2. ephiós handheld ERG provides the
ophthalmologist the capability to
evaluate function of the retinal cells
and help detecting malfunction in
the eye retina. The instrument
follows ISCEV Standard. The
instrument can be used as a stand
alone unit for test & analysis. The
curves are stored in the instrument
memory for easy transfer to the
personal computer for later
analysis.
Applications
¢ For clinical purpose ephiós ERG is used in following cases:
¡ Confirmation of Ophthalmologic & Neurological Diseases
¡ Unexplained Visual Loss
¡ Paediatric Neurology Key benefits at a glance
¡ Opacities in Media
¡ Monitoring Health – Toxicity — Ergonomic design for single
¡ Detection of the Disease or Carrier States of Inherited Visual Disorders
¡ Quantitative Assessment of Visual Disease handed operation
¡ Assessment of Retinal and Optic Nerve Function Following Trauma
— Quick and effective test of the
¢ For Ophthalmologic & Neurological Research the instrument can be used for a number of retinal function in a screening test
different applications where the retinal cell functions are tested. This could be evaluation of
eye diseases, neurological diseases, side effects of drugs, basic research of retinal cells, before and after surgery.
and basic knowledge of retinal function.
— Easy operation for ERG on
¢ The instrument is also a very good teaching tool for understanding of the retinal function prematures, young children or
and electrophysiology.
patients in bed.
¢ In pharmaceutical industries the instrument is used to evaluate the side effects of drugs
on the central nervous system. — Portable for ERG test in or outside
of the clinic. No PC needed.
Technical Specification*
Patient safety: No connections to the mains during operation.
Keyboard: easy to use with 5 keypads: 1. Flash, 2. Send (memory, printer, computer, display),
— Battery powered, fully integrated
3. Functions, 4. step down, 5. step up ERG instrument.
Changeable functions: Flash frequency Manual, 0.5, 1, 30 Hz
Flash intensities: 0.003, 0.006, 0.012, 0.024, 0.048, 0.096, 0.19, 0.38, 0.76, 1.5, 3, 6, 12, 24, 48 cds/m2
Number of responses for averaging: up to 1024. — No connections to mains needed
Sample/display time: 50, 150, 200, 500 ms
during test.
Amplifier: 2 channels
Gain 4000 (ERG), 20000 (VEP) for
Noise <2uV p-p (1-100Hz) — Instrument display allows for direct
Band pass 0.5 – 300Hz
CMRR >90dB interpretation and decision.
Imp. >108 ohm
Memory: up to 10 patients, 10 curves for each, can be stored internally for downloading to a
— Stores up to 100 curves for later
printer or computer afterwards.
Battery: 7.4V, 1.8 Ah rechargeable. transfer to PC.
Display: 75x53 mm, automatic size correction of signals.
Evaluation: automatic evaluation of curve amplitude and implicit time
Stimuli: White LEDs. Intensity from 0.003 Cd*s/m2 to 48 Cd*s/m2 in 15 steps of 0.3log units — Direct ready to go handling from
measured at 2cm from front of stimulator tube. buttons on instrument.
Scotopic 3.0ERG = flash intensity 3cd*s/m2.
Background light 30 Cd/m2 for photopic test
Communication Interface: USB or RS232 serial for out-prints and computer connection. — Flash stimulator placed in front.
Accessories: computer, printer, computer software for evaluation of signals, electrodes and more.
* Specifications are subject to change without notice.
Automatic artefact rejection. Averaging of curves during test (1 - 1024).
Automatic estimated B-wave amplitude during test.
PC software for transfer of results to computer.
ephiós ab
C/o Andersson C/o Swedish Trade Council - Exportrådet
Kompanigatan 19 Embassy of Sweden
S-60209 Norrköping Nyaya Marg, Chanakyapuri
Sweden New Delhi - 110 021, INDIA
www.ephios.com info@ephios.com