SlideShare a Scribd company logo
1 of 75
VENTELATOR SETTINGS
DR- ATEIA EL MEHDOWI
:
CONTENTS
• Manipulation of ventilators .
• Components of ventilator .
• Initial ventilator settings .
• Safety considerations in mechanical ventilators .
• Golden rules in mechanical ventilation .
Manipulation of ventilators
The adjustment of ventilators requires recognition of:
1. Clinical and metabolic condition of the patient.
2. Characteristics of the machine.
3. A base line ABG is absolutely essential to determine
the starting point for treatment of hypercapnia and
acidosis and/or hypoxia.
The PaCO2 level dose not have the same importance as
pH level, since patient consciousness varies with acute
versus chronic hypercapnia.
There should be a manual resuscitation bag at the
bedside of every patient receiving mechanical ventilation.
Ventilators are a method of life-support; if the ventilator
stops working ??????
Physician
Nurse
RT
Patient
Ventilator
Success of mechanical ventilation
COMPONENTS OF VENTILATORS
Grouped into 4 categories:
1. Controls.
3. Alarms
2. Monitors.
4. External features & accessories.
CONTROLS
CONTROLS
The number and type of controls in specific
ventilator depend mostly on their purpose.
 Critical Care
 Neonatal Intensive Care
 Pediatric Intensive Care Unit
 Post Anesthesia Care Unit (PACU)
 Home
 MRI
 Transport – Ground/Air/Hospital
 OR
A. Short-term ventilators: have fewer controls.
CONTROLS
They are more
difficult to operate.
A. Long-term ventilators:
They have large variety of controls
They have great
flexibility in clinical
application
CONTROLS
CONTROLS
CONTROLS
2- Microprocessor-controlled ventilators
Microprocessor-controlled ventilators
• Pressure.
• Volume (Vm & Vt ).
• Rate.
• FIO2.
• I/E ratio or I/E time.
• Insp. Flow.
• Sensitivity.
• Insp. Pause (inflation hold).
• PEEP.
• Sigh (volume & frequency).
CONTROLS
 Tidal volume (Vt)
The clinician sets:
 Frequency
 Trigger sensitivity.
 PEEP
VCV
(PCV)
The clinician sets:
 Peak inspiratory pressure (PIP)
. (Complete control of peak pressure).
 Frequency
 Inspiratory time (It).
 Rise time
 Trigger sensitivity.
 PEEP
e.g.. VCVs do not provide an independent pressure control.
• Control on ventilators may provide an approximate values.
To confirm the accuracy of controls, an effective monitoring
capability is needed.
e.g.. PCVs do not provide an independent volume control.
• There are several functions for the ventilators have no controls,
yet these must be monitored
CONTROLS ..
INITIAL VENTILATOR SETTINGS
Mode of ventilation
• The initial mode of ventilation should be the assist
control (A/C) mode.
• In the A/C mode, the work of breathing is reduced to
only that amount of inspiration needed to trigger the
machine's inspiratory cycle.
• A second possible advantage is that cycling the
ventilator into the inspiratory phase maintains normal
ventilatory activity and, therefore, prevents atrophy of
the respiratory muscles.
The potential disadvantages of the A/C mode include
respiratory alkalosis.
A second possible disadvantage is that every breath is
a full, preset, positive-pressure breath that influences
venous return to the right heart and, possibly, to
global cardiac output.
Nevertheless, the A/C mode is the best initial mode of
ventilation and may be switched to another option if
hypotension occurs or hypocarbia is evident on the
first arterial blood gas analysis.
Fractional Inspired Oxygen (FIO2)
The fractional inspired oxygen is the amount of oxygen
delivered to the patient.
It can range from 21% (room air) to 100%.
It’s recommended that the FIO2 be set at 1.0 (100%)
upon the initiation of mechanical ventilation to allow
the patient to get used to the ventilator without
experiencing hypoxia.
A short period on FIO2 of 100% is not
dangerous to the patient on mechanical
ventilation and offers the clinician several
advantages:
• protects the patient against hypoxemia
if unrecognized problems develop from
the intubation procedure.
• PaO2 measured on an FIO2 of
100%, the clinician can easily
calculate the next desired FIO2 and
quickly estimate the shunt fraction.
When the clinician wishes to change the PaO2,
2 variables might be adjusted
FIO2 and the amount of PEEP.
When using supplemental FIO2, improvement in
oxygenation cannot reliably be achieved with changes
in rate or tidal volume.
With the target PaO2 identified, the FIO2 can be
adjusted using the formula:
New FIO2=(old FIO2 X desired PaO2)/measured PaO2.
The degree of shunt on 100% FIO2 can be
estimated by a rough rule of thumb.
The measured PaO2 is subtracted from 700 mm
Hg. For each 100 mm Hg difference, a 5% shunt
exists.
A shunt of approximately 25% requires the use of
PEEP.
Positive end-expiratory pressure [PEEP]
PEEP therapy can be effective when used in patients
with a diffuse lung disease that results in an acute
decrease in functional residual capacity (FRC)
when used to treat patients with a diffuse lung disease,
PEEP should improve compliance, decrease dead space,
and decrease the intrapulmonary shunt effect. The most
significant benefit of PEEP is that the patient can
maintain an adequate PaO2 at a lower, safer
concentration of oxygen (<60%), thereby reducing the
risk of oxygen toxicity.
No recommendations exist for adding external PEEP
during initial ventilator setup to satisfy misguided
attempts to supply prophylactic PEEP or physiologic
PEEP.
Most clinicians use the lowest amount of positive
pressure that provides an adequate PaO2 with a safe
FIO2.
PEEP
PEEP principally is used to lower the risks of oxygen
toxicity and is employed when a safe PaO2 cannot be
achieved at 60% FIO2.
Initiated at 5 cm H2O, PEEP usually is increased in
3 cm H2O increments, while evaluating the effect on
oxygenation every 15-20 minutes.
A PEEP level of 10 cm H2O rarely causes
hemodynamic problems in the absence of
intravascular volume depletion.
changes in PEEP may not be reflected by changes in
arterial blood gases for 20-30 minutes so changes
in the PEEP setting should usually not be made
faster than this
PEEP over 20 cm is rarely beneficial and usually
results in additional pressure-induced lung injury
BEST PEEP ?
The exact amount of PEEP applied is very
controversial
?????????????????????????????????
Pressure (cm H2O)
0 50
Volume
(ml)
1500
0
Alveoli elastic limit
UIP
LIP Alveoli open
Alveolar collapse
PEEP
Pressure limit
Pressure-volume curve
The titration of PEEP above the lower
inflection point  increases lung compliance
“best PEEP”.
The titration of PIP below the upper inflection
point  prevention of over destintion
Pressure-Volume loope
400
600
200
0 5 10 20 30 40 50
Press
Vt
300
500
700
100
Lower inflection point
Upper inflection point
PIP
PEEP
Pressure-Volume loope
The respiratory rate is the number of breaths
that
the ventilator delivers to the patient each
minute.
The rate chosen depends on the :
1. Tidal volume.
2. Type of pulmonary pathology.
3. patient’s target PaCO2.
Respiratory Rate (RR)
Patients with restrictive lung disease usually
tolerate a range of 12-20 breaths/minute.
Patients with normal pulmonary mechanics can
tolerate a rate of 8-12 breaths/minute.
For patients with obstructive lung disease, the
rate should be set at 6-8 breaths/minute to avoid
the development of auto-PEEP and
hyperventilation, or “blowing off CO2.”. Patients
with obstructive lung disease often adapt to a
higher PaCO2, so lowering it back to the “normal”
range of 35-45 mm Hg may not be beneficial.
The respiratory rate parameters are set above
and below this number and the alarm will
then sound if the patient’s actual rate is
outside of the desired range.
The usual setting is 5-12 cc/kg.
based on compliance, resistance, and type of
pathology.
Tidal Volume (VT)
• Patients with normal lungs can tolerate a tidal
volume of 10-12 cc/kg,
• Patients with restrictive lung disease may
need a tidal volume of 5-8 cc/kg.
Tidal Volume (VT)
Males Ideal body weigh IBW =50+2.3 (Height (inches) – 60 )
Females Ideal body weigh IBW =50+2.3 (Height (inches) – 60 )
The tidal volume alarm level are set above and below
the desired number, and the alarm will sound if the
patient’s actual tidal volume is outside of the desired
range.
Tidal Volume (VT)
Double-checking the selected tidal volume
After a tidal volume is selected, the peak airway and
Plateau pressures pressure should be determined
peak airway pressure <45 cm H2O.
plateau pressure is <30-35 cm H2O.
Tow Pressure level :
 Small values, which just to compensate for
resistance of the ETT.
 Large values, that make the spontaneous breath
equivalent to triggered mandatory breath (PCV).
Pressure suppot PS
Pressure
support
PS only eliminates the work precisely at a given flow.
Above and below that flow, PS under compensates for
resistance or over compensates.
(Weaning)
ATC
ATC compensation can overcome resistance
regardless of patient flow demand.
The I:E ratio is usually set at
1:2 or 1:1.5
to approximate the normal
physiology of inspiration and
expiration.
Occasionally, a longer inspiratory than
expiratory time is desired to allow more time to
oxygenate the patient’s lungs. This is called
inverse ratio ventilation.
Inspiratory:Expiratory (I:E) Ratio
The pressure limit regulates the amount of pressure
the VCV can generate to deliver the preset tidal volume.
Because high pressures can cause lung injury,
it’s recommended that the plateau pressure not exceed
35 cm H20.
If this limit is reached, the ventilator stops delivering the
breath and alarms.
This may be an indication that the patient’s airway is
obstructed with mucus. It can also be caused by the
patient coughing, biting on the ETT, breathing against
the ventilator, or by a kink in the ventilator tubing.
Pressure Limit
The flow rate is the speed with which the tidal volume is
delivered.
Flow Rate
The usual setting is 40-100 liters per minute.
The sensitivity determines the amount of effort
required by the patient to initiate inspiration.
Sensitivity/Trigger
It can be:
Pressure triggering 1 – 2 cm H2O
flow triggering. 1- 5 L/min
Flow triggering is a better setting for patients
who can breathe spontaneously because it
reduces the work of breathing.
time
pressure
trigger-
threshold
• Very fast rising
Depending on rise time, the beginning of the flow will be
• or slower
Rise time
For adults the range is
0-0.4 seconds
The ventilator can be programmed to deliver an
occasional sigh with a larger tidal volume. The use
of frequent sighs was popular during the 1970’s
because it was thought that it prevented collapse
of the alveoli (atelectasis), which can result from
the patient constantly inspiring the same volume
of gas.
However, recently there has been concern that
the increased pressure produced in the alveoli
may heighten the risk of the alveoli rupturing
and causing pneumothorax.
Sigh
Summary of initial ventilator set up
• A/C Mode
• Tidal volume depending on lung status
Normal lungs - 12 cc/kg ideal body weight
COPD - 10 cc/kg ideal body weight
ARDS - 6-8 cc/kg ideal body weight
• Rate of 10-12 breaths per minute
• FIO2 of 100%
• Sighs rarely needed
• PEEP only as indicated after first arterial blood gas
Shunt greater than 25%
Inability to oxygenate with safe FIO2
What is ASV?
Conventional
Tp
Rate
Psup
Pinsp
Ti
Te @
FiO2
PEEP
Vt
CMV
SIMV
PCV
PSV
Goal 3: Simple to use
Modes
Controls
What is ASV?
Conventional
Tp
Rate
Psup
Pinsp
Ti
Te @
FiO2
PEEP
Vt
CMV
SIMV
PCV
PSV
ASV
Goal 3: Simple to use
Controls
What is ASV?
Conventional
Tp
Rate
Psup
Pinsp
Ti
Te @
FiO2
PEEP
Vt
CMV
SIMV
PCV
PSV
ASV
Goal 3: Simple to use
What is ASV?
Conventional
Tp
Rate
Psup
Pinsp
Ti
Te @
FiO2
PEEP
Vt
CMV
SIMV
PCV
PSV
ASV
Goal 3: Simple to use
Tp
Rate Pinsp
Ti Te
PF
FiO2
PEEP
Vt
CMV
SIMV
PCV
PSV
Alveolar Ventilation Oxy.
Conventional
FiO2
PEEP
MV
Automatic
Mechanical ventilation
2. MONITORS
Ventilators are designed to
monitor many components of
the patient’s respiratory status.
Patient monitoring
peak , mean , plateau and PEEP.
Pressure:
Vt, Vm, sigh volume.
Volume:
Frequency
(mandatory &
spontaneous).
MONITORS
I/E ratio It, Et.
FIO2.
MONITORS
Lung mechanics
Monitoring
Flow-time curve
Pressure-time curve
Volume-pressure loop
Flow-volume loop
Nigative
inspiratory
force (NIF)
AutoPEEP
Airway Resistance
Lung compliance
3. Alarms
Help
3. ALARMS
To ensure the safety and effectiveness of MV
electric or pneumatic alarms are installed on
ventilator to signal both visually and audibly
the presence of undesirable conditions.
Problem solving
• pCO2 too high !!!!!!!!!!!!!!
• pO2 too low !!!!!!!!!!!!!!!!!!!!!!!!!!
• PIP too high !!!!!!!!!!!!!!!!!!!!!!!
????????????????????????????????
 Safety considerations in mechanical ventilators
 Back up ventilation
 Low resistance non compliant breathing
circuit
 Limits on closed loop adjustments
 Reliable alarm system
 Battery for electrical failure
Golden rules in mechanical ventilation
5- Adapt the ventilator to the patient instead of
adapting the patient to the ventilator.
2- A mode doesn’t make the lung healthy it
wins only time.
3- Keep healthy parts of the lung healthy.
4- Maintain and support spontaneous
breathing in ventilated patient.
1- Start the process of weaning the moment you
start mechanical ventilation.
8- Avoid airway pressure more than 35
cmH2O
6- Employ tidal volume of 6 – 8 ml/kg Wt.
9- Don’t be too shy with PEEP.
10- Set always the trigger at the highest sensitivity
(avoid autocycling).
7- Give 7 – 8 cmH2O pressure support for
compensation of the tube resistance.
Vt PS PAP PEEP Triggering
pCO2 Too High
• Patient’s minute ventilation is too low.
• Increase rate or TV or both.
• If using PC ventilation, increase PIP.
• If PIP too high, increase the rate instead.
• If air-trapping is occurring, decrease the rate and the I-
time and increase the TV to allow complete exhalation.
pO2 Too Low
• Increase either the FiO2 or the mean airway
pressure (MAP).
• Try to avoid FiO2 >70%.
• Increasing the PEEP is the most efficient way
of increasing the MAP in the PICU.
• Can also increase the I-time to increase the
MAP (PC).
• Can increase the PIP in Pressure Control to
increase the MAP, but this generally doesn’t
add much at rates <30bpm.
PIP Too High
• Decrease the PIP (PC) or the TV (VC).
• Increase the I-time (VC).
• Change to another mode of ventilation. Generally,
pressure control achieves the same TV at a lower PIP
than volume control.
• If the high PIP is due to high airway resistance, generally
the lung is protected from barotrauma unless air-trapping
occurs.
Acute Deterioration
• DIFFERENTIAL DIAGNOSES
o Pneumothorax
o Right main stem intubation
o Pneumonia
o Pulmonary edema
o Airway occlusion
o Ventilator malfunction
o Mucus plugging
o Air leak
o hypoperfusion

More Related Content

Similar to 2.Ventilator Settings by Dr.Ateia.pptx

INITIAL VENTILATOR SETTINGS 4 nurses
INITIAL VENTILATOR SETTINGS 4 nursesINITIAL VENTILATOR SETTINGS 4 nurses
INITIAL VENTILATOR SETTINGS 4 nurses
Manish Masih
 
04 venugopal.
04 venugopal.04 venugopal.
04 venugopal.
Duy Quang
 

Similar to 2.Ventilator Settings by Dr.Ateia.pptx (20)

Mechanical Ventilation
Mechanical VentilationMechanical Ventilation
Mechanical Ventilation
 
Basics of Mechanical Ventilation
Basics of Mechanical VentilationBasics of Mechanical Ventilation
Basics of Mechanical Ventilation
 
Chapter-2.ppt
Chapter-2.pptChapter-2.ppt
Chapter-2.ppt
 
Chapter-2.ppt
Chapter-2.pptChapter-2.ppt
Chapter-2.ppt
 
Mechanical ventilation
Mechanical ventilationMechanical ventilation
Mechanical ventilation
 
Basic ventilator management
Basic ventilator managementBasic ventilator management
Basic ventilator management
 
Mechanical ventilation منتدى تمريض مستشفى غزة الاوروب
Mechanical ventilation منتدى تمريض مستشفى غزة الاوروبMechanical ventilation منتدى تمريض مستشفى غزة الاوروب
Mechanical ventilation منتدى تمريض مستشفى غزة الاوروب
 
Basic modes of mv
Basic modes of mvBasic modes of mv
Basic modes of mv
 
Mechanical Ventilation (1).ppt
Mechanical Ventilation (1).pptMechanical Ventilation (1).ppt
Mechanical Ventilation (1).ppt
 
INITIAL VENTILATOR SETTINGS 4 nurses
INITIAL VENTILATOR SETTINGS 4 nursesINITIAL VENTILATOR SETTINGS 4 nurses
INITIAL VENTILATOR SETTINGS 4 nurses
 
Basics of mechanical ventilation
Basics of mechanical ventilationBasics of mechanical ventilation
Basics of mechanical ventilation
 
Mechanical ventilator, common modes, indications,nursing responsibilities
Mechanical ventilator, common modes, indications,nursing responsibilities Mechanical ventilator, common modes, indications,nursing responsibilities
Mechanical ventilator, common modes, indications,nursing responsibilities
 
Modern Ventilator Management.pptx
Modern Ventilator Management.pptxModern Ventilator Management.pptx
Modern Ventilator Management.pptx
 
04 venugopal.
04 venugopal.04 venugopal.
04 venugopal.
 
Mechanical Ventilation Cheat Book for Internal Medicine Residents
Mechanical Ventilation Cheat Book for Internal Medicine ResidentsMechanical Ventilation Cheat Book for Internal Medicine Residents
Mechanical Ventilation Cheat Book for Internal Medicine Residents
 
VENTILATOR BASICS.pdf
VENTILATOR BASICS.pdfVENTILATOR BASICS.pdf
VENTILATOR BASICS.pdf
 
Modern Ventilator Management.pptx
Modern Ventilator Management.pptxModern Ventilator Management.pptx
Modern Ventilator Management.pptx
 
Mechanical Ventilator
Mechanical VentilatorMechanical Ventilator
Mechanical Ventilator
 
Care of child requiring long term ventilation
Care of child requiring long term ventilationCare of child requiring long term ventilation
Care of child requiring long term ventilation
 
Mv medumat standard2_arht_wood_v2
Mv medumat standard2_arht_wood_v2Mv medumat standard2_arht_wood_v2
Mv medumat standard2_arht_wood_v2
 

Recently uploaded

Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan 087776558899
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
amritaverma53
 
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
chanderprakash5506
 

Recently uploaded (20)

Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in ChennaiChennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICEBhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
 
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
 

2.Ventilator Settings by Dr.Ateia.pptx

  • 2.
  • 3. : CONTENTS • Manipulation of ventilators . • Components of ventilator . • Initial ventilator settings . • Safety considerations in mechanical ventilators . • Golden rules in mechanical ventilation .
  • 4. Manipulation of ventilators The adjustment of ventilators requires recognition of: 1. Clinical and metabolic condition of the patient. 2. Characteristics of the machine. 3. A base line ABG is absolutely essential to determine the starting point for treatment of hypercapnia and acidosis and/or hypoxia. The PaCO2 level dose not have the same importance as pH level, since patient consciousness varies with acute versus chronic hypercapnia.
  • 5. There should be a manual resuscitation bag at the bedside of every patient receiving mechanical ventilation. Ventilators are a method of life-support; if the ventilator stops working ??????
  • 7. COMPONENTS OF VENTILATORS Grouped into 4 categories: 1. Controls. 3. Alarms 2. Monitors. 4. External features & accessories.
  • 9. CONTROLS The number and type of controls in specific ventilator depend mostly on their purpose.  Critical Care  Neonatal Intensive Care  Pediatric Intensive Care Unit  Post Anesthesia Care Unit (PACU)  Home  MRI  Transport – Ground/Air/Hospital  OR
  • 10. A. Short-term ventilators: have fewer controls. CONTROLS
  • 11. They are more difficult to operate. A. Long-term ventilators: They have large variety of controls They have great flexibility in clinical application CONTROLS
  • 16. • Pressure. • Volume (Vm & Vt ). • Rate. • FIO2. • I/E ratio or I/E time. • Insp. Flow. • Sensitivity. • Insp. Pause (inflation hold). • PEEP. • Sigh (volume & frequency). CONTROLS
  • 17.  Tidal volume (Vt) The clinician sets:  Frequency  Trigger sensitivity.  PEEP VCV
  • 18. (PCV) The clinician sets:  Peak inspiratory pressure (PIP) . (Complete control of peak pressure).  Frequency  Inspiratory time (It).  Rise time  Trigger sensitivity.  PEEP
  • 19. e.g.. VCVs do not provide an independent pressure control. • Control on ventilators may provide an approximate values. To confirm the accuracy of controls, an effective monitoring capability is needed. e.g.. PCVs do not provide an independent volume control. • There are several functions for the ventilators have no controls, yet these must be monitored CONTROLS ..
  • 20.
  • 21. INITIAL VENTILATOR SETTINGS Mode of ventilation • The initial mode of ventilation should be the assist control (A/C) mode. • In the A/C mode, the work of breathing is reduced to only that amount of inspiration needed to trigger the machine's inspiratory cycle. • A second possible advantage is that cycling the ventilator into the inspiratory phase maintains normal ventilatory activity and, therefore, prevents atrophy of the respiratory muscles.
  • 22. The potential disadvantages of the A/C mode include respiratory alkalosis. A second possible disadvantage is that every breath is a full, preset, positive-pressure breath that influences venous return to the right heart and, possibly, to global cardiac output. Nevertheless, the A/C mode is the best initial mode of ventilation and may be switched to another option if hypotension occurs or hypocarbia is evident on the first arterial blood gas analysis.
  • 23. Fractional Inspired Oxygen (FIO2) The fractional inspired oxygen is the amount of oxygen delivered to the patient. It can range from 21% (room air) to 100%. It’s recommended that the FIO2 be set at 1.0 (100%) upon the initiation of mechanical ventilation to allow the patient to get used to the ventilator without experiencing hypoxia.
  • 24. A short period on FIO2 of 100% is not dangerous to the patient on mechanical ventilation and offers the clinician several advantages: • protects the patient against hypoxemia if unrecognized problems develop from the intubation procedure. • PaO2 measured on an FIO2 of 100%, the clinician can easily calculate the next desired FIO2 and quickly estimate the shunt fraction.
  • 25. When the clinician wishes to change the PaO2, 2 variables might be adjusted FIO2 and the amount of PEEP. When using supplemental FIO2, improvement in oxygenation cannot reliably be achieved with changes in rate or tidal volume. With the target PaO2 identified, the FIO2 can be adjusted using the formula: New FIO2=(old FIO2 X desired PaO2)/measured PaO2.
  • 26. The degree of shunt on 100% FIO2 can be estimated by a rough rule of thumb. The measured PaO2 is subtracted from 700 mm Hg. For each 100 mm Hg difference, a 5% shunt exists. A shunt of approximately 25% requires the use of PEEP.
  • 27. Positive end-expiratory pressure [PEEP] PEEP therapy can be effective when used in patients with a diffuse lung disease that results in an acute decrease in functional residual capacity (FRC) when used to treat patients with a diffuse lung disease, PEEP should improve compliance, decrease dead space, and decrease the intrapulmonary shunt effect. The most significant benefit of PEEP is that the patient can maintain an adequate PaO2 at a lower, safer concentration of oxygen (<60%), thereby reducing the risk of oxygen toxicity.
  • 28. No recommendations exist for adding external PEEP during initial ventilator setup to satisfy misguided attempts to supply prophylactic PEEP or physiologic PEEP. Most clinicians use the lowest amount of positive pressure that provides an adequate PaO2 with a safe FIO2. PEEP
  • 29. PEEP principally is used to lower the risks of oxygen toxicity and is employed when a safe PaO2 cannot be achieved at 60% FIO2. Initiated at 5 cm H2O, PEEP usually is increased in 3 cm H2O increments, while evaluating the effect on oxygenation every 15-20 minutes. A PEEP level of 10 cm H2O rarely causes hemodynamic problems in the absence of intravascular volume depletion.
  • 30. changes in PEEP may not be reflected by changes in arterial blood gases for 20-30 minutes so changes in the PEEP setting should usually not be made faster than this PEEP over 20 cm is rarely beneficial and usually results in additional pressure-induced lung injury
  • 31. BEST PEEP ? The exact amount of PEEP applied is very controversial ?????????????????????????????????
  • 32. Pressure (cm H2O) 0 50 Volume (ml) 1500 0 Alveoli elastic limit UIP LIP Alveoli open Alveolar collapse PEEP Pressure limit Pressure-volume curve
  • 33. The titration of PEEP above the lower inflection point  increases lung compliance “best PEEP”. The titration of PIP below the upper inflection point  prevention of over destintion Pressure-Volume loope
  • 34. 400 600 200 0 5 10 20 30 40 50 Press Vt 300 500 700 100 Lower inflection point Upper inflection point PIP PEEP Pressure-Volume loope
  • 35.
  • 36. The respiratory rate is the number of breaths that the ventilator delivers to the patient each minute. The rate chosen depends on the : 1. Tidal volume. 2. Type of pulmonary pathology. 3. patient’s target PaCO2. Respiratory Rate (RR)
  • 37. Patients with restrictive lung disease usually tolerate a range of 12-20 breaths/minute. Patients with normal pulmonary mechanics can tolerate a rate of 8-12 breaths/minute. For patients with obstructive lung disease, the rate should be set at 6-8 breaths/minute to avoid the development of auto-PEEP and hyperventilation, or “blowing off CO2.”. Patients with obstructive lung disease often adapt to a higher PaCO2, so lowering it back to the “normal” range of 35-45 mm Hg may not be beneficial.
  • 38. The respiratory rate parameters are set above and below this number and the alarm will then sound if the patient’s actual rate is outside of the desired range.
  • 39. The usual setting is 5-12 cc/kg. based on compliance, resistance, and type of pathology. Tidal Volume (VT) • Patients with normal lungs can tolerate a tidal volume of 10-12 cc/kg, • Patients with restrictive lung disease may need a tidal volume of 5-8 cc/kg.
  • 40. Tidal Volume (VT) Males Ideal body weigh IBW =50+2.3 (Height (inches) – 60 ) Females Ideal body weigh IBW =50+2.3 (Height (inches) – 60 )
  • 41. The tidal volume alarm level are set above and below the desired number, and the alarm will sound if the patient’s actual tidal volume is outside of the desired range. Tidal Volume (VT)
  • 42. Double-checking the selected tidal volume After a tidal volume is selected, the peak airway and Plateau pressures pressure should be determined peak airway pressure <45 cm H2O. plateau pressure is <30-35 cm H2O.
  • 43. Tow Pressure level :  Small values, which just to compensate for resistance of the ETT.  Large values, that make the spontaneous breath equivalent to triggered mandatory breath (PCV). Pressure suppot PS
  • 44. Pressure support PS only eliminates the work precisely at a given flow. Above and below that flow, PS under compensates for resistance or over compensates.
  • 45. (Weaning) ATC ATC compensation can overcome resistance regardless of patient flow demand.
  • 46. The I:E ratio is usually set at 1:2 or 1:1.5 to approximate the normal physiology of inspiration and expiration. Occasionally, a longer inspiratory than expiratory time is desired to allow more time to oxygenate the patient’s lungs. This is called inverse ratio ventilation. Inspiratory:Expiratory (I:E) Ratio
  • 47. The pressure limit regulates the amount of pressure the VCV can generate to deliver the preset tidal volume. Because high pressures can cause lung injury, it’s recommended that the plateau pressure not exceed 35 cm H20. If this limit is reached, the ventilator stops delivering the breath and alarms. This may be an indication that the patient’s airway is obstructed with mucus. It can also be caused by the patient coughing, biting on the ETT, breathing against the ventilator, or by a kink in the ventilator tubing. Pressure Limit
  • 48. The flow rate is the speed with which the tidal volume is delivered. Flow Rate The usual setting is 40-100 liters per minute.
  • 49. The sensitivity determines the amount of effort required by the patient to initiate inspiration. Sensitivity/Trigger It can be: Pressure triggering 1 – 2 cm H2O flow triggering. 1- 5 L/min Flow triggering is a better setting for patients who can breathe spontaneously because it reduces the work of breathing. time pressure trigger- threshold
  • 50. • Very fast rising Depending on rise time, the beginning of the flow will be • or slower Rise time For adults the range is 0-0.4 seconds
  • 51. The ventilator can be programmed to deliver an occasional sigh with a larger tidal volume. The use of frequent sighs was popular during the 1970’s because it was thought that it prevented collapse of the alveoli (atelectasis), which can result from the patient constantly inspiring the same volume of gas. However, recently there has been concern that the increased pressure produced in the alveoli may heighten the risk of the alveoli rupturing and causing pneumothorax. Sigh
  • 52. Summary of initial ventilator set up • A/C Mode • Tidal volume depending on lung status Normal lungs - 12 cc/kg ideal body weight COPD - 10 cc/kg ideal body weight ARDS - 6-8 cc/kg ideal body weight • Rate of 10-12 breaths per minute • FIO2 of 100% • Sighs rarely needed • PEEP only as indicated after first arterial blood gas Shunt greater than 25% Inability to oxygenate with safe FIO2
  • 53. What is ASV? Conventional Tp Rate Psup Pinsp Ti Te @ FiO2 PEEP Vt CMV SIMV PCV PSV Goal 3: Simple to use Modes Controls
  • 54. What is ASV? Conventional Tp Rate Psup Pinsp Ti Te @ FiO2 PEEP Vt CMV SIMV PCV PSV ASV Goal 3: Simple to use Controls
  • 55. What is ASV? Conventional Tp Rate Psup Pinsp Ti Te @ FiO2 PEEP Vt CMV SIMV PCV PSV ASV Goal 3: Simple to use
  • 56. What is ASV? Conventional Tp Rate Psup Pinsp Ti Te @ FiO2 PEEP Vt CMV SIMV PCV PSV ASV Goal 3: Simple to use
  • 57. Tp Rate Pinsp Ti Te PF FiO2 PEEP Vt CMV SIMV PCV PSV Alveolar Ventilation Oxy. Conventional FiO2 PEEP MV Automatic Mechanical ventilation
  • 58. 2. MONITORS Ventilators are designed to monitor many components of the patient’s respiratory status.
  • 60. peak , mean , plateau and PEEP. Pressure:
  • 61. Vt, Vm, sigh volume. Volume: Frequency (mandatory & spontaneous). MONITORS
  • 62. I/E ratio It, Et. FIO2. MONITORS
  • 63. Lung mechanics Monitoring Flow-time curve Pressure-time curve Volume-pressure loop Flow-volume loop Nigative inspiratory force (NIF) AutoPEEP Airway Resistance Lung compliance
  • 65. 3. ALARMS To ensure the safety and effectiveness of MV electric or pneumatic alarms are installed on ventilator to signal both visually and audibly the presence of undesirable conditions.
  • 66.
  • 67. Problem solving • pCO2 too high !!!!!!!!!!!!!! • pO2 too low !!!!!!!!!!!!!!!!!!!!!!!!!! • PIP too high !!!!!!!!!!!!!!!!!!!!!!! ????????????????????????????????
  • 68.  Safety considerations in mechanical ventilators  Back up ventilation  Low resistance non compliant breathing circuit  Limits on closed loop adjustments  Reliable alarm system  Battery for electrical failure
  • 69. Golden rules in mechanical ventilation
  • 70. 5- Adapt the ventilator to the patient instead of adapting the patient to the ventilator. 2- A mode doesn’t make the lung healthy it wins only time. 3- Keep healthy parts of the lung healthy. 4- Maintain and support spontaneous breathing in ventilated patient. 1- Start the process of weaning the moment you start mechanical ventilation.
  • 71. 8- Avoid airway pressure more than 35 cmH2O 6- Employ tidal volume of 6 – 8 ml/kg Wt. 9- Don’t be too shy with PEEP. 10- Set always the trigger at the highest sensitivity (avoid autocycling). 7- Give 7 – 8 cmH2O pressure support for compensation of the tube resistance. Vt PS PAP PEEP Triggering
  • 72. pCO2 Too High • Patient’s minute ventilation is too low. • Increase rate or TV or both. • If using PC ventilation, increase PIP. • If PIP too high, increase the rate instead. • If air-trapping is occurring, decrease the rate and the I- time and increase the TV to allow complete exhalation.
  • 73. pO2 Too Low • Increase either the FiO2 or the mean airway pressure (MAP). • Try to avoid FiO2 >70%. • Increasing the PEEP is the most efficient way of increasing the MAP in the PICU. • Can also increase the I-time to increase the MAP (PC). • Can increase the PIP in Pressure Control to increase the MAP, but this generally doesn’t add much at rates <30bpm.
  • 74. PIP Too High • Decrease the PIP (PC) or the TV (VC). • Increase the I-time (VC). • Change to another mode of ventilation. Generally, pressure control achieves the same TV at a lower PIP than volume control. • If the high PIP is due to high airway resistance, generally the lung is protected from barotrauma unless air-trapping occurs.
  • 75. Acute Deterioration • DIFFERENTIAL DIAGNOSES o Pneumothorax o Right main stem intubation o Pneumonia o Pulmonary edema o Airway occlusion o Ventilator malfunction o Mucus plugging o Air leak o hypoperfusion