HERFST NOVEMBER 2013
Dr Guy odent
Implantable Electronic Cardiac Devices Historical Aspects

Senning and
Elmquist
1st implant of an
electronic PM

Mirowski

Furman

Development of
the 1st ICD –
implant in dogs

1st endocardiac
PM

1932
Hyman

1958

1964

1970

1st report of CRT

1980’s

1994

RECENTLY
Heart Failure
control

Home Monitoring
DEVICES

• Pacemaker
• Defibrillator
• Remote monitoring
• Biventriculaire pacing of
resynchronisatietherapie
TELECARDIOLOGIE
PACEMAKER

AICD BIVENT PM

MONITORING DEVICE

• Remote follow-up : opvragen van device gebeurt vanop afstand, in
communicatie met zorgverlener
• Remote monitoring : op regelmatig basis wordt gegevens
doorgestuurd naar ziekenhuis . Alarmen kunnen ingesteld worden
• Continuous monitoring : huidige systemen sturen geen gegevens
door op continue basis en gegevens worden ook niet op continue
basis gelezen
DEVICES

• Pacemaker
• Defibrillator
• Remote monitoring
• Biventriculaire pacing of
resynchronisatietherapie
TAAK VAN EEN PACEMAKER : PACING
Stimulatie van het hart betekent
• output van een elektrische stimulatie impuls

... Met voldoende energie
• via een elektrode

Resulterend in een depolarisatie van het myocardium en contractie van het hart.
PACEMAKER PULSE

Pulseamplitude
(Volt)

Pulse width
(ms)
8
TAAK VAN EEN PACEMAKER : SENSING
Detectie of Sensing…
...is de mogelijkheid van een pacemaker om het eigen hartritme (signalen van het hart) te detecteren
(sensen)

via de geconnecteerde elektrode
PULSE GENERATOR

10
Feed Through

Defib. protection

Capacitors

Sensor
Crystal
Telemetry Coil
X-ray marker
11
LEAD CONNECTORS

12
IS-1 CONNECTOR IN THE HEADER

13
TIGHTENING THE SET SCREW

14
LEADS
Atriaal ---- ventriculair
Endocavitair ----- epicardiaal
Unipolair ----- bipolair

15
UNIPOLAIRE CONFIGURATIE

● Grote spikes op EKG
● Grote antenne – kans op storing

IPG

I
BIPOLAIRE CONFIGURATIE
•

Kleine spikes op EKG

•

Kleine antenne – minder storing

•

Locale detectie

IPG

I
POLARITY CONFIGURATION

Bi-polar

Uni-polar
18
UNI-POLAR OR BI-POLAR LEADS?
Advantages Bi-polar

Advantages Uni-polar
• Smaller diameter
• (Larger “spikes” on ECG)

• Less risk of myopotential oversensing
• Less risk of musclestim.

• Less risk of Crosstalk and Far Field
Sensing

With a bipolar lead, you can program the device
to e.g. stimulate unipolar but sense bipolar*.
* In Microny, polarity is not programmable

19
Far-field sensing
Sensing intrinsic
activity in the
other chamber

Crosstalk
Sensing pacemakeractivity in the other chamber
20
FIXATION METHODS

Passive

Active
21
PASSIVE AND ACTIVE FIXATION

22
J-shaped lead
(1421T)
MYOCARDIAL LEAD : EPICARDIALE LEAD
Door chirurgen tijdens operatie te plaatsen

24
1 KAMER PACING (VVI)
• Een elektrode in het ventrikel

• Bijgevolg enkel stimulatie en
• detectie in het ventrikel

• Permanente vkf
• Technisch niet mogelijk
Tweede lead te plaatsen
2-KAMER PACING
• Twee elektroden in het hart,

• Één in het atrium en één in het ventrikel
• Bijgevolg stimulatie en
• detectie in het atrium en het ventrikel
PACEMAKER IMPLANT TECHNIQUE - OVERVIEW
surgeons / cardiologists /
electrophysiologists

• Implanting physicians:

• Implant facilities:

Operating room / Cardiac Catheter lab / EP

lab
• Equipment
• Lead Placement
• Threshold measurements
• Pocket Closure
POCKET CREATION

Photos courtesy of Malcolm Clarke, M.D.
TRANSVENOUS ACCESS – CHOICE OF VEINS
•Left Internal Jugular
•Left External Jugular
•Left Subclavian
•Left Axillary
•Left Cephalic
STEPS IN SURGERY
• Opening of SC tissue and formation of pacemaker
pocket by blunt dissection
• Venous access
• Lead Placement

• Pacemaker Connection to Lead(s)
• Pocket Closure
PUNCTURE TECHNIQUE

5 - Insertion of the electrode
6 - Removal of the introducer
VENOUS ACCESS

Veins:
Jugular (exteral, internal)

Subclavian
Cephalic
PUNCTURE TECHNIQUE FOR SINGLE LEAD INSERTION

1 - Puncture, 2 - Insertion of the Seldinger wire, 3 - Introducer with dilator,
4 - Removal of dilator and Seldinger wire, insertion of the electrode
PULSE GENERATOR IMPLANTATION SITES

Left
Right
Abdominal
PLACEMENT OF THE ATRIAL ELECTRODE

Placement of the atrial electrode at the lateral
anterior wall of the right atrium
PLACEMENT OF THE VENTRICULAR ELECTRODE

Placement via the right outflow tract respectively the pulmonary artery
LEAD PLACEMENT
•

Ventricular Lead

• Right Ventricular Apex (RVA) or Right Ventricular
Outflow Tract (RVOT)
- Ventricular Bradycardia Pacing
- Sensing Intrinsic Rhythm
•

Atrial Lead

• Right Atrial Appendage or Atrial Septal Wall
- Atrial Pacing

- Atrial Sensing
VENTRICULAR LEAD PLACEMENT
VENTRICULAR LEAD - RV APEX PA
& LATERAL CHEST X-RAY
PACEMAKER POCKET
Skin
Pacemaker

Pectoralis muscle

Subcutaneous
tissue

Fascia
CONNECTION TO PACEMAKER
CLOSING THE INCISION
POTENTIAL PEROPERATIVE COMPLICATIONS
• Bleeding
• Pneumothroax
• Difficult anatomy
• Difficulties in positioning the electrode(s)
• Arrhythmias
VVI PACING : 1 LEAD IN RECHTER VENTRIKEL
2-KAMER PACING (DDD)
WAT KAN EN WAT MAG?
•

De patiënt kan normaal veel van zijn vroegere activiteiten hervatten.

•

Vermijd echter, of bespreek met cardioloog:

• Competitie sport
• ruwe contactsporten zoals rugby.

• Activiteiten met veel schokken zoals paardrijden of botsauto’s
• Inspannende lichaamsoefeningen met uw bovenlichaam zoals gewichtheffen
• Rugzakken met dunne draagriemen, beste is rugzak met riem om middel ter
ondersteuning
ZAL IK HET VOELEN ALS MIJN PACEMAKER WERKT?
•

De meeste mensen voelen dit niet.

• Vooral overgang tussen intrinsiek en gestimuleerd
• Korte PR intervallen (voelt aan als kramp)
• Let op bij magneet oplegging
• Hogere hartritme tijdens rust/slaap
• PAC en PVC (overslag)
•

De geleverde elektrische energie is zeer klein.

• Let op voor spierstimulatie
• Let op voor diafragma stimulatie
•

Indien u iets voelt, kan u zich best wenden tot uw cardioloog, die kan de waarneming bespreken
en eventueel gepaste actie ondernemen.
Post OP complicaties
●
●
●
●

Bloeding
Veneuze trombose
Infectie
Erosie pacemaker
Wat gebeurt er tijdens een pacemakercontrole?
●
●
●
●

Controle 6 of 12 maandelijks
Anamnese
ECG
Ondervragen pacemaker
○ Bekijken programmatie
○ Bekijken diagnostische gegevens
○ Meten: intracardiale signalen, stimulatiedrempel,
impedantie, batterij status
● Bijkomende onderzoeken op basis van de
bevindingen
• Continue ventriculaire pacing in rechter ventikel is
schadelijk op lange termijn ( hartfalen ) : Algoritmes
op ventr . Pacing te beperken
ELECTIVE REPLACEMENT INDICATOR

Magnet
test rate

•Dialog/Sensolog

100

•Multilog/Sensorithm
•Microny
•Regency

85

•Affinity
•Entity
•Integrity

•Identity

60

•ADx Family
2.8
BOL

2.4
ERI

2.0 Battery voltage V
EOL
53
ELECTRICAL RECORDINGS
Surface
ECG
Cellpotential

IEGM

54
SENSITIVITY, ATRIAL CHANNEL

P-wave
~3 mV

Atrial fibrillation
~0.2-1.0 mV

Noise
~0.1 mV

Sensitivity
0.3 mV
0.1

Sensitivity
1.5 mV

0.2

0.3

0.4

0.5 1.0
0.75

1.5

2.0

3.0

4.0

55
CLASSIC THRESHOLD DEVELOPMENT OVER
TIME
Threshold
(Volts)

Output
Threshold

2 weeks

2-3 months

Time

Implantation
56
DE MAGNEET MODUS VAN EEN PACEMAKER
Schakelt de sensing uit: VOO DOO of AOO Modus
en Stimuleert aan een vastgesteld ritme
Terminologie:
BOL = Beginning of life (meestal > 90 ppm)
EOL = End of life
ERT = Elective replacement time
ERI = Elective replacement indicator: indication ERT is reached
Let op: Kijk naar EKG bij magneet oplegging!
Gebruik: Freq. = indicatie voor batt. Conditie
Controle van de efficiëntie van de spikes.
MAAR VANDAAG…

•

De beste programmer!
AANDACHTSPUNTEN OP HET WERK.
•

Een normaal leven lijden

•

Als de patiënt op het werk in contact komt met zware elektrische apparatuur kan die
elektromagnetische interferentie veroorzaken die de werking van de pacemaker
beïnvloed. Wees voorzichtig met:

•
•
•
•
•

Elektrische apparatuur in slechte staat
Elektrische apparatuur die veel energie produceert zoals generatoren
Booglasapparatuur
medische apparatuur waaronder MRI, therapeutische straling, …
metaaldetectoren en beveiligingssystemen in winkels en op vliegvelden
•

AANDACHTSPUNTENinIN HET zoals magnetrons, mixers, broodroosters,
DAGELIJKS LEVEN.
De meeste huishoudelijke apparaten goede staat
televisies, videorecorders, elektrische dekens, kacheltjes, openingssystemen voor
garagedeuren,…

•

Kantoorapparatuur en de meeste medische apparaten zoals ultrasoon, CT scan,
mammografieën, fluoroscopie

•

GSM gebruik kan een pacemaker beïnvloeden:

• Regel: 10 centimeter van pacemaker blijven.

• St. Jude Medical pacemakers zijn GSM-proof!
MAG IK AUTORIJDEN?
• De kandidaat met een ingeplante pacemaker is niet rijgeschikt tijdens de
maand die volgt op de implanting van de pacemaker of het vervangen van de
pacemakerelektrode.
• Bij het vervangen van enkel de pacemaker kan de kandidaat door de
behandelende cardioloog onmiddellijk rijgeschikt worden bevonden
PACEMAKERSYNDROOM
EENKAMERSYSTEEM ----- SINUSAAL RITME
•

Verlies van AV synchroniciteit door eenkamerstimatulatie in ventrikel ( vvi )

•

Verminderde inspanningscapaciteit door verminderde cardiac output

•

Atriale contractie tegen gesloten atrioventriculaire klep

•

Subjectief : duizeligheid, vermoeidheid, host, dyspnoe, orthopnoe, eigenaardig gevoel in
de keel
DEVICES

• Pacemaker

• Defibrillator
• Remote monitoring

• Biventriculaire pacing of
resynchronisatietherapie
WAT DOET EEN ICD?
Maligne ventriculaire aritmieën
dectecteren en behandelen
MONOMORPHIC VT
Animation

*

EKG Characteristics

• Origin:

Ventricles (Single Focus)

• Mechanism:

Reentry Initiated by abnormal

Automaticity or Triggered activity
• Characteristics:

Rapid, wide, and regular QRS
POLYMORPHIC VT
Animation

*

• Origin:

Ventricles (Wandering Single Focus)

• Mechanism:

Reentry with movement in the circui
Initiated by Abnormal Automaticity o
Triggered activity
Wide and irregular QRS Complex that
changes in axis

• Characteristics:
VENTRICULAR FIBRILLATION (VF)
Animation

*

• Origin:

Ventricle

• Mechanism:

Multiple Wavelets of reentry

• Characteristics: Irregular with no discrete QRS
ICD EVOLUTION
ICD EVOLUTION
Evolution of ICD Therapy and Adoption: 1980 to Present

Number of Worldwide ICD Implants Per Year
(Ref: Corporate Market Share Database)
THE ICD SYSTEM

ICD System

=

+

+

Medtronic products shown
TRANSVENOUS LEADS

Placement

Single Coil

Dual Coil
ICD

How High Voltage Is Created

Pacing
Voltages

For Pacing

Battery
3.2
Volts

For Defibrillation

Voltage
Multiplier
and
Transformer

High
Voltage
Capacitor

High Energy
Shocks
(Up to 800 Volts)
AICD INDICATIES
Cardioverter Defibrillator
Number Needed to Treat

NNTx years = 100 / (% Mortality in Control Group – % Mortality in Treatment Group)

Drug Therapy

amiodarone

ICD Therapy
simvastatin

Metoprolol
succinate
captopril

(5 Yr)

(2.4 Yr)

(3 Yr)

(3 Yr)

(3.5 Yr)

(1 Yr)

(6 Yr)

(2 Yr)
SHOCK
Wat gebeurt er tijdens de controle van een
defibrillator?
●
●
●
●

Controle 6 maandelijks
Anamnese
ECG
Ondervragen defibrillator
○ Bekijken programmatie
○ Bekijken diagnostische gegevens
○ Meten: intracardiale signalen, stimulatiedrempel,
impedantie, batterij status
● Bijkomende onderzoeken op basis van de
bevindingen
Ondervragen defibrillator
● De huidige defib’s kunnen al heel wat tests zelf
gedaan hebben en kunnen eventuele problemen zelf
vaststellen.
Alerts
● Heeft de patiënt nog een eigen ritme? sensingtest
● Zijn de weerstanden op de elektroden stabiel?
● Wat is de minimumenergie nodig om te stimuleren?
Drempeltest. (atriaal en ventriculair)
● Is er nog voldoende energie in de batterij tot de
volgende controle? Resterende levensduur.
DE MAGNEET MODUS VAN EEN DEFIBRILLATOR
Schakelt de TACHY-therapie uit zolang de magneet
op Defibrillator ligt !
Wordt wel eens gebruikt tijdens een ingreep of
ingeval de defibrillator shockt als het niet moet.
Is geen follow-up-middel!
AANDACHTSPUNTEN: MRI MET
DEFIBRILLATOREN IS NIET MOGELIJK!
AANDACHTSPUNTEN OP HET WERK.
•

Een normaal leven lijden

•

Als de patiënt op het werk in contact komt met zware elektrische apparatuur kan die
elektromagnetische interferentie veroorzaken die de werking van de pacemaker
beïnvloed. Wees voorzichtig met:

•
•
•
•
•

Elektrische apparatuur in slechte staat
Elektrische apparatuur die veel energie produceert zoals generatoren
Booglasapparatuur
medische apparatuur waaronder MRI, therapeutische straling, …
metaaldetectoren en beveiligingssystemen in winkels en op vliegvelden
AANDACHTSPUNTEN IN HET DAGELIJKS LEVEN.
•

De meeste huishoudelijke apparaten in goede staat zoals magnetrons, mixers, broodroosters,
televisies, videorecorders, elektrische dekens, kacheltjes, openingssystemen voor
garagedeuren,…

•

Kantoorapparatuur en de meeste medische apparaten zoals ultrasoon, CT scan,
mammografieën, fluoroscopie

•

GSM gebruik kan een pacemaker beïnvloeden:

• Regel: 10 centimeter van pacemaker blijven.

• St. Jude Medical pacemakers zijn GSM-proof!
IMPLICATIES OP RIJBEWIJS
De kandidaat met een ingeplante automatische defibrillator is NIET
rijgeschikt.
• De kandidaat kan evenwel na een periode van één of drie maand, te rekenen
vanaf de implanting, rijgeschikt worden bevonden op basis van een recent
verslag, afgeleverd door de cardioloog van het geneeskundig centrum dat
de ingreep heeft uitgevoerd. De een maand is enkel van toepassing in het
geval de patiënt geen cardiac arrest heeft doorgemaakt. Tijdens deze
periode van één of drie maand en mag er geen stroomstoot geweest zijn die
het hartritme kan aantasten, indien wel is 3 maand terug van toepassing
BEEPING
•

ACID kan beepen

•

- lead problemen

•

- batterijprobemen

•

- na interventie
DEVICES

• Pacemaker
• Defibrillator

• Remote monitoring
• Biventriculaire pacing of
resynchronisatietherapie
HEART MONITORING OPTIONS
SYNCOPE OCCURS INFREQUENTLY
LONG-TERM MONITORING IS NEEDED

12-Lead 10 Seconds
Holter
ELR

2 - 3Days
7 Days- 30 days
3 Days to 30+ Days

MCOT ELR

Up to 36 Months

ILR

ILR = insertable loop recorder,
ELR=External loop recorder

MCOT= mobile cardiac outpatient telemetry
EXTERNAL ‘LOOP’ RECORDERS (ELRS)

7-day ‘Patch’
ELR

Up to 1 month recording
INSERTABLE CARDIAC MONITORS (ICM)
Reveal® System,
Medtronic Inc.
Minneapolis, MN
-manual/auto trigger
-remote download (CareLink®)

Confirm®,
St Jude Medical
St Paul, MN

-manual/auto trigger
-remote download
(Merlin®)
SYMPTOM-RHYTHM CORRELATION:
ICM PERMITS BOTH PATIENT-TRIGGERED AND AUTOMATIC
ACTIVATION

Auto
Activation
Point
Patient
Activation
Point

Medtronic data on file
INSERTABLE CARDIAC MONITOR (ICM)

• Subcutaneous ECG monitor
• 3-year battery

• Records, stores, and may transmit remotely
• Patient-activated and/or automatically-activated
• Longevity of current ILRs up to 36 months

• Indications
1. Patients with clinical syndromes or situations at increased risk for cardiac
arrhythmias
2. Patients who experience transient symptoms such as dizziness, palpitation,
syncope and chest pain, that may suggest a cardiac arrhythmia

• palpitations, cryptogenic stroke, ‘refractory’ epilepsy

Medtronic, Inc. Reveal® XT 9529 Insertable Cardiac Monitor Clinician Manual. 2011.
Remote Monitoring System

Patient

Transmitter

* By landline phone or GSM network

Wireless
Service center Data sent to
transmission*
the physician
REVEAL INDICATIES
• Onuitgelegde syncopes

• Vermoeden van ernstige ritmestoornissen

• 1 keer per week of na een event worden gegevens
doorgestuurd door patiënt
DEVICES

• Pacemaker

• Defibrillator
• Remote monitoring

• Biventriculaire pacing of
resynchronisatietherapie
VLBTB : QRS DUUR >120 MSEC, RSR’
PATROON LINKERPRECORDIALEN EN I EN AVL
CRT PITFALLS
SELECTIE PATIËNTEN
•

Responders – non responders

•

Technical challenging

•

- succes rate 93%

•

- complicaties 4 – 5% ( overlijden 0,3% ) : vene dissectie, perforatie, tamponnade,
pneumothorax

•

- duur ingreep ----- infectierisico
MRI COMPATIBILITEIT
• Pacemaker, aicd, biventr pacemaker : niet mri
compatibel
• Laatste generatie PM soms wel : steeds navragen (
toestel dient geherprogrammeerd te worden , bepaalde
zones van lichaam, bepaalde MRI toestellen )
• Toestellen voor remote monitoring : in principe geen
probleem
Cardiale
patiënt
vliegtuigreis
VLIEGTUIGREIZEN EN DEVICES
•

PM en AICD kunnen onderhevig zijn aan electromagnetische interferentie

•

Airport security :

- gates : induceren magnetische veld. Erdoor wandelen geen probleem. Alarm kan wel
afgaan. Stilstaan erin wordt afgeraden
- hand-held metal detector : ICD te vermijden
•

Electromagn. Interferentie tijdens vlucht is te verwaarlozen.
BIVENTRICULAIRE PACING : INDICATIES
CRT
BIVENTRICULAIRE PACING : INDICATIES
PACING INDICATIONS
Herfst november 2013finalpdf

More Related Content

Featured

How Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental HealthHow Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental Health
ThinkNow
 
Social Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie InsightsSocial Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie Insights
Kurio // The Social Media Age(ncy)
 

Featured (20)

2024 State of Marketing Report – by Hubspot
2024 State of Marketing Report – by Hubspot2024 State of Marketing Report – by Hubspot
2024 State of Marketing Report – by Hubspot
 
Everything You Need To Know About ChatGPT
Everything You Need To Know About ChatGPTEverything You Need To Know About ChatGPT
Everything You Need To Know About ChatGPT
 
Product Design Trends in 2024 | Teenage Engineerings
Product Design Trends in 2024 | Teenage EngineeringsProduct Design Trends in 2024 | Teenage Engineerings
Product Design Trends in 2024 | Teenage Engineerings
 
How Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental HealthHow Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental Health
 
AI Trends in Creative Operations 2024 by Artwork Flow.pdf
AI Trends in Creative Operations 2024 by Artwork Flow.pdfAI Trends in Creative Operations 2024 by Artwork Flow.pdf
AI Trends in Creative Operations 2024 by Artwork Flow.pdf
 
Skeleton Culture Code
Skeleton Culture CodeSkeleton Culture Code
Skeleton Culture Code
 
PEPSICO Presentation to CAGNY Conference Feb 2024
PEPSICO Presentation to CAGNY Conference Feb 2024PEPSICO Presentation to CAGNY Conference Feb 2024
PEPSICO Presentation to CAGNY Conference Feb 2024
 
Content Methodology: A Best Practices Report (Webinar)
Content Methodology: A Best Practices Report (Webinar)Content Methodology: A Best Practices Report (Webinar)
Content Methodology: A Best Practices Report (Webinar)
 
How to Prepare For a Successful Job Search for 2024
How to Prepare For a Successful Job Search for 2024How to Prepare For a Successful Job Search for 2024
How to Prepare For a Successful Job Search for 2024
 
Social Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie InsightsSocial Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie Insights
 
Trends In Paid Search: Navigating The Digital Landscape In 2024
Trends In Paid Search: Navigating The Digital Landscape In 2024Trends In Paid Search: Navigating The Digital Landscape In 2024
Trends In Paid Search: Navigating The Digital Landscape In 2024
 
5 Public speaking tips from TED - Visualized summary
5 Public speaking tips from TED - Visualized summary5 Public speaking tips from TED - Visualized summary
5 Public speaking tips from TED - Visualized summary
 
ChatGPT and the Future of Work - Clark Boyd
ChatGPT and the Future of Work - Clark Boyd ChatGPT and the Future of Work - Clark Boyd
ChatGPT and the Future of Work - Clark Boyd
 
Getting into the tech field. what next
Getting into the tech field. what next Getting into the tech field. what next
Getting into the tech field. what next
 
Google's Just Not That Into You: Understanding Core Updates & Search Intent
Google's Just Not That Into You: Understanding Core Updates & Search IntentGoogle's Just Not That Into You: Understanding Core Updates & Search Intent
Google's Just Not That Into You: Understanding Core Updates & Search Intent
 
How to have difficult conversations
How to have difficult conversations How to have difficult conversations
How to have difficult conversations
 
Introduction to Data Science
Introduction to Data ScienceIntroduction to Data Science
Introduction to Data Science
 
Time Management & Productivity - Best Practices
Time Management & Productivity -  Best PracticesTime Management & Productivity -  Best Practices
Time Management & Productivity - Best Practices
 
The six step guide to practical project management
The six step guide to practical project managementThe six step guide to practical project management
The six step guide to practical project management
 
Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...
Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...
Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...
 

Herfst november 2013finalpdf

  • 2. Implantable Electronic Cardiac Devices Historical Aspects Senning and Elmquist 1st implant of an electronic PM Mirowski Furman Development of the 1st ICD – implant in dogs 1st endocardiac PM 1932 Hyman 1958 1964 1970 1st report of CRT 1980’s 1994 RECENTLY Heart Failure control Home Monitoring
  • 3. DEVICES • Pacemaker • Defibrillator • Remote monitoring • Biventriculaire pacing of resynchronisatietherapie
  • 4.
  • 5. TELECARDIOLOGIE PACEMAKER AICD BIVENT PM MONITORING DEVICE • Remote follow-up : opvragen van device gebeurt vanop afstand, in communicatie met zorgverlener • Remote monitoring : op regelmatig basis wordt gegevens doorgestuurd naar ziekenhuis . Alarmen kunnen ingesteld worden • Continuous monitoring : huidige systemen sturen geen gegevens door op continue basis en gegevens worden ook niet op continue basis gelezen
  • 6. DEVICES • Pacemaker • Defibrillator • Remote monitoring • Biventriculaire pacing of resynchronisatietherapie
  • 7. TAAK VAN EEN PACEMAKER : PACING Stimulatie van het hart betekent • output van een elektrische stimulatie impuls ... Met voldoende energie • via een elektrode Resulterend in een depolarisatie van het myocardium en contractie van het hart.
  • 9. TAAK VAN EEN PACEMAKER : SENSING Detectie of Sensing… ...is de mogelijkheid van een pacemaker om het eigen hartritme (signalen van het hart) te detecteren (sensen) via de geconnecteerde elektrode
  • 13. IS-1 CONNECTOR IN THE HEADER 13
  • 14. TIGHTENING THE SET SCREW 14
  • 15. LEADS Atriaal ---- ventriculair Endocavitair ----- epicardiaal Unipolair ----- bipolair 15
  • 16. UNIPOLAIRE CONFIGURATIE ● Grote spikes op EKG ● Grote antenne – kans op storing IPG I
  • 17. BIPOLAIRE CONFIGURATIE • Kleine spikes op EKG • Kleine antenne – minder storing • Locale detectie IPG I
  • 19. UNI-POLAR OR BI-POLAR LEADS? Advantages Bi-polar Advantages Uni-polar • Smaller diameter • (Larger “spikes” on ECG) • Less risk of myopotential oversensing • Less risk of musclestim. • Less risk of Crosstalk and Far Field Sensing With a bipolar lead, you can program the device to e.g. stimulate unipolar but sense bipolar*. * In Microny, polarity is not programmable 19
  • 20. Far-field sensing Sensing intrinsic activity in the other chamber Crosstalk Sensing pacemakeractivity in the other chamber 20
  • 22. PASSIVE AND ACTIVE FIXATION 22
  • 24. MYOCARDIAL LEAD : EPICARDIALE LEAD Door chirurgen tijdens operatie te plaatsen 24
  • 25. 1 KAMER PACING (VVI) • Een elektrode in het ventrikel • Bijgevolg enkel stimulatie en • detectie in het ventrikel • Permanente vkf • Technisch niet mogelijk Tweede lead te plaatsen
  • 26. 2-KAMER PACING • Twee elektroden in het hart, • Één in het atrium en één in het ventrikel • Bijgevolg stimulatie en • detectie in het atrium en het ventrikel
  • 27. PACEMAKER IMPLANT TECHNIQUE - OVERVIEW surgeons / cardiologists / electrophysiologists • Implanting physicians: • Implant facilities: Operating room / Cardiac Catheter lab / EP lab • Equipment • Lead Placement • Threshold measurements • Pocket Closure
  • 28. POCKET CREATION Photos courtesy of Malcolm Clarke, M.D.
  • 29. TRANSVENOUS ACCESS – CHOICE OF VEINS •Left Internal Jugular •Left External Jugular •Left Subclavian •Left Axillary •Left Cephalic
  • 30. STEPS IN SURGERY • Opening of SC tissue and formation of pacemaker pocket by blunt dissection • Venous access • Lead Placement • Pacemaker Connection to Lead(s) • Pocket Closure
  • 31. PUNCTURE TECHNIQUE 5 - Insertion of the electrode 6 - Removal of the introducer
  • 32. VENOUS ACCESS Veins: Jugular (exteral, internal) Subclavian Cephalic
  • 33. PUNCTURE TECHNIQUE FOR SINGLE LEAD INSERTION 1 - Puncture, 2 - Insertion of the Seldinger wire, 3 - Introducer with dilator, 4 - Removal of dilator and Seldinger wire, insertion of the electrode
  • 34. PULSE GENERATOR IMPLANTATION SITES Left Right Abdominal
  • 35. PLACEMENT OF THE ATRIAL ELECTRODE Placement of the atrial electrode at the lateral anterior wall of the right atrium
  • 36. PLACEMENT OF THE VENTRICULAR ELECTRODE Placement via the right outflow tract respectively the pulmonary artery
  • 37. LEAD PLACEMENT • Ventricular Lead • Right Ventricular Apex (RVA) or Right Ventricular Outflow Tract (RVOT) - Ventricular Bradycardia Pacing - Sensing Intrinsic Rhythm • Atrial Lead • Right Atrial Appendage or Atrial Septal Wall - Atrial Pacing - Atrial Sensing
  • 39. VENTRICULAR LEAD - RV APEX PA & LATERAL CHEST X-RAY
  • 43. POTENTIAL PEROPERATIVE COMPLICATIONS • Bleeding • Pneumothroax • Difficult anatomy • Difficulties in positioning the electrode(s) • Arrhythmias
  • 44.
  • 45.
  • 46. VVI PACING : 1 LEAD IN RECHTER VENTRIKEL
  • 48. WAT KAN EN WAT MAG? • De patiënt kan normaal veel van zijn vroegere activiteiten hervatten. • Vermijd echter, of bespreek met cardioloog: • Competitie sport • ruwe contactsporten zoals rugby. • Activiteiten met veel schokken zoals paardrijden of botsauto’s • Inspannende lichaamsoefeningen met uw bovenlichaam zoals gewichtheffen • Rugzakken met dunne draagriemen, beste is rugzak met riem om middel ter ondersteuning
  • 49. ZAL IK HET VOELEN ALS MIJN PACEMAKER WERKT? • De meeste mensen voelen dit niet. • Vooral overgang tussen intrinsiek en gestimuleerd • Korte PR intervallen (voelt aan als kramp) • Let op bij magneet oplegging • Hogere hartritme tijdens rust/slaap • PAC en PVC (overslag) • De geleverde elektrische energie is zeer klein. • Let op voor spierstimulatie • Let op voor diafragma stimulatie • Indien u iets voelt, kan u zich best wenden tot uw cardioloog, die kan de waarneming bespreken en eventueel gepaste actie ondernemen.
  • 50. Post OP complicaties ● ● ● ● Bloeding Veneuze trombose Infectie Erosie pacemaker
  • 51. Wat gebeurt er tijdens een pacemakercontrole? ● ● ● ● Controle 6 of 12 maandelijks Anamnese ECG Ondervragen pacemaker ○ Bekijken programmatie ○ Bekijken diagnostische gegevens ○ Meten: intracardiale signalen, stimulatiedrempel, impedantie, batterij status ● Bijkomende onderzoeken op basis van de bevindingen • Continue ventriculaire pacing in rechter ventikel is schadelijk op lange termijn ( hartfalen ) : Algoritmes op ventr . Pacing te beperken
  • 52. ELECTIVE REPLACEMENT INDICATOR Magnet test rate •Dialog/Sensolog 100 •Multilog/Sensorithm •Microny •Regency 85 •Affinity •Entity •Integrity •Identity 60 •ADx Family 2.8 BOL 2.4 ERI 2.0 Battery voltage V EOL 53
  • 54. SENSITIVITY, ATRIAL CHANNEL P-wave ~3 mV Atrial fibrillation ~0.2-1.0 mV Noise ~0.1 mV Sensitivity 0.3 mV 0.1 Sensitivity 1.5 mV 0.2 0.3 0.4 0.5 1.0 0.75 1.5 2.0 3.0 4.0 55
  • 55. CLASSIC THRESHOLD DEVELOPMENT OVER TIME Threshold (Volts) Output Threshold 2 weeks 2-3 months Time Implantation 56
  • 56. DE MAGNEET MODUS VAN EEN PACEMAKER Schakelt de sensing uit: VOO DOO of AOO Modus en Stimuleert aan een vastgesteld ritme Terminologie: BOL = Beginning of life (meestal > 90 ppm) EOL = End of life ERT = Elective replacement time ERI = Elective replacement indicator: indication ERT is reached Let op: Kijk naar EKG bij magneet oplegging! Gebruik: Freq. = indicatie voor batt. Conditie Controle van de efficiëntie van de spikes.
  • 58. AANDACHTSPUNTEN OP HET WERK. • Een normaal leven lijden • Als de patiënt op het werk in contact komt met zware elektrische apparatuur kan die elektromagnetische interferentie veroorzaken die de werking van de pacemaker beïnvloed. Wees voorzichtig met: • • • • • Elektrische apparatuur in slechte staat Elektrische apparatuur die veel energie produceert zoals generatoren Booglasapparatuur medische apparatuur waaronder MRI, therapeutische straling, … metaaldetectoren en beveiligingssystemen in winkels en op vliegvelden
  • 59. • AANDACHTSPUNTENinIN HET zoals magnetrons, mixers, broodroosters, DAGELIJKS LEVEN. De meeste huishoudelijke apparaten goede staat televisies, videorecorders, elektrische dekens, kacheltjes, openingssystemen voor garagedeuren,… • Kantoorapparatuur en de meeste medische apparaten zoals ultrasoon, CT scan, mammografieën, fluoroscopie • GSM gebruik kan een pacemaker beïnvloeden: • Regel: 10 centimeter van pacemaker blijven. • St. Jude Medical pacemakers zijn GSM-proof!
  • 60. MAG IK AUTORIJDEN? • De kandidaat met een ingeplante pacemaker is niet rijgeschikt tijdens de maand die volgt op de implanting van de pacemaker of het vervangen van de pacemakerelektrode. • Bij het vervangen van enkel de pacemaker kan de kandidaat door de behandelende cardioloog onmiddellijk rijgeschikt worden bevonden
  • 61. PACEMAKERSYNDROOM EENKAMERSYSTEEM ----- SINUSAAL RITME • Verlies van AV synchroniciteit door eenkamerstimatulatie in ventrikel ( vvi ) • Verminderde inspanningscapaciteit door verminderde cardiac output • Atriale contractie tegen gesloten atrioventriculaire klep • Subjectief : duizeligheid, vermoeidheid, host, dyspnoe, orthopnoe, eigenaardig gevoel in de keel
  • 62. DEVICES • Pacemaker • Defibrillator • Remote monitoring • Biventriculaire pacing of resynchronisatietherapie
  • 63. WAT DOET EEN ICD? Maligne ventriculaire aritmieën dectecteren en behandelen
  • 64. MONOMORPHIC VT Animation * EKG Characteristics • Origin: Ventricles (Single Focus) • Mechanism: Reentry Initiated by abnormal Automaticity or Triggered activity • Characteristics: Rapid, wide, and regular QRS
  • 65. POLYMORPHIC VT Animation * • Origin: Ventricles (Wandering Single Focus) • Mechanism: Reentry with movement in the circui Initiated by Abnormal Automaticity o Triggered activity Wide and irregular QRS Complex that changes in axis • Characteristics:
  • 66. VENTRICULAR FIBRILLATION (VF) Animation * • Origin: Ventricle • Mechanism: Multiple Wavelets of reentry • Characteristics: Irregular with no discrete QRS
  • 68. ICD EVOLUTION Evolution of ICD Therapy and Adoption: 1980 to Present Number of Worldwide ICD Implants Per Year (Ref: Corporate Market Share Database)
  • 69. THE ICD SYSTEM ICD System = + + Medtronic products shown
  • 71. ICD How High Voltage Is Created Pacing Voltages For Pacing Battery 3.2 Volts For Defibrillation Voltage Multiplier and Transformer High Voltage Capacitor High Energy Shocks (Up to 800 Volts)
  • 73. Cardioverter Defibrillator Number Needed to Treat NNTx years = 100 / (% Mortality in Control Group – % Mortality in Treatment Group) Drug Therapy amiodarone ICD Therapy simvastatin Metoprolol succinate captopril (5 Yr) (2.4 Yr) (3 Yr) (3 Yr) (3.5 Yr) (1 Yr) (6 Yr) (2 Yr)
  • 74.
  • 75. SHOCK
  • 76. Wat gebeurt er tijdens de controle van een defibrillator? ● ● ● ● Controle 6 maandelijks Anamnese ECG Ondervragen defibrillator ○ Bekijken programmatie ○ Bekijken diagnostische gegevens ○ Meten: intracardiale signalen, stimulatiedrempel, impedantie, batterij status ● Bijkomende onderzoeken op basis van de bevindingen
  • 77. Ondervragen defibrillator ● De huidige defib’s kunnen al heel wat tests zelf gedaan hebben en kunnen eventuele problemen zelf vaststellen. Alerts ● Heeft de patiënt nog een eigen ritme? sensingtest ● Zijn de weerstanden op de elektroden stabiel? ● Wat is de minimumenergie nodig om te stimuleren? Drempeltest. (atriaal en ventriculair) ● Is er nog voldoende energie in de batterij tot de volgende controle? Resterende levensduur.
  • 78. DE MAGNEET MODUS VAN EEN DEFIBRILLATOR Schakelt de TACHY-therapie uit zolang de magneet op Defibrillator ligt ! Wordt wel eens gebruikt tijdens een ingreep of ingeval de defibrillator shockt als het niet moet. Is geen follow-up-middel!
  • 80. AANDACHTSPUNTEN OP HET WERK. • Een normaal leven lijden • Als de patiënt op het werk in contact komt met zware elektrische apparatuur kan die elektromagnetische interferentie veroorzaken die de werking van de pacemaker beïnvloed. Wees voorzichtig met: • • • • • Elektrische apparatuur in slechte staat Elektrische apparatuur die veel energie produceert zoals generatoren Booglasapparatuur medische apparatuur waaronder MRI, therapeutische straling, … metaaldetectoren en beveiligingssystemen in winkels en op vliegvelden
  • 81. AANDACHTSPUNTEN IN HET DAGELIJKS LEVEN. • De meeste huishoudelijke apparaten in goede staat zoals magnetrons, mixers, broodroosters, televisies, videorecorders, elektrische dekens, kacheltjes, openingssystemen voor garagedeuren,… • Kantoorapparatuur en de meeste medische apparaten zoals ultrasoon, CT scan, mammografieën, fluoroscopie • GSM gebruik kan een pacemaker beïnvloeden: • Regel: 10 centimeter van pacemaker blijven. • St. Jude Medical pacemakers zijn GSM-proof!
  • 82. IMPLICATIES OP RIJBEWIJS De kandidaat met een ingeplante automatische defibrillator is NIET rijgeschikt. • De kandidaat kan evenwel na een periode van één of drie maand, te rekenen vanaf de implanting, rijgeschikt worden bevonden op basis van een recent verslag, afgeleverd door de cardioloog van het geneeskundig centrum dat de ingreep heeft uitgevoerd. De een maand is enkel van toepassing in het geval de patiënt geen cardiac arrest heeft doorgemaakt. Tijdens deze periode van één of drie maand en mag er geen stroomstoot geweest zijn die het hartritme kan aantasten, indien wel is 3 maand terug van toepassing
  • 83. BEEPING • ACID kan beepen • - lead problemen • - batterijprobemen • - na interventie
  • 84. DEVICES • Pacemaker • Defibrillator • Remote monitoring • Biventriculaire pacing of resynchronisatietherapie
  • 85. HEART MONITORING OPTIONS SYNCOPE OCCURS INFREQUENTLY LONG-TERM MONITORING IS NEEDED 12-Lead 10 Seconds Holter ELR 2 - 3Days 7 Days- 30 days 3 Days to 30+ Days MCOT ELR Up to 36 Months ILR ILR = insertable loop recorder, ELR=External loop recorder MCOT= mobile cardiac outpatient telemetry
  • 86. EXTERNAL ‘LOOP’ RECORDERS (ELRS) 7-day ‘Patch’ ELR Up to 1 month recording
  • 87. INSERTABLE CARDIAC MONITORS (ICM) Reveal® System, Medtronic Inc. Minneapolis, MN -manual/auto trigger -remote download (CareLink®) Confirm®, St Jude Medical St Paul, MN -manual/auto trigger -remote download (Merlin®)
  • 88. SYMPTOM-RHYTHM CORRELATION: ICM PERMITS BOTH PATIENT-TRIGGERED AND AUTOMATIC ACTIVATION Auto Activation Point Patient Activation Point Medtronic data on file
  • 89. INSERTABLE CARDIAC MONITOR (ICM) • Subcutaneous ECG monitor • 3-year battery • Records, stores, and may transmit remotely • Patient-activated and/or automatically-activated • Longevity of current ILRs up to 36 months • Indications 1. Patients with clinical syndromes or situations at increased risk for cardiac arrhythmias 2. Patients who experience transient symptoms such as dizziness, palpitation, syncope and chest pain, that may suggest a cardiac arrhythmia • palpitations, cryptogenic stroke, ‘refractory’ epilepsy Medtronic, Inc. Reveal® XT 9529 Insertable Cardiac Monitor Clinician Manual. 2011.
  • 90. Remote Monitoring System Patient Transmitter * By landline phone or GSM network Wireless Service center Data sent to transmission* the physician
  • 91.
  • 92. REVEAL INDICATIES • Onuitgelegde syncopes • Vermoeden van ernstige ritmestoornissen • 1 keer per week of na een event worden gegevens doorgestuurd door patiënt
  • 93. DEVICES • Pacemaker • Defibrillator • Remote monitoring • Biventriculaire pacing of resynchronisatietherapie
  • 94.
  • 95.
  • 96. VLBTB : QRS DUUR >120 MSEC, RSR’ PATROON LINKERPRECORDIALEN EN I EN AVL
  • 98.
  • 99.
  • 100.
  • 101.
  • 102.
  • 103. SELECTIE PATIËNTEN • Responders – non responders • Technical challenging • - succes rate 93% • - complicaties 4 – 5% ( overlijden 0,3% ) : vene dissectie, perforatie, tamponnade, pneumothorax • - duur ingreep ----- infectierisico
  • 104. MRI COMPATIBILITEIT • Pacemaker, aicd, biventr pacemaker : niet mri compatibel • Laatste generatie PM soms wel : steeds navragen ( toestel dient geherprogrammeerd te worden , bepaalde zones van lichaam, bepaalde MRI toestellen ) • Toestellen voor remote monitoring : in principe geen probleem
  • 106. VLIEGTUIGREIZEN EN DEVICES • PM en AICD kunnen onderhevig zijn aan electromagnetische interferentie • Airport security : - gates : induceren magnetische veld. Erdoor wandelen geen probleem. Alarm kan wel afgaan. Stilstaan erin wordt afgeraden - hand-held metal detector : ICD te vermijden • Electromagn. Interferentie tijdens vlucht is te verwaarlozen.
  • 107.
  • 109. CRT