Introduction
Permanent polymers of first generation drug-eluting stents
(DES) are imputed to be a contributing factor in vessel wall
inflammation, positive remodeling, stent malapposition
and ultimately late stent thrombosis (1-3). New DES
with biodegradable polymers and DES with no polymers
have been developed and tested in preclinical and clinical
experiments to overcome the limitations of first generation
DES with durable polymers.
Polymeric coatings was thought to be essential for
effective local drug delivery to the vessel wall, since its
absence has been related to target vessel failure (4). However,
many recent studies with surface-modified polymer-free
DES showed that the elimination of polymers might not
compromise the anti-restenosis efficacy of DES (5,6).
Accordingly, new polymer-free DES has been deve loped
by a number of stent manufactures by texturizing, creating
grooves, niches, or a microporous stent surface (7,8).
This article describes the experimental results of the
novel Focus np stent (Envision Scientific Pvt. Ltd., Gujarat,
Original Article
A new polymer-free drug-eluting stent with nanocarriers eluting
sirolimus from stent-plus-balloon compared with bare-metal stent
and with biolimus A9 eluting stent in porcine coronary arteries
Celso K. Takimura1
, Micheli Z. Galon1
, Paulo S. Gutierrez1
, Prakash Sojitra2
, Ashwin Vyas2
, Manish Doshi2
,
Pedro A. Lemos1
1
Department of Interventional Cardiology, Heart Institute (InCor), University of São Paulo Medical School, São Paulo, Brazil; 2
Envision Scientific
Pvt. Ltd., Office 1-3, Silver Palm II, Opposite Snehmilan Garden, Kadampalli, Nanpura, Surat, Gujarat, India
Correspondence to: Celso K. Takimura, MD, PhD. Heart Institute (InCor) University of São Paulo Medical School, Av. Dr. Eneas Carvalho de Aguiar
44, Bloco I, 3o andar, São Paulo 04503-000, Brazil. Email: celsotakimura@hotmail.com.
Background: Permanent polymers in first generation drug-eluting stent (DES) have been imputed to be a
possible cause of persistent inflammation, remodeling, malapposition and late stent thrombosis. We aim to
describe the in vivo experimental result of a new polymer-free DES eluting sirolimus from stent-plus-balloon
(Focus np stent, Envision Scientific) compared with a bare-metal stent (BMS) (Amazonia CroCo, Minvasys)
and with a biolimus A9 eluting stent (Biomatrix, Biosensors).
Methods: In 10 juvenile pigs, 23 coronary stents were implanted in the coronary arteries (8 Amazonia
CroCo, 8 Focus np, and 7 Biomatrix). At 28-day follow-up, optical coherence tomography (OCT) and
histology were used to evaluate neointimal hyperplasia and healing response.
Results: According to OCT analysis, Focus np stents had a greater lumen area and less neointimal
hyperplasia response than BMS and Biomatrix had. Histomorphometry results showed less neointimal
hyperplasia in Focus np than in BMS. Histology showed a higher fibrin deposition in Biomatrix stent
compared to Focus np and BMS.
Conclusions: The new polymer-free DES with sirolimus eluted from stent-plus-balloon demonstrated
safety and reduced neointimal proliferation compared with the BMS and Biomatrix stents at 28-day follow-
up in this porcine coronary model. This new polymer-free DES is promising and warrants further clinical
studies.
Keywords: Neointimal; restenosis; stent; vascular injury
Submitted Feb 09, 2015. Accepted for publication Mar 16, 2015.
doi: 10.3978/j.issn.2223-3652.2015.03.08
View this article at: http://dx.doi.org/10.3978/j.issn.2223-3652.2015.03.08
114 Takimura et al. Polymer-free Focus np stent
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
India), a polymer-free stent developed with nanotechnology
that uses nanosized phospholipid particles to deliver
sirolimus from a combined balloon-plus-stent platform.
Methods
Animal study
Animal experiments were in compliance with the Guide
for the Care and Use of Laboratory Animals (Institute of
Laboratory Animal Resources, Commission on Life
Sciences and National Research Council. National Academy
Press, Washington, D.C., 1996), and the study protocol
was approved by the Institutional Ethics Committee. Stents
were implanted in 10 juvenile domestic pigs, weighing 23 to
30 kg, and animals were followed for 28 days. Oral aspirin
(100 mg) and clopidogrel (300 mg loading dose; 75 mg daily
dose) were administered starting 1 day before the procedure
and continued daily until euthanasia. Animals were
sedated and anesthetized with an intramuscular solution of
midazolam (0.5 mg/kg) and ketamine (3 mg/kg) followed by
intravenous thiopental (25 mg/kg). Mechanical ventilation
was established and anesthesia maintained with inhaled
isoflurane (2% to 5%) throughout the procedure. Arterial
access was achieved by a femoral cut-down with a general
sterile technique, and coronary angiography was then
performed using 6F guiding catheters after intracoronary
administration of nitroglycerin (200 μg) and intravenous
heparin (7,500 IU). Three types of stents were implanted: (I)
cobalt-chromium polymer-free sirolimus eluting Focus np
stent; (II) stainless steel biolimus A9-eluting Biomatrix stent
(Biosensors International, Singapore) and (III) a cobalt-
chromium bare metal stent (BMS) Amazonia CroCo®
stent, (Minvasys, Paris, France). All stents were 3.0 mm
in diameter, and stent length was 16 mm for polymer-free
sirolimus eluting Focus np stent, 14 mm for biolimus A9
eluting Biomatrix stent, and 16 mm for Amazonia CroCo®
BMS. For each animal, the study stents were deployed
according to a randomization chart (1:1:1) in the major
branches of the coronary arteries (left anterior descending
coronary artery, left circumflex artery, and right coronary
artery) in a non-tapered, non-bifurcation, non-angulated
segment of 3.0 mm in diameter by visual estimation. The
stents were implanted through inflation of the delivery
balloons for approximately 30 seconds in all study groups.
The guiding catheter was used as a reference to obtain
a 1.1:1 to 1.2:1 stent-to-artery ratio compared with the
baseline vessel diameter and to induce moderate vessel
injury (9). Animals were allowed to recover and received
appropriate postoperative care. At 28-day follow-up, control
coronary angiography and optical coherence tomography
(OCT) were performed, and the animals under deep
anesthesia were euthanized with a lethal dose of potassium
chloride.
Design of Focus np stent
The Focus np stent is a cobalt-chromium L-605 stent with
strut thickness of 73 µm. It is crimped onto a COPAN
Co-Polyamide delivery balloon. Nanoparticles with a
mean diameter of 210 nm, containing sirolimus (108 µg of
sirolimus on a 3.0×16 mm stent), were synthesized using
an ultrasonic homogenizer and applied using an inert gas-
assisted spray process in stents already crimped onto the
balloon delivery system. Consequently, the coating was
uniformly distributed over the stent-plus-balloon surface,
with only the abluminal face of the stent struts having the
coating applied (Figure 1). Focus np stent thus, is a polymer-
free cobalt-chromium stent delivering sirolimus through
nanoparticles from balloon and abluminal stent strut
face. The manufacturing and coating process of sirolimus
nanoparticles to metallic stent platform and the balloon-
catheter delivery system is described in detail elsewhere (10).
In a previous experiment in iliac arteries of rabbits treated
with a 60 s dilatation with sirolimus-coated balloons, these
nanoparticles were found initially in the endothelial surface,
then in the medial layer and finally in the deeper stratum
like the adventitia (11). In vitro drug release analyzed using
HPLC revealed that sirolimus is delivered from the stent
with an initial burst followed by a prolonged release for up
to 40 days. In-tissue release of sirolimus from phospholipid
Figure 1 Polymer-free Focus np stent assessed by scanning electron
microscopy (SEM) (100× magnification).
115Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
nanoparticles is longer because of encapsulation in drug
delivery matrix (10).
Description of control stents
Amazonia CroCo (Minvasys, France) is a cobalt-chromium
BMS with a thin strut thickness of 73 µm and a design that
combines intermediate and open cells. This stent is used
as the metallic platform in the Amazonia Pax paclitaxel
polymer-free DES (12).
Biomatrix (Biosensors International, Singapore) is a
DES that consists of a stainless steel platform with a strut
thickness of 112 µm and a quadrature link design. It is
coated exclusively in the abluminal stent surface with a
biodegradable polymer (PLA, polylactic acid) and biolimus
A9 (218 µg on a 3.0×14 mm stent). This biodegradable
coating completely dissolves into carbon dioxide and
water after a 6- to 9-month period. Biolimus A9 is a
semi-synthetic sirolimus analogue with 10 times higher
lipophilicity and similar potency as sirolimus (13,14).
Optical coherence tomography (OCT)
At 28-day follow-up and after control angiography, a 0.014-
inch guidewire was inserted into the coronary arteries, and
an occlusion balloon (Helios™, LightLab Imaging) was
advanced distally to the treatment sites. The 0.014-inch
wire was then removed and replaced for a 0.016-inch OCT
catheter (Image Wire™, LightLab Imaging, Westford,
MA, USA). The Helios™ balloon (LightLab®
Imaging)
was pulled back proximally to the stent and inflated with
a 1:1 saline:contrast solution until occlusion of the artery.
Then, saline was flushed to clear the lumen from blood and
an automatic pullback of the OCT catheter was initiated at
a speed of 1 mm/s with recording of OCT images. These
images were analyzed for thrombus, strut malapposition and
endothelialization.
A strut was considered malapposed if the distance from
the endoluminal surface of the stent to the vessel wall
was higher than the sum of the metal and, if present, the
polymer thickness. The cutoff points used for each stent
type were: BMS (Amazonia CroCo) 73 µm, polymer-free
sirolimus-eluting stent (Focus np) 73 µm, biolimus-eluting
stent (Biomatrix) 120 µm. A stent strut was defined as
covered, if there was a visible layer of tissue covering it.
Semi-automatic measurements were made with
proprietary software for off-line analysis with the M2 System
(LightLab Imaging, Westford, MA, USA) to determine
lumen area, stent area, neointimal thickness (distance from
stent strut until lumen), neointimal area (stent area minus
lumen area) and percentage of neointimal area (ratio of
neointimal area and stent area, multiplied by 100) (15).
Light microscopy
Immediately after euthanasia, the hearts were excised
and the coronaries were pressure-perfused (~100 mm Hg)
via the ascending aorta with 0.9% saline, to clear the
blood, followed by perfusion of 10% buffered formalin
for a minimum of 1 hour. The stented arterial segments
were then excised from the heart by dissection and fixed
by immersion in 10% formalin overnight. The artery-
stent specimens were dehydrated with ethanol solutions
of increasing concentrations and embedded in methyl
methacrylate resin. A total of three cross-sections (proximal,
mid, and distal) were obtained from each vessel on a
rotary microtome (cut thickness 3.5 μm) and stained with
hematoxylin and eosin and Verhoeff stains for elastic fibers.
Microscopic digital morphometry with Leica QWin®
software (Leica Microsystems, Wetzlan, Germany) was
used to measure areas of the external elastic lamina (EEL),
internal elastic lamina (IEL), stent, and lumen. Media area
was defined as EEL area—EL area. Neointimal hyperplasia
area was defined as IEL area—lumen area. Percent
neointimal burden was defined as 100× (neointimal area/
IEL area). Neointimal thickness was measured at the stent
strut and at the inter-strut space.
The cross-sections were examined for semi-quantitative
analysis of endothelialization, inflammation, fibrin
content, angiogenesis, and vessel injury according to
previously described scoring schemes (16-19). Stent
endothelialization score was defined as the extent of the
circumference of the arterial lumen covered by endothelial
cells and graded from 1 to 3 (1=25%, 2=25% to 75%,
3≥75%). Inflammation was graded as 0, none; 1, scattered
inflammatory cells; 2, inflammatory cells encompassing
50% of a strut in at least 25% to 50% of the circumference
of the artery; 3, inflammatory cells surrounding a strut in
at least 25% to 50% of the circumference of the artery.
Inflammation modified was graded as 0, no inflammatory
cells surrounding the strut; 1, light, non-circumferential
lymphohistocytic infiltrate surrounding the strut; 2,
localized, moderate to dense cellular aggregate surrounding
the strut non-circumferentially; 3, circumferential dense
lymphohistiocytic cell infiltration of the strut. The intimal
fibrin content was graded as 0, no fibrin deposition; 1,
116 Takimura et al. Polymer-free Focus np stent
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
focal residual fibrin involving any portion of the artery or
moderate fibrin deposition adjacent to the strut involving
<25% of the circumference of the artery; 2, moderate fibrin
involving >25% of the circumference of the artery or heavy
deposition involving <25% the circumference of the artery;
or 3, heavy deposition of fibrin involving >25% of the
circumference of the artery. The injury score (Schwartz) was
graded as 0, internal elastic lamina intact, media compressed
but not lacerated; 1, internal elastic lamina lacerated,
media typically compressed but not lacerated; 2, internal
elastic lacerated, media visibly lacerated, external elastic
lamina intact but compressed or 3, external elastic lamina
lacerated, typically large lacerations of media extending
through the external elastic lamina, stent struts sometimes
residing in adventitia. The injury score (Gunn) was graded
as 0, no impression of metal upon media; 1, deformation of
the internal elastic lamina by <45°; 2, deformation of the
internal elastic lamina by >45°; 3, rupture of the internal
elastic lamina or 4, rupture of the external elastic lamina (i.e.,
complete medial rupture).
Statistical analysis
Data are presented as mean ± SD and/or median (interquartile
range; IQR). The analyzed per stent individual OCT frames
and histologic cross-sections were treated as independent
samples (7). Differences in continuous variables among
stent groups were evaluated by using parametric (analysis
of variance) or non-parametric (Kruskal-Wallis) statistical
tests as needed. Histologic ordinal variable results were
overall compared for differences among stent groups with
independent-samples Kruskal-Wallis test. The statistical
software SPSS 20.0 was used to analyze data, and a priori
P value ≤0.05 was considered significant.
Results
Twenty-three stents (8 Amazonia CroCo, 8 Focus np,
and 7 Biomatrix) were successfully implanted in the
coronary arteries of 10 swine. Mean balloon pressure to
implant the stents was equal among groups (P=0.895).
Nine of 10 animals survived until the follow-up phase of
the study. There was one early death (pig number 2) that
occurred during the second day after the stent implantation
procedure (Biomatrix stent implanted in the left anterior
descending artery and Amazonia CroCo in the left
circumflex artery). In this animal, a stent was implanted too
distally in the coronary tree, resulting in dissection at the
distal stent edge that was managed by prolonged balloon
inflation with final TIMI III flow. Pathology revealed the
presence of a recent intraluminal thrombus in all stents.
The remaining 9 pigs with 21 stents (7 Amazonia CroCo,
8 Focus np, and 6 Biomatrix) were included in the study for
final analysis. Stent distribution is summarized in Table 1. At
28-day follow-up, all stents in surviving animals were patent
at angiography and could be imaged with OCT.
OCT findings
OCT quantitative analysis is summarized in Table 2, and
illustrative example OCT images of each group are shown
in Figure 2 (upper panel). At 28 days, lumen area was larger in
Focus np compared with BMS and Biomatrix. Neointimal
hyperplasia thickness was reduced in Focus np compared
with BMS and Biomatrix. NIH area and percentage NIH
were smaller in Focus np than BMS and Biomatrix. Biomatrix
stent showed either a less-pronounced neointimal thickness
or hyperplasia area than BMS. Stent struts were completely
covered and well apposed in all stent groups.
Histomorphometric and semiquantitative histologic analysis
The histomorphometric analysis is summarized in Table 3,
and illustrative example histologic images of each group
are shown in Figure 2 (bottom panel). Lumen area, internal
Table 1 Stent distribution
Animal N° BMS Focus np Biomatrix
1 LAD RCA N/A
2* LCX N/A LAD
3 RCA LAD LCX
4 N/A LCX RCA
5 LAD RCA N/A
6 LCX N/A LAD
7 RCA LCX LAD
8 N/A LCX RCA
9 LAD RCA N/A
10 LCX RCA LAD
Balloon pressure (ATM) 11.6±3.7 12.0±1.9 11.0±2.9
*, Early animal death, excluded from stent study analysis. BMS
indicates Amazonia CroCo®
. LAD, left anterior descending
artery; LCX, left coronary circumflex artery; RCA, right
coronary artery; N/A, not applicable.
117Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
Table 2 Optical coherence tomography quantitative analysis
Variables BMS Focus np Biomatrix
Overall
P value
Post-hoc comparisons,
P<0.05
Stent, n 7 8 6 – –
OCT frame images/
stent struts, n
63/728 72/707 54/499 – –
Lumen area, mm² 4.93±1.71
5.38 (4.2-6.16)
6.58±1.65
6.90 (5.30-8.0)
5.94±1.49
6.49 (4.28-7.0)
<0.001 Focus np > BMS & Biomatrix
Stent area, mm² 7.70±1.14
7.35 (6.43-8.77)
7.96±0.97
7.97 (6.91-8.92)
7.76±0.98
8.04 (7.18-8.47)
0.705 –
NIH thickness, mm 0.26±1.17
0.19 (0.10-0.38)
0.12±0.07
0.08 (0.07-0.19)
0.18±0.09
0.16 (0.11-0.26)
<0.001 Focus np < BMS & Biomatrix;
Biomatrix < BMS
NIH area, mm² 2.76±1.66
1.92 (1.19-4.53)
1.38±0.91
1.0 (0.82-2.25)
1.81±0.80
1.81 (1.11-2.34)
<0.001 Focus np < BMS & Biomatrix;
Biomatrix < BMS
Percent NIH, % 35.81±22.14
26.30 (16.19-51.89)
18.23±13.89
12.36 (10.33-28.83)
24.26±12.06
21.85 (13.78-36.89)
<0.001 Focus np < BMS & Biomatrix;
Biomatrix < BMS
Uncovered struts 0 0 0 >0.999
Malapposed struts 0 0 0 >0.999
Data are presented as mean ± SD and median (interquartile range). NIH, neointimal hyperplasia.
Figure 2 Representative OCT images (upper panel) and histologic images (bottom panel) 28 days after porcine coronary stent implantation of
Amazonia CroCo BMS (left), Focus np stent (middle), and Biomatrix stent (right). OCT, optical coherence tomography; BMS, bare-metal stent.
1 mm 1 mm 1 mm
1 mm
118 Takimura et al. Polymer-free Focus np stent
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
elastic lamina area, external elastic lamina area, stent area,
and neointimal thickness at strut and interstrut were not
different among groups. The Focus np stent had a smaller
neointimal area and percentage neointimal area compared
with BMS only.
Semiquantitative histologic analysis revealed that Focus
np stents were associated with an intermediate degree of
fibrin content, smaller than the Biomatrix but higher than
the control BMS (Table 4). In addition, Focus np stents also
had an intermediate degree of angiogenesis, smaller than
Table 3 Histomorphometric analysis
Variables BMS Focus np Biomatrix
Overall
P value
Post hoc comparisons,
P<0.05
Stents, n 7 8 6
Sections, n 20 22 15
Lumen area, mm2
3.36±1.01
3.41 (2.67-3.97)
3.83±1.05
4.06 (2.96-4.62)
3.69±1.13
3.69 (2.64-4.65)
0.363 N/A
IEL area, mm2
4.67±0.89
4.61 (4.07-5.34)
4.85±0.97
5.10 (4.46-5.51)
5.08±0.85
5.17 (4.54-5.79)
0.430 N/A
EEL area, mm2
5.83±0.93
5.66 (5.17-6.44)
5.88±1.11
6.25 (5.30-6.67)
6.06±1.01
6.19 (5.33-6.94)
0.797 N/A
Stent area, mm2
4.92±0.78
4.79 (4.42-5.48)
4.95±0.94
5.25 (4.48-5.58)
5.22±0.86
5.41 (4.59-5.83)
0.556 N/A
NIH area, mm2
1.56±0.60
1.55 (0.96-1.88)
1.11±0.50
0.97 (0.72-1.38)
1.52±0.52
1.53 (1.11-1.96)
0.021 Focus np < BMS
Percent NIH, % 32.5±13.6
36.89 (18.89-39.25)
23.4±11.9
18.71 (14.45-29.77)
30.5±12.6
33.25 (18.82-42.45)
0.046 Focus np < BMS
NIH thickness at strut, mm 0.18±0.10
0.20 (0.08-0.22)
0.15±0.12
0.10 (0.08-0.17)
0.17±0.08
0.15 (0.08-0.27)
0.591 N/A
NIH thickness interstrut, mm 0.17±0.09
0.18 (0.09-0.21)
0.13±0.08
0.10 (0.07-0.18)
0.16±0.89
0.16 (0.06-0.24)
0.393 N/A
Data are presented as mean ± SD and median (Interquartile range). IEL, indicates internal elastic lamina; EEL, external elastic
lamina; NIH, neointimal hyperplasia; N/A, not applicable.
Table 4 Semiquantitative histologic analysis
Variables BMS Focus np Biomatrix Overall P value
Stents, n 7 8 6
Sections, n 19 20 16
Endothelialization 3 3 3 >0.999
Inflammation 0.75±0.44 0.53±0.62 0.62±0.50 0.511
Inflammation modified 0.75±0.44 0.47±0.51 0.62±0.50 0.408
Fibrin 0.19±0.40 0.47±0.51 1.69±0.94 <0.001
Smooth muscle cells 3.00±0.00 2.76±0.75 2.62±0.65 0.070
Angiogenesis 1.56±0.81 1.41±1.06 0.77±0.59 0.025
Injury score (Schwartz) 1.25±0.57 1.00±0.35 1.54±0.51 0.017
Injury score (Gunn) 2.31±0.47 2.06±0.42 3.00±0.40 <0.001
BMS, bare-metal stent.
119Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
the BMS but higher than the Biomatrix stent. Even though
the delivery pressure and the tentative balloon:artery ratio
(1.1:1.0) was the same for all implants, the vessel injury
score was higher in Biomatrix stent group.
Discussion
In this preclinical experiment, the polymer-free Focus
np balloon-plus-stent coated with sirolimus-containing
nanoparticles had an excellent safety profile, yielding good
results at 28 days after implantation in pig coronaries
evaluated by OCT and histology. Due to a unique patented
manufacture process that allows Focus np to elute sirolimus
from a stent-plus-balloon platform, therefore delivering
drug to the entire vessel circumference, there is a possibility
that this device might be more effective in inhibiting
neointimal growth than existing commercial DES eluting
antiproliferative drugs from stent surface alone. Indeed, in
the present study, the antiproliferative effect of Focus np
assessed by OCT was significantly higher than the clinically
available Biomatrix DES, which elutes drug exclusively
from the abluminal stent strut surface.
Related to the comparison of the Focus np stent with
the Biomatrix stent, although these two stents are markedly
different regarding metallic platform, presence or absence
of polymer, and differing antiproliferative agents, the
Biomatrix stent data were useful as a “positive” control
group, enabling us a direct comparison with this successful
clinically used DES.
A relevant finding in our study was the higher fibrin
content in the histologic sections of the Biomatrix stent
compared to the Focus np stent and to BMS. This suggests
a more delayed healing process in the biolimus A9 eluting
stent. Of note, in a preclinical experiment comparing the
high-dose biolimus A9 stent (225 µg/14 mm), low-dose
biolimus A9 (112 µg/14 mm) polymer-free Biofreedom
stent (Biosensors International, Singapore), polymer-
coated sirolimus-eluting Cypher stent, and BMS in porcine
coronary arteries, histological evaluation showed a higher
fibrin deposition for all these 3 DES groups than BMS at
28 days (7).
Sirolimus is a lipophilic drug with potent antiproliferative
and immunosuppressant effects. There are some technical
difficulties in coating a balloon or a balloon-plus-stent and
effectively delivering this drug into the arterial wall tissue
without a durable or a biodegradable polymer. In a sirolimus
eluting polymer-free DES, a fraction of the sirolimus load
can be removed from the device when passing through the
hemostatic valve, delivery catheter, calcified lesions, or be
simply washed out by the blood stream. Encapsulation of
sirolimus in nanoparticles to deliver it to the vessel wall
through stent-plus-balloon is an attractive technology as the
plaque area treated involves both balloon and stent area in
contact with the vessel inner wall (stent edges included).
Many polymer-free stents have been investigated
in clinical studies, including BioFreedom (Biosensors
International, Singapore) (20), Cre8 (CID, Italy) (21),
Yukon Choice PF (Translumina, Germany) (22), Nile
Pax (Minvasys, France) (23) and VESTAsync stent (MIV
Therapeutics, USA) (24). The strategies for drug binding
and release without the use of polymers consists of loading
antiproliferative agents (biolimus A9, amphilimus, paclitaxel,
sirolimus or a combination of sirolimus and probucol) on
microporous or reservoirs on stent surface.
The results of this preclinical study with Focus np stent
warrants initiation of clinical studies as a First-in-Man (FIM)
study and randomized clinical trials comparing it with
currently approved DES.
Study limitations
The pig model of coronary stent implantation is well-
known to be more suited for safety analysis than for efficacy
evaluation of new devices. Therefore, the differences in
neointimal proliferation seen among the groups should be
interpreted with caution. The dissimilarities of the stent
types and their respective delivery systems may account
for some of the differences in the injury scores. It is not
possible to rule out that some of the differences in the
vascular response (e.g., neointimal proliferation, fibrin
deposition) might have been related to a discrepancy in
baseline vessel injury. An ideal experiment to assess the
vessel response to this new technology—nanocarriers
eluting sirolimus from stent-plus-balloon—would be testing
stent types with exactly the same delivery balloons, metallic
designs and stent lengths, and two DES releasing the same
antiproliferative agent with the same dosage but different
releasing drug mechanisms.
The occlusive method (time-domain OCT, M2 System,
LightLab Imaging, Westford, MA, USA) was used in this
experimental study. This method, recently replaced by
the current non-occlusive method (Fourier-domain OCT,
C7XR, St. Jude Medical), was extensively used in preclinical
and clinical studies. Comparing these two OCT methods
in an in vivo non-stented human coronary artery study, the
occlusive method tended to reduce lumen dimension by
120 Takimura et al. Polymer-free Focus np stent
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
13%, with equivalent OCT image quality (25). In our study
of stented porcine coronary arteries, the use of the occlusive
method was well suited to analyze in-stent neointimal
hyperplasia as the OCT NIH area and percentage NIH
results were concordant with histomorphometry (26).
Our sample size was small, but the carefully selected
arterial segments to accommodate the unique diameter of
stents (3.0 mm) available for implantation, and the analysis
of OCT frames and histologic sections as independent
samples reduced the need of a larger sample size.
Conclusions
The Focus np stent with polymer-free sirolimus eluted
from stent-plus-balloon demonstrated safety and a reduced
intimal proliferation compared with BMS and Biomatrix
stent in the present 28-day follow-up porcine coronary
model. The new nanocarrier technology eluting sirolimus
from stent-plus-balloon is promising and warrants
confirmation of its safety and efficacy in future clinical
studies.
Acknowledgements
We thank Richard Silva, Leonora Loppnow and Marcio
Chaves for technical assistance and Ann Morcos (www.
morcosmedia.com) for reviewing the manuscript. The study
was supported in part by Envision Scientific PVT Ltd.
Disclosure: Celso K. Takimura, Micheli Z. Galon, Paulo
S. Gutierrez and Pedro A. Lemos declare no conflict of
interest. Prakash Sojitra, Ashwin Vyas and Manish Doshi
works for Envision Scientific PVT Ltd.
References
1.	 Virmani R, Guagliumi G, Farb A, et al. Localized
hypersensitivity and late coronary thrombosis secondary
to a sirolimus-eluting stent: should we be cautious?
Circulation 2004;109:701-5.
2.	 Cook S, Ladich E, Nakazawa G, et al. Correlation of
intravascular ultrasound findings with histopathological
analysis of thrombus aspirates in patients with very late
drug-eluting stent thrombosis. Circulation 2009;120:391-9.
3.	 Lüscher TF, Steffel J, Eberli FR, et al. Drug-eluting stent
and coronary thrombosis: biological mechanisms and
clinical implications. Circulation 2007;115:1051-8.
4.	 Lansky AJ, Costa RA, Mintz GS, et al. Non-polymer-
based paclitaxel-coated coronary stents for the treatment
of patients with de novo coronary lesions: angiographic
follow-up of the DELIVER clinical trial. Circulation
2004;109:1948-54.
5.	 Waksman R, Pakala R, Baffour R, et al. In vivo comparison
of a polymer-free Biolimus A9-eluting stent with a
biodegradable polymer-based Biolimus A9 eluting stent
and a bare metal stent in balloon denuded and radiated
hypercholesterolemic rabbit iliac arteries. Catheter
Cardiovasc Interv 2012;80:429-36.
6.	 King L, Byrne RA, Mehilli J, et al. Five-year clinical
outcomes of a polymer-free sirolimus-eluting stent
versus a permanent polymer paclitaxel-eluting
stent: final results of the intracoronary stenting and
angiographic restenosis - test equivalence between
two drug-eluting stents (ISAR-TEST) trial. Catheter
Cardiovasc Interv 2013;81:E23-8.
7.	 Tada N, Virmani R, Grant G, et al. Polymer-free biolimus
a9-coated stent demonstrates more sustained intimal
inhibition, improved healing, and reduced inflammation
compared with a polymer-coated sirolimus-eluting
cypher stent in a porcine model. Circ Cardiovasc Interv
2010;3:174-83.
8.	 Costa JR Jr, Abizaid A, Costa R, et al. Preliminary results
of the hydroxyapatite nonpolymer-based sirolimus-
eluting stent for the treatment of single de novo coronary
lesions a first-in-human analysis of a third-generation
drug-eluting stent system. JACC Cardiovasc Interv
2008;1:545-51.
9.	 Schwartz RS, Edelman E, Virmani R, et al. Drug-
eluting stents in preclinical studies: updated consensus
recommendations for preclinical evaluation. Circ
Cardiovasc Interv 2008;1:143-53.
10.	 Lemos PA, Farooq V, Takimura CK, et al. Emerging
technologies: polymer-free phospholipid encapsulated
sirolimus nanocarriers for the controlled release of
drug from a stent-plus-balloon or a stand-alone balloon
catheter. EuroIntervention 2013;9:148-56.
11.	 Yasdani SN, Sherdiwala DP, Prakash S, et al. A temporal
assessment of drug distribution following local balloon
delivery of nanoparticle sirolimus. JACC 2011;58:B6.
12.	 Capodanno D, Dipasqua F, Tamburino C. Novel drug-
eluting stents in the treatment of de novo coronary lesions.
Vasc Health Risk Manag 2011;7:103-18.
13.	 Grube E, Hauptmann KE, Buellesfeld L, et al. Six-
month results of a randomized study to evaluate
safety and efficacy of a Biolimus A9 eluting stent with
a biodegradable polymer coating. EuroIntervention
2005;1:53-7.
121Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015
© Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org
Cite this article as: Takimura CK, Galon MZ, Gutierrez
PS, Sojitra P, Vyas A, Doshi M, Lemos PA. A new polymer-
free drug-eluting stent with nanocarriers eluting sirolimus
from stent-plus-balloon compared with bare-metal stent and
with biolimus A9 eluting stent in porcine coronary arteries.
Cardiovasc Diagn Ther 2015;5(2):113-121. doi: 10.3978/
j.issn.2223-3652.2015.03.08
14.	 Serruys PW, Farooq V, Kalesan B, et al. Improved
safety and reduction in stent thrombosis associated with
biodegradable polymer-based biolimus-eluting stents
versus durable polymer-based sirolimus-eluting stents in
patients with coronary artery disease: final 5-year report
of the LEADERS (Limus Eluted From A Durable Versus
ERodable Stent Coating) randomized, noninferiority trial.
JACC Cardiovasc Interv 2013;6:777-89.
15.	 Bezerra HG, Costa MA, Guagliumi G, et al. Intracoronary
optical coherence tomography: a comprehensive review
clinical and research applications. JACC Cardiovasc Interv
2009;2:1035-46.
16.	 Suzuki T, Kopia G, Hayashi S, et al. Stent-based delivery
of sirolimus reduces neointimal formation in a porcine
coronary model. Circulation 2001;104:1188-93.
17.	 Carter AJ, Aggarwal M, Kopia GA, et al. Long-term
effects of polymer-based, slow-release, sirolimus-eluting
stents in a porcine coronary model. Cardiovasc Res
2004;63:617-24.
18.	 Lowe HC, Schwartz RS, Mac Neill BD, et al. The porcine
coronary model of in-stent restenosis: current status in
the era of drug-eluting stents. Catheter Cardiovasc Interv
2003;60:515-23.
19.	 Kornowski R, Hong MK, Tio FO, et al. In-stent
restenosis: contributions of inflammatory responses and
arterial injury to neointimal hyperplasia. J Am Coll Cardiol
1998;31:224-30.
20.	 Grube E, Mueller R, Schuler G, et al. TCT-46
Comparison of Polymer-Free BioFreedom™ Stents with
Durable Polymer Taxus Liberté™ Stents: 3-Year Results
from the BioFreedom First-In-Man Trial. J Am Coll
Cardiol 2012;60. doi:10.1016/j.jacc.2012.08.055.
21.	 Carrié D, Berland J, Verheye S, et al. A multicenter
randomized trial comparing amphilimus- with paclitaxel-
eluting stents in de novo native coronary artery lesions. J
Am Coll Cardiol 2012;59:1371-6.
22.	 Byrne RA, Kufner S, Tiroch K, et al. Randomised trial
of three rapamycin-eluting stents with different coating
strategies for the reduction of coronary restenosis: 2-year
follow-up results. Heart 2009;95:1489-94.
23.	 Costa RA, Abizaid A, Abizaid AS, et al. Procedural and
early clinical outcomes of patients with de novo coronary
bifurcation lesions treated with the novel Nile PAX
dedicated bifurcation polymer-free paclitaxel coated
stents: results from the prospective, multicentre, non-
randomised BIPAX clinical trial. EuroIntervention
2012;7:1301-9.
24.	 Costa JR Jr, Abizaid A, Costa R, et al. 1-year results of
the hydroxyapatite polymer-free sirolimus-eluting stent
for the treatment of single de novo coronary lesions:
the VESTASYNC I trial. JACC Cardiovasc Interv
2009;2:422-7.
25.	 Gonzalo N, Serruys PW, García-García HM, et al.
Quantitative ex vivo and in vivo comparison of lumen
dimensions measured by optical coherence tomography
and intravascular ultrasound in human coronary arteries.
Rev Esp Cardiol 2009;62:615-24.
26.	 Murata A, Wallace-Bradley D, Tellez A, et al. Accuracy
of optical coherence tomography in the evaluation of
neointimal coverage after stent implantation. JACC
Cardiovasc Imaging 2010;3:76-84.

More Related Content

What's hot

Study of cell therapy assisted regeneration of cartilage
Study of cell therapy assisted regeneration of cartilageStudy of cell therapy assisted regeneration of cartilage
Study of cell therapy assisted regeneration of cartilageSourabh Harankhedkar
 
The sages manual of hernia repair
The sages manual of hernia repairThe sages manual of hernia repair
The sages manual of hernia repairSpringer
 
Presentation8 16 10[1]
Presentation8 16 10[1]Presentation8 16 10[1]
Presentation8 16 10[1]jfreshour
 
EAO Stockholm easygraft
EAO Stockholm easygraftEAO Stockholm easygraft
EAO Stockholm easygraftMinas Leventis
 
Orthopaedic tissue engineering
Orthopaedic tissue engineeringOrthopaedic tissue engineering
Orthopaedic tissue engineeringrajusvmc
 
Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...
Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...
Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...Jaideep Sahni
 
Simultaneous vertical guided bone regeneration and guided tissue regeneration...
Simultaneous vertical guided bone regeneration and guided tissue regeneration...Simultaneous vertical guided bone regeneration and guided tissue regeneration...
Simultaneous vertical guided bone regeneration and guided tissue regeneration...threea3a
 
V Bobic - Stem Cells - BKS Cardiff 030217
V Bobic - Stem Cells - BKS Cardiff 030217V Bobic - Stem Cells - BKS Cardiff 030217
V Bobic - Stem Cells - BKS Cardiff 030217Vladimir Bobic
 
Catalog Osteobiol 2020 EN
Catalog Osteobiol 2020 ENCatalog Osteobiol 2020 EN
Catalog Osteobiol 2020 ENterradent
 
Bone regeneration and substitutes
Bone regeneration and substitutes  Bone regeneration and substitutes
Bone regeneration and substitutes kyaw tint
 
EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...
EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...
EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...Abu-Hussein Muhamad
 
Periodontal regeneration
Periodontal regenerationPeriodontal regeneration
Periodontal regenerationVijeta12
 

What's hot (20)

Idr ppt
Idr pptIdr ppt
Idr ppt
 
Study of cell therapy assisted regeneration of cartilage
Study of cell therapy assisted regeneration of cartilageStudy of cell therapy assisted regeneration of cartilage
Study of cell therapy assisted regeneration of cartilage
 
Stemcells in Orthopaedic suergery.
Stemcells  in Orthopaedic suergery.Stemcells  in Orthopaedic suergery.
Stemcells in Orthopaedic suergery.
 
The sages manual of hernia repair
The sages manual of hernia repairThe sages manual of hernia repair
The sages manual of hernia repair
 
Presentation8 16 10[1]
Presentation8 16 10[1]Presentation8 16 10[1]
Presentation8 16 10[1]
 
Cellulose nanofiber
Cellulose nanofiberCellulose nanofiber
Cellulose nanofiber
 
EAO Stockholm easygraft
EAO Stockholm easygraftEAO Stockholm easygraft
EAO Stockholm easygraft
 
Journal club
Journal clubJournal club
Journal club
 
Orthopaedic tissue engineering
Orthopaedic tissue engineeringOrthopaedic tissue engineering
Orthopaedic tissue engineering
 
Iatriki etireia teliko
Iatriki etireia telikoIatriki etireia teliko
Iatriki etireia teliko
 
Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...
Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...
Effect of Seeding Density of Mesenchymal Stem Cells on Cell Proliferation and...
 
Simultaneous vertical guided bone regeneration and guided tissue regeneration...
Simultaneous vertical guided bone regeneration and guided tissue regeneration...Simultaneous vertical guided bone regeneration and guided tissue regeneration...
Simultaneous vertical guided bone regeneration and guided tissue regeneration...
 
V Bobic - Stem Cells - BKS Cardiff 030217
V Bobic - Stem Cells - BKS Cardiff 030217V Bobic - Stem Cells - BKS Cardiff 030217
V Bobic - Stem Cells - BKS Cardiff 030217
 
Jc 7
Jc 7Jc 7
Jc 7
 
Catalog Osteobiol 2020 EN
Catalog Osteobiol 2020 ENCatalog Osteobiol 2020 EN
Catalog Osteobiol 2020 EN
 
Bone regeneration and substitutes
Bone regeneration and substitutes  Bone regeneration and substitutes
Bone regeneration and substitutes
 
Dental Soft tissue regeneration
Dental Soft tissue regeneration Dental Soft tissue regeneration
Dental Soft tissue regeneration
 
EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...
EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...
EXTRACTION AND IMMEDIATE IMPLANT PLACEMENT USING A COMBINED PRF, AND PROVISIO...
 
Periodontal regeneration
Periodontal regenerationPeriodontal regeneration
Periodontal regeneration
 
Orthopedic applications of stem cells
Orthopedic applications of stem cells Orthopedic applications of stem cells
Orthopedic applications of stem cells
 

Similar to A new polymer-free drug-eluting stent with nanocarriers eluting sirolimus from stent-plus-balloon compared with bare-metal stent and with biolimus A9 eluting stent in porcine coronary arteries

New generations of drug eluting
New generations of drug elutingNew generations of drug eluting
New generations of drug elutingRamachandra Barik
 
Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...
Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...
Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...Chaichuk Sergiy
 
JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...
JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...
JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...Urvashi Sodvadiya
 
Stman et-al-2012-clinical_implant_dentistry_and_related_research
 Stman et-al-2012-clinical_implant_dentistry_and_related_research Stman et-al-2012-clinical_implant_dentistry_and_related_research
Stman et-al-2012-clinical_implant_dentistry_and_related_researchGantumur DDS, PhD
 
Technique for Placement of Oxidized Titanium Implants by Oded Bahat
Technique for Placement of Oxidized Titanium Implants by Oded BahatTechnique for Placement of Oxidized Titanium Implants by Oded Bahat
Technique for Placement of Oxidized Titanium Implants by Oded BahatOded Bahat
 
Biovascular scaffolds - current status 2015
Biovascular scaffolds -  current status 2015Biovascular scaffolds -  current status 2015
Biovascular scaffolds - current status 2015Anshul Kumar Gupta
 
Tissue engineering of stents
Tissue engineering of stentsTissue engineering of stents
Tissue engineering of stentsDr. sreeremya S
 
Supraflex ultra thin coronary stent DES
Supraflex ultra thin coronary stent DESSupraflex ultra thin coronary stent DES
Supraflex ultra thin coronary stent DESGennaro Broya de Lucia
 
Use of grafts & alloplastic material in maxillofacial trauma
Use of grafts & alloplastic material in maxillofacial traumaUse of grafts & alloplastic material in maxillofacial trauma
Use of grafts & alloplastic material in maxillofacial traumaDr. SHEETAL KAPSE
 
Coronarystents phpapp02
Coronarystents phpapp02Coronarystents phpapp02
Coronarystents phpapp02Saurabh Gupta
 
Vascular Closure Devices
Vascular Closure DevicesVascular Closure Devices
Vascular Closure DevicesAsad Moosa
 
Changes of the peri implant soft tissue thickness after grafting with collage...
Changes of the peri implant soft tissue thickness after grafting with collage...Changes of the peri implant soft tissue thickness after grafting with collage...
Changes of the peri implant soft tissue thickness after grafting with collage...Ashok Kumar
 
1362573078 dr. rajdeep agrawal
1362573078 dr. rajdeep agrawal1362573078 dr. rajdeep agrawal
1362573078 dr. rajdeep agrawaldfsimedia
 

Similar to A new polymer-free drug-eluting stent with nanocarriers eluting sirolimus from stent-plus-balloon compared with bare-metal stent and with biolimus A9 eluting stent in porcine coronary arteries (20)

New generations of drug eluting
New generations of drug elutingNew generations of drug eluting
New generations of drug eluting
 
Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...
Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...
Local drug delivery with balloons eluting a liquid paclitaxel formulation. Yo...
 
Biovascular scaffolds
Biovascular scaffolds Biovascular scaffolds
Biovascular scaffolds
 
JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...
JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...
JOURNAL CLUB: “Comparative Evaluation of Postoperative Pain and Success Rate ...
 
Stman et-al-2012-clinical_implant_dentistry_and_related_research
 Stman et-al-2012-clinical_implant_dentistry_and_related_research Stman et-al-2012-clinical_implant_dentistry_and_related_research
Stman et-al-2012-clinical_implant_dentistry_and_related_research
 
Technique for Placement of Oxidized Titanium Implants by Oded Bahat
Technique for Placement of Oxidized Titanium Implants by Oded BahatTechnique for Placement of Oxidized Titanium Implants by Oded Bahat
Technique for Placement of Oxidized Titanium Implants by Oded Bahat
 
The story of coronary stent
The story of coronary stentThe story of coronary stent
The story of coronary stent
 
revascularization.pdf
revascularization.pdfrevascularization.pdf
revascularization.pdf
 
Biovascular scaffolds - current status 2015
Biovascular scaffolds -  current status 2015Biovascular scaffolds -  current status 2015
Biovascular scaffolds - current status 2015
 
poster final
poster finalposter final
poster final
 
Tissue engineering of stents
Tissue engineering of stentsTissue engineering of stents
Tissue engineering of stents
 
Supraflex ultra thin coronary stent DES
Supraflex ultra thin coronary stent DESSupraflex ultra thin coronary stent DES
Supraflex ultra thin coronary stent DES
 
Use of grafts & alloplastic material in maxillofacial trauma
Use of grafts & alloplastic material in maxillofacial traumaUse of grafts & alloplastic material in maxillofacial trauma
Use of grafts & alloplastic material in maxillofacial trauma
 
Coronarystents phpapp02
Coronarystents phpapp02Coronarystents phpapp02
Coronarystents phpapp02
 
Vascular Closure Devices
Vascular Closure DevicesVascular Closure Devices
Vascular Closure Devices
 
20141213_04_FoinTR_AOP201507_v3
20141213_04_FoinTR_AOP201507_v320141213_04_FoinTR_AOP201507_v3
20141213_04_FoinTR_AOP201507_v3
 
Changes of the peri implant soft tissue thickness after grafting with collage...
Changes of the peri implant soft tissue thickness after grafting with collage...Changes of the peri implant soft tissue thickness after grafting with collage...
Changes of the peri implant soft tissue thickness after grafting with collage...
 
Biosolve II
Biosolve IIBiosolve II
Biosolve II
 
ISI
ISIISI
ISI
 
1362573078 dr. rajdeep agrawal
1362573078 dr. rajdeep agrawal1362573078 dr. rajdeep agrawal
1362573078 dr. rajdeep agrawal
 

More from Paul Schoenhagen

The World Hypertension League: where now and where to in salt reduction
The World Hypertension League: where now and where to in salt reductionThe World Hypertension League: where now and where to in salt reduction
The World Hypertension League: where now and where to in salt reductionPaul Schoenhagen
 
Developing a national salt reduction strategy for Mongolia
Developing a national salt reduction strategy for MongoliaDeveloping a national salt reduction strategy for Mongolia
Developing a national salt reduction strategy for MongoliaPaul Schoenhagen
 
Sodium content in processed foods in Argentina: compliance with the national law
Sodium content in processed foods in Argentina: compliance with the national lawSodium content in processed foods in Argentina: compliance with the national law
Sodium content in processed foods in Argentina: compliance with the national lawPaul Schoenhagen
 
Salt reduction and hypertension in China: a concise state-of-the-art review
Salt reduction and hypertension in China: a concise state-of-the-art reviewSalt reduction and hypertension in China: a concise state-of-the-art review
Salt reduction and hypertension in China: a concise state-of-the-art reviewPaul Schoenhagen
 
WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...
WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...
WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...Paul Schoenhagen
 
An overview of salt intake reduction efforts in the Gulf Cooperation Council ...
An overview of salt intake reduction efforts in the Gulf Cooperation Council ...An overview of salt intake reduction efforts in the Gulf Cooperation Council ...
An overview of salt intake reduction efforts in the Gulf Cooperation Council ...Paul Schoenhagen
 
Salt intake reduction efforts: advances and challenges
Salt intake reduction efforts: advances and challengesSalt intake reduction efforts: advances and challenges
Salt intake reduction efforts: advances and challengesPaul Schoenhagen
 
Goal attainments and their discrepancies for low density lipoprotein choleste...
Goal attainments and their discrepancies for low density lipoprotein choleste...Goal attainments and their discrepancies for low density lipoprotein choleste...
Goal attainments and their discrepancies for low density lipoprotein choleste...Paul Schoenhagen
 
Improved non-calcified plaque delineation on coronary CT angiography by sonog...
Improved non-calcified plaque delineation on coronary CT angiography by sonog...Improved non-calcified plaque delineation on coronary CT angiography by sonog...
Improved non-calcified plaque delineation on coronary CT angiography by sonog...Paul Schoenhagen
 
Hemodynamic assessment of partial mechanical circulatory support: data derive...
Hemodynamic assessment of partial mechanical circulatory support: data derive...Hemodynamic assessment of partial mechanical circulatory support: data derive...
Hemodynamic assessment of partial mechanical circulatory support: data derive...Paul Schoenhagen
 
Spontaneous coronary artery dissection—A review
Spontaneous coronary artery dissection—A reviewSpontaneous coronary artery dissection—A review
Spontaneous coronary artery dissection—A reviewPaul Schoenhagen
 
microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...
microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...
microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...Paul Schoenhagen
 
Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...Paul Schoenhagen
 
Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...
Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...
Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...Paul Schoenhagen
 
Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...Paul Schoenhagen
 
Cardiac imaging in prosthetic paravalvular leaks
Cardiac imaging in prosthetic paravalvular leaksCardiac imaging in prosthetic paravalvular leaks
Cardiac imaging in prosthetic paravalvular leaksPaul Schoenhagen
 

More from Paul Schoenhagen (16)

The World Hypertension League: where now and where to in salt reduction
The World Hypertension League: where now and where to in salt reductionThe World Hypertension League: where now and where to in salt reduction
The World Hypertension League: where now and where to in salt reduction
 
Developing a national salt reduction strategy for Mongolia
Developing a national salt reduction strategy for MongoliaDeveloping a national salt reduction strategy for Mongolia
Developing a national salt reduction strategy for Mongolia
 
Sodium content in processed foods in Argentina: compliance with the national law
Sodium content in processed foods in Argentina: compliance with the national lawSodium content in processed foods in Argentina: compliance with the national law
Sodium content in processed foods in Argentina: compliance with the national law
 
Salt reduction and hypertension in China: a concise state-of-the-art review
Salt reduction and hypertension in China: a concise state-of-the-art reviewSalt reduction and hypertension in China: a concise state-of-the-art review
Salt reduction and hypertension in China: a concise state-of-the-art review
 
WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...
WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...
WHO’s supported interventions on salt intake reduction in the sub-Saharan Afr...
 
An overview of salt intake reduction efforts in the Gulf Cooperation Council ...
An overview of salt intake reduction efforts in the Gulf Cooperation Council ...An overview of salt intake reduction efforts in the Gulf Cooperation Council ...
An overview of salt intake reduction efforts in the Gulf Cooperation Council ...
 
Salt intake reduction efforts: advances and challenges
Salt intake reduction efforts: advances and challengesSalt intake reduction efforts: advances and challenges
Salt intake reduction efforts: advances and challenges
 
Goal attainments and their discrepancies for low density lipoprotein choleste...
Goal attainments and their discrepancies for low density lipoprotein choleste...Goal attainments and their discrepancies for low density lipoprotein choleste...
Goal attainments and their discrepancies for low density lipoprotein choleste...
 
Improved non-calcified plaque delineation on coronary CT angiography by sonog...
Improved non-calcified plaque delineation on coronary CT angiography by sonog...Improved non-calcified plaque delineation on coronary CT angiography by sonog...
Improved non-calcified plaque delineation on coronary CT angiography by sonog...
 
Hemodynamic assessment of partial mechanical circulatory support: data derive...
Hemodynamic assessment of partial mechanical circulatory support: data derive...Hemodynamic assessment of partial mechanical circulatory support: data derive...
Hemodynamic assessment of partial mechanical circulatory support: data derive...
 
Spontaneous coronary artery dissection—A review
Spontaneous coronary artery dissection—A reviewSpontaneous coronary artery dissection—A review
Spontaneous coronary artery dissection—A review
 
microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...
microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...
microRNA-based diagnostics and therapy in cardiovascular disease—Summing up t...
 
Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...
 
Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...
Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...
Carotid intima-media thickness (CIMT) and carotid plaques in young Nepalese p...
 
Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...Impact of statins and beta-blocker therapy on mortality after coronary artery...
Impact of statins and beta-blocker therapy on mortality after coronary artery...
 
Cardiac imaging in prosthetic paravalvular leaks
Cardiac imaging in prosthetic paravalvular leaksCardiac imaging in prosthetic paravalvular leaks
Cardiac imaging in prosthetic paravalvular leaks
 

Recently uploaded

Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 

Recently uploaded (20)

Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 

A new polymer-free drug-eluting stent with nanocarriers eluting sirolimus from stent-plus-balloon compared with bare-metal stent and with biolimus A9 eluting stent in porcine coronary arteries

  • 1. © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org Introduction Permanent polymers of first generation drug-eluting stents (DES) are imputed to be a contributing factor in vessel wall inflammation, positive remodeling, stent malapposition and ultimately late stent thrombosis (1-3). New DES with biodegradable polymers and DES with no polymers have been developed and tested in preclinical and clinical experiments to overcome the limitations of first generation DES with durable polymers. Polymeric coatings was thought to be essential for effective local drug delivery to the vessel wall, since its absence has been related to target vessel failure (4). However, many recent studies with surface-modified polymer-free DES showed that the elimination of polymers might not compromise the anti-restenosis efficacy of DES (5,6). Accordingly, new polymer-free DES has been deve loped by a number of stent manufactures by texturizing, creating grooves, niches, or a microporous stent surface (7,8). This article describes the experimental results of the novel Focus np stent (Envision Scientific Pvt. Ltd., Gujarat, Original Article A new polymer-free drug-eluting stent with nanocarriers eluting sirolimus from stent-plus-balloon compared with bare-metal stent and with biolimus A9 eluting stent in porcine coronary arteries Celso K. Takimura1 , Micheli Z. Galon1 , Paulo S. Gutierrez1 , Prakash Sojitra2 , Ashwin Vyas2 , Manish Doshi2 , Pedro A. Lemos1 1 Department of Interventional Cardiology, Heart Institute (InCor), University of São Paulo Medical School, São Paulo, Brazil; 2 Envision Scientific Pvt. Ltd., Office 1-3, Silver Palm II, Opposite Snehmilan Garden, Kadampalli, Nanpura, Surat, Gujarat, India Correspondence to: Celso K. Takimura, MD, PhD. Heart Institute (InCor) University of São Paulo Medical School, Av. Dr. Eneas Carvalho de Aguiar 44, Bloco I, 3o andar, São Paulo 04503-000, Brazil. Email: celsotakimura@hotmail.com. Background: Permanent polymers in first generation drug-eluting stent (DES) have been imputed to be a possible cause of persistent inflammation, remodeling, malapposition and late stent thrombosis. We aim to describe the in vivo experimental result of a new polymer-free DES eluting sirolimus from stent-plus-balloon (Focus np stent, Envision Scientific) compared with a bare-metal stent (BMS) (Amazonia CroCo, Minvasys) and with a biolimus A9 eluting stent (Biomatrix, Biosensors). Methods: In 10 juvenile pigs, 23 coronary stents were implanted in the coronary arteries (8 Amazonia CroCo, 8 Focus np, and 7 Biomatrix). At 28-day follow-up, optical coherence tomography (OCT) and histology were used to evaluate neointimal hyperplasia and healing response. Results: According to OCT analysis, Focus np stents had a greater lumen area and less neointimal hyperplasia response than BMS and Biomatrix had. Histomorphometry results showed less neointimal hyperplasia in Focus np than in BMS. Histology showed a higher fibrin deposition in Biomatrix stent compared to Focus np and BMS. Conclusions: The new polymer-free DES with sirolimus eluted from stent-plus-balloon demonstrated safety and reduced neointimal proliferation compared with the BMS and Biomatrix stents at 28-day follow- up in this porcine coronary model. This new polymer-free DES is promising and warrants further clinical studies. Keywords: Neointimal; restenosis; stent; vascular injury Submitted Feb 09, 2015. Accepted for publication Mar 16, 2015. doi: 10.3978/j.issn.2223-3652.2015.03.08 View this article at: http://dx.doi.org/10.3978/j.issn.2223-3652.2015.03.08
  • 2. 114 Takimura et al. Polymer-free Focus np stent © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org India), a polymer-free stent developed with nanotechnology that uses nanosized phospholipid particles to deliver sirolimus from a combined balloon-plus-stent platform. Methods Animal study Animal experiments were in compliance with the Guide for the Care and Use of Laboratory Animals (Institute of Laboratory Animal Resources, Commission on Life Sciences and National Research Council. National Academy Press, Washington, D.C., 1996), and the study protocol was approved by the Institutional Ethics Committee. Stents were implanted in 10 juvenile domestic pigs, weighing 23 to 30 kg, and animals were followed for 28 days. Oral aspirin (100 mg) and clopidogrel (300 mg loading dose; 75 mg daily dose) were administered starting 1 day before the procedure and continued daily until euthanasia. Animals were sedated and anesthetized with an intramuscular solution of midazolam (0.5 mg/kg) and ketamine (3 mg/kg) followed by intravenous thiopental (25 mg/kg). Mechanical ventilation was established and anesthesia maintained with inhaled isoflurane (2% to 5%) throughout the procedure. Arterial access was achieved by a femoral cut-down with a general sterile technique, and coronary angiography was then performed using 6F guiding catheters after intracoronary administration of nitroglycerin (200 μg) and intravenous heparin (7,500 IU). Three types of stents were implanted: (I) cobalt-chromium polymer-free sirolimus eluting Focus np stent; (II) stainless steel biolimus A9-eluting Biomatrix stent (Biosensors International, Singapore) and (III) a cobalt- chromium bare metal stent (BMS) Amazonia CroCo® stent, (Minvasys, Paris, France). All stents were 3.0 mm in diameter, and stent length was 16 mm for polymer-free sirolimus eluting Focus np stent, 14 mm for biolimus A9 eluting Biomatrix stent, and 16 mm for Amazonia CroCo® BMS. For each animal, the study stents were deployed according to a randomization chart (1:1:1) in the major branches of the coronary arteries (left anterior descending coronary artery, left circumflex artery, and right coronary artery) in a non-tapered, non-bifurcation, non-angulated segment of 3.0 mm in diameter by visual estimation. The stents were implanted through inflation of the delivery balloons for approximately 30 seconds in all study groups. The guiding catheter was used as a reference to obtain a 1.1:1 to 1.2:1 stent-to-artery ratio compared with the baseline vessel diameter and to induce moderate vessel injury (9). Animals were allowed to recover and received appropriate postoperative care. At 28-day follow-up, control coronary angiography and optical coherence tomography (OCT) were performed, and the animals under deep anesthesia were euthanized with a lethal dose of potassium chloride. Design of Focus np stent The Focus np stent is a cobalt-chromium L-605 stent with strut thickness of 73 µm. It is crimped onto a COPAN Co-Polyamide delivery balloon. Nanoparticles with a mean diameter of 210 nm, containing sirolimus (108 µg of sirolimus on a 3.0×16 mm stent), were synthesized using an ultrasonic homogenizer and applied using an inert gas- assisted spray process in stents already crimped onto the balloon delivery system. Consequently, the coating was uniformly distributed over the stent-plus-balloon surface, with only the abluminal face of the stent struts having the coating applied (Figure 1). Focus np stent thus, is a polymer- free cobalt-chromium stent delivering sirolimus through nanoparticles from balloon and abluminal stent strut face. The manufacturing and coating process of sirolimus nanoparticles to metallic stent platform and the balloon- catheter delivery system is described in detail elsewhere (10). In a previous experiment in iliac arteries of rabbits treated with a 60 s dilatation with sirolimus-coated balloons, these nanoparticles were found initially in the endothelial surface, then in the medial layer and finally in the deeper stratum like the adventitia (11). In vitro drug release analyzed using HPLC revealed that sirolimus is delivered from the stent with an initial burst followed by a prolonged release for up to 40 days. In-tissue release of sirolimus from phospholipid Figure 1 Polymer-free Focus np stent assessed by scanning electron microscopy (SEM) (100× magnification).
  • 3. 115Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015 © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org nanoparticles is longer because of encapsulation in drug delivery matrix (10). Description of control stents Amazonia CroCo (Minvasys, France) is a cobalt-chromium BMS with a thin strut thickness of 73 µm and a design that combines intermediate and open cells. This stent is used as the metallic platform in the Amazonia Pax paclitaxel polymer-free DES (12). Biomatrix (Biosensors International, Singapore) is a DES that consists of a stainless steel platform with a strut thickness of 112 µm and a quadrature link design. It is coated exclusively in the abluminal stent surface with a biodegradable polymer (PLA, polylactic acid) and biolimus A9 (218 µg on a 3.0×14 mm stent). This biodegradable coating completely dissolves into carbon dioxide and water after a 6- to 9-month period. Biolimus A9 is a semi-synthetic sirolimus analogue with 10 times higher lipophilicity and similar potency as sirolimus (13,14). Optical coherence tomography (OCT) At 28-day follow-up and after control angiography, a 0.014- inch guidewire was inserted into the coronary arteries, and an occlusion balloon (Helios™, LightLab Imaging) was advanced distally to the treatment sites. The 0.014-inch wire was then removed and replaced for a 0.016-inch OCT catheter (Image Wire™, LightLab Imaging, Westford, MA, USA). The Helios™ balloon (LightLab® Imaging) was pulled back proximally to the stent and inflated with a 1:1 saline:contrast solution until occlusion of the artery. Then, saline was flushed to clear the lumen from blood and an automatic pullback of the OCT catheter was initiated at a speed of 1 mm/s with recording of OCT images. These images were analyzed for thrombus, strut malapposition and endothelialization. A strut was considered malapposed if the distance from the endoluminal surface of the stent to the vessel wall was higher than the sum of the metal and, if present, the polymer thickness. The cutoff points used for each stent type were: BMS (Amazonia CroCo) 73 µm, polymer-free sirolimus-eluting stent (Focus np) 73 µm, biolimus-eluting stent (Biomatrix) 120 µm. A stent strut was defined as covered, if there was a visible layer of tissue covering it. Semi-automatic measurements were made with proprietary software for off-line analysis with the M2 System (LightLab Imaging, Westford, MA, USA) to determine lumen area, stent area, neointimal thickness (distance from stent strut until lumen), neointimal area (stent area minus lumen area) and percentage of neointimal area (ratio of neointimal area and stent area, multiplied by 100) (15). Light microscopy Immediately after euthanasia, the hearts were excised and the coronaries were pressure-perfused (~100 mm Hg) via the ascending aorta with 0.9% saline, to clear the blood, followed by perfusion of 10% buffered formalin for a minimum of 1 hour. The stented arterial segments were then excised from the heart by dissection and fixed by immersion in 10% formalin overnight. The artery- stent specimens were dehydrated with ethanol solutions of increasing concentrations and embedded in methyl methacrylate resin. A total of three cross-sections (proximal, mid, and distal) were obtained from each vessel on a rotary microtome (cut thickness 3.5 μm) and stained with hematoxylin and eosin and Verhoeff stains for elastic fibers. Microscopic digital morphometry with Leica QWin® software (Leica Microsystems, Wetzlan, Germany) was used to measure areas of the external elastic lamina (EEL), internal elastic lamina (IEL), stent, and lumen. Media area was defined as EEL area—EL area. Neointimal hyperplasia area was defined as IEL area—lumen area. Percent neointimal burden was defined as 100× (neointimal area/ IEL area). Neointimal thickness was measured at the stent strut and at the inter-strut space. The cross-sections were examined for semi-quantitative analysis of endothelialization, inflammation, fibrin content, angiogenesis, and vessel injury according to previously described scoring schemes (16-19). Stent endothelialization score was defined as the extent of the circumference of the arterial lumen covered by endothelial cells and graded from 1 to 3 (1=25%, 2=25% to 75%, 3≥75%). Inflammation was graded as 0, none; 1, scattered inflammatory cells; 2, inflammatory cells encompassing 50% of a strut in at least 25% to 50% of the circumference of the artery; 3, inflammatory cells surrounding a strut in at least 25% to 50% of the circumference of the artery. Inflammation modified was graded as 0, no inflammatory cells surrounding the strut; 1, light, non-circumferential lymphohistocytic infiltrate surrounding the strut; 2, localized, moderate to dense cellular aggregate surrounding the strut non-circumferentially; 3, circumferential dense lymphohistiocytic cell infiltration of the strut. The intimal fibrin content was graded as 0, no fibrin deposition; 1,
  • 4. 116 Takimura et al. Polymer-free Focus np stent © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org focal residual fibrin involving any portion of the artery or moderate fibrin deposition adjacent to the strut involving <25% of the circumference of the artery; 2, moderate fibrin involving >25% of the circumference of the artery or heavy deposition involving <25% the circumference of the artery; or 3, heavy deposition of fibrin involving >25% of the circumference of the artery. The injury score (Schwartz) was graded as 0, internal elastic lamina intact, media compressed but not lacerated; 1, internal elastic lamina lacerated, media typically compressed but not lacerated; 2, internal elastic lacerated, media visibly lacerated, external elastic lamina intact but compressed or 3, external elastic lamina lacerated, typically large lacerations of media extending through the external elastic lamina, stent struts sometimes residing in adventitia. The injury score (Gunn) was graded as 0, no impression of metal upon media; 1, deformation of the internal elastic lamina by <45°; 2, deformation of the internal elastic lamina by >45°; 3, rupture of the internal elastic lamina or 4, rupture of the external elastic lamina (i.e., complete medial rupture). Statistical analysis Data are presented as mean ± SD and/or median (interquartile range; IQR). The analyzed per stent individual OCT frames and histologic cross-sections were treated as independent samples (7). Differences in continuous variables among stent groups were evaluated by using parametric (analysis of variance) or non-parametric (Kruskal-Wallis) statistical tests as needed. Histologic ordinal variable results were overall compared for differences among stent groups with independent-samples Kruskal-Wallis test. The statistical software SPSS 20.0 was used to analyze data, and a priori P value ≤0.05 was considered significant. Results Twenty-three stents (8 Amazonia CroCo, 8 Focus np, and 7 Biomatrix) were successfully implanted in the coronary arteries of 10 swine. Mean balloon pressure to implant the stents was equal among groups (P=0.895). Nine of 10 animals survived until the follow-up phase of the study. There was one early death (pig number 2) that occurred during the second day after the stent implantation procedure (Biomatrix stent implanted in the left anterior descending artery and Amazonia CroCo in the left circumflex artery). In this animal, a stent was implanted too distally in the coronary tree, resulting in dissection at the distal stent edge that was managed by prolonged balloon inflation with final TIMI III flow. Pathology revealed the presence of a recent intraluminal thrombus in all stents. The remaining 9 pigs with 21 stents (7 Amazonia CroCo, 8 Focus np, and 6 Biomatrix) were included in the study for final analysis. Stent distribution is summarized in Table 1. At 28-day follow-up, all stents in surviving animals were patent at angiography and could be imaged with OCT. OCT findings OCT quantitative analysis is summarized in Table 2, and illustrative example OCT images of each group are shown in Figure 2 (upper panel). At 28 days, lumen area was larger in Focus np compared with BMS and Biomatrix. Neointimal hyperplasia thickness was reduced in Focus np compared with BMS and Biomatrix. NIH area and percentage NIH were smaller in Focus np than BMS and Biomatrix. Biomatrix stent showed either a less-pronounced neointimal thickness or hyperplasia area than BMS. Stent struts were completely covered and well apposed in all stent groups. Histomorphometric and semiquantitative histologic analysis The histomorphometric analysis is summarized in Table 3, and illustrative example histologic images of each group are shown in Figure 2 (bottom panel). Lumen area, internal Table 1 Stent distribution Animal N° BMS Focus np Biomatrix 1 LAD RCA N/A 2* LCX N/A LAD 3 RCA LAD LCX 4 N/A LCX RCA 5 LAD RCA N/A 6 LCX N/A LAD 7 RCA LCX LAD 8 N/A LCX RCA 9 LAD RCA N/A 10 LCX RCA LAD Balloon pressure (ATM) 11.6±3.7 12.0±1.9 11.0±2.9 *, Early animal death, excluded from stent study analysis. BMS indicates Amazonia CroCo® . LAD, left anterior descending artery; LCX, left coronary circumflex artery; RCA, right coronary artery; N/A, not applicable.
  • 5. 117Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015 © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org Table 2 Optical coherence tomography quantitative analysis Variables BMS Focus np Biomatrix Overall P value Post-hoc comparisons, P<0.05 Stent, n 7 8 6 – – OCT frame images/ stent struts, n 63/728 72/707 54/499 – – Lumen area, mm² 4.93±1.71 5.38 (4.2-6.16) 6.58±1.65 6.90 (5.30-8.0) 5.94±1.49 6.49 (4.28-7.0) <0.001 Focus np > BMS & Biomatrix Stent area, mm² 7.70±1.14 7.35 (6.43-8.77) 7.96±0.97 7.97 (6.91-8.92) 7.76±0.98 8.04 (7.18-8.47) 0.705 – NIH thickness, mm 0.26±1.17 0.19 (0.10-0.38) 0.12±0.07 0.08 (0.07-0.19) 0.18±0.09 0.16 (0.11-0.26) <0.001 Focus np < BMS & Biomatrix; Biomatrix < BMS NIH area, mm² 2.76±1.66 1.92 (1.19-4.53) 1.38±0.91 1.0 (0.82-2.25) 1.81±0.80 1.81 (1.11-2.34) <0.001 Focus np < BMS & Biomatrix; Biomatrix < BMS Percent NIH, % 35.81±22.14 26.30 (16.19-51.89) 18.23±13.89 12.36 (10.33-28.83) 24.26±12.06 21.85 (13.78-36.89) <0.001 Focus np < BMS & Biomatrix; Biomatrix < BMS Uncovered struts 0 0 0 >0.999 Malapposed struts 0 0 0 >0.999 Data are presented as mean ± SD and median (interquartile range). NIH, neointimal hyperplasia. Figure 2 Representative OCT images (upper panel) and histologic images (bottom panel) 28 days after porcine coronary stent implantation of Amazonia CroCo BMS (left), Focus np stent (middle), and Biomatrix stent (right). OCT, optical coherence tomography; BMS, bare-metal stent. 1 mm 1 mm 1 mm 1 mm
  • 6. 118 Takimura et al. Polymer-free Focus np stent © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org elastic lamina area, external elastic lamina area, stent area, and neointimal thickness at strut and interstrut were not different among groups. The Focus np stent had a smaller neointimal area and percentage neointimal area compared with BMS only. Semiquantitative histologic analysis revealed that Focus np stents were associated with an intermediate degree of fibrin content, smaller than the Biomatrix but higher than the control BMS (Table 4). In addition, Focus np stents also had an intermediate degree of angiogenesis, smaller than Table 3 Histomorphometric analysis Variables BMS Focus np Biomatrix Overall P value Post hoc comparisons, P<0.05 Stents, n 7 8 6 Sections, n 20 22 15 Lumen area, mm2 3.36±1.01 3.41 (2.67-3.97) 3.83±1.05 4.06 (2.96-4.62) 3.69±1.13 3.69 (2.64-4.65) 0.363 N/A IEL area, mm2 4.67±0.89 4.61 (4.07-5.34) 4.85±0.97 5.10 (4.46-5.51) 5.08±0.85 5.17 (4.54-5.79) 0.430 N/A EEL area, mm2 5.83±0.93 5.66 (5.17-6.44) 5.88±1.11 6.25 (5.30-6.67) 6.06±1.01 6.19 (5.33-6.94) 0.797 N/A Stent area, mm2 4.92±0.78 4.79 (4.42-5.48) 4.95±0.94 5.25 (4.48-5.58) 5.22±0.86 5.41 (4.59-5.83) 0.556 N/A NIH area, mm2 1.56±0.60 1.55 (0.96-1.88) 1.11±0.50 0.97 (0.72-1.38) 1.52±0.52 1.53 (1.11-1.96) 0.021 Focus np < BMS Percent NIH, % 32.5±13.6 36.89 (18.89-39.25) 23.4±11.9 18.71 (14.45-29.77) 30.5±12.6 33.25 (18.82-42.45) 0.046 Focus np < BMS NIH thickness at strut, mm 0.18±0.10 0.20 (0.08-0.22) 0.15±0.12 0.10 (0.08-0.17) 0.17±0.08 0.15 (0.08-0.27) 0.591 N/A NIH thickness interstrut, mm 0.17±0.09 0.18 (0.09-0.21) 0.13±0.08 0.10 (0.07-0.18) 0.16±0.89 0.16 (0.06-0.24) 0.393 N/A Data are presented as mean ± SD and median (Interquartile range). IEL, indicates internal elastic lamina; EEL, external elastic lamina; NIH, neointimal hyperplasia; N/A, not applicable. Table 4 Semiquantitative histologic analysis Variables BMS Focus np Biomatrix Overall P value Stents, n 7 8 6 Sections, n 19 20 16 Endothelialization 3 3 3 >0.999 Inflammation 0.75±0.44 0.53±0.62 0.62±0.50 0.511 Inflammation modified 0.75±0.44 0.47±0.51 0.62±0.50 0.408 Fibrin 0.19±0.40 0.47±0.51 1.69±0.94 <0.001 Smooth muscle cells 3.00±0.00 2.76±0.75 2.62±0.65 0.070 Angiogenesis 1.56±0.81 1.41±1.06 0.77±0.59 0.025 Injury score (Schwartz) 1.25±0.57 1.00±0.35 1.54±0.51 0.017 Injury score (Gunn) 2.31±0.47 2.06±0.42 3.00±0.40 <0.001 BMS, bare-metal stent.
  • 7. 119Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015 © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org the BMS but higher than the Biomatrix stent. Even though the delivery pressure and the tentative balloon:artery ratio (1.1:1.0) was the same for all implants, the vessel injury score was higher in Biomatrix stent group. Discussion In this preclinical experiment, the polymer-free Focus np balloon-plus-stent coated with sirolimus-containing nanoparticles had an excellent safety profile, yielding good results at 28 days after implantation in pig coronaries evaluated by OCT and histology. Due to a unique patented manufacture process that allows Focus np to elute sirolimus from a stent-plus-balloon platform, therefore delivering drug to the entire vessel circumference, there is a possibility that this device might be more effective in inhibiting neointimal growth than existing commercial DES eluting antiproliferative drugs from stent surface alone. Indeed, in the present study, the antiproliferative effect of Focus np assessed by OCT was significantly higher than the clinically available Biomatrix DES, which elutes drug exclusively from the abluminal stent strut surface. Related to the comparison of the Focus np stent with the Biomatrix stent, although these two stents are markedly different regarding metallic platform, presence or absence of polymer, and differing antiproliferative agents, the Biomatrix stent data were useful as a “positive” control group, enabling us a direct comparison with this successful clinically used DES. A relevant finding in our study was the higher fibrin content in the histologic sections of the Biomatrix stent compared to the Focus np stent and to BMS. This suggests a more delayed healing process in the biolimus A9 eluting stent. Of note, in a preclinical experiment comparing the high-dose biolimus A9 stent (225 µg/14 mm), low-dose biolimus A9 (112 µg/14 mm) polymer-free Biofreedom stent (Biosensors International, Singapore), polymer- coated sirolimus-eluting Cypher stent, and BMS in porcine coronary arteries, histological evaluation showed a higher fibrin deposition for all these 3 DES groups than BMS at 28 days (7). Sirolimus is a lipophilic drug with potent antiproliferative and immunosuppressant effects. There are some technical difficulties in coating a balloon or a balloon-plus-stent and effectively delivering this drug into the arterial wall tissue without a durable or a biodegradable polymer. In a sirolimus eluting polymer-free DES, a fraction of the sirolimus load can be removed from the device when passing through the hemostatic valve, delivery catheter, calcified lesions, or be simply washed out by the blood stream. Encapsulation of sirolimus in nanoparticles to deliver it to the vessel wall through stent-plus-balloon is an attractive technology as the plaque area treated involves both balloon and stent area in contact with the vessel inner wall (stent edges included). Many polymer-free stents have been investigated in clinical studies, including BioFreedom (Biosensors International, Singapore) (20), Cre8 (CID, Italy) (21), Yukon Choice PF (Translumina, Germany) (22), Nile Pax (Minvasys, France) (23) and VESTAsync stent (MIV Therapeutics, USA) (24). The strategies for drug binding and release without the use of polymers consists of loading antiproliferative agents (biolimus A9, amphilimus, paclitaxel, sirolimus or a combination of sirolimus and probucol) on microporous or reservoirs on stent surface. The results of this preclinical study with Focus np stent warrants initiation of clinical studies as a First-in-Man (FIM) study and randomized clinical trials comparing it with currently approved DES. Study limitations The pig model of coronary stent implantation is well- known to be more suited for safety analysis than for efficacy evaluation of new devices. Therefore, the differences in neointimal proliferation seen among the groups should be interpreted with caution. The dissimilarities of the stent types and their respective delivery systems may account for some of the differences in the injury scores. It is not possible to rule out that some of the differences in the vascular response (e.g., neointimal proliferation, fibrin deposition) might have been related to a discrepancy in baseline vessel injury. An ideal experiment to assess the vessel response to this new technology—nanocarriers eluting sirolimus from stent-plus-balloon—would be testing stent types with exactly the same delivery balloons, metallic designs and stent lengths, and two DES releasing the same antiproliferative agent with the same dosage but different releasing drug mechanisms. The occlusive method (time-domain OCT, M2 System, LightLab Imaging, Westford, MA, USA) was used in this experimental study. This method, recently replaced by the current non-occlusive method (Fourier-domain OCT, C7XR, St. Jude Medical), was extensively used in preclinical and clinical studies. Comparing these two OCT methods in an in vivo non-stented human coronary artery study, the occlusive method tended to reduce lumen dimension by
  • 8. 120 Takimura et al. Polymer-free Focus np stent © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org 13%, with equivalent OCT image quality (25). In our study of stented porcine coronary arteries, the use of the occlusive method was well suited to analyze in-stent neointimal hyperplasia as the OCT NIH area and percentage NIH results were concordant with histomorphometry (26). Our sample size was small, but the carefully selected arterial segments to accommodate the unique diameter of stents (3.0 mm) available for implantation, and the analysis of OCT frames and histologic sections as independent samples reduced the need of a larger sample size. Conclusions The Focus np stent with polymer-free sirolimus eluted from stent-plus-balloon demonstrated safety and a reduced intimal proliferation compared with BMS and Biomatrix stent in the present 28-day follow-up porcine coronary model. The new nanocarrier technology eluting sirolimus from stent-plus-balloon is promising and warrants confirmation of its safety and efficacy in future clinical studies. Acknowledgements We thank Richard Silva, Leonora Loppnow and Marcio Chaves for technical assistance and Ann Morcos (www. morcosmedia.com) for reviewing the manuscript. The study was supported in part by Envision Scientific PVT Ltd. Disclosure: Celso K. Takimura, Micheli Z. Galon, Paulo S. Gutierrez and Pedro A. Lemos declare no conflict of interest. Prakash Sojitra, Ashwin Vyas and Manish Doshi works for Envision Scientific PVT Ltd. References 1. Virmani R, Guagliumi G, Farb A, et al. Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious? Circulation 2004;109:701-5. 2. Cook S, Ladich E, Nakazawa G, et al. Correlation of intravascular ultrasound findings with histopathological analysis of thrombus aspirates in patients with very late drug-eluting stent thrombosis. Circulation 2009;120:391-9. 3. Lüscher TF, Steffel J, Eberli FR, et al. Drug-eluting stent and coronary thrombosis: biological mechanisms and clinical implications. Circulation 2007;115:1051-8. 4. Lansky AJ, Costa RA, Mintz GS, et al. Non-polymer- based paclitaxel-coated coronary stents for the treatment of patients with de novo coronary lesions: angiographic follow-up of the DELIVER clinical trial. Circulation 2004;109:1948-54. 5. Waksman R, Pakala R, Baffour R, et al. In vivo comparison of a polymer-free Biolimus A9-eluting stent with a biodegradable polymer-based Biolimus A9 eluting stent and a bare metal stent in balloon denuded and radiated hypercholesterolemic rabbit iliac arteries. Catheter Cardiovasc Interv 2012;80:429-36. 6. King L, Byrne RA, Mehilli J, et al. Five-year clinical outcomes of a polymer-free sirolimus-eluting stent versus a permanent polymer paclitaxel-eluting stent: final results of the intracoronary stenting and angiographic restenosis - test equivalence between two drug-eluting stents (ISAR-TEST) trial. Catheter Cardiovasc Interv 2013;81:E23-8. 7. Tada N, Virmani R, Grant G, et al. Polymer-free biolimus a9-coated stent demonstrates more sustained intimal inhibition, improved healing, and reduced inflammation compared with a polymer-coated sirolimus-eluting cypher stent in a porcine model. Circ Cardiovasc Interv 2010;3:174-83. 8. Costa JR Jr, Abizaid A, Costa R, et al. Preliminary results of the hydroxyapatite nonpolymer-based sirolimus- eluting stent for the treatment of single de novo coronary lesions a first-in-human analysis of a third-generation drug-eluting stent system. JACC Cardiovasc Interv 2008;1:545-51. 9. Schwartz RS, Edelman E, Virmani R, et al. Drug- eluting stents in preclinical studies: updated consensus recommendations for preclinical evaluation. Circ Cardiovasc Interv 2008;1:143-53. 10. Lemos PA, Farooq V, Takimura CK, et al. Emerging technologies: polymer-free phospholipid encapsulated sirolimus nanocarriers for the controlled release of drug from a stent-plus-balloon or a stand-alone balloon catheter. EuroIntervention 2013;9:148-56. 11. Yasdani SN, Sherdiwala DP, Prakash S, et al. A temporal assessment of drug distribution following local balloon delivery of nanoparticle sirolimus. JACC 2011;58:B6. 12. Capodanno D, Dipasqua F, Tamburino C. Novel drug- eluting stents in the treatment of de novo coronary lesions. Vasc Health Risk Manag 2011;7:103-18. 13. Grube E, Hauptmann KE, Buellesfeld L, et al. Six- month results of a randomized study to evaluate safety and efficacy of a Biolimus A9 eluting stent with a biodegradable polymer coating. EuroIntervention 2005;1:53-7.
  • 9. 121Cardiovascular Diagnosis and Therapy, Vol 5, No 2 April 2015 © Cardiovascular Diagnosis and Therapy. All rights reserved. Cardiovasc Diagn Ther 2015;5(2):113-121www.thecdt.org Cite this article as: Takimura CK, Galon MZ, Gutierrez PS, Sojitra P, Vyas A, Doshi M, Lemos PA. A new polymer- free drug-eluting stent with nanocarriers eluting sirolimus from stent-plus-balloon compared with bare-metal stent and with biolimus A9 eluting stent in porcine coronary arteries. Cardiovasc Diagn Ther 2015;5(2):113-121. doi: 10.3978/ j.issn.2223-3652.2015.03.08 14. Serruys PW, Farooq V, Kalesan B, et al. Improved safety and reduction in stent thrombosis associated with biodegradable polymer-based biolimus-eluting stents versus durable polymer-based sirolimus-eluting stents in patients with coronary artery disease: final 5-year report of the LEADERS (Limus Eluted From A Durable Versus ERodable Stent Coating) randomized, noninferiority trial. JACC Cardiovasc Interv 2013;6:777-89. 15. Bezerra HG, Costa MA, Guagliumi G, et al. Intracoronary optical coherence tomography: a comprehensive review clinical and research applications. JACC Cardiovasc Interv 2009;2:1035-46. 16. Suzuki T, Kopia G, Hayashi S, et al. Stent-based delivery of sirolimus reduces neointimal formation in a porcine coronary model. Circulation 2001;104:1188-93. 17. Carter AJ, Aggarwal M, Kopia GA, et al. Long-term effects of polymer-based, slow-release, sirolimus-eluting stents in a porcine coronary model. Cardiovasc Res 2004;63:617-24. 18. Lowe HC, Schwartz RS, Mac Neill BD, et al. The porcine coronary model of in-stent restenosis: current status in the era of drug-eluting stents. Catheter Cardiovasc Interv 2003;60:515-23. 19. Kornowski R, Hong MK, Tio FO, et al. In-stent restenosis: contributions of inflammatory responses and arterial injury to neointimal hyperplasia. J Am Coll Cardiol 1998;31:224-30. 20. Grube E, Mueller R, Schuler G, et al. TCT-46 Comparison of Polymer-Free BioFreedom™ Stents with Durable Polymer Taxus Liberté™ Stents: 3-Year Results from the BioFreedom First-In-Man Trial. J Am Coll Cardiol 2012;60. doi:10.1016/j.jacc.2012.08.055. 21. Carrié D, Berland J, Verheye S, et al. A multicenter randomized trial comparing amphilimus- with paclitaxel- eluting stents in de novo native coronary artery lesions. J Am Coll Cardiol 2012;59:1371-6. 22. Byrne RA, Kufner S, Tiroch K, et al. Randomised trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis: 2-year follow-up results. Heart 2009;95:1489-94. 23. Costa RA, Abizaid A, Abizaid AS, et al. Procedural and early clinical outcomes of patients with de novo coronary bifurcation lesions treated with the novel Nile PAX dedicated bifurcation polymer-free paclitaxel coated stents: results from the prospective, multicentre, non- randomised BIPAX clinical trial. EuroIntervention 2012;7:1301-9. 24. Costa JR Jr, Abizaid A, Costa R, et al. 1-year results of the hydroxyapatite polymer-free sirolimus-eluting stent for the treatment of single de novo coronary lesions: the VESTASYNC I trial. JACC Cardiovasc Interv 2009;2:422-7. 25. Gonzalo N, Serruys PW, García-García HM, et al. Quantitative ex vivo and in vivo comparison of lumen dimensions measured by optical coherence tomography and intravascular ultrasound in human coronary arteries. Rev Esp Cardiol 2009;62:615-24. 26. Murata A, Wallace-Bradley D, Tellez A, et al. Accuracy of optical coherence tomography in the evaluation of neointimal coverage after stent implantation. JACC Cardiovasc Imaging 2010;3:76-84.