Probe Card Manufacturing Unveiling The Technology Behind Semiconductor Testing
Treatment accessories
1. Treatment Accessories
1. Beam Directional Devices - Indicate beam direction and location of
isocenter
Lasers
Cross hairs
Gradicule,
BB trays, field light and mirror assembly
Digital readouts for gantry and collimator angles
Electron cones and collimators
Front and back pointers
Breast bridge - After borders are determined, place bridge
to skin marks and you will find field separation)
2. Beam Modifiers - modify the absorption and scattering
characteristics of the beam
Standard Wedges
Dynamic wedges
Custom tissue compensators (copper brass, lead &
Lucite)
Bolus materials
Cerrobend blocks
MLC
Latching mechanisms for interlocking accessories such as cones, trays, and
standard wedges should be checked monthly
Fabricated Cerrobend blocks may be checked for voids using radiographic
imaging
When pouring, care must be taken to prevent air bubble formation
Bolus materials should be checked regularly for cleanliness and voids, and for
appropriate thickness
Treatment accessories need to be checked for physical integrity and reliability
such as tears, leaks, and broken parts