ICRP Publication 129
Authors on behalf of ICRP
M.M. Rehani, R. Gupta, S. Bartling, G. C. Sharp, R. Pauwels, T. Berris, J. M. Boone
2
3
 This slide set copies information from ICRP
Publication 129 in Main Points and in
Recommendations
 Readers of this slide set are advised to go through
the full publication for detailed information as this
material is only a glimpse, rather than being a
comprehensive presentation of contents
4
1. Introduction
2. CBCT Technology
3. The Biological Effects of Radiation
4. Principles of Radiological Protection for Patients and
Workers
5. Assessing Patient Doses in CBCT
6. Optimisation of Protection of Patients and Workers in
CBCT
7. Radiation Dose Management in Specific Applications of
CBCT
8. Training Considerations for CBCT
9. Quality Assurance Programmes
10. Recommendations
5
 CBCT extends the use of CT to areas that were not
typically associated with CT imaging in the past, e.g.
surgery, dental and otolaryngology (ear-nose-throat,
ENT) imaging, angiography suites, radiotherapy
treatment vaults, and orthopaedic imaging.
 The associated radiological protection issues are
substantially different from those of conventional CT.
 The perception that CBCT involves lower doses was
only true in initial applications. CBCT is now used
widely by specialists who have little or no training in
radiological protection.
6
ICRP P129 p.13 and p 9
 The manufacturers of CBCT scanners have
invested considerable effort into meeting the
electrical and mechanical safety requirements of
the users. Similar diligence is needed for issues
related to radiation dose and radiological
protection.
 This document provides a basis to develop
informed decisions and to direct the usage of
CBCT for optimising the trade-off between
clinical benefit and radiation risk.
7
 The ICRP emphasises that protection should be
optimised not only for whole-body exposures, but
also for exposures to specific tissues, especially
those of the lens of the eye, thyroid, breast, heart,
and cerebrovascular system.
 Equipment used for both fluoroscopy and CBCT
should provide aggregate dose indices for individual
patients during the entire procedure through
electronic display on the operator console and a
radiation dose structured report (RDSR file).
8
 Optimisation of both patient and worker doses,
particularly when the worker has to be near the modality,
is important wherein monitoring of doses become an
essential tool. Recording, reporting and tracking of
radiation dose for an individual patient should be made
possible in a consistent manner across vendors.
 Low dose protocols may be sufficient for diagnostic
procedures focused on high-contrast structures, such as
lung, bones, dental and maxillofacial scans, ENT scans
(paranasal sinuses, skull, temporal bone), interventional
material, and contrast-enhanced vessels (angiographic
interventions).
9
 Protocols with higher dose should only be
selected if visualisation of low-contrast
structures such as intracranial haemorrhage,
soft-tissue tumours, or abscesses are the
primary focus.
 Most interventional and intra-procedural C-arm
CBCT systems can scan an angular range
spanning 180o to 240o + the cone angle of the x-
ray beam. Localised critical organs, such as the
thyroid, eyes, female breasts and gonads, should
be on the “detector side” of the arc, whenever
possible. 10
 Clinical need permitting, every effort should be made
by users to ensure the volume of interest is fully
incorporated in the field of view (FOV) provided by
the CBCT scanners while radiosensitive organs are
placed outside the FOV.
 The aim of CBCT should be to answer a specific
diagnostic or intra-operative question vis-à-vis other
imaging modalities and not to obtain image quality
that rivals MDCT. The decision by the referring
practitioner to utilise CBCT should be made in
consultation with the imaging professional.
11
 There is a need to provide checks and balances,
for example, dose check alerts implemented in
CT in recent years, to avoid unintended high
patient exposure as compared to locally defined
reference values.
 Methods which provide reliable estimates of
patient eye dose under practical situations
should be established and utilised.
 The user of CBCT in interventions can influence
the radiation dose imparted to the patient
significantly by judiciously using a “low-image-
quality or low dose” vs. a “high-image-quality or
high dose” scan protocol.
12
 In radiotherapy, justified use of CBCT has
potential uses at different stages of therapy such
as: pre-treatment verification of patient position
and target volume localisation, evaluation of
non-rigid misalignments, such as flexion of the
spine or anatomic changes in soft tissue, and
during or after treatment to verify that the patient
position has remained stable throughout the
procedure. Low-dose CBCT protocols should be
used for pre-treatment alignment of bony
structures.
13
 Many machines were initially only capable of
fluoroscopy, but can now additionally perform CBCT.
Because of the improved clinical information in
CBCT and its ability to remove overlying structures,
the user may be tempted to over-utilise the CBCT
mode. Users should judiciously use CBCT mode.
 In orthopaedics, justified use of CBCT can help in
assessing the 2 dimensional position of fractures
and implants with respect to the bony anatomy,
especially in situations where fluoroscopy alone is
insufficient, and thus can help in patient dose
management.
14
 In urology, low-dose CBCT protocols should be
used when imaging high-contrast structures,
such as calcified kidney stones or device
placements.
 Dental and maxillofacial CBCT scans should be
justified, considering alternative imaging
modalities. Once justified, it should be optimised
to obtain images with minimal radiation dose
without compromising the diagnostic
information.
15
 The level of training in radiological protection
should be commensurate with the level of
expected radiation exposure (ICRP, 2009).
 All personnel intending to use CBCT for
diagnostic purposes should be trained in the
same manner as for diagnostic CT and those
intending to perform interventional CBCT should
be in trained in same manner as for
interventional CT.
16
1. Expanded availability and newer applications have
put CBCT technology in the hands of medical
professionals who do not traditionally use CT.
ICRP’s radiological protection principles and
recommendations provided in earlier publications,
particularly Publications 87 and 102 apply to these
newer applications and should be adhered to.
17
2. As many applications of CBCT involve patient
doses similar to MDCT, the room layout and
shielding requirements in such cases need to be
similar to protect workers adequately.
3. Medical practitioners bear the responsibility for
making sure that each CBCT examination is
justified and appropriate.
18
4. When referring a patient for a diagnostic CBCT
examination, the referring practitioner should be aware
of the strengths and weaknesses for CBCT vis-à-vis
MDCT, magnetic resonance imaging, and other
competing imaging modalities. The decision to utilise
CBCT should be made in consultation with an imaging
professional.
5. Manufacturers are challenged to practice standardised
methods for dosimetry and dose display in CBCT in
conformance with international recommendations such
as ICRU. Unfortunately, at present, there is wide
variation in dose quantities and units being displayed in
CBCT machines. The users are often unable to
compare doses among different scanners or protocols.
19
6. Use of CBCT systems for both fluoroscopy and
tomography pose new challenges in quantitating
radiation dose. There is a need to develop
methods that aggregate exposures to individual
patients during the entire procedure that may utilise
a combination of fluoroscopy and CBCT during a
given examination.
7. Recording, reporting and tracking of radiation dose
for an individual patient should be made possible in
a consistent manner across vendors.
20
8. There is a need to provide checks and balances
(e.g. dose check alerts implemented in CT in
recent years), to avoid unintended high patient
exposure as compared with locally defined
reference values.
9. Positioning radiosensitive organs such as the
thyroid, the lens of the eye, breasts and gonads on
the detector side during the partial rotation scan is
a useful feature in CBCT that needs to be utilised
for radiological protection of these organs.
21
10. Many machines were initially only capable of
fluoroscopy, but can now additionally perform
CBCT. Because of the improved clinical information
on CBCT, and its ability to remove overlying
structures, a user may be tempted to over utilise
the CBCT mode. Users must understand that the
CBCT function of their system is not a low-dose
“fluoroscopy run” and use this mode judiciously.
22
ICRP, 2015. Radiological Protection in Cone Beam
Computed Tomography (CBCT). ICRP Publication 129.
Ann. ICRP .
23
www.icrp.org
24

More Related Content

Similar to ICRP P129 Provides Guidance on Radiological Protection in CBCT

safe imaging.pptx
safe imaging.pptxsafe imaging.pptx
safe imaging.pptxAbinVl1
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTMelissa McClement
 
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaPatient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaRamzee Small
 
Clinical aspects and applications of high dose-rate brachytherapy
Clinical aspects and applications of high dose-rate brachytherapyClinical aspects and applications of high dose-rate brachytherapy
Clinical aspects and applications of high dose-rate brachytherapyTeresa Muñoz Migueláñez
 
Cbct is not the imaging technique of choice for comprehensive orthodontic ass...
Cbct is not the imaging technique of choice for comprehensive orthodontic ass...Cbct is not the imaging technique of choice for comprehensive orthodontic ass...
Cbct is not the imaging technique of choice for comprehensive orthodontic ass...Nielsen Pereira
 
Current and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutalaCurrent and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutalaKesho Conference
 
SOP CONFERENCE PROTOCOLS FOR BEGINNERS
SOP CONFERENCE PROTOCOLS FOR BEGINNERSSOP CONFERENCE PROTOCOLS FOR BEGINNERS
SOP CONFERENCE PROTOCOLS FOR BEGINNERSKanhu Charan
 
MR recommendations gec-estro
MR recommendations gec-estroMR recommendations gec-estro
MR recommendations gec-estroRadiaion Matters
 
Clinical quality assurance in Radiotherapy
Clinical quality assurance in RadiotherapyClinical quality assurance in Radiotherapy
Clinical quality assurance in RadiotherapyBharti Devnani
 
BRACHYTHERAPY IN ORAL CAVITY
BRACHYTHERAPY IN ORAL CAVITYBRACHYTHERAPY IN ORAL CAVITY
BRACHYTHERAPY IN ORAL CAVITYIsha Jaiswal
 
Trends in Radiation Protection of PET/CT imaging
Trends in Radiation Protection of PET/CT imagingTrends in Radiation Protection of PET/CT imaging
Trends in Radiation Protection of PET/CT imaging@Saudi_nmc
 
Brachytherapy IGABT
Brachytherapy IGABTBrachytherapy IGABT
Brachytherapy IGABTPuneet Seth
 
Radiation Oncology in 21st Century - Changing the Paradigms
Radiation Oncology in 21st Century - Changing the ParadigmsRadiation Oncology in 21st Century - Changing the Paradigms
Radiation Oncology in 21st Century - Changing the ParadigmsApollo Hospitals
 
12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO procedures12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO proceduresEuro CTO Club
 

Similar to ICRP P129 Provides Guidance on Radiological Protection in CBCT (20)

safe imaging.pptx
safe imaging.pptxsafe imaging.pptx
safe imaging.pptx
 
R0907 pa dosect_pr
R0907 pa dosect_prR0907 pa dosect_pr
R0907 pa dosect_pr
 
IGRT APP.pdf
IGRT APP.pdfIGRT APP.pdf
IGRT APP.pdf
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRT
 
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of GuyanaPatient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
 
Clinical aspects and applications of high dose-rate brachytherapy
Clinical aspects and applications of high dose-rate brachytherapyClinical aspects and applications of high dose-rate brachytherapy
Clinical aspects and applications of high dose-rate brachytherapy
 
Cbct is not the imaging technique of choice for comprehensive orthodontic ass...
Cbct is not the imaging technique of choice for comprehensive orthodontic ass...Cbct is not the imaging technique of choice for comprehensive orthodontic ass...
Cbct is not the imaging technique of choice for comprehensive orthodontic ass...
 
Trauma audit presentation
Trauma audit presentationTrauma audit presentation
Trauma audit presentation
 
Current and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutalaCurrent and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutala
 
SOP CONFERENCE PROTOCOLS FOR BEGINNERS
SOP CONFERENCE PROTOCOLS FOR BEGINNERSSOP CONFERENCE PROTOCOLS FOR BEGINNERS
SOP CONFERENCE PROTOCOLS FOR BEGINNERS
 
MR recommendations gec-estro
MR recommendations gec-estroMR recommendations gec-estro
MR recommendations gec-estro
 
Lung sbrt ppt
Lung  sbrt pptLung  sbrt ppt
Lung sbrt ppt
 
Clinical quality assurance in Radiotherapy
Clinical quality assurance in RadiotherapyClinical quality assurance in Radiotherapy
Clinical quality assurance in Radiotherapy
 
BRACHYTHERAPY IN ORAL CAVITY
BRACHYTHERAPY IN ORAL CAVITYBRACHYTHERAPY IN ORAL CAVITY
BRACHYTHERAPY IN ORAL CAVITY
 
Trends in Radiation Protection of PET/CT imaging
Trends in Radiation Protection of PET/CT imagingTrends in Radiation Protection of PET/CT imaging
Trends in Radiation Protection of PET/CT imaging
 
Brachytherapy IGABT
Brachytherapy IGABTBrachytherapy IGABT
Brachytherapy IGABT
 
IMRT
IMRTIMRT
IMRT
 
Radiation Oncology in 21st Century - Changing the Paradigms
Radiation Oncology in 21st Century - Changing the ParadigmsRadiation Oncology in 21st Century - Changing the Paradigms
Radiation Oncology in 21st Century - Changing the Paradigms
 
Elective ii
Elective iiElective ii
Elective ii
 
12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO procedures12:25 Vano - Decreasing X-ray exposure in CTO procedures
12:25 Vano - Decreasing X-ray exposure in CTO procedures
 

More from poojamuley7

Tobacco day.pptx
Tobacco day.pptxTobacco day.pptx
Tobacco day.pptxpoojamuley7
 
Lecture 14 Optimization Dental radiology.ppt
Lecture 14 Optimization Dental radiology.pptLecture 14 Optimization Dental radiology.ppt
Lecture 14 Optimization Dental radiology.pptpoojamuley7
 
summaryofcbct-161216070820.pptx
summaryofcbct-161216070820.pptxsummaryofcbct-161216070820.pptx
summaryofcbct-161216070820.pptxpoojamuley7
 
Doman - Numbers.pptx
Doman - Numbers.pptxDoman - Numbers.pptx
Doman - Numbers.pptxpoojamuley7
 
Doman - Farm Animals.pptx
Doman - Farm Animals.pptxDoman - Farm Animals.pptx
Doman - Farm Animals.pptxpoojamuley7
 
6 human body organs.pptx
6 human body organs.pptx6 human body organs.pptx
6 human body organs.pptxpoojamuley7
 
Panoramic-radiography.ppt
Panoramic-radiography.pptPanoramic-radiography.ppt
Panoramic-radiography.pptpoojamuley7
 

More from poojamuley7 (7)

Tobacco day.pptx
Tobacco day.pptxTobacco day.pptx
Tobacco day.pptx
 
Lecture 14 Optimization Dental radiology.ppt
Lecture 14 Optimization Dental radiology.pptLecture 14 Optimization Dental radiology.ppt
Lecture 14 Optimization Dental radiology.ppt
 
summaryofcbct-161216070820.pptx
summaryofcbct-161216070820.pptxsummaryofcbct-161216070820.pptx
summaryofcbct-161216070820.pptx
 
Doman - Numbers.pptx
Doman - Numbers.pptxDoman - Numbers.pptx
Doman - Numbers.pptx
 
Doman - Farm Animals.pptx
Doman - Farm Animals.pptxDoman - Farm Animals.pptx
Doman - Farm Animals.pptx
 
6 human body organs.pptx
6 human body organs.pptx6 human body organs.pptx
6 human body organs.pptx
 
Panoramic-radiography.ppt
Panoramic-radiography.pptPanoramic-radiography.ppt
Panoramic-radiography.ppt
 

Recently uploaded

💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Miss joya
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoybabeytanya
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call girls in Ahmedabad High profile
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night EnjoyCall Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoynarwatsonia7
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 

Recently uploaded (20)

💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night EnjoyCall Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
Call Girls Yelahanka Bangalore 📲 9907093804 💞 Full Night Enjoy
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 

ICRP P129 Provides Guidance on Radiological Protection in CBCT

  • 1. ICRP Publication 129 Authors on behalf of ICRP M.M. Rehani, R. Gupta, S. Bartling, G. C. Sharp, R. Pauwels, T. Berris, J. M. Boone
  • 2. 2
  • 3. 3
  • 4.  This slide set copies information from ICRP Publication 129 in Main Points and in Recommendations  Readers of this slide set are advised to go through the full publication for detailed information as this material is only a glimpse, rather than being a comprehensive presentation of contents 4
  • 5. 1. Introduction 2. CBCT Technology 3. The Biological Effects of Radiation 4. Principles of Radiological Protection for Patients and Workers 5. Assessing Patient Doses in CBCT 6. Optimisation of Protection of Patients and Workers in CBCT 7. Radiation Dose Management in Specific Applications of CBCT 8. Training Considerations for CBCT 9. Quality Assurance Programmes 10. Recommendations 5
  • 6.  CBCT extends the use of CT to areas that were not typically associated with CT imaging in the past, e.g. surgery, dental and otolaryngology (ear-nose-throat, ENT) imaging, angiography suites, radiotherapy treatment vaults, and orthopaedic imaging.  The associated radiological protection issues are substantially different from those of conventional CT.  The perception that CBCT involves lower doses was only true in initial applications. CBCT is now used widely by specialists who have little or no training in radiological protection. 6 ICRP P129 p.13 and p 9
  • 7.  The manufacturers of CBCT scanners have invested considerable effort into meeting the electrical and mechanical safety requirements of the users. Similar diligence is needed for issues related to radiation dose and radiological protection.  This document provides a basis to develop informed decisions and to direct the usage of CBCT for optimising the trade-off between clinical benefit and radiation risk. 7
  • 8.  The ICRP emphasises that protection should be optimised not only for whole-body exposures, but also for exposures to specific tissues, especially those of the lens of the eye, thyroid, breast, heart, and cerebrovascular system.  Equipment used for both fluoroscopy and CBCT should provide aggregate dose indices for individual patients during the entire procedure through electronic display on the operator console and a radiation dose structured report (RDSR file). 8
  • 9.  Optimisation of both patient and worker doses, particularly when the worker has to be near the modality, is important wherein monitoring of doses become an essential tool. Recording, reporting and tracking of radiation dose for an individual patient should be made possible in a consistent manner across vendors.  Low dose protocols may be sufficient for diagnostic procedures focused on high-contrast structures, such as lung, bones, dental and maxillofacial scans, ENT scans (paranasal sinuses, skull, temporal bone), interventional material, and contrast-enhanced vessels (angiographic interventions). 9
  • 10.  Protocols with higher dose should only be selected if visualisation of low-contrast structures such as intracranial haemorrhage, soft-tissue tumours, or abscesses are the primary focus.  Most interventional and intra-procedural C-arm CBCT systems can scan an angular range spanning 180o to 240o + the cone angle of the x- ray beam. Localised critical organs, such as the thyroid, eyes, female breasts and gonads, should be on the “detector side” of the arc, whenever possible. 10
  • 11.  Clinical need permitting, every effort should be made by users to ensure the volume of interest is fully incorporated in the field of view (FOV) provided by the CBCT scanners while radiosensitive organs are placed outside the FOV.  The aim of CBCT should be to answer a specific diagnostic or intra-operative question vis-à-vis other imaging modalities and not to obtain image quality that rivals MDCT. The decision by the referring practitioner to utilise CBCT should be made in consultation with the imaging professional. 11
  • 12.  There is a need to provide checks and balances, for example, dose check alerts implemented in CT in recent years, to avoid unintended high patient exposure as compared to locally defined reference values.  Methods which provide reliable estimates of patient eye dose under practical situations should be established and utilised.  The user of CBCT in interventions can influence the radiation dose imparted to the patient significantly by judiciously using a “low-image- quality or low dose” vs. a “high-image-quality or high dose” scan protocol. 12
  • 13.  In radiotherapy, justified use of CBCT has potential uses at different stages of therapy such as: pre-treatment verification of patient position and target volume localisation, evaluation of non-rigid misalignments, such as flexion of the spine or anatomic changes in soft tissue, and during or after treatment to verify that the patient position has remained stable throughout the procedure. Low-dose CBCT protocols should be used for pre-treatment alignment of bony structures. 13
  • 14.  Many machines were initially only capable of fluoroscopy, but can now additionally perform CBCT. Because of the improved clinical information in CBCT and its ability to remove overlying structures, the user may be tempted to over-utilise the CBCT mode. Users should judiciously use CBCT mode.  In orthopaedics, justified use of CBCT can help in assessing the 2 dimensional position of fractures and implants with respect to the bony anatomy, especially in situations where fluoroscopy alone is insufficient, and thus can help in patient dose management. 14
  • 15.  In urology, low-dose CBCT protocols should be used when imaging high-contrast structures, such as calcified kidney stones or device placements.  Dental and maxillofacial CBCT scans should be justified, considering alternative imaging modalities. Once justified, it should be optimised to obtain images with minimal radiation dose without compromising the diagnostic information. 15
  • 16.  The level of training in radiological protection should be commensurate with the level of expected radiation exposure (ICRP, 2009).  All personnel intending to use CBCT for diagnostic purposes should be trained in the same manner as for diagnostic CT and those intending to perform interventional CBCT should be in trained in same manner as for interventional CT. 16
  • 17. 1. Expanded availability and newer applications have put CBCT technology in the hands of medical professionals who do not traditionally use CT. ICRP’s radiological protection principles and recommendations provided in earlier publications, particularly Publications 87 and 102 apply to these newer applications and should be adhered to. 17
  • 18. 2. As many applications of CBCT involve patient doses similar to MDCT, the room layout and shielding requirements in such cases need to be similar to protect workers adequately. 3. Medical practitioners bear the responsibility for making sure that each CBCT examination is justified and appropriate. 18
  • 19. 4. When referring a patient for a diagnostic CBCT examination, the referring practitioner should be aware of the strengths and weaknesses for CBCT vis-à-vis MDCT, magnetic resonance imaging, and other competing imaging modalities. The decision to utilise CBCT should be made in consultation with an imaging professional. 5. Manufacturers are challenged to practice standardised methods for dosimetry and dose display in CBCT in conformance with international recommendations such as ICRU. Unfortunately, at present, there is wide variation in dose quantities and units being displayed in CBCT machines. The users are often unable to compare doses among different scanners or protocols. 19
  • 20. 6. Use of CBCT systems for both fluoroscopy and tomography pose new challenges in quantitating radiation dose. There is a need to develop methods that aggregate exposures to individual patients during the entire procedure that may utilise a combination of fluoroscopy and CBCT during a given examination. 7. Recording, reporting and tracking of radiation dose for an individual patient should be made possible in a consistent manner across vendors. 20
  • 21. 8. There is a need to provide checks and balances (e.g. dose check alerts implemented in CT in recent years), to avoid unintended high patient exposure as compared with locally defined reference values. 9. Positioning radiosensitive organs such as the thyroid, the lens of the eye, breasts and gonads on the detector side during the partial rotation scan is a useful feature in CBCT that needs to be utilised for radiological protection of these organs. 21
  • 22. 10. Many machines were initially only capable of fluoroscopy, but can now additionally perform CBCT. Because of the improved clinical information on CBCT, and its ability to remove overlying structures, a user may be tempted to over utilise the CBCT mode. Users must understand that the CBCT function of their system is not a low-dose “fluoroscopy run” and use this mode judiciously. 22
  • 23. ICRP, 2015. Radiological Protection in Cone Beam Computed Tomography (CBCT). ICRP Publication 129. Ann. ICRP . 23