Technical Information
Liquisys M CLM223/253
Conductivity/Resistivity Measurement
Transmitter for conductive and inductive sensors
Application
The modular design of the Liquisys M CLM223/253 allows
easy adaption of the transmitter to a variety of customer
requirements. Starting with the basic version for "measurement
and alarm generation", the transmitter can be equipped with
additional software and hardware modules for special
applications. These modules can also be retrofitted as required.
Application
• Ultrapure water
• Water treatment
• Ion exchanger, reverse osmosis
• Cooling water desalinization
• Sewage
Your benefits
• Field or panel-mounted housing
• Universal application
• Simple handling
– Logically arranged menu structure
– Ultrasimple two-point calibration
• Safe operation
– Overvoltage (lightning) protection
– Direct access for manual contact control
– User-defined alarm configuration
The basic unit can be extended with:
• 2 or 4 additional contacts for use as:
– Limit contacts (also for temperature)
– P(ID) controller
– Timer for simple rinse processes
– Complete cleaning with Chemoclean
• Plus package:
– User defined current output characteristics
– Automatic cleaning trigger on alarm or limit violation
– Ultrapure water monitoring acc. to USP (United States
Pharmacopeia) and EP (European Pharmacopoeia)
(conductive)
– Polarization detection (conductive)
– Concentration measurement
– Temperature compensation via table
– Process Check System (PCS): live check of the sensor
– Adaptive calibration with installation factor (inductive)
• HART or PROFIBUS-PA/-DP
• 2nd current output for temperature, main measured value or
actuating variable
• Current input for flow rate monitoring with controller shut off
or for feedforward control
Liquisys M CLM223/253
2 Endress+Hauser
Function and system design
Features of the basic version Conductive or inductive
Two instrument versions for measurement with conductive (two electrode) sensors or inductive sensors are
available. The use of inductive sensors that are less sensitive to soiling than conductive sensors is recommended
for high conductivity measurement, concentration measurement or adhering media.
Measuring of conductivity and resistivity (conductive)
This is selected via the menu. During measurement, the value measured can be displayed in the other
measuring mode. The temperature is displayed at the same time or, if desired, not shown at all.
Temperature compensation
The following temperature compensation selections are available:
• Linear
• NaCl curve according to IEC 746
• Ultrapure water NaCl (neutral compensation)
• Ultrapure water HCl (acid compensation, also ammonia)
The reference temperature is user defined, the standard value is 25 °C (77 °F).
Configuration
Different alarms are required depending on application and operator. Therefore the transmitter permits
independent configuration of the alarm contact and error current for each individual error. Unnecessary
or undesirable alarms can be suppressed in this manner. Up to four contacts can be used as limit contacts
(also for temperature), to implement a P(ID) controller or for cleaning functions.
Direct manual operation of the contacts (bypassing the menu) provides quick access to limit, control or
cleaning contacts, permitting speedy correction of deviations.
The serial numbers of the instrument and modules and the order code can be called up on the display.
The cell constant can be edited and even calibrated for demanding special applications.
Additional functions of the
plus package
Current output configuration
In order to output wide measuring ranges while still achieving a high resolution in specific ranges, the current
output can be configured as required via a table. This permits bilinear or quasi-logarithmic curves, etc.
Polarization detection
Polarization effects in the boundary layer between the sensor and the medium to be measured limit the
measuring range of conductive conductivity sensors.
The transmitter can detect polarization effects using an innovative, intelligent signal evaluation process.
Live check
The live check issues an alarm when the sensor signal does not change over a defined period of time. This may
be caused by blocking, passivation, separation from the process, etc.
Ultrapure water monitoring acc. to USP (United States Pharmacopeia) and EP (European
Pharmacopoeia)
Ultrapure water monitoring according to USP <645> and EP means that the uncompensated conductivity and
the temperature are measured and compared to a table.
The transmitter (conductive with additional contacts) comes with the following functions:
• Monitoring of "Water for Injection" (WFI) according to USP and EP
• Monitoring of "Highly Purified Water" (HPW) according to EP
• Monitoring of "Purified Water" (PW) according to EP
The user-adjustable pre-alarm indicates undesirable operating values in due time. Full compliance with USP
or EP requires the use of a precisely calibrated sensor, for example, the CLS16.
Concentration measurement
The conversion from conductivity to concentration is effected using four user-definable concentration
curves. This permits concentrations to be displayed in %, ppm, mg/l or TDS (total dissolved solids).
Liquisys M CLM223/253
Endress+Hauser 3
Adaptive calibration for determination of the installation factor (inductive)
Inductive measuring sensors must normally be installed in pipes at a required minimum distance from the pipe
wall. If this minimum distance is not observed, the measured value changes. The built-in adaptive calibration
using the installation factor allows you to compensate for this once the sensor is installed.
Second current output The second current output can be configured for temperature, main measured value (conductivity, resistivity,
concentration) or actuating variable.
Current input The current input of the transmitter allows two different applications: controller shut-down in case of lower
flow rate violation or total failure in the main flow as well as feedforward control. Both functions are also
combinable.
Measuring system A complete measuring systems comprises:
• The transmitter Liquisys M CLM223 or CLM253
• A sensor with or without an integrated temperature sensor
• A measuring cable CYK71 (conductive), CPK9 for Condumax H CLS16 or CLK5 (inductive)
Options: extension cable, junction box VBM
a0008937
Complete measuring system Liquisys M CLM223/253
1
2
3
4
Conductive sensor CLS15
Liquisys M CLM253
Liquisys M CLM223
Inductive sensor CLS54
5
6
7
Inductive sensor CLS50
Conductive sensor CLS21
Immersion assembly CLA111
ENDRESS+HAUSER
LIQUISYS S
CAL REL
REL1
REL1
ALARM
REL2
REL2
E
ENDRESS+HAUSER
LIQUISYS S
CAL REL
REL1
REL1
ALARM
REL2
REL2
E
1
2
3
4
5
6
7
Liquisys M CLM223/253
4 Endress+Hauser
Input
Measured variables Conductivity, resistivity, temperature
Measuring range
Cable specification
Cell constant
Temperature sensors Pt 100, Pt 1000, NTC 30K
Measuring frequency
Binary inputs
Current input 4 to 20 mA, galvanically separated
Load: 260 Ω at 20 mA (voltage drop 5.2 V)
Output
Output signal 0/4 to 20 mA, galvanically separated, active
Conductivity (conductive): 0 to 600 mS/cm (uncompensated)
Conductivity (inductive): 0 to 2000 mS/cm (uncompensated)
Resistivity: 0 to 200 MΩ⋅cm
Concentration: 0 to 9999 (%, ppm, mg/l, TDS)
Temperature: -35 to +250 °C (-31 to +482 °F)
Cable length (conductive): conductivity: max. 100 m (328.1 ft) (CYK71)
resistivity: max 15 m (49.22 ft) (CYK71)
Cable length (inductive): max 55 m (180.46 ft) (CLK5)
Cable resistance CYK71: 165 Ω/km (conductivity measurement)
Adjustable cell constant: k = 0.0025 to 99.99 cm-1
Conductivity, resistivity (conductive): 170 Hz to 2 kHz
Conductivity (inductive): 2 kHz
Voltage: 10 to 50 V
Power consumption: max. 10 mA
HART
Signalkodierung Frequency Shift Keying (FSK) + 0,5 mA über Stromausgangssignal
Datenübertragungsrate 1200 Baud
Galvanische Trennung ja
PROFIBUS PA
Signalkodierung Manchester Bus Powered (MBP)
Datenübertragungsrate 31,25 kBit/s, Spannungsmodus
Galvanische Trennung ja (IO-Module)
PROFIBUS DP
Signalkodierung RS485
Datenübertragungsrate 9,6 kBd, 19,2 kBd, 93,75 kBd, 187,5 kBd, 500 kBd, 1,5 MBd
Galvanische Trennung ja (IO-Module)
Liquisys M CLM223/253
Endress+Hauser 5
Signal on alarm 2.4 or 22 mA in case of an error
Load maximum 500 Ω
Linearization transmission
behaviour
Resolution max. 700 digits/mA
Min. distance for
0 / 4 to 20 mA signal
Isolation voltage max. 350 VRMS/500 V DC
Overvoltage protection according to EN 61000-4-5
Auxiliary voltage output
Contact outputs
Limit contactor
Controller
Alarm
Conductivity:
Resistivity:
Concentration:
Actuating variable:
Temperature:
adjustable
adjustable
adjustable
adjustable
adjustable
Conductivity:
Measured value 0 to 1.999 μS/cm 0.2 μS/cm
Measured value 0 to 19.99 μS/cm 2 μS/cm
Measured value 20 to 199.9 μS/cm 20 μS/cm
Measured value 200 to 1999 μS/cm 200 μS/cm
Measured value 2 to 19.99 mS/cm 2 mS/cm
Measured value 20 to 2000 mS/cm 20 mS/cm
Resistivity
Measured value 0 to 199.9 kΩ⋅cm 20 kΩ⋅cm
Measured value 200 to 1999 kΩ⋅cm 200 kΩ⋅cm
Measured value 2 to 19.99 MΩ⋅cm 2.0 MΩ⋅cm
Measured value 20 to 200 MΩ⋅cm 20 MΩ⋅cm
Concentration no minimum distance
Temperature 15 °C
Output voltage: 15 V ± 0.6
Output current: max. 10 mA
Switching current with ohmic load (cos ϕ = 1): max. 2 A
Switching current with inductive load (cos ϕ = 0.4): max. 2 A
Switching voltage: max. 250 V AC, 30 V DC
Switching power with ohmic load (cos ϕ = 1): max. 500 VA AC, 60 W DC
Switching power with inductive load (cos ϕ = 0.4): max. 500 VA AC, 60 W DC
Pickup/dropout delay: 0 to 2000 s
Function (adjustable): pulse length/pulse frequency controller
Controller response: PID
Control gain Kp: 0.01 to 20.00
Integral action time Tn: 0.0 to 999.9 min
Derivative action time Tv: 0.0 to 999.9 min
Period for pulse length controller: 0.5 to 999.9 s
Frequency for pulse frequency controller: 60 to 180 min-1
Basic load: 0 to 40% of max. set value
Function (selectable): Latching/momentary contact
Alarm threshold adjustment range: Conductivity, resistivity, concentration,
temperature, USP, EP: complete measuring range
Alarm delay: 0 to 2000 s (min)
Liquisys M CLM223/253
6 Endress+Hauser
Protocol specific data HART
Manufacturer ID 11h
Device type code 0092h (ind. measured), 0093h (cond. measured)
Transmitter specific revision 0001h
HART specification 5.0
DD files www.products.endress.com/profibus
Load HART 250 Ω
Device variables None (dynamic variables PV, SV, only)
Features supported -
PROFIBUS PA
Manufacturer ID 11h
Ident number 1515h
Device revision 11h
Profile version 2.0
GSD files www.products.endress.com/profibus
GSD file version
Output values Main value, temperature value
Input values Display value of PLC
Features supported Device locking: The device can be locked by hardware or software.
PROFIBUS DP
Manufacturer ID 11h
Ident number 1521h
Profile version 2.0
GSD files www.products.endress.com/profibus
GSD file version
Output values Main value, temperature value
Input values Display value of PLC
Features supported Device locking: The device can be locked by hardware or software.
Liquisys M CLM223/253
Endress+Hauser 7
Power supply
Electrical connection
The instrument has protection class II and is generally operated without protective earth connection. To ensure
the measuring stability and the function for conductive sensors you have to connect the outer screen of the
sensor cable to the PE terminal.
C07-CLM2x3xx-04-06-00-xx-001.eps
Electrical connetion of the transmitter
A
B
C
D
E
F
G
H
Sensor (conductive)
Sensor (inductive)
Temperature sensor
Signal output 1 conductivity
Signal output 2 variable
Binary input 1 (Hold)
Binary input 2 (Chemoclean)
Aux. voltage output
I
J
K
L
M
N
O
Alarm (current-free contact position)
Relay 1 (current-free contact position)
Relay 2 (current-free contact position)
Relay 3 (current-free contact position)
Relay 4 (current-free contact position)
Current input 4 ... 20 mA
Power supply
11
12
13
47
48
49
57
58
59
15 V
10-50 V
81
82
10-50 V
93
94
41
42
43
Lf
mA
31
32
mA
33
34
51
52
53
54
55
56
S
84
83
15
16
84
PE
83
L+
L–
N
L1
~ =
PE
23 +
24 -
mA
15
16
85+
86 _
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
Liquisys M CLM223/253
8 Endress+Hauser
Sensor connection You require screened special measuring cables to connect conductivity sensors to the transmitter. To extend
the measuring cable, use junction box and extension cable (see accessories).
Supply voltage Depending on ordered version:
100/115/230 V AC +10/-15 %, 48 to 62 Hz
24 V AC/DC +20/-15 %
Fieldbus connection
Power consumption max. 7.5 VA
Mains protection Fine-wire fuse, medium-slow blow 250 V/3.15 A
C07-CLM2x3xx-04-06-00-xx-007.eps
Connection of conductive sensors
a0006611
Connection of inductive sensors
GN WH YE
GN WH YE
BK
BK
CYK71/CPK9
J
83 16 15 84
2
3 1
11 12 13
PE
RD WH
CLK 5
J
83 16 15 84
11 12 13
GN WH YE BU RD RD BU
S
HART
Supply voltage n/a, active current outputs
Integrated reverse voltage protection n/a, active current outputs
PROFIBUS PA
Supply voltage 9 V to 32 V, max. 35 V
Polarity sensitive no
FISCO/FNICO compliant acc. to IEC 60079-27 no
PROFIBUS DP
Supply voltage 9 V to 32 V, max. 35 V
Polarity sensitive n/a
FISCO/FNICO compliant acc. to IEC 60079-27 no
Liquisys M CLM223/253
Endress+Hauser 9
Performance characteristics
Reference temperature 25 °C (77 °F); adjustable for the compensation of the medium temperature
Resolution
Maximum measured error1)
Repeatability1
Temperature compensation
Temperature offset ±5 °C; for the adjustment of the temperature display
Conductivity:
Temperature:
depending on the measuring range: 0.001 μS/cm to
1.999 μS/cm and k ≤ 0.5 cm-1
0.1 °C
1) acc. to IEC 746-1, for nominal operating conditions
Conductivity:
Display: max. 0.5 % of measured value ± 4 digits
Conductivity signal output: max. 0.75 % of current output range
Resistivity:
Display: max. 0.5 % of measured value ± 4 digits
Resistivity signal output: max. 0.75 % of current output range
Temperature:
Display: max. 1.0 % of measuring range
Temperature signal output: max. 1.25 % of current output range
Conductivity: max. 0.2 % of measured value ± 2 digits
Resistivity: max. 0.2 % of measured value ± 2 digits
Range: -35 to +250 °C (-31 to +482 °F)
Types of compensation: uncompensated, linear, NaCl, table;
conductive only: ultrapure water NaCl, ultrapure water HCl
Liquisys M CLM223/253
10 Endress+Hauser
Installation
Installation instructions
a0005733
Field instrument
a0005737
Mounting on pipes
1 - 3 Mounting screws and mounting plate
CAL REL
REL1
REL1
ALARM
REL2
REL2
E
Pg 7
Pg 16
M
5
11/
0.43
Ø
6
/
Ø
0.24
115 / 4.53
157
/
6.18
247
/
9.72
70
/
2.76
70 / 2.76
154 / 6.06
170 / 6.69
mm / inch
Pg 13.5
1
2
3
1
1
3
Ø
max.
60
/
2.36
mm / inch
Liquisys M CLM223/253
Endress+Hauser 11
a0005734
View into the field instrument
1
2
3
4
Removable electronics box
Partition plate
Terminal blocks
Fuse
a0005736
Wall mounting of the field instrument
1 Mounting holes
2 Protecting cap
a0005738
Mounting of the field instrument with mounting post and weather protection cover
1 - 3 Mounting holes
1
2
3
4
1
2
154 / 6.06
mm / inch
Ø
6
/
Ø
0.24
3
1
2
3
Liquisys M CLM223/253
12 Endress+Hauser
a0005735
Dimensions panel-mounted instrument
a0005739
Installation of the panel-mounted instrument
1 Wall of control cabinet
2 Gasket
3 Tensioning screws
* Required installation depth
90 / 3.54
139
/
5.47
96
/
3.78
6 /
0.24
mm / inch
92 / 3.62
149.5
/
5.89
1
2
3
3
3
92 / 3.62
+0.5 +0.02
92
/
3.62
+0.5
+0.02
90
/
3.54
57
/
2.24
33
/
1.30
approx. 25 /
0.98
139 / 5.47
max. 45 /
1.77 6 /
0.24
96
/
3.78
92
/
3.62
approx. 165 / 6.50*
mm / inch
Liquisys M CLM223/253
Endress+Hauser 13
Environment
Ambient temperature -10 to +55 °C (+14 to +131 °F)
Storage temperature –25 to +65 °C (-13 to +149 °F)
Electromagnetic compatibility Interference emission and interference immunity as per EN 61326-1:2006, EN 61326-2-3:2006
Ingress protection
Electrical safety according EN/IEC 61010-1:2001, Installation Category II, for use up to 2000 m above sea level
CSA Apparatus with CSA General Purpose Approval are certified for indoor use.
Relative humidity 10 to 95%, non-condensing
Pollution degree The product is suitable for pollution degree 2.
Mechanical construction
Dimensions
Weight
Materials
Terminals
Panel mounted instrument: IP 54 (front), IP 30 (housing)
Field instrument: IP 65 / tightness acc. to NEMA 4X
Panel mounted instrument: 96 x 96 x 145 mm (3.78 x 3.78 x 5.71 inches)
Installation depth: approx. 165 mm (6.50")
Field instrument: 247 x 170 x 115 mm (9.72 x 6.69 x 4.53 inches)
Panel mounted instrument: max. 0.7 kg (1.5 lb)
Field instrument: max. 2.3 kg (5.1 lb)
Housing of panel mounted instrument: Polycarbonate
Field housing: ABS PC FR
Front membrane: Polyester, UV-resistant
Cross section max. 2.5 mm2
Liquisys M CLM223/253
14 Endress+Hauser
Operability
Operating concept All instrument control functions are arranged in a logical menu structure. Following access code entry, the
individual parameters can be easily selected and modified as needed.
Display elements
C07-CLM2x3xx-19-06-00-en-001.eps
Operating elements
1 LC display for display of measured values, configuration data and current menu field
2 Field for user labeling
3 4 main control keys for calibration and instrument configuration
4 Key for switching between automatic/manual operation of the relays
5 LED indicators for limit contactor relay (switch status)
6 LED indicator for alarm function
7 Display of active contact and key for relay switching in manual mode
The display simultaneously shows the current measured value and the temperature - the essential process data.
Brief information texts in the configuration menu provide assistance with parameter configuration.
Certificates and approvals
4 symbol Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with the legal
requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.
Ex approval for zone 2 Application of transmitter as related electrical equipment in non-hazardous area or in simple pressurized
apparatus; application of sensor in hazardous area zone 2
CSA General Purpose C.M2.3-..2...
C.M2.3-..3...
C.M2.3-..7...
12.15
mS/cm
R232
1
2
3
4
5
6
7
Setpoint
C.M2.3-..4...
C.M223-..6...
C.M253-..A...
ATEX II (3)G (Ex nL) IIC
Liquisys M CLM223/253
Endress+Hauser 15
Ordering information
Product structure
Additional functions of the
Plus package
• Current output table to cover large ranges with varying resolution, fields O33x
• Process Check System (PCS): live check of the sensor, function group P
• Ultrapure water monitoring for "Water for injection" (WFI) and "Purified water" (PW) acc. to United States
Pharmacopeia (USP) and European Pharmacopoeia (EP) with pre-alarm (conductive, additional contacts
necessary), fields R26x and R27x
• Polarization detection (conductive), function group P
• Concentration measurement, function group K
• Temperature compensation via coefficient table, function group T
• Adaptive calibration with installation factor (inductive), fields C13x
• Automatic cleaning function start, field F8
Input, software
CD Conductivity/resistivity measurement (conductive two-electrode sensor)
CS Conductivity/resistivity measurement (conductive two-electrode sensor) with additional functions (Plus package)
ID Conductivity measurement (inductive sensor)
IS Conductivity measurement (inductive sensor) with additional functions (Plus package)
Power supply, approval
A 24 V AC/DC; ATEX II (3)G (Ex nL) IIC (CLM253 only)
0 230 V AC
1 115 V AC
2 230 V AC; CSA Gen. Purp.
3 115 V AC; CSA Gen. Purp.
4 230 V AC; ATEX II 3G [Ex nL] IIC
5 100 V AC
6 24 V AC/DC; ATEX II 3G [Ex nL] IIC (CLM223 only)
7 24 V AC/DC; CSA Gen. Purp.
8 24 V AC/DC
Output
0 1 x 20 mA, primary value
1 2 x 20 mA, primary value + secondary value
3 PROFIBUS PA
4 PROFIBUS DP
5 1 x 20 mA, primary value, HART
6 2 x 20 mA, primary value, HART + secondary value
Additional contacts
05 not selected
10 2 relays (limit/P(ID)/timer)
15 4 relays (limit/P(ID)/Chemoclean) (not with PROFIBUS DP)
16 4 relays (limit/P(ID)/timer) (not with PROFIBUS DP)
20 1 x 4 ... 20 mA input + 2 relays (limit/P(ID)/timer)
25 1 x 4 ... 20 mA input + 4 relays (limit/P(ID)/Chemoclean) (not with PROFIBUS DP)
26 1 x 4 ... 20 mA input + 4 relays (limit/P(ID)/timer) (not with PROFIBUS DP)
Additional features (CLM223 only)
PRL Protective layer
Marking
1 Tagging (Tag), see additional spec.
CLM253-
complete order code
CLM223-
Liquisys M CLM223/253
16 Endress+Hauser
Scope of delivery The delivery of the field instrument includes:
• 1 transmitter CLM253
• 1 plug-in screw terminal
• 1 cable gland Pg 7
• 1 cable gland Pg 16 reduced
• 2 cable glands Pg 13.5
• 1 Operating Instructions BA193C/07/EN
• versions with HART communication:
1 Operating Instructions Field Communication with HART, BA208C/07/EN
• versions with PROFIBUS communication:
1 Operating Instructions Field Communication with PROFIBUS PA/DP, BA209C/07/EN
• versions with explosion protection for hazardous area zone II (ATEX II 3G):
Safety Instructions for use in explosion-hazardous areas, XA194C/07/A3
The delivery of the panel mounted instrument includes:
• 1 transmitter CLM223
• 1 set of plug-in screw terminals
• 2 tensioning screws
• 1 Operating Instructions BA193C/07/EN
• versions with HART communication:
1 Operating Instructions Field Communication with HART, BA208C/07/EN
• versions with PROFIBUS communication:
1 Operating Instructions Field Communication with PROFIBUS PA/DP, BA209C/07/EN
• versions with explosion protection for hazardous area zone II (ATEX II 3G):
Safety Instructions for use in explosion-hazardous areas, XA194C/07/A3
Accessories
Sensors • Condumax W CLS12
Conductive conductivity sensor for standard, Ex and high-temperature applications;
Ordering acc. to version, see Technical Information TI082C/07/en
• Condumax W CLS13
Conductive conductivity sensor for standard, Ex and high-temperature applications;
Ordering acc. to version, see Technical Information TI083C/07/en
• Condumax W CLS15
Conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex);
Ordering acc. to version, see Technical Information TI109C/07/en
• Condumax H CLS16
Hygienic conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex);
Ordering acc. to version, see Technical Information TI227C/07/en
• Condumax W CLS19
Conductive conductivity sensor for pure and ultra-pure water applications;
Ordering acc. to version, see Technical Information TI110C/07/en
• Condumax W CLS21
Conductive conductivity sensor for applications with middle to high conductivity
(incl. Ex); Ordering acc. to version, see Technical Information TI085C/07/en
• Indumax P CLS50
Inductive conductivity sensor for standard, Ex and high-temperature applications
Ordering acc. to version, see Technical Information (TI118C/07/en)
• Indumax H CLS52
Inductive conductivity sensor with short response time for food applications
Ordering acc. to version, see Technical Information (TI167C/07/en)
• Indumax H CLS54
Inductive conductivity sensor in certified hygienic design for applications in:
food, beverages, pharmaceuticals and biotechnology
Ordering acc. to version, see Technical Information (TI400C/07/en)
Liquisys M CLM223/253
Endress+Hauser 17
Connection accessories CYK71 measuring cable
• Non-terminated cable for the connection of sensors (e.g. conductivity sensors) or the extension of sensor
cables
• Sold by the meter, order numbers:
– non-Ex version, black: 50085333
– Ex version, blue: 51506616
Extension cable CLK5
• For inductive conductivity sensors, for extension via the VBM junction box, sold by the meter
• Order no.: 50085473
Junction box VBM
• For cable extension
• 10 terminals
• Cable entries: 2 x Pg 13.5 or 2 x NPT ½"
• Material: aluminum
• Ingress protection: IP 65 (i NEMA 4X)
• Order numbers:
– cable entries Pg 13.5: 50003987
– cable entries NPT ½": 51500177
• Four-pole metal plug M12 for fieldbus connection
order no. 51502184
C07-CM12xxx-02-06-00-xx-002.eps
M12 plug with socket
n.c.
n.c.
A (PA-)
A (PA-)
S
S
B (PA+)
B (PA+)
A (PA-)
S (PE)
B (PA+)
45
1.77”
7
0.28”
150
5.91”
Liquisys M CLM223/253
18 Endress+Hauser
Mounting accessories CYY101 weather protection cover for field devices, absolutely essential if operating the unit outdoors
• Material: stainless steel 1.4031 (AISI 304)
• Order No. CYY101-A
Post mounting kit
• For mounting of field housing on horizontal or vertical pipes (Ø max. 60 mm (2.36"))
• Material: stainless steel 1.4301
• order no. 50086842
a0005676
Post mounting kit
a0012427
Weather protection cover for field devices
* Wall and post mounting
** Rail mounting
70 / 2.76
90 / 3.54
Ø 6 / 0.24
Ø
8
/
0.31
70
/
2.76
90
/
3.54
M6
60
/
2.36
90 / 3.54
70 / 2.76
Ø
8
/
0.31
30
/
1.18
Ø max. 60 / 2.36
mm / inch
Liquisys M CLM223/253
Endress+Hauser 19
CYY102 universal post
• Square pipe for mounting transmitters
• Material: stainless steel 1.4301 (AISI 304)
• Order No. CYY102-A
a0005742
Universal post
Buffer solutions Precision calibration solutions, acc. to SRM (Standard reference material) of NIST, reference temperature 25 °C
(77 °F), with temperature table
• CLY11-A, 74.0 μS/cm, 500 ml (16.9 fl.oz); order no. 50081902
• CLY11-B, 149.6 μS/cm, 500 ml (16.9 fl.oz); order no. 50081903
• CLY11-C, 1.406 mS/cm, 500 ml (16.9 fl.oz); order no. 50081904
• CLY11-D, 12.64 mS/cm, 500 ml (16.9 fl.oz); order no. 50081905
• CLY11-E, 107.0 mS/cm, 500 ml (16.9 fl.oz); order no. 50081906
Optoscope Optoscope
• Interface between transmitter and PC / laptop for service purposes.
• The Windows software "Scopeware" required for the PC or laptop is supplied with the Optoscope. The
Optoscope is supplied in a sturdy plastic case with all the accessories required.
• Order no. 51500650
1495
/
58.86
60 x 60 /
2.36 x 2.36
120 /
4.72
150
/
5.91
mm/inch
TI193C/07/EN/13.11
71130248
FM+SGML 6.0

More Related Content

Similar to Liquisys m clm223 253 endress+hauser datasheet-conductivity-resistivity measurement

Liquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitter
Liquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitterLiquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitter
Liquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitterENVIMART
 
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensorsCondumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensorsENVIMART
 
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensorsCondumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensorsENVIMART
 
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...Belilove Company-Engineers
 
Flowphant t dtt31, dtt35 endress+hauser datasheet-flow switch
Flowphant t dtt31, dtt35 endress+hauser datasheet-flow switchFlowphant t dtt31, dtt35 endress+hauser datasheet-flow switch
Flowphant t dtt31, dtt35 endress+hauser datasheet-flow switchENVIMART
 
ATI-Q46H-C4-Datasheet.pdf
ATI-Q46H-C4-Datasheet.pdfATI-Q46H-C4-Datasheet.pdf
ATI-Q46H-C4-Datasheet.pdfENVIMART
 
Toshniwal - Thermal mass flow meters and controller
Toshniwal - Thermal mass flow meters and controllerToshniwal - Thermal mass flow meters and controller
Toshniwal - Thermal mass flow meters and controllerSangeethaS99
 
Kidde Alarmline SenseTek-NL
Kidde Alarmline SenseTek-NLKidde Alarmline SenseTek-NL
Kidde Alarmline SenseTek-NLHans Bronkhorst
 
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygenOxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygenENVIMART
 
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
Katalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamika
Katalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamikaKatalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamika
Katalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamikaPT. Tridinamika Jaya Instrument
 
Data Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME AC
Data Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME ACData Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME AC
Data Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME ACPT. Siwali Swantika
 
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 

Similar to Liquisys m clm223 253 endress+hauser datasheet-conductivity-resistivity measurement (20)

Turbidity and Suspended Solids Transmitter-COM223/253
Turbidity and Suspended Solids Transmitter-COM223/253Turbidity and Suspended Solids Transmitter-COM223/253
Turbidity and Suspended Solids Transmitter-COM223/253
 
Dissolved Oxygen Transmitter-Liquisys M COM223/253
Dissolved Oxygen Transmitter-Liquisys M COM223/253Dissolved Oxygen Transmitter-Liquisys M COM223/253
Dissolved Oxygen Transmitter-Liquisys M COM223/253
 
Liquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitter
Liquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitterLiquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitter
Liquisys m ccm223 ccm253 endress+hauser datasheet-chlorine transmitter
 
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensorsCondumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
 
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensorsCondumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
 
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
ECD Model T23 pH, ORP, Specific Ion, Dissolved Oxygen, Conduc􏰁tivity Transmit...
 
Dissolved Oxygen Sensors-Oxymax COS22/22D
Dissolved Oxygen Sensors-Oxymax COS22/22DDissolved Oxygen Sensors-Oxymax COS22/22D
Dissolved Oxygen Sensors-Oxymax COS22/22D
 
Flowphant t dtt31, dtt35 endress+hauser datasheet-flow switch
Flowphant t dtt31, dtt35 endress+hauser datasheet-flow switchFlowphant t dtt31, dtt35 endress+hauser datasheet-flow switch
Flowphant t dtt31, dtt35 endress+hauser datasheet-flow switch
 
ATI-Q46H-C4-Datasheet.pdf
ATI-Q46H-C4-Datasheet.pdfATI-Q46H-C4-Datasheet.pdf
ATI-Q46H-C4-Datasheet.pdf
 
Toshniwal - Thermal mass flow meters and controller
Toshniwal - Thermal mass flow meters and controllerToshniwal - Thermal mass flow meters and controller
Toshniwal - Thermal mass flow meters and controller
 
Kidde Alarmline SenseTek-NL
Kidde Alarmline SenseTek-NLKidde Alarmline SenseTek-NL
Kidde Alarmline SenseTek-NL
 
Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13
 
pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41
 
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygenOxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos51 d endress+hauser datasheet-measurement of the dissolved oxygen
 
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
 
Katalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamika
Katalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamikaKatalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamika
Katalog fluke-oscilloscope-the-toughest-scopes-ever-built-by-tridinamika
 
Data Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME AC
Data Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME ACData Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME AC
Data Teknis Gossen Metrawatt Installation Tester : PROFITEST PRIME | PRIME AC
 
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
 
Dissolved Oxygen Sensors-Oxymax COS51D
Dissolved Oxygen Sensors-Oxymax COS51DDissolved Oxygen Sensors-Oxymax COS51D
Dissolved Oxygen Sensors-Oxymax COS51D
 
Siemens sitrans LUT400 series
Siemens sitrans LUT400 seriesSiemens sitrans LUT400 series
Siemens sitrans LUT400 series
 

More from ENVIMART

Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfZinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfNickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfManganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfLead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfIron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfIron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfChromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfCopper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfCopper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfCopper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfCobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfCadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfArsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfAluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfENVIMART
 
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfENVIMART
 
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfENVIMART
 
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfENVIMART
 
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfENVIMART
 
B12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfB12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfENVIMART
 

More from ENVIMART (20)

Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfZinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfNickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfManganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfLead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfIron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfIron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
 
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfChromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfCopper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfCopper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfCopper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfCobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfCadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfArsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
 
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfAluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
 
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
 
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
 
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
 
B12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfB12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdf
 

Recently uploaded

See How do animals kill their prey for food
See How do animals kill their prey for foodSee How do animals kill their prey for food
See How do animals kill their prey for fooddrsk203
 
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Kalighat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Roomdivyansh0kumar0
 
Joint GBIF Biodiversa+ symposium in Helsinki on 2024-04-16
Joint GBIF Biodiversa+ symposium in  Helsinki on 2024-04-16Joint GBIF Biodiversa+ symposium in  Helsinki on 2024-04-16
Joint GBIF Biodiversa+ symposium in Helsinki on 2024-04-16Dag Endresen
 
VIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service Gorakhpur
VIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service GorakhpurVIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service Gorakhpur
VIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service GorakhpurSuhani Kapoor
 
ENVIRONMENTAL LAW ppt on laws of environmental law
ENVIRONMENTAL LAW ppt on laws of environmental lawENVIRONMENTAL LAW ppt on laws of environmental law
ENVIRONMENTAL LAW ppt on laws of environmental lawnitinraj1000000
 
Philippines-Native-Chicken.pptx file copy
Philippines-Native-Chicken.pptx file copyPhilippines-Native-Chicken.pptx file copy
Philippines-Native-Chicken.pptx file copyKristineRoseCorrales
 
885MTAMount DMU University Bachelor's Diploma in Education
885MTAMount DMU University Bachelor's Diploma in Education885MTAMount DMU University Bachelor's Diploma in Education
885MTAMount DMU University Bachelor's Diploma in Educationz xss
 
(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量sehgh15heh
 
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170simranguptaxx69
 
Abu Dhabi Sea Beach Visitor Community pp
Abu Dhabi Sea Beach Visitor Community ppAbu Dhabi Sea Beach Visitor Community pp
Abu Dhabi Sea Beach Visitor Community pp202215407
 
Spiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsSpiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsprasan26
 
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girlsMumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girlsPooja Nehwal
 
Hi FI Call Girl Ahmedabad 7397865700 Independent Call Girls
Hi FI Call Girl Ahmedabad 7397865700 Independent Call GirlsHi FI Call Girl Ahmedabad 7397865700 Independent Call Girls
Hi FI Call Girl Ahmedabad 7397865700 Independent Call Girlsssuser7cb4ff
 
VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130
VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130
VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130Suhani Kapoor
 
Sustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and ChallengesSustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and ChallengesDr. Salem Baidas
 
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service NashikRussian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashikranjana rawat
 

Recently uploaded (20)

See How do animals kill their prey for food
See How do animals kill their prey for foodSee How do animals kill their prey for food
See How do animals kill their prey for food
 
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Kalighat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
 
Joint GBIF Biodiversa+ symposium in Helsinki on 2024-04-16
Joint GBIF Biodiversa+ symposium in  Helsinki on 2024-04-16Joint GBIF Biodiversa+ symposium in  Helsinki on 2024-04-16
Joint GBIF Biodiversa+ symposium in Helsinki on 2024-04-16
 
VIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service Gorakhpur
VIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service GorakhpurVIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service Gorakhpur
VIP Call Girl Gorakhpur Aashi 8250192130 Independent Escort Service Gorakhpur
 
ENVIRONMENTAL LAW ppt on laws of environmental law
ENVIRONMENTAL LAW ppt on laws of environmental lawENVIRONMENTAL LAW ppt on laws of environmental law
ENVIRONMENTAL LAW ppt on laws of environmental law
 
Philippines-Native-Chicken.pptx file copy
Philippines-Native-Chicken.pptx file copyPhilippines-Native-Chicken.pptx file copy
Philippines-Native-Chicken.pptx file copy
 
885MTAMount DMU University Bachelor's Diploma in Education
885MTAMount DMU University Bachelor's Diploma in Education885MTAMount DMU University Bachelor's Diploma in Education
885MTAMount DMU University Bachelor's Diploma in Education
 
(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(RIYA) Kalyani Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
 
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
 
Call Girls In Dhaula Kuan꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCe
Call Girls In Dhaula Kuan꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCeCall Girls In Dhaula Kuan꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCe
Call Girls In Dhaula Kuan꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCe
 
Abu Dhabi Sea Beach Visitor Community pp
Abu Dhabi Sea Beach Visitor Community ppAbu Dhabi Sea Beach Visitor Community pp
Abu Dhabi Sea Beach Visitor Community pp
 
Spiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsSpiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnids
 
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girlsMumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
 
FULL ENJOY Call Girls In kashmiri gate (Delhi) Call Us 9953056974
FULL ENJOY Call Girls In  kashmiri gate (Delhi) Call Us 9953056974FULL ENJOY Call Girls In  kashmiri gate (Delhi) Call Us 9953056974
FULL ENJOY Call Girls In kashmiri gate (Delhi) Call Us 9953056974
 
Gandhi Nagar (Delhi) 9953330565 Escorts, Call Girls Services
Gandhi Nagar (Delhi) 9953330565 Escorts, Call Girls ServicesGandhi Nagar (Delhi) 9953330565 Escorts, Call Girls Services
Gandhi Nagar (Delhi) 9953330565 Escorts, Call Girls Services
 
Hi FI Call Girl Ahmedabad 7397865700 Independent Call Girls
Hi FI Call Girl Ahmedabad 7397865700 Independent Call GirlsHi FI Call Girl Ahmedabad 7397865700 Independent Call Girls
Hi FI Call Girl Ahmedabad 7397865700 Independent Call Girls
 
VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130
VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130
VIP Call Girls Service Chaitanyapuri Hyderabad Call +91-8250192130
 
Sustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and ChallengesSustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and Challenges
 
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service NashikRussian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
 

Liquisys m clm223 253 endress+hauser datasheet-conductivity-resistivity measurement

  • 1. TI193C/07/EN/13.11 71130248 Technical Information Liquisys M CLM223/253 Conductivity/Resistivity Measurement Transmitter for conductive and inductive sensors Application The modular design of the Liquisys M CLM223/253 allows easy adaption of the transmitter to a variety of customer requirements. Starting with the basic version for "measurement and alarm generation", the transmitter can be equipped with additional software and hardware modules for special applications. These modules can also be retrofitted as required. Application • Ultrapure water • Water treatment • Ion exchanger, reverse osmosis • Cooling water desalinization • Sewage Your benefits • Field or panel-mounted housing • Universal application • Simple handling – Logically arranged menu structure – Ultrasimple two-point calibration • Safe operation – Overvoltage (lightning) protection – Direct access for manual contact control – User-defined alarm configuration The basic unit can be extended with: • 2 or 4 additional contacts for use as: – Limit contacts (also for temperature) – P(ID) controller – Timer for simple rinse processes – Complete cleaning with Chemoclean • Plus package: – User defined current output characteristics – Automatic cleaning trigger on alarm or limit violation – Ultrapure water monitoring acc. to USP (United States Pharmacopeia) and EP (European Pharmacopoeia) (conductive) – Polarization detection (conductive) – Concentration measurement – Temperature compensation via table – Process Check System (PCS): live check of the sensor – Adaptive calibration with installation factor (inductive) • HART or PROFIBUS-PA/-DP • 2nd current output for temperature, main measured value or actuating variable • Current input for flow rate monitoring with controller shut off or for feedforward control
  • 2. Liquisys M CLM223/253 2 Endress+Hauser Function and system design Features of the basic version Conductive or inductive Two instrument versions for measurement with conductive (two electrode) sensors or inductive sensors are available. The use of inductive sensors that are less sensitive to soiling than conductive sensors is recommended for high conductivity measurement, concentration measurement or adhering media. Measuring of conductivity and resistivity (conductive) This is selected via the menu. During measurement, the value measured can be displayed in the other measuring mode. The temperature is displayed at the same time or, if desired, not shown at all. Temperature compensation The following temperature compensation selections are available: • Linear • NaCl curve according to IEC 746 • Ultrapure water NaCl (neutral compensation) • Ultrapure water HCl (acid compensation, also ammonia) The reference temperature is user defined, the standard value is 25 °C (77 °F). Configuration Different alarms are required depending on application and operator. Therefore the transmitter permits independent configuration of the alarm contact and error current for each individual error. Unnecessary or undesirable alarms can be suppressed in this manner. Up to four contacts can be used as limit contacts (also for temperature), to implement a P(ID) controller or for cleaning functions. Direct manual operation of the contacts (bypassing the menu) provides quick access to limit, control or cleaning contacts, permitting speedy correction of deviations. The serial numbers of the instrument and modules and the order code can be called up on the display. The cell constant can be edited and even calibrated for demanding special applications. Additional functions of the plus package Current output configuration In order to output wide measuring ranges while still achieving a high resolution in specific ranges, the current output can be configured as required via a table. This permits bilinear or quasi-logarithmic curves, etc. Polarization detection Polarization effects in the boundary layer between the sensor and the medium to be measured limit the measuring range of conductive conductivity sensors. The transmitter can detect polarization effects using an innovative, intelligent signal evaluation process. Live check The live check issues an alarm when the sensor signal does not change over a defined period of time. This may be caused by blocking, passivation, separation from the process, etc. Ultrapure water monitoring acc. to USP (United States Pharmacopeia) and EP (European Pharmacopoeia) Ultrapure water monitoring according to USP <645> and EP means that the uncompensated conductivity and the temperature are measured and compared to a table. The transmitter (conductive with additional contacts) comes with the following functions: • Monitoring of "Water for Injection" (WFI) according to USP and EP • Monitoring of "Highly Purified Water" (HPW) according to EP • Monitoring of "Purified Water" (PW) according to EP The user-adjustable pre-alarm indicates undesirable operating values in due time. Full compliance with USP or EP requires the use of a precisely calibrated sensor, for example, the CLS16. Concentration measurement The conversion from conductivity to concentration is effected using four user-definable concentration curves. This permits concentrations to be displayed in %, ppm, mg/l or TDS (total dissolved solids).
  • 3. Liquisys M CLM223/253 Endress+Hauser 3 Adaptive calibration for determination of the installation factor (inductive) Inductive measuring sensors must normally be installed in pipes at a required minimum distance from the pipe wall. If this minimum distance is not observed, the measured value changes. The built-in adaptive calibration using the installation factor allows you to compensate for this once the sensor is installed. Second current output The second current output can be configured for temperature, main measured value (conductivity, resistivity, concentration) or actuating variable. Current input The current input of the transmitter allows two different applications: controller shut-down in case of lower flow rate violation or total failure in the main flow as well as feedforward control. Both functions are also combinable. Measuring system A complete measuring systems comprises: • The transmitter Liquisys M CLM223 or CLM253 • A sensor with or without an integrated temperature sensor • A measuring cable CYK71 (conductive), CPK9 for Condumax H CLS16 or CLK5 (inductive) Options: extension cable, junction box VBM a0008937 Complete measuring system Liquisys M CLM223/253 1 2 3 4 Conductive sensor CLS15 Liquisys M CLM253 Liquisys M CLM223 Inductive sensor CLS54 5 6 7 Inductive sensor CLS50 Conductive sensor CLS21 Immersion assembly CLA111 ENDRESS+HAUSER LIQUISYS S CAL REL REL1 REL1 ALARM REL2 REL2 E ENDRESS+HAUSER LIQUISYS S CAL REL REL1 REL1 ALARM REL2 REL2 E 1 2 3 4 5 6 7
  • 4. Liquisys M CLM223/253 4 Endress+Hauser Input Measured variables Conductivity, resistivity, temperature Measuring range Cable specification Cell constant Temperature sensors Pt 100, Pt 1000, NTC 30K Measuring frequency Binary inputs Current input 4 to 20 mA, galvanically separated Load: 260 Ω at 20 mA (voltage drop 5.2 V) Output Output signal 0/4 to 20 mA, galvanically separated, active Conductivity (conductive): 0 to 600 mS/cm (uncompensated) Conductivity (inductive): 0 to 2000 mS/cm (uncompensated) Resistivity: 0 to 200 MΩ⋅cm Concentration: 0 to 9999 (%, ppm, mg/l, TDS) Temperature: -35 to +250 °C (-31 to +482 °F) Cable length (conductive): conductivity: max. 100 m (328.1 ft) (CYK71) resistivity: max 15 m (49.22 ft) (CYK71) Cable length (inductive): max 55 m (180.46 ft) (CLK5) Cable resistance CYK71: 165 Ω/km (conductivity measurement) Adjustable cell constant: k = 0.0025 to 99.99 cm-1 Conductivity, resistivity (conductive): 170 Hz to 2 kHz Conductivity (inductive): 2 kHz Voltage: 10 to 50 V Power consumption: max. 10 mA HART Signalkodierung Frequency Shift Keying (FSK) + 0,5 mA über Stromausgangssignal Datenübertragungsrate 1200 Baud Galvanische Trennung ja PROFIBUS PA Signalkodierung Manchester Bus Powered (MBP) Datenübertragungsrate 31,25 kBit/s, Spannungsmodus Galvanische Trennung ja (IO-Module) PROFIBUS DP Signalkodierung RS485 Datenübertragungsrate 9,6 kBd, 19,2 kBd, 93,75 kBd, 187,5 kBd, 500 kBd, 1,5 MBd Galvanische Trennung ja (IO-Module)
  • 5. Liquisys M CLM223/253 Endress+Hauser 5 Signal on alarm 2.4 or 22 mA in case of an error Load maximum 500 Ω Linearization transmission behaviour Resolution max. 700 digits/mA Min. distance for 0 / 4 to 20 mA signal Isolation voltage max. 350 VRMS/500 V DC Overvoltage protection according to EN 61000-4-5 Auxiliary voltage output Contact outputs Limit contactor Controller Alarm Conductivity: Resistivity: Concentration: Actuating variable: Temperature: adjustable adjustable adjustable adjustable adjustable Conductivity: Measured value 0 to 1.999 μS/cm 0.2 μS/cm Measured value 0 to 19.99 μS/cm 2 μS/cm Measured value 20 to 199.9 μS/cm 20 μS/cm Measured value 200 to 1999 μS/cm 200 μS/cm Measured value 2 to 19.99 mS/cm 2 mS/cm Measured value 20 to 2000 mS/cm 20 mS/cm Resistivity Measured value 0 to 199.9 kΩ⋅cm 20 kΩ⋅cm Measured value 200 to 1999 kΩ⋅cm 200 kΩ⋅cm Measured value 2 to 19.99 MΩ⋅cm 2.0 MΩ⋅cm Measured value 20 to 200 MΩ⋅cm 20 MΩ⋅cm Concentration no minimum distance Temperature 15 °C Output voltage: 15 V ± 0.6 Output current: max. 10 mA Switching current with ohmic load (cos ϕ = 1): max. 2 A Switching current with inductive load (cos ϕ = 0.4): max. 2 A Switching voltage: max. 250 V AC, 30 V DC Switching power with ohmic load (cos ϕ = 1): max. 500 VA AC, 60 W DC Switching power with inductive load (cos ϕ = 0.4): max. 500 VA AC, 60 W DC Pickup/dropout delay: 0 to 2000 s Function (adjustable): pulse length/pulse frequency controller Controller response: PID Control gain Kp: 0.01 to 20.00 Integral action time Tn: 0.0 to 999.9 min Derivative action time Tv: 0.0 to 999.9 min Period for pulse length controller: 0.5 to 999.9 s Frequency for pulse frequency controller: 60 to 180 min-1 Basic load: 0 to 40% of max. set value Function (selectable): Latching/momentary contact Alarm threshold adjustment range: Conductivity, resistivity, concentration, temperature, USP, EP: complete measuring range Alarm delay: 0 to 2000 s (min)
  • 6. Liquisys M CLM223/253 6 Endress+Hauser Protocol specific data HART Manufacturer ID 11h Device type code 0092h (ind. measured), 0093h (cond. measured) Transmitter specific revision 0001h HART specification 5.0 DD files www.products.endress.com/profibus Load HART 250 Ω Device variables None (dynamic variables PV, SV, only) Features supported - PROFIBUS PA Manufacturer ID 11h Ident number 1515h Device revision 11h Profile version 2.0 GSD files www.products.endress.com/profibus GSD file version Output values Main value, temperature value Input values Display value of PLC Features supported Device locking: The device can be locked by hardware or software. PROFIBUS DP Manufacturer ID 11h Ident number 1521h Profile version 2.0 GSD files www.products.endress.com/profibus GSD file version Output values Main value, temperature value Input values Display value of PLC Features supported Device locking: The device can be locked by hardware or software.
  • 7. Liquisys M CLM223/253 Endress+Hauser 7 Power supply Electrical connection The instrument has protection class II and is generally operated without protective earth connection. To ensure the measuring stability and the function for conductive sensors you have to connect the outer screen of the sensor cable to the PE terminal. C07-CLM2x3xx-04-06-00-xx-001.eps Electrical connetion of the transmitter A B C D E F G H Sensor (conductive) Sensor (inductive) Temperature sensor Signal output 1 conductivity Signal output 2 variable Binary input 1 (Hold) Binary input 2 (Chemoclean) Aux. voltage output I J K L M N O Alarm (current-free contact position) Relay 1 (current-free contact position) Relay 2 (current-free contact position) Relay 3 (current-free contact position) Relay 4 (current-free contact position) Current input 4 ... 20 mA Power supply 11 12 13 47 48 49 57 58 59 15 V 10-50 V 81 82 10-50 V 93 94 41 42 43 Lf mA 31 32 mA 33 34 51 52 53 54 55 56 S 84 83 15 16 84 PE 83 L+ L– N L1 ~ = PE 23 + 24 - mA 15 16 85+ 86 _ A B C D E F G H I J K L M N O
  • 8. Liquisys M CLM223/253 8 Endress+Hauser Sensor connection You require screened special measuring cables to connect conductivity sensors to the transmitter. To extend the measuring cable, use junction box and extension cable (see accessories). Supply voltage Depending on ordered version: 100/115/230 V AC +10/-15 %, 48 to 62 Hz 24 V AC/DC +20/-15 % Fieldbus connection Power consumption max. 7.5 VA Mains protection Fine-wire fuse, medium-slow blow 250 V/3.15 A C07-CLM2x3xx-04-06-00-xx-007.eps Connection of conductive sensors a0006611 Connection of inductive sensors GN WH YE GN WH YE BK BK CYK71/CPK9 J 83 16 15 84 2 3 1 11 12 13 PE RD WH CLK 5 J 83 16 15 84 11 12 13 GN WH YE BU RD RD BU S HART Supply voltage n/a, active current outputs Integrated reverse voltage protection n/a, active current outputs PROFIBUS PA Supply voltage 9 V to 32 V, max. 35 V Polarity sensitive no FISCO/FNICO compliant acc. to IEC 60079-27 no PROFIBUS DP Supply voltage 9 V to 32 V, max. 35 V Polarity sensitive n/a FISCO/FNICO compliant acc. to IEC 60079-27 no
  • 9. Liquisys M CLM223/253 Endress+Hauser 9 Performance characteristics Reference temperature 25 °C (77 °F); adjustable for the compensation of the medium temperature Resolution Maximum measured error1) Repeatability1 Temperature compensation Temperature offset ±5 °C; for the adjustment of the temperature display Conductivity: Temperature: depending on the measuring range: 0.001 μS/cm to 1.999 μS/cm and k ≤ 0.5 cm-1 0.1 °C 1) acc. to IEC 746-1, for nominal operating conditions Conductivity: Display: max. 0.5 % of measured value ± 4 digits Conductivity signal output: max. 0.75 % of current output range Resistivity: Display: max. 0.5 % of measured value ± 4 digits Resistivity signal output: max. 0.75 % of current output range Temperature: Display: max. 1.0 % of measuring range Temperature signal output: max. 1.25 % of current output range Conductivity: max. 0.2 % of measured value ± 2 digits Resistivity: max. 0.2 % of measured value ± 2 digits Range: -35 to +250 °C (-31 to +482 °F) Types of compensation: uncompensated, linear, NaCl, table; conductive only: ultrapure water NaCl, ultrapure water HCl
  • 10. Liquisys M CLM223/253 10 Endress+Hauser Installation Installation instructions a0005733 Field instrument a0005737 Mounting on pipes 1 - 3 Mounting screws and mounting plate CAL REL REL1 REL1 ALARM REL2 REL2 E Pg 7 Pg 16 M 5 11/ 0.43 Ø 6 / Ø 0.24 115 / 4.53 157 / 6.18 247 / 9.72 70 / 2.76 70 / 2.76 154 / 6.06 170 / 6.69 mm / inch Pg 13.5 1 2 3 1 1 3 Ø max. 60 / 2.36 mm / inch
  • 11. Liquisys M CLM223/253 Endress+Hauser 11 a0005734 View into the field instrument 1 2 3 4 Removable electronics box Partition plate Terminal blocks Fuse a0005736 Wall mounting of the field instrument 1 Mounting holes 2 Protecting cap a0005738 Mounting of the field instrument with mounting post and weather protection cover 1 - 3 Mounting holes 1 2 3 4 1 2 154 / 6.06 mm / inch Ø 6 / Ø 0.24 3 1 2 3
  • 12. Liquisys M CLM223/253 12 Endress+Hauser a0005735 Dimensions panel-mounted instrument a0005739 Installation of the panel-mounted instrument 1 Wall of control cabinet 2 Gasket 3 Tensioning screws * Required installation depth 90 / 3.54 139 / 5.47 96 / 3.78 6 / 0.24 mm / inch 92 / 3.62 149.5 / 5.89 1 2 3 3 3 92 / 3.62 +0.5 +0.02 92 / 3.62 +0.5 +0.02 90 / 3.54 57 / 2.24 33 / 1.30 approx. 25 / 0.98 139 / 5.47 max. 45 / 1.77 6 / 0.24 96 / 3.78 92 / 3.62 approx. 165 / 6.50* mm / inch
  • 13. Liquisys M CLM223/253 Endress+Hauser 13 Environment Ambient temperature -10 to +55 °C (+14 to +131 °F) Storage temperature –25 to +65 °C (-13 to +149 °F) Electromagnetic compatibility Interference emission and interference immunity as per EN 61326-1:2006, EN 61326-2-3:2006 Ingress protection Electrical safety according EN/IEC 61010-1:2001, Installation Category II, for use up to 2000 m above sea level CSA Apparatus with CSA General Purpose Approval are certified for indoor use. Relative humidity 10 to 95%, non-condensing Pollution degree The product is suitable for pollution degree 2. Mechanical construction Dimensions Weight Materials Terminals Panel mounted instrument: IP 54 (front), IP 30 (housing) Field instrument: IP 65 / tightness acc. to NEMA 4X Panel mounted instrument: 96 x 96 x 145 mm (3.78 x 3.78 x 5.71 inches) Installation depth: approx. 165 mm (6.50") Field instrument: 247 x 170 x 115 mm (9.72 x 6.69 x 4.53 inches) Panel mounted instrument: max. 0.7 kg (1.5 lb) Field instrument: max. 2.3 kg (5.1 lb) Housing of panel mounted instrument: Polycarbonate Field housing: ABS PC FR Front membrane: Polyester, UV-resistant Cross section max. 2.5 mm2
  • 14. Liquisys M CLM223/253 14 Endress+Hauser Operability Operating concept All instrument control functions are arranged in a logical menu structure. Following access code entry, the individual parameters can be easily selected and modified as needed. Display elements C07-CLM2x3xx-19-06-00-en-001.eps Operating elements 1 LC display for display of measured values, configuration data and current menu field 2 Field for user labeling 3 4 main control keys for calibration and instrument configuration 4 Key for switching between automatic/manual operation of the relays 5 LED indicators for limit contactor relay (switch status) 6 LED indicator for alarm function 7 Display of active contact and key for relay switching in manual mode The display simultaneously shows the current measured value and the temperature - the essential process data. Brief information texts in the configuration menu provide assistance with parameter configuration. Certificates and approvals 4 symbol Declaration of conformity The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives. The manufacturer confirms successful testing of the product by affixing the 4 symbol. Ex approval for zone 2 Application of transmitter as related electrical equipment in non-hazardous area or in simple pressurized apparatus; application of sensor in hazardous area zone 2 CSA General Purpose C.M2.3-..2... C.M2.3-..3... C.M2.3-..7... 12.15 mS/cm R232 1 2 3 4 5 6 7 Setpoint C.M2.3-..4... C.M223-..6... C.M253-..A... ATEX II (3)G (Ex nL) IIC
  • 15. Liquisys M CLM223/253 Endress+Hauser 15 Ordering information Product structure Additional functions of the Plus package • Current output table to cover large ranges with varying resolution, fields O33x • Process Check System (PCS): live check of the sensor, function group P • Ultrapure water monitoring for "Water for injection" (WFI) and "Purified water" (PW) acc. to United States Pharmacopeia (USP) and European Pharmacopoeia (EP) with pre-alarm (conductive, additional contacts necessary), fields R26x and R27x • Polarization detection (conductive), function group P • Concentration measurement, function group K • Temperature compensation via coefficient table, function group T • Adaptive calibration with installation factor (inductive), fields C13x • Automatic cleaning function start, field F8 Input, software CD Conductivity/resistivity measurement (conductive two-electrode sensor) CS Conductivity/resistivity measurement (conductive two-electrode sensor) with additional functions (Plus package) ID Conductivity measurement (inductive sensor) IS Conductivity measurement (inductive sensor) with additional functions (Plus package) Power supply, approval A 24 V AC/DC; ATEX II (3)G (Ex nL) IIC (CLM253 only) 0 230 V AC 1 115 V AC 2 230 V AC; CSA Gen. Purp. 3 115 V AC; CSA Gen. Purp. 4 230 V AC; ATEX II 3G [Ex nL] IIC 5 100 V AC 6 24 V AC/DC; ATEX II 3G [Ex nL] IIC (CLM223 only) 7 24 V AC/DC; CSA Gen. Purp. 8 24 V AC/DC Output 0 1 x 20 mA, primary value 1 2 x 20 mA, primary value + secondary value 3 PROFIBUS PA 4 PROFIBUS DP 5 1 x 20 mA, primary value, HART 6 2 x 20 mA, primary value, HART + secondary value Additional contacts 05 not selected 10 2 relays (limit/P(ID)/timer) 15 4 relays (limit/P(ID)/Chemoclean) (not with PROFIBUS DP) 16 4 relays (limit/P(ID)/timer) (not with PROFIBUS DP) 20 1 x 4 ... 20 mA input + 2 relays (limit/P(ID)/timer) 25 1 x 4 ... 20 mA input + 4 relays (limit/P(ID)/Chemoclean) (not with PROFIBUS DP) 26 1 x 4 ... 20 mA input + 4 relays (limit/P(ID)/timer) (not with PROFIBUS DP) Additional features (CLM223 only) PRL Protective layer Marking 1 Tagging (Tag), see additional spec. CLM253- complete order code CLM223-
  • 16. Liquisys M CLM223/253 16 Endress+Hauser Scope of delivery The delivery of the field instrument includes: • 1 transmitter CLM253 • 1 plug-in screw terminal • 1 cable gland Pg 7 • 1 cable gland Pg 16 reduced • 2 cable glands Pg 13.5 • 1 Operating Instructions BA193C/07/EN • versions with HART communication: 1 Operating Instructions Field Communication with HART, BA208C/07/EN • versions with PROFIBUS communication: 1 Operating Instructions Field Communication with PROFIBUS PA/DP, BA209C/07/EN • versions with explosion protection for hazardous area zone II (ATEX II 3G): Safety Instructions for use in explosion-hazardous areas, XA194C/07/A3 The delivery of the panel mounted instrument includes: • 1 transmitter CLM223 • 1 set of plug-in screw terminals • 2 tensioning screws • 1 Operating Instructions BA193C/07/EN • versions with HART communication: 1 Operating Instructions Field Communication with HART, BA208C/07/EN • versions with PROFIBUS communication: 1 Operating Instructions Field Communication with PROFIBUS PA/DP, BA209C/07/EN • versions with explosion protection for hazardous area zone II (ATEX II 3G): Safety Instructions for use in explosion-hazardous areas, XA194C/07/A3 Accessories Sensors • Condumax W CLS12 Conductive conductivity sensor for standard, Ex and high-temperature applications; Ordering acc. to version, see Technical Information TI082C/07/en • Condumax W CLS13 Conductive conductivity sensor for standard, Ex and high-temperature applications; Ordering acc. to version, see Technical Information TI083C/07/en • Condumax W CLS15 Conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex); Ordering acc. to version, see Technical Information TI109C/07/en • Condumax H CLS16 Hygienic conductive conductivity sensor for pure and ultra-pure water applications (incl. Ex); Ordering acc. to version, see Technical Information TI227C/07/en • Condumax W CLS19 Conductive conductivity sensor for pure and ultra-pure water applications; Ordering acc. to version, see Technical Information TI110C/07/en • Condumax W CLS21 Conductive conductivity sensor for applications with middle to high conductivity (incl. Ex); Ordering acc. to version, see Technical Information TI085C/07/en • Indumax P CLS50 Inductive conductivity sensor for standard, Ex and high-temperature applications Ordering acc. to version, see Technical Information (TI118C/07/en) • Indumax H CLS52 Inductive conductivity sensor with short response time for food applications Ordering acc. to version, see Technical Information (TI167C/07/en) • Indumax H CLS54 Inductive conductivity sensor in certified hygienic design for applications in: food, beverages, pharmaceuticals and biotechnology Ordering acc. to version, see Technical Information (TI400C/07/en)
  • 17. Liquisys M CLM223/253 Endress+Hauser 17 Connection accessories CYK71 measuring cable • Non-terminated cable for the connection of sensors (e.g. conductivity sensors) or the extension of sensor cables • Sold by the meter, order numbers: – non-Ex version, black: 50085333 – Ex version, blue: 51506616 Extension cable CLK5 • For inductive conductivity sensors, for extension via the VBM junction box, sold by the meter • Order no.: 50085473 Junction box VBM • For cable extension • 10 terminals • Cable entries: 2 x Pg 13.5 or 2 x NPT ½" • Material: aluminum • Ingress protection: IP 65 (i NEMA 4X) • Order numbers: – cable entries Pg 13.5: 50003987 – cable entries NPT ½": 51500177 • Four-pole metal plug M12 for fieldbus connection order no. 51502184 C07-CM12xxx-02-06-00-xx-002.eps M12 plug with socket n.c. n.c. A (PA-) A (PA-) S S B (PA+) B (PA+) A (PA-) S (PE) B (PA+) 45 1.77” 7 0.28” 150 5.91”
  • 18. Liquisys M CLM223/253 18 Endress+Hauser Mounting accessories CYY101 weather protection cover for field devices, absolutely essential if operating the unit outdoors • Material: stainless steel 1.4031 (AISI 304) • Order No. CYY101-A Post mounting kit • For mounting of field housing on horizontal or vertical pipes (Ø max. 60 mm (2.36")) • Material: stainless steel 1.4301 • order no. 50086842 a0005676 Post mounting kit a0012427 Weather protection cover for field devices * Wall and post mounting ** Rail mounting 70 / 2.76 90 / 3.54 Ø 6 / 0.24 Ø 8 / 0.31 70 / 2.76 90 / 3.54 M6 60 / 2.36 90 / 3.54 70 / 2.76 Ø 8 / 0.31 30 / 1.18 Ø max. 60 / 2.36 mm / inch
  • 19. Liquisys M CLM223/253 Endress+Hauser 19 CYY102 universal post • Square pipe for mounting transmitters • Material: stainless steel 1.4301 (AISI 304) • Order No. CYY102-A a0005742 Universal post Buffer solutions Precision calibration solutions, acc. to SRM (Standard reference material) of NIST, reference temperature 25 °C (77 °F), with temperature table • CLY11-A, 74.0 μS/cm, 500 ml (16.9 fl.oz); order no. 50081902 • CLY11-B, 149.6 μS/cm, 500 ml (16.9 fl.oz); order no. 50081903 • CLY11-C, 1.406 mS/cm, 500 ml (16.9 fl.oz); order no. 50081904 • CLY11-D, 12.64 mS/cm, 500 ml (16.9 fl.oz); order no. 50081905 • CLY11-E, 107.0 mS/cm, 500 ml (16.9 fl.oz); order no. 50081906 Optoscope Optoscope • Interface between transmitter and PC / laptop for service purposes. • The Windows software "Scopeware" required for the PC or laptop is supplied with the Optoscope. The Optoscope is supplied in a sturdy plastic case with all the accessories required. • Order no. 51500650 1495 / 58.86 60 x 60 / 2.36 x 2.36 120 / 4.72 150 / 5.91 mm/inch