LEAK CURRENT HiTESTER
ST5540, ST5541
Leak Current Measurement - Essential to Electrical Safety
•	Uninterrupted polarity switching function dramatically reduces cycle time
•	Support for rated currents up to 20 A gives the instrument more than adequate capability for testing
products designed to comply with new standards
For medical-use electrical devices and essential to electrical safety
Compliance with IEC 60601-1:2005 Ed 3.0, JIS T 0601-1:2012
	 (*Starting on June 1, 2012, medical electrical equipment sold in the EU must comply). Model ST5540 comply with IEC 60601-1:2005
+A1:2012 (Ed 3.1), and IEC 62353 of 2017
Compliance with Electrical Appliances and Materials Safety Act,
	 JIS, IEC, and UL standards for general-use electrical devices
2
General-use
electrical devices
(Medical) IEC standards
IEC 60601-1: 2005 +A1: 2012
IEC 60601-1: 1988 A2: 1995, etc.
(Medical) JIS standards
JIS T0601-1: 2012 supplement 1: 2014, 1999
■ Comparison of ST5540 Functionality
ST5540 compliance
ST5540
Medical-use
electrical devices
JIS
standards
Electrical
Appliances and
Materials Safety Act
IEC
standards
UL
standards
In order to prevent the danger of electric shock, electrical devices use power supplies that are isolated from parts of
the device that may come into contact with the body. However, it is impossible to achieve infinite insulation resistance.
Some leak current always exists, and its magnitude changes as the insulation degrades over time. The LEAK CURRENT
HiTESTER ST5540/ST5541 provides an easy-to-operate solution for measuring leak current in electrical devices,
making it eminently suitable for use in an extensive array of applications, ranging from production lines to equipment
maintenance and inspections.
*Compliance with medical IEC standards became mandatory in June 2012.
*Medical JIS standards were revised in June 2012.
Complies with all standards (suitable for use with all networks)
Leak current parameters as defined for medical-use electrical devices include ground leak current, contact current, patient leak
current, and patient measurement current. The ST5540 provides a single solution for measuring all of these leak current variants.
For Medical Electrical Devices
Complies with IEC 60601-1:2005 +A1:2012 (Ed 3.1)
ST5540
Measurement mode Category Standard compliance
●	 Patient leak current (between parts of device that come into contact with
patient and ground)
●	 Patient leak current (external SIP/SOP voltage)
●	 Patient leak current (external voltage at specific F-type applied part)
●	 Patient leak current (current resulting from external voltage at parts of
device that come into contact with patients)
●	 Patient measurement current
●	 Total patient leak current (between parts of device that come into contact
with patient and ground)
●	 Total patient leak current (external SIP/SOP voltage)
●	 Total patient leak current (external voltage at specific F-type applied part)
●	 Total patient leak current (current resulting from external voltage at parts
of device that come into contact with patient)
•	 Medical industry (Japan Association for
Clinical Engineering Technologists, etc.)
•	 Medical device manufacturers and dealers
•	 Medical device repair and maintenance
businesses
•	 Hospitals
•	 IEC60601-1 (Ed 3.1)
•	 IEC60990
•	 IEC62353
●	 Contact current (between device enclosure and lines)
●	 Contact current (between device enclosure and ground)
●	 Contact current (between device enclosure and device enclosure)
●	 Ground leak current
●	 Free current measurement
•	 Public agencies
•	 Electric vehicle manufacturers
•	 Manufacturers of general electrical devices
•	 Household appliance industry
•	 Information device industry
•	 Electrical vehicle standards UL
2231-1 and UL 2231-2
•	 Electrical Appliances and
Materials Safety Act
•	 IEC, JIS, and UL standards
*The ST5540 also complies with old standards.
3
Electrical Appliances and Materials Safety Act
Ministerial ordinance defining technical standards for electrical devices
UL standards
UL2231-1: 2002, UL2231-1: 2012 (revision 2016)
UL2231-2: 2002, UL2231-2: 2012 (revision 2016)
UL1492: 1996 (revision 2013)
UL492: 1996, etc.
IEC standards
IEC60990: 2016
IEC61010-1: 2010 +A1: 2016
IEC60950-1: 2005 +A1: 2009 +A2: 2013
IEC60065: 2014
JIS standards
JIS B8561: (2007), JIS C9250: 1992 +A1: 2007, etc.
ST5541 compliance
Measurement of General-use Electrical Devices
ST5541 provides standard support for standard-compliant networks (excluding medical-use electrical devices).
For Standard- and Regulation-compliance
ST5541
A single, robust solution for leak
current measurement
ST5541
General-use
electrical devices
■ Comparison of ST5541 Functionality
Measurement mode Category Standard compliance
●	 Contact current (between device enclosure and lines)
●	 Contact current (between device enclosure and ground)
●	 Contact current (between device enclosure and device enclosure)
●	 Ground leak current
●	 Free current measurement
•	 Public agencies
•	 Electric vehicle manufacturers
•	 Manufacturers of general electrical devices
•	 Household appliance industry
•	 Information device industry
•	 Electrical vehicle standards UL
2231-1 and UL 2231-2
•	 Electrical Appliances and Materials
Safety Act
•	 IEC, JIS, and UL standards
There are various standards in place concerning networks (body simulated resistance), and a standard-compliant
network is required in order to make measurements.
4
The ST5540/ST5541 can switch polarity without stopping
the supply of power to the device under measurement.
Old models require that the device under measurement
be turned off and then back on again when switching
polarity, but the ST5540 and ST5541 let you progress
smoothly to the next testing process.
■ Safety conductor current measurement function
The ST5540/ST5541 can perform safety conductor current
measurement as defined in standards such as IEC 60990 and IEC 60950-1.
■ Automatic measurement functionality
Simple operation allows you to switch power supply polarity
and automatically make measurements with the target device
in the normal and single-fault states, displaying the peak
values. You can also set the measurement time and wait
time. These capabilities help reduce operation time.
ST5540/ST5541 Features
■ Uninterrupted polarity switching function
The ability to conduct tests without turning off the power when
switching the power supply polarity dramatically reduces cycle times.
■ Improved test reliability
Blown fuse check function
When measurement starts, the instrument checks for unintentional
probe misalignment using of a preconfigured lower limit setting.
L
N
E
*Only the ground is
switched, allowing
an uninterrupted
supply of power.
AC
■ 110% voltage application jack
The instrument’s 110% voltage application jack,
which is used during testing of medical devices,
outputs the target device line power supply voltage
as-is. The polarity can be switched (ST5540 only).
■ Circuit breaker for device under measurement
The instrument’s workbench-type design features a terminal block and a circuit
breaker on the front panel, making it deal for embedding in test lines and
simplifying connectivity with the device being measured, even while rack-mounted.
■ Simple, interactive operation
The ST5540/ST5541 uses a touch panel that lets you
configure settings by touching selections in response to
information displayed on the panel, keeping operation simple.
Peak value display
Displays the type of power supply fault and
the peak value for the leak current, which
varies with target device operation.
Allowable value
The maximum allowable value under the
standard in question is automatically set.
Settings can also be changed as desired by
the user.
Data storage
Measurement data:
For up to 100 target devices
Measurement conditions:
Up to 30 sets
Current measured value Judgment result based on set
allowable value
Power supply polarity/device status/measurement current
[Measurement screen]
Measurement data (peak values) can be stored in
the instrument’s built-in memory. Saved data can
be checked on the stored data reference screen
after measurement is complete. Data can be
stored for up to 100 test targets, with each target
being identified by a registered device name and
control number. Additionally, the instrument can
store a maximum of 2,000 peak value data points.
Together, these capabilities eliminate the need to
jot down measured values at the measurement site.
■ Save measurement data for 100 devices
■ Ability to store up to 30 sets of measurement conditions
The instrument can save and load up to 30 sets of
measurement conditions, allowing you to immediately
switch between conditions.
ST5540
ST5541
*Requires isolation transformer. (not included)
5
ST5540 internal circuitry
Connection terminals
( Line power supply terminal block for device under test )
Expandability for the Future
S10 terminal : Switch for connecting a function ground terminal to the measurement power supply
system’s ground point
Allows connectivity to ground to be configured during leak current measurement.
S12 terminal : Switch for connecting the parts of device that come into contact with the patient to
the measurement power supply circuit’s ground point
Allows connectivity to ground to be configured during leak current measurement.
S13 terminal : Switch for connecting contactable metallic parts that are not protectively grounded to
the ground line
Allows connectivity to ground to be configured during leak current measurement.
E terminal : Connected to the E (earth) LINE IN terminal.
This terminal is always connected and cannot be configured.
■ All the switch terminals needed for standard-compliant
measurement
The ST5540/ST5541 provides the terminals needed to perform leak current
measurement in compliance with IEC 60601-1 and JIS T0601-1, eliminating the need
for the user to provide external switches.
S10
S12
S13
E
Power supply polarity switching unit
Circuit for simulating a single wire break
in the power supply
Circuit for simulating a
ground wire break
Device under
measurement
AC
*S10, S12, S13, and E are available on the ST5540 only.
■ Standard USB interface
The ST5540/ST5541’s standard USB interface simplifies auto-
matic testing on manufacturing lines and in similar installa-
tions.
■ Separation of the instrument’s power supply and target
device lines
The instrument’s power supply and target device line power supply
are separated, helping prevent damage due to the inadvertent input of
an incorrect supply voltage. There’s no need to change the ST5540/
ST5541’s supply voltage, even if the target device’s supply voltages
changes.
■ Support for rated currents of up to 20 A
The ST5540/ST5541 supports currents of up to 20 A and voltages of up to 250
V. Its ability to accommodate large currents allows it to be used with a more
extensive range of target products, including devices from new fields such as
electric vehicles and household appliances.
Start of measurement and loading of measurement conditions can be controlled from
an external source. Additionally, judgment results, test signals, and other data can
be output, making it possible to use the instrument to develop an automated line.
In addition to outputting judgment results for each measurement
item, the instrument also provides T-FAIL output, which is generated
continuously once a FAIL result is encountered during automatic testing.
■External control via EXT I/O EXT I/O description
Input signals Active-low input
Max. pplied voltage EXT.DCV terminal input voltage
High level EXT.DCV terminal input voltage or open
Low level 0.3 VDC or less
Output signal Open collector output
Max. load voltage 24 VDC (when not using the EXT.DCV terminal)
Max. output current 60 mA DC per signal (low level)
●Output
TEST : Outputs low continuously during automatic measurement.
MEAS : Outputs the measurement count automatic measurement and measurement of multiple items.
PASS : Outputs the PASS judgment result for each measurement item.
FAIL : Outputs the FAIL judgment result for each measurement item.
LOW : Generates continuous output once a low signal is encountered during automatic testing.
T-FAIL : Generates continuous output once a FAILresult is encountered during automatic testing.
INT.DCV : Generates internal 5 VDC output (not isolated from internal circuitry).
INT.GND : Generates internal GND output (same as the case ground level).
●Input
START : Starts automatic measurement at low.
STOP : Forcibly terminates measurement at low.
LOAD (0 to 4) : Loads saved panels (30 panels).
EXT.DCV : Accepts external power supply input from 5 VDC to 24 VDC.
EXT.COM : Accepts external COM input.
KEYLOCK : Disables switches other than the start switch.
Inlet
Outlet
F 1 : Connectivity not supported for USB memory. Data communications only.
2 : Cannot be used while an RS-232C connection is active.
■ Standard RS-232C port
The ST5540/ST5541’s standard RS-232C port can be used to
control the instrument from a computer and to print data using
the 9442 printer (option).
6
■ Leak current measurement unit
ST5540/ST5541 Specifications
■ Network (body simulated resistance)
■ General specifications
*1 When using AC measurement mode, the high-pass filter frequency characteristics (fc = 4 Hz)
are added.
*2 ST5540 only.
■ Accuracy (current measurement unit)
■ Temperature and humidity range within which accuracy is guaranteed: 23°C ±5°C, 80% RH or less, non-condensing
■ Temperature coefficient: Add 0.1 × basic accuracy × (T-23) for operating temperature T (°C) ■ Warm-up time: 20 min
■ The range within which accuracy is guaranteed when using Network D and Network F (full-scale value for each range) is approximately 1/1.5 and 1/2, respectively.
■ Calculated value when the voltage is detected across both ends of a network consisting of non-inductive resistance with a theoretical value of 1 kΩ
■ The following accuracy values also apply when using voltmeter mode.
● Measurement mode: AC*1 / AC+DC ● Measurement: mode DC
● Measurement mode: AC peak*3
● Current monitor accuracy (Measurement methods: Average value
response, rms calculation)
● Voltage monitor accuracy
Range
Guaranteed
accuracy range
Resolution
Accuracy
0.1Hz≤f<15Hz*2 DC≤f≤100kHz 100kHz<f≤1MHz
50.00mA From 4 mA 10 μA
±(4.0%rdg.+10dgt.) ±(2.0%rdg.+6dgt.) ±(2.0%rdg.+10dgt.)
5.000mA From 400 μA 1 μA
500.0μA From 40 μA  0.1 μA
50.00μA From 4 μA 0.01 μA ±4.0%f.s. ±2.0%f.s. ±2.0%f.s.
Range
Guaranteed
accuracy range
Resolution
Accuracy
15Hz≤f≤10kHz 10kHz<f≤100kHz 100kHz<f≤1MHz
75.0mA From 8 mA 100 μA
±(2.0%rdg.+6dgt.)
±5.0%f.s.
±15.0%f.s.
10.00mA From 0.8 mA 10 μA
1.000mA From 100 μA  1 μA ±2.5%f.s.
500.0 μA From 40 μA 0.1 μA ±2.5%f.s. ±20.0%f.s.
Range
Guaranteed
accuracy range
Resolution Accuracy
50.00mA From 4 mA 10 μA
±(2.0%rdg.+6dgt.)
5.000mA From 400 μA 1 μA
500.0 μA From 40 μA 0.1 μA
50.00 μA From 4 μA 0.01 μA ±2.0%f.s.
Range
Guaranteed accuracy
range
Resolution Accuracy
300.0 V 85 V*4 to 275V 0.1 V ±(5.0%rdg.+10dgt.)
Range
Guaranteed accuracy
range
Resolution Accuracy
300.0 V From 85 V*5 0.1 V ±(2.0%rdg.+5dgt.)
Measurement current DC / AC / AC+DC / AC peak
Allowable measurement current Max. 50 mA (DC / AC / AC+DC mode)
Max. 75 mA (AC peak mode)
Measurement ranges DC / AC / AC+DC mode:50 μA/500 μA/5 mA/50 mA
AC peak mode:500 μA/1 mA/10 mA/75 mA
Range switching AUTO/HOLD
Trigger method Manual: Generates trigger automatically internally, free-run measurement.
Automatic: Starts measurement based on external start signal.
Measurement terminals T1 terminal, T2 terminal (with built-in fuse holder), T3
terminal (110% voltage application terminal: ST5540 only)
(*Step-up isolation transformer required for 110% application.)
Measurement methods Measurement of voltage drop across body simulated resistance points
Calculation and display of current values
True rms measurement
Measurement unit floats relative to instrument ground.
A/D conversion method ΔΣ method (20-bit)
Instrument-to-ground capacitance 200 pF or less (between T1/T2 terminal and case ground)
Input resistance 1 MΩ ±1% (single-end input)
Notincludingvoltagemeasurementunit,bodysimulatedresistance(currentdetectioncircuit)
Input capacitance 150 pF or less (between T1 and T2 terminals)
(f = 100 kHz, isolated network circuit, including cables)
CMRR (between T1 and
T2 terminals and case)
60 dB or greater at 60 Hz / 60 dB or greater at 10 kHz
40 dB or greater at 100 kHz / 40 dB or greater at 1 MHz
(Isolated from network circuit with fuse shorted)
● Medical-use electrical devices:
Network B (ST5540 only)
Basic measurement element: 1 kΩ
Filter: 10 kΩ + 15 nF
● Electrical Appliances and
Materials Safety Act: Network A
Basic measurement element: 1 kΩ
Filter: 10 kΩ + 11.22 nF + 579 Ω
● IEC 60990: Network C Basic measurement element: 1.5 kΩ + 500 Ω
Filter 1: 10 kΩ + 22 nF
Filter 2: 10 kΩ + (20 kΩ + 6.2 nF) //9.1 nF
● UL: Network D Basic measurement element: 1.5 kΩ //0.15 μF
● General-purpose 1: Network E Basic measurement element: 1 kΩ
● General-purpose 2: Network F Basic measurement element: 2 kΩ
● IEC 61010-1: Network G Basic measurement element: 375 Ω + 500 Ω
Filter: 375 Ω //0.22 μF + 500 Ω
● Safety conductor current Basic measurement element (35 Ω)
Outlet max. allowable
leak current
50 mA
Dielectric strength Between power supply terminals and protective ground: 1.39 kVAC (5 mA), 15 sec
Between measurement terminals and power supply terminals: 2.30 kVAC (10 mA), 15 sec
Between measurement terminals and control circuit: 2.30 kVAC (10 mA), 15 sec
Standard compliance EMC: EN 61326
Safety: EN 61010
Conductive RF 3% f.s. or less at 3 V
Magnetic field effects (Representative value when conducting measurements in
the AC 500 μA range)
Accessories ST5540: One set of L2200 test leads (one red, one black) + one
red L2200 test lead
ST5541: One set of L2200 test leads (one red, one black)
Enclosure Probe 9195 × 1 , CD-ROM × 1 (USB Driver)
power cord × 3 (1 for instrument and 2 for measuring instrument line supply use)
spare fuse × 1 (250 V F 50 mAL, measurement use)
Dimensions Approx. 320 (W) × 110 (H) × 253 (D) mm
Mass Approx. 4.5 kg
Range
Guaranteed accuracy
range
Resolution
Accuracy
DC,15Hz≤f≤100kHz 100kHz<f≤1MHz
50.00 mA 12.00 mAto 50.00 mA 10 μA ±(2.0%rdg.+6dgt.) ±(5.0%rdg.+20dgt.)
10.00mA 1.30 mAto 13.00 mA 10 μA ±(2.0%rdg.+6dgt.) ±(5.0%rdg.+20dgt.)
Range
Guaranteed accuracy
range
Resolution
Accuracy
15Hz≤f≤10kHz 1kHz<f≤100kHz 100kHz<f≤1MHz
75.0 mA 12.0 mAto 75.00 mA 100 μA ±(2.0%rdg.+6dgt.) ±5.0%f.s. ±25.0%f.s.
10.00 mA 1.30 mAto 13.00 mA 10 μA ±2.5%f.s. ±5.0%f.s. ±25.0%f.s.
■ Safety conductor current accuracy
● Measurement mode: DC / AC*4 / AC+DC ● Measurement mode: AC peak
*3 Setting not available with Network A, B, or C (when filter off).
*4 Voltages of less than 80 V are displayed as “Less than 80 V.”
*5 Currents of less than 0.5 A are displayed as “Less than 0.5 A.”
(Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Display 320 × 240 dot matrix LCD (with backlight)
Control 6 × 6 matrix touch panel
Operating temperature and
humidity range
0°C to 40°C, 80% RH or less (non-condensing)
Storage temperature and
humidity range
-10°C to 50°C, 80% RH or less (non-condensing)
Temperature and humidity
range within which accuracy
is guaranteed
23°C ±5°C, 80% RH or less (non-condensing)
Guaranteed accuracy period 1 year
Operating location Indoor use at an elevation not exceeding 2,000 m
Instrument power supply 100/120/220/240 VAC, as specified by customer
Rated power supply frequency: 50/60 Hz
Rated power: 30 VA
Line power supply for device
being measured and outlet
Rated supply voltage: 100 to 250 VAC
Rated power supply frequency: 50/60 Hz
Rated current: Input, terminal block: 20 A
Output, terminal block: 20 A
7
AC GROUNDING HiTESTER 3157 Protective Ground Tester Indispensable for Standards
Certification
Standard compliance (examples)
● IEC 60065
● IEC 60335-1
● IEC 60601-1
● IEC 60950-1
● IEC 61010-1
● JIS-C1010-1
● Electrical Appliances and Materials
Safety Act
● UL (various applicable standards)
Main ground line
Ground branch line
Ground branch line
Medical outlet
Medical ground terminal
Grounding electrode
Medical ground
center
Medical room
■ Measuring between the grounding center and grounding terminal
Verify that the electrical resistance between the medical outlet’s grounding
electrode connector or medical ground terminal and medical ground center is
less than or equal to 0.1 Ω by applying a current of approximately 25 A with an
AC current with a no-load voltage of 6 V or less and measuring the resistance
using the voltage droop method.
*	This measurement requires an extension cable (available separately). The extension cable is a
special-order item; please contact your HIOKI distributor for more information.
Combination of Instruments for Leak Current Testing and
Safety Conductor Testing
The following are key parts of any safety inspection of electrical equipment:
•	 Leak current test: Measure with the ST5540 and ST5541.
•	 Safety conductor test (also known as a ground line resistance test or ground conductor test): Measure
with the 3157.
The 3157 can also be used for conducting measurements under the JIS T 1022:2006 safety standard for
hospital electrical equipment.
For JIS T 1022:2006 Measurement at Hospitals Electrical devices
Safety Standard for Hospital Electrical Equipment
3157
option
option
Printer
options
PRINTER 9442
	 Forprintingnumerical
values112mm(4.41in)
paperwidth
AC ADAPTER
9443-02
	 ForthePrinter9442,
EUtype
CONNECTION CABLE
9446
	 ForthePrinter9442,25pin
-9pin,1.5m(4.92ft)length
*When using the Printer 9442, also purchase the RS-232C Interface 9593-02
Other
SAFETYTESTDATA
MANAGEMENTSOFTWARE9267
	 For PC control application
software
Input/Output
cords
CURRENTAPPLYPROBE9297
	 With switch, 1.48m (4.86 ft)
length
CURRENT PROBE 9296
	 Alligator clip, 1.45m (4.76 ft)
length
REMOTE CONTROL
BOX (DUAL) 9614
	 For Start/Stop control,
1.5m (4.92 ft) cord length
REMOTE CONTROL
BOX (SINGLE) 9613
	 For Start/Stop control,
1.5m (4.92 ft) cord length
*The 3157-01’s handshake function cannot be used with the RS-232C Cable 9638
GP-IB CONNECTOR
CABLE 9151-02
	2 m (6.56 ft) length
GP-IB INTERFACE
9518-02
	For the 3157-01, built in
type
RS-232C INTERFACE
9593-02
	For the 3157-01, built in
type, not CE marked
RECORDING PAPER
1196
	 ForthePrinter9442,112mm
(4.41in)×25m(82.03ft),10
rolls/set
•	 Easily perform protective continuity testing in compliance with international
safety standards and laws
	 -1) Protective continuity resistance measurement for medical devices and gen-
eral electrical devices
	 -2) Ground connectivity testing when installing electrical machine tools and dis-
tribution panels
	 -3) Testing of protective grounding and isopotential grounding work for medical
equipment
	 -4) Evaluation of contact status using large currents
•	 Feedback control system that is capable of applying a stable current even with a
fluctuating load
•	 Soft-start function that checks the connection to the device under test before
applying the current
EXT I/O
Interface insertion slot
Model : AC GROUNDING HiTESTER 3157
Model No. (Order Code) (Note)
3157-01 (100-120 / 200-240 VAC switching)
Note: This instrument is not capable of performing measurement by itself. Please purchase
two Current probe 9296 units or one Current probe 9296 and one Current apply probe 9297,
depending on your measurement application.:
HEADQUARTERS
	81 Koizumi
	 Ueda, Nagano 386-1192 Japan
	www.hioki.com
HIOKI USA CORPORATION
	 TEL +1-609-409-9109 FAX +1-609-409-9108
	 hioki@hiokiusa.com / www.hiokiusa.com
All information correct as of May 21, 2018. All specifications are subject to change without notice. ST5540E12-85E Printed in Japan
DISTRIBUTED BY
HIOKI (Shanghai) SALES & TRADING CO., LTD.
	 TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
	 info@hioki.com.cn / www.hioki.cn
HIOKI SINGAPORE PTE.LTD.
	 TEL +65-6634-7677 FAX +65-6634-7477
	 info-sg@hioki.com.sg / www.hioki.com.sg
HIOKI KOREA CO., LTD.
	 TEL +82-2-2183-8847 FAX +82-2-2183-3360
	 info-kr@hioki.co.jp / www.hiokikorea.com
HIOKI EUROPE GmbH
	 TEL +49-6173-31856-0 FAX +49-6173-31856-25
	 hioki@hioki.eu / www.hioki.com
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
● Measurement date ● Allowable values
● Instrument name ● Maximum value
● Control number ● Judgment results
● Class (applied part) ● Measurement current (AC, DC, AC+DC, AC peak)
● Network ● Power supply polarity (normal, reversed)
● Measurement mode ● Instrument status (normal, ground line broken)
● Filter settings
The optional PRINTER 9442 can be used to print data via the instrument’s RS-232C interface, providing a convenient way to attach a
hard copy of test data.
Print method: Thermal serial dot
Paper width/print speed: 112 mm/52.5 cps
Power supply: AC ADAPTER 9443 or included nickel-
metal-hydride battery (good for approx. 3,000 lines of
print when fully charged with the 9443 adapter)
Dimensions: Approx. 160 (W) × 66.5 (H) × 170 (D) mm
Mass: Approx. 580 g
*CONNECTION CABLE 9444 and AC ADAPTER 9443 are required in order to connect the 9442 printer.
■ Printing saved data
Saved measurement data is displayed (pressing the print key within the
same data unit causes all data in the data unit to be printed).
Printable data (printed data can be selected from the following)
PRINTER 9442
Convenience
■ Leak current tester supplies
*Some standards require use of an isolation transformer.
Product inquiries should be directed to:
Isolation transformer model numbers
	 100 to 110 V (Japan): HSW-2KSP
	 240 to 264 V (overseas): HSW-5KSP
For more information: Tokyo Rikosha Co., Ltd.
Phone: +81-48-856-3851 (reception)
http://www.tokyorikosha.co.jp
Standards require use of an isolation transformer when
measuring medical-use electrical devices. Please purchase a
transformer with the necessary rated capacity.
Isolation transformer
Safety
standards
Example printout
Date
Instrument name
Control number
Connection class and applied part
Network
Measurement mode
Filter
Allowable value (upper limit)
Allowable value (lower limit)
Maximum value
Judgment result
Polarity
Status
Model : LEAK CURRENT HiTESTER ST5540
Model No. (Order Code) (Note)
ST5540 (For medical-use and electrical devices)
Model : LEAK CURRENT HiTESTER ST5541
Model No. (Order Code) (Note)
ST5541 (For electrical devices)
PC
communication
RS-232C CABLE 9637
	 For the PC, 9pin - 9pin,
cross, 1.8m (5.91 ft) length
RS-232C CABLE 9638
	 For the PC, 9pin - 25pin,
cross, 1.8m (5.91 ft) length
Printer
options
PRINTER 9442
	 Forprintingnumerical
values112mm(4.41in)
paperwidth
RECORDING
PAPER 1196
	 For the Printer 9442,
112 mm (4.41 in) × 25 m
(82.03 ft), 10 rolls/set
AC ADAPTER
9443-02
	 ForthePrinter9442,
EUtype
CONNECTION
CABLE 9444
	 ForthePrinter9442,
9pin-9pin,1.5m
(4.92ft)length
Test
probes
*The L2200 and the 9195 are bundled
ENCLOSURE
PROBE 9195
	 For the ST5540 series,
3156/3155
TEST LEAD L2200
	 70 cm (2.30ft) length, detachable
large alligator clips or needle tips are
bundled, CAT IV 600V, CAT III 1000V
ST5540, ST5541 shared options

More Related Content

What's hot

Gtec AVR(Stabilizer)
Gtec AVR(Stabilizer)Gtec AVR(Stabilizer)
Gtec AVR(Stabilizer)
Jerry Ashong
 
Iec 61439 for_low_voltage_switchgear_and
Iec 61439 for_low_voltage_switchgear_andIec 61439 for_low_voltage_switchgear_and
Iec 61439 for_low_voltage_switchgear_and
ssuserfd2587
 

What's hot (19)

Practical Medium and High Voltage Testing of Electrical Equipment for Enginee...
Practical Medium and High Voltage Testing of Electrical Equipment for Enginee...Practical Medium and High Voltage Testing of Electrical Equipment for Enginee...
Practical Medium and High Voltage Testing of Electrical Equipment for Enginee...
 
Electronic Calibrator
Electronic CalibratorElectronic Calibrator
Electronic Calibrator
 
Instrumentation
InstrumentationInstrumentation
Instrumentation
 
IRJET - Overheat Protection of Induction Motor
IRJET - Overheat Protection of Induction MotorIRJET - Overheat Protection of Induction Motor
IRJET - Overheat Protection of Induction Motor
 
ZIPPTORK Torque controller system for air tools
ZIPPTORK Torque controller system for air toolsZIPPTORK Torque controller system for air tools
ZIPPTORK Torque controller system for air tools
 
What is test method of impulse limiting voltage on signal SPD (impulse voltag...
What is test method of impulse limiting voltage on signal SPD (impulse voltag...What is test method of impulse limiting voltage on signal SPD (impulse voltag...
What is test method of impulse limiting voltage on signal SPD (impulse voltag...
 
IRJET- Patient Monitoring System
IRJET- Patient Monitoring SystemIRJET- Patient Monitoring System
IRJET- Patient Monitoring System
 
Electrical measuring instruments
Electrical measuring instrumentsElectrical measuring instruments
Electrical measuring instruments
 
ZIPPTORK Torque control system
ZIPPTORK Torque control systemZIPPTORK Torque control system
ZIPPTORK Torque control system
 
Benefits of using iec 61439 standard in electrical busbar systems
Benefits of using iec 61439 standard in electrical busbar systemsBenefits of using iec 61439 standard in electrical busbar systems
Benefits of using iec 61439 standard in electrical busbar systems
 
Temperature controllerai208v76
Temperature controllerai208v76Temperature controllerai208v76
Temperature controllerai208v76
 
ZIPPTORK Torque Control System - 2019 May update
ZIPPTORK Torque Control System - 2019 May updateZIPPTORK Torque Control System - 2019 May update
ZIPPTORK Torque Control System - 2019 May update
 
PTP35
PTP35PTP35
PTP35
 
Benefits of using iec 61439 standard in electrical busbar systems
Benefits of using iec 61439 standard in electrical busbar systemsBenefits of using iec 61439 standard in electrical busbar systems
Benefits of using iec 61439 standard in electrical busbar systems
 
Best Electronics Measurement Equipments Manufacturers In Pune, India
Best Electronics Measurement Equipments Manufacturers In Pune, IndiaBest Electronics Measurement Equipments Manufacturers In Pune, India
Best Electronics Measurement Equipments Manufacturers In Pune, India
 
Voltmeter Calibration
Voltmeter CalibrationVoltmeter Calibration
Voltmeter Calibration
 
Condition Monitoring of Rotating Equipment Considering the Cause and Effects ...
Condition Monitoring of Rotating Equipment Considering the Cause and Effects ...Condition Monitoring of Rotating Equipment Considering the Cause and Effects ...
Condition Monitoring of Rotating Equipment Considering the Cause and Effects ...
 
Gtec AVR(Stabilizer)
Gtec AVR(Stabilizer)Gtec AVR(Stabilizer)
Gtec AVR(Stabilizer)
 
Iec 61439 for_low_voltage_switchgear_and
Iec 61439 for_low_voltage_switchgear_andIec 61439 for_low_voltage_switchgear_and
Iec 61439 for_low_voltage_switchgear_and
 

Similar to Hioki st5540 e12-85e_leakage_current_tester

Electrical hazards2 (1)
 Electrical hazards2 (1) Electrical hazards2 (1)
Electrical hazards2 (1)
jcpdrz
 
power quality-analyzer-pw3198-90
power quality-analyzer-pw3198-90power quality-analyzer-pw3198-90
power quality-analyzer-pw3198-90
Nanan Ginanjar
 

Similar to Hioki st5540 e12-85e_leakage_current_tester (20)

Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
Hioki pq3198 pq3100_e4-11m_power_quality_analyzer
 
3102.1
3102.13102.1
3102.1
 
Dec2001 contactor
Dec2001   contactorDec2001   contactor
Dec2001 contactor
 
01 - Contactor - Fuji Electric
 01 - Contactor - Fuji Electric 01 - Contactor - Fuji Electric
01 - Contactor - Fuji Electric
 
F04844043
F04844043F04844043
F04844043
 
TESCO Tuesday: Test Switch Operation & Accessories
TESCO Tuesday: Test Switch Operation & AccessoriesTESCO Tuesday: Test Switch Operation & Accessories
TESCO Tuesday: Test Switch Operation & Accessories
 
Electrical hazards2 (1)
 Electrical hazards2 (1) Electrical hazards2 (1)
Electrical hazards2 (1)
 
power quality-analyzer-pw3198-90
power quality-analyzer-pw3198-90power quality-analyzer-pw3198-90
power quality-analyzer-pw3198-90
 
Hioki pw3198 power_quality_analyzer_datasheet
Hioki pw3198 power_quality_analyzer_datasheetHioki pw3198 power_quality_analyzer_datasheet
Hioki pw3198 power_quality_analyzer_datasheet
 
Electrical Hazards in the Workplace
Electrical Hazards in the WorkplaceElectrical Hazards in the Workplace
Electrical Hazards in the Workplace
 
Thông số kĩ thuật siemens monitoring devices
Thông số kĩ thuật siemens monitoring devicesThông số kĩ thuật siemens monitoring devices
Thông số kĩ thuật siemens monitoring devices
 
M4100
M4100M4100
M4100
 
Catalog ELCB Fuji Electric dòng G-TWIN Mới Nhất
Catalog ELCB Fuji Electric dòng G-TWIN Mới NhấtCatalog ELCB Fuji Electric dòng G-TWIN Mới Nhất
Catalog ELCB Fuji Electric dòng G-TWIN Mới Nhất
 
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
 
Dienhathe.com monitoring devices
Dienhathe.com monitoring devicesDienhathe.com monitoring devices
Dienhathe.com monitoring devices
 
Biomedical electrical system testin.pdf
Biomedical electrical system testin.pdfBiomedical electrical system testin.pdf
Biomedical electrical system testin.pdf
 
08 - Circuit Protector - Fuji Electric
08 - Circuit Protector - Fuji Electric08 - Circuit Protector - Fuji Electric
08 - Circuit Protector - Fuji Electric
 
Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413
Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413
Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413
 
Catalog Contactor Fuji Electric - Beeteco.com
Catalog Contactor Fuji Electric - Beeteco.comCatalog Contactor Fuji Electric - Beeteco.com
Catalog Contactor Fuji Electric - Beeteco.com
 
Catalog contactor fuji-electric Mới Nhất - Hạo Phương
Catalog contactor fuji-electric Mới Nhất - Hạo PhươngCatalog contactor fuji-electric Mới Nhất - Hạo Phương
Catalog contactor fuji-electric Mới Nhất - Hạo Phương
 

More from NIHON DENKEI SINGAPORE

More from NIHON DENKEI SINGAPORE (20)

Quantum Computer - Low noise signal processing solution
Quantum Computer - Low noise signal processing solutionQuantum Computer - Low noise signal processing solution
Quantum Computer - Low noise signal processing solution
 
MTI Japan Magnetic Field Measurement Instruments
MTI Japan Magnetic Field Measurement InstrumentsMTI Japan Magnetic Field Measurement Instruments
MTI Japan Magnetic Field Measurement Instruments
 
catalog-en-20190708_safety_training_machine.pdf
catalog-en-20190708_safety_training_machine.pdfcatalog-en-20190708_safety_training_machine.pdf
catalog-en-20190708_safety_training_machine.pdf
 
Ono sokki ds3000 Data Station DS-3000 series Sound and Vibration Real-time...
Ono sokki ds3000  Data Station  DS-3000 series  Sound and Vibration Real-time...Ono sokki ds3000  Data Station  DS-3000 series  Sound and Vibration Real-time...
Ono sokki ds3000 Data Station DS-3000 series Sound and Vibration Real-time...
 
Hioki data logger selection
Hioki data logger selectionHioki data logger selection
Hioki data logger selection
 
HIOKI Current Series sensors probes_e1-1_zb
HIOKI Current  Series sensors probes_e1-1_zbHIOKI Current  Series sensors probes_e1-1_zb
HIOKI Current Series sensors probes_e1-1_zb
 
Hioki Field measuring instruments 2022 catalogue
Hioki Field measuring instruments 2022 catalogueHioki Field measuring instruments 2022 catalogue
Hioki Field measuring instruments 2022 catalogue
 
Hioki Electrical measuring instruments 2022 catalogue
Hioki Electrical measuring instruments 2022 catalogueHioki Electrical measuring instruments 2022 catalogue
Hioki Electrical measuring instruments 2022 catalogue
 
9322 Hioki Differential Probe
9322 Hioki Differential Probe9322 Hioki Differential Probe
9322 Hioki Differential Probe
 
KG Electronic load DC power supply Ripple and Noise meter
KG Electronic load DC power supply Ripple and Noise meterKG Electronic load DC power supply Ripple and Noise meter
KG Electronic load DC power supply Ripple and Noise meter
 
HIOKI Power Analyzer Pw8001
HIOKI Power Analyzer  Pw8001 HIOKI Power Analyzer  Pw8001
HIOKI Power Analyzer Pw8001
 
ONO SOKKI Elevator Speedometer EC-2100
ONO SOKKI Elevator Speedometer  EC-2100ONO SOKKI Elevator Speedometer  EC-2100
ONO SOKKI Elevator Speedometer EC-2100
 
KIKUSUI PBZ Series Intelligent Bi-Polar Power Supplies
KIKUSUI PBZ Series Intelligent Bi-Polar Power Supplies KIKUSUI PBZ Series Intelligent Bi-Polar Power Supplies
KIKUSUI PBZ Series Intelligent Bi-Polar Power Supplies
 
Rion product for_the_automotive_industry
Rion product for_the_automotive_industry Rion product for_the_automotive_industry
Rion product for_the_automotive_industry
 
HIOKI BYPASS DIODE TESTER FT4310
HIOKI BYPASS DIODE TESTER FT4310HIOKI BYPASS DIODE TESTER FT4310
HIOKI BYPASS DIODE TESTER FT4310
 
TEAC Load cell catalog (JAPAN)
TEAC Load cell catalog  (JAPAN) TEAC Load cell catalog  (JAPAN)
TEAC Load cell catalog (JAPAN)
 
ADCMT DC Voltage Current Source/Monitor Catalog 6253 DENKEI
ADCMT DC Voltage Current Source/Monitor  Catalog 6253  DENKEIADCMT DC Voltage Current Source/Monitor  Catalog 6253  DENKEI
ADCMT DC Voltage Current Source/Monitor Catalog 6253 DENKEI
 
RIGOL Programmable linear DC power supply DP800 Series
RIGOL  Programmable linear DC power supply DP800 SeriesRIGOL  Programmable linear DC power supply DP800 Series
RIGOL Programmable linear DC power supply DP800 Series
 
Rz catalogue cn AIKOH 推拉力计
Rz catalogue cn AIKOH 推拉力计Rz catalogue cn AIKOH 推拉力计
Rz catalogue cn AIKOH 推拉力计
 
Aikoh General Catalogue in Japanese catj201907
Aikoh General Catalogue in Japanese  catj201907Aikoh General Catalogue in Japanese  catj201907
Aikoh General Catalogue in Japanese catj201907
 

Recently uploaded

Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Victor Rentea
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 

Recently uploaded (20)

Elevate Developer Efficiency & build GenAI Application with Amazon Q​
Elevate Developer Efficiency & build GenAI Application with Amazon Q​Elevate Developer Efficiency & build GenAI Application with Amazon Q​
Elevate Developer Efficiency & build GenAI Application with Amazon Q​
 
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost SavingRepurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
 
Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..
 
Platformless Horizons for Digital Adaptability
Platformless Horizons for Digital AdaptabilityPlatformless Horizons for Digital Adaptability
Platformless Horizons for Digital Adaptability
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
 
How to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected WorkerHow to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected Worker
 
AWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of TerraformAWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of Terraform
 
ICT role in 21st century education and its challenges
ICT role in 21st century education and its challengesICT role in 21st century education and its challenges
ICT role in 21st century education and its challenges
 
Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...
 
Apidays New York 2024 - APIs in 2030: The Risk of Technological Sleepwalk by ...
Apidays New York 2024 - APIs in 2030: The Risk of Technological Sleepwalk by ...Apidays New York 2024 - APIs in 2030: The Risk of Technological Sleepwalk by ...
Apidays New York 2024 - APIs in 2030: The Risk of Technological Sleepwalk by ...
 
FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024
 
Artificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : UncertaintyArtificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : Uncertainty
 
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
 
Six Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal OntologySix Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal Ontology
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 

Hioki st5540 e12-85e_leakage_current_tester

  • 1. LEAK CURRENT HiTESTER ST5540, ST5541 Leak Current Measurement - Essential to Electrical Safety • Uninterrupted polarity switching function dramatically reduces cycle time • Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards For medical-use electrical devices and essential to electrical safety Compliance with IEC 60601-1:2005 Ed 3.0, JIS T 0601-1:2012 (*Starting on June 1, 2012, medical electrical equipment sold in the EU must comply). Model ST5540 comply with IEC 60601-1:2005 +A1:2012 (Ed 3.1), and IEC 62353 of 2017 Compliance with Electrical Appliances and Materials Safety Act, JIS, IEC, and UL standards for general-use electrical devices
  • 2. 2 General-use electrical devices (Medical) IEC standards IEC 60601-1: 2005 +A1: 2012 IEC 60601-1: 1988 A2: 1995, etc. (Medical) JIS standards JIS T0601-1: 2012 supplement 1: 2014, 1999 ■ Comparison of ST5540 Functionality ST5540 compliance ST5540 Medical-use electrical devices JIS standards Electrical Appliances and Materials Safety Act IEC standards UL standards In order to prevent the danger of electric shock, electrical devices use power supplies that are isolated from parts of the device that may come into contact with the body. However, it is impossible to achieve infinite insulation resistance. Some leak current always exists, and its magnitude changes as the insulation degrades over time. The LEAK CURRENT HiTESTER ST5540/ST5541 provides an easy-to-operate solution for measuring leak current in electrical devices, making it eminently suitable for use in an extensive array of applications, ranging from production lines to equipment maintenance and inspections. *Compliance with medical IEC standards became mandatory in June 2012. *Medical JIS standards were revised in June 2012. Complies with all standards (suitable for use with all networks) Leak current parameters as defined for medical-use electrical devices include ground leak current, contact current, patient leak current, and patient measurement current. The ST5540 provides a single solution for measuring all of these leak current variants. For Medical Electrical Devices Complies with IEC 60601-1:2005 +A1:2012 (Ed 3.1) ST5540 Measurement mode Category Standard compliance ● Patient leak current (between parts of device that come into contact with patient and ground) ● Patient leak current (external SIP/SOP voltage) ● Patient leak current (external voltage at specific F-type applied part) ● Patient leak current (current resulting from external voltage at parts of device that come into contact with patients) ● Patient measurement current ● Total patient leak current (between parts of device that come into contact with patient and ground) ● Total patient leak current (external SIP/SOP voltage) ● Total patient leak current (external voltage at specific F-type applied part) ● Total patient leak current (current resulting from external voltage at parts of device that come into contact with patient) • Medical industry (Japan Association for Clinical Engineering Technologists, etc.) • Medical device manufacturers and dealers • Medical device repair and maintenance businesses • Hospitals • IEC60601-1 (Ed 3.1) • IEC60990 • IEC62353 ● Contact current (between device enclosure and lines) ● Contact current (between device enclosure and ground) ● Contact current (between device enclosure and device enclosure) ● Ground leak current ● Free current measurement • Public agencies • Electric vehicle manufacturers • Manufacturers of general electrical devices • Household appliance industry • Information device industry • Electrical vehicle standards UL 2231-1 and UL 2231-2 • Electrical Appliances and Materials Safety Act • IEC, JIS, and UL standards *The ST5540 also complies with old standards.
  • 3. 3 Electrical Appliances and Materials Safety Act Ministerial ordinance defining technical standards for electrical devices UL standards UL2231-1: 2002, UL2231-1: 2012 (revision 2016) UL2231-2: 2002, UL2231-2: 2012 (revision 2016) UL1492: 1996 (revision 2013) UL492: 1996, etc. IEC standards IEC60990: 2016 IEC61010-1: 2010 +A1: 2016 IEC60950-1: 2005 +A1: 2009 +A2: 2013 IEC60065: 2014 JIS standards JIS B8561: (2007), JIS C9250: 1992 +A1: 2007, etc. ST5541 compliance Measurement of General-use Electrical Devices ST5541 provides standard support for standard-compliant networks (excluding medical-use electrical devices). For Standard- and Regulation-compliance ST5541 A single, robust solution for leak current measurement ST5541 General-use electrical devices ■ Comparison of ST5541 Functionality Measurement mode Category Standard compliance ● Contact current (between device enclosure and lines) ● Contact current (between device enclosure and ground) ● Contact current (between device enclosure and device enclosure) ● Ground leak current ● Free current measurement • Public agencies • Electric vehicle manufacturers • Manufacturers of general electrical devices • Household appliance industry • Information device industry • Electrical vehicle standards UL 2231-1 and UL 2231-2 • Electrical Appliances and Materials Safety Act • IEC, JIS, and UL standards There are various standards in place concerning networks (body simulated resistance), and a standard-compliant network is required in order to make measurements.
  • 4. 4 The ST5540/ST5541 can switch polarity without stopping the supply of power to the device under measurement. Old models require that the device under measurement be turned off and then back on again when switching polarity, but the ST5540 and ST5541 let you progress smoothly to the next testing process. ■ Safety conductor current measurement function The ST5540/ST5541 can perform safety conductor current measurement as defined in standards such as IEC 60990 and IEC 60950-1. ■ Automatic measurement functionality Simple operation allows you to switch power supply polarity and automatically make measurements with the target device in the normal and single-fault states, displaying the peak values. You can also set the measurement time and wait time. These capabilities help reduce operation time. ST5540/ST5541 Features ■ Uninterrupted polarity switching function The ability to conduct tests without turning off the power when switching the power supply polarity dramatically reduces cycle times. ■ Improved test reliability Blown fuse check function When measurement starts, the instrument checks for unintentional probe misalignment using of a preconfigured lower limit setting. L N E *Only the ground is switched, allowing an uninterrupted supply of power. AC ■ 110% voltage application jack The instrument’s 110% voltage application jack, which is used during testing of medical devices, outputs the target device line power supply voltage as-is. The polarity can be switched (ST5540 only). ■ Circuit breaker for device under measurement The instrument’s workbench-type design features a terminal block and a circuit breaker on the front panel, making it deal for embedding in test lines and simplifying connectivity with the device being measured, even while rack-mounted. ■ Simple, interactive operation The ST5540/ST5541 uses a touch panel that lets you configure settings by touching selections in response to information displayed on the panel, keeping operation simple. Peak value display Displays the type of power supply fault and the peak value for the leak current, which varies with target device operation. Allowable value The maximum allowable value under the standard in question is automatically set. Settings can also be changed as desired by the user. Data storage Measurement data: For up to 100 target devices Measurement conditions: Up to 30 sets Current measured value Judgment result based on set allowable value Power supply polarity/device status/measurement current [Measurement screen] Measurement data (peak values) can be stored in the instrument’s built-in memory. Saved data can be checked on the stored data reference screen after measurement is complete. Data can be stored for up to 100 test targets, with each target being identified by a registered device name and control number. Additionally, the instrument can store a maximum of 2,000 peak value data points. Together, these capabilities eliminate the need to jot down measured values at the measurement site. ■ Save measurement data for 100 devices ■ Ability to store up to 30 sets of measurement conditions The instrument can save and load up to 30 sets of measurement conditions, allowing you to immediately switch between conditions. ST5540 ST5541 *Requires isolation transformer. (not included)
  • 5. 5 ST5540 internal circuitry Connection terminals ( Line power supply terminal block for device under test ) Expandability for the Future S10 terminal : Switch for connecting a function ground terminal to the measurement power supply system’s ground point Allows connectivity to ground to be configured during leak current measurement. S12 terminal : Switch for connecting the parts of device that come into contact with the patient to the measurement power supply circuit’s ground point Allows connectivity to ground to be configured during leak current measurement. S13 terminal : Switch for connecting contactable metallic parts that are not protectively grounded to the ground line Allows connectivity to ground to be configured during leak current measurement. E terminal : Connected to the E (earth) LINE IN terminal. This terminal is always connected and cannot be configured. ■ All the switch terminals needed for standard-compliant measurement The ST5540/ST5541 provides the terminals needed to perform leak current measurement in compliance with IEC 60601-1 and JIS T0601-1, eliminating the need for the user to provide external switches. S10 S12 S13 E Power supply polarity switching unit Circuit for simulating a single wire break in the power supply Circuit for simulating a ground wire break Device under measurement AC *S10, S12, S13, and E are available on the ST5540 only. ■ Standard USB interface The ST5540/ST5541’s standard USB interface simplifies auto- matic testing on manufacturing lines and in similar installa- tions. ■ Separation of the instrument’s power supply and target device lines The instrument’s power supply and target device line power supply are separated, helping prevent damage due to the inadvertent input of an incorrect supply voltage. There’s no need to change the ST5540/ ST5541’s supply voltage, even if the target device’s supply voltages changes. ■ Support for rated currents of up to 20 A The ST5540/ST5541 supports currents of up to 20 A and voltages of up to 250 V. Its ability to accommodate large currents allows it to be used with a more extensive range of target products, including devices from new fields such as electric vehicles and household appliances. Start of measurement and loading of measurement conditions can be controlled from an external source. Additionally, judgment results, test signals, and other data can be output, making it possible to use the instrument to develop an automated line. In addition to outputting judgment results for each measurement item, the instrument also provides T-FAIL output, which is generated continuously once a FAIL result is encountered during automatic testing. ■External control via EXT I/O EXT I/O description Input signals Active-low input Max. pplied voltage EXT.DCV terminal input voltage High level EXT.DCV terminal input voltage or open Low level 0.3 VDC or less Output signal Open collector output Max. load voltage 24 VDC (when not using the EXT.DCV terminal) Max. output current 60 mA DC per signal (low level) ●Output TEST : Outputs low continuously during automatic measurement. MEAS : Outputs the measurement count automatic measurement and measurement of multiple items. PASS : Outputs the PASS judgment result for each measurement item. FAIL : Outputs the FAIL judgment result for each measurement item. LOW : Generates continuous output once a low signal is encountered during automatic testing. T-FAIL : Generates continuous output once a FAILresult is encountered during automatic testing. INT.DCV : Generates internal 5 VDC output (not isolated from internal circuitry). INT.GND : Generates internal GND output (same as the case ground level). ●Input START : Starts automatic measurement at low. STOP : Forcibly terminates measurement at low. LOAD (0 to 4) : Loads saved panels (30 panels). EXT.DCV : Accepts external power supply input from 5 VDC to 24 VDC. EXT.COM : Accepts external COM input. KEYLOCK : Disables switches other than the start switch. Inlet Outlet F 1 : Connectivity not supported for USB memory. Data communications only. 2 : Cannot be used while an RS-232C connection is active. ■ Standard RS-232C port The ST5540/ST5541’s standard RS-232C port can be used to control the instrument from a computer and to print data using the 9442 printer (option).
  • 6. 6 ■ Leak current measurement unit ST5540/ST5541 Specifications ■ Network (body simulated resistance) ■ General specifications *1 When using AC measurement mode, the high-pass filter frequency characteristics (fc = 4 Hz) are added. *2 ST5540 only. ■ Accuracy (current measurement unit) ■ Temperature and humidity range within which accuracy is guaranteed: 23°C ±5°C, 80% RH or less, non-condensing ■ Temperature coefficient: Add 0.1 × basic accuracy × (T-23) for operating temperature T (°C) ■ Warm-up time: 20 min ■ The range within which accuracy is guaranteed when using Network D and Network F (full-scale value for each range) is approximately 1/1.5 and 1/2, respectively. ■ Calculated value when the voltage is detected across both ends of a network consisting of non-inductive resistance with a theoretical value of 1 kΩ ■ The following accuracy values also apply when using voltmeter mode. ● Measurement mode: AC*1 / AC+DC ● Measurement: mode DC ● Measurement mode: AC peak*3 ● Current monitor accuracy (Measurement methods: Average value response, rms calculation) ● Voltage monitor accuracy Range Guaranteed accuracy range Resolution Accuracy 0.1Hz≤f<15Hz*2 DC≤f≤100kHz 100kHz<f≤1MHz 50.00mA From 4 mA 10 μA ±(4.0%rdg.+10dgt.) ±(2.0%rdg.+6dgt.) ±(2.0%rdg.+10dgt.) 5.000mA From 400 μA 1 μA 500.0μA From 40 μA  0.1 μA 50.00μA From 4 μA 0.01 μA ±4.0%f.s. ±2.0%f.s. ±2.0%f.s. Range Guaranteed accuracy range Resolution Accuracy 15Hz≤f≤10kHz 10kHz<f≤100kHz 100kHz<f≤1MHz 75.0mA From 8 mA 100 μA ±(2.0%rdg.+6dgt.) ±5.0%f.s. ±15.0%f.s. 10.00mA From 0.8 mA 10 μA 1.000mA From 100 μA  1 μA ±2.5%f.s. 500.0 μA From 40 μA 0.1 μA ±2.5%f.s. ±20.0%f.s. Range Guaranteed accuracy range Resolution Accuracy 50.00mA From 4 mA 10 μA ±(2.0%rdg.+6dgt.) 5.000mA From 400 μA 1 μA 500.0 μA From 40 μA 0.1 μA 50.00 μA From 4 μA 0.01 μA ±2.0%f.s. Range Guaranteed accuracy range Resolution Accuracy 300.0 V 85 V*4 to 275V 0.1 V ±(5.0%rdg.+10dgt.) Range Guaranteed accuracy range Resolution Accuracy 300.0 V From 85 V*5 0.1 V ±(2.0%rdg.+5dgt.) Measurement current DC / AC / AC+DC / AC peak Allowable measurement current Max. 50 mA (DC / AC / AC+DC mode) Max. 75 mA (AC peak mode) Measurement ranges DC / AC / AC+DC mode:50 μA/500 μA/5 mA/50 mA AC peak mode:500 μA/1 mA/10 mA/75 mA Range switching AUTO/HOLD Trigger method Manual: Generates trigger automatically internally, free-run measurement. Automatic: Starts measurement based on external start signal. Measurement terminals T1 terminal, T2 terminal (with built-in fuse holder), T3 terminal (110% voltage application terminal: ST5540 only) (*Step-up isolation transformer required for 110% application.) Measurement methods Measurement of voltage drop across body simulated resistance points Calculation and display of current values True rms measurement Measurement unit floats relative to instrument ground. A/D conversion method ΔΣ method (20-bit) Instrument-to-ground capacitance 200 pF or less (between T1/T2 terminal and case ground) Input resistance 1 MΩ ±1% (single-end input) Notincludingvoltagemeasurementunit,bodysimulatedresistance(currentdetectioncircuit) Input capacitance 150 pF or less (between T1 and T2 terminals) (f = 100 kHz, isolated network circuit, including cables) CMRR (between T1 and T2 terminals and case) 60 dB or greater at 60 Hz / 60 dB or greater at 10 kHz 40 dB or greater at 100 kHz / 40 dB or greater at 1 MHz (Isolated from network circuit with fuse shorted) ● Medical-use electrical devices: Network B (ST5540 only) Basic measurement element: 1 kΩ Filter: 10 kΩ + 15 nF ● Electrical Appliances and Materials Safety Act: Network A Basic measurement element: 1 kΩ Filter: 10 kΩ + 11.22 nF + 579 Ω ● IEC 60990: Network C Basic measurement element: 1.5 kΩ + 500 Ω Filter 1: 10 kΩ + 22 nF Filter 2: 10 kΩ + (20 kΩ + 6.2 nF) //9.1 nF ● UL: Network D Basic measurement element: 1.5 kΩ //0.15 μF ● General-purpose 1: Network E Basic measurement element: 1 kΩ ● General-purpose 2: Network F Basic measurement element: 2 kΩ ● IEC 61010-1: Network G Basic measurement element: 375 Ω + 500 Ω Filter: 375 Ω //0.22 μF + 500 Ω ● Safety conductor current Basic measurement element (35 Ω) Outlet max. allowable leak current 50 mA Dielectric strength Between power supply terminals and protective ground: 1.39 kVAC (5 mA), 15 sec Between measurement terminals and power supply terminals: 2.30 kVAC (10 mA), 15 sec Between measurement terminals and control circuit: 2.30 kVAC (10 mA), 15 sec Standard compliance EMC: EN 61326 Safety: EN 61010 Conductive RF 3% f.s. or less at 3 V Magnetic field effects (Representative value when conducting measurements in the AC 500 μA range) Accessories ST5540: One set of L2200 test leads (one red, one black) + one red L2200 test lead ST5541: One set of L2200 test leads (one red, one black) Enclosure Probe 9195 × 1 , CD-ROM × 1 (USB Driver) power cord × 3 (1 for instrument and 2 for measuring instrument line supply use) spare fuse × 1 (250 V F 50 mAL, measurement use) Dimensions Approx. 320 (W) × 110 (H) × 253 (D) mm Mass Approx. 4.5 kg Range Guaranteed accuracy range Resolution Accuracy DC,15Hz≤f≤100kHz 100kHz<f≤1MHz 50.00 mA 12.00 mAto 50.00 mA 10 μA ±(2.0%rdg.+6dgt.) ±(5.0%rdg.+20dgt.) 10.00mA 1.30 mAto 13.00 mA 10 μA ±(2.0%rdg.+6dgt.) ±(5.0%rdg.+20dgt.) Range Guaranteed accuracy range Resolution Accuracy 15Hz≤f≤10kHz 1kHz<f≤100kHz 100kHz<f≤1MHz 75.0 mA 12.0 mAto 75.00 mA 100 μA ±(2.0%rdg.+6dgt.) ±5.0%f.s. ±25.0%f.s. 10.00 mA 1.30 mAto 13.00 mA 10 μA ±2.5%f.s. ±5.0%f.s. ±25.0%f.s. ■ Safety conductor current accuracy ● Measurement mode: DC / AC*4 / AC+DC ● Measurement mode: AC peak *3 Setting not available with Network A, B, or C (when filter off). *4 Voltages of less than 80 V are displayed as “Less than 80 V.” *5 Currents of less than 0.5 A are displayed as “Less than 0.5 A.” (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year) Display 320 × 240 dot matrix LCD (with backlight) Control 6 × 6 matrix touch panel Operating temperature and humidity range 0°C to 40°C, 80% RH or less (non-condensing) Storage temperature and humidity range -10°C to 50°C, 80% RH or less (non-condensing) Temperature and humidity range within which accuracy is guaranteed 23°C ±5°C, 80% RH or less (non-condensing) Guaranteed accuracy period 1 year Operating location Indoor use at an elevation not exceeding 2,000 m Instrument power supply 100/120/220/240 VAC, as specified by customer Rated power supply frequency: 50/60 Hz Rated power: 30 VA Line power supply for device being measured and outlet Rated supply voltage: 100 to 250 VAC Rated power supply frequency: 50/60 Hz Rated current: Input, terminal block: 20 A Output, terminal block: 20 A
  • 7. 7 AC GROUNDING HiTESTER 3157 Protective Ground Tester Indispensable for Standards Certification Standard compliance (examples) ● IEC 60065 ● IEC 60335-1 ● IEC 60601-1 ● IEC 60950-1 ● IEC 61010-1 ● JIS-C1010-1 ● Electrical Appliances and Materials Safety Act ● UL (various applicable standards) Main ground line Ground branch line Ground branch line Medical outlet Medical ground terminal Grounding electrode Medical ground center Medical room ■ Measuring between the grounding center and grounding terminal Verify that the electrical resistance between the medical outlet’s grounding electrode connector or medical ground terminal and medical ground center is less than or equal to 0.1 Ω by applying a current of approximately 25 A with an AC current with a no-load voltage of 6 V or less and measuring the resistance using the voltage droop method. * This measurement requires an extension cable (available separately). The extension cable is a special-order item; please contact your HIOKI distributor for more information. Combination of Instruments for Leak Current Testing and Safety Conductor Testing The following are key parts of any safety inspection of electrical equipment: • Leak current test: Measure with the ST5540 and ST5541. • Safety conductor test (also known as a ground line resistance test or ground conductor test): Measure with the 3157. The 3157 can also be used for conducting measurements under the JIS T 1022:2006 safety standard for hospital electrical equipment. For JIS T 1022:2006 Measurement at Hospitals Electrical devices Safety Standard for Hospital Electrical Equipment 3157 option option Printer options PRINTER 9442 Forprintingnumerical values112mm(4.41in) paperwidth AC ADAPTER 9443-02 ForthePrinter9442, EUtype CONNECTION CABLE 9446 ForthePrinter9442,25pin -9pin,1.5m(4.92ft)length *When using the Printer 9442, also purchase the RS-232C Interface 9593-02 Other SAFETYTESTDATA MANAGEMENTSOFTWARE9267 For PC control application software Input/Output cords CURRENTAPPLYPROBE9297 With switch, 1.48m (4.86 ft) length CURRENT PROBE 9296 Alligator clip, 1.45m (4.76 ft) length REMOTE CONTROL BOX (DUAL) 9614 For Start/Stop control, 1.5m (4.92 ft) cord length REMOTE CONTROL BOX (SINGLE) 9613 For Start/Stop control, 1.5m (4.92 ft) cord length *The 3157-01’s handshake function cannot be used with the RS-232C Cable 9638 GP-IB CONNECTOR CABLE 9151-02 2 m (6.56 ft) length GP-IB INTERFACE 9518-02 For the 3157-01, built in type RS-232C INTERFACE 9593-02 For the 3157-01, built in type, not CE marked RECORDING PAPER 1196 ForthePrinter9442,112mm (4.41in)×25m(82.03ft),10 rolls/set • Easily perform protective continuity testing in compliance with international safety standards and laws -1) Protective continuity resistance measurement for medical devices and gen- eral electrical devices -2) Ground connectivity testing when installing electrical machine tools and dis- tribution panels -3) Testing of protective grounding and isopotential grounding work for medical equipment -4) Evaluation of contact status using large currents • Feedback control system that is capable of applying a stable current even with a fluctuating load • Soft-start function that checks the connection to the device under test before applying the current EXT I/O Interface insertion slot Model : AC GROUNDING HiTESTER 3157 Model No. (Order Code) (Note) 3157-01 (100-120 / 200-240 VAC switching) Note: This instrument is not capable of performing measurement by itself. Please purchase two Current probe 9296 units or one Current probe 9296 and one Current apply probe 9297, depending on your measurement application.:
  • 8. HEADQUARTERS 81 Koizumi Ueda, Nagano 386-1192 Japan www.hioki.com HIOKI USA CORPORATION TEL +1-609-409-9109 FAX +1-609-409-9108 hioki@hiokiusa.com / www.hiokiusa.com All information correct as of May 21, 2018. All specifications are subject to change without notice. ST5540E12-85E Printed in Japan DISTRIBUTED BY HIOKI (Shanghai) SALES & TRADING CO., LTD. TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360 info@hioki.com.cn / www.hioki.cn HIOKI SINGAPORE PTE.LTD. TEL +65-6634-7677 FAX +65-6634-7477 info-sg@hioki.com.sg / www.hioki.com.sg HIOKI KOREA CO., LTD. TEL +82-2-2183-8847 FAX +82-2-2183-3360 info-kr@hioki.co.jp / www.hiokikorea.com HIOKI EUROPE GmbH TEL +49-6173-31856-0 FAX +49-6173-31856-25 hioki@hioki.eu / www.hioki.com Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies. ● Measurement date ● Allowable values ● Instrument name ● Maximum value ● Control number ● Judgment results ● Class (applied part) ● Measurement current (AC, DC, AC+DC, AC peak) ● Network ● Power supply polarity (normal, reversed) ● Measurement mode ● Instrument status (normal, ground line broken) ● Filter settings The optional PRINTER 9442 can be used to print data via the instrument’s RS-232C interface, providing a convenient way to attach a hard copy of test data. Print method: Thermal serial dot Paper width/print speed: 112 mm/52.5 cps Power supply: AC ADAPTER 9443 or included nickel- metal-hydride battery (good for approx. 3,000 lines of print when fully charged with the 9443 adapter) Dimensions: Approx. 160 (W) × 66.5 (H) × 170 (D) mm Mass: Approx. 580 g *CONNECTION CABLE 9444 and AC ADAPTER 9443 are required in order to connect the 9442 printer. ■ Printing saved data Saved measurement data is displayed (pressing the print key within the same data unit causes all data in the data unit to be printed). Printable data (printed data can be selected from the following) PRINTER 9442 Convenience ■ Leak current tester supplies *Some standards require use of an isolation transformer. Product inquiries should be directed to: Isolation transformer model numbers 100 to 110 V (Japan): HSW-2KSP 240 to 264 V (overseas): HSW-5KSP For more information: Tokyo Rikosha Co., Ltd. Phone: +81-48-856-3851 (reception) http://www.tokyorikosha.co.jp Standards require use of an isolation transformer when measuring medical-use electrical devices. Please purchase a transformer with the necessary rated capacity. Isolation transformer Safety standards Example printout Date Instrument name Control number Connection class and applied part Network Measurement mode Filter Allowable value (upper limit) Allowable value (lower limit) Maximum value Judgment result Polarity Status Model : LEAK CURRENT HiTESTER ST5540 Model No. (Order Code) (Note) ST5540 (For medical-use and electrical devices) Model : LEAK CURRENT HiTESTER ST5541 Model No. (Order Code) (Note) ST5541 (For electrical devices) PC communication RS-232C CABLE 9637 For the PC, 9pin - 9pin, cross, 1.8m (5.91 ft) length RS-232C CABLE 9638 For the PC, 9pin - 25pin, cross, 1.8m (5.91 ft) length Printer options PRINTER 9442 Forprintingnumerical values112mm(4.41in) paperwidth RECORDING PAPER 1196 For the Printer 9442, 112 mm (4.41 in) × 25 m (82.03 ft), 10 rolls/set AC ADAPTER 9443-02 ForthePrinter9442, EUtype CONNECTION CABLE 9444 ForthePrinter9442, 9pin-9pin,1.5m (4.92ft)length Test probes *The L2200 and the 9195 are bundled ENCLOSURE PROBE 9195 For the ST5540 series, 3156/3155 TEST LEAD L2200 70 cm (2.30ft) length, detachable large alligator clips or needle tips are bundled, CAT IV 600V, CAT III 1000V ST5540, ST5541 shared options