Laparoscopic Suturing System Industrial
Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
Date: Oct 16 2016
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
2 | P a g e
Contents
Table of Figures.............................................................................................................................................4
Scope.............................................................................................................................................................5
Problem to Solve...........................................................................................................................................5
Specifications and Assumptions................................................................................................................5
Suturing Method...........................................................................................................................................6
1. Deploy Stabilizing Augers..................................................................................................................6
2. Close the Wound with a Curved Suture Needle ...............................................................................6
3. Needle Tip Exchange Grasper ...........................................................................................................7
4. Needle Tip Grasper Retracts Curved Needle Tip ..............................................................................9
5. Curved Tube is Retracted..................................................................................................................9
Curved Tube Needle Handle Controls Overview ........................................................................................11
1. Trigger Interlock..............................................................................................................................11
2. Stainless Cable with Bicycle Style Pull Anchors...............................................................................12
Needle Tip Exchange Grasper Controls Overview ......................................................................................16
1. Needle Grasper Default State .........................................................................................................17
2. Needle Grasper Open State ............................................................................................................18
Auger Tissue Stabilizer Controls Overview .................................................................................................19
1. Deploying Tissue Augers .................................................................................................................19
2. Removing Augers ............................................................................................................................20
Manufacturability Overview .......................................................................................................................21
Curved Needle Hypodermic Tube...........................................................................................................21
Laparoscopic Suture System Main Body.................................................................................................22
Needle Tube Torque Spindle...................................................................................................................22
Curved Needle Tube and Torque Spindle Welded Assembly .................................................................23
Curved Needle Tip and Suture Assembly................................................................................................24
Grasping Needle Tip Collet......................................................................................................................24
Tissue Stabilizing Auger...........................................................................................................................27
Final Discussions .........................................................................................................................................28
Time Constrained Development Results.................................................................................................28
Delivered Therapy Considerations..........................................................................................................28
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
3 | P a g e
User Interface Considerations ................................................................................................................28
Manufacturability Considerations ..........................................................................................................28
Cleaning and Sterilization Considerations ..............................................................................................28
Design for Manufacturing Considerations..............................................................................................28
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
4 | P a g e
Table of Figures
Figure 1 Wound with Access Section View...................................................................................................5
Figure 2 Tissue Stabilizing Augers Deployed.................................................................................................6
Figure 3 Needle Extended Through the Wound ...........................................................................................7
Figure 4 Needle Tip Exchange Grasper Extended.........................................................................................7
Figure 5 Needle Tip Exchange Grasper Extended Section View ...................................................................8
Figure 6 Needle Tip Exchange Grasper Contracted Section View ................................................................8
Figure 7 Needle Tip Exchange Grasper Retracts Curved Needle Tip Section View ......................................9
Figure 8 Suture System Retracting out of Access Port................................................................................10
Figure 9 Curved Needle Handle Safety Interlock Position Section View ....................................................11
Figure 10 Curved Needle Handle Safety Interlock Position Detailed Section View....................................11
Figure 11 Curved Needle Handle Safety Pin Depressed Section View........................................................12
Figure 12 Bicycle Style Cable with Pull Anchors..........................................................................................12
Figure 13 Curved Needle Handle Trigger to Intermediate Pulley Cable Path.............................................13
Figure 14 Curved Needle Intermediate Pulley to Actuation Pin Cable Path...............................................14
Figure 15 Bicycle Style Cable with Three Pull Anchors ...............................................................................14
Figure 16 Curved Tube Needle Actuation Pin.............................................................................................15
Figure 17 Complete Laparoscopic Suturing System....................................................................................16
Figure 18 Needle Grasper Spring Default State..........................................................................................17
Figure 19 Needle Grasper Collet in Closed State Section View ..................................................................17
Figure 20 Needle Grasper Spring Depressed State.....................................................................................18
Figure 21 Needle Grasper Collet in Open State Section View ....................................................................18
Figure 22 Tissue Auger Detail View.............................................................................................................19
Figure 23 Deploying Tissue Augers .............................................................................................................19
Figure 24 Auger Control Knob Detailed View .............................................................................................20
Figure 25 Auger Control Knob Bottomed Out ............................................................................................20
Figure 26 Curved Needle Hypodermic Tube...............................................................................................21
Figure 27 Laparoscopic Suture System Main Body.....................................................................................22
Figure 28 Needle Tube Torque Spindle.......................................................................................................22
Figure 29 Curved Needle to Torque Spindle Welded Assembly .................................................................23
Figure 30 Laser Star i960 Welding Station..................................................................................................23
Figure 31 Curved Needle Tip.......................................................................................................................24
Figure 32 Grasping Needle Tip Collet Step 1 Section View.........................................................................24
Figure 33 Grasping Needle Tip Collet Step 2...............................................................................................25
Figure 34 Grasping Needle Tip Collet Step 3...............................................................................................25
Figure 35 Needle Grasping Collet is Laser Welded to a Control Tube........................................................26
Figure 36 Tissue Auger................................................................................................................................27
Figure 37 Auger Forming Mandrel..............................................................................................................27
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
5 | P a g e
Scope
The purpose of this document is to provide a technical design overview of a laparoscopic suturing
system. The document is made in response to a technical problem presented by a client to develop a
prototype suturing system.
Problem to Solve
Design a system that can send a single suture across an open wound with limited access as shown in
Figure 1.
Specifications and Assumptions
 Wound size: 8mm x 3mm
 Access size: 1”ID x 4” length, like a laparoscopic access point
 Visual guidance is provided with a separate system, so assume good visibility of the wound is
available
 There are organs below the wound, so the suturing solution must not puncture too deep below
the bottom of the wound
 The solution will not include electronics, pneumatics or nittinol alloys
Figure 1 Wound with Access Section View
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
6 | P a g e
Suturing Method
In this section of the document the suturing method is reviewed by focusing on the therapy that is
delivered to the wound site. The approach to the suturing system resembles how a surgeon would
suture a wound using a curved needle.
1. Deploy Tissue Stabilizing Augers
 Augers are deployed around the wound in order to stabilize the tissue before a curved needle
can be used as shown in Figure 2
 Note the position of the pink tip of the stowed curved tube needle
Figure 2 Tissue Stabilizing Augers Deployed
2. Close the Wound with a Curved Suture Needle
 A curved needle is driven into the tissue as shown Figure 3
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
7 | P a g e
Figure 3 Needle Extended Through the Wound
3. Needle Tip Exchange Grasper
 The needle tip exchange grasper is extended over the tip of the needle as shown in Figure 4 and
Figure 5
 Note: the actual suture is not modeled in the images shown below
Figure 4 Needle Tip Exchange Grasper Extended
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
8 | P a g e
Figure 5 Needle Tip Exchange Grasper Extended Section View
 The collet of the needle tip exchange grasper is closed around the needle tip edge as shown in
Figure 6
Figure 6 Needle Tip Exchange Grasper Contracted Section View
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
9 | P a g e
4. Needle Tip Grasper Retracts Curved Needle Tip
 The needle tip is pulled away from the curved tube with the collet as shown Figure 7.
 Note the curved tube needle body is left in place to allow for easier suture routing around the
wound site.
Figure 7 Needle Tip Exchange Grasper Retracts Curved Needle Tip Section View
5. Curved Tube is Retracted
 Once the suture is retrieved to the top of the access port, the curved tube body is retracted into
nominal position.
 The tissue stabilizing augers are retracted.
 The suture system can now be retracted out of the access port as shown Figure 8.
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
10 | P a g e
Figure 8 Suture System Retracting out of Access Port
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
11 | P a g e
Curved Tube Needle Handle Controls Overview
The curved needle handle controls were inspired by Medtronics LigaSure System,
http://www.medtronic.com/covidien/products/vessel-sealing/ligasure-5-mm-blunt-tip-37-cm. The
curved tube needle handle is intended to be operated by the left hand.
1. Trigger Interlock
 To minimize accidental deployment of the curved needle a trigger safety interlock is used as
shown in Figure 9 and Figure 10.
Figure 9 Curved Needle Handle Safety Interlock Position Section View
Figure 10 Curved Needle Handle Safety Interlock Position Detailed Section View
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
12 | P a g e
 In order to actuate the trigger, a push button must be depressed in order to disable the safety
pin as shown Figure 11.
 Note the safety pin is sprung such that when the trigger is retracted to normal position the
safety pin automatically deploys.
 Note the disarm push button is also sprung so that the user has to actively hold down the safety
over-ride button with their thumb in order to actuate the curved tube needle.
Figure 11 Curved Needle Handle Safety Pin Depressed Section View
2. Stainless Cable with Bicycle Style Pull Anchors
 All curved tube needle actuations are performed using a set of multi-strand stainless steel cables
with bicycle style anchors as shown in Figure 12.
Figure 12 Bicycle Style Cable with Pull Anchors
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
13 | P a g e
 The trigger pulley rotates the intermediate pulley by a cable (highlighted blue cable path shown
in Figure 13).
 A secondary cable (also highlighted in blue Figure 14) is used to move the curved needle
actuation pin as shown in Figure 16.
Figure 13 Curved Needle Handle Trigger to Intermediate Pulley Cable Path
 The curved needle is actuated by a second cable that has three pull anchors attached to it as
shown in Figure 14 and Figure 15.
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
14 | P a g e
Figure 14 Curved Needle Intermediate Pulley to Actuation Pin Cable Path
 Note that the pull cable shown in Figure 15 has a hole in the center of the middle anchor. The
hole in the center of the anchor is used to actuate the curved needle pin shown in Figure 16.
Figure 15 Bicycle Style Cable with Three Pull Anchors
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
15 | P a g e
 The green pulleys shown in Figure 16 are used to reduce the friction in the mechanism as the
cables go around sharp corners.
Figure 16 Curved Tube Needle Actuation Pin
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
16 | P a g e
Needle Tip Exchange Grasper Controls Overview
The user controls for the needle grasper are very similar to a tri-finger ring syringe controls, like Qosina
part number: C1068, http://www.qosina.com/control-syringe-with-solid-ring-plunger-and-grip-with-
male-luer-lock-c1068
The user is intended to operate the needle tip exchange grasper with their right hand while
simultaneously holding the needle handle with their left hand as shown in Figure 17.
Figure 17 Complete Laparoscopic Suturing System
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
17 | P a g e
1. Needle Grasper Default State
 A spring shown in Figure 18 is used to maintain the grasper collet in a default closed state as
shown in Figure 19
Figure 18 Needle Grasper Spring Default State
Figure 19 Needle Grasper Collet in Closed State Section View
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
18 | P a g e
2. Needle Grasper Open State
 By depressing the spring on the needle grasper handle as shown in Figure 20, relative movement
of the tubes allows the grasper collet to open as shown in Figure 21.
 Note the grasping collet is manufactured such that its default state is open and only by pulling it
into a tube it becomes closed. Manufacturability of the grasping collet is reviewed in
subsequent sections of this document.
Figure 20 Needle Grasper Spring Depressed State
Figure 21 Needle Grasper Collet in Open State Section View
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
19 | P a g e
Auger Tissue Stabilizer Controls Overview
Control method for augers is inspired by various endoscopic procedures.
Figure 22 Tissue Auger Detail View
1. Deploying Tissue Augers
 Tissue augers are deployed by rotating clockwise the auger control knobs with your right hand
as shown in Figure 23 and Figure 24.
Figure 23 Deploying Tissue Augers
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
20 | P a g e
Figure 24 Auger Control Knob Detailed View
 Tissue augers are fully deployed once the tissue control knobs bottom out as shown in Figure
25.
Figure 25 Auger Control Knob Bottomed Out
2. Removing Augers
 Auger removal is done by rotating the auger control knobs counterclockwise.
 Once the augers are fully unwound from the tissue they are safe to pull out of the handle.
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
21 | P a g e
Manufacturability Overview
The laparoscopic suturing system is currently modeled as an easily machinable prototype. The system is
modeled as a CNC machined unit because the first step in validating the concepts shown in this
document is to construct a functioning prototype. In this section of the document the manufacturability
of only the non-trivial components will be reviewed.
Curved Needle Hypodermic Tube
The curved hypodermic needle tube shown in Figure 26, is made from 14 Gauge thin wall stainless steel
tube stock. In order to create this round shape without crushing the walls of the tube a round mandrel
must be installed inside of the tube prior to bending. Popular mandrel choices are as follows:
 Nittinol wire
 Stainless steel wire with a Teflon sleeve
 Music wire
Figure 26 Curved Needle Hypodermic Tube
Once the mandrel is installed inside of the tube, the tube may be bent around a piece of round bar with
the required diameter to form this shape. Once the shape is formed, the mandrel is removed.
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
22 | P a g e
Laparoscopic Suture System Main Body
This profile is created using a custom aluminum extrusion. The mounting holes are added in a post
process CNC milling operation. This component can also be modified so that only a CNC milling
operation is required, but it will require a few more components.
Figure 27 Laparoscopic Suture System Main Body
Needle Tube Torque Spindle
The needle tube torque spindle shown in Figure 28 is created from stainless steel round stock using a
CNC lathe to create a round shape that looks like a spin top. The spin top shape is then installed into a
CNC mill and the final profile outline is made.
Figure 28 Needle Tube Torque Spindle
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
23 | P a g e
Curved Needle Tube and Torque Spindle Welded Assembly
Figure 29 Curved Needle to Torque Spindle Welded Assembly
The curved needle tube and torque spindle are welded together at the location illustrated in Figure 29
using a laser welder. Laser Star i960 portable laser weld station shown in Figure 30 would perform well
in this type of application.
Figure 30 Laser Star i960 Welding Station
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
24 | P a g e
Curved Needle Tip and Suture Assembly
The curved needle tip is created using a CNC lathe as shown Figure 31. A suture is inserted and crimped
into place using a compression collet tool.
Figure 31 Curved Needle Tip
Grasping Needle Tip Collet
The grasping needle tip collet is manufactured from 17-4 PH Stainless Steel. 17-4 PH stainless steel is a
good material for this application because it is corrosion resistant, has a high yield stress(good for
flexures), and because it can be heat treated after it is machined and formed into to a desired shape.
This is a common method of creating a geometry that is not easily manufactured using conventional
machining methods.
 Collet is first machined using a CNC lathe as shown in Figure 32
Figure 32 Grasping Needle Tip Collet Step 1 Section View
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
25 | P a g e
 Collet is then laser cut to obtain thin slits as shown in Figure 33.
Figure 33 Grasping Needle Tip Collet Step 2
 The collet is then formed and heat treated at 1150degF for a few hours and quenched in still air
to take on the final shape and hardness as shown in Figure 34.
Figure 34 Grasping Needle Tip Collet Step 3
The grasping needle collet is welded to the tube using a laser welder as shown in Figure 35.
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
26 | P a g e
Figure 35 Needle Grasping Collet is Laser Welded to a Control Tube
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
27 | P a g e
Tissue Stabilizing Auger
Tissue augers as shown in Figure 36 are created by bending wire around a round mandrel as shown in
Figure 37.
Figure 36 Tissue Auger
Figure 37 Auger Forming Mandrel
Laparoscopic Suturing System Industrial Design Overview
Prepared by: Aleksandar Tegzes, M.Eng, P.Eng
28 | P a g e
Final Discussions
Time Constrained Development Results
The solution presented in this document provides a very complete solution with detailed
implementation notes of each the systems used. Keeping in mind that this solution took less than 24hrs
to develop and document, this is not an unreasonable first step towards a final product.
Delivered Therapy Considerations
In the current implementation thin wire (0.5mm diameter) augers are used to stabilize tissue before a
curved needle is driven through the wound. The cost/benefit consideration here is that the augers are
not introducing a significant wound to the area, but they are minimizing tissue movement which greatly
improves accuracy of needle deployment. There are many endoscopic procedures that use similar size
augers to perform wound closing operations, like OverStich product by Apollo Endosurgery:
http://apolloendo.com/overstitch/.
User Interface Considerations
The document presents preliminary user interface control considerations and assumptions. The primary
curved needle operating handle is intended to be held by the left hand and auxiliary operations that
demand slightly higher hand dexterity are operated by the right hand. Bench top testing of early
prototypes would be valuable proving ground to validate the assumption of this arrangement.
Manufacturability Considerations
Manufacturing solutions to non-trivial components of this assembly are also presented in this
document. The complexity of all of the non-trivial components in this assembly is fairly low. Assuming a
laser welder, a CNC lathe and a CNC mill are available for prototyping, a well-trained hands-on engineer
can program and fabricate all of the parts needed to construct the first prototype within 7-10 working
days.
Cleaning and Sterilization Considerations
Currently all of the components in the prototype are intended to be built from stainless steel or
aluminum. Material selection of the production components will vary depending on how the product is
intended to be marketed to the users, as either a single use or a many use sterilizable tool.
Design for Manufacturing Considerations
Once the first prototypes are built and their functionality is verified, more time could be afforded
designing more ergonomic user interface handles. Many parts can also be made from injection molded
plastic. More time could also be spent removing visible screws from the assembly; this can easily be
accomplished with thin plastic snap in place exterior panels.

More Related Content

What's hot

Mod mag m3000 manual badger meter electromagnetic flow meter_m-series
 Mod mag m3000 manual badger meter electromagnetic flow meter_m-series Mod mag m3000 manual badger meter electromagnetic flow meter_m-series
Mod mag m3000 manual badger meter electromagnetic flow meter_m-seriesENVIMART
 
Mod mag m7600 manual badger meter electromagnetic flow meter_m-series
 Mod mag m7600 manual badger meter electromagnetic flow meter_m-series Mod mag m7600 manual badger meter electromagnetic flow meter_m-series
Mod mag m7600 manual badger meter electromagnetic flow meter_m-seriesENVIMART
 
Smith randall 15-rolling-element-bearing-diagnostics-cwu
Smith randall 15-rolling-element-bearing-diagnostics-cwuSmith randall 15-rolling-element-bearing-diagnostics-cwu
Smith randall 15-rolling-element-bearing-diagnostics-cwuProto Gonzales Rique
 
Mod mag m1000 manual badger meter electromagnetic flow meter_m-series
 Mod mag m1000 manual badger meter electromagnetic flow meter_m-series Mod mag m1000 manual badger meter electromagnetic flow meter_m-series
Mod mag m1000 manual badger meter electromagnetic flow meter_m-seriesENVIMART
 
Dynasonics dxn manual badger meter portable clamp-on ultrasonic flow meter
Dynasonics dxn manual badger meter portable clamp-on ultrasonic flow meterDynasonics dxn manual badger meter portable clamp-on ultrasonic flow meter
Dynasonics dxn manual badger meter portable clamp-on ultrasonic flow meterENVIMART
 
Inflight data acquisition system
Inflight data acquisition systemInflight data acquisition system
Inflight data acquisition systemSyamim Idris
 
New holland w80 tc wheel loader service repair manual instant download
New holland w80 tc wheel loader service repair manual instant downloadNew holland w80 tc wheel loader service repair manual instant download
New holland w80 tc wheel loader service repair manual instant downloaduufsjkefkksemm
 
New holland w50 tc compact wheel loader service repair manual
New holland w50 tc compact wheel loader service repair manualNew holland w50 tc compact wheel loader service repair manual
New holland w50 tc compact wheel loader service repair manualfhsejkdkmem
 
New holland w70 tc wheel loader service repair manual
New holland w70 tc wheel loader service repair manualNew holland w70 tc wheel loader service repair manual
New holland w70 tc wheel loader service repair manualujdfjskkmmed
 
New holland w60 tc wheel loader service repair manual instant download
New holland w60 tc wheel loader service repair manual instant downloadNew holland w60 tc wheel loader service repair manual instant download
New holland w60 tc wheel loader service repair manual instant downloaduufsjkefkksemm
 
New holland w70 tc compact wheel loader service repair manual
New holland w70 tc compact wheel loader service repair manualNew holland w70 tc compact wheel loader service repair manual
New holland w70 tc compact wheel loader service repair manualfjjskekmdmme
 

What's hot (11)

Mod mag m3000 manual badger meter electromagnetic flow meter_m-series
 Mod mag m3000 manual badger meter electromagnetic flow meter_m-series Mod mag m3000 manual badger meter electromagnetic flow meter_m-series
Mod mag m3000 manual badger meter electromagnetic flow meter_m-series
 
Mod mag m7600 manual badger meter electromagnetic flow meter_m-series
 Mod mag m7600 manual badger meter electromagnetic flow meter_m-series Mod mag m7600 manual badger meter electromagnetic flow meter_m-series
Mod mag m7600 manual badger meter electromagnetic flow meter_m-series
 
Smith randall 15-rolling-element-bearing-diagnostics-cwu
Smith randall 15-rolling-element-bearing-diagnostics-cwuSmith randall 15-rolling-element-bearing-diagnostics-cwu
Smith randall 15-rolling-element-bearing-diagnostics-cwu
 
Mod mag m1000 manual badger meter electromagnetic flow meter_m-series
 Mod mag m1000 manual badger meter electromagnetic flow meter_m-series Mod mag m1000 manual badger meter electromagnetic flow meter_m-series
Mod mag m1000 manual badger meter electromagnetic flow meter_m-series
 
Dynasonics dxn manual badger meter portable clamp-on ultrasonic flow meter
Dynasonics dxn manual badger meter portable clamp-on ultrasonic flow meterDynasonics dxn manual badger meter portable clamp-on ultrasonic flow meter
Dynasonics dxn manual badger meter portable clamp-on ultrasonic flow meter
 
Inflight data acquisition system
Inflight data acquisition systemInflight data acquisition system
Inflight data acquisition system
 
New holland w80 tc wheel loader service repair manual instant download
New holland w80 tc wheel loader service repair manual instant downloadNew holland w80 tc wheel loader service repair manual instant download
New holland w80 tc wheel loader service repair manual instant download
 
New holland w50 tc compact wheel loader service repair manual
New holland w50 tc compact wheel loader service repair manualNew holland w50 tc compact wheel loader service repair manual
New holland w50 tc compact wheel loader service repair manual
 
New holland w70 tc wheel loader service repair manual
New holland w70 tc wheel loader service repair manualNew holland w70 tc wheel loader service repair manual
New holland w70 tc wheel loader service repair manual
 
New holland w60 tc wheel loader service repair manual instant download
New holland w60 tc wheel loader service repair manual instant downloadNew holland w60 tc wheel loader service repair manual instant download
New holland w60 tc wheel loader service repair manual instant download
 
New holland w70 tc compact wheel loader service repair manual
New holland w70 tc compact wheel loader service repair manualNew holland w70 tc compact wheel loader service repair manual
New holland w70 tc compact wheel loader service repair manual
 

Viewers also liked

1.29.16 Facebook Creative Ad Test
1.29.16 Facebook Creative Ad Test1.29.16 Facebook Creative Ad Test
1.29.16 Facebook Creative Ad TestRyan Walsh
 
Goalsetting 1
Goalsetting 1Goalsetting 1
Goalsetting 1ionlyspy
 
Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ...
 Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ... Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ...
Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ...NSEAkure
 
VALUEADDITIONS 04-02-2016
VALUEADDITIONS 04-02-2016VALUEADDITIONS 04-02-2016
VALUEADDITIONS 04-02-2016BNI Exponential
 
Bentonite for construction
Bentonite for construction Bentonite for construction
Bentonite for construction Sharian Raj
 
Shortage of Skilled Labour in Construction Industry of Sri Lanka
Shortage of Skilled Labour in Construction Industry of Sri LankaShortage of Skilled Labour in Construction Industry of Sri Lanka
Shortage of Skilled Labour in Construction Industry of Sri LankaPrasan Premathilaka
 
How to deliver an extremely boring presentation
How to deliver an extremely boring presentationHow to deliver an extremely boring presentation
How to deliver an extremely boring presentationSameh Shehata
 
An Excellent Boring Presentation
An Excellent Boring PresentationAn Excellent Boring Presentation
An Excellent Boring PresentationIshtiaque Zico
 
PPT ON TRENCHLESS TECHNOLOGY
PPT ON TRENCHLESS TECHNOLOGY PPT ON TRENCHLESS TECHNOLOGY
PPT ON TRENCHLESS TECHNOLOGY MD AMANULLAH
 

Viewers also liked (14)

Kalpatarumulund
KalpatarumulundKalpatarumulund
Kalpatarumulund
 
1.29.16 Facebook Creative Ad Test
1.29.16 Facebook Creative Ad Test1.29.16 Facebook Creative Ad Test
1.29.16 Facebook Creative Ad Test
 
Goalsetting 1
Goalsetting 1Goalsetting 1
Goalsetting 1
 
Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ...
 Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ... Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ...
Performance Evaluation of an Auger-fer Chipping Machine Adapter for cassava ...
 
Augur Press Kit
Augur Press KitAugur Press Kit
Augur Press Kit
 
VALUEADDITIONS 04-02-2016
VALUEADDITIONS 04-02-2016VALUEADDITIONS 04-02-2016
VALUEADDITIONS 04-02-2016
 
Aes
AesAes
Aes
 
Bentonite for construction
Bentonite for construction Bentonite for construction
Bentonite for construction
 
Tools using for piling
Tools using for pilingTools using for piling
Tools using for piling
 
Aperio® Intro 2016
Aperio® Intro 2016Aperio® Intro 2016
Aperio® Intro 2016
 
Shortage of Skilled Labour in Construction Industry of Sri Lanka
Shortage of Skilled Labour in Construction Industry of Sri LankaShortage of Skilled Labour in Construction Industry of Sri Lanka
Shortage of Skilled Labour in Construction Industry of Sri Lanka
 
How to deliver an extremely boring presentation
How to deliver an extremely boring presentationHow to deliver an extremely boring presentation
How to deliver an extremely boring presentation
 
An Excellent Boring Presentation
An Excellent Boring PresentationAn Excellent Boring Presentation
An Excellent Boring Presentation
 
PPT ON TRENCHLESS TECHNOLOGY
PPT ON TRENCHLESS TECHNOLOGY PPT ON TRENCHLESS TECHNOLOGY
PPT ON TRENCHLESS TECHNOLOGY
 

Similar to Laparoscopic Semi-Automatic Suturing System Overview Oct 16 2016 by Alex Tegzes

Report on "STRUCTURAL MECHANICAL PERFORMANCE EVALUATION AND HEALTH MONITORING"
Report on "STRUCTURAL MECHANICAL  PERFORMANCE EVALUATION AND  HEALTH MONITORING"Report on "STRUCTURAL MECHANICAL  PERFORMANCE EVALUATION AND  HEALTH MONITORING"
Report on "STRUCTURAL MECHANICAL PERFORMANCE EVALUATION AND HEALTH MONITORING"Abdul Majid
 
Study and simulation of gait cycle of a planar five link kanwar et al. 2015
Study and simulation of gait cycle of a planar five link kanwar et al. 2015Study and simulation of gait cycle of a planar five link kanwar et al. 2015
Study and simulation of gait cycle of a planar five link kanwar et al. 2015Nemish Kanwar
 
Optimization of an Energy-Generating Turnstile
Optimization of an Energy-Generating TurnstileOptimization of an Energy-Generating Turnstile
Optimization of an Energy-Generating TurnstileWayne Smith
 
Etx 90 w-autostar man
Etx 90 w-autostar manEtx 90 w-autostar man
Etx 90 w-autostar manlsd999
 
Solutions to apostol analysis problems vol 2
Solutions to apostol analysis problems vol 2Solutions to apostol analysis problems vol 2
Solutions to apostol analysis problems vol 2Carlos Fuentes
 
Online Electronic Shopping Project Report Final Year
Online Electronic Shopping Project Report Final YearOnline Electronic Shopping Project Report Final Year
Online Electronic Shopping Project Report Final YearAmit Verma
 
Fackrell_S_s1400430_CT4017_Report_Group_2
Fackrell_S_s1400430_CT4017_Report_Group_2Fackrell_S_s1400430_CT4017_Report_Group_2
Fackrell_S_s1400430_CT4017_Report_Group_2Sean Fackrell
 
Beej Guide Network Programming
Beej Guide Network ProgrammingBeej Guide Network Programming
Beej Guide Network ProgrammingSriram Raj
 
Wireshark user's guide
Wireshark user's guideWireshark user's guide
Wireshark user's guideGió Lào
 
Capstone Final Report
Capstone Final ReportCapstone Final Report
Capstone Final ReportVaibhav Menon
 
Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...
Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...
Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...SANTIAGO PABLO ALBERTO
 
T Series Core Router Architecture Review (Whitepaper)
T Series Core Router Architecture Review (Whitepaper)T Series Core Router Architecture Review (Whitepaper)
T Series Core Router Architecture Review (Whitepaper)Juniper Networks
 
Drive size users_manual
Drive size users_manualDrive size users_manual
Drive size users_manualPeruvian Navy
 
59582162 dpr
59582162 dpr59582162 dpr
59582162 dprablaze7
 
Orbiter 2010 manual.
Orbiter 2010 manual.Orbiter 2010 manual.
Orbiter 2010 manual.JacekKupras
 

Similar to Laparoscopic Semi-Automatic Suturing System Overview Oct 16 2016 by Alex Tegzes (20)

Report on "STRUCTURAL MECHANICAL PERFORMANCE EVALUATION AND HEALTH MONITORING"
Report on "STRUCTURAL MECHANICAL  PERFORMANCE EVALUATION AND  HEALTH MONITORING"Report on "STRUCTURAL MECHANICAL  PERFORMANCE EVALUATION AND  HEALTH MONITORING"
Report on "STRUCTURAL MECHANICAL PERFORMANCE EVALUATION AND HEALTH MONITORING"
 
Study and simulation of gait cycle of a planar five link kanwar et al. 2015
Study and simulation of gait cycle of a planar five link kanwar et al. 2015Study and simulation of gait cycle of a planar five link kanwar et al. 2015
Study and simulation of gait cycle of a planar five link kanwar et al. 2015
 
Optimization of an Energy-Generating Turnstile
Optimization of an Energy-Generating TurnstileOptimization of an Energy-Generating Turnstile
Optimization of an Energy-Generating Turnstile
 
Etx 90 w-autostar man
Etx 90 w-autostar manEtx 90 w-autostar man
Etx 90 w-autostar man
 
Solutions to apostol analysis problems vol 2
Solutions to apostol analysis problems vol 2Solutions to apostol analysis problems vol 2
Solutions to apostol analysis problems vol 2
 
Online Electronic Shopping Project Report Final Year
Online Electronic Shopping Project Report Final YearOnline Electronic Shopping Project Report Final Year
Online Electronic Shopping Project Report Final Year
 
MSc_Dissertation
MSc_DissertationMSc_Dissertation
MSc_Dissertation
 
Fackrell_S_s1400430_CT4017_Report_Group_2
Fackrell_S_s1400430_CT4017_Report_Group_2Fackrell_S_s1400430_CT4017_Report_Group_2
Fackrell_S_s1400430_CT4017_Report_Group_2
 
AuthorCopy
AuthorCopyAuthorCopy
AuthorCopy
 
AutoBox Rev B
AutoBox Rev BAutoBox Rev B
AutoBox Rev B
 
Beej Guide Network Programming
Beej Guide Network ProgrammingBeej Guide Network Programming
Beej Guide Network Programming
 
Feasibility report
Feasibility reportFeasibility report
Feasibility report
 
Lfa
LfaLfa
Lfa
 
Wireshark user's guide
Wireshark user's guideWireshark user's guide
Wireshark user's guide
 
Capstone Final Report
Capstone Final ReportCapstone Final Report
Capstone Final Report
 
Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...
Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...
Electrónica: Estudio de diseño y construcción de una reductora con cambio de ...
 
T Series Core Router Architecture Review (Whitepaper)
T Series Core Router Architecture Review (Whitepaper)T Series Core Router Architecture Review (Whitepaper)
T Series Core Router Architecture Review (Whitepaper)
 
Drive size users_manual
Drive size users_manualDrive size users_manual
Drive size users_manual
 
59582162 dpr
59582162 dpr59582162 dpr
59582162 dpr
 
Orbiter 2010 manual.
Orbiter 2010 manual.Orbiter 2010 manual.
Orbiter 2010 manual.
 

Laparoscopic Semi-Automatic Suturing System Overview Oct 16 2016 by Alex Tegzes

  • 1. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng Date: Oct 16 2016
  • 2. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 2 | P a g e Contents Table of Figures.............................................................................................................................................4 Scope.............................................................................................................................................................5 Problem to Solve...........................................................................................................................................5 Specifications and Assumptions................................................................................................................5 Suturing Method...........................................................................................................................................6 1. Deploy Stabilizing Augers..................................................................................................................6 2. Close the Wound with a Curved Suture Needle ...............................................................................6 3. Needle Tip Exchange Grasper ...........................................................................................................7 4. Needle Tip Grasper Retracts Curved Needle Tip ..............................................................................9 5. Curved Tube is Retracted..................................................................................................................9 Curved Tube Needle Handle Controls Overview ........................................................................................11 1. Trigger Interlock..............................................................................................................................11 2. Stainless Cable with Bicycle Style Pull Anchors...............................................................................12 Needle Tip Exchange Grasper Controls Overview ......................................................................................16 1. Needle Grasper Default State .........................................................................................................17 2. Needle Grasper Open State ............................................................................................................18 Auger Tissue Stabilizer Controls Overview .................................................................................................19 1. Deploying Tissue Augers .................................................................................................................19 2. Removing Augers ............................................................................................................................20 Manufacturability Overview .......................................................................................................................21 Curved Needle Hypodermic Tube...........................................................................................................21 Laparoscopic Suture System Main Body.................................................................................................22 Needle Tube Torque Spindle...................................................................................................................22 Curved Needle Tube and Torque Spindle Welded Assembly .................................................................23 Curved Needle Tip and Suture Assembly................................................................................................24 Grasping Needle Tip Collet......................................................................................................................24 Tissue Stabilizing Auger...........................................................................................................................27 Final Discussions .........................................................................................................................................28 Time Constrained Development Results.................................................................................................28 Delivered Therapy Considerations..........................................................................................................28
  • 3. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 3 | P a g e User Interface Considerations ................................................................................................................28 Manufacturability Considerations ..........................................................................................................28 Cleaning and Sterilization Considerations ..............................................................................................28 Design for Manufacturing Considerations..............................................................................................28
  • 4. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 4 | P a g e Table of Figures Figure 1 Wound with Access Section View...................................................................................................5 Figure 2 Tissue Stabilizing Augers Deployed.................................................................................................6 Figure 3 Needle Extended Through the Wound ...........................................................................................7 Figure 4 Needle Tip Exchange Grasper Extended.........................................................................................7 Figure 5 Needle Tip Exchange Grasper Extended Section View ...................................................................8 Figure 6 Needle Tip Exchange Grasper Contracted Section View ................................................................8 Figure 7 Needle Tip Exchange Grasper Retracts Curved Needle Tip Section View ......................................9 Figure 8 Suture System Retracting out of Access Port................................................................................10 Figure 9 Curved Needle Handle Safety Interlock Position Section View ....................................................11 Figure 10 Curved Needle Handle Safety Interlock Position Detailed Section View....................................11 Figure 11 Curved Needle Handle Safety Pin Depressed Section View........................................................12 Figure 12 Bicycle Style Cable with Pull Anchors..........................................................................................12 Figure 13 Curved Needle Handle Trigger to Intermediate Pulley Cable Path.............................................13 Figure 14 Curved Needle Intermediate Pulley to Actuation Pin Cable Path...............................................14 Figure 15 Bicycle Style Cable with Three Pull Anchors ...............................................................................14 Figure 16 Curved Tube Needle Actuation Pin.............................................................................................15 Figure 17 Complete Laparoscopic Suturing System....................................................................................16 Figure 18 Needle Grasper Spring Default State..........................................................................................17 Figure 19 Needle Grasper Collet in Closed State Section View ..................................................................17 Figure 20 Needle Grasper Spring Depressed State.....................................................................................18 Figure 21 Needle Grasper Collet in Open State Section View ....................................................................18 Figure 22 Tissue Auger Detail View.............................................................................................................19 Figure 23 Deploying Tissue Augers .............................................................................................................19 Figure 24 Auger Control Knob Detailed View .............................................................................................20 Figure 25 Auger Control Knob Bottomed Out ............................................................................................20 Figure 26 Curved Needle Hypodermic Tube...............................................................................................21 Figure 27 Laparoscopic Suture System Main Body.....................................................................................22 Figure 28 Needle Tube Torque Spindle.......................................................................................................22 Figure 29 Curved Needle to Torque Spindle Welded Assembly .................................................................23 Figure 30 Laser Star i960 Welding Station..................................................................................................23 Figure 31 Curved Needle Tip.......................................................................................................................24 Figure 32 Grasping Needle Tip Collet Step 1 Section View.........................................................................24 Figure 33 Grasping Needle Tip Collet Step 2...............................................................................................25 Figure 34 Grasping Needle Tip Collet Step 3...............................................................................................25 Figure 35 Needle Grasping Collet is Laser Welded to a Control Tube........................................................26 Figure 36 Tissue Auger................................................................................................................................27 Figure 37 Auger Forming Mandrel..............................................................................................................27
  • 5. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 5 | P a g e Scope The purpose of this document is to provide a technical design overview of a laparoscopic suturing system. The document is made in response to a technical problem presented by a client to develop a prototype suturing system. Problem to Solve Design a system that can send a single suture across an open wound with limited access as shown in Figure 1. Specifications and Assumptions  Wound size: 8mm x 3mm  Access size: 1”ID x 4” length, like a laparoscopic access point  Visual guidance is provided with a separate system, so assume good visibility of the wound is available  There are organs below the wound, so the suturing solution must not puncture too deep below the bottom of the wound  The solution will not include electronics, pneumatics or nittinol alloys Figure 1 Wound with Access Section View
  • 6. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 6 | P a g e Suturing Method In this section of the document the suturing method is reviewed by focusing on the therapy that is delivered to the wound site. The approach to the suturing system resembles how a surgeon would suture a wound using a curved needle. 1. Deploy Tissue Stabilizing Augers  Augers are deployed around the wound in order to stabilize the tissue before a curved needle can be used as shown in Figure 2  Note the position of the pink tip of the stowed curved tube needle Figure 2 Tissue Stabilizing Augers Deployed 2. Close the Wound with a Curved Suture Needle  A curved needle is driven into the tissue as shown Figure 3
  • 7. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 7 | P a g e Figure 3 Needle Extended Through the Wound 3. Needle Tip Exchange Grasper  The needle tip exchange grasper is extended over the tip of the needle as shown in Figure 4 and Figure 5  Note: the actual suture is not modeled in the images shown below Figure 4 Needle Tip Exchange Grasper Extended
  • 8. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 8 | P a g e Figure 5 Needle Tip Exchange Grasper Extended Section View  The collet of the needle tip exchange grasper is closed around the needle tip edge as shown in Figure 6 Figure 6 Needle Tip Exchange Grasper Contracted Section View
  • 9. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 9 | P a g e 4. Needle Tip Grasper Retracts Curved Needle Tip  The needle tip is pulled away from the curved tube with the collet as shown Figure 7.  Note the curved tube needle body is left in place to allow for easier suture routing around the wound site. Figure 7 Needle Tip Exchange Grasper Retracts Curved Needle Tip Section View 5. Curved Tube is Retracted  Once the suture is retrieved to the top of the access port, the curved tube body is retracted into nominal position.  The tissue stabilizing augers are retracted.  The suture system can now be retracted out of the access port as shown Figure 8.
  • 10. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 10 | P a g e Figure 8 Suture System Retracting out of Access Port
  • 11. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 11 | P a g e Curved Tube Needle Handle Controls Overview The curved needle handle controls were inspired by Medtronics LigaSure System, http://www.medtronic.com/covidien/products/vessel-sealing/ligasure-5-mm-blunt-tip-37-cm. The curved tube needle handle is intended to be operated by the left hand. 1. Trigger Interlock  To minimize accidental deployment of the curved needle a trigger safety interlock is used as shown in Figure 9 and Figure 10. Figure 9 Curved Needle Handle Safety Interlock Position Section View Figure 10 Curved Needle Handle Safety Interlock Position Detailed Section View
  • 12. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 12 | P a g e  In order to actuate the trigger, a push button must be depressed in order to disable the safety pin as shown Figure 11.  Note the safety pin is sprung such that when the trigger is retracted to normal position the safety pin automatically deploys.  Note the disarm push button is also sprung so that the user has to actively hold down the safety over-ride button with their thumb in order to actuate the curved tube needle. Figure 11 Curved Needle Handle Safety Pin Depressed Section View 2. Stainless Cable with Bicycle Style Pull Anchors  All curved tube needle actuations are performed using a set of multi-strand stainless steel cables with bicycle style anchors as shown in Figure 12. Figure 12 Bicycle Style Cable with Pull Anchors
  • 13. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 13 | P a g e  The trigger pulley rotates the intermediate pulley by a cable (highlighted blue cable path shown in Figure 13).  A secondary cable (also highlighted in blue Figure 14) is used to move the curved needle actuation pin as shown in Figure 16. Figure 13 Curved Needle Handle Trigger to Intermediate Pulley Cable Path  The curved needle is actuated by a second cable that has three pull anchors attached to it as shown in Figure 14 and Figure 15.
  • 14. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 14 | P a g e Figure 14 Curved Needle Intermediate Pulley to Actuation Pin Cable Path  Note that the pull cable shown in Figure 15 has a hole in the center of the middle anchor. The hole in the center of the anchor is used to actuate the curved needle pin shown in Figure 16. Figure 15 Bicycle Style Cable with Three Pull Anchors
  • 15. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 15 | P a g e  The green pulleys shown in Figure 16 are used to reduce the friction in the mechanism as the cables go around sharp corners. Figure 16 Curved Tube Needle Actuation Pin
  • 16. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 16 | P a g e Needle Tip Exchange Grasper Controls Overview The user controls for the needle grasper are very similar to a tri-finger ring syringe controls, like Qosina part number: C1068, http://www.qosina.com/control-syringe-with-solid-ring-plunger-and-grip-with- male-luer-lock-c1068 The user is intended to operate the needle tip exchange grasper with their right hand while simultaneously holding the needle handle with their left hand as shown in Figure 17. Figure 17 Complete Laparoscopic Suturing System
  • 17. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 17 | P a g e 1. Needle Grasper Default State  A spring shown in Figure 18 is used to maintain the grasper collet in a default closed state as shown in Figure 19 Figure 18 Needle Grasper Spring Default State Figure 19 Needle Grasper Collet in Closed State Section View
  • 18. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 18 | P a g e 2. Needle Grasper Open State  By depressing the spring on the needle grasper handle as shown in Figure 20, relative movement of the tubes allows the grasper collet to open as shown in Figure 21.  Note the grasping collet is manufactured such that its default state is open and only by pulling it into a tube it becomes closed. Manufacturability of the grasping collet is reviewed in subsequent sections of this document. Figure 20 Needle Grasper Spring Depressed State Figure 21 Needle Grasper Collet in Open State Section View
  • 19. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 19 | P a g e Auger Tissue Stabilizer Controls Overview Control method for augers is inspired by various endoscopic procedures. Figure 22 Tissue Auger Detail View 1. Deploying Tissue Augers  Tissue augers are deployed by rotating clockwise the auger control knobs with your right hand as shown in Figure 23 and Figure 24. Figure 23 Deploying Tissue Augers
  • 20. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 20 | P a g e Figure 24 Auger Control Knob Detailed View  Tissue augers are fully deployed once the tissue control knobs bottom out as shown in Figure 25. Figure 25 Auger Control Knob Bottomed Out 2. Removing Augers  Auger removal is done by rotating the auger control knobs counterclockwise.  Once the augers are fully unwound from the tissue they are safe to pull out of the handle.
  • 21. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 21 | P a g e Manufacturability Overview The laparoscopic suturing system is currently modeled as an easily machinable prototype. The system is modeled as a CNC machined unit because the first step in validating the concepts shown in this document is to construct a functioning prototype. In this section of the document the manufacturability of only the non-trivial components will be reviewed. Curved Needle Hypodermic Tube The curved hypodermic needle tube shown in Figure 26, is made from 14 Gauge thin wall stainless steel tube stock. In order to create this round shape without crushing the walls of the tube a round mandrel must be installed inside of the tube prior to bending. Popular mandrel choices are as follows:  Nittinol wire  Stainless steel wire with a Teflon sleeve  Music wire Figure 26 Curved Needle Hypodermic Tube Once the mandrel is installed inside of the tube, the tube may be bent around a piece of round bar with the required diameter to form this shape. Once the shape is formed, the mandrel is removed.
  • 22. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 22 | P a g e Laparoscopic Suture System Main Body This profile is created using a custom aluminum extrusion. The mounting holes are added in a post process CNC milling operation. This component can also be modified so that only a CNC milling operation is required, but it will require a few more components. Figure 27 Laparoscopic Suture System Main Body Needle Tube Torque Spindle The needle tube torque spindle shown in Figure 28 is created from stainless steel round stock using a CNC lathe to create a round shape that looks like a spin top. The spin top shape is then installed into a CNC mill and the final profile outline is made. Figure 28 Needle Tube Torque Spindle
  • 23. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 23 | P a g e Curved Needle Tube and Torque Spindle Welded Assembly Figure 29 Curved Needle to Torque Spindle Welded Assembly The curved needle tube and torque spindle are welded together at the location illustrated in Figure 29 using a laser welder. Laser Star i960 portable laser weld station shown in Figure 30 would perform well in this type of application. Figure 30 Laser Star i960 Welding Station
  • 24. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 24 | P a g e Curved Needle Tip and Suture Assembly The curved needle tip is created using a CNC lathe as shown Figure 31. A suture is inserted and crimped into place using a compression collet tool. Figure 31 Curved Needle Tip Grasping Needle Tip Collet The grasping needle tip collet is manufactured from 17-4 PH Stainless Steel. 17-4 PH stainless steel is a good material for this application because it is corrosion resistant, has a high yield stress(good for flexures), and because it can be heat treated after it is machined and formed into to a desired shape. This is a common method of creating a geometry that is not easily manufactured using conventional machining methods.  Collet is first machined using a CNC lathe as shown in Figure 32 Figure 32 Grasping Needle Tip Collet Step 1 Section View
  • 25. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 25 | P a g e  Collet is then laser cut to obtain thin slits as shown in Figure 33. Figure 33 Grasping Needle Tip Collet Step 2  The collet is then formed and heat treated at 1150degF for a few hours and quenched in still air to take on the final shape and hardness as shown in Figure 34. Figure 34 Grasping Needle Tip Collet Step 3 The grasping needle collet is welded to the tube using a laser welder as shown in Figure 35.
  • 26. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 26 | P a g e Figure 35 Needle Grasping Collet is Laser Welded to a Control Tube
  • 27. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 27 | P a g e Tissue Stabilizing Auger Tissue augers as shown in Figure 36 are created by bending wire around a round mandrel as shown in Figure 37. Figure 36 Tissue Auger Figure 37 Auger Forming Mandrel
  • 28. Laparoscopic Suturing System Industrial Design Overview Prepared by: Aleksandar Tegzes, M.Eng, P.Eng 28 | P a g e Final Discussions Time Constrained Development Results The solution presented in this document provides a very complete solution with detailed implementation notes of each the systems used. Keeping in mind that this solution took less than 24hrs to develop and document, this is not an unreasonable first step towards a final product. Delivered Therapy Considerations In the current implementation thin wire (0.5mm diameter) augers are used to stabilize tissue before a curved needle is driven through the wound. The cost/benefit consideration here is that the augers are not introducing a significant wound to the area, but they are minimizing tissue movement which greatly improves accuracy of needle deployment. There are many endoscopic procedures that use similar size augers to perform wound closing operations, like OverStich product by Apollo Endosurgery: http://apolloendo.com/overstitch/. User Interface Considerations The document presents preliminary user interface control considerations and assumptions. The primary curved needle operating handle is intended to be held by the left hand and auxiliary operations that demand slightly higher hand dexterity are operated by the right hand. Bench top testing of early prototypes would be valuable proving ground to validate the assumption of this arrangement. Manufacturability Considerations Manufacturing solutions to non-trivial components of this assembly are also presented in this document. The complexity of all of the non-trivial components in this assembly is fairly low. Assuming a laser welder, a CNC lathe and a CNC mill are available for prototyping, a well-trained hands-on engineer can program and fabricate all of the parts needed to construct the first prototype within 7-10 working days. Cleaning and Sterilization Considerations Currently all of the components in the prototype are intended to be built from stainless steel or aluminum. Material selection of the production components will vary depending on how the product is intended to be marketed to the users, as either a single use or a many use sterilizable tool. Design for Manufacturing Considerations Once the first prototypes are built and their functionality is verified, more time could be afforded designing more ergonomic user interface handles. Many parts can also be made from injection molded plastic. More time could also be spent removing visible screws from the assembly; this can easily be accomplished with thin plastic snap in place exterior panels.