QUALITY CONTROL IN
LABORATORY
WHAT IS QC ?
• QC is a technique that is used to detect &
correct errors before they result in a defective
product or service.
• It refers to the measures that must be
included during each assay run to verify that
the test is working properly.
• QC must be practical, achievable & affordable.
TERMINOLOGY
• PRECISION: This indicates how close test
measurements to each other when the same
test is run on the same sample repeatedly.
• Precision does not imply accuracy.
• Precision does not imply accuracy.
• ACCURACY: How close to the true value a
measurement is.
• The closer to the actual value, the more
accurate.
• STANDARD: This is a substance of constant
composition of sufficient purity to be used for
comparison purpose.
• CONTROL: This is a sample i.e. chemically &
physically similar to the unknown specimen.
physically similar to the unknown specimen.
• STANDARD DEVIATION: This is a statistical
expression of scatter or dispersion of values
around a central average value.
Calculation of Standard Deviation
Calculation of Standard Deviation
Coefficient of Variation
Coefficient of Variation
• The Coefficient of Variation (CV) is the
standard Deviation (SD) expressed as a
percentage of the mean
-Also known as Relative Standard
deviation (RSD)
• CV % = SD x 100
Mean
CALCULATION OF MEAN
Data set
(30.0, 32.0, 31.5, 33.5, 32.0, 33.0, 29.0,29.5, 31.0,
32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0,
32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0,
35.0, 32.5.) mg/dL
The sum of the values (X1 + X2 + X3 … X20)
divided by the number (n) of observations
The mean of these 20 observations is (639.5 ÷ 20)
= 32.0 mg/dL
P
PRECISE & INACCURATE
RECISE & INACCURATE
IMPRECISE & INACCURATE
IMPRECISE & INACCURATE
PRECISE & ACCURATE
PRECISE & ACCURATE
L. J. CHART
• L-J chart is a graph that quality control data is
plotted on to give a visual indication whether
a laboratory test is working well.
• It is named after S.LEVEY & E.R.JENNINGS in
1950.
lj with Manufacturer & Lab Calculated Mean & SD
213
234
255
276
297
318
339
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Glucose (Jan 2011) with Manufacturer Mean & Ranges (SD – 21 ; CV- 7.6%)
Level 3
266.58
269.31
272.04
274.77
277.5
280.23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Glucose (Jan 2011) with Lab defined Mean & Ranges(SD – 2.73; CV- 0.99%)
Level 3
Normal Distribution
Normal Distribution
• All values are symmetrically distributed around the mean
• Characteristic “bell-shaped” curve
• Assumed for all quality control statistics
INTERNAL QUALITY CONTROL
• Use of standard glassware , reagent
, equipment
• Well trained staff
• Selection of accurate & precise
method.
• At least one primary std. is included
with each batch of unknown specimen
analysis.
• Occasionally a different primary std . of
higher concentration is included to
check the reliability of routine primary
higher concentration is included to
check the reliability of routine primary
std.
• The batch result are accepted if the
values of control sera are within first SD.
Types of Errors
• Intrinsic : due to an imprecise &
inaccurate method.
• Systemic : all high values , all low
values.
• Random : technical errors.
• Intrinsic errors can be eliminated by
A) selecting precise & accurate method.
B) by using fresh batches of reagents.
• systemic errors can be eliminated by using diff.
concentration of primary std. & analyzing
concentration of primary std. & analyzing
control serum.
• Technical with a proper analytical application
can be correct random errors.
Systematic vs. Random Errors
Systematic vs. Random Errors
Systematic Error
Avoidable error due to
controllable variables in a
measurement.
Random Errors
Unavoidable errors that
are always present in any
measurement. Impossible
measurement. measurement. Impossible
to eliminate
PREVENTIVE PHASE
• Collection of specimen
• Separation of serum
• Specimen analysis
• Calculation of test values etc..
RETROSPECTIVE PHASE
• Optimum condition variance
• Routine condition variance
OPTIMUM CONDITION VARIANCE
• Refers to the results obtained under
optimum conditions i.e. by using
i)freshly prepared reagents
ii)by using standardized glassware
ROUTINE CONDITIONS VARIANCE
• Refers to the results obtained by using routine
requirements i.e. by using routinely stored
reagents and glassware in regular use.
WESTGARD RULES
• 1-2S
• 1-3S
• 2-2S
• 2-2S
• R-4S
• 4-1S
• 10X
• Multi control QC rules (WESTGARD RULES)
given by Dr. James Westgard of the University of
Wisconsin in an article in 1981 on laboratory
quality control that set the basis for evaluating
analytical run quality for medical laboratories.
• The Westgard system -based on the principles of
statistical process control used in manufacturing
statistical process control used in manufacturing
nationwide since the 1950s
• Six basic rules in the Westgard scheme: 1-3s, 2-
2s, R-4s, 1-2s, 4-1s, and 10x. These rules are used
individually or in combination (multi-rule) to
evaluate the quality of analytical runs.
• Detect random or systematic errors
• Warning 12SD or 1-2s:
It is violated if the single IQC
value exceeds the mean by ± 2SD.
• Rejection 22SD or 2-2s:
• This rule detects systematic error and is applied within and across
runs.
• It is violated within the run when two consecutive control values
exceed the "same" (mean + 2s or mean - 2s) limit.
• The rule is violated across runs when the previous value for a
particular control level exceeds the "same"
(mean + 2s or mean - 2s) limit.
(mean + 2s or mean - 2s) limit.
Within run violation Across run violation
• Rejection 13SD or 1-3s:
• It is violated when the single IQC
value exceeds the mean by ±3SD.
• This rule is applied within control
material only.
• The 1-3s rule identifies unacceptable
random error or possibly the
beginning of a large systematic error.
• Rejection 41SD or 4-1s:
It is violated if four consecutive IQC
values exceed the same mean plus 1s
or the same mean minus 1s control limit.
• Rejection 10x:
• This rule detects systematic bias and is applied both within and
across control materials.
• It is violated across control materials if the last 10 consecutive
values, regardless of control level, are on the same side of the
mean.
• The rule is violated within the control materials if the last 10
values for the same control level are on the same side of the
mean.
Westgard Procedure Flowchart
Westgard Procedure Flowchart
EXTERNAL QUALITY CONTROL
• All the participating laboratories daily
analyze the same lot of control material.
• The results are tabulated monthly & sent
the sponsoring groups for the data
the sponsoring groups for the data
analysis.
• Summary reports are prepared by the
program sponsor & are distributed to all
participating laboratories .
• The mean of values of all reference
laboratories is taken as the “ true “ or
correct value & is used for comparison
with the individual laboratory reported
values.
values.
• If the difference between the reported
value & the true value is statistically
significant then the reporting lab is
alerted.
THANK YOU

More Related Content

Similar to safety_in_laboratory.pdf

Quality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryQuality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryDr. Bikash Kumar Chaudhury
 
Quality control
Quality controlQuality control
Quality controlSKYFALL
 
(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptxDr.Rajeev Ranjan
 
medical -laboratory- quality- control --
medical -laboratory- quality- control --medical -laboratory- quality- control --
medical -laboratory- quality- control --zahraaali64
 
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptxANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptxDr. Jagroop Singh
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryDipesh Tamrakar
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Adel Elazab Elged
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control NotesLeah Molai
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJAMamunReja1
 
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Ola Elgaddar
 
Method validation terms quality control and assurance
Method validation terms quality control and assurance Method validation terms quality control and assurance
Method validation terms quality control and assurance MuhammadImranMirza2
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .pptABRARAHMED767665
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Randox
 

Similar to safety_in_laboratory.pdf (20)

Quality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryQuality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing Laboratory
 
Quality assurance
Quality assuranceQuality assurance
Quality assurance
 
Quality control
Quality controlQuality control
Quality control
 
Clinical lab qc sethu
Clinical lab qc sethuClinical lab qc sethu
Clinical lab qc sethu
 
(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx
 
medical -laboratory- quality- control --
medical -laboratory- quality- control --medical -laboratory- quality- control --
medical -laboratory- quality- control --
 
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptxANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistry
 
Quality control chemical
Quality control chemicalQuality control chemical
Quality control chemical
 
Quality control.pptx
Quality control.pptxQuality control.pptx
Quality control.pptx
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control Notes
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJA
 
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012
 
Method validation terms quality control and assurance
Method validation terms quality control and assurance Method validation terms quality control and assurance
Method validation terms quality control and assurance
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
 
ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
 
Quality Control in Laboratory
Quality Control in LaboratoryQuality Control in Laboratory
Quality Control in Laboratory
 
QA-QC-PSK.pdf
QA-QC-PSK.pdfQA-QC-PSK.pdf
QA-QC-PSK.pdf
 

Recently uploaded

Low Rate Call Girls Patna Anika 8250192130 Independent Escort Service Patna
Low Rate Call Girls Patna Anika 8250192130 Independent Escort Service PatnaLow Rate Call Girls Patna Anika 8250192130 Independent Escort Service Patna
Low Rate Call Girls Patna Anika 8250192130 Independent Escort Service Patnamakika9823
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call girls in Ahmedabad High profile
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls ServiceMiss joya
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoybabeytanya
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...Miss joya
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...indiancallgirl4rent
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 

Recently uploaded (20)

Low Rate Call Girls Patna Anika 8250192130 Independent Escort Service Patna
Low Rate Call Girls Patna Anika 8250192130 Independent Escort Service PatnaLow Rate Call Girls Patna Anika 8250192130 Independent Escort Service Patna
Low Rate Call Girls Patna Anika 8250192130 Independent Escort Service Patna
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 

safety_in_laboratory.pdf

  • 2. WHAT IS QC ? • QC is a technique that is used to detect & correct errors before they result in a defective product or service. • It refers to the measures that must be included during each assay run to verify that the test is working properly. • QC must be practical, achievable & affordable.
  • 3. TERMINOLOGY • PRECISION: This indicates how close test measurements to each other when the same test is run on the same sample repeatedly. • Precision does not imply accuracy. • Precision does not imply accuracy. • ACCURACY: How close to the true value a measurement is. • The closer to the actual value, the more accurate.
  • 4. • STANDARD: This is a substance of constant composition of sufficient purity to be used for comparison purpose. • CONTROL: This is a sample i.e. chemically & physically similar to the unknown specimen. physically similar to the unknown specimen. • STANDARD DEVIATION: This is a statistical expression of scatter or dispersion of values around a central average value.
  • 5. Calculation of Standard Deviation Calculation of Standard Deviation
  • 6. Coefficient of Variation Coefficient of Variation • The Coefficient of Variation (CV) is the standard Deviation (SD) expressed as a percentage of the mean -Also known as Relative Standard deviation (RSD) • CV % = SD x 100 Mean
  • 7. CALCULATION OF MEAN Data set (30.0, 32.0, 31.5, 33.5, 32.0, 33.0, 29.0,29.5, 31.0, 32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0, 32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0, 35.0, 32.5.) mg/dL The sum of the values (X1 + X2 + X3 … X20) divided by the number (n) of observations The mean of these 20 observations is (639.5 ÷ 20) = 32.0 mg/dL
  • 11. L. J. CHART • L-J chart is a graph that quality control data is plotted on to give a visual indication whether a laboratory test is working well. • It is named after S.LEVEY & E.R.JENNINGS in 1950.
  • 12.
  • 13.
  • 14. lj with Manufacturer & Lab Calculated Mean & SD 213 234 255 276 297 318 339 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Glucose (Jan 2011) with Manufacturer Mean & Ranges (SD – 21 ; CV- 7.6%) Level 3 266.58 269.31 272.04 274.77 277.5 280.23 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Glucose (Jan 2011) with Lab defined Mean & Ranges(SD – 2.73; CV- 0.99%) Level 3
  • 15. Normal Distribution Normal Distribution • All values are symmetrically distributed around the mean • Characteristic “bell-shaped” curve • Assumed for all quality control statistics
  • 16.
  • 17. INTERNAL QUALITY CONTROL • Use of standard glassware , reagent , equipment • Well trained staff • Selection of accurate & precise method.
  • 18. • At least one primary std. is included with each batch of unknown specimen analysis. • Occasionally a different primary std . of higher concentration is included to check the reliability of routine primary higher concentration is included to check the reliability of routine primary std. • The batch result are accepted if the values of control sera are within first SD.
  • 19. Types of Errors • Intrinsic : due to an imprecise & inaccurate method. • Systemic : all high values , all low values. • Random : technical errors.
  • 20. • Intrinsic errors can be eliminated by A) selecting precise & accurate method. B) by using fresh batches of reagents. • systemic errors can be eliminated by using diff. concentration of primary std. & analyzing concentration of primary std. & analyzing control serum. • Technical with a proper analytical application can be correct random errors.
  • 21. Systematic vs. Random Errors Systematic vs. Random Errors Systematic Error Avoidable error due to controllable variables in a measurement. Random Errors Unavoidable errors that are always present in any measurement. Impossible measurement. measurement. Impossible to eliminate
  • 22. PREVENTIVE PHASE • Collection of specimen • Separation of serum • Specimen analysis • Calculation of test values etc..
  • 23. RETROSPECTIVE PHASE • Optimum condition variance • Routine condition variance
  • 24. OPTIMUM CONDITION VARIANCE • Refers to the results obtained under optimum conditions i.e. by using i)freshly prepared reagents ii)by using standardized glassware
  • 25. ROUTINE CONDITIONS VARIANCE • Refers to the results obtained by using routine requirements i.e. by using routinely stored reagents and glassware in regular use.
  • 26. WESTGARD RULES • 1-2S • 1-3S • 2-2S • 2-2S • R-4S • 4-1S • 10X
  • 27. • Multi control QC rules (WESTGARD RULES) given by Dr. James Westgard of the University of Wisconsin in an article in 1981 on laboratory quality control that set the basis for evaluating analytical run quality for medical laboratories. • The Westgard system -based on the principles of statistical process control used in manufacturing statistical process control used in manufacturing nationwide since the 1950s • Six basic rules in the Westgard scheme: 1-3s, 2- 2s, R-4s, 1-2s, 4-1s, and 10x. These rules are used individually or in combination (multi-rule) to evaluate the quality of analytical runs. • Detect random or systematic errors
  • 28. • Warning 12SD or 1-2s: It is violated if the single IQC value exceeds the mean by ± 2SD.
  • 29. • Rejection 22SD or 2-2s: • This rule detects systematic error and is applied within and across runs. • It is violated within the run when two consecutive control values exceed the "same" (mean + 2s or mean - 2s) limit. • The rule is violated across runs when the previous value for a particular control level exceeds the "same" (mean + 2s or mean - 2s) limit. (mean + 2s or mean - 2s) limit. Within run violation Across run violation
  • 30. • Rejection 13SD or 1-3s: • It is violated when the single IQC value exceeds the mean by ±3SD. • This rule is applied within control material only. • The 1-3s rule identifies unacceptable random error or possibly the beginning of a large systematic error. • Rejection 41SD or 4-1s: It is violated if four consecutive IQC values exceed the same mean plus 1s or the same mean minus 1s control limit.
  • 31. • Rejection 10x: • This rule detects systematic bias and is applied both within and across control materials. • It is violated across control materials if the last 10 consecutive values, regardless of control level, are on the same side of the mean. • The rule is violated within the control materials if the last 10 values for the same control level are on the same side of the mean.
  • 33. EXTERNAL QUALITY CONTROL • All the participating laboratories daily analyze the same lot of control material. • The results are tabulated monthly & sent the sponsoring groups for the data the sponsoring groups for the data analysis. • Summary reports are prepared by the program sponsor & are distributed to all participating laboratories .
  • 34. • The mean of values of all reference laboratories is taken as the “ true “ or correct value & is used for comparison with the individual laboratory reported values. values. • If the difference between the reported value & the true value is statistically significant then the reporting lab is alerted.