EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

INTRAOCULAR PRESSURE (IOP) MEASUREMENT WITH DIATON® PEN
TONOMETER

INTRODUCTION:
DIATON® Tonometer Pen ( BiCOM Inc, USA) is Non-corneal, Non-contact, Transpalpebral
(through eyelid) handheld tonometer.
DIATON® PEN TONOMETER is a handheld, pen-shaped device used for measuring
intraocular pressures (IOP). The DIATON Pen tonometer calculates pressure by measuring the
response of a free-falling floater, as it rebounds against the tarsal plate of the eyelid and the
sclera.
The patient is positioned so that the tip of the device and lid are overlying sclera, in either sitting
or supine position.
Non-corneal and transpalpebral tonometry does not involve contact with the cornea and does not
require topical anesthetic during routine use. Contact lenses do not need to be taken out and
central corneal thickness (CCT) and other corneal properties do not influence the IOP
measurement.
In Emergency Department and Acute Care settings IOP measurement is indicated in cases of
acute eye injury, acute eye pain, red eye, and acute loss of visual acuity. Normal intraocular
pressure ranges between 10-20 mmHg.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

EQUIPMENT, SUPPLIES & MAINTENANCE:

To measure intraocular pressure with a DIATON® Tonometer Pen, the following equipment is
needed: a pen tonometer, alcohol swab – to wipe off the tip of the tonometer between patients.
NO need for topical anesthetic (either properacaine or tetracaine) or tip covers since the
measurement is taken over upper eyelid.
DIATON® Tonometer Pens are battery operated devices.
Tonometer will indicate “U” on the display if the battery is low. Batteries are button CR2032.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

The parts of the tonometer are: the stainless steel tip, the operation button, and an LCD display.
The operation button is used for turning the device ON and OFF and to obtain AVERAGE aget
measurements.
The LCD screen provides information about the sequence of readings, possible errors and tells
you the pressure reading.

CALIBRATION:
DIATON® Tonometer Pen comes with a carry case which includes Pressure Selector Testing
Device. To check for calibration, simply take the measurement on the Testing Plate. Results
should be within ±2mm of 20mm. LCD display will also indicate any other errors or if the
device needs to be cleaned.
Frequency of testing – once a day.

MEASURING IOP:
The patient should be relaxed and seated at the edge of the chair and reclining back, with head
titled back as horizontal as possible or lying down flat prior to beginning. Tight shirt collars
should be loosened. Patient should extend a thumb forward at appx 45º to center the eye. User
should guide the patients to be in the correct position.
Putting the rod into working position
The working position of the rod is its fixed inside the device (the rod is not seen in the tip
area).
All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

Fig. 1

If the rod is seen in the tip area, it's necessary to put it into working position.
Keep the tonometer vertically with its tip downwards, as it is shown in fig.1.

Fig. 2

Turn the tonometer with its tip upwards, as it is shown in fig. 2.
All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

The rod automatically falls inside the tonometer.

Fig. 3

Turn the tonometer with its tip downwards.
The rod is fixed inside the device, as it is shown in fig.3; it means that it is in the proper
working position.

Turning the tonometer ON

Fig. 4
Press momentarily the OPERATION button once.
"0000" symbol appears on the display as it is shown in fig. 4.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

If after turning ON the tonometer produces an interrupted sound signal it means that its body deviates
from vertical position. Move the device so it reaches the vertical position of the tonometer and the
sound signal will stop.

IOP measuring must be realized with the tonometer's position being strictly vertical!
To get the reliable results it is important to learn how to reach the vertical position of the
tonometer before starting IOP measurement in patients.

Practicing with the tonometer on the test plate

Fig. 5

The testing device is built in the tonometer's case (Fig.5). Its main assignment is checking the
tonometer's capacity for work before starting the operation and when necessary.
Use the testing device to easily find the vertical position and practice Test measurements.
Prepare the tonometer for work — the rod is in the working position, the tonometer is turned
ON.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

Fig. 6
Place the tonometer's tip in the indents of the testing device as it is shown in fig.6.
Reach the tonometer's vertical position when there's no interrupted sound signal.

Fig. 7
Fluently lower the tonometer's body down until the rod falls. The display must show the test
value from 18 to 22, as it is shown in fig.7. The checking is finished.

At other test values the tonometer is considered inoperative.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

During checking the tonometer's capacity for work keep it in the vertical position, there should
be no sound signal. If after the measurement the display shows L symbol it means that user took
the measurement while the device was beeping. It is necessary to put the rod into the working
position again and repeat the operation on the testing device.

Measuring of IOP
It is possible to measure IOP with the patient being in sitting or reclining position. Horizontal
position of the patient's head is required.
Disinfect the tip of the tonometer with alcohol swab according to the directions of the Operation
Manual.

Fig. 8a

Fig. 8b

If you are Right Handed - Stand at the LEFT hand side of the patient as it is shown in fig.8.
Put the rod into working position, turn the tonometer ON.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

Fig. 9

Place and fix the patient's glance using the test object (extend patients Thumb forward) the
patient's glance line being oriented approximately at the angle of 45° as it is shown in fig.9.

Fig. 10

Stretch the upper eyelid with a finger of a free hand so that the edge of the upper eyelid
coincides with the limb (fig.10). Keep the eyelid in this position.
Do not press the eyeball.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

Fig. 11

Place the edge of your palm with the tonometer on the patient's forehead. With the other hand
keep holding the eyelid in the necessary position (edge of the lid should be 1mm above the edge
of the limbus) (fig.11).

Place the tonometer's tip on the
eyelid in such a way that the front
part of the tip is as close to the front
edge of the upper eyelid as possible
without touching the eyelashes.
(fig.12, 13).
The influence zone of the tonometer's
rod must be the part of sclera
corresponding to сorona ciliaris in
12 o'clock meridian. Secure the
tonometer's vertical position (there's
no sound signal).
Fig. 12

Gently lower the tonometer's body
keeping its vertical position until the
rod falls on the eyelid which is
accompanied with the short sound
signal.

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

Fig. 13
Do NOT let Eyelid slide onto the cornea during IOP measuring
IOP measurement is reliable only in sclera area!
The tonometer's display shows the number of a single IOP measurement
(fig.14), the digital result is recorded automatically in the tonometer's
memory.
Fig. 14

Fig. 15

Keeping the eyelid in the same position tilt the tonometer back and
forward so the rod in the working position and proceed with the
measurement series until you hear one long sound signal (or two long sound
signals). Press the OPERATION button and you'll get the digital IOP value
on the display (fig.15).

During carrying out the measuring series do not look aside the display showings!
You will see the IOP value only after finishing the measuring series and pressing the
OPERATION button.

Interpretation of the obtained IOP values
Display indication
Symbol "А" and IOP
value in nonflickering mode.

Result evaluation
The result is reliable

Notes
The IOP measuring of the eye
being studied is finished

Symbol "А" in a
The result should be considered as If necessary carry out the
flickering mode and approximate, but if IOP is equal or new series of IOP measuring
IOP value in a
less than 20 mm Hg it can be
strictly following the
All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)
flickering mode.

considered as reliable.

Symbol "А 00" in a
flickering mode.

The result is considered erroneous. Carry out the new series of
IOP measuring strictly
following the methodology.

methodology.

The tonometer is turned OFF by momentarily pressing of the OPERATION button, or it is
turned OFF automatically in 30 seconds.
Attention! If you keep the OPERATION button pressed for more than 3 seconds the tonometer
enters the auxiliary mode the auxiliary information being displayed. It is easy to be corrected;
the detailed information can be found in Part II of Operation Manual. (To get out of this mode:
Press and hold the ON button until the countdown starts 1-7, Release once you reach #7. )
Continue taking measurements on another eye.

Possible mistakes during IOP measuring
The most common mistakes

These are the most common mistakes during mastering the measuring methodology and lead to
significant underestimation of the tonometry results.
Incorrect!
The measurement is taken not on the sclera (but on
the limb or on the cornea).

Correct!
The edge of the upper eyelid coincides
with the limb

All Rights Reserved. Copyright 2013
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

The situation can arise due the following reasons:
•

•

You did not coincide the patient's upper eyelid with the edge of the limb. The eyelid slightly
moved to the limb even before the measurement that's why the IOP measurement was realized
not in the sclera area but on the limb or on the cornea.
You did not fix the eyelid with the finger so during the measurement under the pressure of the
rod the eyelid moved to the cornea and the measurement was carried out not in the sclera
area.

Incorrect!
During the IOP measurement the tonometer's tip is
placed beyond the eyelid cartilage that is not close
to the ciliary edge but indented for 1 mm or more.

Correct!
The front part of the tip is close to the front
edge of the upper eyelid without touching
the eyelashes.

In this case the result underestimation is connected with the influence of the eyelid viscosity
because this influence increases significantly while moving further from the ciliary edge and
disfigures the tonometry results.
Please pay attention once again to the correct position of the eyelid and the tonometer's tipthis is the most important condition of receiving the reliable results!

Other mistakes
- nonhorizontal position of the patient's head
- the neck's squeeze with the collar
- the prolonged throwing the head back while there's the pathology of the
All Rights Reserved. Copyright 2013

Result
underestimation
Result overestimation
EMERGENCY MEDICINE UPDATE:
TONOMETER DIATON® SELECTED FOR
EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER)

spine's cervical section

Result overestimation

- non-vertical position of the tonometer

Result
underestimation

- the edge of the eyelid is above the corneal limb
- the eyelid's extrophy while pulling it strongly

Result
underestimation
Result
underestimation

More detailed information on IOP measuring methodology and the meaning of the displayed
symbols you can find in the Operation Manual and in the Training Video at:
http://tonometerdiaton.com/index.php?do=home.training

SUMMARY
DIATON® Tonometer Pen ( BiCOM Inc, USA) transpalpebral (through the eyelid, non-corneal)
tonometry is an easy, quick and accurate way of measuring intraocular pressures (IOP) in
patients presenting with acute eye problems. By becoming proficient with this skill, important
information can be obtained which may aid in the appropriate triage and management of these
challenging eye patients.
More information about Diaton tonometer can be obtained here:
http://www.TonometerDiaton.com

All Rights Reserved. Copyright 2013

More Related Content

Viewers also liked

Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)
Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)
Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)Uswatun Hasanah
 
Τα τείχη της Κωνσταντινούπολης
Τα τείχη της ΚωνσταντινούποληςΤα τείχη της Κωνσταντινούπολης
Τα τείχη της Κωνσταντινούποληςtsakalilia
 
Τα οχυρωματικά έργα της Κωνσταντινούπολης.
Τα οχυρωματικά έργα της Κωνσταντινούπολης.Τα οχυρωματικά έργα της Κωνσταντινούπολης.
Τα οχυρωματικά έργα της Κωνσταντινούπολης.tsakalilia
 
The power of language
The power of languageThe power of language
The power of languageFallana77
 
AQA GCSE Geography Physical Case Studies
AQA GCSE Geography Physical Case StudiesAQA GCSE Geography Physical Case Studies
AQA GCSE Geography Physical Case StudiesJoe Bush
 

Viewers also liked (6)

Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)
Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)
Pedomanpenskoranuasxfisika 121210081848-phpapp01 (1)
 
Τα τείχη της Κωνσταντινούπολης
Τα τείχη της ΚωνσταντινούποληςΤα τείχη της Κωνσταντινούπολης
Τα τείχη της Κωνσταντινούπολης
 
Τα οχυρωματικά έργα της Κωνσταντινούπολης.
Τα οχυρωματικά έργα της Κωνσταντινούπολης.Τα οχυρωματικά έργα της Κωνσταντινούπολης.
Τα οχυρωματικά έργα της Κωνσταντινούπολης.
 
441 1
441 1441 1
441 1
 
The power of language
The power of languageThe power of language
The power of language
 
AQA GCSE Geography Physical Case Studies
AQA GCSE Geography Physical Case StudiesAQA GCSE Geography Physical Case Studies
AQA GCSE Geography Physical Case Studies
 

Similar to ER IOP Measurement with Diaton Tonometer

Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511Tenmars Việt Nam
 
Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460
Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460
Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460Tenmars Việt Nam
 
ZMPCRM071000.12.01 User's Manual
ZMPCRM071000.12.01 User's  ManualZMPCRM071000.12.01 User's  Manual
ZMPCRM071000.12.01 User's ManualPainezee Specialist
 
O-Two eSeries e700 Ventilator fFunction and Operation
O-Two eSeries e700 Ventilator fFunction and OperationO-Two eSeries e700 Ventilator fFunction and Operation
O-Two eSeries e700 Ventilator fFunction and OperationRian Lins
 
Liemke Merlin Clip On 35 Instruction Manual | Optics Trade
Liemke Merlin Clip On 35 Instruction Manual | Optics TradeLiemke Merlin Clip On 35 Instruction Manual | Optics Trade
Liemke Merlin Clip On 35 Instruction Manual | Optics TradeOptics-Trade
 
Manual de usuario ultrasonido focalizado
Manual de usuario ultrasonido focalizadoManual de usuario ultrasonido focalizado
Manual de usuario ultrasonido focalizadoJorgeCobo11
 
Hướng dẫn sử dụng máy đo nhiệt độ Testo 610
Hướng dẫn sử dụng máy đo nhiệt độ Testo 610Hướng dẫn sử dụng máy đo nhiệt độ Testo 610
Hướng dẫn sử dụng máy đo nhiệt độ Testo 610Tenmars Việt Nam
 
Tonometry by dr. mamta meena
Tonometry by dr. mamta meenaTonometry by dr. mamta meena
Tonometry by dr. mamta meenaAIMS Education
 
Goldman applanation tonometry
 Goldman applanation tonometry Goldman applanation tonometry
Goldman applanation tonometryHossein Mirzaie
 
InfiRay Thermal E3 Max V2 Instruction Manual - Optics Trade
InfiRay Thermal E3 Max V2 Instruction Manual - Optics TradeInfiRay Thermal E3 Max V2 Instruction Manual - Optics Trade
InfiRay Thermal E3 Max V2 Instruction Manual - Optics TradeOptics-Trade
 

Similar to ER IOP Measurement with Diaton Tonometer (20)

Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511
 
Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460
Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460
Hướng dẫn sử dụng máy đo tốc độ vòng quay testo 460
 
ZMPCRM071000.12.01 User's Manual
ZMPCRM071000.12.01 User's  ManualZMPCRM071000.12.01 User's  Manual
ZMPCRM071000.12.01 User's Manual
 
O-Two eSeries e700 Ventilator fFunction and Operation
O-Two eSeries e700 Ventilator fFunction and OperationO-Two eSeries e700 Ventilator fFunction and Operation
O-Two eSeries e700 Ventilator fFunction and Operation
 
Liemke Merlin Clip On 35 Instruction Manual | Optics Trade
Liemke Merlin Clip On 35 Instruction Manual | Optics TradeLiemke Merlin Clip On 35 Instruction Manual | Optics Trade
Liemke Merlin Clip On 35 Instruction Manual | Optics Trade
 
TONOMETRY
TONOMETRYTONOMETRY
TONOMETRY
 
Different types of Tonometry
Different types of TonometryDifferent types of Tonometry
Different types of Tonometry
 
Part_1a.ppt
Part_1a.pptPart_1a.ppt
Part_1a.ppt
 
Iop
IopIop
Iop
 
Manual de usuario ultrasonido focalizado
Manual de usuario ultrasonido focalizadoManual de usuario ultrasonido focalizado
Manual de usuario ultrasonido focalizado
 
Hướng dẫn sử dụng máy đo nhiệt độ Testo 610
Hướng dẫn sử dụng máy đo nhiệt độ Testo 610Hướng dẫn sử dụng máy đo nhiệt độ Testo 610
Hướng dẫn sử dụng máy đo nhiệt độ Testo 610
 
Custom hearing aid operations manual
Custom hearing aid operations manualCustom hearing aid operations manual
Custom hearing aid operations manual
 
Tonometry
 Tonometry Tonometry
Tonometry
 
Instruction manual sendo_advance_english
Instruction manual sendo_advance_englishInstruction manual sendo_advance_english
Instruction manual sendo_advance_english
 
Instruction manual sendo_primo_english
Instruction manual sendo_primo_englishInstruction manual sendo_primo_english
Instruction manual sendo_primo_english
 
Tonometry by dr. mamta meena
Tonometry by dr. mamta meenaTonometry by dr. mamta meena
Tonometry by dr. mamta meena
 
Goldman applanation tonometry
 Goldman applanation tonometry Goldman applanation tonometry
Goldman applanation tonometry
 
Bpm 210 um blood pressure monitor
Bpm 210 um blood pressure monitorBpm 210 um blood pressure monitor
Bpm 210 um blood pressure monitor
 
Erbe icc-80-50-icc-service-manual
Erbe icc-80-50-icc-service-manualErbe icc-80-50-icc-service-manual
Erbe icc-80-50-icc-service-manual
 
InfiRay Thermal E3 Max V2 Instruction Manual - Optics Trade
InfiRay Thermal E3 Max V2 Instruction Manual - Optics TradeInfiRay Thermal E3 Max V2 Instruction Manual - Optics Trade
InfiRay Thermal E3 Max V2 Instruction Manual - Optics Trade
 

Recently uploaded

Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...chandars293
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...indiancallgirl4rent
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 

Recently uploaded (20)

Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 

ER IOP Measurement with Diaton Tonometer

  • 1. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) INTRAOCULAR PRESSURE (IOP) MEASUREMENT WITH DIATON® PEN TONOMETER INTRODUCTION: DIATON® Tonometer Pen ( BiCOM Inc, USA) is Non-corneal, Non-contact, Transpalpebral (through eyelid) handheld tonometer. DIATON® PEN TONOMETER is a handheld, pen-shaped device used for measuring intraocular pressures (IOP). The DIATON Pen tonometer calculates pressure by measuring the response of a free-falling floater, as it rebounds against the tarsal plate of the eyelid and the sclera. The patient is positioned so that the tip of the device and lid are overlying sclera, in either sitting or supine position. Non-corneal and transpalpebral tonometry does not involve contact with the cornea and does not require topical anesthetic during routine use. Contact lenses do not need to be taken out and central corneal thickness (CCT) and other corneal properties do not influence the IOP measurement. In Emergency Department and Acute Care settings IOP measurement is indicated in cases of acute eye injury, acute eye pain, red eye, and acute loss of visual acuity. Normal intraocular pressure ranges between 10-20 mmHg. All Rights Reserved. Copyright 2013
  • 2. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) EQUIPMENT, SUPPLIES & MAINTENANCE: To measure intraocular pressure with a DIATON® Tonometer Pen, the following equipment is needed: a pen tonometer, alcohol swab – to wipe off the tip of the tonometer between patients. NO need for topical anesthetic (either properacaine or tetracaine) or tip covers since the measurement is taken over upper eyelid. DIATON® Tonometer Pens are battery operated devices. Tonometer will indicate “U” on the display if the battery is low. Batteries are button CR2032. All Rights Reserved. Copyright 2013
  • 3. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) The parts of the tonometer are: the stainless steel tip, the operation button, and an LCD display. The operation button is used for turning the device ON and OFF and to obtain AVERAGE aget measurements. The LCD screen provides information about the sequence of readings, possible errors and tells you the pressure reading. CALIBRATION: DIATON® Tonometer Pen comes with a carry case which includes Pressure Selector Testing Device. To check for calibration, simply take the measurement on the Testing Plate. Results should be within ±2mm of 20mm. LCD display will also indicate any other errors or if the device needs to be cleaned. Frequency of testing – once a day. MEASURING IOP: The patient should be relaxed and seated at the edge of the chair and reclining back, with head titled back as horizontal as possible or lying down flat prior to beginning. Tight shirt collars should be loosened. Patient should extend a thumb forward at appx 45º to center the eye. User should guide the patients to be in the correct position. Putting the rod into working position The working position of the rod is its fixed inside the device (the rod is not seen in the tip area). All Rights Reserved. Copyright 2013
  • 4. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) Fig. 1 If the rod is seen in the tip area, it's necessary to put it into working position. Keep the tonometer vertically with its tip downwards, as it is shown in fig.1. Fig. 2 Turn the tonometer with its tip upwards, as it is shown in fig. 2. All Rights Reserved. Copyright 2013
  • 5. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) The rod automatically falls inside the tonometer. Fig. 3 Turn the tonometer with its tip downwards. The rod is fixed inside the device, as it is shown in fig.3; it means that it is in the proper working position. Turning the tonometer ON Fig. 4 Press momentarily the OPERATION button once. "0000" symbol appears on the display as it is shown in fig. 4. All Rights Reserved. Copyright 2013
  • 6. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) If after turning ON the tonometer produces an interrupted sound signal it means that its body deviates from vertical position. Move the device so it reaches the vertical position of the tonometer and the sound signal will stop. IOP measuring must be realized with the tonometer's position being strictly vertical! To get the reliable results it is important to learn how to reach the vertical position of the tonometer before starting IOP measurement in patients. Practicing with the tonometer on the test plate Fig. 5 The testing device is built in the tonometer's case (Fig.5). Its main assignment is checking the tonometer's capacity for work before starting the operation and when necessary. Use the testing device to easily find the vertical position and practice Test measurements. Prepare the tonometer for work — the rod is in the working position, the tonometer is turned ON. All Rights Reserved. Copyright 2013
  • 7. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) Fig. 6 Place the tonometer's tip in the indents of the testing device as it is shown in fig.6. Reach the tonometer's vertical position when there's no interrupted sound signal. Fig. 7 Fluently lower the tonometer's body down until the rod falls. The display must show the test value from 18 to 22, as it is shown in fig.7. The checking is finished. At other test values the tonometer is considered inoperative. All Rights Reserved. Copyright 2013
  • 8. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) During checking the tonometer's capacity for work keep it in the vertical position, there should be no sound signal. If after the measurement the display shows L symbol it means that user took the measurement while the device was beeping. It is necessary to put the rod into the working position again and repeat the operation on the testing device. Measuring of IOP It is possible to measure IOP with the patient being in sitting or reclining position. Horizontal position of the patient's head is required. Disinfect the tip of the tonometer with alcohol swab according to the directions of the Operation Manual. Fig. 8a Fig. 8b If you are Right Handed - Stand at the LEFT hand side of the patient as it is shown in fig.8. Put the rod into working position, turn the tonometer ON. All Rights Reserved. Copyright 2013
  • 9. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) Fig. 9 Place and fix the patient's glance using the test object (extend patients Thumb forward) the patient's glance line being oriented approximately at the angle of 45° as it is shown in fig.9. Fig. 10 Stretch the upper eyelid with a finger of a free hand so that the edge of the upper eyelid coincides with the limb (fig.10). Keep the eyelid in this position. Do not press the eyeball. All Rights Reserved. Copyright 2013
  • 10. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) Fig. 11 Place the edge of your palm with the tonometer on the patient's forehead. With the other hand keep holding the eyelid in the necessary position (edge of the lid should be 1mm above the edge of the limbus) (fig.11). Place the tonometer's tip on the eyelid in such a way that the front part of the tip is as close to the front edge of the upper eyelid as possible without touching the eyelashes. (fig.12, 13). The influence zone of the tonometer's rod must be the part of sclera corresponding to сorona ciliaris in 12 o'clock meridian. Secure the tonometer's vertical position (there's no sound signal). Fig. 12 Gently lower the tonometer's body keeping its vertical position until the rod falls on the eyelid which is accompanied with the short sound signal. All Rights Reserved. Copyright 2013
  • 11. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) Fig. 13 Do NOT let Eyelid slide onto the cornea during IOP measuring IOP measurement is reliable only in sclera area! The tonometer's display shows the number of a single IOP measurement (fig.14), the digital result is recorded automatically in the tonometer's memory. Fig. 14 Fig. 15 Keeping the eyelid in the same position tilt the tonometer back and forward so the rod in the working position and proceed with the measurement series until you hear one long sound signal (or two long sound signals). Press the OPERATION button and you'll get the digital IOP value on the display (fig.15). During carrying out the measuring series do not look aside the display showings! You will see the IOP value only after finishing the measuring series and pressing the OPERATION button. Interpretation of the obtained IOP values Display indication Symbol "А" and IOP value in nonflickering mode. Result evaluation The result is reliable Notes The IOP measuring of the eye being studied is finished Symbol "А" in a The result should be considered as If necessary carry out the flickering mode and approximate, but if IOP is equal or new series of IOP measuring IOP value in a less than 20 mm Hg it can be strictly following the All Rights Reserved. Copyright 2013
  • 12. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) flickering mode. considered as reliable. Symbol "А 00" in a flickering mode. The result is considered erroneous. Carry out the new series of IOP measuring strictly following the methodology. methodology. The tonometer is turned OFF by momentarily pressing of the OPERATION button, or it is turned OFF automatically in 30 seconds. Attention! If you keep the OPERATION button pressed for more than 3 seconds the tonometer enters the auxiliary mode the auxiliary information being displayed. It is easy to be corrected; the detailed information can be found in Part II of Operation Manual. (To get out of this mode: Press and hold the ON button until the countdown starts 1-7, Release once you reach #7. ) Continue taking measurements on another eye. Possible mistakes during IOP measuring The most common mistakes These are the most common mistakes during mastering the measuring methodology and lead to significant underestimation of the tonometry results. Incorrect! The measurement is taken not on the sclera (but on the limb or on the cornea). Correct! The edge of the upper eyelid coincides with the limb All Rights Reserved. Copyright 2013
  • 13. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) The situation can arise due the following reasons: • • You did not coincide the patient's upper eyelid with the edge of the limb. The eyelid slightly moved to the limb even before the measurement that's why the IOP measurement was realized not in the sclera area but on the limb or on the cornea. You did not fix the eyelid with the finger so during the measurement under the pressure of the rod the eyelid moved to the cornea and the measurement was carried out not in the sclera area. Incorrect! During the IOP measurement the tonometer's tip is placed beyond the eyelid cartilage that is not close to the ciliary edge but indented for 1 mm or more. Correct! The front part of the tip is close to the front edge of the upper eyelid without touching the eyelashes. In this case the result underestimation is connected with the influence of the eyelid viscosity because this influence increases significantly while moving further from the ciliary edge and disfigures the tonometry results. Please pay attention once again to the correct position of the eyelid and the tonometer's tipthis is the most important condition of receiving the reliable results! Other mistakes - nonhorizontal position of the patient's head - the neck's squeeze with the collar - the prolonged throwing the head back while there's the pathology of the All Rights Reserved. Copyright 2013 Result underestimation Result overestimation
  • 14. EMERGENCY MEDICINE UPDATE: TONOMETER DIATON® SELECTED FOR EMERGENCY DEPARTMENT (ED) / EMERGENCY ROOM (ER) spine's cervical section Result overestimation - non-vertical position of the tonometer Result underestimation - the edge of the eyelid is above the corneal limb - the eyelid's extrophy while pulling it strongly Result underestimation Result underestimation More detailed information on IOP measuring methodology and the meaning of the displayed symbols you can find in the Operation Manual and in the Training Video at: http://tonometerdiaton.com/index.php?do=home.training SUMMARY DIATON® Tonometer Pen ( BiCOM Inc, USA) transpalpebral (through the eyelid, non-corneal) tonometry is an easy, quick and accurate way of measuring intraocular pressures (IOP) in patients presenting with acute eye problems. By becoming proficient with this skill, important information can be obtained which may aid in the appropriate triage and management of these challenging eye patients. More information about Diaton tonometer can be obtained here: http://www.TonometerDiaton.com All Rights Reserved. Copyright 2013