• WHAT IS CVP?
• The central venous pressure is the pressure measured in the central veins
close to the heart.
• It indicates mean atrial pressure and is frequently used as an estimate of
right ventricular preload.
• CVP reflect the amount of blood returning to the heart and the ability of
the heart to pump blood into arterial system
• It is the pressure measured at junction of right atrium and SVC.
• It reflect the driving force for filling of the right atrium and ventricle.
• Normal CVP in awake, spontaneously breathing patient : 1-7 mmHg(or 5-
10 cmH2O)
• Mechanical ventilation: 3-5 cmH2O higher
• FACTOR AFFECTING CENTRAL VENOUS PRESSURE:
FACTOR INCREASING CVP FACTOR DECREASING CVP
• Raised intrathoracic pressure
• Circulatory overload
• Venoconstriction
• Impaired cardiac function
• SVC obstruction
• IPPV
• Reduced intrathoracic pressure
• Hypovolemia
• Venodilation
• CVP MONITORING
• In CVP monitoring, we insert a catheter through a vein and advances it until
its tip lies in or near the right atrium
• Because no major valves lies at the junction of the vena cava and right
atrium, pressure at end diastole reflects back to the catheter.
• When connected to manometer, the catheter measures CVP an index of right
ventricular function.
• CVP monitoring helps to assess cardiac function, to evaluate venous return
to the heart, and to indirectly gauge how well the heart is pumping.
• METHODS TO MEASURE CVP
1. Indirect method:
 Inspection of jugular venous pulsation in the neck.
2. Direct method
 fluid filled manometer connected to central venous catheter.
 Calibrated transducer.
1. Inspection of jugular venous pulsation in the neck:
• No valve between right atrium and internal jugular vein
• Degree of distension and venous wave form reflects information about
cardiac function.
2. Fluid filled manometer connected to central venous catheter
• CVP is measured using a column of water in a marked manometer.
• CVP is height of the column in cm of H2O when column is at the level of
right atrium
• ADVANTADE: simplicity to measure
• DISADVANTAGE: 1) inability to analyze the CVP waveform
2) relative slow response of water column to change in
intrathoracic pressure
• METHOD:
• With the CV line in place, position the patient flat.
• Align the base of the manometer with the determined zero reference point
by using leveling device.
• Because CVP reflect right atrium pressure, you must align the right atrium
with the zero mark on manometer
• To find right atrium locate the fourth intercostal space at the midaxillary
line.
• Mark the appropriate place on patients chest so that all subsequent
recordings will be made using same location.
• Attach water manometer to pint stand or place it to the next to patient’s
chest.
• Make sure the zero reference point is level with right atrium.
• A 3 way tap is used to connect the manometer to in IV drip set on one
side and via extension tubing filled with IV fluid to patient on other
side
• Check that the CVP catheter tubing is not kinked or blocked, and IV
fluid can easily flushed in and blood can easily aspirated from CVP
line.
• The 3 way tap is then turned so that it is open to the fluid bag and
manometer but closed to the patient, allowing manometer column to
fill with fluid.
• Once manometer has filled adequately the 3 way tap is turned again-
this time so it is open to patient and manometer, but close to the
fluid bag.
• The fluid level within the manometer column will fall to the level of
the CVP, the value of which can be read on manometer scale which is
marked in CM of water therefor giving value of CVP in cmH2O.
• The fluid level will continue to rise and fall slightly with respiration
and average reading should be recorded.
3. Calibrated transducer
• Automated, electronic pressure monitor.
• Pressure wave form displayed on an oscilloscope or paper.
• ADVANTAGE: 1)more accurate
2)direct observation of waveform
• METHOD:
• The transducer is fixed at the level of right atrium and connected to patient’s
CVP catheter via fluid filled extension tubing.
• The transducer than ‘zeroed, to atmospheric pressure by turning its 3 way
tap so that it is open to the transducer and to room air but closed to patient.
• The 3 way tap is then turned so that it is now closed to room air and open
between patient and transducer.
• A continue CVP reading measured in mmHg
CVP WAVEFORM
CVP waveform
Waveform component Phase of cardiac cycle Mechanical event
a wave End diastole Atrial contraction
c wave Early systole Isovolumetric
ventricular contraction,
tricuspid bulge in right
atrium
v wave Late systole Systolic filling of atrium
x decent Mid systole Atrial relaxation
Y decent Early diastole Early ventricular filling
• COMLICATION
• Haemorrhage
• Pneumothorax
• Air embolism
• Arteriovenous fistula
• Adjacent organ puncture
• Thrombosis
• Thrombo embolism
• Skin infection and necrosis
• sepsis
INTRAARTERIAL BLOOD PRESSURE MONITORING
• IBP measurement is often considered to be the gold standard of blood
pressure measurement.
• Despite its increased risk, cost, and need for technical expertise for
placement and management, its utility in providing crucial and timely
information outweighs its risks in many cases.
• INDICATION
• Continuous, real time BP monitoring
• Planned pharmacological or mechanical cardiovascular manipulation
• Repeated blood sampling
• Failure of indirect arterial bp monitoring
• Supplementary diagnostic information from the arterial waveform
• BASIC PRINCIPLE
• The pressure waveform of the arterial pulse is transmitted via the
column of fluid, to a pressure transducer where it is converted into
electrical signal.
• The electrical signal is then processed, amplified and converted into a
visual display by a microprocessor.
ARTERIAL PRESSURE MONITORING SITE
1. radial artery – most commonly used
2. Ulnar artery
3. Brachial artery
4. Axillary artery
5. Femoral artery
6. Dorsalis pedis artery
• COMPLICATION
1. Hemorrhage
2. Distal ischemia
3. Pseudoaneurysm
4. Arteriovenous fistula
5. Infection
6. Peripheral neuropathy
• COMPONENT OF AN IBP SYSTEM:
1. Intra arterial cannula
2. Fluid filled tubing
3. Transducer
4. Infusion/flushing system
5. Signal processor, amplifier and display
• PHYSICAL PRINCIPLES
• A wave is disturbance that travel through a medium, transferring
energy but not matter.
• One of the simplest waveforms is the sine wave.
• Fourier analysis:
• The arterial waveform is clearly not simple sine wave but it can be
broken down into a series of many component sine waves.
• The process of analyzing a complex waveform in terms of its
constituent sine waves is called Fourier analysis.
• Property:
• Natural frequency
• Damping coefficient
• NATURAL FREQUENCY
• The natural frequency of a system determines how rapid the system
oscillates after a stimulus.
• It is important that IBP system has a very high natural frequency at least
eight timed the fundamental frequency of arterial waveform(pulse rate).
• Therefore, for a system to remain accurate at heart rate of up to 180bpm, its
natural frequency must be at least: (180bpm *8)/60sec=24Hz
• The natural frequency of a system may be increased by
1. Reducing the length and compliance of tubing
2. Reducing the density of the fluid used in tubing
3. Increase the diameter of tubing
4. Commercially available system -200Hz
• DAMPING:
• Anything that reduces energy in an oscillating system will reduce the
amplitude of oscillation. This is termed damping.
• Some degree of damping required in for all system(critical damping), But if
excessive(overdamping) or insufficient(underdamping) the output will be
adversely affected.
• The damping coefficient reflects forces acting on the system and
determines how rapidly it returns to rest after a stimulation.
• FAST FLUSH TEST
• Provide a convenient bedside method for determining dynamic response of
system.
• Natural frequency is inversely proportional to the time between adjacent
oscillation peaks
• The damping coefficient can be calculated mathematically, but it is usually
determined graphically from amplitude ratio
CVP.pptx

More Related Content

Similar to CVP.pptx

Swan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdfSwan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdframbhoopal1
 
MONITORING HEMODYNAMIC STATUS-1.pptx
MONITORING HEMODYNAMIC STATUS-1.pptxMONITORING HEMODYNAMIC STATUS-1.pptx
MONITORING HEMODYNAMIC STATUS-1.pptxRizwanUllah151255
 
Hemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patientsHemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patientsV4Veeru25
 
Invasive Cardio-vascular Monitoring.pptx
Invasive Cardio-vascular Monitoring.pptxInvasive Cardio-vascular Monitoring.pptx
Invasive Cardio-vascular Monitoring.pptxMalik Mohammad
 
Basic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxBasic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxamitkalirawana07
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...prateek gupta
 
cardiacoutputmonitoring-190708221224.pdf
cardiacoutputmonitoring-190708221224.pdfcardiacoutputmonitoring-190708221224.pdf
cardiacoutputmonitoring-190708221224.pdfaishabajwa8081
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoringmauryaramgopal
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoringPratik Tantia
 
cardiac output pptx
cardiac output pptxcardiac output pptx
cardiac output pptxananya nanda
 
Advances in haemodynamic monitoring
Advances in haemodynamic monitoringAdvances in haemodynamic monitoring
Advances in haemodynamic monitoringMohamed Abdulrazik
 
Central venous pressure monitoring - Pooja Murkar
Central  venous  pressure  monitoring - Pooja MurkarCentral  venous  pressure  monitoring - Pooja Murkar
Central venous pressure monitoring - Pooja Murkarpooja murkar
 
Pulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusionPulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusionNahas N
 
Intra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.pptIntra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.ppttaimourali64
 
Hemodynamic_Monitoring_I_ABP_CVP_Ao.ppt
Hemodynamic_Monitoring_I_ABP_CVP_Ao.pptHemodynamic_Monitoring_I_ABP_CVP_Ao.ppt
Hemodynamic_Monitoring_I_ABP_CVP_Ao.pptsavitri49
 
HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...
HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...
HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...Bassel Ericsoussi, MD
 

Similar to CVP.pptx (20)

Swan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdfSwan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdf
 
MONITORING HEMODYNAMIC STATUS-1.pptx
MONITORING HEMODYNAMIC STATUS-1.pptxMONITORING HEMODYNAMIC STATUS-1.pptx
MONITORING HEMODYNAMIC STATUS-1.pptx
 
Hemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patientsHemodynamic monitoring of critically ill patients
Hemodynamic monitoring of critically ill patients
 
ANES 1502 - M9 PPT: Hemodynamic Monitoring
ANES 1502 - M9 PPT: Hemodynamic MonitoringANES 1502 - M9 PPT: Hemodynamic Monitoring
ANES 1502 - M9 PPT: Hemodynamic Monitoring
 
Invasive Cardio-vascular Monitoring.pptx
Invasive Cardio-vascular Monitoring.pptxInvasive Cardio-vascular Monitoring.pptx
Invasive Cardio-vascular Monitoring.pptx
 
Basic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxBasic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptx
 
Cvp
CvpCvp
Cvp
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
 
cardiacoutputmonitoring-190708221224.pdf
cardiacoutputmonitoring-190708221224.pdfcardiacoutputmonitoring-190708221224.pdf
cardiacoutputmonitoring-190708221224.pdf
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoring
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
cardiac output pptx
cardiac output pptxcardiac output pptx
cardiac output pptx
 
Advances in haemodynamic monitoring
Advances in haemodynamic monitoringAdvances in haemodynamic monitoring
Advances in haemodynamic monitoring
 
Echocardiography
EchocardiographyEchocardiography
Echocardiography
 
Central venous pressure monitoring - Pooja Murkar
Central  venous  pressure  monitoring - Pooja MurkarCentral  venous  pressure  monitoring - Pooja Murkar
Central venous pressure monitoring - Pooja Murkar
 
Pulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusionPulsatile vs non pulsatile perfusion
Pulsatile vs non pulsatile perfusion
 
Intra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.pptIntra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.ppt
 
Hemodynamic_Monitoring_I_ABP_CVP_Ao.ppt
Hemodynamic_Monitoring_I_ABP_CVP_Ao.pptHemodynamic_Monitoring_I_ABP_CVP_Ao.ppt
Hemodynamic_Monitoring_I_ABP_CVP_Ao.ppt
 
ICU management of ECMO pt
ICU management of ECMO ptICU management of ECMO pt
ICU management of ECMO pt
 
HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...
HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...
HEMODYNAMICS MONITORING IN CRITICALLY ILL PATIENTS: ASSESSMENT OF FLUID STATU...
 

More from SwatiChoudhary97

RESPIRATORY FUNCTION DURING ANAESTHESIA.pptx
RESPIRATORY FUNCTION DURING ANAESTHESIA.pptxRESPIRATORY FUNCTION DURING ANAESTHESIA.pptx
RESPIRATORY FUNCTION DURING ANAESTHESIA.pptxSwatiChoudhary97
 
Spinal and Epidural Anaesthesia.pptx
Spinal and Epidural Anaesthesia.pptxSpinal and Epidural Anaesthesia.pptx
Spinal and Epidural Anaesthesia.pptxSwatiChoudhary97
 
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptxECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptxSwatiChoudhary97
 
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptxECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptxSwatiChoudhary97
 
734_Visual_Pathway_lesions.pptx
734_Visual_Pathway_lesions.pptx734_Visual_Pathway_lesions.pptx
734_Visual_Pathway_lesions.pptxSwatiChoudhary97
 

More from SwatiChoudhary97 (10)

RESPIRATORY FUNCTION DURING ANAESTHESIA.pptx
RESPIRATORY FUNCTION DURING ANAESTHESIA.pptxRESPIRATORY FUNCTION DURING ANAESTHESIA.pptx
RESPIRATORY FUNCTION DURING ANAESTHESIA.pptx
 
Spinal and Epidural Anaesthesia.pptx
Spinal and Epidural Anaesthesia.pptxSpinal and Epidural Anaesthesia.pptx
Spinal and Epidural Anaesthesia.pptx
 
SMR ppt.pptx
SMR ppt.pptxSMR ppt.pptx
SMR ppt.pptx
 
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptxECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
 
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptxECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
ECG MONITORING AND ECG CHANGES OF INTRAOPERATIVE MYOCARDIAL.pptx
 
Gonioscopy
Gonioscopy Gonioscopy
Gonioscopy
 
734_Visual_Pathway_lesions.pptx
734_Visual_Pathway_lesions.pptx734_Visual_Pathway_lesions.pptx
734_Visual_Pathway_lesions.pptx
 
visual Pathway.pptx
visual Pathway.pptxvisual Pathway.pptx
visual Pathway.pptx
 
visual Pathway.pptx
visual Pathway.pptxvisual Pathway.pptx
visual Pathway.pptx
 
slbasics.pdf
slbasics.pdfslbasics.pdf
slbasics.pdf
 

Recently uploaded

Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiNehru place Escorts
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Servicenarwatsonia7
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 

Recently uploaded (20)

Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 

CVP.pptx

  • 1. • WHAT IS CVP? • The central venous pressure is the pressure measured in the central veins close to the heart. • It indicates mean atrial pressure and is frequently used as an estimate of right ventricular preload. • CVP reflect the amount of blood returning to the heart and the ability of the heart to pump blood into arterial system • It is the pressure measured at junction of right atrium and SVC. • It reflect the driving force for filling of the right atrium and ventricle. • Normal CVP in awake, spontaneously breathing patient : 1-7 mmHg(or 5- 10 cmH2O) • Mechanical ventilation: 3-5 cmH2O higher
  • 2. • FACTOR AFFECTING CENTRAL VENOUS PRESSURE: FACTOR INCREASING CVP FACTOR DECREASING CVP • Raised intrathoracic pressure • Circulatory overload • Venoconstriction • Impaired cardiac function • SVC obstruction • IPPV • Reduced intrathoracic pressure • Hypovolemia • Venodilation
  • 3. • CVP MONITORING • In CVP monitoring, we insert a catheter through a vein and advances it until its tip lies in or near the right atrium • Because no major valves lies at the junction of the vena cava and right atrium, pressure at end diastole reflects back to the catheter. • When connected to manometer, the catheter measures CVP an index of right ventricular function. • CVP monitoring helps to assess cardiac function, to evaluate venous return to the heart, and to indirectly gauge how well the heart is pumping.
  • 4. • METHODS TO MEASURE CVP 1. Indirect method:  Inspection of jugular venous pulsation in the neck. 2. Direct method  fluid filled manometer connected to central venous catheter.  Calibrated transducer.
  • 5. 1. Inspection of jugular venous pulsation in the neck: • No valve between right atrium and internal jugular vein • Degree of distension and venous wave form reflects information about cardiac function.
  • 6.
  • 7. 2. Fluid filled manometer connected to central venous catheter • CVP is measured using a column of water in a marked manometer. • CVP is height of the column in cm of H2O when column is at the level of right atrium • ADVANTADE: simplicity to measure • DISADVANTAGE: 1) inability to analyze the CVP waveform 2) relative slow response of water column to change in intrathoracic pressure
  • 8. • METHOD: • With the CV line in place, position the patient flat. • Align the base of the manometer with the determined zero reference point by using leveling device. • Because CVP reflect right atrium pressure, you must align the right atrium with the zero mark on manometer • To find right atrium locate the fourth intercostal space at the midaxillary line. • Mark the appropriate place on patients chest so that all subsequent recordings will be made using same location. • Attach water manometer to pint stand or place it to the next to patient’s chest. • Make sure the zero reference point is level with right atrium.
  • 9. • A 3 way tap is used to connect the manometer to in IV drip set on one side and via extension tubing filled with IV fluid to patient on other side • Check that the CVP catheter tubing is not kinked or blocked, and IV fluid can easily flushed in and blood can easily aspirated from CVP line. • The 3 way tap is then turned so that it is open to the fluid bag and manometer but closed to the patient, allowing manometer column to fill with fluid. • Once manometer has filled adequately the 3 way tap is turned again- this time so it is open to patient and manometer, but close to the fluid bag.
  • 10. • The fluid level within the manometer column will fall to the level of the CVP, the value of which can be read on manometer scale which is marked in CM of water therefor giving value of CVP in cmH2O. • The fluid level will continue to rise and fall slightly with respiration and average reading should be recorded.
  • 11.
  • 12. 3. Calibrated transducer • Automated, electronic pressure monitor. • Pressure wave form displayed on an oscilloscope or paper. • ADVANTAGE: 1)more accurate 2)direct observation of waveform
  • 13. • METHOD: • The transducer is fixed at the level of right atrium and connected to patient’s CVP catheter via fluid filled extension tubing. • The transducer than ‘zeroed, to atmospheric pressure by turning its 3 way tap so that it is open to the transducer and to room air but closed to patient. • The 3 way tap is then turned so that it is now closed to room air and open between patient and transducer. • A continue CVP reading measured in mmHg
  • 15. CVP waveform Waveform component Phase of cardiac cycle Mechanical event a wave End diastole Atrial contraction c wave Early systole Isovolumetric ventricular contraction, tricuspid bulge in right atrium v wave Late systole Systolic filling of atrium x decent Mid systole Atrial relaxation Y decent Early diastole Early ventricular filling
  • 16.
  • 17.
  • 18. • COMLICATION • Haemorrhage • Pneumothorax • Air embolism • Arteriovenous fistula • Adjacent organ puncture • Thrombosis • Thrombo embolism • Skin infection and necrosis • sepsis
  • 19. INTRAARTERIAL BLOOD PRESSURE MONITORING • IBP measurement is often considered to be the gold standard of blood pressure measurement. • Despite its increased risk, cost, and need for technical expertise for placement and management, its utility in providing crucial and timely information outweighs its risks in many cases.
  • 20. • INDICATION • Continuous, real time BP monitoring • Planned pharmacological or mechanical cardiovascular manipulation • Repeated blood sampling • Failure of indirect arterial bp monitoring • Supplementary diagnostic information from the arterial waveform
  • 21. • BASIC PRINCIPLE • The pressure waveform of the arterial pulse is transmitted via the column of fluid, to a pressure transducer where it is converted into electrical signal. • The electrical signal is then processed, amplified and converted into a visual display by a microprocessor.
  • 22. ARTERIAL PRESSURE MONITORING SITE 1. radial artery – most commonly used 2. Ulnar artery 3. Brachial artery 4. Axillary artery 5. Femoral artery 6. Dorsalis pedis artery
  • 23. • COMPLICATION 1. Hemorrhage 2. Distal ischemia 3. Pseudoaneurysm 4. Arteriovenous fistula 5. Infection 6. Peripheral neuropathy
  • 24. • COMPONENT OF AN IBP SYSTEM: 1. Intra arterial cannula 2. Fluid filled tubing 3. Transducer 4. Infusion/flushing system 5. Signal processor, amplifier and display
  • 25.
  • 26. • PHYSICAL PRINCIPLES • A wave is disturbance that travel through a medium, transferring energy but not matter. • One of the simplest waveforms is the sine wave. • Fourier analysis: • The arterial waveform is clearly not simple sine wave but it can be broken down into a series of many component sine waves. • The process of analyzing a complex waveform in terms of its constituent sine waves is called Fourier analysis. • Property: • Natural frequency • Damping coefficient
  • 27. • NATURAL FREQUENCY • The natural frequency of a system determines how rapid the system oscillates after a stimulus. • It is important that IBP system has a very high natural frequency at least eight timed the fundamental frequency of arterial waveform(pulse rate). • Therefore, for a system to remain accurate at heart rate of up to 180bpm, its natural frequency must be at least: (180bpm *8)/60sec=24Hz
  • 28. • The natural frequency of a system may be increased by 1. Reducing the length and compliance of tubing 2. Reducing the density of the fluid used in tubing 3. Increase the diameter of tubing 4. Commercially available system -200Hz
  • 29. • DAMPING: • Anything that reduces energy in an oscillating system will reduce the amplitude of oscillation. This is termed damping. • Some degree of damping required in for all system(critical damping), But if excessive(overdamping) or insufficient(underdamping) the output will be adversely affected. • The damping coefficient reflects forces acting on the system and determines how rapidly it returns to rest after a stimulation.
  • 30.
  • 31. • FAST FLUSH TEST • Provide a convenient bedside method for determining dynamic response of system. • Natural frequency is inversely proportional to the time between adjacent oscillation peaks • The damping coefficient can be calculated mathematically, but it is usually determined graphically from amplitude ratio