K.K.D.Ramesh
Jr Medical Physicist
American Oncology Institute
Hyderabad
TG-51 PROTOCOL
Most of the Clinical studies and Retrospective
data analysis have proved that the dose delivered
to tumor must be within ±5% of the prescribed
dose to achieve meaningful and acceptable tumor
control . This requires all parameters that could
effect the delivered dose must be accurate within
±3% or better.
TG-51 PROTOCOL
For this a number of protocols for beam dosimetry based
on the Exposure ,Air Kerma, or Absorbed dose to water
calibration of Ion chambers are available, for
traceability and uniformity.
Among this TG-51 is a Protocol for the practice of
reference dosimetry. It applies to photon with nominal
energies between Co60 to 50 MV and for ë beams of
nominal energies between 4 and 50 MeV.
It uses Ion chambers calibrated in terms of absorbed dose
to water in a Co60( N DW
Q).
The absorbed dose to water standard is assumed to be
rugged standard of dosimetry as it is based on Water
Calorimetry, Graphite Calorimetery, Ionometry and
Chemical dosimetry.
The absorbed dose to water based formalism is relatively more accurate and simple to use and it is
possible to determine directly the absorbed dose to water from Electrometer reading using a simple
formalism.
DW
Q=M* N DW
CO60 *KQ(GY)
DW
Q – Absorbed dose to water at a user quality beam”Q”(GY)
M - Fully corrected Electrometer reading(c)
N DW
CO60 - Usually absorbed dose calibration factors will be
obtained for reference condition in a Co60 beam. To obtain calibration
factor ( N DW
Q) for user beam,
KQ- Converts the absorbed dose to water calibration factor for a Co60 beam
in to the calibration factor for an arbitrary beam quality ’Q’ which can be
for Ȣ or ë. It is a chamber specific.
N DW
Q=N DW
CO60 *KQ(GY/C)
DW
Q=M* N DW
CO60 *KQ(GY))
For ë:
KQ contains 2 components
KQ=Pgr
Q *K’*R50 Kecal=Pgr
Q*KR50
KR50-Chamber specific factor & it depends on the quality for which the absorbed dose calibration factor was
obtained and the user beam quality ‘Q’ as specified by R50.
Kecal- It is a Ȣ to ë conversion factor, is fixed for a given chamber and its value is one for an ë beam quality
Q ecal Value is available from the table for a given chamber at given quality of beamR50 -7.5cm.
K’R50- It is a beam quality dependent and converts N DW
Qcal in to N DW
Q .this values are obtained
from the graph for Parallel plate &Cylindrical chambers.
Pgr
Q- It is necessary for cylindrical chambers to correct for gradient effect at the
reference depth.
Absorbed dose to Water for ë
KQ=Pgr
Q *K’R50* Kecal=Pgr
Q*KR50
Pgr
Q=Mraw(dref+0.5rcav)/Mraw(dref)
DW
Q=M*Pgr
Q *K’R50* Kecal* N DW
CO60(GY)
M=Mu*Pion*Ptp*Pelc*Ppol
Pion=1-(v1/V2)2/(M1/M2)-(V1/V2)2 , V2=V1/2
Ptp=(273.2+T/273.2+220C)(1013.2/P)
Pelc= Electrometer calibration factor
Ppol=(M++M-)/2(M+)
M=Mu*Pion*Ptp*Pelc*Ppol
Beam Quality Specifications for Photons
Influence quantity Reference value/characteristics
Phantom material Water
Chamber type Cylindrical
Measurement depth zref PDD(10)
Reference point of On the central axis at the centre of the the chamber
cavity volume
Position of the reference At 0.6 rcav above measurement depth
point of the chamber
SSD/SCD 100 cm
Field size 10 cm × 10 cm
Reference Conditions for Photon Dosimetry
Influence quantity Reference value/characteristics
Phantom material Water
Chamber type Cylindrical
Measurement depth zref 10 g/cm2
Reference point of On the central axis at the centre of the the chamber
cavity volume
Position of the reference At the measurement depth zref
point of the chamber
SSD/SCD 100 cm
Field size 10 cm × 10 cm
Photons reference Dosimetry
10 cm (depth)
100 cm (SSD)
10 x 10 cm2
Electro-
meter
Water Phantom
Ion chamber
Experimental Set-up : SSD
Water
Beam quality specification for ë
Influence quantity Reference value/characteristics
Phantom material water
Chamber type PP or cylindrical
Plane parallel (PP)
Reference point of PP - on the inner surface of the
the chamber window at its centre
Cylindrical - on the central axis at the
centre of the cavity volume
Position of the reference PP - at the point of interest
point of the chamber Cylindrical : 0.5 rcyl deeper than the
point of interest
SSD 100 cm
Field size 10 cm × 10 cm
at phantom surface 20 cm × 20 cm - R50 > 8.5 g/cm2
When using an ionisation chamber, the measured quantity is R50,ion. The R50 is
obtained using
R50=1.029R50,ion-0.063(2≤ R50,ion ≤10cm)
R50=1.059R50,ion-0.37( R50,ion >10cm)
When using detectors other ion chambers (e. g. diode, diamond, etc.) the measured
quantity is R50
waterproofing for ion chamber ( if needed) <1mm PMMA
water phantom (at least 30x30x30 cm3)
lead foil for photons 10MV and above
1 mm  20
Reference conditions for ë
Influence quantity Reference value/characteristic
Phantom material water
Chamber type PP or cylindrical - R50  4 g/cm2
Plane parallel (PP) - R50 < 4 g/cm2
Measurement depth zref = (0.6 R50 - 0.1) g/cm2
Reference point of PP - on the inner surface of the
the chamber window at its centre
Cylindrical - on the central axis at the
centre of the cavity volume
Position of the reference PP - at zref
point of the chamber Cylindrical : 0.5 rcyl deeper than zref
SSD 100 cm
Field size 10 cm × 10 cm or
at phantom surface 20 cm × 20 cm - R50 > 8.5 g/cm2

More Related Content

What's hot

QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxQUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxSuryaSuganthan2
 
Cavity theory-Radiation physics
Cavity theory-Radiation physicsCavity theory-Radiation physics
Cavity theory-Radiation physicsKathiravan E M
 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXVictor Ekpo
 
Icru – 83 dr. upasna
Icru – 83  dr. upasnaIcru – 83  dr. upasna
Icru – 83 dr. upasnaUpasna Saxena
 
Quality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and DosimetryQuality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and Dosimetrymahbubul hassan
 
Plan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranPlan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranKiran Ramakrishna
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesSantam Chakraborty
 
planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapyfondas vakalis
 
Monte carlo Technique - An algorithm for Radiotherapy Calculations
Monte carlo Technique - An algorithm for Radiotherapy CalculationsMonte carlo Technique - An algorithm for Radiotherapy Calculations
Monte carlo Technique - An algorithm for Radiotherapy CalculationsSambasivaselli R
 
ICRU reports 50 and 62
ICRU reports 50 and 62ICRU reports 50 and 62
ICRU reports 50 and 62Bharti Devnani
 
Robust Challenges of Bladder Protocol management ,Knowledge & Understanding
Robust Challenges of Bladder Protocol management ,Knowledge & UnderstandingRobust Challenges of Bladder Protocol management ,Knowledge & Understanding
Robust Challenges of Bladder Protocol management ,Knowledge & UnderstandingSubrata Roy
 
4D Scan and Respiratory Gating
4D Scan and Respiratory Gating4D Scan and Respiratory Gating
4D Scan and Respiratory GatingKothanda Raman
 

What's hot (20)

Treatment plannings i kiran
Treatment plannings i   kiranTreatment plannings i   kiran
Treatment plannings i kiran
 
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxQUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
 
Session 3 measure beam profile rt
Session 3 measure beam profile rtSession 3 measure beam profile rt
Session 3 measure beam profile rt
 
Cavity theory-Radiation physics
Cavity theory-Radiation physicsCavity theory-Radiation physics
Cavity theory-Radiation physics
 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
 
Isodose lines
Isodose linesIsodose lines
Isodose lines
 
Icru – 83 dr. upasna
Icru – 83  dr. upasnaIcru – 83  dr. upasna
Icru – 83 dr. upasna
 
Flattening filter Free
Flattening filter FreeFlattening filter Free
Flattening filter Free
 
Quality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and DosimetryQuality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and Dosimetry
 
Plan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranPlan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr Kiran
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
 
Beam Direction
Beam DirectionBeam Direction
Beam Direction
 
planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapy
 
Dosimetry
DosimetryDosimetry
Dosimetry
 
Monte carlo Technique - An algorithm for Radiotherapy Calculations
Monte carlo Technique - An algorithm for Radiotherapy CalculationsMonte carlo Technique - An algorithm for Radiotherapy Calculations
Monte carlo Technique - An algorithm for Radiotherapy Calculations
 
ICRU reports 50 and 62
ICRU reports 50 and 62ICRU reports 50 and 62
ICRU reports 50 and 62
 
Robust Challenges of Bladder Protocol management ,Knowledge & Understanding
Robust Challenges of Bladder Protocol management ,Knowledge & UnderstandingRobust Challenges of Bladder Protocol management ,Knowledge & Understanding
Robust Challenges of Bladder Protocol management ,Knowledge & Understanding
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
4D Scan and Respiratory Gating
4D Scan and Respiratory Gating4D Scan and Respiratory Gating
4D Scan and Respiratory Gating
 
Dose Distribution Measurement (part 1)
Dose Distribution Measurement (part 1)Dose Distribution Measurement (part 1)
Dose Distribution Measurement (part 1)
 

Similar to TG-51 Ramesh presentation

TRS 398 code of practice for low KV xrays
TRS 398 code of practice for low KV xraysTRS 398 code of practice for low KV xrays
TRS 398 code of practice for low KV xraysNithin Sadanandan
 
Out put calibration of co60
Out put calibration of co60Out put calibration of co60
Out put calibration of co60KangkanDas3
 
Trs 398 code of practice for heavy ion
Trs 398 code of practice for heavy ionTrs 398 code of practice for heavy ion
Trs 398 code of practice for heavy ionNithin Sadanandan
 
Basics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptxBasics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptxTaushifulHoque
 
Resumen del Tg-51 en idioma ingles con 65 diapositivas
Resumen del Tg-51 en idioma ingles con 65 diapositivasResumen del Tg-51 en idioma ingles con 65 diapositivas
Resumen del Tg-51 en idioma ingles con 65 diapositivasFranklinFiallos
 
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...SELINA SRAVANTHI
 
Physical society presentation ashik.pptx
Physical society presentation ashik.pptxPhysical society presentation ashik.pptx
Physical society presentation ashik.pptxAshikurRahman204838
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validationMohiniTawade
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumetsChowdaryPavani
 
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSPERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSBhuvi palaniswamy
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instrumentsSolairajan A
 
Larry phillips plasma etching for cleaning and surface preparation of niobi...
Larry phillips   plasma etching for cleaning and surface preparation of niobi...Larry phillips   plasma etching for cleaning and surface preparation of niobi...
Larry phillips plasma etching for cleaning and surface preparation of niobi...thinfilmsworkshop
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Malarvili14
 
McGill Ozone Contactor Design Project
McGill Ozone Contactor Design ProjectMcGill Ozone Contactor Design Project
McGill Ozone Contactor Design ProjectNicholas Mead-Fox
 
Calibration of analytical instruments
Calibration of analytical instrumentsCalibration of analytical instruments
Calibration of analytical instrumentsVigneshVicky470
 
Reverse phsase chromatography 2
Reverse phsase chromatography 2Reverse phsase chromatography 2
Reverse phsase chromatography 2Amol Sagulale
 

Similar to TG-51 Ramesh presentation (20)

TRS 398 code of practice for low KV xrays
TRS 398 code of practice for low KV xraysTRS 398 code of practice for low KV xrays
TRS 398 code of practice for low KV xrays
 
Out put calibration of co60
Out put calibration of co60Out put calibration of co60
Out put calibration of co60
 
Trs 398 code of practice for heavy ion
Trs 398 code of practice for heavy ionTrs 398 code of practice for heavy ion
Trs 398 code of practice for heavy ion
 
Basics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptxBasics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptx
 
Resumen del Tg-51 en idioma ingles con 65 diapositivas
Resumen del Tg-51 en idioma ingles con 65 diapositivasResumen del Tg-51 en idioma ingles con 65 diapositivas
Resumen del Tg-51 en idioma ingles con 65 diapositivas
 
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
 
Physical society presentation ashik.pptx
Physical society presentation ashik.pptxPhysical society presentation ashik.pptx
Physical society presentation ashik.pptx
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
Dosimetry concepts and dosimeters
Dosimetry concepts and dosimetersDosimetry concepts and dosimeters
Dosimetry concepts and dosimeters
 
Session 2 radiotherapy
Session 2 radiotherapySession 2 radiotherapy
Session 2 radiotherapy
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumets
 
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSPERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
 
Larry phillips plasma etching for cleaning and surface preparation of niobi...
Larry phillips   plasma etching for cleaning and surface preparation of niobi...Larry phillips   plasma etching for cleaning and surface preparation of niobi...
Larry phillips plasma etching for cleaning and surface preparation of niobi...
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
 
Prez Baljaa09
Prez Baljaa09Prez Baljaa09
Prez Baljaa09
 
McGill Ozone Contactor Design Project
McGill Ozone Contactor Design ProjectMcGill Ozone Contactor Design Project
McGill Ozone Contactor Design Project
 
Calibration of analytical instruments
Calibration of analytical instrumentsCalibration of analytical instruments
Calibration of analytical instruments
 
Ragab R 1 - UEI Day 1 - Kochi Jan18
Ragab R 1 - UEI Day 1 - Kochi Jan18Ragab R 1 - UEI Day 1 - Kochi Jan18
Ragab R 1 - UEI Day 1 - Kochi Jan18
 
Reverse phsase chromatography 2
Reverse phsase chromatography 2Reverse phsase chromatography 2
Reverse phsase chromatography 2
 

More from K.K.D.RAMESH -

Applications of Physics in Medical industry
Applications of Physics in Medical industryApplications of Physics in Medical industry
Applications of Physics in Medical industryK.K.D.RAMESH -
 
Physics aspects of Acuros algorithm-1
Physics aspects of Acuros algorithm-1Physics aspects of Acuros algorithm-1
Physics aspects of Acuros algorithm-1K.K.D.RAMESH -
 
PRECISION RADIO THERAPY Modified
PRECISION RADIO THERAPY ModifiedPRECISION RADIO THERAPY Modified
PRECISION RADIO THERAPY ModifiedK.K.D.RAMESH -
 

More from K.K.D.RAMESH - (6)

Applications of Physics in Medical industry
Applications of Physics in Medical industryApplications of Physics in Medical industry
Applications of Physics in Medical industry
 
Physics aspects of Acuros algorithm-1
Physics aspects of Acuros algorithm-1Physics aspects of Acuros algorithm-1
Physics aspects of Acuros algorithm-1
 
DISCOVERY OF GOD(3)
DISCOVERY OF GOD(3)DISCOVERY OF GOD(3)
DISCOVERY OF GOD(3)
 
PRECISION RADIO THERAPY Modified
PRECISION RADIO THERAPY ModifiedPRECISION RADIO THERAPY Modified
PRECISION RADIO THERAPY Modified
 
AWARNESS (1)
AWARNESS (1)AWARNESS (1)
AWARNESS (1)
 
rpad seminar
rpad seminarrpad seminar
rpad seminar
 

TG-51 Ramesh presentation

  • 1. K.K.D.Ramesh Jr Medical Physicist American Oncology Institute Hyderabad TG-51 PROTOCOL
  • 2. Most of the Clinical studies and Retrospective data analysis have proved that the dose delivered to tumor must be within ±5% of the prescribed dose to achieve meaningful and acceptable tumor control . This requires all parameters that could effect the delivered dose must be accurate within ±3% or better. TG-51 PROTOCOL
  • 3. For this a number of protocols for beam dosimetry based on the Exposure ,Air Kerma, or Absorbed dose to water calibration of Ion chambers are available, for traceability and uniformity. Among this TG-51 is a Protocol for the practice of reference dosimetry. It applies to photon with nominal energies between Co60 to 50 MV and for ë beams of nominal energies between 4 and 50 MeV.
  • 4. It uses Ion chambers calibrated in terms of absorbed dose to water in a Co60( N DW Q). The absorbed dose to water standard is assumed to be rugged standard of dosimetry as it is based on Water Calorimetry, Graphite Calorimetery, Ionometry and Chemical dosimetry.
  • 5. The absorbed dose to water based formalism is relatively more accurate and simple to use and it is possible to determine directly the absorbed dose to water from Electrometer reading using a simple formalism. DW Q=M* N DW CO60 *KQ(GY) DW Q – Absorbed dose to water at a user quality beam”Q”(GY) M - Fully corrected Electrometer reading(c) N DW CO60 - Usually absorbed dose calibration factors will be obtained for reference condition in a Co60 beam. To obtain calibration factor ( N DW Q) for user beam, KQ- Converts the absorbed dose to water calibration factor for a Co60 beam in to the calibration factor for an arbitrary beam quality ’Q’ which can be for Ȣ or ë. It is a chamber specific. N DW Q=N DW CO60 *KQ(GY/C) DW Q=M* N DW CO60 *KQ(GY))
  • 6. For ë: KQ contains 2 components KQ=Pgr Q *K’*R50 Kecal=Pgr Q*KR50 KR50-Chamber specific factor & it depends on the quality for which the absorbed dose calibration factor was obtained and the user beam quality ‘Q’ as specified by R50. Kecal- It is a Ȣ to ë conversion factor, is fixed for a given chamber and its value is one for an ë beam quality Q ecal Value is available from the table for a given chamber at given quality of beamR50 -7.5cm. K’R50- It is a beam quality dependent and converts N DW Qcal in to N DW Q .this values are obtained from the graph for Parallel plate &Cylindrical chambers. Pgr Q- It is necessary for cylindrical chambers to correct for gradient effect at the reference depth. Absorbed dose to Water for ë KQ=Pgr Q *K’R50* Kecal=Pgr Q*KR50 Pgr Q=Mraw(dref+0.5rcav)/Mraw(dref) DW Q=M*Pgr Q *K’R50* Kecal* N DW CO60(GY)
  • 7. M=Mu*Pion*Ptp*Pelc*Ppol Pion=1-(v1/V2)2/(M1/M2)-(V1/V2)2 , V2=V1/2 Ptp=(273.2+T/273.2+220C)(1013.2/P) Pelc= Electrometer calibration factor Ppol=(M++M-)/2(M+) M=Mu*Pion*Ptp*Pelc*Ppol
  • 8. Beam Quality Specifications for Photons Influence quantity Reference value/characteristics Phantom material Water Chamber type Cylindrical Measurement depth zref PDD(10) Reference point of On the central axis at the centre of the the chamber cavity volume Position of the reference At 0.6 rcav above measurement depth point of the chamber SSD/SCD 100 cm Field size 10 cm × 10 cm
  • 9. Reference Conditions for Photon Dosimetry Influence quantity Reference value/characteristics Phantom material Water Chamber type Cylindrical Measurement depth zref 10 g/cm2 Reference point of On the central axis at the centre of the the chamber cavity volume Position of the reference At the measurement depth zref point of the chamber SSD/SCD 100 cm Field size 10 cm × 10 cm
  • 10. Photons reference Dosimetry 10 cm (depth) 100 cm (SSD) 10 x 10 cm2 Electro- meter Water Phantom Ion chamber Experimental Set-up : SSD Water
  • 11. Beam quality specification for ë Influence quantity Reference value/characteristics Phantom material water Chamber type PP or cylindrical Plane parallel (PP) Reference point of PP - on the inner surface of the the chamber window at its centre Cylindrical - on the central axis at the centre of the cavity volume Position of the reference PP - at the point of interest point of the chamber Cylindrical : 0.5 rcyl deeper than the point of interest SSD 100 cm Field size 10 cm × 10 cm at phantom surface 20 cm × 20 cm - R50 > 8.5 g/cm2
  • 12. When using an ionisation chamber, the measured quantity is R50,ion. The R50 is obtained using R50=1.029R50,ion-0.063(2≤ R50,ion ≤10cm) R50=1.059R50,ion-0.37( R50,ion >10cm) When using detectors other ion chambers (e. g. diode, diamond, etc.) the measured quantity is R50
  • 13. waterproofing for ion chamber ( if needed) <1mm PMMA water phantom (at least 30x30x30 cm3) lead foil for photons 10MV and above 1 mm  20
  • 14. Reference conditions for ë Influence quantity Reference value/characteristic Phantom material water Chamber type PP or cylindrical - R50  4 g/cm2 Plane parallel (PP) - R50 < 4 g/cm2 Measurement depth zref = (0.6 R50 - 0.1) g/cm2 Reference point of PP - on the inner surface of the the chamber window at its centre Cylindrical - on the central axis at the centre of the cavity volume Position of the reference PP - at zref point of the chamber Cylindrical : 0.5 rcyl deeper than zref SSD 100 cm Field size 10 cm × 10 cm or at phantom surface 20 cm × 20 cm - R50 > 8.5 g/cm2