You Ca...
Fun all Day Call Girls in Jaipur   9332606886  High Profile Call Girls You Ca...Fun all Day Call Girls in Jaipur   9332606886  High Profile Call Girls You Ca...
Fun all Day Call Girls in Jaipur 9332606886 High Profile Call Girls You Ca...
 
💞 Safe And Secure Call Girls Agra Call Girls Service Just Call 🍑👄6378878445 🍑...
💞 Safe And Secure Call Girls Agra Call Girls Service Just Call 🍑👄6378878445 🍑...💞 Safe And Secure Call Girls Agra Call Girls Service Just Call 🍑👄6378878445 🍑...
💞 Safe And Secure Call Girls Agra Call Girls Service Just Call 🍑👄6378878445 🍑...
 
社内勉強会資料_Object Recognition as Next Token Prediction
社内勉強会資料_Object Recognition as Next Token Prediction社内勉強会資料_Object Recognition as Next Token Prediction
社内勉強会資料_Object Recognition as Next Token Prediction
 
Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...
Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...
Sonagachi * best call girls in Kolkata | ₹,9500 Pay Cash 8005736733 Free Home...
 
Vastral Call Girls Book Now 7737669865 Top Class Escort Service Available
Vastral Call Girls Book Now 7737669865 Top Class Escort Service AvailableVastral Call Girls Book Now 7737669865 Top Class Escort Service Available
Vastral Call Girls Book Now 7737669865 Top Class Escort Service Available
 
TrafficWave Generator Will Instantly drive targeted and engaging traffic back...
TrafficWave Generator Will Instantly drive targeted and engaging traffic back...TrafficWave Generator Will Instantly drive targeted and engaging traffic back...
TrafficWave Generator Will Instantly drive targeted and engaging traffic back...
 
Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...
Top profile Call Girls In Latur [ 7014168258 ] Call Me For Genuine Models We ...
 
Ranking and Scoring Exercises for Research
Ranking and Scoring Exercises for ResearchRanking and Scoring Exercises for Research
Ranking and Scoring Exercises for Research
 
Call Girls In GOA North Goa +91-8588052666 Direct Cash Escorts Service
Call Girls In GOA North Goa +91-8588052666 Direct Cash Escorts ServiceCall Girls In GOA North Goa +91-8588052666 Direct Cash Escorts Service
Call Girls In GOA North Goa +91-8588052666 Direct Cash Escorts Service
 
Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...
Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...
Belur $ Female Escorts Service in Kolkata (Adult Only) 8005736733 Escort Serv...
 
Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...
Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...
Kalyani ? Call Girl in Kolkata | Service-oriented sexy call girls 8005736733 ...
 
Gomti Nagar & best call girls in Lucknow | 9548273370 Independent Escorts & D...
Gomti Nagar & best call girls in Lucknow | 9548273370 Independent Escorts & D...Gomti Nagar & best call girls in Lucknow | 9548273370 Independent Escorts & D...
Gomti Nagar & best call girls in Lucknow | 9548273370 Independent Escorts & D...
 
Vadodara 💋 Call Girl 7737669865 Call Girls in Vadodara Escort service book now
Vadodara 💋 Call Girl 7737669865 Call Girls in Vadodara Escort service book nowVadodara 💋 Call Girl 7737669865 Call Girls in Vadodara Escort service book now
Vadodara 💋 Call Girl 7737669865 Call Girls in Vadodara Escort service book now
 
Oral Sex Call Girls Kashmiri Gate Delhi Just Call 👉👉 📞 8448380779 Top Class C...
Oral Sex Call Girls Kashmiri Gate Delhi Just Call 👉👉 📞 8448380779 Top Class C...Oral Sex Call Girls Kashmiri Gate Delhi Just Call 👉👉 📞 8448380779 Top Class C...
Oral Sex Call Girls Kashmiri Gate Delhi Just Call 👉👉 📞 8448380779 Top Class C...
 
Digital Transformation Playbook by Graham Ware
Digital Transformation Playbook by Graham WareDigital Transformation Playbook by Graham Ware
Digital Transformation Playbook by Graham Ware
 
Abortion pills in Doha {{ QATAR }} +966572737505) Get Cytotec
Abortion pills in Doha {{ QATAR }} +966572737505) Get CytotecAbortion pills in Doha {{ QATAR }} +966572737505) Get Cytotec
Abortion pills in Doha {{ QATAR }} +966572737505) Get Cytotec
 
Nirala Nagar / Cheap Call Girls In Lucknow Phone No 9548273370 Elite Escort S...
Nirala Nagar / Cheap Call Girls In Lucknow Phone No 9548273370 Elite Escort S...Nirala Nagar / Cheap Call Girls In Lucknow Phone No 9548273370 Elite Escort S...
Nirala Nagar / Cheap Call Girls In Lucknow Phone No 9548273370 Elite Escort S...
 

Principle Scanning Electron Microscopy (SEM)

  • 1. MENA3100 Scanning Electron Microscopy (SEM) Anders Werner Bredvei Skilbred Harald Fjeld
  • 2. MENA3100 Outline • What can we use a SEM for? • How do we get an image? • Electron beam-sample interactions • Signals that can be used to characterize the microstructure – Secondary electrons – Backscattered electrons – X-rays • Components of the SEM • Some comments on resolution • Summary
  • 3. MENA3100 The most versatile instrument for a materials scientist? What can we study in a SEM? • Topography and morphology • Chemistry • Crystallography • Orientation of grains • In-situ experiments: – Reactions with atmosphere – Effects of temperature “Easy” sample preparation!! “Big” samples!
  • 4. MENA3100 Topography and morphology • High depth of focus Image: Camilla Kongshaug, UiO Image: Christian Kjølseth, UiO
  • 5. MENA3100 Depth of focus Optical microscopy vs SEM • A SEM typically has orders of magnitude better depth of focus than a optical microscope making SEM suitable for studying rough surfaces • The higher magnification, the lower depth of focus Screw length: ~ 0.6 cm Images: the A to Z of Materials
  • 7. MENA3100 In-situ imaging • A modern SEM can be equipped with various accessories, e.g. a hot stage
  • 8. MENA3100 In-situ imaging: oxidation of steel at high temperatures • 800 °C, pH2O = 667 Pa • Formation of Cr2O3 Images: Anders W. B. Skilbred, UiO 2 min 10 min 90 min
  • 9. MENA3100 How do we get an image? • In brief: we shoot high-energy electrons and analyze the outcoming electrons/x-rays Electrons in Electrons out or: x-rays out
  • 11. MENA3100 How do we get an image? 156 electrons! Image Detector Electron gun 288 electrons!
  • 12. MENA3100 Electron beam-sample interactions • The incident electron beam is scattered in the sample, both elastically and inelastically • This gives rise to various signals that we can detect (more on that on next slide) • Interaction volume increases with increasing acceleration voltage and decreases with increasing atomic number Images: Smith College Northampton, Massachusetts
  • 13. MENA3100 Signals from the sample Incoming electrons Secondary electrons Backscattered electrons Auger electrons X-rays Cathodo- luminescence (light) Sample
  • 14. MENA3100 Image: see the text book, page 281.
  • 15. MENA3100 Image: Department of Geology and Geophysics, Louisiana State University Where does the signals come from? • Diameter of the interaction volume is larger than the electron spot  resolution is poorer than the size of the electron spot
  • 16. MENA3100 Secondary electrons (SE) • Generated from the collision between the incoming electrons and the loosely bonded outer electrons • Low energy electrons (~10-50 eV) • Only SE generated close to surface escape (topographic information is obtained) • Number of SE is greater than the number of incoming electrons • We differentiate between SE1 and SE2
  • 17. MENA3100 SE1 • The secondary electrons that are generated by the incoming electron beam as they enter the surface • High resolution signal with a resolution which is only limited by the electron beam diameter
  • 18. MENA3100 SE2 • The secondary electrons that are generated by the backscattered electrons that have returned to the surface after several inelastic scattering events • SE2 come from a surface area that is bigger than the spot from the incoming electrons  resolution is poorer than for SE1 exclusively Sample surface Incoming electrons SE2
  • 19. MENA3100 Factors that affect SE emission 1. Work function of the surface 2. Beam energy and beam current • Electron yield goes through a maximum at low acc. voltage, then decreases with increasing acc. voltage (page 283) Incident electron energy / kV Secondary electron yield
  • 20. MENA3100 Factors that affect SE emission 3. Atomic number (Z) • More SE2 are created with increasing Z • The Z-dependence is more pronounced at lower beam energies 4. The local curvature of the surface (the most important factor) Image: Smith College Northampton, Massachusetts
  • 21. MENA3100 High resolution image setup • By placing the secondary electron detector inside the lens, mainly SE1 are detected – Resolution of 1 – 2 nm is possible Image: see the text book (page 286)
  • 22. MENA3100 Backscattered electrons (BSE) • A fraction of the incident electrons is retarded by the electro-magnetic field of the nucleus and if the scattering angle is greater than 180 ° the electron can escape from the surface
  • 23. MENA3100 Backscattered electrons (BSE) • High energy electrons (elastic scattering) • Fewer BSE than SE • We differentiate between BSE1 and BSE2 Image: See the text book, page 281.
  • 25. MENA3100 BSE as a function of atomic number Image: University of Cape Town • For phases containing more than one element, it is the average atomic number that determines the backscatter coefficient h
  • 26. MENA3100 Factors that affect BSE emission • Direction of the irritated surface – more electrons will hit the BSE detector when the surface is aligned towards the BSE detector • Average atomic number • When you want to study differences in atomic numbers the sample should be as levelled as possible (sample preparation is an issue!)
  • 27. MENA3100 BSE vs SE Images: Greg Meeker, USGS
  • 28. MENA3100 X-rays • Photons not electrons • Each element has a fingerprint X-ray signal • Poorer spatial resolution than BSE and SE • Relatively few X-ray signals are emitted and the detector is inefficient  relatively long signal collecting times are needed
  • 29. MENA3100 X-rays • Most common spectrometer: EDS (energy-dispersive spectrometer) • Signal overlap can be a problem • We can analyze our sample in different modes – spot analysis – line scan – chemical concentration map (elemental mapping)
  • 30. MENA3100 Considerations when using EDS • Dead time – some twenty-thirty percent is ok • Statistics – Signal-to-noise ratio • Drift in electron beam with time • Build-up of a carbonaceous contamination film after extended periods of electron probe irradiation
  • 31. MENA3100 Components of the instrument • electron gun (filament) • electromagnetic optics • scan coils • sample stage • detectors • vacuum system • computer hardware and software (not trivial!!)
  • 32. MENA3100 Electron guns • We want many electrons per time unit per area (high current density) and as small electron spot as possible • Traditional guns: thermionic electron gun (electrons are emitted when a solid is heated) – W-wire, LaB6-crystal • Modern: field emission guns (FEG) (cold guns, a strong electric field is used to extract electrons) – Single crystal of W, etched to a thin tip
  • 33. MENA3100 Electron guns • With field emission guns we get a smaller spot and higher current densities compared to thermionic guns • Vacuum requirements are tougher for a field emission guns Single crystal of LaB6 Tungsten wire Field emission tip
  • 34. MENA3100 Detectors Image: Anders W. B. Skilbred, UiO Secondary electron detector: (Everhart-Thornley) Backscattered electron detector: (Solid-State Detector)
  • 35. MENA3100 Our traditional detectors • Secondary electrons: Everhart-Thornley Detector • Backscattered electrons: Solid State Detector • X-rays: Energy dispersive spectrometer (EDS)
  • 36. MENA3100 Why do we need vacuum? • Chemical (corrosion!!) and thermal stability is necessary for a well-functioning filament (gun pressure) – A field emission gun requires ~ 10-10 Torr – LaB6: ~ 10-6 Torr • The signal electrons must travel from the sample to the detector (chamber pressure) – Vacuum requirements is dependant of the type of detector
  • 37. MENA3100 Environmental SEM: ESEM • Traditional SEM chamber pressure: ~ 10-6 Torr • ESEM: 0.08 – 30 Torr • Various gases can be used • Requires different SE detector
  • 38. MENA3100 Why ESEM? • To image challenging samples such as: – insulating samples – vacuum-sensitive samples (e.g. biological samples) – irradiation-sensitive samples (e.g. thin organic films) – “wet” samples (oily, dirty, greasy) • To study and image chemical and physical processes in-situ such as: – mechanical stress-testing – oxidation of metals – hydration/dehydration (e.g. watching paint dry)
  • 39. MENA3100 Our instrument: Quanta 200, FEI • Field emission gun, but no in-lens detector • ESEM • Can be equipped with various accessories to perform in-situ experiments (more on this on next slide)
  • 40. MENA3100 Accessories on our Quanta 200: ◦ GAD – Gaseous Analytical Detector → for X-ray analysis in gaseous environments ◦ GSED – Gaseous Secondary Electron Detector → 500 μm aperture, allowing 20 Torr chamber pressure ◦ Hot stage GSED → Must be used at temperatures above 500°C ◦ EBSD – Electron Backscatter Diffraction → Grain orientation, grain and subgrain structures, phase identification, micro textures ◦ Hot stages – 1000°C and 1500°C ▪ ETD – Everhart-Thornley Detector → Secondary electron detector ▪ LFD – Large Field Detector → used in low vacuum and ESEM mode (SE) ▪ SSD-BSD – Solid State Backscattered Detector → Backscatter electrons ▪ EDS – Energy dispersive spectroscopy → X-ray analysis Images: FEI
  • 41. MENA3100 Some comments on resolution • Best resolution that can be obtained: size of the electron spot on the sample surface – The introduction of FEG has dramatically improved the resolution of SEM’s • The volume from which the signal electrons are formed defines the resolution – SE image has higher resolution than a BSE image • Scanning speed: – a weak signal requires slow speed to improve signal- to-noise ratio – when doing a slow scan drift in the electron beam can affect the accuracy of the analysis
  • 42. MENA3100 What have we not covered in this presentation • Sample preparation • Details on the electromagnetic optics • Alternative imaging modes: – Cathodoluminescence – Electron beam induces current – Orientation imaging microscopy (available at UiO) • Focused ion beam microscopy
  • 43. MENA3100 Summary • The scanning electron microscope is a versatile instrument that can be used for many purposes and can be equipped with various accessories • An electron probe is scanned across the surface of the sample and detectors interpret the signal as a function of time • A resolution of 1 – 2 nm can be obtained when operated in a high resolution setup • The introduction of ESEM and the field emission gun have simplified the imaging of challenging samples
  • 44. MENA3100 Summary • Signals: – Secondary electrons (SE): mainly topography • Low energy electrons, high resolution • Surface signal dependent on curvature – Backscattered electrons (BSE): mainly chemistry • High energy electrons • “Bulk” signal dependent on atomic number – X-rays: chemistry • Longer recording times are needed