SlideShare a Scribd company logo
1 of 57
Download to read offline
Optimizing The
Sterilization Process
Tutor: Eamonn Hoxey
How can I ask questions?
Questions & Answers
Optimizing The Sterilization Process
Sterilization
• Sterilization processes need to be properly controlled by the
management of the responsible organization
• Procedures for validation and routine control of the sterilization
process and the sterile barrier system should be within the
organization's QMS
• Compliance with ISO 13485 provides a basis for exercising that
control
• EN 556 - Requirements for designating sterile medical devices
• EN ISO 11135 - Ethylene oxide
• EN ISO 11137 - Radiation
• EN ISO 14937 - General criteria
• EN ISO 17665 - Moist Heat
• EN ISO 20857 - Dry Heat
• EN ISO 14160 - Liquid chemical sterilizing agents
• ISO 22441 - Low temperature vaporized hydrogen peroxide
• EN ISO 25424 - Low Temperature Steam Formaldehyde
Standards focus
(102 - 10-6 = 8D Reduction)
Sterility Assurance
2
0
-2
-4
-6
Probability of there
being a survivor on an
individual unit
Number of survivors
on each unit
Time
10
10
10
10
10
Countable
number of
organisms
surviving on
each unit
Sterilizing treatment
106
True
bioburden
remaining
on
each
unit
Probability
of
a
surviving
organism
remaining
on
any
individual
unit
102
10-2
10-4
10-0
10-6
104
Bioburden estimate
EN ISO 11737-1
Test of
Sterility
EN ISO
11737-2
1 or more
organisms
surviving on some
units in sample of
practical size (up
to 100 units)
Impractical to
sample for
surviving
organisms
Impossible to sample
for surviving organisms
Sterility Assurance
Nordic Pharmacopeia Addendum 1970
“Sterile drugs must be
prepared and sterilised
under conditions that aim at
such a result that in one
million units there will be no
more than one living
microorganism”.
“The preparation procedure
shall be controlled regularly
in order to ensure a low
content of microorganisms
prior to sterilization.”
4.1 For a terminally-sterilized medical device to be
designated "STERILE", the theoretical probability of there
being a viable micro-organism present on/in the device
shall be equal to or less than 1x10-6.
EN 556-1: Requirements for terminally
sterilized medical devices.
EN 556-1: Requirements
4.2: Compliance shall be shown by the manufacturer or
supplier through provision of documentation and records
which demonstrate that the devices have been subjected to a
validated sterilization process fulfilling 4.1
EN 556-1: Requirements contd
NOTE 1: Evidence that a medical device is sterile comes from: i) the
initial validation of the sterilization process and subsequent
revalidations that demonstrate the acceptability of the process,
and ii) information gathered during the routine control and
monitoring which demonstrates that the validated process has
been delivered in practice.
NOTE 2: The achievement of sterility is predicted from the
bioburden level on products, the resistance of the
microorganisms comprising the bioburden and the extent of
treatment imposed during sterilization
Format of sterilization standards
• Common format
• based on ISO 14937:2000
• Sterilization of health care products - General requirements for characterization of a
sterilizing agent and the development, validation and routine control of a sterilization
process for medical devices.
• ISO 14937 revision published 2009
• Consistent definitions
• based on ISO TS 11139 - Terminology
• Common Quality Management System Elements
• based on reference to ISO 13485
 Introduction
 Scope
 Normative references
 Definitions
 Quality management system elements
 Sterilizing agent characterization
 Process / equipment characterization
 Product definition
 Process definition
 Validation
 Routine control and monitoring
 Product release from sterilization
 Maintaining process effectiveness
High level structure of sterilization standards
Process definition
The purpose of this activity is to obtain a detailed specification
for the sterilization process to be applied to defined product
without compromising the safety, quality and performance of
that product.
Process Definition
- development activities
- mechanical and biological safety
- process residuals
- product compatibility
- limits on resterilization
Process Definition
General approaches to process definition
1. Develop a specific process for a particular product
2. Demonstrate effectiveness of process already in use
For each of these approaches, the microbicidal performance can be
determined using:
– Natural bioburden
– Reference microorganisms
– Conservative approach
• 12 D
• Half cycle
• Reference process
Process optimization
• Opportunity to tailor the sterilization process to
– Reduce undesirable effects of the sterilization process
• Physical or chemical deterioration
• Residuals
– Reduce processing time
– Reduce costs
Sterilization - Process Definition
Sterilization processes therefore must achieve a maximal SAL of 10-6
Seen that achievement of a maximal SAL of 10-6 by a process is dependant
on the numbers and type of organisms present
Approaches to demonstrating a maximal SAL of 10-6
• Conservative or overkill approach
• Reference process
• Use of a reference organism and a knowledge of the bioburden
• Inactivation of the bioburden in its natural state
Conservative approach
Based on inactivation kinetics of a
highly resistant spore present in
high numbers
Sterilizing treatment
106
102
10-2
10-4
10-0
10-6
104
Product bioburden
Product bioburden
inactivation kinetics
Sterilization - Process Definition
Combined bioburden and biological
indicator method.
Based on inactivation of an
organism of know resistance
greater than the natural bioburden
and the numbers of organisms on
the product
SAL 10-6
Biological indicators
Control of Sterilization
Control of sterilization processes requires integration of process
variables.
A biological indicator integrates process variables
Variables
Process
Dose received
Irradiation
Time, Temperature
Dry heat
Time, Temperature, Moisture
Moist heat
Time, Temperature, EO concentration, Humidity
Ethylene Oxide
• Physical/chemical monitoring – mandatory
• Chemical indicators – not mandatory
• Biological indicators – not mandatory
Control of Sterilization
Biological Indicators (BIs)
Definition
A test system containing viable
microorganisms providing a specified
resistance to a specified sterilization
process
ISO 11139 3.29
• Selected microorganisms with known, usually high,
resistance to specific sterilization processes
• Prepared on a carrier, e.g. paper, sand, cotton yarn or metal
coupon = inoculated carrier
• Inoculated carrier in primary packaging = biological indicator
(BI)
• Ready for use and with a defined resistance to specified
sterilization process
Description of a Biological Indicator
Examples of process development
approaches
• Conservative approach – Ethylene oxide
• Reference process – Moist heat
• Reference organism and knowledge of bioburden –
Ethylene oxide
• Natural bioburden – Ionizing radiation
Ethylene oxide – Process definition
Annex A - D value determination
• Survivor curve construction
• Fraction negative (Most Probable Numbers)
• All BIs at each time increment must be exposed to same conditions
Annex B - Conservative process definition - (Half-cycle)
• All BIs at each time increment must be exposed to same conditions
• Inactivation of 12 log cycles
‒ Demonstration of 6 ‘spore log reduction’ from 106 to 100 –then double
exposure time
‒ Demonstration of 8 ‘spore log reduction’ from 106 to 10-2 -then add 50%
to exposure time
• Complete inactivation of 106 biological indicator – then double exposure time
Ethylene Oxide Process Definition -
Microbiological PQ
Half cycle approach
106
100
10-2
10-6
x x
Demonstration of 12 log cycles of reduction
106
100
10-2
10-6
x x
Demonstration of 12 log cycles of reduction
106
100
10-2
10-6
x x/2
Conservative approach
Based on inactivation kinetics of a
highly resistant spore present in high
numbers
Sterilizing treatment
106
102
10-2
10-4
10-0
10-6
104
Product bioburden
Product bioburden
inactivation kinetics
Ethylene oxide Sterilization – Combined BI and
bioburden process definition
Combined bioburden and biological
indicator method.
Based on inactivation of an organism
of know resistance greater than the
natural bioburden and the numbers of
organisms on the product
SAL 10-6
Moist heat sterilization – Reference processes
Moist heat reference processes
Perkins (1956)
Temperature Time
1320C 2 min
1250C 8 min
1210C 12 min
1180C 18 min
1160C 30 min
MRC (1959)
Temperature Time
1340C 3 min
1260C 10 min
1210C 15 min
Added additional (safety) period to allow to for
Deviations in steam quality
Moist heat - Sterility assurance level attained by different “accepted”
cycles
34
10-6
106
100
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
G stearothermophilus
spores D121 1.5min
Natural bioburden D121
often a few seconds
SAL v’s BI with a
count of 106
SAL at F0 = 8
SAL at 1210
for 15 mins
10-12
Cycle time in
mins at 1210
Radiation Sterilization - Dose setting methods
Dose Setting Methods – Principles ..cont..
 The underlying concept for establishing a ‘product specific
dose’ is the use of naturally contaminated product items and a
measurement of the radiation resistance of micro-organisms
as they occur on the product.
Dose Setting Methods – Product specific
 ISO 11137
– Method 1
– Method 2
 Use of Method 1 requires substantially less resource than that
required for Method 2. Most manufacturers prefer to use
Method 1 initially
Dose Setting Methods
 Method 1:
– Dose setting using bioburden information
 Method 2:
– Dose setting using fraction positive information from
incremental dosing to determine extrapolation factor –
measure of radiation resistance
Dose Setting Methods - Principles
• The extent of treatment required for a sterilization process (T) is a
function of:
– the number of micro-organisms present prior to processing (N)
– the distribution of resistances of these micro-organisms (R)
– the required Sterility Assurance Level (SAL)
• Or
T= f(N, R, SAL)
Radiation Sterilization Dose Setting
Method 1
Dose Setting Method 1
The choice of the sterilizing dose is based on experimental
verification that the resistance to radiation of the bioburden on
the device is less than that of a microbial population having a
standard distribution of resistances
Dose Setting Method 1 – ISO 11137-2
Dose Setting Method 1 – ISO 11137-2
 Using the Standard Distribution of Resistances (SDR), solutions
to the equation T= f(N, R, SAL) have been generated for a
range of levels of bioburden and SAL values.
 These are tabulated in ISO 11137-2
Radiation Sterilization Dose Setting -
Method 2
Dose Setting Method 2
 Requires considerably more experimentation than Method 1
 Provides a measurement of the radiation resistance of micro-
organisms as they occur on product units.
Dose Setting Method 2
 Two experiments performed to establish estimates of:
− the radiation dose giving an SAL of 10-2;
– the D10 value of micro-organisms surviving this dose
– Based on these data, the dose for an SAL of 10-2 can be
extrapolated to that required for an SAL of 10-6
Dose Setting Method 2
• Two experiments
– Incremental Dose Experiment;
– 20 items exposed to each of 9 incremental doses. Items are
tested individually for sterility. Data used to estimate 10-2
SAL dose and resistance of organisms surviving that dose.
– Verification Dose Experiment;
– 100 items irradiated at 10-2 SAL dose to confirm dose at
which 1 in 100 items non-sterile.
Log viable
microorganisms
Probability of
one surviving
microorganism
or Sterility
Assurance
Level (SAL)
6
5
4
3
2
1
0
-1
-2
-3
-4
-5
-6
10-1
10-2
10-3
10-4
10-5
10-6
Dose kGy
Incremental Dose experiment
Log viable
microorganisms
Probability of
one surviving
microorganism
or Sterility
Assurance
Level (SAL)
6
5
4
3
2
1
0
-1
-2
-3
-4
-5
-6
10-1
10-2
10-3
10-4
10-5
10-6
Dose kGy
Sterilization dose
Verification Dose experiment
Dose Setting Method 2
• Calculation
Sterilizing Dose is calculated by extrapolation to the
required SAL from the verified 10-2 SAL dose using the
estimate of the resistance of the micro-organisms
surviving that dose.
Radiation Sterilization - Dose Substantiation
Dose substantiation - Principles
T= f(N, R, SAL)
• In the case of substantiation of 25kGy:
– T = 25kGy
– N is determined in bioburden studies
– SAL = 10-6
• Solve the equation for the maximum resistance that can be tolerated to achieve SAL =
10-6 with 25 kGy
25kGy= f(Bioburden, R, 10-6)
or
R=f(Bioburden, 25kGY, 10-6)
• Verify that this resistance is not exceeded
ISO TS 13004: 2013 - ISO 13004:2022
• Sterilization of health care products – Radiation –
Substantiation of a selected sterilization dose: Method VDmax
SD
• Covers substantiation of sterilization doses from 17.5 to 35
kGy at 2.5 kGy increments
• Adopted as a European Technical Specification
• Not a harmonized European standard
• Intended to gain experience before considering for
inclusion in a future revision of ISO 11137-2
Process Optimization - Conclusion
• Different approaches to process definition provide different degrees
of process optimization
– Radiation Method 2
– Radiation Method 1
– Radiation Dose Substantiation (Vdmax
SD)
– Ethylene oxide combined BI/bioburden approach
– Moist heat reference process
– Ethylene oxide 12D approach
– Ethylene oxide half-cycle
Increasing
optimization
Ethylene Oxide Sterilization
To Learn More About ETO Sterilisation Please Join Us
When : 17TH – 20TH April 2023
Monday – Thursday Half days 1pm – 5pm (GMT)
Where: Virtual Instructor led
Promo Code: EARLYBIRD10
Trainer: Eamonn Hoxey
Optimizing  the Sterilization Process in compliance with standards

More Related Content

Similar to Optimizing the Sterilization Process in compliance with standards

Sterilization; infection control, sandipta banerjee
Sterilization; infection control, sandipta banerjee Sterilization; infection control, sandipta banerjee
Sterilization; infection control, sandipta banerjee Sandipta Banerjee
 
Cleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfCleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfLdiaSabino
 
STERILIZATION.pptx
STERILIZATION.pptxSTERILIZATION.pptx
STERILIZATION.pptxsyedumair76
 
industrial sterilization
industrial sterilizationindustrial sterilization
industrial sterilizationNilesh Utpure
 
Sterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMPSterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMPIqra Shafeeq
 
Evaluation of various /certified fixed orthodontic courses by Indian dental a...
Evaluation of various /certified fixed orthodontic courses by Indian dental a...Evaluation of various /certified fixed orthodontic courses by Indian dental a...
Evaluation of various /certified fixed orthodontic courses by Indian dental a...Indian dental academy
 
Validation of sterilization
Validation of sterilizationValidation of sterilization
Validation of sterilizationKashf Rasheed
 
sterilization.1.pptx
sterilization.1.pptxsterilization.1.pptx
sterilization.1.pptxHafeezahmad41
 
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANINGCONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANINGShankar Maind Patil
 
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationEU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationMerck Life Sciences
 
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationEU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationMilliporeSigma
 
Equipment validation
Equipment validationEquipment validation
Equipment validationGaurav Kr
 
Eto sterilization slides
Eto sterilization slidesEto sterilization slides
Eto sterilization slidesIrfan Ali
 
Cleaning validation presentation(1)
Cleaning validation presentation(1)Cleaning validation presentation(1)
Cleaning validation presentation(1)Randheer Dubey
 

Similar to Optimizing the Sterilization Process in compliance with standards (20)

Autoclave validation maliba
Autoclave validation malibaAutoclave validation maliba
Autoclave validation maliba
 
sterilization (2).pptx
sterilization (2).pptxsterilization (2).pptx
sterilization (2).pptx
 
Industrial sterilization
Industrial sterilizationIndustrial sterilization
Industrial sterilization
 
Sterilization; infection control, sandipta banerjee
Sterilization; infection control, sandipta banerjee Sterilization; infection control, sandipta banerjee
Sterilization; infection control, sandipta banerjee
 
Cleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfCleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdf
 
STERILIZATION.pptx
STERILIZATION.pptxSTERILIZATION.pptx
STERILIZATION.pptx
 
industrial sterilization
industrial sterilizationindustrial sterilization
industrial sterilization
 
Sterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMPSterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMP
 
Evaluation of various /certified fixed orthodontic courses by Indian dental a...
Evaluation of various /certified fixed orthodontic courses by Indian dental a...Evaluation of various /certified fixed orthodontic courses by Indian dental a...
Evaluation of various /certified fixed orthodontic courses by Indian dental a...
 
Validation of sterilization
Validation of sterilizationValidation of sterilization
Validation of sterilization
 
sterilization.1.pptx
sterilization.1.pptxsterilization.1.pptx
sterilization.1.pptx
 
Control of microbial growth
Control of microbial growthControl of microbial growth
Control of microbial growth
 
4_2018_04_07!06_25_24_AM (1).pptx
4_2018_04_07!06_25_24_AM (1).pptx4_2018_04_07!06_25_24_AM (1).pptx
4_2018_04_07!06_25_24_AM (1).pptx
 
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANINGCONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
 
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationEU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
 
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationEU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
 
Equipment validation
Equipment validationEquipment validation
Equipment validation
 
Eto sterilization slides
Eto sterilization slidesEto sterilization slides
Eto sterilization slides
 
Cleaning Validation "Part 1"
Cleaning Validation "Part 1"Cleaning Validation "Part 1"
Cleaning Validation "Part 1"
 
Cleaning validation presentation(1)
Cleaning validation presentation(1)Cleaning validation presentation(1)
Cleaning validation presentation(1)
 

Recently uploaded

High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxDeepakSakkari2
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Introduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptxIntroduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptxvipinkmenon1
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝soniya singh
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidNikhilNagaraju
 
Heart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxHeart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxPoojaBan
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AIabhishek36461
 
power system scada applications and uses
power system scada applications and usespower system scada applications and uses
power system scada applications and usesDevarapalliHaritha
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZTE
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSCAESB
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...VICTOR MAESTRE RAMIREZ
 
IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024Mark Billinghurst
 

Recently uploaded (20)

High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptx
 
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
Introduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptxIntroduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptx
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfid
 
Heart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxHeart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptx
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AI
 
power system scada applications and uses
power system scada applications and usespower system scada applications and uses
power system scada applications and uses
 
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentation
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...
 
IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024
 

Optimizing the Sterilization Process in compliance with standards

  • 2. How can I ask questions? Questions & Answers
  • 4. Sterilization • Sterilization processes need to be properly controlled by the management of the responsible organization • Procedures for validation and routine control of the sterilization process and the sterile barrier system should be within the organization's QMS • Compliance with ISO 13485 provides a basis for exercising that control
  • 5. • EN 556 - Requirements for designating sterile medical devices • EN ISO 11135 - Ethylene oxide • EN ISO 11137 - Radiation • EN ISO 14937 - General criteria • EN ISO 17665 - Moist Heat • EN ISO 20857 - Dry Heat • EN ISO 14160 - Liquid chemical sterilizing agents • ISO 22441 - Low temperature vaporized hydrogen peroxide • EN ISO 25424 - Low Temperature Steam Formaldehyde Standards focus
  • 6. (102 - 10-6 = 8D Reduction) Sterility Assurance 2 0 -2 -4 -6 Probability of there being a survivor on an individual unit Number of survivors on each unit Time 10 10 10 10 10
  • 7. Countable number of organisms surviving on each unit Sterilizing treatment 106 True bioburden remaining on each unit Probability of a surviving organism remaining on any individual unit 102 10-2 10-4 10-0 10-6 104 Bioburden estimate EN ISO 11737-1 Test of Sterility EN ISO 11737-2 1 or more organisms surviving on some units in sample of practical size (up to 100 units) Impractical to sample for surviving organisms Impossible to sample for surviving organisms Sterility Assurance
  • 8. Nordic Pharmacopeia Addendum 1970 “Sterile drugs must be prepared and sterilised under conditions that aim at such a result that in one million units there will be no more than one living microorganism”. “The preparation procedure shall be controlled regularly in order to ensure a low content of microorganisms prior to sterilization.”
  • 9. 4.1 For a terminally-sterilized medical device to be designated "STERILE", the theoretical probability of there being a viable micro-organism present on/in the device shall be equal to or less than 1x10-6. EN 556-1: Requirements for terminally sterilized medical devices.
  • 10. EN 556-1: Requirements 4.2: Compliance shall be shown by the manufacturer or supplier through provision of documentation and records which demonstrate that the devices have been subjected to a validated sterilization process fulfilling 4.1
  • 11. EN 556-1: Requirements contd NOTE 1: Evidence that a medical device is sterile comes from: i) the initial validation of the sterilization process and subsequent revalidations that demonstrate the acceptability of the process, and ii) information gathered during the routine control and monitoring which demonstrates that the validated process has been delivered in practice. NOTE 2: The achievement of sterility is predicted from the bioburden level on products, the resistance of the microorganisms comprising the bioburden and the extent of treatment imposed during sterilization
  • 12. Format of sterilization standards • Common format • based on ISO 14937:2000 • Sterilization of health care products - General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices. • ISO 14937 revision published 2009 • Consistent definitions • based on ISO TS 11139 - Terminology • Common Quality Management System Elements • based on reference to ISO 13485
  • 13.  Introduction  Scope  Normative references  Definitions  Quality management system elements  Sterilizing agent characterization  Process / equipment characterization  Product definition  Process definition  Validation  Routine control and monitoring  Product release from sterilization  Maintaining process effectiveness High level structure of sterilization standards
  • 14. Process definition The purpose of this activity is to obtain a detailed specification for the sterilization process to be applied to defined product without compromising the safety, quality and performance of that product.
  • 15. Process Definition - development activities - mechanical and biological safety - process residuals - product compatibility - limits on resterilization Process Definition
  • 16. General approaches to process definition 1. Develop a specific process for a particular product 2. Demonstrate effectiveness of process already in use For each of these approaches, the microbicidal performance can be determined using: – Natural bioburden – Reference microorganisms – Conservative approach • 12 D • Half cycle • Reference process
  • 17. Process optimization • Opportunity to tailor the sterilization process to – Reduce undesirable effects of the sterilization process • Physical or chemical deterioration • Residuals – Reduce processing time – Reduce costs
  • 18. Sterilization - Process Definition Sterilization processes therefore must achieve a maximal SAL of 10-6 Seen that achievement of a maximal SAL of 10-6 by a process is dependant on the numbers and type of organisms present Approaches to demonstrating a maximal SAL of 10-6 • Conservative or overkill approach • Reference process • Use of a reference organism and a knowledge of the bioburden • Inactivation of the bioburden in its natural state
  • 19. Conservative approach Based on inactivation kinetics of a highly resistant spore present in high numbers Sterilizing treatment 106 102 10-2 10-4 10-0 10-6 104 Product bioburden Product bioburden inactivation kinetics Sterilization - Process Definition Combined bioburden and biological indicator method. Based on inactivation of an organism of know resistance greater than the natural bioburden and the numbers of organisms on the product SAL 10-6
  • 21. Control of Sterilization Control of sterilization processes requires integration of process variables. A biological indicator integrates process variables Variables Process Dose received Irradiation Time, Temperature Dry heat Time, Temperature, Moisture Moist heat Time, Temperature, EO concentration, Humidity Ethylene Oxide
  • 22. • Physical/chemical monitoring – mandatory • Chemical indicators – not mandatory • Biological indicators – not mandatory Control of Sterilization
  • 23. Biological Indicators (BIs) Definition A test system containing viable microorganisms providing a specified resistance to a specified sterilization process ISO 11139 3.29
  • 24. • Selected microorganisms with known, usually high, resistance to specific sterilization processes • Prepared on a carrier, e.g. paper, sand, cotton yarn or metal coupon = inoculated carrier • Inoculated carrier in primary packaging = biological indicator (BI) • Ready for use and with a defined resistance to specified sterilization process Description of a Biological Indicator
  • 25. Examples of process development approaches • Conservative approach – Ethylene oxide • Reference process – Moist heat • Reference organism and knowledge of bioburden – Ethylene oxide • Natural bioburden – Ionizing radiation
  • 26. Ethylene oxide – Process definition
  • 27. Annex A - D value determination • Survivor curve construction • Fraction negative (Most Probable Numbers) • All BIs at each time increment must be exposed to same conditions Annex B - Conservative process definition - (Half-cycle) • All BIs at each time increment must be exposed to same conditions • Inactivation of 12 log cycles ‒ Demonstration of 6 ‘spore log reduction’ from 106 to 100 –then double exposure time ‒ Demonstration of 8 ‘spore log reduction’ from 106 to 10-2 -then add 50% to exposure time • Complete inactivation of 106 biological indicator – then double exposure time Ethylene Oxide Process Definition - Microbiological PQ
  • 29. Demonstration of 12 log cycles of reduction 106 100 10-2 10-6 x x
  • 30. Demonstration of 12 log cycles of reduction 106 100 10-2 10-6 x x/2
  • 31. Conservative approach Based on inactivation kinetics of a highly resistant spore present in high numbers Sterilizing treatment 106 102 10-2 10-4 10-0 10-6 104 Product bioburden Product bioburden inactivation kinetics Ethylene oxide Sterilization – Combined BI and bioburden process definition Combined bioburden and biological indicator method. Based on inactivation of an organism of know resistance greater than the natural bioburden and the numbers of organisms on the product SAL 10-6
  • 32. Moist heat sterilization – Reference processes
  • 33. Moist heat reference processes Perkins (1956) Temperature Time 1320C 2 min 1250C 8 min 1210C 12 min 1180C 18 min 1160C 30 min MRC (1959) Temperature Time 1340C 3 min 1260C 10 min 1210C 15 min Added additional (safety) period to allow to for Deviations in steam quality
  • 34. Moist heat - Sterility assurance level attained by different “accepted” cycles 34 10-6 106 100 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 G stearothermophilus spores D121 1.5min Natural bioburden D121 often a few seconds SAL v’s BI with a count of 106 SAL at F0 = 8 SAL at 1210 for 15 mins 10-12 Cycle time in mins at 1210
  • 35. Radiation Sterilization - Dose setting methods
  • 36. Dose Setting Methods – Principles ..cont..  The underlying concept for establishing a ‘product specific dose’ is the use of naturally contaminated product items and a measurement of the radiation resistance of micro-organisms as they occur on the product.
  • 37. Dose Setting Methods – Product specific  ISO 11137 – Method 1 – Method 2  Use of Method 1 requires substantially less resource than that required for Method 2. Most manufacturers prefer to use Method 1 initially
  • 38. Dose Setting Methods  Method 1: – Dose setting using bioburden information  Method 2: – Dose setting using fraction positive information from incremental dosing to determine extrapolation factor – measure of radiation resistance
  • 39. Dose Setting Methods - Principles • The extent of treatment required for a sterilization process (T) is a function of: – the number of micro-organisms present prior to processing (N) – the distribution of resistances of these micro-organisms (R) – the required Sterility Assurance Level (SAL) • Or T= f(N, R, SAL)
  • 40. Radiation Sterilization Dose Setting Method 1
  • 41. Dose Setting Method 1 The choice of the sterilizing dose is based on experimental verification that the resistance to radiation of the bioburden on the device is less than that of a microbial population having a standard distribution of resistances
  • 42. Dose Setting Method 1 – ISO 11137-2
  • 43. Dose Setting Method 1 – ISO 11137-2  Using the Standard Distribution of Resistances (SDR), solutions to the equation T= f(N, R, SAL) have been generated for a range of levels of bioburden and SAL values.  These are tabulated in ISO 11137-2
  • 44. Radiation Sterilization Dose Setting - Method 2
  • 45. Dose Setting Method 2  Requires considerably more experimentation than Method 1  Provides a measurement of the radiation resistance of micro- organisms as they occur on product units.
  • 46. Dose Setting Method 2  Two experiments performed to establish estimates of: − the radiation dose giving an SAL of 10-2; – the D10 value of micro-organisms surviving this dose – Based on these data, the dose for an SAL of 10-2 can be extrapolated to that required for an SAL of 10-6
  • 47. Dose Setting Method 2 • Two experiments – Incremental Dose Experiment; – 20 items exposed to each of 9 incremental doses. Items are tested individually for sterility. Data used to estimate 10-2 SAL dose and resistance of organisms surviving that dose. – Verification Dose Experiment; – 100 items irradiated at 10-2 SAL dose to confirm dose at which 1 in 100 items non-sterile.
  • 48. Log viable microorganisms Probability of one surviving microorganism or Sterility Assurance Level (SAL) 6 5 4 3 2 1 0 -1 -2 -3 -4 -5 -6 10-1 10-2 10-3 10-4 10-5 10-6 Dose kGy Incremental Dose experiment
  • 49. Log viable microorganisms Probability of one surviving microorganism or Sterility Assurance Level (SAL) 6 5 4 3 2 1 0 -1 -2 -3 -4 -5 -6 10-1 10-2 10-3 10-4 10-5 10-6 Dose kGy Sterilization dose Verification Dose experiment
  • 50. Dose Setting Method 2 • Calculation Sterilizing Dose is calculated by extrapolation to the required SAL from the verified 10-2 SAL dose using the estimate of the resistance of the micro-organisms surviving that dose.
  • 51. Radiation Sterilization - Dose Substantiation
  • 52. Dose substantiation - Principles T= f(N, R, SAL) • In the case of substantiation of 25kGy: – T = 25kGy – N is determined in bioburden studies – SAL = 10-6 • Solve the equation for the maximum resistance that can be tolerated to achieve SAL = 10-6 with 25 kGy 25kGy= f(Bioburden, R, 10-6) or R=f(Bioburden, 25kGY, 10-6) • Verify that this resistance is not exceeded
  • 53. ISO TS 13004: 2013 - ISO 13004:2022 • Sterilization of health care products – Radiation – Substantiation of a selected sterilization dose: Method VDmax SD • Covers substantiation of sterilization doses from 17.5 to 35 kGy at 2.5 kGy increments • Adopted as a European Technical Specification • Not a harmonized European standard • Intended to gain experience before considering for inclusion in a future revision of ISO 11137-2
  • 54. Process Optimization - Conclusion • Different approaches to process definition provide different degrees of process optimization – Radiation Method 2 – Radiation Method 1 – Radiation Dose Substantiation (Vdmax SD) – Ethylene oxide combined BI/bioburden approach – Moist heat reference process – Ethylene oxide 12D approach – Ethylene oxide half-cycle Increasing optimization
  • 55.
  • 56. Ethylene Oxide Sterilization To Learn More About ETO Sterilisation Please Join Us When : 17TH – 20TH April 2023 Monday – Thursday Half days 1pm – 5pm (GMT) Where: Virtual Instructor led Promo Code: EARLYBIRD10 Trainer: Eamonn Hoxey