- 1. © 2003 Terumo Cardiovascular Systems
51
CDI™ Blood Parameter Monitoring System 500
Optical fluorescence and reflectance-based in-line system to continuously
monitor eleven critical blood gas parameters with lab-quality accuracy.
SPECIFICATIONS
Displayed Parameters Operating Range Resolution
pH 6.8 – 8.0 pH units 0.01 pH units
PCO2 10 – 80 mmHg (1 – 11 kPa) 1 mmHg (0.1 kPa)
PO2 10 – 500 mmHg (1 – 67 kPa) 1 mmHg (0.1 kPa)
K+ 1.0 – 8.0 mmole/L 0.1 mmole/L
Temperature 10° – 45° Celsius 0.1° Celsius
Oxygen saturation (SO2) 60% – 100% 1%
Hematocrit (Hct) 15% – 45% 1%
Total hemoglobin (Hgb) 5 – 15 g/dL 0.1 g/dL
Oxygen consumption 10 – 400 mL/min 1 mL/min
BE -25 – 25 mEq/L 1 mEq/L
HCO3 0 – 50 mEq/L 1 mEq/L
Blood flow 0 – 9.9 L/min 0.1 L/min
Physical Specifications Monitor Calibrator
Height 11.0" (27.9 cm) 12.5" (31.8 cm)
Width 12.5" (31.8 cm) 8.0" (20.3 cm)
Depth 6.0" (15.2 cm) 8.0" (20.3 cm)
Weight 16.1 lbs (7.3 kg) 8.4 lbs (3.8 kg)
Monitor power requirements and specifications
100 – 240 VAC, 50/60 Hz
12V lead acid backup battery
Data output port: RS-232 serial interface
Pumping systems input port: RS-232/RS-485 serial interface
Model CDI510H shunt sensor
Sterile, heparin-treated
Priming volume 1.2 mL
System measurement cycle time
pH, PCO2, PO2 = one measurement per second
K+ = one measurement per six seconds
SO2, Hct, Hgb = one measurement per eighteen milliseconds
System display update: Every six seconds
ORDERING INFORMATION
CDI™ Blood Parameter Monitoring System 500
DESCRIPTION CATALOG NO. UNITS/CASE
Monitor Configurations
Monitor with one blood parameter module ................................................................ 500A .................... 1
Monitor with one blood parameter module and one Hct/Sat probe ........... 500AHCT .................... 1
Monitor with two blood parameter modules ........................................................... 500AV .................... 1
Monitor with two blood parameter modules and one Hct/Sat probe ..... 500AVHCT .................... 1
continued, next page
CDI Model 540 calibrator
CDI System 500 monitor
FEATURES
Continuous measurements and
calculations
Lab-quality accuracy
Easy-to-use in-line shunt sensor
Color LCD display
Integral printer
Automated 2-point gas tonometered
calibration
CDI™ Blood Parameter Monitoring System 500
Blood
Parameter
Monitoring
Systems
Monitoring
Systems
- 2. © 2003 Terumo Cardiovascular Systems
52
CDI™ Blood Parameter Monitoring System 500, continued
DESCRIPTION CATALOG NO. UNITS/CASE
Calibrator .............................................................................................................................540 .................... 1
Accessories for Use with CDI System 500
Gas A, calibration gas for use with Calibrator 540 ............................................... CDI506 .................... 1
Gas B, calibration gas for use with Calibrator 540 ............................................... CDI507 .................... 1
Printer paper .......................................................................................................................... 7310 .................... 5
Monitor pole clamp, 7.0" (17.8 cm) arm length, calibrator mount ................. CDI517 .................... 1
Monitor pole clamp, 4.5" (11.4 cm) arm length ..................................................... CDI518 .................... 1
Cable head bracket .......................................................................................................... CDI519 .................... 1
Disposable Sensors for Use with CDI System 500
Shunt sensor for use with System 500, heparin treated .................................. CDI510H .................. 20
Disposable H/S Cuvettes for Use with
CDI Systems 500 and 100
1/4" x 1/4" .............................................................................................................................. 6914 .................. 20
3/8" x 3/8" .............................................................................................................................. 6913 .................. 20
1/2" x 1/2" .............................................................................................................................. 6912 .................. 20
1/4" x 1/4" with heparin treatment3,8
....................................................................... 6914H .................. 20
3/8" x 3/8" with heparin treatment3,8
....................................................................... 6913H .................. 20
1/2" x 1/2" with heparin treatment3,8
....................................................................... 6912H .................. 20
1/4" x 1/4" with 6" (15.2 cm) extension tube............................................................. 6934 .................. 10
3/8" x 3/8" with 6" (15.2 cm) extension tube............................................................. 6933 .................. 10
1/2" x 1/2" with 6" (15.2 cm) extension tube............................................................. 6932 .................. 10
Monitoring
Systems
Blood
Parameter
Monitoring
Systems
CDI H/S cuvette
CDI System 500
shunt sensor, in-line
3
Made to order. Expect delivery in 4-6 weeks.
8